      "cfres_id": "70470",
      "product_res_number": "Z-0403-2009",
      "event_date_initiated": "2008-02-21",
      "event_date_posted": "2009-01-15",
      "recall_status": "Terminated",
      "event_date_terminated": "2009-01-15",
      "res_event_number": "47848",
      "product_code": "GET",
      "product_description": "ConMed Linvatec  Hall Surgical Blades, Sterile STER RPLCMNT BLADE/STRYKER RECIP,  Catalog # 00505213500\n\n\nConMed Linvatec Hall (R) Surgical Blades are intended for use in small bone or large bone orthopedic surgeries.  These blades are designed to cut bone in an oscillating plane.",
      "code_info": "BBD22605, BBD23365, BBD30513, BBD31988, BBD34449, and BBD35377.",
      "firm_fei_number": "1017294",
      "recalling_firm": "Linvatec Corp.",
      "address_1": "11311 Concept Boulevard",
      "city": "Largo",
      "state": "FL",
      "postal_code": "33773",
      "additional_info_contact": "727-392-6464",
      "reason_for_recall": "Sterility compromised.  There is a possibility that certain catalog number and lots of ConMed Linvatec Hall Surgical Blades, may have a hole(s) in the corner(s) of the blister package(tray), thereby potentially compromising the sterility of the devices.",
      "root_cause_description": "Process control",
      "action": "An Urgent Medical Device Recall letter was issued.  Customers must check their facility's inventory immediately for the catalog and lot numbers in the attached list.  They also need to complete and sign the enclosed Reply Form and mail or fax the Reply Form to ConMed Linvatec using the enclosed self addressed postage paid envelope or fax 727-319-5701.  Customers have an option to either re-sterilize the products using the re-sterilization instructions in the letter or they can return the devices to ConMed Linvatec for a credit using the enclosed preprinted FedEx Air Bill with ConMed Linvatec's Bill Number and the return goods number.  Recall is being conducted to the distributor and end user level.",
      "product_quantity": "447",
      "distribution_pattern": "Worldwide Distribution",
      "openfda": {
        "k_number": [
          "K953432",
          "K800382",
          "K952463",
          "K832188",
          "K885066",
          "K790195",
          "K832186",
          "K840370",
          "K872951",
          "K914753",
          "K952300",
          "K904578",
          "K863447",
          "K910071",
          "K832187",
          "K831675",
          "K953433",
          "K843498",
          "K833187",
          "K954076",
          "K955174",
          "K942987",
          "K960473"
        ],
        "registration_number": [
          "",
          "3015453963",
          "3005180920",
          "3017210488",
          "3010656982",
          "3018094310",
          "1058584",
          "3003587896",
          "3013893019",
          "3003120897",
          "3003079267",
          "3032391",
          "1017294",
          "3004962786",
          "9610612",
          "1828288",
          "1423662",
          "3015276580",
          "3022320321",
          "3011495120",
          "1000606483",
          "3008110533",
          "3008787426",
          "1836161",
          "3014461810",
          "3039169546",
          "2020601",
          "1649518",
          "1828464",
          "3010141593",
          "1625507",
          "3004940685",
          "2916714",
          "8043750",
          "3005061536",
          "3030956508",
          "3007305485",
          "1018470",
          "3025936823",
          "1320894",
          "2647580",
          "1030489",
          "3009599228",
          "1219930",
          "3004264602",
          "8030607"
        ],
        "fei_number": [
          "",
          "3015453963",
          "3002948288",
          "1000219637",
          "3005180920",
          "3001236277",
          "3002750084",
          "3017210488",
          "3010656982",
          "3018094310",
          "3003587896",
          "3013893019",
          "3003120897",
          "3003079267",
          "3006509795",
          "1017294",
          "3004962786",
          "1828288",
          "1000477302",
          "3002299608",
          "3015276580",
          "3022320321",
          "3011495120",
          "1000606483",
          "3008110533",
          "3008787426",
          "3014461810",
          "3039169546",
          "2020601",
          "1828464",
          "1625507",
          "3010141593",
          "3002806315",
          "3004940685",
          "2916714",
          "3000215273",
          "3005061536",
          "3030956508",
          "3007305485",
          "1018470",
          "3025936823",
          "2647580",
          "3009599228",
          "3001451571",
          "1219930",
          "3004264602"
        ],
        "device_name": "Motor, Surgical Instrument, Pneumatic Powered",
        "medical_specialty_description": "General, Plastic Surgery",
        "regulation_number": "878.4820",
        "device_class": "1"
      }
    }
  ]
}