  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-08-20",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "70453",
      "product_res_number": "Z-0386-2009",
      "event_date_initiated": "2008-02-21",
      "event_date_posted": "2009-01-15",
      "recall_status": "Terminated",
      "event_date_terminated": "2009-01-15",
      "res_event_number": "47848",
      "product_code": "GET",
      "product_description": "ConMed Linvatec  Hall Surgical Blades, Sterile ST OSC BLDE,  Catalog # 00507155200, 00507155300, 00507114600, 00507155100, 00507155600, 00507155000, 00507112500, 00507112600, and 00507112700.\n\n\nConMed Linvatec Hall (R) Surgical Blades are intended for use in small bone or large bone orthopedic surgeries.  These blades are designed to cut bone in an oscillating plane.",
      "code_info": "BBD27997, BBD33248, BBD34458, BBD26279, BBD23462, BBD26021, BBD27537, BBD22412, BBD23258, BBD26861, BBD28114, BBD29762, BBD30515, BBD31766, BBD33228, BD35380, BBD37129, BBD38466, BBD26020, BBD23453, BBD25525, BBD27531, BBD29420, BBD33035, BBD33036, BBD35906, BBD22162 through BBD22164, BBD27529, BBD29958, BBD29960, BBD 29963, BBD29964, BBD32800, BBD32804, BBD33037, BBD33291, BBD39455, BBD43139, BBD21965, BBD23044, BBD23823, BBD26016, BBD26018, BBD27533, BBD30781 through BBD30785, BBD31753, BBD35194, BBD34918, BBD35409, BBD40489, BBD23465, BBD26015, BBD26022, BBD26273, BBD27211, BBD32189, BBD33038, BBD34562, BBD35908, BBD36136, and BBD36457.",
      "firm_fei_number": "1017294",
      "recalling_firm": "Linvatec Corp.",
      "address_1": "11311 Concept Boulevard",
      "city": "Largo",
      "state": "FL",
      "postal_code": "33773",
      "additional_info_contact": "727-392-6464",
      "reason_for_recall": "Sterility compromised.  There is a possibility that certain catalog number and lots of ConMed Linvatec Hall Surgical Blades, may have a hole(s) in the corner(s) of the blister package(tray), thereby potentially compromising the sterility of the devices.",
      "root_cause_description": "Process control",
      "action": "An Urgent Medical Device Recall letter was issued.  Customers must check their facility's inventory immediately for the catalog and lot numbers in the attached list.  They also need to complete and sign the enclosed Reply Form and mail or fax the Reply Form to ConMed Linvatec using the enclosed self addressed postage paid envelope or fax 727-319-5701.  Customers have an option to either re-sterilize the products using the re-sterilization instructions in the letter or they can return the devices to ConMed Linvatec for a credit using the enclosed preprinted FedEx Air Bill with ConMed Linvatec's Bill Number and the return goods number.  Recall is being conducted to the distributor and end user level.",
      "product_quantity": "17635",
      "distribution_pattern": "Worldwide Distribution",
      "openfda": {
        "k_number": [
          "K953432",
          "K800382",
          "K952463",
          "K832188",
          "K885066",
          "K790195",
          "K832186",
          "K840370",
          "K872951",
          "K914753",
          "K952300",
          "K904578",
          "K863447",
          "K910071",
          "K832187",
          "K831675",
          "K953433",
          "K843498",
          "K833187",
          "K954076",
          "K955174",
          "K942987",
          "K960473"
        ],
        "registration_number": [
          "",
          "3015453963",
          "3005180920",
          "3017210488",
          "3010656982",
          "3018094310",
          "1058584",
          "3003587896",
          "3013893019",
          "3003120897",
          "3003079267",
          "3032391",
          "1017294",
          "3004962786",
          "9610612",
          "1828288",
          "1423662",
          "3015276580",
          "3022320321",
          "3011495120",
          "1000606483",
          "3008110533",
          "3008787426",
          "1836161",
          "3014461810",
          "3039169546",
          "2020601",
          "1649518",
          "1828464",
          "3010141593",
          "1625507",
          "3004940685",
          "2916714",
          "8043750",
          "3005061536",
          "3030956508",
          "3007305485",
          "1018470",
          "3025936823",
          "1320894",
          "2647580",
          "1030489",
          "3009599228",
          "1219930",
          "3004264602",
          "8030607"
        ],
        "fei_number": [
          "",
          "3015453963",
          "3002948288",
          "1000219637",
          "3005180920",
          "3001236277",
          "3002750084",
          "3017210488",
          "3010656982",
          "3018094310",
          "3003587896",
          "3013893019",
          "3003120897",
          "3003079267",
          "3006509795",
          "1017294",
          "3004962786",
          "1828288",
          "1000477302",
          "3002299608",
          "3015276580",
          "3022320321",
          "3011495120",
          "1000606483",
          "3008110533",
          "3008787426",
          "3014461810",
          "3039169546",
          "2020601",
          "1828464",
          "1625507",
          "3010141593",
          "3002806315",
          "3004940685",
          "2916714",
          "3000215273",
          "3005061536",
          "3030956508",
          "3007305485",
          "1018470",
          "3025936823",
          "2647580",
          "3009599228",
          "3001451571",
          "1219930",
          "3004264602"
        ],
        "device_name": "Motor, Surgical Instrument, Pneumatic Powered",
        "medical_specialty_description": "General, Plastic Surgery",
        "regulation_number": "878.4820",
        "device_class": "1"
      }
    }
  ]
}