{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-08-13",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "68356",
      "product_res_number": "Z-0356-2009",
      "event_date_initiated": "2008-07-18",
      "event_date_posted": "2008-12-11",
      "recall_status": "Terminated",
      "event_date_terminated": "2009-07-07",
      "res_event_number": "44839",
      "product_code": "DTQ",
      "k_numbers": [
        "K022947"
      ],
      "product_description": "Terumo Advanced Perfusion System 1 (4 inch diameter Roller Pump, Gray; Model 816570).\n\nExtracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures.",
      "code_info": "Serial numbers 1001 through 1008 and 4009 through 4269.",
      "firm_fei_number": "1828100",
      "recalling_firm": "Terumo Cardiovascular Systems Corporation",
      "address_1": "6200 Jackson Rd",
      "city": "Ann Arbor",
      "state": "MI",
      "postal_code": "48103",
      "additional_info_contact": "800-521-2818",
      "reason_for_recall": "Overspeed and underspeed errors, erratic pump behavior, jerky operation at low RPMs, pump instability, pump slowdowns and pump stops due to overspeed max events may occur due to a grease leak onto the motor speed encoder disk.",
      "root_cause_description": "Component design/selection",
      "action": "Consignees were notified of the problem via recall letter dated  7/18/08.  The firm began conducting field upgrades on the roller pumps in 07/08.  For questions or concerns, contact Terumo CVS Customer Service at 1-800-521-2818.",
      "product_quantity": "269",
      "distribution_pattern": "Worldwide Distribution including USA and countries of  Australia, Belgium. Canada, Columbia, Dominican Republic, Egypt, Germany, Honduras, Hong Kong, India, Indonesia, Israel, Japan, Jordan, Kuwait, Malaysia, Mexico, Pakistan, Philippines, Russia, Saudi Arabia, Singapore, South Africa, South Korea, Taiwan, Thailand, Turkey, United Arab Emirates and Viet Nam.",
      "openfda": {
        "k_number": [
          "K870976",
          "K022947",
          "K221373",
          "K172220",
          "K942582",
          "K091007",
          "K133598",
          "K002116",
          "K915183",
          "K232291",
          "K801097",
          "K842138",
          "K093882",
          "K102726",
          "K822068",
          "K982014",
          "K863541",
          "K851615",
          "K080832",
          "K772001",
          "K163531",
          "K221895",
          "K770002",
          "K813034",
          "K210130",
          "K950990",
          "K871131",
          "K090006",
          "K103762",
          "K173834",
          "K002118",
          "K202125",
          "K070213",
          "K060053",
          "K062396",
          "K071318",
          "K021694",
          "K153376",
          "K020571",
          "K964017",
          "K030141",
          "K862973",
          "K131041",
          "K851614",
          "K011088",
          "K863960",
          "K861388",
          "K943803",
          "K052555",
          "K092463",
          "K151349",
          "K190282"
        ],
        "registration_number": [
          "9611109",
          "3019807891",
          "3002807090",
          "1828100",
          "3011554160",
          "3007427881",
          "8010762",
          "3042990265",
          "3009507273",
          "3010163695",
          "3006073153",
          "3001589655",
          "3010703925",
          "1218444",
          "3005941719",
          "2916596",
          "9680841",
          "2184009"
        ],
        "fei_number": [
          "3010703925",
          "3019807891",
          "1828100",
          "3011554160",
          "3007427881",
          "3009649876",
          "3042990265",
          "3002807111",
          "3009507273",
          "3010163695",
          "3005426550",
          "1000116158",
          "3006073153",
          "3001589655",
          "3006003617",
          "1218444",
          "2916596",
          "3002808267"
        ],
        "device_name": "Console, Heart-Lung Machine, Cardiopulmonary Bypass",
        "medical_specialty_description": "Cardiovascular",
        "regulation_number": "870.4220",
        "device_class": "2"
      }
    }
  ]
}