  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-08-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "66398",
      "product_res_number": "Z-0526-2008",
      "event_date_initiated": "2007-10-02",
      "event_date_posted": "2008-04-10",
      "recall_status": "Terminated",
      "event_date_terminated": "2010-12-13",
      "res_event_number": "45880",
      "product_code": "GKI",
      "k_numbers": [
        "K050191"
      ],
      "product_description": "BD FACS Sample Prep Assistant II, Part Number: 337170;\nProduct is manufactured and distributed by BD Biosciences, San Jose, CA  95131",
      "code_info": "Serial Numbers: T0246 to T0290 (except for T0274,T0275,T0277 and T0279)",
      "firm_fei_number": "2916837",
      "recalling_firm": "BD Biosciences",
      "address_1": "2350 Qume Dr",
      "city": "San Jose",
      "state": "CA",
      "postal_code": "95131-1807",
      "additional_info_contact": "Kim Ky \t\n408-954-2610",
      "reason_for_recall": "Mixing Performance:  SPA II devices manufactured with the affected product motor control board may exhibit suboptimal mixing performance.",
      "root_cause_description": "Software change control",
      "action": "On 10/1/07, all consignees were notified via Important Product Information letter and also an on-site visit by the firm's sales representative.  The letter informed users of the affected product and provided instructions on the recall.  For additional concerns or questions contact BD Customer Support Center at 1-877-232-8995.",
      "distribution_pattern": "Worldwide Distribution --- USA including states of:  NJ, IL, TX, PA, NY, MN, and AZ, and countries of Australia, Belgium, Hong Kong, Trinidad and Tobago.\n\n\n\nAustralia, Belgium, Hong Kong, Trinidad and Tobago.",
      "openfda": {}
    }
  ]
}