  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-08-13",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "65617",
      "product_res_number": "Z-0294-2008",
      "event_date_initiated": "2007-10-17",
      "event_date_posted": "2007-11-27",
      "recall_status": "Terminated",
      "event_date_terminated": "2007-11-27",
      "res_event_number": "45492",
      "product_code": "JZL",
      "k_numbers": [
        "K050996"
      ],
      "product_description": "Medtronic 0-arm Imaging System, Product Catalog Number: B1-700-00027, Medtronic Navigation, Inc., Littleton, MA 01460\n(software version 3.0)",
      "code_info": "System Serial Numbers: 103, 104, 109, 117, 119 (Germany) 125, and 144 (Canada)",
      "firm_fei_number": "1000517638",
      "recalling_firm": "Medtronic Navigation, Inc",
      "address_1": "826 Coal Creek Circle",
      "city": "Louisville",
      "state": "CO",
      "postal_code": "80027-9710",
      "additional_info_contact": "720-890-3200 Ext. 3303",
      "reason_for_recall": "Reversed Instrument Position : a defect in the software version 3.0 causes an error in the position of surgical instruments and relative motion are reversed 180' on the StealthStation image.  (If the StealthStation is not used, or if the patient orientation is prone or supine, the problem does not occur)",
      "root_cause_description": "Software change control",
      "action": "Medtronic Navigation notified customers by letter on October 17, 2007. Medtronic Field service will perform upgrade (software version 3.0.1) at user site.",
      "product_quantity": "7 units",
      "distribution_pattern": "Worldwide: USA including states of CO, IN, NE, and MN and countries of  Canada and Germany",
      "openfda": {}
    }
  ]
}