      "cfres_id": "64565",
      "product_res_number": "Z-0682-2007",
      "event_date_initiated": "2007-03-01",
      "event_date_posted": "2007-09-28",
      "recall_status": "Terminated",
      "event_date_terminated": "2012-05-01",
      "res_event_number": "37498",
      "product_code": "JAA",
      "k_numbers": [
        "K021049",
        "K022069",
        "K024012"
      ],
      "product_description": "OEC 9800 Plus (Fluorotrak), Image-intensified fluoroscopic x-ray system, GE Healthcare, Salt Lake City, UT.",
      "code_info": "Serial number 8S-2339-N.",
      "firm_fei_number": "1720753",
      "recalling_firm": "GE OEC Medical Systems, Inc",
      "address_1": "384 Wright Brothers Dr",
      "city": "Salt Lake City",
      "state": "UT",
      "postal_code": "84116-2862",
      "additional_info_contact": "Karen Tolson\n847-501-1884",
      "reason_for_recall": "During routine service, primary and/or secondary collimators may not have been properly reinstalled.",
      "root_cause_description": "Other",
      "action": "Customers were notified by letter on 03/01/2007. They were told to continue using the machines and that an inspection would be scheduled immediately to verify that collimators had been correctly installed.",
      "product_quantity": "1",
      "distribution_pattern": "Nationwide and Germany.",
      "openfda": {
        "k_number": [
          "K864837",
          "K951261",
          "K882175",
          "K871119",
          "K840390",
          "K820184",
          "K894913",
          "K882176",
          "K810716",
          "K875157",
          "K210469",
          "K950971",
          "K771298",
          "K853984",
          "K896331",
          "K924879",
          "K132904",
          "K982319",
          "K903821",
          "K920550",
          "K220871",
          "K953113",
          "K080859",
          "K833230",
          "K050866",
          "K022911",
          "K941611",
          "K973067",
          "K827489",
          "K992603",
          "K893644",
          "K926434",
          "K872061",
          "K883985",
          "K973744",
          "K854953",
          "K922977",
          "K943805",
          "K840364",
          "K864584",
          "K010392",
          "K922013",
          "K915405",
          "K883372",
          "K955687",
          "K051025",
          "K020055",
          "K180589",
          "K914446",
          "K840191",
          "K110785",
          "K874901",
          "K160301",
          "K142341",
          "K855213",
          "K883580",
          "K931268",
          "K950631",
          "K141541",
          "K203087",
          "K932445",
          "K990055",
          "K934481",
          "K833237",
          "K800639",
          "K895730",
          "K873985",
          "K101491",
          "K972442",
          "K953950",
          "K812963",
          "K923585",
          "K202235",
          "K801166",
          "K920721",
          "K902981",
          "K943532",
          "K792225",
          "K770337",
          "K820178",
          "K912896",
          "K010187",
          "K993896",
          "K862120",
          "K993038",
          "K163210",
          "K943447",
          "K901153",
          "K912380",
          "K922274",
          "K864749",
          "K123097",
          "K812775",
          "K061028",
          "K851031",
          "K974058",
          "K944973",
          "K813158",
          "K905694",
          "K040870",
          "K915592",
          "K952456",
          "K191877",
          "K970608",
          "K811295",
          "K884578",
          "K880527",
          "K943272",
          "K081621",
          "K253103",
          "K953974",
          "K874905",
          "K895546",
          "K923107",
          "K843972",
          "K924164",
          "K982028",
          "K871127",
          "K840225",
          "K895394",
          "K081091",
          "K111292",
          "K842217",
          "K913886",
          "K132106",
          "K982345",
          "K970947",
          "K843235",
          "K842191",
          "K952493",
          "K922012",
          "K910384",
          "K010772",
          "K242948",
          "K200396",
          "K945460",
          "K893475",
          "K103038",
          "K131766",
          "K973073",
          "K840621",
          "K810189",
          "K961193",
          "K021021",
          "K121293",
          "K052610",
          "K954945",
          "K920746",
          "K933472",
          "K936136",
          "K871911",
          "K933299",
          "K022565",
          "K191713",
          "K914525",
          "K232526",
          "K791830",
          "K961207",
          "K812561",
          "K926560",
          "K930885",
          "K951358",
          "K942674",
          "K781711",
          "K834287",
          "K900893",
          "K923453",
          "K946081",
          "K902005",
          "K191504",
          "K051602",
          "K961782",
          "K902927",
          "K870429",
          "K936236",
          "K920720",
          "K770333",
          "K954590",
          "K904986",
          "K852346",
          "K925251",
          "K871743",
          "K203010",
          "K004037",
          "K802150",
          "K203428",
          "K923454",
          "K191310",
          "K082830",
          "K902983",
          "K081491",
          "K802953",
          "K933473",
          "K081582",
          "K911454",
          "K133553",
          "K810364",
          "K130318",
          "K942019",
          "K926064",
          "K912341",
          "K840134",
          "K833530",
          "K131075",
          "K212837",
          "K132289",
          "K024012",
          "K042698",
          "K801948",
          "K801320",
          "K902583",
          "K913012",
          "K912470",
          "K140380",
          "K112054",
          "K904012",
          "K761285",
          "K850959",
          "K882311",
          "K942841",
          "K942264",
          "K812217",
          "K883161",
          "K901405",
          "K901671",
          "K884598",
          "K950781",
          "K843920",
          "K923137",
          "K200965",
          "K961690",
          "K834286",
          "K893476",
          "K941917",
          "K811094",
          "K771485",
          "K962606",
          "K861785",
          "K962140",
          "K173395",
          "K833641",
          "K961140",
          "K981227",
          "K232910",
          "K884028",
          "K012424",
          "K940295",
          "K182834",
          "K964322",
          "K212890",
          "K122768",
          "K950683",
          "K893131",
          "K902982",
          "K890745",
          "K964990",
          "K000987",
          "K843917",
          "K233945",
          "K905111",
          "K050692",
          "K151485",
          "K890878",
          "K902980",
          "K901323",
          "K212145",
          "K870869",
          "K190373",
          "K813217",
          "K884317",
          "K011624",
          "K242488",
          "K863853",
          "K902535",
          "K931941",
          "K844375",
          "K934626",
          "K963617",
          "K771096",
          "K051136",
          "K070390",
          "K926122",
          "K922240",
          "K933791",
          "K133434",
          "K914414",
          "K833531",
          "K161019",
          "K150306",
          "K943156",
          "K022899",
          "K963888",
          "K180651",
          "K905278",
          "K933100",
          "K922011",
          "K233380",
          "K251650",
          "K981414",
          "K960266"
        ],
        "registration_number": [
          "",
          "1000221974",
          "3012624817",
          "3011950073",
          "3005172103",
          "3026391023",
          "1018470",
          "3012761544",
          "3010162973",
          "3005496266",
          "1125622",
          "3043048939",
          "3012127758",
          "1048735",
          "1061158",
          "3004911384",
          "2126677",
          "3031503890",
          "3003775006",
          "3008632256",
          "3004077520",
          "9616491",
          "3011015597",
          "3008284470",
          "3027373189",
          "3023003446",
          "3014391847",
          "1052701",
          "3010703925",
          "1054713",
          "3005254598",
          "3014343100",
          "3036666444",
          "3033525987",
          "3035950510",
          "1225677",
          "3021559833",
          "3003768277",
          "3010202439",
          "3001722928",
          "3008044777",
          "3005843418",
          "9613445",
          "9681146",
          "3016619943",
          "3003202425",
          "3033527282",
          "3006972752",
          "1720753",
          "1038913",
          "2936485",
          "3010685285",
          "3003314115",
          "1000188474",
          "3009144915",
          "3011416394",
          "3021059265",
          "3006544299",
          "1717855",
          "3010462278",
          "3015223550",
          "3042175844",
          "1000181430",
          "3005877899",
          "3018094310",
          "3011491794",
          "3033527770",
          "3015437189",
          "3005810333",
          "3000126629",
          "3021060474",
          "2027062",
          "9611343",
          "3017375271",
          "3004785967",
          "3014273597",
          "2243057",
          "3030412764",
          "3009715136",
          "3030089787",
          "3014925393",
          "3001556265",
          "3008496839",
          "1043214",
          "3016438694",
          "1047843",
          "3027339477",
          "3025442726",
          "3009196578",
          "3009077524",
          "3011713200",
          "3003927316",
          "3003477135",
          "3009963139",
          "3004754211",
          "3033525986",
          "9613347",
          "8021091",
          "3009048384",
          "3004977335",
          "9610903",
          "3007143236",
          "3003935342",
          "3038614464",
          "1055236",
          "3004026651",
          "3009973505",
          "1056553",
          "3003768251",
          "3036255956",
          "2133772",
          "3003600763",
          "3013099020",
          "9710018",
          "9611388",
          "2020550",
          "3011158427",
          "9614698",
          "2031966",
          "1037885",
          "1723170",
          "3019878714",
          "3010398867",
          "1054241",
          "2320762",
          "3015276088",
          "3012939903",
          "3014279513",
          "3014634069",
          "3005706667",
          "8020997",
          "3002466018",
          "8030112",
          "8030233",
          "9680001",
          "2027467",
          "3015531959",
          "1037955"
        ],
        "fei_number": [
          "",
          "1000221974",
          "3012624817",
          "3011950073",
          "3005172103",
          "3026391023",
          "1018470",
          "1000284313",
          "3012761544",
          "3005496266",
          "3043048939",
          "3012127758",
          "1048735",
          "3004911384",
          "2126677",
          "3002807041",
          "3031503890",
          "3003775006",
          "3008632256",
          "3004077520",
          "3011015597",
          "3008284470",
          "3027373189",
          "3023003446",
          "3014391847",
          "3007475226",
          "3010703925",
          "3005254598",
          "3002807092",
          "3014343100",
          "3002808139",
          "1000517638",
          "3036666444",
          "3033525987",
          "3035950510",
          "1000582291",
          "3021559833",
          "3001722928",
          "3008044777",
          "3005843418",
          "3000255482",
          "3016619943",
          "3003202425",
          "3002808445",
          "3033527282",
          "3006972752",
          "1720753",
          "1038913",
          "2936485",
          "3010685285",
          "1000112877",
          "3003314115",
          "1000188474",
          "3009144915",
          "3002952981",
          "3000209794",
          "3011416394",
          "3021059265",
          "3006544299",
          "1717855",
          "3010462278",
          "3015223550",
          "1000132370",
          "3042175844",
          "1000181430",
          "3005877899",
          "3018094310",
          "3002808382",
          "3011491794",
          "3033527770",
          "3015437189",
          "1000114111",
          "3005810333",
          "3000126629",
          "3021060474",
          "2027062",
          "3017375271",
          "3004785967",
          "3014273597",
          "2243057",
          "3030412764",
          "3009715136",
          "3030089787",
          "3014925393",
          "3001556265",
          "3008496839",
          "1000112968",
          "1043214",
          "3016438694",
          "1047843",
          "3027339477",
          "3002808188",
          "3009196578",
          "3009077524",
          "3025442726",
          "3003340745",
          "3011713200",
          "3003927316",
          "3003477135",
          "3009963139",
          "3004754211",
          "3033525986",
          "3009048384",
          "3004977335",
          "3002806833",
          "3007143236",
          "3003935342",
          "3038614464",
          "3005430103",
          "1000511585",
          "3009973505",
          "3002175190",
          "3003768251",
          "3036255956",
          "3002807880",
          "3003600763",
          "3002807145",
          "3013099020",
          "2020550",
          "3011158427",
          "1037885",
          "3019878714",
          "1000222289",
          "3010398867",
          "3015276088",
          "3001451453",
          "3002980729",
          "3005116291",
          "3012939903",
          "3014279513",
          "3014634069",
          "3005706667",
          "3002466018",
          "3002808461",
          "2027467",
          "3015531959",
          "1037955",
          "3003040625"
        ],
        "device_name": "System, X-Ray, Fluoroscopic, Image-Intensified",
        "medical_specialty_description": "Radiology",
        "regulation_number": "892.1650",
        "device_class": "2"
      }
    }
  ]
}