{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-08-20",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "64556",
      "product_res_number": "Z-0045-2008",
      "event_date_initiated": "2007-09-26",
      "event_date_posted": "2007-10-18",
      "recall_status": "Terminated",
      "event_date_terminated": "2008-01-15",
      "res_event_number": "44940",
      "product_code": "F",
      "product_description": "Microtainer Safety flow Lancets; Yellow\nSterile Use once and discard;\n2.2mm Blade Length;\n1.0mm blade width;\nBecton Dickinson Vacutainer System;\nBecton Dickinson and Company;\nFranklin Lakes, NJ 07417",
      "code_info": "Catalog number 365759; Lot nuimbers: 2340323, 3017522, 3049342, 3069725, 3136072, 3149410, 3220675, 3213535, 2319394.",
      "firm_fei_number": "2243072",
      "recalling_firm": "Becton Dickinson & Company",
      "address_1": "1 Becton Dr",
      "city": "Franklin Lakes",
      "state": "NJ",
      "postal_code": "07417-1815",
      "additional_info_contact": "201-847-7186",
      "reason_for_recall": "Potential for the blade and holder assembly to become loose and then separate from the main body of the lancet during the process of disposal.",
      "root_cause_description": "Process design",
      "action": "Recall communication issued on or about 9/28/2007.",
      "product_quantity": "1,978,000 units",
      "distribution_pattern": "Product distributed worldwide to Australia, Brazil, Canada, China, Colombia, Guatemala, Hong Kong, India, Indonesia, Mexico, Malaysia, New Zealand, Pakistan, Philippines, Singapore, Thailand, and United States.",
      "openfda": {}
    }
  ]
}