{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-08-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "52219",
      "product_res_number": "Z-0172-2008",
      "event_date_initiated": "2007-08-02",
      "event_date_posted": "2007-11-28",
      "recall_status": "Terminated",
      "event_date_terminated": "2009-10-05",
      "res_event_number": "37912",
      "product_code": "LYJ",
      "pma_numbers": [
        "P970003"
      ],
      "product_description": "Cyberonics VNS Therapy Bipolar Lead, Model #302; Sterile, Single use only; Distributed by Cyberonics, Inc., Houston, TX  77058.",
      "code_info": "All manufactured and distributed leads.",
      "firm_fei_number": "1000117004",
      "recalling_firm": "Cyberonics, Inc",
      "address_1": "100 Cyberonics Blvd",
      "city": "Houston",
      "state": "TX",
      "postal_code": "77058-2069",
      "additional_info_contact": "800-332-1375 Ext. 223",
      "reason_for_recall": "Dissolution/Fractures to the leads of the VNS Therapy System",
      "root_cause_description": "Process control",
      "action": "Cyberonics issued a correction notice (Safety Alert) to physicians in June of 2007, informing them of the present labeling and including updated labeling to provide instructions on how to detect lead fractures and how to minimize risk to patients.",
      "product_quantity": "34,719 units.",
      "distribution_pattern": "Worldwide Distribution - USA and countries of Argentina, Australia, Austria, Belgium, Canada, Costa Rica, Cyprus, Czech Rep, Denmark, Finland, France, Germany, Greece, Hong Kong, Hungary, Iceland, India, Ireland, Israel, Italy, Jordan, Korea, Malta,  Mexico, Netherlands, Northern Ireland, Norway, Poland, Slovenia, South Africa, Spain, Sweden, Switzerland, Turkey, United Kingdom, and Venezuela.",
      "openfda": {
        "pma_number": [
          "P970003"
        ],
        "device_name": "Stimulator, Autonomic Nerve, Implanted For Epilepsy",
        "medical_specialty_description": "Unknown",
        "regulation_number": "",
        "device_class": "3",
        "registration_number": [
          "1644487"
        ],
        "fei_number": [
          "1000117004"
        ]
      }
    }
  ]
}