  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-08-20",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "51835",
      "product_res_number": "Z-0816-2007",
      "event_date_initiated": "2006-04-28",
      "event_date_posted": "2007-05-18",
      "recall_status": "Terminated",
      "event_date_terminated": "2008-02-24",
      "res_event_number": "37838",
      "product_code": "EDH",
      "k_numbers": [
        "K974877"
      ],
      "product_description": "Cranex Tome panoramic x-ray unit with Spiral Tomography for cross-sectional imaging.  manufactured by SOREDEX, PO Box 250, 00031 Helsinki, Finland.",
      "code_info": "Serial numbers X51146, X51147, X51148, X51154, X51155, X51156, X51170, X51171, X61188",
      "firm_fei_number": "3005383085",
      "recalling_firm": "Instrumentarium Dental, Inc.",
      "address_1": "300 West Edgerton Avenue",
      "city": "Milwaukee",
      "state": "WI",
      "postal_code": "53207",
      "reason_for_recall": "There may be a material defect in the up/down movement axle of the unit.  This may cause the upper part of the unit to suddenly come down.",
      "root_cause_description": "Other",
      "action": "A recall letter, dated 04/28/2006, was sent to affected customers .  The letter describes the issue, and requests that no patient operation is permitted until the unit is checked by authorized service technician.  If the service technician noted wear of the screw, the assembly was replaced.",
      "product_quantity": "9",
      "distribution_pattern": "Worldwide, including USA, Canada, Bermuda, and Colombia.",
      "openfda": {}
    }
  ]
}