      "cfres_id": "50124",
      "product_res_number": "Z-0498-2007",
      "event_date_initiated": "2006-08-17",
      "event_date_posted": "2007-02-17",
      "recall_status": "Terminated",
      "event_date_terminated": "2008-06-06",
      "res_event_number": "37215",
      "product_code": "J",
      "product_description": "Anaerotest for microbiology\nSpecification: Performance Test  Anerobic: white within 4-6 hours; aerobic: blue with 20 minutes\nMerck KgaA\n64271 Darmstadt, Germany\n1.19034.0002  17.08.06  1 unit\nUPC 4 022536 667928",
      "code_info": "Item number 1.19034.0002 Lot # OC553145",
      "firm_fei_number": "2244894",
      "recalling_firm": "EMD Chemicals Inc",
      "address_1": "480 S Democrat Rd",
      "city": "Gibbstown",
      "state": "NJ",
      "postal_code": "08027-1239",
      "additional_info_contact": "Joseph Scheollkopff\n856-423-6300",
      "reason_for_recall": "Lot # OC553145 of Anaerotest failed to remain in specification. QC indicator shows an anaerobic environment even if oxygen is present.",
      "root_cause_description": "Other",
      "action": "A Recall Letter was issued by EMD Chemicals on 8/17/2006 to discontinue product usage and for product return.",
      "product_quantity": "8 rolls",
      "distribution_pattern": "Product was only sent to one US customer: Becton Dickinson, Cayey. Pureto Rico.",