{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-08-20",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "49967",
      "product_res_number": "Z-0420-2007",
      "event_date_initiated": "2006-08-29",
      "event_date_posted": "2007-02-08",
      "recall_status": "Terminated",
      "event_date_terminated": "2012-04-19",
      "res_event_number": "37083",
      "product_code": "GKZ",
      "k_numbers": [
        "K030828"
      ],
      "product_description": "CXP Analysis Software, Versions 2.0 & 2.1: Part #722396 CXP Analysis Software Kit V2.1, Single User; Part #722395 CXP Analysis Software Kit V2.1, Single User Upgrade; Part #722397 CXP Analysis Software Kit V2.1, 3 User Pack; Part #722398 CXP Analysis Software Kit V2.1, 5 User Pack; Part #722399 CXP Analysis Software Kit V2. 1, 15 User Pack; Part #\n722593 CXP Analysis Software Kit V2.1, 5 Network User Pack\nPart #722594 CXP Analysis Software Kit V2.1, 10 Network User Pack; Part #722595 CXP Analysis Software Kit V2.1, 15 Network User Pack; Part #629640 CXP Analysis Software Kit V2.0, Single User; Part #629639 CXP Analysis Software Kit V2.0, Single User Upgrade; Part #629641 CXP Analysis Software Kit V2.0, 3 User Pack; Part #629642 CXP Analysis Software Kit V2.0, 5 User Pack; Part #629643 CXP Analysis Software Kit V2.0, 15 User Pack; Part #629644 CXP Analysis Software Kit V2.0, 5 Network User Pack; Part #629645             CXP Analysis Software Kit V2.0, 10 Network User Pack;         Part #629646 CXP Analysis Software Kit V2.0, 15 Network User Pack.  The Cytomics FC 500 is a system for the qualitative and quantitative measurement of biological and physical properties of cells and other particles.",
      "code_info": "Part Numbers: 722396, 722395, 722397, 722398, 722399,          722593, 722594, 722595, 629640, 629639, 629641, 629642,      629643, 629644, 629645, 629646.   Software, Versions 2.0 & 2.1",
      "firm_fei_number": "2050012",
      "recalling_firm": "Beckman Coulter Inc",
      "address_1": "200 S Kraemer Blvd",
      "city": "Brea",
      "state": "CA",
      "postal_code": "92822-6208",
      "additional_info_contact": "Nora Zerounian\n714-961-3634",
      "reason_for_recall": "Two workflow scenarios associated with renaming regions may produce incorrect results.",
      "root_cause_description": "Other",
      "action": "A Product Corrective Action (PCA) letter was sent on the week of August 29, 2006 to all accounts that have an FC500 or have ordered CXP or MXP software to inform them that a) incorrect statistics are generated when a polygonal region with vertical re-entrant vertices is renamed using the Region Properties dialog box. b)when a region is copied from one plot to another, and then the new region is renamed, the statistics are not updated with the new region name. (Letter was sent by US mail).  Letter includes immediate user preventive action to remedy both issues.  A response form is included for consignees to return to recalling firm.",
      "product_quantity": "200 units nationwide",
      "distribution_pattern": "Nationwide & Canada",
      "openfda": {
        "k_number": [
          "K991454",
          "K053308",
          "K914014",
          "K882370",
          "K980635",
          "K841852",
          "K081930",
          "K791108",
          "K935720",
          "K936023",
          "K884605",
          "K160429",
          "K893852",
          "K923237",
          "K022161",
          "K190898",
          "K873256",
          "K961623",
          "K950057",
          "K053497",
          "K844645",
          "K022668",
          "K032039",
          "K003796",
          "K982166",
          "K211840",
          "K954594",
          "K050436",
          "K925528",
          "K251249",
          "K953340",
          "K040725",
          "K963263",
          "K974360",
          "K002646",
          "K071141",
          "K940003",
          "K925344",
          "K980614",
          "K911385",
          "K962633",
          "K945692",
          "K042724",
          "K901186",
          "K883753",
          "K950742",
          "K071681",
          "K900684",
          "K061150",
          "K922084",
          "K950568",
          "K153538",
          "K011342",
          "K151133",
          "K022200",
          "K171655",
          "K953598",
          "K903665",
          "K960531",
          "K021241",
          "K060423",
          "K914848",
          "K950482",
          "K913305",
          "K955909",
          "K012904",
          "K954688",
          "K893423",
          "K162414",
          "K042251",
          "K960427",
          "K884346",
          "K901187",
          "K901189",
          "K965053",
          "K030144",
          "K921279",
          "K945756",
          "K913192",
          "K990327",
          "K945212",
          "K960887",
          "K032000",
          "K842636",
          "K852988",
          "K922704",
          "K770701",
          "K883177",
          "K790816",
          "K032214",
          "K170353",
          "K761208",
          "K953643",
          "K925909",
          "K100489",
          "K844910",
          "K910283",
          "K992849",
          "K102064",
          "K991839",
          "K222148",
          "K900078",
          "K853304",
          "K981568",
          "K833600",
          "K961610",
          "K030263",
          "K890491",
          "K890349",
          "K780577",
          "K050719",
          "K954188",
          "K243348",
          "K970836",
          "K961278",
          "K050057",
          "K922744",
          "K250943",
          "K140911",
          "K993356",
          "K090810",
          "K012934",
          "K072869",
          "K240252",
          "K920544",
          "K090013",
          "K943894",
          "K930309",
          "K112755",
          "K990311",
          "K920949",
          "K800533",
          "K951100",
          "K926124",
          "K840195",
          "K870233",
          "K040026",
          "K042173",
          "K061616",
          "K875047",
          "K770869",
          "K971152",
          "K800121",
          "K854368",
          "K964856",
          "K950918",
          "K253212",
          "K003677",
          "K915071",
          "K243283",
          "K943284",
          "K961435",
          "K821186",
          "K961439",
          "K871837",
          "K951073",
          "K891276",
          "K982172",
          "K971736",
          "K991402",
          "K945837",
          "K060375",
          "K232946",
          "K885028",
          "K955673",
          "K220031",
          "K952182",
          "K030828",
          "K980858",
          "K030408",
          "K032524",
          "K000611",
          "K973483",
          "K992511",
          "K240402",
          "K944127",
          "K945678",
          "K060656",
          "K041882",
          "K854242",
          "K873734",
          "K181288",
          "K062087",
          "K180020",
          "K913259",
          "K181475",
          "K782050",
          "K901188",
          "K971205",
          "K803219",
          "K960827",
          "K022528",
          "K880616",
          "K982231",
          "K193124",
          "K030799",
          "K941713",
          "K071967",
          "K892205",
          "K874763",
          "K893853",
          "K032342",
          "K942797",
          "K024002",
          "K970326",
          "K932530",
          "K910473",
          "K120771",
          "K964754",
          "K071143",
          "K063407",
          "K900593",
          "K982802",
          "K141161",
          "K950626",
          "K992875",
          "K874427",
          "K882612",
          "K954570",
          "K885093",
          "K944128",
          "K991539",
          "K110381",
          "K855117",
          "K895127",
          "K953448",
          "K230887",
          "K972354",
          "K914744",
          "K251371",
          "K160538",
          "K884020",
          "K953302",
          "K172604",
          "K950917",
          "K770550",
          "K930913",
          "K933475",
          "K951496",
          "K961701",
          "K991142",
          "K946335",
          "K041074",
          "K872761",
          "K081526",
          "K051199",
          "K951632",
          "K761207",
          "K081495",
          "K883997",
          "K162977",
          "K850354",
          "K932532",
          "K921567",
          "K964988",
          "K043215",
          "K143348",
          "K983562",
          "K922181",
          "K950916",
          "K880507",
          "K982167",
          "K910834",
          "K950342",
          "K981342",
          "K875346",
          "K855116",
          "K052905",
          "K896873",
          "K933486",
          "K933478",
          "K061574",
          "K965232",
          "K953116",
          "K911626",
          "K925255",
          "K182886",
          "K942525",
          "K923790",
          "K924587",
          "K050758",
          "K190294",
          "K944126",
          "K081890",
          "K945205",
          "K040073",
          "K852371",
          "K833618",
          "K853029",
          "K914846",
          "K962768",
          "K904209",
          "K932531",
          "K941171",
          "K925274",
          "K143577",
          "K946055",
          "K963963",
          "K791041",
          "K882579",
          "K964618",
          "K992444",
          "K962305",
          "K081213",
          "K093394",
          "K230878",
          "K853027",
          "K030291",
          "K091313",
          "K010765",
          "K111534",
          "K955715",
          "K922745",
          "K960072",
          "K051459",
          "K090346",
          "K970742",
          "K973634",
          "K895917",
          "K032013",
          "K950625",
          "K990352",
          "K971110",
          "K770244",
          "K964974",
          "K874838",
          "K030513",
          "K050421",
          "K922085",
          "K963954",
          "K955504",
          "K061667",
          "K791903",
          "K831271",
          "K082091",
          "K953498",
          "K932030",
          "K950919",
          "K874147",
          "K943967",
          "K823355",
          "K923773",
          "K871600",
          "K884006",
          "K950949",
          "K141681",
          "K933878",
          "K894179",
          "K112605",
          "K200828",
          "K944253",
          "K240636",
          "K090967",
          "K051122",
          "K883710",
          "K900276",
          "K930148",
          "K990172",
          "K770680",
          "K030606",
          "K081610",
          "K020496",
          "K951459",
          "K010953",
          "K210346",
          "K952789",
          "K050589",
          "K882557",
          "K904390",
          "K911793",
          "K083420",
          "K102644",
          "K904253",
          "K000897",
          "K900802",
          "K071562",
          "K182389",
          "K864579",
          "K991605",
          "K051215"
        ],
        "registration_number": [
          "3007113174",
          "3027394506",
          "3003601075",
          "3009495621",
          "3006412379",
          "3016090200",
          "3039523350",
          "3020704087",
          "3007349751",
          "2221819",
          "8023018",
          "1221433",
          "2432235",
          "9610099",
          "9613959",
          "3002809144",
          "3003423869",
          "3006198300",
          "2032112",
          "3009156722",
          "3013059683",
          "8040564",
          "1036362",
          "1063851",
          "2914537",
          "3009297077",
          "3007750372",
          "3010421384",
          "3006606827",
          "3009834076",
          "3013445220",
          "1181121",
          "2919069",
          "3011765035",
          "3009092698",
          "1061932",
          "1415939",
          "3010960914",
          "3014053517",
          "3003589956",
          "1064608",
          "1950302",
          "3016438761",
          "1017835",
          "3009711478",
          "1000292377",
          "3004008703",
          "3010324880",
          "8020888",
          "3007603826",
          "3009189893",
          "3000148879",
          "3038625004",
          "3005333358",
          "2182501",
          "3008096179",
          "3003630693",
          "3006946274",
          "3013608230",
          "1832816",
          "3008425728",
          "3008352382",
          "3002759288"
        ],
        "fei_number": [
          "3007113174",
          "3027394506",
          "3003601075",
          "3009495621",
          "3006412379",
          "3016090200",
          "3039523350",
          "3020704087",
          "3007349751",
          "2221819",
          "3003076831",
          "1221433",
          "2432235",
          "3000203293",
          "3002807272",
          "3002809144",
          "3003071743",
          "3003423869",
          "3006198300",
          "3013059683",
          "3009156722",
          "1036362",
          "2914537",
          "3009297077",
          "3007750372",
          "3010421384",
          "3006606827",
          "3009834076",
          "3013445220",
          "1181121",
          "2919069",
          "3011765035",
          "3009092698",
          "1415939",
          "3010960914",
          "3014053517",
          "3003589956",
          "1950302",
          "3016438761",
          "1017835",
          "3009711478",
          "1000292377",
          "3004008703",
          "3010324880",
          "3002806778",
          "3007603826",
          "3009189893",
          "3000148879",
          "3038625004",
          "3005333358",
          "2182501",
          "3008096179",
          "3003630693",
          "6137",
          "3006946274",
          "3013608230",
          "1000442352",
          "3007019572",
          "1832816",
          "3002808301",
          "3008425728",
          "3008352382",
          "3002759288"
        ],
        "device_name": "Counter, Differential Cell",
        "medical_specialty_description": "Hematology",
        "regulation_number": "864.5220",
        "device_class": "2"
      }
    }
  ]
}