  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-08-20",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "49741",
      "product_res_number": "Z-0485-2007",
      "event_date_initiated": "2006-10-09",
      "event_date_posted": "2007-01-11",
      "recall_status": "Terminated",
      "event_date_terminated": "2012-04-26",
      "res_event_number": "37003",
      "product_code": "FMT",
      "k_numbers": [
        "K970432"
      ],
      "product_description": "Fisher & Paykel IW930 CosyCot  Infant Radiant Warmer (with integrated bed) fitted with 9001W005J Elevator Module",
      "code_info": "9893JEU00110,     9993JEU00004,     9993JEU00018,     9993JEU00019,     9993JEU00021,     9993JEU00081,     9993JEU00082,     9993JEU00083,     9993JEU00084,     9993JEU00139,     9993JEU00140,     9993JEU00150,     9993JEU00151,     9993JEU00204,     9993JEU00205,     9993JEU00332,     9993JEU00333,     9993JEU00424,     9993JEU00448,     9993JEU00449,     9993JEU00450,     9993JEU00494,     9993JEU00536,     9993JEU00537,     9993JEU00605,     9993JEU00606,     9993JEU00662,     9993JEU00663,     9993JEU00669,     9993JEU00670,     0093JEU00064,     0093JEU00065,     0093JEU00066,     0093JEU00067,     0093JEU00076,     0093JEU00077,     0093JEU00078,     0093JEU00079,     0093JEU00086,     0093JEU00087,     0093JEU00088,     0093JEU00131,     0093JEU00132,     0093JEU00133,     0093JEU00134,     0093JEU00200,     0093JEU00266,     0093JEU00305,     0093JEU00306,     0093JEU00307,     0093JEU00308,     0093JEU00309,     0093JEU00310,     0093JEU00345,     0093JEU00346,  2000-93JEU00402,  2000-93JEU00436,  2000-93JEU00437,  2000-93JEU00438,  2000-93JEU00439,  2000-93JEU00499,  2000-93JEU00500,  2000-93JEU00505,  2000-93JEU00506,  2000-93JEU00507,  2000-93JEU00508,  2000-93JEU00541,  2000-93JEU00602,  2000-93JEU00603,  2000-93JEU00604,  2000-93JEU00605,  2000-93JEU00606,  2000-93JEU00607,  2000-93JEU00608,  2000-93JEU00609,  2000-93JEU00610,  2001-93JEU00048,  2001-93JEU00059,  2001-93JEU00060,  2001-93JEU00061,  2001-93JEU00062,  2001-93JEU00063,  2001-93JEU00064,  2001-93JEU00277,  2001-93JEU00293,  2001-93JEU00294,  2001-93JEU00295,  2001-93JEU00304,  2001-93JEU00305,  2001-93JEU00306,  2001-93JEU00403,  2001-93JEU00404,  2001-93JEU00405,  2001-93JEU00406,  2001-93JEU00407,  2001-93JEU00408,  2001-93JEU00409,  2001-93JEU00435,  2001-93JEU00436,  2001-93JEU00437,  2001-93JEU00438,  2001-93JEU00439,  2001-93JEU00440,  2001-93JEU00441,  2001-93JEU00442,  2001-93JEU00443,  2001-93JEU00444,  2001-93JEU00445,  2001-93JEU00446,  2001-93JEU00447,  2001-93JEU00448,  2001-93JEU00449,  2001-93JEU00450,  2001-93JEU00451,  2001-93JEU00452,     010827000538,     010907000568,     011002000604,     011002000605,     011002000606,     011019000663,     011019000664,     011019000665,     011019000666,     011106000715,     011106000716,     011115000743,     011115000744,     011115000745,     011115000746,     011116000747,     011116000748,     011116000749,     011116000750,     011119000751,     011119000752,     011217000830,     011217000831,     011217000832,     011217000833,     011217000834,     020130000101,     020130000102,     020130000103,     020218000186,     020218000187,     020218000188,     020218000189,     020218000190,     020221000204,     020221000205,     020221000206,     020221000207,     020221000208,     020226000217,     020226000218, 020403000292, 020405000317, 020405000318, 20405000319, 20405000320, 020405000321, 020405000322 , 020510000390, 020510000391, 020510000392 , 020510000393, 020510000394, 020523000410, 020523000411, 020523000412, 020523000413, 020606000420, 020611000432, 020611000433, 020611000434, 020611000435, 020611000436, 020611000437, 20611000438, 020611000439, 020611000440, 020611000441, 020611000442, 020611000443, 020612000444, 020801000544, 020801000545, 020802000546, 020802000547, 020802000548, 020802000549, 020802000550, 020802000551, 020904000736, 020911000676, 020912000682, 020913000689, 020916000699, 020916000700, 020917000701, 020917000702, 020917000703, 020917000704, 020917000705, 020917000706, 020917000707, 020917000708, 020918000709, 020918000710, 020924000732, 020924000733, 020924000734, 020924000735, 021004000751, 021008000758, 021008000759, 021105000876, 021105000879, 021105000880, 021107000890, 021107000891, 021112000910, 021112000911, 021112000912, 030113000041, 030113000042, 030113000043, 030113000044, 030113000045, 030117000074, 030117000075, 030117000076, 030117000077, 030120000078, 030120000079, 030120000080, 030120000081, 030305000241, 030307000245, 030307000246, 030307000247, 030307000248, 030307000249, 030307000250, 030307000251, 030307000252, 030313000253, 030313000254, 030313000255, 030408000300, 030408000301, 030408000302, 030505000356, 030505000357, 030512000384, 030512000385, 030512000386, 030512000387, 030512000388, 030512000389, 030512000390, 030616000477, 030616000478, 030618000494, 030618000495, 030618000496, 030618000497, 030618000498, 030618000499, 030618000500, 030618000501, 030708000558, 030708000559, 30708000560, 30708000560, 030711000580, 030711000581, 030711000582, 030711000583, 030717000599, 030722000611, 030802000635, 030802000636, 030804000637, 030812000662, 030905000702, 030905000703, 030905000704, 030912000718, 030922000747, 031003000780, 031003000781, 031015000814, 031015000815, 031015000816, 031015000817, 031021000841, 031021000842, 031021000843, 031021000844, 031028000862, 031028000863, 031104000876, 031106000891, 031106000892, 031120000964, 031125001001, 031125001002, 031125001003, 031125001004, 031125001005, 031217001112, 040112000001, 040112000002, 040112000003, 040112000004, 040112000006, 040112000007, 040112000008, 040112000009, 040112000010, 040112000011, 040119000023, 040119000024, 040120000030, 040120000031, 040204000112, 040204000113 & 040204000114",
      "firm_fei_number": "3002806988",
      "recalling_firm": "Fisher & Paykel Healthcare Ltd",
      "address_1": "15 Maurice Paykel Place",
      "city": "Auckland",
      "country": "New Zealand",
      "reason_for_recall": "This firm has received Product Complaint Reports concerning bending of the pivot plate. The elevator base of the CosyCot Infant Warmer incorporates a pivot plate which forms part of the mechanism allowing the CosyCot to raise and lower in height.  A bent pivot plate generally results in the Warmer being unable to elevate properly.  Other component failures may result.",
      "root_cause_description": "Other",
      "action": "Fisher & Paykel Healthcare commenced sending out Product Correction Notifications letter via U.S. Postal Service certified   \nmail October 9, 2006. This Product Correction Notice  explained the nature of the problem and include a reminder of the maximum recommended weight of accessories. Two FAX Response Forms were provided, Part A for initial receipt of the Product Correction Notice and Part B for confirmation of the completed upgrade.",
      "product_quantity": "273 units in US & 9 in Canada",
      "distribution_pattern": "Nationwide and Canada",
      "openfda": {
        "k_number": [
          "K853926",
          "K871178",
          "K935609",
          "K862921",
          "K921547",
          "K912921",
          "K840792",
          "K201628",
          "K812745",
          "K853351",
          "K971366",
          "K191425",
          "K871594",
          "K020966",
          "K881713",
          "K971198",
          "K120642",
          "K964401",
          "K971695",
          "K861781",
          "K925006",
          "K020219",
          "K101804",
          "K813408",
          "K851449",
          "K790137",
          "K946005",
          "K884283",
          "K915831",
          "K872742",
          "K871177",
          "K840858",
          "K913945",
          "K830748",
          "K950143",
          "K952800",
          "K884160",
          "K970074",
          "K931529",
          "K972885",
          "K840857",
          "K850069",
          "K885350",
          "K952801",
          "K950422",
          "K971461",
          "K072157",
          "K122267",
          "K243606",
          "K952807",
          "K003335",
          "K884285",
          "K033040",
          "K833802",
          "K881670",
          "K060461",
          "K200319",
          "K173516",
          "K952804",
          "K926030",
          "K090697",
          "K875270",
          "K790948",
          "K121625",
          "K865012",
          "K952803",
          "K910915",
          "K884280",
          "K942905",
          "K852943",
          "K940951",
          "K952805",
          "K120937",
          "K102227",
          "K863208",
          "K953489",
          "K213551",
          "K970686",
          "K942904",
          "K832637",
          "K921766",
          "K963058",
          "K002355",
          "K070377",
          "K970432",
          "K802365",
          "K123309",
          "K853794",
          "K934677",
          "K960080",
          "K012229",
          "K230278"
        ],
        "registration_number": [
          "3030733800",
          "",
          "3012348048",
          "3004050971",
          "1055236",
          "3010838917",
          "1000617661",
          "3014220748",
          "3001104093",
          "1317188",
          "9617277",
          "3005934741",
          "9710602",
          "1057358",
          "3005652389",
          "9613793",
          "1043214",
          "3021559257",
          "1625392",
          "2510954",
          "2027062",
          "3008729547",
          "3007734888",
          "1314417",
          "1226711",
          "3036470679",
          "1054241",
          "9611500",
          "1000604079",
          "9611451",
          "1054713",
          "1034876",
          "2085602",
          "3012086037"
        ],
        "fei_number": [
          "3030733800",
          "",
          "1000114111",
          "3012348048",
          "3004050971",
          "3023530182",
          "3010838917",
          "1000208608",
          "1000617661",
          "1000112877",
          "3014220748",
          "1317188",
          "3001104093",
          "3005934741",
          "3003338236",
          "3004846435",
          "1043214",
          "3021559257",
          "1000112968",
          "3006469671",
          "2510954",
          "2027062",
          "3008729547",
          "3003513129",
          "3007734888",
          "1314417",
          "3000210616",
          "3036470679",
          "1000604079",
          "3002806988",
          "1034876",
          "1000058335",
          "3012086037",
          "3003190524"
        ],
        "device_name": "Warmer, Infant Radiant",
        "medical_specialty_description": "General Hospital",
        "regulation_number": "880.5130",
        "device_class": "2"
      }
    }
  ]
}