  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-08-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "49460",
      "product_res_number": "Z-0504-2007",
      "event_date_initiated": "2006-11-09",
      "event_date_posted": "2007-02-15",
      "recall_status": "Terminated",
      "event_date_terminated": "2007-06-20",
      "res_event_number": "36855",
      "product_code": "IXR",
      "product_description": "Stille Sonesta Video Fluoroscopy Table 6210; manufactured by Stille AB, Solna, Sweden.",
      "code_info": "Serial Numbers:  12-70000 - 12-70017, 00010, 00012, 00013 - 00022, 00024 - 00027, 00030 - 00034, 00036 - 00040, 00043 - 00044, 00045 - 00049, 00053 - 00054, 00057 - 00065, 00067 - 00069, 00071.  Outside U.S. Serial Numbers: 12-70018 - 12-70019, 23-70000 - 23-70009, 23-70008, 23-700010, 00011, 00023, 00028 - 00029, 00050 - 00052, 00055 - 00056, 00066, 00072, 00073, 00074.",
      "firm_fei_number": "1000516389",
      "recalling_firm": "Stille-Sonesta, Inc",
      "address_1": "2220 Canton St Apt 209 2220 Canton St",
      "address_2": "Apt 209",
      "city": "Dallas",
      "state": "TX",
      "postal_code": "75201-5932",
      "reason_for_recall": "Bolts may back out of their joints causing the back section of the table to tilt.",
      "root_cause_description": "Other",
      "action": "Notifications letters sent to consignees on 11/20/06.  Users are instructed to check the bolts located under the table. The Notice includes instructions with pictures for checking the bolts and tightening them if necessary.  Stille AB has developed a kit for replacing the bolt that is involved and will send a service engineer out to the site to do the replacement.",
      "product_quantity": "66 units.",
      "distribution_pattern": "Nationwide.",
      "openfda": {
        "k_number": [
          "K854478",
          "K922364",
          "K911267",
          "K853175",
          "K912741",
          "K850948",
          "K883162",
          "K896119",
          "K880529",
          "K922591",
          "K781532",
          "K953556",
          "K963851",
          "K771293",
          "K945842",
          "K792645",
          "K920742",
          "K971128",
          "K973039",
          "K943581",
          "K944987",
          "K884531",
          "K860913",
          "K951176",
          "K950402",
          "K940721",
          "K914417",
          "K961022",
          "K882019",
          "K950981",
          "K911098",
          "K792326",
          "K910838",
          "K920722",
          "K970640",
          "K925175",
          "K924326",
          "K892226",
          "K840226",
          "K905152",
          "K893673",
          "K896324",
          "K760466",
          "K854486",
          "K902989",
          "K905265",
          "K912586",
          "K905501",
          "K843926",
          "K903422",
          "K870707",
          "K000412",
          "K854085",
          "K912996",
          "K874114",
          "K874113",
          "K914165",
          "K883153",
          "K935444",
          "K965105",
          "K791812",
          "K954356",
          "K901654",
          "K931563",
          "K920728",
          "K965096",
          "K915853",
          "K953557",
          "K883863",
          "K934574",
          "K896583",
          "K770334",
          "K840227",
          "K903395",
          "K905153",
          "K913147",
          "K920726"
        ],
        "registration_number": [
          "3007912863",
          "3033527770",
          "3002808176",
          "3013846070",
          "3003768251",
          "3000126629",
          "9613445",
          "1223864",
          "2243057",
          "3012010520",
          "2529571",
          "3009077524",
          "9616014",
          "3010496070",
          "3003927316",
          "3015276088",
          "8022065",
          "3008284470",
          "3033525986",
          "3004977335",
          "9610903",
          "1225473",
          "8030233"
        ],
        "fei_number": [
          "3002808139",
          "1000121370",
          "1000305755",
          "3007912863",
          "3033527770",
          "3002808176",
          "3013846070",
          "3002175190",
          "3003768251",
          "3000126629",
          "3002807145",
          "2243057",
          "3012010520",
          "3009077524",
          "3010496070",
          "3003927316",
          "3003434037",
          "3015276088",
          "3002675731",
          "3001237319",
          "3008284470",
          "3033525986",
          "3004977335"
        ],
        "device_name": "Table, Radiographic, Tilting",
        "medical_specialty_description": "Radiology",
        "regulation_number": "892.1980",
        "device_class": "2"
      }
    }
  ]
}