  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-08-20",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "49169",
      "product_res_number": "Z-0636-2007",
      "event_date_initiated": "2006-10-02",
      "event_date_posted": "2007-03-17",
      "recall_status": "Terminated",
      "event_date_terminated": "2012-03-26",
      "res_event_number": "36685",
      "k_numbers": [
        "K033357"
      ],
      "product_description": "Brilliance CT System, Big Bore Configuration, Model #728243. (version 2.2.1)",
      "code_info": "Serial numbers 7009, 7056, 7060, 7066, 7069, 7072, 7079, 7088, 7111, 7112, 7113, 7114, 7117, 7118, 7119, 7121, 7125.",
      "firm_fei_number": "1525965",
      "recalling_firm": "Philips Medical Systems (Cleveland) Inc",
      "address_1": "595 Miner Road",
      "city": "Cleveland",
      "state": "OH",
      "postal_code": "44143-2131",
      "additional_info_contact": "George X. Kambic, Ph.D.\n440-483-2557",
      "reason_for_recall": "Two conditions have been identified with the Philips Medical System Brilliance Bore ,version 2.2.1 Tumor Localization, that may lead the user to misinterpret the displayed data.",
      "root_cause_description": "Other",
      "action": "Customer advisory notification letters dated 10/4/06 and 10/25/06 were sent to consignees providing short term work around instructions.  Software modifications installed by the recalling firm will follow as the long term correction.",
      "product_quantity": "17 units",
      "distribution_pattern": "Class II Recall-Nationwide Distribution-USA including states of KY, MD, TX, RI, NJ, MA, LA, VA, PA, SC, CA, OH, and FL.",
      "openfda": {}
    }
  ]
}