      "cfres_id": "48809",
      "product_res_number": "Z-0100-2007",
      "event_date_initiated": "2006-10-12",
      "event_date_posted": "2006-10-27",
      "recall_status": "Terminated",
      "event_date_terminated": "2012-08-30",
      "res_event_number": "36515",
      "product_code": "GCJ",
      "k_numbers": [
        "K943327"
      ],
      "product_description": "ConMed Laparoscopic Instrument: Endoweave Disposable Grasper, 5mm x 43cm length, REF/Product Code 60-6045-435.  --- The product is distributed sterile in heat-sealed tray and labeled with an expiration date that pertains only to the sterility.",
      "code_info": "REF/Product Code 60-6045-435, all lot codes manufactured between 08/25/2003 and 03/24/2005. [Lot codes on unopened boxes and packaging contain a lot code in the following form: for example 0409271 - 04 represents the year 2004, 09 represents the month of the year, 27 represents the day of the month, and the last digit 1 represents the manufacturing shift code. Lot codes are located on the device in this form: for example 40927 - 4 represents the year 2004, 09 represents the month of the year, and 27 represents the day of the month.]",
      "firm_fei_number": "3001451571",
      "recalling_firm": "Conmed Corporation",
      "address_1": "525 French Road",
      "city": "Utica",
      "state": "NY",
      "postal_code": "13502-5945",
      "additional_info_contact": "Ms. Patricia Cotter\n315-624-3533",
      "reason_for_recall": "The grasper jaws broke during laparoscopic procedures at the junction of the jaw and the tube.",
      "root_cause_description": "Nonconforming Material/Component",
      "action": "On 10/12/06, ConMed sent \"URGENT DEVICE RECALL\" packages (letter dated 10/11/06) to the domestic accounts via UPS Ground, and via UPS Express for the International accounts. Devices are to be removed from use and returned.",
      "product_quantity": "13 units",
      "distribution_pattern": "Worldwide",
      "openfda": {
        "k_number": [
          "K250411",
          "K130435",
          "K211234",
          "K973700",
          "K120278",
          "K990028",
          "K934725",
          "K962005",
          "K935293",
          "K926078",
          "K954174",
          "K952842",
          "K143299",
          "K943958",
          "K970900",
          "K261139",
          "K162077",
          "K210248",
          "K943117",
          "K925958",
          "K140593",
          "K210495",
          "K090792",
          "K913172",
          "K930617",
          "K213818",
          "K991734",
          "K970054",
          "K933025",
          "K020270",
          "K101937",
          "K943786",
          "K942703",
          "K041827",
          "K061180",
          "K260710",
          "K183675",
          "K203717",
          "K960655",
          "K983527",
          "K953903",
          "K211250",
          "K992612",
          "K921457",
          "K053382",
          "K190831",
          "K912502",
          "K010519",
          "K023261",
          "K954108",
          "K942910",
          "K952934",
          "K221533",
          "K911817",
          "K950753",
          "K123325",
          "K911478",
          "K922751",
          "K923726",
          "K931706",
          "K133203",
          "K991377",
          "K932692",
          "K934158",
          "K193161",
          "K943970",
          "K183264",
          "K932229",
          "K130121",
          "K990894",
          "K934098",
          "K252858",
          "K151609",
          "K103509",
          "K914547",
          "K970917",
          "K123102",
          "K172672",
          "K945101",
          "K913762",
          "K240174",
          "K943172",
          "K955404",
          "K060629",
          "K931705",
          "K932648",
          "K952959",
          "K002542",
          "K991228",
          "K935382",
          "K253541",
          "K031891",
          "K933067",
          "K933661",
          "K181608",
          "K925198",
          "K100442",
          "K933169",
          "K930301",
          "K993045",
          "K232519",
          "K935115",
          "K935762",
          "K142603",
          "K953440",
          "K031593",
          "K112349",
          "K936262",
          "K952977",
          "K130826",
          "K973794",
          "K122299",
          "K232157",
          "K012771",
          "K000202",
          "K921208",
          "K221004",
          "K220108",
          "K972298",
          "K142427",
          "K926223",
          "K052740",
          "K944467",
          "K946241",
          "K151356",
          "K922655",
          "K933789",
          "K062326",
          "K091697",
          "K941286",
          "K221861",
          "K931898",
          "K192918",
          "K011359",
          "K032889",
          "K933387",
          "K201243",
          "K973232",
          "K221365",
          "K132785",
          "K221591",
          "K933812",
          "K060182",
          "K090677",
          "K944974",
          "K982623",
          "K911164",
          "K231725",
          "K973795",
          "K942740",
          "K950286",
          "K981940",
          "K072674",
          "K123728",
          "K011189",
          "K953243",
          "K150705",
          "K934534",
          "K943671",
          "K081995",
          "K953377",
          "K951692",
          "K922797",
          "K082322",
          "K944418",
          "K102646",
          "K072715",
          "K942201",
          "K141272",
          "K935277",
          "K152149",
          "K941747",
          "K921040",
          "K930935",
          "K944536",
          "K925147",
          "K922319",
          "K951086",
          "K980022",
          "K182606",
          "K062779",
          "K260197",
          "K963329",
          "K972178",
          "K920040",
          "K941500",
          "K932697",
          "K992437",
          "K951878",
          "K151036",
          "K062050",
          "K940900",
          "K964123",
          "K092584",
          "K950214",
          "K951387",
          "K925499",
          "K932321",
          "K963046",
          "K990379",
          "K033842",
          "K190645",
          "K180579",
          "K093624",
          "K161720",
          "K201151",
          "K010620",
          "K936154",
          "K013165",
          "K944414",
          "K932604",
          "K191177",
          "K940155",
          "K926016",
          "K952264",
          "K243399",
          "K150633",
          "K980350",
          "K964334",
          "K983561",
          "K925602",
          "K191930",
          "K170084",
          "K914690",
          "K093785",
          "K930667",
          "K912097",
          "K961343",
          "K922243",
          "K202592",
          "K181395",
          "K070712",
          "K001169",
          "K093194",
          "K950076",
          "K953121",
          "K010489",
          "K912836",
          "K992324",
          "K243439",
          "K922685",
          "K952955",
          "K940199",
          "K934528",
          "K182160",
          "K963115",
          "K911488",
          "K241739",
          "K931274",
          "K931878",
          "K240635",
          "K930567",
          "K943253",
          "K080257",
          "K950747",
          "K020336",
          "K934022",
          "K951625",
          "K962147",
          "K250551",
          "K941276",
          "K232701",
          "K910246",
          "K935986",
          "K770711",
          "K955490",
          "K950500",
          "K914526",
          "K971850",
          "K202303",
          "K941027",
          "K190788",
          "K221217",
          "K994066",
          "K170683",
          "K952900",
          "K220872",
          "K051073",
          "K193026",
          "K932553",
          "K932441",
          "K212511",
          "K231400",
          "K973068",
          "K012539",
          "K934163",
          "K014005",
          "K971738",
          "K945369",
          "K023735",
          "K020310",
          "K940875",
          "K123847",
          "K914377",
          "K990691",
          "K952951",
          "K972504",
          "K101671",
          "K222504",
          "K232880",
          "K233327",
          "K950243",
          "K123110",
          "K143640",
          "K933421",
          "K926479",
          "K241530",
          "K152619",
          "K250130",
          "K231003",
          "K101193",
          "K945979",
          "K952954",
          "K935270",
          "K933049",
          "K950612",
          "K914098",
          "K082291",
          "K260177",
          "K141592",
          "K960637",
          "K993688",
          "K222130",
          "K981941",
          "K990912",
          "K931783",
          "K930605",
          "K941024",
          "K913500",
          "K252714",
          "K172789",
          "K230605",
          "K943307",
          "K930716",
          "K923528",
          "K952957",
          "K950336",
          "K013701",
          "K142547",
          "K952814",
          "K933192",
          "K933345",
          "K203603",
          "K983566",
          "K931275",
          "K921336",
          "K261249",
          "K912319",
          "K152400",
          "K170598",
          "K992288",
          "K951522",
          "K210584",
          "K991928",
          "K974771",
          "K013617",
          "K934027",
          "K151117",
          "K223923",
          "K936114",
          "K961430",
          "K984434",
          "K944954",
          "K974139",
          "K141594",
          "K925129",
          "K930607",
          "K931237",
          "K110257",
          "K823581",
          "K991171",
          "K950985",
          "K214046",
          "K944082",
          "K020915",
          "K132375",
          "K946023",
          "K030512",
          "K171726",
          "K131375",
          "K020307",
          "K954160",
          "K043541",
          "K942053",
          "K071488",
          "K181791",
          "K964258",
          "K083638",
          "K913810",
          "K915317",
          "K061666",
          "K012527",
          "K933895",
          "K974767",
          "K030240",
          "K212055",
          "K932977",
          "K945266",
          "K942601",
          "K031382",
          "K251923",
          "K121796",
          "K201198",
          "K945953",
          "K252985",
          "K973249",
          "K920595",
          "K943614",
          "K924728",
          "K230491",
          "K060534",
          "K924022",
          "K222575",
          "K023053",
          "K962799",
          "K010754",
          "K021456",
          "K981531",
          "K071756",
          "K930887",
          "K972471",
          "K173917",
          "K930466",
          "K992541",
          "K926089",
          "K931707",
          "K190029",
          "K953348",
          "K170030",
          "K994256",
          "K945474",
          "K934750",
          "K073719",
          "K091930",
          "K033644",
          "K973046",
          "K952902",
          "K946330",
          "K952932",
          "K925041",
          "K251209",
          "K201134",
          "K964171",
          "K120468",
          "K931962",
          "K010220",
          "K000424",
          "K060096",
          "K945457",
          "K941348",
          "K922405",
          "K171939",
          "K921925",
          "K951090",
          "K953438",
          "K942168",
          "K201526",
          "K942358",
          "K950231",
          "K171741",
          "K915544",
          "K912492",
          "K970369",
          "K930467",
          "K981700",
          "K973909",
          "K933896",
          "K952443",
          "K131115",
          "K933009",
          "K932627",
          "K240831",
          "K023630",
          "K952953",
          "K921922",
          "K083194",
          "K925824",
          "K934280",
          "K042923",
          "K934023",
          "K950744",
          "K914919",
          "K912343",
          "K944672",
          "K223116",
          "K011016",
          "K173272",
          "K200228",
          "K121893",
          "K935604",
          "K211148",
          "K221781",
          "K092821",
          "K080654",
          "K011538",
          "K934134",
          "K915099",
          "K231358",
          "K161421",
          "K941249",
          "K942202",
          "K041131",
          "K920599",
          "K931142",
          "K972927",
          "K121701",
          "K936097",
          "K945170",
          "K050829",
          "K133845",
          "K003871",
          "K954864",
          "K001597",
          "K934347",
          "K941541",
          "K934737",
          "K935207",
          "K141715",
          "K122526",
          "K000669",
          "K935878",
          "K926583",
          "K970056",
          "K103741",
          "K032339",
          "K080161",
          "K926010",
          "K914861",
          "K911243",
          "K964930",
          "K970949",
          "K030718",
          "K133375",
          "K922987",
          "K992250",
          "K934527",
          "K830719",
          "K041641",
          "K924035",
          "K912605",
          "K100180",
          "K991416",
          "K935826",
          "K030422",
          "K073009",
          "K962193",
          "K912247",
          "K934024",
          "K962651",
          "K062209",
          "K003217",
          "K943327",
          "K913469",
          "K120012",
          "K980129",
          "K934157",
          "K952935",
          "K911589",
          "K922082",
          "K113203",
          "K990414",
          "K251743",
          "K171475",
          "K945297",
          "K945520",
          "K926335",
          "K981999",
          "K123280",
          "K924931",
          "K911149",
          "K926080",
          "K951546",
          "K201434",
          "K143091",
          "K973543",
          "K990493",
          "K011257",
          "K933536",
          "K934338",
          "K934279",
          "K950502",
          "K202551",
          "K930569",
          "K935999",
          "K915286",
          "K942636",
          "K060051",
          "K924284",
          "K924792",
          "K951646",
          "K161017",
          "K232866",
          "K960878",
          "K924281",
          "K934078",
          "K942574",
          "K093783",
          "K934021",
          "K983538",
          "K913870",
          "K911490",
          "K241052",
          "K931425",
          "K100045",
          "K151990",
          "K010346",
          "K934026",
          "K932979",
          "K931278",
          "K950658",
          "K953410",
          "K030940",
          "K942306",
          "K072749",
          "K933243",
          "K091869",
          "K952958",
          "K000337",
          "K933750",
          "K935279",
          "K932806",
          "K941773",
          "K930082",
          "K971150",
          "K970605",
          "K950959",
          "K112659",
          "K201805",
          "K935709",
          "K953561",
          "K910985",
          "K050163",
          "K981598",
          "K133272",
          "K933330",
          "K971334",
          "K013551",
          "K940347",
          "K221740",
          "K930848",
          "K933725",
          "K951535",
          "K935786",
          "K222828",
          "K191295",
          "K931238",
          "K092789",
          "K945059",
          "K941967",
          "K984355",
          "K031956",
          "K944224",
          "K961492",
          "K013952",
          "K921260",
          "K930817",
          "K031720",
          "K941950",
          "K260414",
          "K172618",
          "K170103",
          "K022393",
          "K932033",
          "K081169",
          "K932995",
          "K912980",
          "K202582",
          "K911157",
          "K951251",
          "K932540",
          "K241913",
          "K153773",
          "K926444",
          "K930464",
          "K930352",
          "K251994",
          "K160801",
          "K233840",
          "K923982",
          "K942416",
          "K931895",
          "K152848",
          "K103510",
          "K945862",
          "K980787",
          "K932695",
          "K212500",
          "K090275",
          "K951990",
          "K072811",
          "K013090",
          "K953376",
          "K965001",
          "K222783",
          "K963184",
          "K060717",
          "K935426",
          "K930254",
          "K923945",
          "K911754",
          "K121380",
          "K140101",
          "K030824",
          "K200310",
          "K933104",
          "K932346",
          "K192643",
          "K951482",
          "K990635",
          "K024294",
          "K913104",
          "K974336",
          "K930813",
          "K974592",
          "K940891",
          "K243008",
          "K060233",
          "K031098",
          "K931879",
          "K912973",
          "K233451",
          "K951480",
          "K912579",
          "K083211",
          "K960301",
          "K170379",
          "K132169",
          "K241361",
          "K082735",
          "K931850",
          "K241111",
          "K073279",
          "K991703",
          "K936195",
          "K990663",
          "K950103",
          "K180977",
          "K070509",
          "K143221",
          "K060777",
          "K082796",
          "K241796",
          "K941401",
          "K233892",
          "K940960",
          "K932796",
          "K153606",
          "K252532",
          "K930933",
          "K991718",
          "K102379",
          "K972699",
          "K950170",
          "K930550",
          "K961224",
          "K894716",
          "K232042",
          "K911730",
          "K100959",
          "K051610",
          "K980079",
          "K955408",
          "K912859",
          "K973430",
          "K940193",
          "K982445",
          "K091350",
          "K970937",
          "K091515",
          "K923219",
          "K912927",
          "K093372",
          "K041795",
          "K192499",
          "K930224",
          "K931649",
          "K162584",
          "K061292",
          "K203255",
          "K950005",
          "K233263",
          "K973698",
          "K922246",
          "K915239",
          "K941390",
          "K950702",
          "K050639",
          "K972280",
          "K193137",
          "K942414",
          "K210088",
          "K023629",
          "K954649",
          "K010569",
          "K911486",
          "K973943",
          "K230216",
          "K934094",
          "K932696",
          "K023095",
          "K172940",
          "K191043",
          "K120061",
          "K940011",
          "K926281",
          "K070198",
          "K931223",
          "K952723",
          "K223759",
          "K964672",
          "K091648",
          "K911744",
          "K951386",
          "K233225",
          "K142903",
          "K991444",
          "K243307",
          "K990785",
          "K930147",
          "K972958",
          "K953419",
          "K934077",
          "K961800",
          "K924916",
          "K953409",
          "K951060",
          "K923547",
          "K925410",
          "K161122",
          "K961175",
          "K193315",
          "K972685",
          "K082619",
          "K950297",
          "K940157",
          "K933388",
          "K963126",
          "K951082",
          "K953418",
          "K080977",
          "K932698",
          "K911494",
          "K925699",
          "K932976",
          "K931778",
          "K935234",
          "K950811",
          "K153218",
          "K212950",
          "K954089",
          "K220895",
          "K942839",
          "K933720",
          "K961182",
          "K933075",
          "K936248",
          "K131688",
          "K990731",
          "K950632",
          "K083218",
          "K945637",
          "K921287",
          "K915854",
          "K963919",
          "K984600",
          "K040244",
          "K950339",
          "K141902",
          "K952278",
          "K200053",
          "K233386",
          "K926181",
          "K943056",
          "K241402",
          "K944299",
          "K915790",
          "K002013",
          "K150253",
          "K926341",
          "K182434",
          "K932060",
          "K223562",
          "K940427",
          "K940687",
          "K100533",
          "K011944",
          "K013695",
          "K242927",
          "K924372",
          "K972984",
          "K940259",
          "K952901",
          "K945890",
          "K213597",
          "K951974",
          "K100548",
          "K042740",
          "K951481",
          "K954011",
          "K932466",
          "K000422",
          "K934185",
          "K001594",
          "K944303",
          "K140662",
          "K202931",
          "K940948",
          "K931276",
          "K201593",
          "K934704",
          "K932282",
          "K934398",
          "K940346",
          "K961940",
          "K152663",
          "K933160",
          "K951252",
          "K925914",
          "K932629",
          "K250124",
          "K921086",
          "K032680",
          "K922435",
          "K914782",
          "K915372",
          "K920611",
          "K254228",
          "K912556",
          "K160172",
          "K931518",
          "K931337",
          "K914869",
          "K163102",
          "K090340",
          "K010033",
          "K934511",
          "K960232",
          "K950746",
          "K932045",
          "K182024",
          "K933076",
          "K932885",
          "K222079",
          "K952936",
          "K972679",
          "K230058",
          "K925882",
          "K933790",
          "K192891",
          "K071292",
          "K012968",
          "K111845",
          "K990237",
          "K150956",
          "K160332",
          "K946239",
          "K934173",
          "K952444",
          "K950726",
          "K172038",
          "K941222",
          "K014158",
          "K033249",
          "K011183",
          "K833189",
          "K972056",
          "K120403",
          "K920068",
          "K925382",
          "K033132",
          "K945394",
          "K913999",
          "K173822",
          "K942924",
          "K222890",
          "K223593",
          "K102839",
          "K943976",
          "K011841",
          "K072613",
          "K930465",
          "K933032",
          "K092504",
          "K992852",
          "K252640",
          "K162836",
          "K942491",
          "K932987",
          "K925149",
          "K061555",
          "K020428",
          "K950862",
          "K071571",
          "K063528",
          "K922526",
          "K050051",
          "K002431",
          "K223718",
          "K110121",
          "K972578",
          "K163654",
          "K911053",
          "K931522",
          "K093449",
          "K931092",
          "K926214",
          "K931259",
          "K941136",
          "K221252",
          "K073170",
          "K970214",
          "K925669",
          "K970888",
          "K923201",
          "K936250",
          "K950126",
          "K924058",
          "K953549",
          "K250728",
          "K954853",
          "K152565",
          "K201641",
          "K000663",
          "K001264",
          "K924149",
          "K062968",
          "K960883",
          "K935109",
          "K945263",
          "K952926",
          "K940731",
          "K141323",
          "K021731",
          "K926282",
          "K940232",
          "K933458",
          "K042741",
          "K932626",
          "K202130",
          "K070102",
          "K951986",
          "K100109",
          "K934165",
          "K930462",
          "K181258",
          "K241401",
          "K940398",
          "K980123",
          "K090631",
          "K960748",
          "K925565",
          "K062727",
          "K935073",
          "K011501",
          "K241360",
          "K962684",
          "K261614",
          "K042412",
          "K153542",
          "K243968",
          "K914752",
          "K933724",
          "K011997",
          "K160393",
          "K943435",
          "K013576",
          "K980023",
          "K990188",
          "K101810",
          "K935071",
          "K921103",
          "K242513",
          "K922424",
          "K933776",
          "K162197",
          "K953123",
          "K934739",
          "K930572",
          "K014250",
          "K260183",
          "K953509",
          "K911484",
          "K960562",
          "K050027",
          "K952956",
          "K002924",
          "K940551",
          "K910132",
          "K211043",
          "K992190",
          "K031847",
          "K254014",
          "K191678",
          "K982073",
          "K031379",
          "K013872",
          "K952421",
          "K946164",
          "K181173",
          "K935243",
          "K103253",
          "K980050",
          "K092684",
          "K933834",
          "K220069",
          "K952785",
          "K935645",
          "K240334",
          "K930180",
          "K931138",
          "K072912",
          "K000545",
          "K252957",
          "K991303",
          "K010007",
          "K983623",
          "K231066",
          "K945545",
          "K933456",
          "K946263",
          "K980623",
          "K922094",
          "K133345",
          "K962395",
          "K932631",
          "K951996",
          "K101077",
          "K961463",
          "K933909",
          "K100346",
          "K944201",
          "K150781",
          "K032676",
          "K080051",
          "K942678",
          "K022897",
          "K253994",
          "K912903",
          "K951173",
          "K930463",
          "K940431",
          "K241502",
          "K992353",
          "K180836",
          "K960903",
          "K180208",
          "K940988",
          "K770348",
          "K200902",
          "K243867",
          "K953481",
          "K160230",
          "K260185",
          "K151822",
          "K940688",
          "K931111",
          "K914997",
          "K911105",
          "K925267",
          "K945751",
          "K980279",
          "K012884",
          "K950278",
          "K941250",
          "K162445",
          "K934025",
          "K935754",
          "K931272",
          "K926110",
          "K252697",
          "K950748",
          "K162387",
          "K930809",
          "K183600",
          "K936279",
          "K913954",
          "K242606",
          "K915360",
          "K974382",
          "K172794",
          "K930511",
          "K170899",
          "K150377",
          "K070158",
          "K942879",
          "K770712",
          "K063457",
          "K940241",
          "K033359",
          "K952906",
          "K940428",
          "K911489",
          "K962104",
          "K942415",
          "K950021",
          "K923055",
          "K161792",
          "K933416",
          "K150127",
          "K101733",
          "K230886",
          "K141593",
          "K032093",
          "K933985",
          "K962020",
          "K050209",
          "K925959",
          "K042834",
          "K971475",
          "K914549",
          "K932022",
          "K193272",
          "K951715",
          "K934153",
          "K934097",
          "K040604",
          "K112358",
          "K060085",
          "K913991",
          "K934916",
          "K192300",
          "K941316",
          "K042069",
          "K012150",
          "K150549",
          "K934937",
          "K912750",
          "K240196",
          "K934076",
          "K924592",
          "K151489",
          "K950277",
          "K932805",
          "K942641",
          "K090979",
          "K933497",
          "K951444",
          "K933353",
          "K944103",
          "K920403",
          "K982013",
          "K925932",
          "K980599",
          "K983363",
          "K973814",
          "K953086",
          "K024251",
          "K162059",
          "K230359",
          "K950503",
          "K924761",
          "K912023",
          "K914084",
          "K954072",
          "K946213",
          "K933520",
          "K944465",
          "K192172",
          "K935442",
          "K182080",
          "K936102",
          "K953611",
          "K963323",
          "K982287",
          "K973139",
          "K190592",
          "K914967",
          "K050650",
          "K112456",
          "K051585",
          "K021290",
          "K941092",
          "K150569",
          "K925624",
          "K222011",
          "K001666",
          "K933897",
          "K231622",
          "K991635",
          "K931928",
          "K014047",
          "K932804",
          "K914470",
          "K112621",
          "K252442",
          "K003431",
          "K973415",
          "K925917",
          "K950198",
          "K941693",
          "K930427",
          "K912159",
          "K940790",
          "K250927",
          "K963760",
          "K912862",
          "K954175",
          "K922571",
          "K000493",
          "K240205",
          "K924011",
          "K935086",
          "K981928",
          "K926480",
          "K943502",
          "K943070",
          "K013611",
          "K080155",
          "K970754",
          "K950063",
          "K213860",
          "K951398",
          "K952899",
          "K940439",
          "K082233",
          "K955076",
          "K152513",
          "K932617",
          "K931262",
          "K931281",
          "K983293",
          "K000769",
          "K925599",
          "K931273",
          "K200517",
          "K950378",
          "K030174",
          "K934155",
          "K023627",
          "K955845",
          "K950745",
          "K231042",
          "K914922",
          "K990724",
          "K102219",
          "K031882",
          "K944309",
          "K925860",
          "K943489",
          "K950562",
          "K251490",
          "K172578",
          "K922608",
          "K250820",
          "K942584",
          "K924048",
          "K003790",
          "K041967",
          "K912933",
          "K942597",
          "K171344",
          "K974366",
          "K051732",
          "K093279",
          "K180114",
          "K923820",
          "K120608",
          "K082156",
          "K911493",
          "K946171",
          "K926079",
          "K151832",
          "K934547",
          "K161838",
          "K011360",
          "K992182",
          "K053311",
          "K091211",
          "K212214",
          "K940141",
          "K942612",
          "K982417",
          "K942817",
          "K003803",
          "K211577",
          "K983279",
          "K943554",
          "K940587",
          "K973436",
          "K943927",
          "K963815",
          "K944870",
          "K983567",
          "K973796",
          "K010870",
          "K862611",
          "K043284",
          "K090902",
          "K252010",
          "K151548",
          "K931018",
          "K914203",
          "K010789",
          "K942819",
          "K120611",
          "K962702",
          "K253978",
          "K921709",
          "K914781",
          "K915298",
          "K142464",
          "K122511",
          "K991382",
          "K952965",
          "K974391",
          "K980604",
          "K260089",
          "K123467",
          "K063859",
          "K973359",
          "K070266",
          "K100195",
          "K944304",
          "K161228",
          "K926104",
          "K260318",
          "K000423",
          "K946089",
          "K934807",
          "K191866",
          "K952980",
          "K930965",
          "K951733",
          "K173919",
          "K153126",
          "K945023",
          "K212736"
        ],
        "registration_number": [
          "",
          "3010860245",
          "3010432890",
          "9610617",
          "2210968",
          "3015336263",
          "3005528784",
          "1058584",
          "3014663814",
          "2029275",
          "1416980",
          "3025482817",
          "3006990600",
          "3017820152",
          "3015183635",
          "3008439467",
          "3014202166",
          "8043396",
          "3005110411",
          "2124215",
          "1047843",
          "1123010",
          "2518417",
          "3009881188",
          "3016678045",
          "3013184361",
          "3015540957",
          "3011052945",
          "3015532525",
          "3007593944",
          "3006161756",
          "9611112",
          "3033043422",
          "3009212837",
          "3038188839",
          "9710058",
          "3003687489",
          "1643264",
          "3005448067",
          "2011171",
          "9612501",
          "3019269298",
          "3012627460",
          "9616067",
          "3045857302",
          "3043226252",
          "1017294",
          "3045816907",
          "3021875817",
          "3002579136",
          "9611102",
          "3013497507",
          "3004044817",
          "3015539532",
          "3009350690",
          "3011416394",
          "8010349",
          "8010935",
          "9616088",
          "3008868758",
          "3007055279",
          "8010257",
          "3012880598",
          "3004608878",
          "3013011598",
          "3004491689",
          "3008191256",
          "9611665",
          "3010452421",
          "3012329552",
          "3010701807",
          "8010456",
          "3027339877",
          "3008729892",
          "3009889060",
          "3006546082",
          "1222773",
          "3002907620",
          "3021336392",
          "9613793",
          "3007345416",
          "3008132398",
          "3003627840",
          "3014334038",
          "1836161",
          "3014004349",
          "3022626486",
          "3008386005",
          "3008770252",
          "3003713161",
          "8022210",
          "1650907",
          "9612075",
          "3030447506",
          "3010437008",
          "3011373951",
          "1032347",
          "1055236",
          "1420032",
          "3005465238",
          "9612086",
          "3009495876",
          "3013445147",
          "3018907196",
          "1317547",
          "1643817",
          "3027242374",
          "3015974593",
          "3010547268",
          "8010372",
          "3007766601",
          "2244478",
          "3010177712",
          "1416891",
          "3016112537",
          "3010564249",
          "3013176080",
          "2530138",
          "3004542960",
          "2916714",
          "3009328501",
          "3005254598",
          "3003915875",
          "3020704423",
          "3003903031",
          "1216677",
          "3026606306",
          "3011137372",
          "3014615697",
          "1222928",
          "3014279513",
          "3015219237",
          "3021140368",
          "3015453963",
          "9612278",
          "1060680",
          "3010707607",
          "3006116468",
          "3009217531",
          "3003418325",
          "1531174",
          "3042940623",
          "9611827",
          "1057358",
          "2434982",
          "3011024991",
          "3004091527",
          "3003604053",
          "3003755939",
          "8010370",
          "1423662",
          "3013247477",
          "2027111",
          "8010026",
          "3010684280",
          "3008114965",
          "3042953929",
          "3009763019",
          "3013868968",
          "3009337690",
          "3016668451",
          "3017196117",
          "2650143",
          "3012883081",
          "3020746799",
          "3004215117",
          "2242352",
          "3011063460",
          "2649614",
          "1220246",
          "2936485",
          "3011277660",
          "3011802799",
          "2135348",
          "2032112",
          "1222997",
          "3008595179",
          "1048735",
          "1056128",
          "3017102971",
          "9615782",
          "3014461810",
          "9611612",
          "3004434200",
          "1220477",
          "1018470",
          "3008309062",
          "3014991167",
          "3012453745",
          "3004122598",
          "3013767421",
          "1421101",
          "3010399422",
          "1219930",
          "3014611142",
          "3013665681",
          "3005369776",
          "3030733800",
          "3001003398",
          "3012494290",
          "3003435550",
          "3011401113",
          "8043746",
          "1320468",
          "9680721",
          "3011221476",
          "3013002341",
          "3001614464",
          "3032391",
          "2031917",
          "3011683674",
          "3042225166",
          "3031240334",
          "3015895045",
          "3010305275",
          "2133810",
          "3033902430",
          "3011986317",
          "3009420598",
          "8010273",
          "1450997",
          "3030926959",
          "2085947",
          "3007111496",
          "9611594",
          "2529647",
          "3015173212",
          "3012998149",
          "3038187466",
          "2134944",
          "3009161688",
          "3007303113",
          "1820334",
          "3008837339",
          "8030607",
          "8044098",
          "9614969",
          "3030153385",
          "1018233",
          "3001722928",
          "3008805497",
          "3016727307",
          "3003775006",
          "2246552",
          "3011782009",
          "8010168",
          "1219602",
          "2022435",
          "1928237",
          "3012460907",
          "2027009",
          "3015612252",
          "3015335150",
          "3004361445",
          "9710649",
          "3002806603",
          "8010591",
          "9681260",
          "3015241782",
          "1221826",
          "3007185043",
          "3017262662",
          "3014579161",
          "3001675293",
          "3015151147",
          "3010155673",
          "3043152784",
          "3010612620",
          "2029015",
          "3015224777",
          "1226146",
          "3010590422",
          "3005099803",
          "3002807315",
          "3006897899",
          "3009057691",
          "8043971",
          "1417592",
          "3014479313",
          "3008044552",
          "3010419931",
          "3011803381",
          "9680515",
          "3009231402",
          "2025816",
          "8040233",
          "3020468145",
          "3010157088",
          "3016096100",
          "3010601984",
          "3014296178",
          "1825034",
          "3010202439",
          "8043903",
          "3011936016",
          "3007305485",
          "3006581049",
          "3006498370",
          "3031594238",
          "3010155648",
          "3005636544",
          "9680837",
          "3010493171",
          "1000121056",
          "3012470322",
          "3025306172",
          "3010055973",
          "1211566",
          "3003898360",
          "3011989923",
          "9710524",
          "3018984",
          "9614641",
          "2523676",
          "3021226419",
          "3012481535",
          "2320767",
          "3003790304",
          "3016804805",
          "1221072",
          "3015231789",
          "9610773",
          "3001556604",
          "3020142503",
          "3025598220",
          "3015895145",
          "3006554912",
          "9681129",
          "3003407244",
          "3010041511",
          "1450662",
          "1061927",
          "2032521",
          "1721983",
          "9680952",
          "2020550",
          "3017636737",
          "3010009693",
          "3014579024",
          "3030839824",
          "1213643",
          "3009051888",
          "3004598644",
          "1526534",
          "1221934",
          "3010703925",
          "3009719225",
          "3017210488",
          "3015489752",
          "1422634",
          "3004135191",
          "1519132",
          "3013155768",
          "3011270181",
          "3005579585",
          "2126670",
          "1423537",
          "3010380518",
          "3013023969",
          "3007695959",
          "9614209",
          "3008571754",
          "3016832577",
          "3007703432",
          "3045813414",
          "3015775837",
          "3014635169",
          "3005809810",
          "1061124",
          "9611503",
          "1124841",
          "3014644276",
          "3035708926",
          "1223422",
          "3013188547",
          "3008383115",
          "3011230048",
          "1718850",
          "3010806653",
          "2032098",
          "3003139373",
          "2954921",
          "3012995405",
          "9612030",
          "3007791599",
          "3016245399",
          "3010043098",
          "3006082230",
          "3024021261",
          "8044057",
          "3012345110",
          "9515015",
          "1646747",
          "2183744",
          "3013501671",
          "3009134855",
          "3014273609",
          "3017619652",
          "2219920",
          "2030624",
          "1054241",
          "1123137",
          "1037955",
          "3005012805",
          "8010047",
          "3010885402",
          "2529846",
          "3009380063",
          "3014263632",
          "3016248634",
          "9610595",
          "3015153469",
          "3020331900",
          "3024137003",
          "3006789154",
          "3004526608",
          "1222274",
          "1219659",
          "3005595283",
          "3008911114",
          "3010806467",
          "3010370500",
          "3036773433",
          "3007487007",
          "3011543740",
          "3013476721",
          "1825146",
          "3008864603",
          "3004378299",
          "3015806723",
          "3008202779",
          "2183319",
          "9610612",
          "9611274",
          "1218882",
          "3015497424",
          "3016607092",
          "3005699784",
          "3011050570",
          "1720747",
          "8010422",
          "2518897",
          "3013700547",
          "3015542154",
          "2028523",
          "1828132",
          "3021355451",
          "3021744308",
          "3002807314",
          "3014769442",
          "3007137643",
          "3007421149",
          "2249651",
          "2647580",
          "3004644065",
          "2183425",
          "3030311186",
          "3010160527",
          "3008843439",
          "1526012",
          "2424366",
          "3044828322",
          "3010155661",
          "3016851379",
          "9680001",
          "3018094310",
          "1000181154",
          "3004142400",
          "9681147",
          "2938071",
          "3021393544",
          "2648727",
          "1056350",
          "3004111573",
          "3043225236",
          "3003724334",
          "3021744379",
          "3007791595",
          "8010697",
          "1423395",
          "3007738736",
          "1527736",
          "3005075853",
          "3014342096",
          "1649518",
          "9612074",
          "9611281",
          "3010120217",
          "3016555232",
          "3021336182",
          "3010291427",
          "3007516837",
          "1320894",
          "3008573567",
          "1220351",
          "1921846",
          "8010877",
          "3003935342",
          "3020703212",
          "9613188",
          "1225984",
          "9613846",
          "1054811",
          "8040510",
          "1225798",
          "2434839",
          "3023852420",
          "3021632375",
          "3030504644",
          "1528319",
          "3002808148",
          "3010400865",
          "8043548",
          "1043214",
          "3009756153",
          "3000237660",
          "8010871",
          "3004464325",
          "3010220244",
          "3002259632",
          "3029968414",
          "9680718",
          "9681834",
          "3031238646",
          "3017119401",
          "9614434",
          "3010331645",
          "3004530184",
          "3014848720",
          "9681191",
          "3013557562",
          "3014262693",
          "1923569",
          "1412854",
          "3009039068",
          "3006638824",
          "3008040391",
          "3013283620",
          "3016527960",
          "2245304",
          "3013077751",
          "3004605321",
          "8010298",
          "9614945",
          "3010400367",
          "8010510",
          "8010487",
          "1421879",
          "3029950834",
          "3018126215",
          "9616567",
          "3004485166",
          "3016667340",
          "3027500817",
          "9616517",
          "3003542872",
          "3012322232",
          "3012818475",
          "1319639",
          "3017390965",
          "3009453258",
          "3012507533",
          "3035366890",
          "3011459664",
          "3010273872",
          "3013944123",
          "2320762",
          "1418479",
          "3030626857",
          "2528981",
          "2431166",
          "8010418",
          "3015207155",
          "3005441393",
          "3020899789",
          "3010605922",
          "8043980"
        ],
        "fei_number": [
          "",
          "3003631551",
          "3010860245",
          "3010432890",
          "4417",
          "3003221505",
          "1000138054",
          "2210968",
          "3002902735",
          "3005747797",
          "3015336263",
          "3005528784",
          "3014663814",
          "3003301977",
          "1416980",
          "3025482817",
          "3006990600",
          "3017820152",
          "3015183635",
          "3008439467",
          "3014202166",
          "3005110411",
          "2124215",
          "1047843",
          "3004028763",
          "3000219468",
          "2518417",
          "3002808327",
          "3009881188",
          "3016678045",
          "3013184361",
          "3015540957",
          "3011052945",
          "3015532525",
          "3007593944",
          "3033043422",
          "3009212837",
          "3038188839",
          "3003687489",
          "1643264",
          "1000137656",
          "3005448067",
          "2011171",
          "1000121298",
          "3002807040",
          "2523865",
          "3019269298",
          "3012627460",
          "3045857302",
          "3043226252",
          "1017294",
          "3045816907",
          "3021875817",
          "3002579136",
          "3013497507",
          "3004044817",
          "3015539532",
          "3002806565",
          "3009350690",
          "3011416394",
          "3008868758",
          "3002808273",
          "3007055279",
          "3012880598",
          "3003944071",
          "3004608878",
          "3006149253",
          "3013011598",
          "3004491689",
          "3008191256",
          "3001595430",
          "3002681132",
          "3007130719",
          "3010452421",
          "3008207730",
          "3012329552",
          "3010701807",
          "3003495548",
          "3027339877",
          "1000121057",
          "3003565435",
          "3002808268",
          "3009889060",
          "3003117688",
          "3006546082",
          "1000114185",
          "3007475226",
          "3001436772",
          "3002907620",
          "3021336392",
          "3007345416",
          "3003627840",
          "3014334038",
          "3014004349",
          "1000512168",
          "3010666513",
          "3022626486",
          "3008386005",
          "3008770252",
          "3003713161",
          "3030447506",
          "3010437008",
          "1000221205",
          "3011373951",
          "1032347",
          "1420032",
          "3005465238",
          "3009495876",
          "3013445147",
          "3018907196",
          "1317547",
          "1643817",
          "3027242374",
          "3015974593",
          "3010547268",
          "1000122656",
          "3006469671",
          "3932",
          "3007766601",
          "3010177712",
          "1416891",
          "1000244733",
          "3016112537",
          "3010564249",
          "3013176080",
          "3004542960",
          "2916714",
          "3009328501",
          "3005254598",
          "3003915875",
          "3007573441",
          "3020704423",
          "3003903031",
          "1216677",
          "3026606306",
          "3002807555",
          "3002808001",
          "3014615697",
          "3002806436",
          "3014279513",
          "3015219237",
          "3021140368",
          "3002807795",
          "3015453963",
          "3005649458",
          "1000219637",
          "3010707607",
          "3006116468",
          "3001236905",
          "3000204839",
          "3009217531",
          "3003418325",
          "3042940623",
          "2434982",
          "1000120768",
          "3011024991",
          "3004091527",
          "3002477156",
          "3003604053",
          "3003755939",
          "1000491840",
          "3013247477",
          "4498",
          "2027111",
          "1000136191",
          "1000514656",
          "3010684280",
          "3008114965",
          "3042953929",
          "3009763019",
          "3013868968",
          "3009337690",
          "3016668451",
          "3017196117",
          "2242352",
          "3012883081",
          "3020746799",
          "3004215117",
          "3011063460",
          "3942",
          "1000080301",
          "2649614",
          "3002807847",
          "1220246",
          "3002516162",
          "3005461046",
          "2936485",
          "3011277660",
          "3011802799",
          "3008595179",
          "3008782585",
          "1048735",
          "3003499455",
          "3017102971",
          "3004058281",
          "3001319476",
          "3014461810",
          "3004434200",
          "3002807798",
          "1220477",
          "4430",
          "1000120548",
          "1018470",
          "3008309062",
          "3014991167",
          "3012453745",
          "3004122598",
          "3013767421",
          "1421101",
          "3010399422",
          "1219930",
          "3014611142",
          "3013665681",
          "1000160458",
          "3005369776",
          "3030733800",
          "3001003398",
          "3012494290",
          "3003435550",
          "3011401113",
          "3011221476",
          "3013002341",
          "3001614464",
          "3002806862",
          "3011683674",
          "3042225166",
          "3031240334",
          "3002963732",
          "3015895045",
          "3002808088",
          "3010305275",
          "3033902430",
          "3011986317",
          "3009420598",
          "3000215273",
          "1450997",
          "3030926959",
          "3007111496",
          "3002807769",
          "3015173212",
          "3002806934",
          "3012998149",
          "3038187466",
          "3009161688",
          "3007303113",
          "3002806575",
          "1820334",
          "3008837339",
          "3003995201",
          "3030153385",
          "1018233",
          "3001722928",
          "3008805497",
          "3016727307",
          "3003775006",
          "1000160162",
          "3002808043",
          "2246552",
          "1219602",
          "2022435",
          "1928237",
          "3012460907",
          "2027009",
          "3015612252",
          "3015335150",
          "3004361445",
          "1000123180",
          "3002806725",
          "3002806603",
          "3002808445",
          "3015241782",
          "1221826",
          "3007185043",
          "3017262662",
          "1000132435",
          "3014579161",
          "3000226659",
          "3001451571",
          "3001675293",
          "3015151147",
          "3010155673",
          "3043152784",
          "3010612620",
          "3015224777",
          "3010590422",
          "3005099803",
          "3002807315",
          "3006897899",
          "1000208624",
          "3009057691",
          "1417592",
          "3014479313",
          "3008044552",
          "3010419931",
          "3011803381",
          "1000499348",
          "3009231402",
          "2025816",
          "3002809480",
          "3020468145",
          "3004455125",
          "3010157088",
          "3016096100",
          "3010601984",
          "3004207793",
          "3014296178",
          "1825034",
          "1000120613",
          "3011936016",
          "3000206435",
          "3007305485",
          "3003056094",
          "3006581049",
          "3006498370",
          "3031594238",
          "1000064530",
          "3002929183",
          "3010155648",
          "3005636544",
          "3000206455",
          "1000453845",
          "3000280067",
          "3010493171",
          "1000121056",
          "3012470322",
          "3025306172",
          "3010055973",
          "3002680121",
          "1211566",
          "3003898360",
          "3011989923",
          "3006509795",
          "3018984",
          "2523676",
          "3021226419",
          "3012481535",
          "3003790304",
          "3002807395",
          "3016804805",
          "1221072",
          "3015231789",
          "3001556604",
          "3020142503",
          "3025598220",
          "3015895145",
          "3006554912",
          "3003407244",
          "3010041511",
          "1450662",
          "3003058448",
          "2020550",
          "3017636737",
          "3010009693",
          "3014579024",
          "3030839824",
          "1213643",
          "3009051888",
          "3004598644",
          "1526534",
          "3002993297",
          "1221934",
          "3010703925",
          "3009719225",
          "3017210488",
          "3015489752",
          "3004135191",
          "1519132",
          "3013155768",
          "3011270181",
          "3005579585",
          "3003445992",
          "2126670",
          "3010380518",
          "3013023969",
          "3007695959",
          "3008571754",
          "3002807091",
          "1000125147",
          "3016832577",
          "3008238512",
          "3007703432",
          "3045813414",
          "3015775837",
          "3014635169",
          "3005809810",
          "3035708926",
          "3000717925",
          "3013188547",
          "3008383115",
          "3002771993",
          "3011230048",
          "1718850",
          "3007290840",
          "3010806653",
          "1000140363",
          "3003139373",
          "3002808173",
          "3012995405",
          "3010043098",
          "3007791599",
          "3016245399",
          "3006082230",
          "3024021261",
          "3012345110",
          "2183744",
          "3013501671",
          "3009134855",
          "3002808127",
          "3014273609",
          "3017619652",
          "2219920",
          "1037955",
          "3005012805",
          "3010885402",
          "1000272060",
          "3009380063",
          "3014263632",
          "3016248634",
          "3015153469",
          "3020331900",
          "3024137003",
          "3006789154",
          "3004526608",
          "1219659",
          "3005595283",
          "1000115048",
          "3008911114",
          "3010806467",
          "3010370500",
          "3036773433",
          "3007487007",
          "3011543740",
          "3013476721",
          "1825146",
          "3008864603",
          "3004378299",
          "3015806723",
          "3008202779",
          "2183319",
          "1218882",
          "3015497424",
          "3016607092",
          "3003040625",
          "2648209",
          "3005699784",
          "2828",
          "3011050570",
          "1720747",
          "3013700547",
          "3015542154",
          "3004595684",
          "3003448211",
          "3002806315",
          "1828132",
          "3021355451",
          "3003276727",
          "3021744308",
          "3002807314",
          "3014769442",
          "3007137643",
          "3007421149",
          "2647580",
          "3004644065",
          "2183425",
          "3030311186",
          "3008843439",
          "1526012",
          "3003076831",
          "2424366",
          "3044828322",
          "1000208608",
          "3010155661",
          "3016851379",
          "3001236277",
          "3002750084",
          "3018094310",
          "1000181154",
          "3004142400",
          "2938071",
          "3021393544",
          "2648727",
          "1000306431",
          "3003249645",
          "3004111573",
          "3043225236",
          "1000136269",
          "3003724334",
          "3021744379",
          "3007791595",
          "3005075853",
          "3007738736",
          "1527736",
          "3014342096",
          "3002964398",
          "3002807095",
          "3010120217",
          "3016555232",
          "3021336182",
          "3010291427",
          "3007516837",
          "3008573567",
          "1220351",
          "1921846",
          "3003935342",
          "3020703212",
          "3002807421",
          "3003574398",
          "3005969827",
          "2434839",
          "3023852420",
          "3021632375",
          "3002808335",
          "4486",
          "3030504644",
          "3002808148",
          "1000123297",
          "3010400865",
          "1043214",
          "1000112968",
          "3009756153",
          "3002807072",
          "3000237660",
          "3000204191",
          "3004464325",
          "3010220244",
          "3002259632",
          "3029968414",
          "3031238646",
          "3017119401",
          "3010331645",
          "3004530184",
          "3014848720",
          "3013557562",
          "3014262693",
          "1923569",
          "1412854",
          "3009039068",
          "3006638824",
          "3008040391",
          "3000203000",
          "3013283620",
          "3005276402",
          "3016527960",
          "2245304",
          "3013077751",
          "3000203066",
          "3004605321",
          "1000112877",
          "3010400367",
          "6166",
          "3002914049",
          "3002299608",
          "3029950834",
          "3018126215",
          "3004485166",
          "3016667340",
          "3027500817",
          "3004084693",
          "3003542872",
          "3012322232",
          "3012818475",
          "3002807646",
          "3017390965",
          "3009453258",
          "3012507533",
          "3035366890",
          "3011459664",
          "3010273872",
          "3004053982",
          "3013944123",
          "1418479",
          "3030626857",
          "2528981",
          "3015207155",
          "3005441393",
          "3020899789",
          "3010605922"
        ],
        "device_name": "Laparoscope, General & Plastic Surgery",
        "medical_specialty_description": "Gastroenterology, Urology",
        "regulation_number": "876.1500",
        "device_class": "2"
      }
    }
  ]
}