      "cfres_id": "43173",
      "product_res_number": "Z-0354-06",
      "event_date_initiated": "2005-11-18",
      "event_date_posted": "2006-01-06",
      "recall_status": "Terminated",
      "event_date_terminated": "2008-04-22",
      "res_event_number": "34129",
      "product_code": "CBK",
      "k_numbers": [
        "K022750"
      ],
      "product_description": "The PL V Continuum ventilator is a microprocessor controlled, compressorbased, mechanical ventilator. It is intended to control or assist breathing by delivering room air to the patient. PL V Continuum utilizes an internal compressor to generate compressed air for delivery to the patient. Breath delivery is controlled by software algorithms. The user interface on PL V Continuum has a membrane keypad with indicator Light Emitting Diodes (LED) for the selection and acceptance of patient settings and for the display of alarm conditions. PL V Continuum is capable of providing the following types of ventilatory support:\n\n. Positive Pressure Ventilation, delivered either invasively (via endotracheal or tracheostomy tube) or non-invasively (via mask or mouthpiece).\n. Assist/Control, Spontaneous Intermittent Mandatory Ventilation (SIMV) or Continuous Positive Airway Pressure (CP AP) modes of ventilation.\n. Volume-Controlled (VC). Available in AlC and SIMV.\n. Pressure-Controlled (PC). Available in AlC and SIMV.\n. Pressure Support (PS). Available in SIMV and SPONT.",
      "code_info": "Domestic Distribution Serial Numbers: VS4700581, VS4700661,  VS4700631,  VS4702578,  VS4702507,  VS4702551,  VS4702583, VS4702589,  VS4702590, VS470259], VS4702574, VS4702580, VS4702566, VS4700824, VS4702593, VS4700835, VS4700536, 1VS4700539, VS4700540, VS4700543, VS4700544, S4700546, VS4700548, VS4700549, VS4700550, VS4700552,VS4700554, VS4700555, VS4700558, VS4700565, VS4700566,VS4700567, VS4700569, VS4700573, VS4700574, VS4700577,VS4700578, VS4700580, VS4700581, VS4700585, VS4700588VS4700590, VS4700593, VS4700594, VS4700595, VS4700596, VS4700605, VS4700607, VS4700608, VS4700613, VS4700614, VS4700615, VS4700617, VS4700620, VS4700622, VS4700624, VS4700625, VS4700632, VS4700635, VS4700637, VS4700642, VS4700643, VS4700646, VS4700647, VS4700649, VS4700650, VS4700653, VS4700660, VS4700661, VS4700663, VS4700668, VS4700670, VS4700673, VS4700681, VS4700687, VS4700693, VS4700694, VS4700698, VS4700699, VS4700708, VS470071O, VS4700711, VS4700718, VS4700719, VS4700721, VS4700724, VS4700725, VS4700729, VS4700730, VS4700736, VS4700738, VS4700739, VS4700740, VS4700743, VS4700744, VS4700750, VS4700751, VS4700754, VS4700760, VS4700761, VS4700762, VS4700767, VS4700770, VS4700772, VS4700773, VS4700774, VS4700775, VS4700776, VS4700777, VS4700780, VS4700781, VS4700782, VS4700783, VS4700785, VS4700786, VS4700789, VS4700790, VS4700791, VS4700796, VS4700797, VS4700798, VS4700802, VS4700804, VS4700808, VS4700809, VS4700811, VS4700813, VS4700814, VS4700816, VS4700817, VS4700818, VS4700819, VS4700821, VS4700822, VS4700823, VS4700827, VS4700828, VS4700829, VS4700832, VS4700833, VS4700834, VS4700836, VS4700837, VS4700838, VS4700840, VS4700841, VS4700842, VS4700843, VS4700844, VS4700846, VS4700847, VS4700849, VS4700850, VS4700852, VS4700853, VS4700854, VS4700855, VS4700856, VS4702501, VS4702502, VS4702505, VS4702512, VS4702516, VS4702519, VS4702523, VS4702524, VS4702525, VS4702528, VS4702531, VS4702532, VS4702535, VS4702543, VS4702544, VS4702545, VS4702548, VS4702549, VS4702552, VS4702553, VS4702555, VS4702556, VS4702557, VS4702567, VS4702568, VS4702570, VS4702571, VS4702572, VS4702573, VS4702575, VS4702586, VS4702594, VS4702599, VS4702600, VS4702604, VS4702612, VS4702613, VS4702623, VS4702636, VS4702647, VS4702650, VS4702670, VS4702677, VS4702688, VS4702689, VS4702709, VS4702734, VS4702747, VS4702748, VS4702750, VS4702776, VS4702784, VS4702800, VS4702805, VS4702809, VS4702811, VS4702813, VS4702825, VS4702835, VS4702848, VS4702854, VS4702856, VS4702857, VS4702859, VS4702883, VS4702885, VS4702981, VS4703079, VS4703103  International Distribution Serial Numbers:  VS4702605, VS4702608, VS4702609, VS4702631, VS4702632, VS4702634, VS4702642, VS4702643, VS4702659, VS4702719, VS4702754, VS4702759, VS4700792, VS4702722, VS4700563, VS4700696, VS4700820, VS4700601, VS4702645, VS4700839, VS4700644, VS4700589, VS4700703, VS4700723, VS4700746, VS4700553, VS4700628, VS4702962, VS4702983, VS4702995, VS4703006, VS4703023, VS4703041, VS4703056, VS4703064, VS4700717, VS4700645, VS4702915, VS4702500, VS4702828, VS4702858",
      "firm_fei_number": "2518422",
      "recalling_firm": "Respironics California Inc",
      "address_1": "2271 Cosmos Ct",
      "city": "Carlsbad",
      "state": "CA",
      "postal_code": "92009-1517",
      "additional_info_contact": "Anil Bhalani\n760-918-7458",
      "reason_for_recall": "During product testing of the PLV Continuum Ventilator's Safety Valve/Pressure Relief Valve (SV/PRV) at our manufacturing facility, Respironics found that when the PLY Continuum Ventilator is operated in the 'Face Up' orientation, the SV /PRV may stick in the open position, which may result in loss of delivered gas volume to the patient.",
      "root_cause_description": "Other",
      "action": "On November 18, 2005 all affected customers received a personal phone call from Respironics explaining the recall situation in detail.  On November 21, 2005, an advisory notice was sent to customers via fax, email, or mail, instructing customers to only use your PLV Continuum Ventilator in the 'Face Foward' orientation.  \n\nThe firm  will contact customers with further information shortly to assure that this situation is positively resolved.\n\nCustomers were instructed to Only use the PL V Continuum Ventilator in the 'Face Forward' orientation.\n\nIn the event that a customer requires the application of the PLV Continuum Ventilator in the Face Up Orientation, Respironics will provide the customer, at no charge, with a rental PLV-l00, PLV-IO2b or PLV-I02 Ventilator for the customer to use until their  PLV Continuum Ventilator is upgraded.\n\nIf additional information is needed the customer was instructed to contact Respironics'' Customer Service at 800 345-6443.",
      "product_quantity": "268",
      "distribution_pattern": "Nationwide, Taiwan, Saudi Arabia, Netherlands, Hong Kong, Canada, Brazil, Australia, and Argentina.",
      "openfda": {
        "k_number": [
          "K922705",
          "K822456",
          "K813253",
          "K051767",
          "K841757",
          "K060549",
          "K222024",
          "K802583",
          "K993449",
          "K002881",
          "K802051",
          "K894322",
          "K792455",
          "K801890",
          "K112754",
          "K051895",
          "K062793",
          "K120639",
          "K123149",
          "K103803",
          "K020746",
          "K844179",
          "K923809",
          "K851509",
          "K023366",
          "K896735",
          "K040540",
          "K042468",
          "K853300",
          "K192307",
          "K010925",
          "K823861",
          "K896269",
          "K241520",
          "K862025",
          "K023350",
          "K854070",
          "K092648",
          "K822103",
          "K844590",
          "K896442",
          "K010793",
          "K895486",
          "K872113",
          "K864105",
          "K915312",
          "K800999",
          "K864129",
          "K911336",
          "K781159",
          "K180098",
          "K914424",
          "K062267",
          "K960010",
          "K161450",
          "K863228",
          "K083688",
          "K111116",
          "K140939",
          "K073694",
          "K862604",
          "K842362",
          "K153046",
          "K001646",
          "K903010",
          "K802118",
          "K021573",
          "K133868",
          "K861432",
          "K032451",
          "K895541",
          "K063404",
          "K041412",
          "K173973",
          "K864431",
          "K060705",
          "K120726",
          "K862712",
          "K982580",
          "K081842",
          "K181695",
          "K893622",
          "K053270",
          "K984535",
          "K062019",
          "K833982",
          "K071526",
          "K913423",
          "K841818",
          "K890921",
          "K923704",
          "K871433",
          "K970165",
          "K053388",
          "K050860",
          "K210384",
          "K772065",
          "K871448",
          "K861707",
          "K812980",
          "K924043",
          "K781926",
          "K981971",
          "K990429",
          "K032922",
          "K051262",
          "K870861",
          "K881420",
          "K812958",
          "K102120",
          "K203449",
          "K241943",
          "K802168",
          "K930017",
          "K061094",
          "K053502",
          "K103625",
          "K193228",
          "K992133",
          "K813254",
          "K072450",
          "K943531",
          "K812642",
          "K896381",
          "K823858",
          "K163127",
          "K842876",
          "K080181",
          "K926292",
          "K243767",
          "K874371",
          "K912533",
          "K103211",
          "K872482",
          "K820393",
          "K140575",
          "K181166",
          "K143035",
          "K102775",
          "K822233",
          "K871009",
          "K001686",
          "K062710",
          "K792564",
          "K112729",
          "K884389",
          "K073179",
          "K051222",
          "K790199",
          "K820248",
          "K860675",
          "K110795",
          "K900314",
          "K900637",
          "K871985",
          "K771039",
          "K953228",
          "K222207",
          "K972833",
          "K240807",
          "K854300",
          "K841393",
          "K901953",
          "K781160",
          "K192604",
          "K111473",
          "K822888",
          "K873549",
          "K123637",
          "K041775",
          "K822040",
          "K090888",
          "K101401",
          "K103318",
          "K041524",
          "K002686",
          "K844081",
          "K953238",
          "K120574",
          "K983219",
          "K093633",
          "K852771",
          "K023135",
          "K914485",
          "K894426",
          "K915488",
          "K902084",
          "K844912",
          "K791033",
          "K151901",
          "K072926",
          "K072588",
          "K970158",
          "K910847",
          "K241135",
          "K170037",
          "K072412",
          "K172875",
          "K040642",
          "K102054",
          "K911804",
          "K011957",
          "K874415",
          "K120670",
          "K082600",
          "K052863",
          "K213098",
          "K093886",
          "K911450",
          "K030341",
          "K030515",
          "K082966",
          "K202970",
          "K183364",
          "K862810",
          "K992788",
          "K902506",
          "K072590",
          "K223120",
          "K862212",
          "K970839",
          "K121886",
          "K791713",
          "K900465",
          "K081670",
          "K780188",
          "K871354",
          "K941309",
          "K922975",
          "K831877",
          "K151814",
          "K853760",
          "K961687",
          "K832700",
          "K010608",
          "K791188",
          "K984379",
          "K802087",
          "K781161",
          "K842122",
          "K021112",
          "K101850",
          "K920002",
          "K130171",
          "K062093",
          "K960861",
          "K900315",
          "K870275",
          "K173060",
          "K023289",
          "K964543",
          "K012691",
          "K914607",
          "K922102",
          "K900744",
          "K093632",
          "K012693",
          "K013642",
          "K792456",
          "K231778",
          "K111146",
          "K980208",
          "K832467",
          "K180295",
          "K141943",
          "K921821",
          "K972203",
          "K082351",
          "K860438",
          "K810893",
          "K841279",
          "K943412",
          "K061090",
          "K840119",
          "K880923",
          "K152068",
          "K811141",
          "K884521",
          "K981371",
          "K895513",
          "K921387",
          "K980642",
          "K030780",
          "K802270",
          "K790809",
          "K121379",
          "K821792",
          "K892886",
          "K010093",
          "K782074",
          "K120931",
          "K092264",
          "K071194",
          "K083050",
          "K063806",
          "K883809",
          "K180779",
          "K070594",
          "K913972",
          "K893786",
          "K982174",
          "K993167",
          "K072885",
          "K922477",
          "K872466",
          "K941308",
          "K090468",
          "K871792",
          "K201874",
          "K150638",
          "K003068",
          "K162832",
          "K873151",
          "K110083",
          "K082724",
          "K840865",
          "K812016",
          "K924930",
          "K821376",
          "K992670",
          "K092148",
          "K040790",
          "K220107",
          "K800125",
          "K920121",
          "K113743",
          "K001208",
          "K843737",
          "K121225",
          "K861488",
          "K943106",
          "K083122",
          "K101643",
          "K972248",
          "K093905",
          "K841529",
          "K151252",
          "K896380",
          "K083693",
          "K984577",
          "K990177",
          "K822113",
          "K901448",
          "K081837",
          "K122560",
          "K863546",
          "K111683",
          "K993071",
          "K131774",
          "K812961",
          "K101803",
          "K897113",
          "K830488",
          "K111610",
          "K960168",
          "K820746",
          "K873215",
          "K790733",
          "K880054",
          "K882764",
          "K921726",
          "K950853",
          "K920113",
          "K922654",
          "K984056",
          "K760663",
          "K081743",
          "K131252",
          "K022679",
          "K022132",
          "K950484",
          "K193056",
          "K201306",
          "K233486",
          "K012177",
          "K844350",
          "K121891",
          "K850374",
          "K043085",
          "K792459",
          "K884832",
          "K894316",
          "K063650",
          "K822835",
          "K100753",
          "K113051",
          "K923444",
          "K863155",
          "K981072",
          "K801697",
          "K883159",
          "K860798",
          "K871398",
          "K843424",
          "K864379",
          "K842688",
          "K955865",
          "K893043",
          "K051550",
          "K896736",
          "K022674",
          "K040221",
          "K981412",
          "K970460",
          "K122438",
          "K252173",
          "K820220",
          "K884520",
          "K081845",
          "K912619",
          "K820132",
          "K041223",
          "K201658",
          "K092135",
          "K003770",
          "K181170",
          "K880721",
          "K800114",
          "K253647",
          "K232365",
          "K024344",
          "K051263",
          "K982910",
          "K840345",
          "K830230",
          "K232836",
          "K120146",
          "K061627",
          "K790605",
          "K942938",
          "K981668",
          "K162738",
          "K243394",
          "K992608",
          "K142679",
          "K181216",
          "K930747",
          "K222644",
          "K800650",
          "K903089",
          "K952880",
          "K251888",
          "K884793",
          "K060647",
          "K111595",
          "K905272",
          "K083526",
          "K901785",
          "K881482",
          "K791186",
          "K093094",
          "K883865",
          "K091238",
          "K101129",
          "K881335",
          "K923992",
          "K842363",
          "K070899",
          "K190533",
          "K052554",
          "K922314",
          "K071212",
          "K983981",
          "K032226",
          "K050597",
          "K811102",
          "K252663",
          "K933262",
          "K251133",
          "K092263",
          "K060862",
          "K905244",
          "K010317",
          "K790732",
          "K900313",
          "K162877",
          "K081521",
          "K974176",
          "K112006",
          "K905235",
          "K092847",
          "K034040",
          "K130178",
          "K960964",
          "K833786",
          "K770132",
          "K821930",
          "K925262",
          "K982132",
          "K040574",
          "K902405",
          "K812953",
          "K073069",
          "K905234",
          "K832508",
          "K201082",
          "K002951",
          "K822635",
          "K902859",
          "K070513",
          "K900745",
          "K895779",
          "K093416",
          "K964540",
          "K853501"
        ],
        "registration_number": [
          "",
          "3010291427",
          "3030733800",
          "3007344957",
          "1000604079",
          "9610849",
          "3020283264",
          "3011190789",
          "3009077524",
          "3004428458",
          "8010939",
          "3018094310",
          "3033571078",
          "3005783425",
          "3003908598",
          "3038562397",
          "9617566",
          "3030149975",
          "3009302812",
          "8021084",
          "3010952312",
          "3008496839",
          "9610531",
          "9615827",
          "3010088563",
          "3007573469",
          "9611295",
          "1722070",
          "3016618143",
          "3032026475",
          "3007549157",
          "9681384",
          "2112667",
          "1220908",
          "3021607937",
          "3016450067",
          "3031369651",
          "3026740087",
          "3018783526",
          "3002797442",
          "3006135941",
          "3003697063",
          "3016723884",
          "3008102042",
          "3004050971",
          "2011171",
          "2085602",
          "2024500",
          "2221819",
          "3010817335",
          "1314417",
          "3004748541",
          "3010834934",
          "3029227324",
          "1066270",
          "8030673",
          "3010317211",
          "1316463",
          "2025816",
          "1222088",
          "3014631252",
          "9611451",
          "3006783791",
          "3003135857",
          "3000126629",
          "3007238566",
          "3011120183",
          "3001400310",
          "9680661",
          "3008363989",
          "3006807694",
          "3014766741",
          "3009107628",
          "9680215",
          "3044156918",
          "3025936435",
          "8030941",
          "1218058",
          "1922553",
          "1924066",
          "3001104093",
          "3033526676",
          "3029922922",
          "3001421318",
          "2112020",
          "2020676",
          "3030446844",
          "3012667060",
          "3009156722",
          "3029906224",
          "3010882708",
          "3006446479",
          "2518436",
          "1219324",
          "8010042",
          "3006087789",
          "3013154126",
          "3008264065",
          "3011205710",
          "9611500",
          "3017520736",
          "3004604967",
          "3021947238",
          "3029845",
          "3029871637",
          "1625392",
          "8020893",
          "3003862188",
          "2518422",
          "3021244152",
          "1225492",
          "3013095415",
          "3010157426",
          "1218704",
          "2134715",
          "3023272766",
          "3009609218",
          "9710602",
          "3003477135",
          "3009129579",
          "3000143502",
          "3007603826",
          "1000122786",
          "3010017366",
          "3011213495",
          "2021710",
          "3008867836",
          "3009490946",
          "3038259592",
          "3006240053"
        ],
        "fei_number": [
          "",
          "3010291427",
          "3030733800",
          "3007344957",
          "1000604079",
          "3020283264",
          "3011190789",
          "3009077524",
          "3004428458",
          "3002807907",
          "3018094310",
          "3033571078",
          "3005783425",
          "3003908598",
          "3038562397",
          "3030149975",
          "3009302812",
          "3010952312",
          "3002807626",
          "3008496839",
          "3010088563",
          "3004213967",
          "3003952420",
          "3016618143",
          "3032026475",
          "3007549157",
          "3002807992",
          "1000524541",
          "2112667",
          "1220908",
          "3021607937",
          "3016450067",
          "3031369651",
          "3026740087",
          "3002807850",
          "3018783526",
          "3003244771",
          "3002797442",
          "3006135941",
          "3003697063",
          "3016723884",
          "1000058335",
          "3004050971",
          "2011171",
          "2024500",
          "2221819",
          "3010817335",
          "1314417",
          "2995",
          "3004748541",
          "3010834934",
          "3029227324",
          "3010317211",
          "3003513129",
          "1316463",
          "2025816",
          "3001237338",
          "1222088",
          "3000210616",
          "3003135857",
          "3006783791",
          "3014631252",
          "3000126629",
          "3007238566",
          "3011120183",
          "3001400310",
          "3008363989",
          "3006807694",
          "3014766741",
          "3001084824",
          "3002807391",
          "1000117260",
          "3009107628",
          "3044156918",
          "3025936435",
          "1218058",
          "1922553",
          "1924066",
          "3002924544",
          "3001104093",
          "3033526676",
          "3029922922",
          "3001421318",
          "2112020",
          "2020676",
          "3030446844",
          "3005979524",
          "3012667060",
          "3009156722",
          "3029906224",
          "3010882708",
          "3023530182",
          "3006446479",
          "3003080079",
          "2518436",
          "3002808440",
          "1219324",
          "3006087789",
          "3013154126",
          "3008264065",
          "3002806988",
          "3011205710",
          "1000115734",
          "3002808696",
          "3017520736",
          "3004604967",
          "3021947238",
          "3002808128",
          "3029871637",
          "3003862188",
          "2518422",
          "3021244152",
          "3013095415",
          "3010157426",
          "3003007937",
          "3000957797",
          "3023272766",
          "3009609218",
          "3003477135",
          "3009129579",
          "3000143502",
          "3007603826",
          "1000122786",
          "3010017366",
          "3011213495",
          "2021710",
          "3008867836",
          "3009490946",
          "3038259592",
          "3006240053"
        ],
        "device_name": "Ventilator, Continuous, Facility Use",
        "medical_specialty_description": "Anesthesiology",
        "regulation_number": "868.5895",
        "device_class": "2"
      }
    }
  ]
}