      "cfres_id": "42764",
      "product_res_number": "Z-0893-06",
      "event_date_initiated": "2005-04-04",
      "event_date_posted": "2006-05-23",
      "recall_status": "Terminated",
      "event_date_terminated": "2007-11-27",
      "res_event_number": "33938",
      "product_code": "LSZ",
      "pma_numbers": [
        "P850064"
      ],
      "product_description": "Life Pulse High Frequency Ventilator (HFV)",
      "code_info": "Serial numbers 2585 and below.",
      "firm_fei_number": "3032745446",
      "recalling_firm": "Bunnell, Inc.",
      "address_1": "436 Lawndale Dr",
      "city": "Salt Lake City",
      "state": "UT",
      "postal_code": "84115-2917",
      "additional_info_contact": "801-467-0811 Ext. 25",
      "reason_for_recall": "Potential loss of air flow or change in cycling of the ventilator. Valve Drive Regulator Board may not function properly at elevated temperatures.  The increase in temperature could cause a reduction of power, causing the solenoid in the valve to function sporadically.",
      "root_cause_description": "Other",
      "action": "Notified consignees by letter on 04/04/2005.",
      "product_quantity": "297 units",
      "distribution_pattern": "Nationwide, Military medical center, Canada, and Czech Republic",
      "openfda": {
        "pma_number": [
          "P890057",
          "P890034",
          "P880013",
          "P850064"
        ],
        "device_name": "Ventilator, High Frequency",
        "medical_specialty_description": "Unknown",
        "regulation_number": "",
        "device_class": "3",
        "registration_number": [
          "2021710",
          "3032745446",
          "8030673",
          "1220908",
          "1314417"
        ],
        "fei_number": [
          "2021710",
          "3032745446",
          "1220908",
          "1314417",
          "3004213967"
        ]
      }
    }
  ]
}