{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-08-13",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "42246",
      "product_res_number": "Z-0134-06",
      "event_date_initiated": "2005-09-29",
      "event_date_posted": "2005-11-15",
      "recall_status": "Terminated",
      "event_date_terminated": "2007-11-29",
      "res_event_number": "33659",
      "product_code": "BRZ",
      "k_numbers": [
        "K830014",
        "K865060",
        "K920736",
        "K941214"
      ],
      "product_description": "LifeShield Latex-Free HEMA Y-Type Blood PlumSet, Nonvented, 122 inch, with 210 Micron Filter, Prepierced Injection Site, Prepierced Secondary Port and Option-Lok for use with Plum Series Infusers; Hospira, Inc., Lake Forest, IL 60045 USA, Made in Costa Rica; list 11664-03",
      "code_info": "list 11664-03, lots 240285H, 261265H, 271075H, 280885H",
      "firm_fei_number": "3013319212",
      "recalling_firm": "Hospira Inc.",
      "address_1": "275 N Field Dr",
      "city": "Lake Forest",
      "state": "IL",
      "postal_code": "60045-2579",
      "additional_info_contact": "Ms. Luann M. Pendy\n224-212-2000",
      "reason_for_recall": "Inlet and outlet tubing on the cassette may be reversed, which would result in reverse flow.",
      "root_cause_description": "Other",
      "action": "Hospira sent product recall letters dated 9/29/05 to the accounts via Federal Express overnight letter on the same date. The letters informed the accounts of the possibility of the inlet and outlet tubing being reversed on the cassette, which would allow the patient''s blood to be drawn up the tubing. The accounts were requested to examine their inventory for the affected lots, quarantine any found and call Hospira Customer Care at 1-877-946-7747 to arrange for return of the product for replacement.",
      "product_quantity": "172,872 sets",
      "distribution_pattern": "Nationwide, including Puerto Rico, and internationally to Australia, Mexico, Canada, Taiwan and Saudi Arabia.",
      "openfda": {
        "k_number": [
          "K822410",
          "K993120",
          "K833294",
          "K853386",
          "K073339",
          "K000685",
          "K083349",
          "K921595",
          "K041496",
          "K220558",
          "K210335",
          "K861523",
          "K852167",
          "K851477",
          "K132905",
          "K864441",
          "K871908",
          "K920788",
          "K840411",
          "K880174",
          "K972843",
          "K841187",
          "K940074",
          "K822409",
          "K854562",
          "K023039",
          "K940055",
          "K841561",
          "K050805",
          "K050115"
        ],
        "registration_number": [
          "2029275",
          "3010934635",
          "2032098",
          "3002808468",
          "3012410077",
          "9612051",
          "3013517171",
          "3010432890",
          "9616066",
          "9612052",
          "1066733",
          "1061124",
          "3003704957",
          "8040700",
          "1416980",
          "3011237770",
          "3004548776",
          "8020785",
          "3006788678",
          "1319639",
          "3013557562",
          "3030447506",
          "3015276780",
          "1018470",
          "1422078",
          "3002907620",
          "1423507",
          "3033536319"
        ],
        "fei_number": [
          "3010934635",
          "3002808468",
          "1000138054",
          "3002910717",
          "3002808169",
          "3012410077",
          "3013517171",
          "3010432890",
          "3002806443",
          "3003704957",
          "1416980",
          "3008002452",
          "3011237770",
          "3004548776",
          "1000306431",
          "1000512168",
          "3006788678",
          "3013557562",
          "3030447506",
          "3015276780",
          "3009291251",
          "1018470",
          "3002907620",
          "3002806457",
          "1000036860",
          "3033536319",
          "3002806588",
          "3003058448"
        ],
        "device_name": "Set, Blood Transfusion",
        "medical_specialty_description": "General Hospital",
        "regulation_number": "880.5440",
        "device_class": "2"
      }
    }
  ]
}