  "results": [
    {
      "cfres_id": "41854",
      "product_res_number": "Z-0239-06",
      "event_date_initiated": "2005-09-01",
      "event_date_posted": "2005-11-30",
      "recall_status": "Terminated",
      "event_date_terminated": "2007-06-27",
      "res_event_number": "33375",
      "product_code": "DPS",
      "k_numbers": [
        "K010770",
        "K011177",
        "K030407",
        "K982279"
      ],
      "product_description": "Advisor Vital Signs Monitor (model 9200), catalog number 92D754335, with Impedance Respiration board (optional circuit board, P/N:  80004B1) installed manufactured with Rev 10 Main boards (P/N:  80002B1) 5-lead 60 Hz ECG, standard SPO2 option, invasive Pressure/Temperature, printer and battery.",
      "code_info": "serial number:  AM05070143",
      "firm_fei_number": "2183502",
      "recalling_firm": "Smiths Medical PM, Inc.",
      "address_1": "N7w22025 Johnson Dr",
      "city": "Waukesha",
      "state": "WI",
      "postal_code": "53186-1856",
      "reason_for_recall": "Potential Safety related problem associated with the pace detect function on some Advisor monitors that have impedance respiration capabilities in that random pace markers may appear when the pace detect function is turned on and the primary lead is set to Lead II, III or any of the augmented leads, AVL, AVR or AVF on the Advisor monitor.",
      "root_cause_description": "Other",
      "action": "Distirbutors and consignees were notified beginning 09/01/05 via email or phone about Safety Action Bulletin 05-004.   Customers were asked to respond to Smiths Medical with a written acknowledgement that they received this Safety Action Bulletin (SAB).  The SAB described the issue, actions they should take and instruction on how to have the monitor repaired.",
      "product_quantity": "1",
      "distribution_pattern": "AZ, CA, KS, KY, NY, TN, TX  OUS to countries such as:  Egypt, France, Iran, Mexico, Peru, Romania, Saudi Arabia and Spain",
      "openfda": {
        "k_number": [
          "K241217",
          "K881939",
          "K251670",
          "K052511",
          "K072353",
          "K914206",
          "K896386",
          "K850021",
          "K950004",
          "K080036",
          "K051534",
          "K933065",
          "K853979",
          "K043074",
          "K914974",
          "K993875",
          "K030535",
          "K901817",
          "K171943",
          "K833458",
          "K771380",
          "K923221",
          "K922927",
          "K113485",
          "K862544",
          "K861992",
          "K973426",
          "K800492",
          "K032200",
          "K093339",
          "K791540",
          "K080884",
          "K133985",
          "K790114",
          "K852770",
          "K781054",
          "K180234",
          "K811016",
          "K800659",
          "K915879",
          "K880357",
          "K093872",
          "K760271",
          "K002473",
          "K790233",
          "K103077",
          "K090936",
          "K043380",
          "K081437",
          "K810357",
          "K813279",
          "K103427",
          "K020366",
          "K981951",
          "K171942",
          "K821758",
          "K844744",
          "K851643",
          "K913374",
          "K050858",
          "K904758",
          "K840440",
          "K897029",
          "K113144",
          "K130409",
          "K813021",
          "K102206",
          "K882197",
          "K091358",
          "K152384",
          "K864009",
          "K860636",
          "K964784",
          "K921331",
          "K013717",
          "K082967",
          "K001492",
          "K760612",
          "K881328",
          "K002822",
          "K860646",
          "K161302",
          "K850322",
          "K012686",
          "K171360",
          "K802190",
          "K182571",
          "K896772",
          "K111159",
          "K090249",
          "K121825",
          "K223670",
          "K062282",
          "K170182",
          "K082312",
          "K832227",
          "K212163",
          "K171517",
          "K900700",
          "K903722",
          "K080933",
          "K951128",
          "K191692",
          "K160746",
          "K802718",
          "K933156",
          "K212112",
          "K760854",
          "K792072",
          "K883402",
          "K003392",
          "K974351",
          "K031557",
          "K800004",
          "K111805",
          "K781679",
          "K832050",
          "K052898",
          "K950018",
          "K863535",
          "K232974",
          "K832130",
          "K912183",
          "K002459",
          "K183369",
          "K952173",
          "K955891",
          "K896417",
          "K113234",
          "K822582",
          "K243305",
          "K863914",
          "K920364",
          "K140020",
          "K811950",
          "K232816",
          "K232035",
          "K112959",
          "K972991",
          "K082380",
          "K761014",
          "K091513",
          "K102854",
          "K874641",
          "K062469",
          "K240229",
          "K871239",
          "K942629",
          "K883903",
          "K880595",
          "K101569",
          "K900272",
          "K180242",
          "K883454",
          "K203786",
          "K122073",
          "K820014",
          "K811121",
          "K093211",
          "K051163",
          "K852829",
          "K232512",
          "K823584",
          "K934439",
          "K830791",
          "K250569",
          "K851633",
          "K021230",
          "K963400",
          "K082173",
          "K791364",
          "K191738",
          "K861359",
          "K860212",
          "K241556",
          "K971038",
          "K974220",
          "K002197",
          "K832014",
          "K113306",
          "K950779",
          "K800005",
          "K173226",
          "K870880",
          "K780307",
          "K861851",
          "K120770",
          "K051126",
          "K974746",
          "K961247",
          "K102378",
          "K955045",
          "K862553",
          "K861281",
          "K870926",
          "K953415",
          "K090212",
          "K980362",
          "K152427",
          "K120865",
          "K841611",
          "K840869",
          "K882472",
          "K801677",
          "K070136",
          "K163607",
          "K875057",
          "K921930",
          "K862748",
          "K850702",
          "K972310",
          "K890791",
          "K222902",
          "K961798",
          "K183322",
          "K844470",
          "K160092",
          "K810988",
          "K885177",
          "K791420",
          "K182093",
          "K123217",
          "K141582",
          "K071168",
          "K252361",
          "K060167",
          "K904355",
          "K780166",
          "K981807",
          "K810154",
          "K143432",
          "K093869",
          "K864978",
          "K232445",
          "K960321",
          "K864700",
          "K210140",
          "K022618",
          "K870380",
          "K850323",
          "K841148",
          "K230834",
          "K884116",
          "K842944",
          "K972996",
          "K974079",
          "K183425",
          "K903774",
          "K170954",
          "K760465",
          "K900598",
          "K093172",
          "K072449",
          "K790739",
          "K910017",
          "K062677",
          "K143436",
          "K110065",
          "K963198",
          "K896064",
          "K012264",
          "K993414",
          "K823488",
          "K122632",
          "K770832",
          "K001034",
          "K101115",
          "K852282",
          "K972199",
          "K141946",
          "K230812",
          "K860734",
          "K011436",
          "K863030",
          "K841254",
          "K864465",
          "K901948",
          "K955416",
          "K082013",
          "K111438",
          "K133622",
          "K032145",
          "K833951",
          "K013367",
          "K802113",
          "K901771",
          "K760265",
          "K872548",
          "K832013",
          "K874326",
          "K903671",
          "K833584",
          "K813022",
          "K963578",
          "K013615",
          "K840571",
          "K884956",
          "K854840",
          "K024283",
          "K052295",
          "K013565",
          "K861850",
          "K022149",
          "K761023",
          "K013564",
          "K012414",
          "K083715",
          "K902368",
          "K201456",
          "K955642",
          "K082946",
          "K200884",
          "K150564",
          "K131503",
          "K103791",
          "K830666",
          "K790869",
          "K892835",
          "K031422",
          "K020708",
          "K131573",
          "K860175",
          "K120188",
          "K884896",
          "K832184",
          "K820067",
          "K112135",
          "K930134",
          "K761198",
          "K890481",
          "K935394",
          "K050074",
          "K881055",
          "K212278",
          "K884662",
          "K791421",
          "K924544",
          "K173952",
          "K770366",
          "K070539",
          "K050620",
          "K232161",
          "K873346",
          "K002074",
          "K810263",
          "K991831",
          "K760268",
          "K122356",
          "K820011",
          "K844392",
          "K101403",
          "K843934",
          "K061123",
          "K231160",
          "K252474",
          "K223013",
          "K193415",
          "K172068",
          "K897121",
          "K123816",
          "K860973",
          "K173830",
          "K760510",
          "K802952",
          "K100813",
          "K790691",
          "K050686",
          "K922291",
          "K181998",
          "K811949",
          "K854234",
          "K860249",
          "K122712",
          "K955113",
          "K823281",
          "K954066",
          "K072217",
          "K110328",
          "K945626",
          "K810104",
          "K884504",
          "K832212",
          "K791201",
          "K760150",
          "K062068",
          "K760266",
          "K760677",
          "K103640",
          "K852181",
          "K070614",
          "K102767",
          "K862381",
          "K112622",
          "K811341",
          "K851662",
          "K042827",
          "K821888",
          "K073625",
          "K873584",
          "K855096",
          "K864614",
          "K914725",
          "K850454",
          "K152863",
          "K030104",
          "K941811",
          "K760926",
          "K884800",
          "K182790",
          "K200036",
          "K802368",
          "K172778",
          "K950989",
          "K770760",
          "K130921",
          "K882198",
          "K803309",
          "K231150",
          "K760542",
          "K832015",
          "K791795",
          "K852281",
          "K896190",
          "K131900",
          "K233266",
          "K052049",
          "K943709",
          "K863116",
          "K010822",
          "K852322",
          "K022152",
          "K791151",
          "K864843",
          "K092576",
          "K904606",
          "K131262",
          "K132068",
          "K162665",
          "K770037",
          "K792229",
          "K053083",
          "K823529",
          "K221141",
          "K160667",
          "K915760",
          "K844101",
          "K821090",
          "K760463",
          "K142105",
          "K031182",
          "K981697",
          "K000645",
          "K964408",
          "K823220",
          "K790794",
          "K160876",
          "K840569",
          "K920020",
          "K142969",
          "K080999",
          "K961401",
          "K883468",
          "K243252",
          "K112431",
          "K062894",
          "K160840",
          "K024278",
          "K131572",
          "K832421",
          "K100127",
          "K042177",
          "K042814",
          "K831878",
          "K011177",
          "K864504",
          "K882254",
          "K884992",
          "K862951",
          "K221260",
          "K830566",
          "K090367",
          "K220535",
          "K933404",
          "K972603",
          "K050674",
          "K851322",
          "K914762",
          "K833470",
          "K142551",
          "K170995",
          "K082709",
          "K102830",
          "K851044",
          "K872658",
          "K810696",
          "K252463",
          "K010955",
          "K923406",
          "K101876",
          "K010505",
          "K142288",
          "K890875",
          "K864580",
          "K780138",
          "K800566",
          "K101867",
          "K930558",
          "K833593",
          "K832883",
          "K152944",
          "K922703",
          "K861067",
          "K973437",
          "K991786",
          "K860790",
          "K831429",
          "K101016",
          "K232823",
          "K021171",
          "K871993",
          "K800593",
          "K221321",
          "K901947",
          "K893660",
          "K903524",
          "K111601",
          "K840020",
          "K863652",
          "K220362",
          "K092010",
          "K210560",
          "K130207",
          "K955546",
          "K060766",
          "K040313",
          "K131819",
          "K001896",
          "K863100",
          "K960449",
          "K152043",
          "K110266",
          "K790300",
          "K970787",
          "K132758",
          "K142693",
          "K012540",
          "K131858",
          "K033488",
          "K864704",
          "K092663",
          "K881475",
          "K152135",
          "K072060",
          "K972746",
          "K894956",
          "K121009",
          "K091397",
          "K865031",
          "K832888",
          "K151135",
          "K946205",
          "K103765",
          "K082124",
          "K170568",
          "K010758",
          "K014262",
          "K090483",
          "K760272",
          "K790939",
          "K822768",
          "K170154",
          "K012263",
          "K810176",
          "K081473",
          "K052303",
          "K901032"
        ],
        "registration_number": [
          "3041276591",
          "3030733800",
          "3043668257",
          "9617277",
          "3012494290",
          "1220477",
          "3014449259",
          "3009077524",
          "3011861727",
          "3003681187",
          "3000130829",
          "3018094310",
          "3001570235",
          "3005783425",
          "3021131596",
          "3014848734",
          "8030978",
          "3009499478",
          "3013188547",
          "3007305624",
          "3012516456",
          "3007184699",
          "3014522540",
          "3010698831",
          "3008642652",
          "3008164770",
          "9611031",
          "3018587883",
          "3004145393",
          "9611295",
          "3016618143",
          "3013557562",
          "3030447506",
          "1319390",
          "1066427",
          "1220908",
          "3013250653",
          "3003294644",
          "3033558564",
          "3018783526",
          "3005569927",
          "3010381606",
          "3045323923",
          "3004186703",
          "3012404027",
          "8030229",
          "3015142815",
          "3011721071",
          "1314417",
          "1218950",
          "3008729547",
          "3013679558",
          "3030154814",
          "3031570367",
          "1651104",
          "3021422553",
          "3016852448",
          "3008881835",
          "3001149827",
          "8030673",
          "1066270",
          "3021559257",
          "3032536454",
          "1316463",
          "3004906872",
          "3013501110",
          "1054713",
          "2080783",
          "3010390468",
          "3003591740",
          "3004582233",
          "2182208",
          "3010097171",
          "3012250536",
          "1930027",
          "3007836437",
          "3000126629",
          "3030125003",
          "3010089768",
          "3008820594",
          "3011651230",
          "3013896610",
          "3003174284",
          "3033536319",
          "3015134086",
          "3012421607",
          "1523530",
          "3010082909",
          "3023412068",
          "2000004234",
          "3004035727",
          "3011353843",
          "3007134734",
          "3016674095",
          "3031158",
          "3008716327",
          "8030245",
          "3035371769",
          "3033959233",
          "3003647588",
          "3043479346",
          "3008293988",
          "3009735945",
          "3030639099",
          "3006979678",
          "3031786139",
          "3008717264",
          "3030446844",
          "3030597723",
          "3007734888",
          "3016931915",
          "3023146633",
          "9611252",
          "3010419931",
          "3009607734",
          "3008483389",
          "3007384703",
          "3009443693",
          "1219324",
          "2433214",
          "3006087789",
          "3009831876",
          "3003857828",
          "3034603279",
          "3006371704",
          "1450057",
          "3009715978",
          "3008457078",
          "3036039911",
          "3001373226",
          "3007770164",
          "9611045",
          "3019388613",
          "3010536822",
          "9610816",
          "3029871637",
          "3007413079",
          "3012027770",
          "3011235999",
          "3014579161",
          "9680510",
          "3043226252",
          "3013757696",
          "3013689228",
          "3005877899",
          "3010157426",
          "3005488716",
          "3007048291",
          "3010342616",
          "3038259592",
          "3003622639",
          "3023272766",
          "3003263092",
          "3020981894",
          "1126271",
          "9710602",
          "3012359877",
          "2084041",
          "3016701404",
          "9615102",
          "3010838917",
          "3003832357",
          "3012528160",
          "3013596024",
          "3014541700",
          "3012302888",
          "3008505660",
          "3011359729",
          "3045498928",
          "3011200896",
          "3013500228",
          "3032109181",
          "1000122786",
          "3004415095",
          "9610105",
          "3005521864",
          "3009183442",
          "3042928308",
          "3008921101",
          "3003971136",
          "3007029079"
        ],
        "fei_number": [
          "3041276591",
          "3030733800",
          "3043668257",
          "3012494290",
          "1220477",
          "3014449259",
          "3009077524",
          "3011861727",
          "3002807720",
          "3003681187",
          "3018094310",
          "3000130829",
          "3001570235",
          "3005783425",
          "3002807739",
          "3021131596",
          "3014848734",
          "3009499478",
          "3013188547",
          "3007305624",
          "3012516456",
          "3007184699",
          "3014522540",
          "3010698831",
          "3008642652",
          "3008164770",
          "3018587883",
          "3004213967",
          "3004145393",
          "3016618143",
          "3013557562",
          "3030447506",
          "1319390",
          "3002807992",
          "3004072282",
          "1220908",
          "3013250653",
          "3003294644",
          "3033558564",
          "3018783526",
          "3005569927",
          "3010381606",
          "3045323923",
          "3004186703",
          "3012404027",
          "3015142815",
          "3011721071",
          "1314417",
          "1218950",
          "3008729547",
          "3013679558",
          "3030154814",
          "3031570367",
          "3021422553",
          "3016852448",
          "3008881835",
          "3001149827",
          "3003513129",
          "3021559257",
          "3032536454",
          "1316463",
          "3004906872",
          "3013501110",
          "2080783",
          "3002806783",
          "3010390468",
          "3003591740",
          "1000271197",
          "3004582233",
          "2182208",
          "3010097171",
          "3012250536",
          "1930027",
          "3007836437",
          "3000126629",
          "3030125003",
          "3000215024",
          "3010089768",
          "3008820594",
          "3011651230",
          "3013896610",
          "1000114111",
          "3003174284",
          "3033536319",
          "3015134086",
          "3012421607",
          "1523530",
          "3010082909",
          "3023412068",
          "2000004234",
          "3004035727",
          "3011353843",
          "3002807674",
          "3003708031",
          "3007134734",
          "3016674095",
          "3008716327",
          "3035371769",
          "3033959233",
          "3003647588",
          "3043479346",
          "3008293988",
          "1000263641",
          "3007513035",
          "3009735945",
          "3030639099",
          "3006979678",
          "3031786139",
          "3008717264",
          "3002807019",
          "3030446844",
          "3030597723",
          "3007734888",
          "3016931915",
          "3023146633",
          "3010419931",
          "3009607734",
          "3008483389",
          "3007384703",
          "3009443693",
          "1219324",
          "2433214",
          "3006087789",
          "3009831876",
          "3002808182",
          "3003857828",
          "3034603279",
          "3006371704",
          "1450057",
          "3009715978",
          "3008457078",
          "3002807167",
          "3036039911",
          "3001373226",
          "3007770164",
          "3019388613",
          "3010536822",
          "3029871637",
          "3007413079",
          "3012027770",
          "3011235999",
          "3014579161",
          "3043226252",
          "3013757696",
          "3013689228",
          "3005877899",
          "3010157426",
          "3005488716",
          "3007048291",
          "3010342616",
          "3038259592",
          "3000957797",
          "3003622639",
          "3023272766",
          "3003263092",
          "3020981894",
          "3003190524",
          "3012359877",
          "2084041",
          "3016701404",
          "3010838917",
          "3003832357",
          "3012528160",
          "3013596024",
          "3014541700",
          "3012302888",
          "3008505660",
          "3011359729",
          "3045498928",
          "3011200896",
          "3002737156",
          "3013500228",
          "3032109181",
          "1000122786",
          "3004415095",
          "3005521864",
          "3009183442",
          "3042928308",
          "3008921101",
          "3003971136",
          "3007029079"
        ],
        "device_name": "Electrocardiograph",
        "medical_specialty_description": "Cardiovascular",
        "regulation_number": "870.2340",
        "device_class": "2"
      }
    }
  ]
}