  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-08-20",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "41034",
      "product_res_number": "Z-0624-06",
      "event_date_initiated": "2005-03-23",
      "event_date_posted": "2006-03-15",
      "recall_status": "Terminated",
      "event_date_terminated": "2012-01-27",
      "res_event_number": "32890",
      "product_code": "MGU",
      "pma_numbers": [
        "P890025"
      ],
      "product_description": "In vitro Test  TDx/TDxFLx Cyclosporine Monoclonal Whole Blood  Reagent Pack 100 test\n\nTDx/TDxFLx Cyclosporine Monoclonal Whole Blood Calibrators are composed of six vials of calibrators A, B, C, D, E, and F containing cyclosporine prepared in human blood component (non-reactive for HBxAg, anti-HCv, anti-HIV-1/HIV-20. \n\nTDx/TDxFLx Cyclosporine Monoclonal Whole Blood Controls are composed of three vials L, M and H containing cyclosporine in human blood componenet (nonreactive for HBsAg, anti-HCV, and anti-HIV-1/HIV-2).",
      "code_info": "Lot 18229Q100 (exp 8/26/2006) Lot 18230Q100 (exp 10/05/2006)",
      "firm_fei_number": "2623532",
      "recalling_firm": "Abbott Health Products, Inc.",
      "address_1": "Hwy 2 Km 58.0 Cruze Davila",
      "city": "Barceloneta",
      "postal_code": "00617",
      "country": "PR",
      "additional_info_contact": "Nydia I. Cancel\n787-846-3500 Ext. 7350",
      "reason_for_recall": "Patient results reported using calibrators made from new primary material may be falsely decreased by 5-8% compared to results using calibrators made from the previous primary material.  An Important Product Information Letter with technical information regarding the Whole Blood Calibrators and Controls, implementing new internal reference standards was issued.",
      "root_cause_description": "Other",
      "action": "Abbott has recently implemented new internal reference standards for TDx/TDxFLx Cyclosporine monoclonal Whole Blood Calibrators and Controls.  A change in performance is associated with the use of calibrator lots 1829Q100 and higher and control lots 18230Q100 and higher, which are the first lots manufactured using the new internal reference standards.  The specific performance changes are identified in the letter, dated March 23, 2005, as compared to current product using the previous standards (calibrators lot 05434Q100 and control lot 03071Q100).",
      "product_quantity": "4,514 kits were distributed",
      "distribution_pattern": "Nationwide, Puerto Rico, Hawaii, Mexico, Guatemala, Costa Rica, Venezuela, Colombia, Chile, Uruguay, Argentina, Brazil, Dominican Republic, Canada, Germany, Japan , Singapore, India, Taiwan, Thailand,  Korea, Australia, Honduras, and Panama,",
      "openfda": {
        "pma_number": [
          "P890025"
        ],
        "device_name": "Fluorescence Polarization Immunoassay For Cyclosporine",
        "medical_specialty_description": "Clinical Chemistry",
        "regulation_number": "862.1235",
        "device_class": "2"
      }
    }
  ]
}