  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-08-13",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "38874",
      "product_res_number": "Z-0256-06",
      "event_date_initiated": "2004-09-07",
      "event_date_posted": "2005-12-01",
      "recall_status": "Terminated",
      "event_date_terminated": "2006-10-03",
      "res_event_number": "31986",
      "product_code": "NDH",
      "k_numbers": [
        "K013468"
      ],
      "product_description": "Long Length Dyax Nail.",
      "code_info": "Catalog No. 1594-4430S Lot K559623",
      "firm_fei_number": "3003070421",
      "recalling_firm": "Stryker Howmedica Osteonics Corp.",
      "address_1": "325 Corporate Dr",
      "city": "Mahwah",
      "state": "NJ",
      "postal_code": "07430-2002",
      "additional_info_contact": "Dominick Cristofolo\n201-831-5247",
      "reason_for_recall": "Stryker became aware that Catalog number 1594-1130S, Lot Code K559623 was manufactured with a lag screw hold angulation of 135 degrees rather than 130 degrees.",
      "root_cause_description": "Other",
      "action": "The firm sent out notification letters via FedEx on 9/7/2004 with return receipt.",
      "product_quantity": "4",
      "distribution_pattern": "The devices were distributed to Stryker branches and agencies.  These are Stryker sales representatives located nationwide.",
      "openfda": {
        "registration_number": [
          "3015542154",
          "3006801265",
          "3035366890",
          "3011683674",
          "3015207155",
          "3013176080",
          "1825034"
        ],
        "fei_number": [
          "3015542154",
          "3006801265",
          "3035366890",
          "3011683674",
          "3015207155",
          "3013176080",
          "1825034"
        ],
        "device_name": "Nail, Fixation, Bone, Metallic",
        "medical_specialty_description": "Orthopedic",
        "regulation_number": "888.3030",
        "device_class": "2"
      }
    }
  ]
}