  "results": [
    {
      "cfres_id": "38328",
      "product_res_number": "Z-0286-06",
      "event_date_initiated": "2005-03-25",
      "event_date_posted": "2005-12-15",
      "recall_status": "Terminated",
      "event_date_terminated": "2011-08-08",
      "res_event_number": "31670",
      "product_code": "LJS",
      "k_numbers": [
        "K920755"
      ],
      "product_description": "16 G / 5F Dual Lumen L-Cath Complete Kit with Introsyte",
      "code_info": "REF 384427",
      "firm_fei_number": "1710034",
      "recalling_firm": "Becton Dickinson Medical Systems",
      "address_1": "9450 S State St",
      "city": "Sandy",
      "state": "UT",
      "postal_code": "84070-3213",
      "additional_info_contact": "801-565-2341",
      "reason_for_recall": "Aspiration (blood return) could not be achieved on some L-Cath 16 gauge catheters.",
      "root_cause_description": "Other",
      "action": "Consignees were notified via telephone and fax letter, beginning 04/25/2005.",
      "product_quantity": "3750 units",
      "distribution_pattern": "AL, CA, CO, FL, GA, IL, IN, KY, LA, MA, MD, MI, MN, NJ, OH, OK, PA, TN, TX, WA. No Government or military consignees. Foreign distribution to Australia, Canada, Japan.",
      "openfda": {
        "k_number": [
          "K003110",
          "K960055",
          "K141634",
          "K841878",
          "K993614",
          "K832774",
          "K910486",
          "K935380",
          "K093252",
          "K072391",
          "K162441",
          "K843450",
          "K950987",
          "K971999",
          "K130897",
          "K921755",
          "K053345",
          "K061179",
          "K093775",
          "K954302",
          "K121990",
          "K161765",
          "K973057",
          "K041088",
          "K163452",
          "K192246",
          "K031626",
          "K021704",
          "K132567",
          "K833915",
          "K130034",
          "K920755",
          "K935145",
          "K882448",
          "K910315",
          "K051102",
          "K110627",
          "K001874",
          "K034020",
          "K883313",
          "K232571",
          "K082481",
          "K892926",
          "K150448",
          "K112744",
          "K980090",
          "K050931",
          "K162443",
          "K113225",
          "K073451",
          "K123617",
          "K030255",
          "K091953",
          "K061289",
          "K232594",
          "K072509",
          "K071875",
          "K961155",
          "K844080",
          "K153469",
          "K954720",
          "K153423",
          "K141151",
          "K070002",
          "K142616",
          "K243484",
          "K832580",
          "K840958",
          "K122545",
          "K871998",
          "K233149",
          "K983544",
          "K093927",
          "K041420",
          "K141633",
          "K152409",
          "K033389",
          "K041849",
          "K072230",
          "K926331",
          "K970269",
          "K170934",
          "K964929",
          "K904334",
          "K061240",
          "K893126",
          "K123759",
          "K170486",
          "K173956",
          "K062425",
          "K121094",
          "K932464",
          "K152819",
          "K972555",
          "K070996",
          "K061936",
          "K051991",
          "K091466",
          "K131687",
          "K130876",
          "K113043",
          "K091586",
          "K241587",
          "K951098",
          "K092040",
          "K172885",
          "K131942",
          "K003642",
          "K903648",
          "K945956",
          "K031129",
          "K051417",
          "K921751",
          "K093366",
          "K964447",
          "K933884",
          "K092347",
          "K100635",
          "K111244",
          "K003682",
          "K932433",
          "K142267",
          "K133264",
          "K945499",
          "K834091",
          "K110914",
          "K180548",
          "K042491",
          "K071361",
          "K962300",
          "K213550",
          "K093406",
          "K952626",
          "K140345",
          "K963806",
          "K051690",
          "K830233",
          "K001901",
          "K050889",
          "K840846",
          "K061986",
          "K260929",
          "K121941",
          "K151985",
          "K210264",
          "K833442",
          "K003839",
          "K252792",
          "K072625",
          "K172899",
          "K252373",
          "K172511",
          "K120980",
          "K993442",
          "K910622",
          "K092306",
          "K091261",
          "K140799",
          "K220363",
          "K895712",
          "K093309",
          "K230669",
          "K851243",
          "K120981",
          "K132334",
          "K101329",
          "K200037",
          "K070615",
          "K831386",
          "K897168",
          "K053501",
          "K052475",
          "K834299",
          "K111906",
          "K932211",
          "K023374",
          "K031718",
          "K953436",
          "K101326",
          "K861563",
          "K930129",
          "K012902",
          "K091924",
          "K970926",
          "K834383",
          "K896598",
          "K130507",
          "K102966",
          "K113277",
          "K152261",
          "K171994",
          "K180546",
          "K161496",
          "K932198",
          "K241910",
          "K042126",
          "K964556",
          "K841443",
          "K141818",
          "K030400",
          "K050542",
          "K180560",
          "K954422",
          "K963097",
          "K153487",
          "K243458",
          "K946250",
          "K041468",
          "K181904",
          "K083920",
          "K840681",
          "K954638",
          "K831587",
          "K072196",
          "K992424",
          "K970434",
          "K040446",
          "K083763",
          "K171931",
          "K930630",
          "K030270",
          "K113808",
          "K911067",
          "K201452",
          "K925236",
          "K080604",
          "K983412",
          "K112896",
          "K051672",
          "K180567",
          "K251212",
          "K830256",
          "K091488",
          "K063363",
          "K953256",
          "K063848",
          "K840742",
          "K131038",
          "K012641",
          "K222232",
          "K172039",
          "K062579",
          "K831845",
          "K834473",
          "K091324",
          "K030415",
          "K953300",
          "K160925",
          "K072518",
          "K944685",
          "K133634",
          "K923653",
          "K121089",
          "K851195",
          "K831802",
          "K993052",
          "K081904",
          "K823553",
          "K091670",
          "K864086",
          "K963215",
          "K882100",
          "K051194",
          "K132885",
          "K961856",
          "K043502",
          "K172397",
          "K930473",
          "K971085",
          "K063240",
          "K102159",
          "K140266",
          "K142889",
          "K961336",
          "K193015",
          "K083873",
          "K926293",
          "K962849",
          "K243941",
          "K143238",
          "K121501",
          "K122882",
          "K963257",
          "K830406",
          "K932430",
          "K951078",
          "K924677",
          "K122423",
          "K954104",
          "K240486",
          "K834225",
          "K070613",
          "K042461",
          "K060887",
          "K150527",
          "K924851",
          "K954409",
          "K010349",
          "K830969",
          "K895543",
          "K050185",
          "K034019"
        ],
        "registration_number": [
          "3003898360",
          "1220477",
          "2245270",
          "3009211636",
          "3011534620",
          "3013666218",
          "3018094310",
          "3035111606",
          "1724474",
          "1048735",
          "2030624",
          "3030148766",
          "1722746",
          "3013557562",
          "1000484830",
          "3004111573",
          "1833566",
          "1319211",
          "3010882065",
          "1721676",
          "3013162291",
          "2011171",
          "9617592",
          "3004753388",
          "1061124",
          "1921846",
          "3033589330",
          "1219930",
          "3011237770",
          "1644312",
          "1721686",
          "3005704822",
          "3024673",
          "3015309643",
          "1424778",
          "3003915875",
          "9612433",
          "1000314502",
          "3012536737",
          "1423537",
          "3021570645",
          "3008236846",
          "3038181770",
          "3007544874",
          "1625425",
          "3010827261",
          "3013750125",
          "3030714660",
          "3007137787",
          "1643817",
          "3006942602",
          "3010041511",
          "3007697249",
          "2527072",
          "3010882708",
          "3013041594",
          "2183744",
          "1647149",
          "3005636544",
          "2246552",
          "1721504",
          "3006082230",
          "3013501772",
          "3006950086",
          "3015537049",
          "3013788631",
          "3010532612",
          "3009888344",
          "3015859709",
          "3004134316",
          "3016472589",
          "3006260740",
          "1718873",
          "1225492",
          "3011137372",
          "9681477",
          "9617604",
          "3006425876",
          "2133772",
          "1018233",
          "3007207906",
          "1319639",
          "2518902",
          "3015225571",
          "1223004",
          "9680794",
          "3013233645",
          "3015060216",
          "3016727301",
          "1820334"
        ],
        "fei_number": [
          "3003898360",
          "1220477",
          "2245270",
          "3009211636",
          "3011534620",
          "3013666218",
          "3018094310",
          "3002807439",
          "3002468371",
          "3035111606",
          "1048735",
          "3003473227",
          "3030148766",
          "3013557562",
          "1000484830",
          "1000117748",
          "3005747797",
          "3004111573",
          "1319211",
          "1000138054",
          "1721676",
          "3013162291",
          "2011171",
          "3004753388",
          "1921846",
          "3033589330",
          "1219930",
          "3011237770",
          "1721686",
          "3000209794",
          "3005704822",
          "3024673",
          "3001237338",
          "3015309643",
          "3003094851",
          "3003915875",
          "1000314502",
          "3012536737",
          "1000306139",
          "1000115408",
          "1000306216",
          "3021570645",
          "3008236846",
          "3038181770",
          "3001236905",
          "3007544874",
          "1625425",
          "3010827261",
          "3013750125",
          "3030714660",
          "3007137787",
          "1643817",
          "3006942602",
          "3010041511",
          "3007697249",
          "2527072",
          "1000512168",
          "3010882708",
          "3013041594",
          "2183744",
          "3005636544",
          "2246552",
          "1721504",
          "3006082230",
          "3013501772",
          "3006950086",
          "3015537049",
          "3013788631",
          "3010532612",
          "3009888344",
          "3015859709",
          "3004134316",
          "3016472589",
          "3006260740",
          "1718873",
          "3006097687",
          "3006425876",
          "3002808464",
          "1000120555",
          "1018233",
          "3007207906",
          "3003737899",
          "2518902",
          "3015225571",
          "3013233645",
          "3015060216",
          "3016727301",
          "3002558692",
          "1820334",
          "3000206435"
        ],
        "device_name": "Catheter, Intravascular, Therapeutic, Long-Term Greater Than 30 Days",
        "medical_specialty_description": "General Hospital",
        "regulation_number": "880.5970",
        "device_class": "2"
      }
    }
  ]
}