  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-08-20",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "37978",
      "product_res_number": "Z-0666-05",
      "event_date_initiated": "2005-03-17",
      "event_date_posted": "2005-03-31",
      "recall_status": "Terminated",
      "event_date_terminated": "2005-09-16",
      "res_event_number": "31489",
      "product_code": "DTQ",
      "k_numbers": [
        "K022947"
      ],
      "product_description": "Terumo Advanced Perfusion System 1 (with 4 inch diameter roller pump); Catalog number 801040.",
      "code_info": "Serial numbers 0028 through 0527 and 0700 through 0874.",
      "firm_fei_number": "1828100",
      "recalling_firm": "Terumo Cardiovascular Systems Corp",
      "address_1": "6200 Jackson Rd",
      "city": "Ann Arbor",
      "state": "MI",
      "postal_code": "48103-9586",
      "additional_info_contact": "800-521-2818",
      "reason_for_recall": "The pumps will stop, show an overspeed error message, and will not restart using routine procedures if the user reduces the pump speed to zero RPM while the pump is operating in pulse mode.",
      "root_cause_description": "Other",
      "action": "Recall letters dated 3/17/05 were sent to customers advising them of the problem, the temporary fix, and that software will be upgraded.",
      "distribution_pattern": "Nationwide, Australia, Belgium, Canada, Egypt, Germany, Hong Kong, India, Indonesia, Japan, Malasia, Mexico, Pakistan, South Africa, South Korea, Taiwan, Thailand and United Arab Emirates.",
      "openfda": {
        "k_number": [
          "K842138",
          "K011088",
          "K052555",
          "K813034",
          "K863960",
          "K090006",
          "K822068",
          "K133598",
          "K021694",
          "K210130",
          "K861388",
          "K002116",
          "K071318",
          "K173834",
          "K943803",
          "K871131",
          "K221373",
          "K862973",
          "K851615",
          "K080832",
          "K870976",
          "K801097",
          "K915183",
          "K151349",
          "K060053",
          "K092463",
          "K190282",
          "K030141",
          "K070213",
          "K091007",
          "K020571",
          "K163531",
          "K772001",
          "K062396",
          "K103762",
          "K002118",
          "K863541",
          "K982014",
          "K102726",
          "K022947",
          "K942582",
          "K770002",
          "K232291",
          "K153376",
          "K221895",
          "K964017",
          "K202125",
          "K950990",
          "K172220",
          "K131041",
          "K851614",
          "K093882"
        ],
        "registration_number": [
          "2916596",
          "9680841",
          "3007427881",
          "1828100",
          "3010163695",
          "3042990265",
          "2184009",
          "1218444",
          "8010762",
          "9611109",
          "3019807891",
          "3006073153",
          "3002807090",
          "3010703925",
          "3009507273",
          "3005941719",
          "3011554160",
          "3001589655"
        ],
        "fei_number": [
          "2916596",
          "1000116158",
          "3007427881",
          "1828100",
          "3010163695",
          "3042990265",
          "1218444",
          "3002808267",
          "3019807891",
          "3006073153",
          "3010703925",
          "3009649876",
          "3006003617",
          "3002807111",
          "3009507273",
          "3011554160",
          "3001589655",
          "3005426550"
        ],
        "device_name": "Console, Heart-Lung Machine, Cardiopulmonary Bypass",
        "medical_specialty_description": "Cardiovascular",
        "regulation_number": "870.4220",
        "device_class": "2"
      }
    }
  ]
}