{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-08-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "36301",
      "product_res_number": "Z-0365-05",
      "event_date_initiated": "2004-11-23",
      "event_date_posted": "2004-12-24",
      "recall_status": "Terminated",
      "event_date_terminated": "2006-12-26",
      "res_event_number": "30481",
      "product_code": "GBK",
      "product_description": "AMPLATZER Exchange System 180-degree curve, order no. 9-EXCH-6F-180/80.  Product is labeled as Sterile EO.  \n\n\nThe AMPLATZER Exchange System is comprised of the identical components as the AMPLATZER Delivery System, which is comprised of delivery sheath, delivery cable, dilator, loading device and pin vise.  The only exception is the dilator component in the Exchange System has a larger inner lumen to allow passage over an AMPLATZER delivery cable.   Exchange Systems are placed onto polyboard backing cards.  The backing card is then placed into a smaller of 2 Tyvek pouches.  The pouch is then placed inside a larger Tyvek pouch.  The kit is assembled at AGA and then sterilized by Steris, Minneapolis, MN.",
      "code_info": "M02B14-09, M02C25-16",
      "firm_fei_number": "3005334138",
      "recalling_firm": "AGA Medical Corporation",
      "address_1": "682 Mendelssohn Ave N",
      "city": "Golden Valley",
      "state": "MN",
      "postal_code": "55427-4306",
      "additional_info_contact": "Jody Raus\n763-513-9227",
      "reason_for_recall": "Tubing utilized in the manufacture of the AMPLATZER Delivery and Exchange System sheaths and dilators was in some cases manufactured by our supplier outside of process parameters; causing a residue to be left on the inside of the tubing.",
      "root_cause_description": "Other",
      "action": "Recall notices were sent via e-mail 11/22/04 to physicians, hospitals and distributors who received affected product.  Hard copies of the Recall Notices were sent to all customers (international and domestic) via Federal Express 11/22-24/04.   A form that is attached to the Recall Notice asks for customers to fill out quantities of affected product and to return the affected product to AGA Medical.",
      "product_quantity": "157",
      "distribution_pattern": "Product has been distributed throughout the US and Internationally, including Canada, Europe and Australia.",
      "openfda": {}
    }
  ]
}