  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-08-20",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "35826",
      "product_res_number": "Z-0374-05",
      "event_date_initiated": "2004-11-05",
      "event_date_posted": "2004-12-30",
      "recall_status": "Terminated",
      "event_date_terminated": "2008-10-29",
      "res_event_number": "30414",
      "product_code": "BTE",
      "k_numbers": [
        "K942599"
      ],
      "product_description": "Warm Air Model 135 Warming Units.",
      "code_info": "All serial numbers.",
      "firm_fei_number": "1516825",
      "recalling_firm": "Cincinnati Sub-Zero Products",
      "address_1": "12011 Mosteller Rd",
      "city": "Cincinnati",
      "state": "OH",
      "postal_code": "45241-1528",
      "additional_info_contact": "Crystal Morrison\n513-772-8810",
      "reason_for_recall": "Burning smell, burned microprocessor board.  The firm is modifying the connectors and wiring gauge configuration of the device.",
      "root_cause_description": "Other",
      "action": "The firm notified consignees by certified letter on November 5, 2004.",
      "product_quantity": "649 units.",
      "distribution_pattern": "The product was distributed throughout the United States and to foreign countries including Saudi Arabia, Taiwan, Japan, and Canada.",
      "openfda": {}
    }
  ]
}