      "cfres_id": "35504",
      "product_res_number": "Z-0149-05",
      "event_date_initiated": "2004-10-08",
      "event_date_posted": "2004-11-02",
      "recall_status": "Terminated",
      "event_date_terminated": "2012-05-15",
      "res_event_number": "30196",
      "product_code": "FMI",
      "k_numbers": [
        "K012736"
      ],
      "product_description": "Argyle TURKEL Safety Paracentesis Procedure Tray \nProduct Number: 8888568006",
      "code_info": "All lots of the following codes with lot numbers in the range of 4086027 through 4275213 (inclusive):",
      "firm_fei_number": "1282497",
      "recalling_firm": "Kendall Healthcare Products Co",
      "address_1": "15 Hampshire St",
      "city": "Mansfield",
      "state": "MA",
      "postal_code": "02048-1113",
      "additional_info_contact": "James Welsh\n508-261-8532",
      "reason_for_recall": "Needle cannula may separate from the hub during use",
      "root_cause_description": "Other",
      "action": "Recall letters requesting product return were mailed to all customers on October 8, 2004",
      "distribution_pattern": "Nationwide",
      "openfda": {
        "k_number": [
          "K052445",
          "K943135",
          "K943626",
          "K222124",
          "K882880",
          "K002180",
          "K890926",
          "K945276",
          "K945273",
          "K182770",
          "K990371",
          "K920737",
          "K102328",
          "K140516",
          "K121908",
          "K250700",
          "K861152",
          "K130212",
          "K080904",
          "K061483",
          "K911037",
          "K132552",
          "K935628",
          "K841742",
          "K983067",
          "K192082",
          "K970370",
          "K121607",
          "K874430",
          "K924793",
          "K073127",
          "K945825",
          "K111424",
          "K981184",
          "K943771",
          "K152597",
          "K844313",
          "K231924",
          "K041399",
          "K141117",
          "K974363",
          "K211634",
          "K970737",
          "K231723",
          "K102584",
          "K962001",
          "K001175",
          "K031453",
          "K961328",
          "K191853",
          "K944680",
          "K172483",
          "K834553",
          "K082908",
          "K913885",
          "K173757",
          "K970383",
          "K183016",
          "K062956",
          "K100209",
          "K840967",
          "K823693",
          "K983750",
          "K943134",
          "K841577",
          "K895075",
          "K922158",
          "K000085",
          "K942930",
          "K090111",
          "K173479",
          "K935514",
          "K061032",
          "K173881",
          "K173279",
          "K925056",
          "K130097",
          "K202319",
          "K936235",
          "K946289",
          "K173803",
          "K210864",
          "K840334",
          "K992729",
          "K904198",
          "K160532",
          "K000657",
          "K210258",
          "K761098",
          "K202492",
          "K895705",
          "K943141",
          "K943575",
          "K152050",
          "K831395",
          "K791156",
          "K884211",
          "K853731",
          "K220890",
          "K033478",
          "K011004",
          "K242632",
          "K220129",
          "K111797",
          "K230635",
          "K002938",
          "K192222",
          "K171982",
          "K952274",
          "K213478",
          "K060007",
          "K063450",
          "K831610",
          "K000387",
          "K770370",
          "K944230",
          "K062940",
          "K943679",
          "K232165",
          "K771207",
          "K172095",
          "K925816",
          "K253047",
          "K113422",
          "K973620",
          "K913141",
          "K250724",
          "K924439",
          "K002213",
          "K041941",
          "K981853",
          "K102994",
          "K031200",
          "K852831",
          "K833705",
          "K260855",
          "K905305",
          "K980768",
          "K110312",
          "K872566",
          "K110850",
          "K022415",
          "K222417",
          "K964654",
          "K133738",
          "K112789",
          "K972705",
          "K911436",
          "K240355",
          "K830242",
          "K954041",
          "K050505",
          "K101026",
          "K222852",
          "K884139",
          "K895432",
          "K062608",
          "K161553",
          "K202681",
          "K960982",
          "K895630",
          "K982029",
          "K090156",
          "K170768",
          "K203792",
          "K842756",
          "K910455",
          "K960933",
          "K932889",
          "K963748",
          "K181447",
          "K770425",
          "K960641",
          "K020921",
          "K973012",
          "K850420",
          "K160805",
          "K220614",
          "K992699",
          "K010188",
          "K011925",
          "K945809",
          "K210748",
          "K001098",
          "K001690",
          "K071567",
          "K990936",
          "K960698",
          "K092746",
          "K003461",
          "K944931",
          "K153388",
          "K934404",
          "K911597",
          "K181369",
          "K243581",
          "K212015",
          "K860781",
          "K880892",
          "K213183",
          "K920717",
          "K000455",
          "K221080",
          "K855248",
          "K854548",
          "K791462",
          "K163093",
          "K842542",
          "K936261",
          "K813649",
          "K071998",
          "K792096",
          "K191643",
          "K874024",
          "K871142",
          "K922179",
          "K200017",
          "K212514",
          "K980227",
          "K992091",
          "K043249",
          "K102010",
          "K050464",
          "K943794",
          "K915409",
          "K210065",
          "K844305",
          "K813648",
          "K012283",
          "K854547",
          "K864310",
          "K945274",
          "K010887",
          "K250658",
          "K980003",
          "K221176",
          "K063142",
          "K240915",
          "K961878",
          "K962040",
          "K895769",
          "K841166",
          "K243309",
          "K962149",
          "K052843",
          "K052195",
          "K946084",
          "K120420",
          "K760780",
          "K900282",
          "K214004",
          "K180395",
          "K003827",
          "K896293",
          "K210232",
          "K980062",
          "K113469",
          "K970547",
          "K834436",
          "K211214",
          "K073705",
          "K813256",
          "K222681",
          "K925086",
          "K152877",
          "K160575",
          "K230043",
          "K943765",
          "K033409",
          "K853604",
          "K150263",
          "K943140",
          "K121632",
          "K921950",
          "K110656",
          "K905664",
          "K950159",
          "K943722",
          "K832858",
          "K131125",
          "K072602",
          "K210059",
          "K191644",
          "K190538",
          "K952837",
          "K942054",
          "K973339",
          "K943605",
          "K231720",
          "K935776",
          "K130487",
          "K170824",
          "K940961",
          "K935480",
          "K874414",
          "K943594",
          "K000369",
          "K063567",
          "K031058",
          "K091140",
          "K973911",
          "K950899",
          "K852615",
          "K123684",
          "K942073",
          "K890127",
          "K120191",
          "K972469",
          "K971763",
          "K870193",
          "K133059",
          "K100812",
          "K935916",
          "K180720",
          "K043490",
          "K863514",
          "K132583",
          "K250284",
          "K884212",
          "K943892",
          "K162516",
          "K061933",
          "K992626",
          "K931019",
          "K771203",
          "K834407",
          "K955772",
          "K071088",
          "K103587",
          "K131358",
          "K243806",
          "K925446",
          "K813545",
          "K923127",
          "K151571",
          "K902422",
          "K161255",
          "K945275",
          "K955235",
          "K874372",
          "K913329",
          "K961088",
          "K952518",
          "K181508",
          "K073188",
          "K051992",
          "K843264",
          "K142086",
          "K122249",
          "K003382",
          "K191488",
          "K072487",
          "K140568",
          "K891919",
          "K934088",
          "K872580",
          "K052474",
          "K854549",
          "K152362",
          "K801108",
          "K963571",
          "K073257",
          "K152339",
          "K972538",
          "K160199",
          "K972536",
          "K973899",
          "K181730",
          "K890473",
          "K052725",
          "K861318",
          "K832904",
          "K901162",
          "K823899",
          "K252886",
          "K202682",
          "K801431",
          "K223286",
          "K790953",
          "K100436",
          "K874602",
          "K042917",
          "K131123",
          "K021475",
          "K946027",
          "K092831",
          "K954500",
          "K841217",
          "K131040",
          "K192057",
          "K160887",
          "K896320",
          "K092016",
          "K881654",
          "K033575",
          "K052561",
          "K842012",
          "K780209",
          "K942179",
          "K961251",
          "K771667",
          "K252631",
          "K890620",
          "K211968",
          "K070440",
          "K910667",
          "K841306",
          "K013782",
          "K102774",
          "K861390",
          "K001572",
          "K071947",
          "K142765",
          "K230959",
          "K943485",
          "K082820",
          "K071785",
          "K043019",
          "K980179",
          "K151341",
          "K032885",
          "K072059",
          "K933274",
          "K913498",
          "K191976",
          "K944121",
          "K970398",
          "K885122",
          "K982097",
          "K946041",
          "K913914",
          "K944757",
          "K110105",
          "K180259",
          "K063466",
          "K840033",
          "K934591",
          "K152166",
          "K943683",
          "K170276",
          "K935843",
          "K981619",
          "K003521",
          "K122971",
          "K861406",
          "K943496",
          "K010038",
          "K941031",
          "K950012",
          "K904295",
          "K062500",
          "K260739",
          "K072320",
          "K254273",
          "K883009",
          "K942937",
          "K193422",
          "K051783",
          "K974006",
          "K961877",
          "K231261",
          "K151311",
          "K873165",
          "K950686",
          "K211293",
          "K011757",
          "K260933",
          "K925234",
          "K110703",
          "K823255",
          "K933674",
          "K032141",
          "K043397",
          "K251447",
          "K964042",
          "K222739",
          "K211716",
          "K951254",
          "K230070",
          "K012646",
          "K832556",
          "K881758",
          "K943057",
          "K841704",
          "K052408",
          "K832102",
          "K982541",
          "K943740",
          "K141749",
          "K951515",
          "K132233",
          "K960963",
          "K152824",
          "K022406",
          "K994197",
          "K152410",
          "K952834",
          "K935505",
          "K790086",
          "K133058",
          "K951736",
          "K003812",
          "K221178",
          "K000714",
          "K895811",
          "K952044",
          "K033373",
          "K830568",
          "K841305",
          "K820911",
          "K171144",
          "K090117",
          "K862610",
          "K231255",
          "K152183",
          "K801107",
          "K112332",
          "K231729",
          "K214075",
          "K873957",
          "K861081",
          "K993125",
          "K223334",
          "K863606",
          "K913317",
          "K042907",
          "K924820",
          "K971924",
          "K233343",
          "K942323",
          "K253622",
          "K001853",
          "K123300",
          "K113186",
          "K051899",
          "K895771",
          "K161170",
          "K223327",
          "K133635",
          "K831063",
          "K213156",
          "K935777",
          "K984576",
          "K831765",
          "K942901",
          "K944114",
          "K971487",
          "K011786",
          "K965248",
          "K812404",
          "K895024",
          "K242741",
          "K211395",
          "K955748",
          "K252908",
          "K845043",
          "K191258",
          "K040531",
          "K100005",
          "K073122",
          "K131954",
          "K132557",
          "K082511",
          "K180588",
          "K943137",
          "K030032",
          "K093043",
          "K963658",
          "K013293",
          "K152514",
          "K181069",
          "K935407",
          "K002403",
          "K233037",
          "K903490",
          "K861560",
          "K950466",
          "K913258",
          "K161950",
          "K050106",
          "K904107",
          "K133100",
          "K122969",
          "K151090",
          "K021315",
          "K062422",
          "K212805",
          "K011984",
          "K010958",
          "K222271",
          "K840944",
          "K910733",
          "K943578",
          "K002394",
          "K062984",
          "K811966",
          "K963351",
          "K112777",
          "K071307",
          "K934435",
          "K143437",
          "K873991",
          "K172938",
          "K930217",
          "K163578",
          "K210399",
          "K113505",
          "K182146",
          "K190962",
          "K924070",
          "K874249",
          "K093109",
          "K012736",
          "K024109",
          "K100124",
          "K894374",
          "K884620",
          "K933601",
          "K151398",
          "K933751",
          "K090426",
          "K945284",
          "K982399",
          "K880746",
          "K080865",
          "K131842",
          "K991904",
          "K942903",
          "K120690",
          "K894372",
          "K003381",
          "K991404",
          "K913386",
          "K182320",
          "K202340",
          "K213407",
          "K852442",
          "K022057",
          "K771209",
          "K221066",
          "K152334",
          "K882064",
          "K841565",
          "K970380",
          "K873569",
          "K150662",
          "K102108",
          "K943139",
          "K940645",
          "K202005",
          "K963595",
          "K881785",
          "K821529",
          "K972279",
          "K870863",
          "K812910",
          "K041153",
          "K801839",
          "K192464",
          "K053470",
          "K220603",
          "K110606",
          "K003406",
          "K001965",
          "K180545",
          "K261829",
          "K942624",
          "K935506",
          "K120661",
          "K823696",
          "K072247",
          "K153706",
          "K171984",
          "K943136",
          "K770528",
          "K110527",
          "K936029",
          "K212095",
          "K962035",
          "K170988",
          "K070988",
          "K000441",
          "K003627",
          "K851103",
          "K213535",
          "K080235",
          "K884792",
          "K223353",
          "K200932",
          "K963830",
          "K881671"
        ],
        "registration_number": [
          "",
          "9611680",
          "3027338968",
          "2320762",
          "3022699058",
          "3015859709",
          "3004054016",
          "3011088743",
          "2020394",
          "1000563940",
          "3006425876",
          "3010392991",
          "3018125942",
          "3012116434",
          "3003915875",
          "3005841027",
          "3003681915",
          "1223004",
          "8021764",
          "3014662855",
          "3014456066",
          "9612057",
          "9610847",
          "3011617315",
          "9681822",
          "3009756153",
          "9612030",
          "9681835",
          "3007384659",
          "3003895440",
          "1313700",
          "3030626857",
          "3014649053",
          "3031233007",
          "3004416208",
          "1649518",
          "3005798905",
          "3006371704",
          "3003310007",
          "3007441619",
          "3015142721",
          "3013481097",
          "1213680",
          "2032098",
          "3000247873",
          "9680222",
          "2025816",
          "3010621964",
          "3027332508",
          "1055927",
          "2243072",
          "2427500",
          "3033536319",
          "3007007790",
          "3010384323",
          "3006850123",
          "3016884191",
          "9614868",
          "9610825",
          "1030451",
          "3008780388",
          "3014747754",
          "3009578786",
          "3003120666",
          "3007895168",
          "3034667025",
          "3013557562",
          "3008102042",
          "3018507299",
          "3012039263",
          "3019944216",
          "3020294761",
          "3030446844",
          "3030714660",
          "3012502601",
          "9680794",
          "2032112",
          "9616791",
          "3010085632",
          "1836161",
          "2183744",
          "9680904",
          "3012187743",
          "3002807350",
          "1710034",
          "3010872792",
          "3015528193",
          "3011137372",
          "3023316",
          "2433012",
          "9617604",
          "3010395155",
          "3002808227",
          "3010515547",
          "3010132111",
          "3017210488",
          "3017770710",
          "3042370483",
          "3018681450",
          "1722458",
          "3006999064",
          "2133772",
          "3023274465",
          "3011198121",
          "3008700497",
          "3038356900",
          "3009498536",
          "3012494290",
          "3010026090",
          "3009888344",
          "3012354669",
          "3010155648",
          "1420054",
          "9610849",
          "8040619",
          "3013429260",
          "3005288183",
          "3024508819",
          "3003902955",
          "3008588205",
          "9681413",
          "3007774504",
          "1058955",
          "1035907",
          "3009504230",
          "1066741",
          "3010280735",
          "3012809732",
          "2024024",
          "3004504731",
          "3024088964",
          "3026211706",
          "3010370650",
          "3009756327",
          "3005246939",
          "3013160284",
          "3007392007",
          "9681387",
          "3045891669",
          "3029328631",
          "3012944064",
          "1319639",
          "3002807315",
          "3018094310",
          "3029755538",
          "3016525500",
          "1043729",
          "9613662",
          "3009328501",
          "3005609751",
          "3014315664",
          "3007590959",
          "3021566676",
          "3010864832",
          "3009498548",
          "3006082230",
          "3015727296",
          "1045254",
          "3004458481",
          "9616157",
          "1043646",
          "3008272066",
          "3007097064",
          "3004417597",
          "3034669683",
          "9615588",
          "1017768",
          "3010220595",
          "3006762440",
          "3010175289",
          "1213809",
          "3007508552",
          "1410248",
          "1623486",
          "3021023752",
          "3027605735",
          "3013764800",
          "3015997711",
          "1450392",
          "2244771",
          "3005674478",
          "3043709866",
          "1220477",
          "1058584",
          "2523676",
          "3021186226",
          "3015140199",
          "3010041511",
          "3017494520",
          "3009077131",
          "3015620563",
          "3013496867",
          "1625685",
          "3013188547",
          "3011689956",
          "1058602",
          "3004961578",
          "3019004087",
          "3005670221",
          "3014654360",
          "2521402",
          "1217831",
          "3021169199",
          "3038110468",
          "3008729549",
          "3004496829",
          "3012646696",
          "1061124",
          "2027487",
          "8043792",
          "3030447506",
          "3030078769",
          "3010830834",
          "3015485283",
          "3012050423",
          "3023359743",
          "3012552122",
          "3029905131",
          "3036722184",
          "3031259044",
          "3010160527",
          "3004748528",
          "3003589911",
          "3010405396",
          "1219611",
          "3014579161",
          "2032810",
          "3009495876",
          "1056186",
          "3025432964",
          "3030733800",
          "3012478173",
          "9614033",
          "3009761558",
          "3013221738",
          "1000118068",
          "3015060216",
          "3012410077",
          "3008009298",
          "3009746425",
          "1036836",
          "3003464075",
          "8010418",
          "1018233",
          "3010413216",
          "3007817041",
          "1018470",
          "3009746061",
          "3013187142",
          "3030089798",
          "1526863",
          "3043226252",
          "1720499",
          "3008381300",
          "3041276591",
          "3013517171",
          "3020831613",
          "3009176340",
          "3003542872",
          "9616482",
          "3008307705",
          "8020040",
          "3023228837",
          "3004143450",
          "3021286079",
          "1054241",
          "3006787036",
          "3000126234",
          "3044026004",
          "3005941719",
          "3003974574",
          "3035642068",
          "3003829571",
          "3011676762",
          "3011453824",
          "1721676",
          "3007279404",
          "3004127962",
          "3016844398",
          "8020347",
          "1055236",
          "1064858",
          "3012461127",
          "9616567",
          "3005405146",
          "1647137",
          "3004111573",
          "1423507",
          "3000268902",
          "3003824562",
          "3030201489",
          "9617594",
          "3009742443",
          "1000113657",
          "3011237770",
          "3002807408",
          "3013160357",
          "3043051314",
          "3020298301",
          "3006948745",
          "9614279",
          "3043531632",
          "2031093",
          "3035111606",
          "3002806603",
          "3010131137",
          "3010966701",
          "3004468406",
          "3001124136",
          "9612086",
          "3013298431",
          "3002806821",
          "3014165712",
          "1930034",
          "1417592",
          "1048735",
          "1423625",
          "3007829657",
          "1928237",
          "3008719222",
          "8010026",
          "3010223109",
          "3012421607",
          "9616074",
          "3010462274",
          "2183771",
          "1000484830",
          "1417485",
          "3007207906",
          "3026478941",
          "3017636737",
          "9611446",
          "3003916417",
          "3030419987",
          "1221765",
          "3001728571",
          "3020460367",
          "1625425",
          "3017509841",
          "3012551896",
          "3014656668",
          "3006142663",
          "3006260740",
          "9617592",
          "1423537",
          "3015724078",
          "3020138744",
          "3042953454",
          "3012438919",
          "3003348846",
          "1000523114",
          "3008449424",
          "8043385",
          "3008789114",
          "3007766601",
          "3014421917",
          "3008729892",
          "3030446758",
          "3014393881",
          "3016904853",
          "3004192065",
          "3015042030",
          "3038294399",
          "1644312",
          "3029901632",
          "3035387162",
          "3004425647",
          "3020976354",
          "3014312726",
          "1222780",
          "3016685880",
          "3015058854",
          "3013633270",
          "1034525",
          "3007939342",
          "3010934635",
          "3015176887",
          "3010377171",
          "1224960",
          "1923569",
          "3010419931",
          "1915484",
          "3000214934",
          "3002682307",
          "3019012929",
          "3004055392",
          "3012104670",
          "1047429",
          "3014680791",
          "1319211",
          "3006942524",
          "1911916",
          "3017654330",
          "8041187",
          "3008998256",
          "8041145",
          "1043214",
          "1063851",
          "9612501",
          "1721183",
          "3011287770",
          "1820334",
          "1125230",
          "1832816",
          "1220905",
          "3014570560",
          "3003895930",
          "3017717188",
          "3009518015",
          "3005202235",
          "3010055973",
          "9681821",
          "1921846",
          "3007617202",
          "3036679495",
          "3006380458",
          "3012854006",
          "3005654106",
          "3017907716",
          "3015176914",
          "3009392898",
          "3004099493",
          "2011171",
          "1721686",
          "3014479313",
          "3004138517",
          "3007344957",
          "3003902757",
          "3002907620",
          "2245270",
          "8020785",
          "3013162291",
          "3023339256",
          "3015142802",
          "3005669815",
          "9615908",
          "1450743",
          "1066444",
          "9613304",
          "1035236",
          "3004605321",
          "3009853488",
          "3014819517",
          "3015955792",
          "3003589279",
          "3012680643",
          "3006973666"
        ],
        "fei_number": [
          "",
          "3027338968",
          "3022699058",
          "3015859709",
          "3004054016",
          "3011088743",
          "2020394",
          "1000563940",
          "3006425876",
          "3010392991",
          "3018125942",
          "3012116434",
          "3003915875",
          "3005841027",
          "3003681915",
          "3000974732",
          "3014662855",
          "3014456066",
          "3002807111",
          "3011617315",
          "3002654698",
          "3009756153",
          "3007384659",
          "3003895440",
          "1313700",
          "3030626857",
          "3014649053",
          "3031233007",
          "3004416208",
          "3005798905",
          "3006371704",
          "3002806575",
          "3003310007",
          "3007441619",
          "3015142721",
          "3013481097",
          "1213680",
          "3000247873",
          "2025816",
          "3010621964",
          "3027332508",
          "2243072",
          "2427500",
          "3033536319",
          "3007007790",
          "3010384323",
          "3006850123",
          "3016884191",
          "3003910417",
          "1030451",
          "3008780388",
          "3014747754",
          "3005612924",
          "3009578786",
          "3003120666",
          "3007895168",
          "3034667025",
          "3008002452",
          "3013557562",
          "3018507299",
          "3012039263",
          "3002468371",
          "3002807555",
          "3019944216",
          "1000138386",
          "3020294761",
          "3030446844",
          "3030714660",
          "3012502601",
          "3010085632",
          "2183744",
          "3012187743",
          "3002807350",
          "1710034",
          "3010872792",
          "3015528193",
          "3003495548",
          "3023316",
          "2433012",
          "3010395155",
          "3003752255",
          "3002808227",
          "3010515547",
          "3010132111",
          "3017210488",
          "3017770710",
          "3042370483",
          "3018681450",
          "3006999064",
          "3023274465",
          "3011198121",
          "3008700497",
          "3038356900",
          "3009498536",
          "3012494290",
          "3010026090",
          "3003005156",
          "3009888344",
          "3012354669",
          "3010155648",
          "1420054",
          "3005288183",
          "3024508819",
          "3003902955",
          "3008588205",
          "3002919051",
          "2828",
          "3007774504",
          "1035907",
          "3009504230",
          "3010862763",
          "3010280735",
          "3012809732",
          "2024024",
          "3004504731",
          "3024088964",
          "3026211706",
          "3005246939",
          "3013160284",
          "3007392007",
          "3004169099",
          "3045891669",
          "3029328631",
          "3012944064",
          "3018094310",
          "3002807315",
          "3029755538",
          "3016525500",
          "1043729",
          "3002808401",
          "3002558692",
          "1000117748",
          "3009328501",
          "3005609751",
          "3014315664",
          "3007590959",
          "3021566676",
          "3009498548",
          "3006082230",
          "3015727296",
          "1045254",
          "3004458481",
          "3004521048",
          "3002924544",
          "1043646",
          "3008272066",
          "3003188701",
          "3007097064",
          "3004417597",
          "3034669683",
          "1017768",
          "3010220595",
          "3006762440",
          "3010175289",
          "1213809",
          "3007508552",
          "3003405786",
          "1410248",
          "1623486",
          "3021023752",
          "3027605735",
          "3013764800",
          "3015997711",
          "1450392",
          "2244771",
          "3005674478",
          "3043709866",
          "1220477",
          "2523676",
          "3021186226",
          "3015140199",
          "3010041511",
          "3017494520",
          "1000112968",
          "3009077131",
          "3015620563",
          "1000174591",
          "3013496867",
          "1625685",
          "3013188547",
          "3011689956",
          "3004961578",
          "3019004087",
          "3005670221",
          "3014654360",
          "1000306068",
          "2521402",
          "1217831",
          "3021169199",
          "3038110468",
          "3008729549",
          "3004496829",
          "3012646696",
          "2027487",
          "3002807969",
          "3030447506",
          "3030078769",
          "3015485283",
          "3012050423",
          "3023359743",
          "1000132332",
          "3012552122",
          "3029905131",
          "3036722184",
          "3031259044",
          "3004748528",
          "3003589911",
          "3010405396",
          "1219611",
          "3014579161",
          "3009495876",
          "3025432964",
          "3030733800",
          "3012478173",
          "3009761558",
          "3013221738",
          "1000118068",
          "3015060216",
          "3012410077",
          "3008009298",
          "3009746425",
          "1036836",
          "3002963732",
          "3009291251",
          "3003282181",
          "3003464075",
          "1018233",
          "3010413216",
          "3007817041",
          "1018470",
          "3009746061",
          "3013187142",
          "3030089798",
          "1526863",
          "3000209794",
          "3043226252",
          "1720499",
          "3008381300",
          "3041276591",
          "3013517171",
          "3020831613",
          "3009176340",
          "3003542872",
          "3008307705",
          "3023228837",
          "3001236277",
          "3004143450",
          "3021286079",
          "3003040625",
          "3006787036",
          "1000138447",
          "1000138054",
          "3000126234",
          "3044026004",
          "3001570557",
          "3003974574",
          "3035642068",
          "3003829571",
          "3011676762",
          "3011453824",
          "1721676",
          "3001316486",
          "3007279404",
          "3002750084",
          "3016844398",
          "3002807040",
          "3004127962",
          "3012461127",
          "3003076831",
          "3000210101",
          "1000219614",
          "3005405146",
          "3004111573",
          "3000268902",
          "3003824562",
          "3030201489",
          "3009742443",
          "1000113657",
          "3011237770",
          "1000512168",
          "3002807408",
          "3013160357",
          "3043051314",
          "1000112877",
          "3020298301",
          "3006948745",
          "3001239292",
          "3043531632",
          "3035111606",
          "3002806603",
          "3010131137",
          "3010966701",
          "3004468406",
          "3001124136",
          "3003131673",
          "3013298431",
          "3002806821",
          "1000442352",
          "3014165712",
          "1930034",
          "1417592",
          "3008397795",
          "1048735",
          "3007829657",
          "1928237",
          "3008719222",
          "3010223109",
          "3012421607",
          "3010462274",
          "2183771",
          "1000484830",
          "1417485",
          "3007207906",
          "3004547341",
          "3017636737",
          "3026478941",
          "3003737899",
          "3003169040",
          "3030419987",
          "1221765",
          "3001728571",
          "3004525092",
          "3020460367",
          "1625425",
          "3002807038",
          "3017509841",
          "3012551896",
          "3007130719",
          "3014656668",
          "3006142663",
          "3006260740",
          "3000204727",
          "3002808269",
          "3015724078",
          "3002807391",
          "3020138744",
          "3042953454",
          "3012438919",
          "3003348846",
          "1000523114",
          "3003560965",
          "3008449424",
          "3008789114",
          "3007766601",
          "3014421917",
          "3002879021",
          "3030446758",
          "3001451379",
          "3014393881",
          "3016904853",
          "3004192065",
          "3003058448",
          "3005747797",
          "3003807466",
          "3002807725",
          "3015042030",
          "3038294399",
          "3029901632",
          "3035387162",
          "3004425647",
          "1000016009",
          "3010165944",
          "3020976354",
          "3014312726",
          "3016685880",
          "3015058854",
          "3004034544",
          "3013633270",
          "3007939342",
          "1000120743",
          "3010934635",
          "3015176887",
          "3010377171",
          "3007019930",
          "1923569",
          "3010419931",
          "1000120555",
          "1915484",
          "3000214934",
          "3002806460",
          "3002682307",
          "1000219637",
          "3019012929",
          "3004055392",
          "3012104670",
          "1047429",
          "3014680791",
          "1319211",
          "3001236549",
          "3002808190",
          "3003767881",
          "3006942524",
          "1911916",
          "3017654330",
          "3008998256",
          "3001236905",
          "1043214",
          "3011287770",
          "1000113359",
          "1820334",
          "1832816",
          "1220905",
          "3014570560",
          "3009518015",
          "3017717188",
          "3005202235",
          "3010055973",
          "3002806934",
          "1921846",
          "3003154916",
          "3007617202",
          "3036679495",
          "1000525253",
          "3006380458",
          "3012854006",
          "3005654106",
          "3017907716",
          "3015176914",
          "3009392898",
          "3004099493",
          "2011171",
          "1721686",
          "3014479313",
          "3004138517",
          "3007344957",
          "3011649541",
          "3002926090",
          "3003902757",
          "3002907620",
          "2245270",
          "3002974635",
          "3002806469",
          "3013162291",
          "3023339256",
          "3015142802",
          "3000191104",
          "1000232458",
          "3005669815",
          "1450743",
          "1035236",
          "3004605321",
          "3009853488",
          "3014819517",
          "3015955792",
          "3003589279",
          "3012680643",
          "3006973666"
        ],
        "device_name": "Needle, Hypodermic, Single Lumen",
        "medical_specialty_description": "General Hospital",
        "regulation_number": "880.5570",
        "device_class": "2"
      }
    }
  ]
}