  "results": [
    {
      "cfres_id": "35202",
      "product_res_number": "Z-0059-05",
      "event_date_initiated": "2004-09-27",
      "event_date_posted": "2004-10-28",
      "recall_status": "Terminated",
      "event_date_terminated": "2004-11-30",
      "res_event_number": "30115",
      "product_code": "FOZ",
      "k_numbers": [
        "K993691"
      ],
      "product_description": "Arrowg+ard Blue Two Lumen CVC Super Kit, Catalog number AK-22802-SK. The kit actually contains Arrowg+ard Blue Plus Two Lumen CVC Super Kit. Catalog number AK-42802-SK, 8 Fr x 16 cm Two Lumen Arrowg+ard Blue Plus Catheters. The plus indicates that the line extension clamps are attached to the catheters.",
      "code_info": "Lot number RF4066579",
      "firm_fei_number": "3015859709",
      "recalling_firm": "Arrow International Inc",
      "address_1": "2400 Bernville Rd",
      "city": "Reading",
      "state": "PA",
      "postal_code": "19605-9607",
      "additional_info_contact": "Jody Cadd\n610-378-0131 Ext. 3318",
      "reason_for_recall": "Labeling error. Kits contains the correct components for AK-42802-SK but may have a package label indicating that it is product number AK-22802-SK",
      "root_cause_description": "Other",
      "action": "The recalling firm issued a recall letter dated 9/27/04 to their accounts informing them of the problem and the need to return the product.",
      "product_quantity": "290 kits",
      "distribution_pattern": "The products were shipped to distributors in KS and VA, medical facilities in AR, FL, and GA, and a sales representative in FL.",
      "openfda": {
        "k_number": [
          "K231239",
          "K853701",
          "K081113",
          "K063469",
          "K772354",
          "K251422",
          "K250292",
          "K790968",
          "K100775",
          "K910564",
          "K191232",
          "K880260",
          "K943023",
          "K081953",
          "K171530",
          "K910565",
          "K940775",
          "K800786",
          "K081690",
          "K854566",
          "K161777",
          "K770403",
          "K834480",
          "K834255",
          "K182870",
          "K962226",
          "K834227",
          "K890435",
          "K112290",
          "K984614",
          "K884259",
          "K130518",
          "K032274",
          "K203069",
          "K231401",
          "K833657",
          "K220626",
          "K121073",
          "K970669",
          "K070451",
          "K102520",
          "K042862",
          "K161866",
          "K944660",
          "K032777",
          "K182538",
          "K851142",
          "K921270",
          "K872998",
          "K801537",
          "K954112",
          "K813345",
          "K872585",
          "K901701",
          "K864038",
          "K861725",
          "K002033",
          "K914528",
          "K022099",
          "K042504",
          "K974764",
          "K955053",
          "K850349",
          "K173354",
          "K875177",
          "K883980",
          "K860474",
          "K172496",
          "K120337",
          "K832624",
          "K133856",
          "K873050",
          "K150109",
          "K960764",
          "K151513",
          "K791918",
          "K953805",
          "K220756",
          "K932789",
          "K051355",
          "K992198",
          "K862153",
          "K940441",
          "K853988",
          "K833079",
          "K201075",
          "K813396",
          "K200373",
          "K932071",
          "K011297",
          "K000087",
          "K840862",
          "K140419",
          "K844193",
          "K862130",
          "K933050",
          "K802918",
          "K945521",
          "K131555",
          "K834393",
          "K241115",
          "K192676",
          "K940634",
          "K070020",
          "K823017",
          "K883415",
          "K931027",
          "K003553",
          "K111939",
          "K904121",
          "K013556",
          "K920828",
          "K941678",
          "K011937",
          "K944410",
          "K001616",
          "K801941",
          "K920635",
          "K905463",
          "K251155",
          "K890751",
          "K944504",
          "K060174",
          "K932947",
          "K944368",
          "K122544",
          "K140504",
          "K842577",
          "K082362",
          "K190001",
          "K893003",
          "K912165",
          "K032292",
          "K170336",
          "K912158",
          "K072419",
          "K040427",
          "K904558",
          "K100282",
          "K160374",
          "K902527",
          "K945053",
          "K770751",
          "K172204",
          "K984189",
          "K993691",
          "K163358",
          "K891087",
          "K950737",
          "K760319",
          "K895191",
          "K830539",
          "K946170",
          "K090502",
          "K220584",
          "K980493",
          "K230616",
          "K870403",
          "K952861",
          "K153280",
          "K082997",
          "K020785",
          "K811171",
          "K981368",
          "K943349",
          "K993518",
          "K884701",
          "K052564",
          "K112542",
          "K984590",
          "K850887",
          "K111366",
          "K260539",
          "K050114",
          "K151698",
          "K894611",
          "K150407",
          "K953811",
          "K844219",
          "K212370",
          "K983916",
          "K896774",
          "K030479",
          "K033500",
          "K923702",
          "K875082",
          "K251654",
          "K951620",
          "K253235",
          "K121488",
          "K140136",
          "K192230",
          "K901263",
          "K110443",
          "K926097",
          "K930048",
          "K850276",
          "K000665",
          "K830541",
          "K162377",
          "K864949",
          "K984055",
          "K160235",
          "K941417",
          "K120839",
          "K861510",
          "K972254",
          "K150611",
          "K052881",
          "K123267",
          "K833092",
          "K010034",
          "K881767",
          "K161779",
          "K261041",
          "K252137",
          "K881976",
          "K770181",
          "K073471",
          "K920133",
          "K002142",
          "K952645",
          "K812159",
          "K940385",
          "K113429",
          "K851101",
          "K003187",
          "K915824",
          "K895734",
          "K972262",
          "K021139",
          "K931729",
          "K971990",
          "K011982",
          "K021557",
          "K955839",
          "K920028",
          "K982805",
          "K010021",
          "K910572",
          "K854308",
          "K092629",
          "K022756",
          "K932946",
          "K240359",
          "K884325",
          "K895942",
          "K942045",
          "K921955",
          "K172506",
          "K011860",
          "K013800",
          "K984059",
          "K832558",
          "K801575",
          "K840183",
          "K992732",
          "K962938",
          "K233529",
          "K011967",
          "K964049",
          "K990390",
          "K073241",
          "K193158",
          "K010410",
          "K972263",
          "K862480",
          "K910566",
          "K821372",
          "K981382",
          "K161681",
          "K243105",
          "K771573",
          "K852078",
          "K851327",
          "K830765",
          "K926544",
          "K850278",
          "K911251",
          "K970736",
          "K770510",
          "K113700",
          "K900116",
          "K840355",
          "K822776",
          "K905415",
          "K013287",
          "K960013",
          "K881215",
          "K060341",
          "K143610",
          "K953120",
          "K760163",
          "K912358",
          "K991406",
          "K903347",
          "K895597",
          "K923323",
          "K771381",
          "K823565",
          "K931073",
          "K070705",
          "K822023",
          "K100974",
          "K924968",
          "K982558",
          "K043446",
          "K821371",
          "K944302",
          "K944367",
          "K010406",
          "K202060",
          "K780350",
          "K850369",
          "K980566",
          "K033853",
          "K861015",
          "K162894",
          "K901640",
          "K952256",
          "K141138",
          "K821679",
          "K920638",
          "K801414",
          "K772011",
          "K833091",
          "K874582",
          "K041345",
          "K770511",
          "K071538",
          "K964050",
          "K182252",
          "K912563",
          "K000235",
          "K883866",
          "K003245",
          "K200266",
          "K080793",
          "K200891",
          "K192493",
          "K894371",
          "K974543",
          "K833135",
          "K920128",
          "K904150",
          "K241845",
          "K851981",
          "K113565",
          "K843033",
          "K242281",
          "K950301",
          "K111236",
          "K052389",
          "K190855",
          "K952757",
          "K232234",
          "K083454",
          "K931668",
          "K910799",
          "K944336",
          "K061250",
          "K152272",
          "K193278",
          "K252677",
          "K935090",
          "K803238",
          "K950284",
          "K213664",
          "K821577",
          "K813374",
          "K112347",
          "K142136",
          "K962577",
          "K990236",
          "K903545",
          "K243403",
          "K926405",
          "K160592",
          "K992532",
          "K091709",
          "K200263",
          "K944968",
          "K821929",
          "K032417",
          "K893546",
          "K840815",
          "K840116",
          "K001795",
          "K993515",
          "K823351",
          "K970644",
          "K981905",
          "K141026",
          "K002786",
          "K893965",
          "K024007",
          "K961552",
          "K000632",
          "K954428",
          "K844113",
          "K780105",
          "K213085",
          "K090973",
          "K163513",
          "K963196",
          "K201717",
          "K963867",
          "K132153",
          "K983409",
          "K830972",
          "K812137",
          "K842133",
          "K123734",
          "K904648",
          "K093546",
          "K790967",
          "K013182",
          "K790914",
          "K221827",
          "K110794",
          "K153298",
          "K220772",
          "K032843",
          "K121007",
          "K914378",
          "K183399",
          "K053120",
          "K964966",
          "K881236",
          "K952058",
          "K113622",
          "K223788",
          "K142797",
          "K192533",
          "K233106",
          "K974015",
          "K033843",
          "K950118",
          "K903732",
          "K860256",
          "K210037",
          "K033250",
          "K954887",
          "K223648",
          "K882242",
          "K905009",
          "K972264",
          "K971991",
          "K073520",
          "K953252",
          "K900056",
          "K895812",
          "K231626",
          "K914180",
          "K970864",
          "K960051",
          "K933467",
          "K252513",
          "K013073",
          "K252398",
          "K090122",
          "K133280",
          "K884806",
          "K030571",
          "K243599",
          "K083429",
          "K032962",
          "K800525",
          "K834073",
          "K861117",
          "K830573",
          "K250682",
          "K071382",
          "K895768",
          "K140270",
          "K884577",
          "K023253",
          "K990308",
          "K841880",
          "K953335",
          "K081420",
          "K964048",
          "K896813",
          "K944511",
          "K860290",
          "K971339",
          "K904122",
          "K021130",
          "K871546",
          "K244059",
          "K853348",
          "K821699",
          "K850398",
          "K850460",
          "K973561",
          "K170283",
          "K223018",
          "K012128",
          "K861479",
          "K902135",
          "K915065",
          "K900263",
          "K000997",
          "K881666",
          "K062108",
          "K210929"
        ],
        "registration_number": [
          "3013764800",
          "",
          "3030733800",
          "9611446",
          "3012494290",
          "1220477",
          "1220905",
          "2245270",
          "9617594",
          "9610849",
          "3009211636",
          "8020889",
          "3014270673",
          "3035111606",
          "3013188547",
          "9610048",
          "1054241",
          "3010621964",
          "2032112",
          "3022518300",
          "9680091",
          "2030624",
          "3008496528",
          "1056186",
          "8020785",
          "9612030",
          "2029015",
          "2320762",
          "3013557562",
          "8021545",
          "3010413216",
          "3002807090",
          "3008403546",
          "3003916417",
          "3036773433",
          "3011241266",
          "1928237",
          "3037918935",
          "3004111573",
          "3012050423",
          "9615908",
          "1055927",
          "8041145",
          "1319211",
          "3012026312",
          "2032098",
          "3002807314",
          "3010882065",
          "3014663754",
          "1721676",
          "3013162291",
          "2024311",
          "1282497",
          "3003442380",
          "2011171",
          "3010384323",
          "3013517171",
          "3012809732",
          "3004785273",
          "3005956311",
          "9617592",
          "1219611",
          "1061124",
          "1921846",
          "9681835",
          "3009750712",
          "3006846316",
          "9616991",
          "3015173212",
          "3011237770",
          "3007616112",
          "1644312",
          "1721686",
          "3003902955",
          "1713468",
          "3004129394",
          "3003674698",
          "3014263632",
          "3005246939",
          "3010131137",
          "3020460367",
          "9680143",
          "2025816",
          "3032916632",
          "9612126",
          "3030626857",
          "1417592",
          "3006621386",
          "9616157",
          "1123137",
          "3007836437",
          "3016904853",
          "9616088",
          "3015997711",
          "3011088743",
          "3013319212",
          "3017129766",
          "1423507",
          "3042370483",
          "3003915875",
          "3001124136",
          "3033536319",
          "9612433",
          "1423537",
          "3012421607",
          "3010934635",
          "3031983053",
          "1647137",
          "1064858",
          "3009758019",
          "3014656749",
          "3010754844",
          "3021570645",
          "3011689956",
          "3004102031",
          "2523676",
          "9616567",
          "1000563940",
          "1625425",
          "1043214",
          "3030714660",
          "3004187702",
          "3006942602",
          "2024024",
          "3010041511",
          "3012039263",
          "1710034",
          "3030446844",
          "3022728142",
          "3014389131",
          "3004784537",
          "3012542015",
          "2527072",
          "3010220595",
          "1824619",
          "3010419931",
          "9610825",
          "3012187635",
          "1035907",
          "8043983",
          "2246552",
          "9610847",
          "3006082230",
          "3002601200",
          "3008410596",
          "1018470",
          "3030574705",
          "3010055973",
          "3010532612",
          "3004519921",
          "3009888344",
          "3005515211",
          "9615588",
          "3015724078",
          "3003184529",
          "3008307705",
          "3015859709",
          "3014579161",
          "3004134316",
          "3043226252",
          "3005987240",
          "3006260740",
          "3022054170",
          "1055236",
          "2243072",
          "3011137372",
          "8041187",
          "3002807350",
          "3002806603",
          "9617604",
          "3006425876",
          "3005941719",
          "1410248",
          "3015058854",
          "1018233",
          "9612086",
          "3013298431",
          "3004122598",
          "3007207906",
          "3016678045",
          "9610987",
          "3031233007",
          "1319639",
          "3012104670",
          "8010026",
          "2518902",
          "1422634",
          "1223004",
          "3000247873",
          "9680794",
          "3009291251",
          "3018507299",
          "3003775072",
          "3005609751",
          "3017509841",
          "3000268902",
          "3003895440",
          "3016727301",
          "1820334",
          "2648045"
        ],
        "fei_number": [
          "3013764800",
          "",
          "3030733800",
          "1000174591",
          "3012494290",
          "3003076831",
          "1220905",
          "1220477",
          "2245270",
          "3009211636",
          "1000160458",
          "3002806482",
          "3002806830",
          "3014270673",
          "3002468371",
          "3035111606",
          "3003282181",
          "3013188547",
          "3003382061",
          "3010621964",
          "3003560965",
          "3022518300",
          "3003169040",
          "3008496528",
          "3013557562",
          "3002974635",
          "3002806575",
          "3005747797",
          "1000117748",
          "3010413216",
          "3008403546",
          "3036773433",
          "3011241266",
          "1928237",
          "3037918935",
          "3004111573",
          "3012050423",
          "1000306068",
          "1319211",
          "3012026312",
          "3002807314",
          "1000138054",
          "3014663754",
          "1721676",
          "3003752255",
          "1000113359",
          "3013162291",
          "2024311",
          "1282497",
          "3003442380",
          "2011171",
          "3010384323",
          "3013517171",
          "3012809732",
          "3004785273",
          "3005956311",
          "1219611",
          "1000514656",
          "1921846",
          "3009750712",
          "3016261803",
          "3006846316",
          "3015173212",
          "3008002452",
          "3011237770",
          "3007616112",
          "1721686",
          "3003902955",
          "1713468",
          "3004129394",
          "3003674698",
          "3014263632",
          "1000132332",
          "3003040625",
          "3002806469",
          "3005246939",
          "3010131137",
          "3020460367",
          "3005846326",
          "2025816",
          "3032916632",
          "3030626857",
          "1417592",
          "1000112877",
          "3006621386",
          "3007836437",
          "3016904853",
          "3008397795",
          "3013319212",
          "3015997711",
          "3011088743",
          "3042370483",
          "3017129766",
          "3006003617",
          "3003915875",
          "3001124136",
          "3002807548",
          "3033536319",
          "3012421607",
          "3010934635",
          "3031983053",
          "3002807391",
          "3009758019",
          "3014656749",
          "3010754844",
          "3021570645",
          "3002902735",
          "3011689956",
          "3004102031",
          "2523676",
          "3001236905",
          "3002807308",
          "1000563940",
          "1625425",
          "1043214",
          "3030714660",
          "3004187702",
          "3006942602",
          "2024024",
          "3010041511",
          "3012039263",
          "1710034",
          "3030446844",
          "3022728142",
          "3014389131",
          "3004784537",
          "3012542015",
          "3003910417",
          "1000512168",
          "3010220595",
          "3010419931",
          "1824619",
          "2527072",
          "3012187635",
          "1035907",
          "2246552",
          "3002806934",
          "3006082230",
          "3002601200",
          "3008410596",
          "1018470",
          "3003807466",
          "3030574705",
          "3010055973",
          "3010532612",
          "3003058448",
          "3009888344",
          "3005515211",
          "3015724078",
          "3003184529",
          "3008307705",
          "3015859709",
          "3014579161",
          "3004134316",
          "3043226252",
          "3005987240",
          "3006260740",
          "3022054170",
          "1000123180",
          "2243072",
          "1000112968",
          "3002807350",
          "3002806603",
          "3006097687",
          "3006425876",
          "3002808464",
          "1410248",
          "3002879021",
          "3015058854",
          "1000120555",
          "1018233",
          "3013298431",
          "1000477110",
          "3004122598",
          "3007207906",
          "3003205539",
          "3002807111",
          "3016678045",
          "3031233007",
          "3012104670",
          "3003737899",
          "2518902",
          "3002807710",
          "3002963732",
          "3000247873",
          "3009291251",
          "3018507299",
          "2828",
          "3003775072",
          "3005609751",
          "3017509841",
          "3000268902",
          "3003895440",
          "3002807725",
          "3016727301",
          "3002558692",
          "1820334",
          "2648045",
          "3000206435"
        ],
        "device_name": "Catheter, Intravascular, Therapeutic, Short-Term Less Than 30 Days",
        "medical_specialty_description": "General Hospital",
        "regulation_number": "880.5200",
        "device_class": "2"
      }
    }
  ]
}