      "cfres_id": "31851",
      "product_res_number": "Z-0655-4",
      "event_date_initiated": "2004-03-01",
      "event_date_posted": "2004-07-20",
      "recall_status": "Terminated",
      "event_date_terminated": "2008-10-24",
      "res_event_number": "28421",
      "product_code": "KNT",
      "product_description": "Ross Hide-A-Port Low Profile Balloon Gastrostomy Tube Kit, 16 Fr., 0.8 cm length, Item Numbers: 54146 and 55628.",
      "code_info": "Lot Numbers: 72038GZ00, 78315GZ00, 80399GZ00,  75201GZ00, 78308GZ00, 79358GZ00.",
      "firm_fei_number": "1000136098",
      "recalling_firm": "Ross Products Division Abbott Laboratories",
      "address_1": "625 Cleveland Ave",
      "city": "Columbus",
      "state": "OH",
      "postal_code": "43215-1754",
      "additional_info_contact": "614-624-3688",
      "reason_for_recall": "The balloons may be defective as indicated by an increased number of early balloon failures, which may result in tube displacement from patient.",
      "root_cause_description": "Other",
      "action": "The firm notified their customers on 3/1/2004 by telephone, fax, and FedEx.  Product will be returned.",
      "distribution_pattern": "Nationwide; Australia, New Zealand, Italy.",
      "openfda": {
        "k_number": [
          "K800918",
          "K924697",
          "K903579",
          "K911897",
          "K853253",
          "K840936",
          "K895246",
          "K844789",
          "K871019",
          "K872921",
          "K020985",
          "K992181",
          "K862573",
          "K142793",
          "K913476",
          "K160567",
          "K120182",
          "K110529",
          "K873343",
          "K946249",
          "K821410",
          "K080679",
          "K193325",
          "K842589",
          "K924308",
          "K123804",
          "K973893",
          "K061021",
          "K861323",
          "K182915",
          "K153506",
          "K820176",
          "K240069",
          "K102572",
          "K101199",
          "K183508",
          "K042300",
          "K965087",
          "K875232",
          "K234033",
          "K891109",
          "K032734",
          "K842808",
          "K861618",
          "K945618",
          "K871691",
          "K961884",
          "K070124",
          "K971230",
          "K832093",
          "K880051",
          "K130611",
          "K920579",
          "K143378",
          "K934772",
          "K994300",
          "K834554",
          "K123659",
          "K844175",
          "K021493",
          "K041034",
          "K885339",
          "K082710",
          "K853384",
          "K833293",
          "K240965",
          "K923692",
          "K102391",
          "K792597",
          "K102741",
          "K852707",
          "K834268",
          "K821409",
          "K042431",
          "K043258",
          "K884275",
          "K841945",
          "K862855",
          "K875171",
          "K961664",
          "K012476",
          "K142548",
          "K112268",
          "K810854",
          "K990127",
          "K896210",
          "K953375",
          "K821906",
          "K760945",
          "K960632",
          "K923168",
          "K860052",
          "K971901",
          "K850664",
          "K880768",
          "K830985",
          "K872771",
          "K936168",
          "K100173",
          "K072476",
          "K922216",
          "K043044",
          "K180605",
          "K791714",
          "K991375",
          "K935045",
          "K831840",
          "K935633",
          "K112434",
          "K253514",
          "K841034",
          "K831167",
          "K860991",
          "K834258",
          "K832672",
          "K812427",
          "K954776",
          "K250389",
          "K841033",
          "K902623",
          "K901134",
          "K910912",
          "K230206",
          "K220218",
          "K970849",
          "K242901",
          "K183577",
          "K924996",
          "K884439",
          "K830998",
          "K081725",
          "K012113",
          "K823475",
          "K810156",
          "K043203",
          "K963113",
          "K853197",
          "K872932",
          "K182248",
          "K031538",
          "K791250",
          "K895919",
          "K832170",
          "K210437",
          "K962554",
          "K910913",
          "K900563",
          "K801558",
          "K012147",
          "K951489",
          "K855097",
          "K915735",
          "K841038",
          "K973409",
          "K022558",
          "K854616",
          "K871528",
          "K092049",
          "K901835",
          "K883351",
          "K895329",
          "K961345",
          "K950017",
          "K915110",
          "K202539",
          "K874936",
          "K860818",
          "K823082",
          "K831970",
          "K221400",
          "K212316",
          "K850134",
          "K162340",
          "K810250",
          "K100366",
          "K913997",
          "K831009",
          "K862091",
          "K812648",
          "K931921",
          "K900324",
          "K082241",
          "K853288",
          "K090559",
          "K832196",
          "K830989",
          "K884309",
          "K875126",
          "K895933",
          "K072550",
          "K811683",
          "K923488",
          "K840864",
          "K902641",
          "K890151",
          "K855098",
          "K242336",
          "K180622",
          "K893888",
          "K872531",
          "K924065",
          "K832927",
          "K813132",
          "K833513",
          "K203133",
          "K932861",
          "K882867",
          "K811061",
          "K901683",
          "K831631",
          "K935198",
          "K833591",
          "K831169",
          "K932295",
          "K133096",
          "K182804",
          "K930456",
          "K131532",
          "K851671",
          "K895616",
          "K861051",
          "K831847",
          "K920262",
          "K821036",
          "K923474",
          "K113719",
          "K133465",
          "K191387",
          "K861309",
          "K971434",
          "K860075",
          "K843392",
          "K880767",
          "K063061",
          "K833687",
          "K960177",
          "K813526",
          "K822380",
          "K910382",
          "K812507",
          "K905164",
          "K884128",
          "K030784",
          "K955248",
          "K081739",
          "K901753",
          "K112863",
          "K914375",
          "K913786",
          "K841428",
          "K001916",
          "K002838",
          "K823195",
          "K142971",
          "K844174",
          "K901574",
          "K220455",
          "K924306",
          "K201322",
          "K830965",
          "K791758",
          "K133881",
          "K021991",
          "K972478",
          "K910911",
          "K093653",
          "K140310",
          "K861552",
          "K983366",
          "K112511",
          "K844745",
          "K924850",
          "K023344",
          "K162110",
          "K871846",
          "K861494",
          "K843305",
          "K831328",
          "K964920",
          "K011467",
          "K840019",
          "K993296",
          "K862489",
          "K831971",
          "K946140",
          "K241185",
          "K211072",
          "K925102",
          "K893540",
          "K861427",
          "K073034",
          "K860589",
          "K932855",
          "K842076",
          "K241039",
          "K881310",
          "K920894",
          "K790387",
          "K833054",
          "K222880",
          "K982894",
          "K910910",
          "K800508",
          "K191340",
          "K133087",
          "K920703",
          "K925061",
          "K802585",
          "K083684",
          "K882218",
          "K801498",
          "K071764",
          "K071579",
          "K954879",
          "K890870",
          "K052903",
          "K922217",
          "K802300",
          "K790612",
          "K011343",
          "K822087",
          "K882918",
          "K900464",
          "K871554",
          "K881978",
          "K870508",
          "K873360",
          "K954429",
          "K043114",
          "K812102",
          "K013162",
          "K152246",
          "K970178",
          "K921652",
          "K233591",
          "K813169",
          "K020379",
          "K822565",
          "K043027",
          "K884127",
          "K061432",
          "K933842",
          "K830986",
          "K851444",
          "K811639",
          "K831168",
          "K874755",
          "K010797",
          "K992494",
          "K841036",
          "K810979",
          "K800955",
          "K853309",
          "K973892",
          "K153461",
          "K161413",
          "K834562",
          "K855121",
          "K843975",
          "K830710",
          "K130674",
          "K841014",
          "K821825",
          "K913603",
          "K072787",
          "K133334",
          "K895696",
          "K792333",
          "K921127",
          "K100288",
          "K932880",
          "K840752",
          "K890552",
          "K850182",
          "K895142",
          "K010798",
          "K990365",
          "K100273",
          "K864364",
          "K162906",
          "K901023",
          "K981323",
          "K841289",
          "K852792",
          "K900590",
          "K051206",
          "K893541",
          "K880201",
          "K841349",
          "K895330",
          "K810499",
          "K132686",
          "K240039",
          "K143088",
          "K072756",
          "K915841",
          "K221898",
          "K940548",
          "K030841",
          "K892439",
          "K953232",
          "K890093",
          "K163092",
          "K922567",
          "K993358",
          "K190084",
          "K092418",
          "K131052",
          "K861599",
          "K080328",
          "K915837",
          "K160509",
          "K841035",
          "K981203",
          "K800815",
          "K222085",
          "K091759",
          "K812279",
          "K122030",
          "K243228",
          "K073598",
          "K962375",
          "K103254",
          "K895706",
          "K083153",
          "K900623",
          "K212407",
          "K133129",
          "K781230",
          "K983476",
          "K852552",
          "K921853",
          "K173355",
          "K993138",
          "K992719",
          "K945612",
          "K941228",
          "K142327",
          "K241169",
          "K822443",
          "K900852",
          "K831969",
          "K082863",
          "K895650",
          "K170319",
          "K811533",
          "K821067",
          "K042112",
          "K920493",
          "K840916",
          "K791268",
          "K874393",
          "K922209",
          "K925261",
          "K070449",
          "K832293",
          "K190114",
          "K843939",
          "K890871",
          "K881162",
          "K132424",
          "K160654",
          "K830119",
          "K133077",
          "K830990",
          "K882682",
          "K142792",
          "K882089",
          "K922998",
          "K030855",
          "K230131",
          "K852363",
          "K170487",
          "K091637",
          "K861090",
          "K921680",
          "K831823",
          "K150350",
          "K133599",
          "K173955",
          "K843567",
          "K852535",
          "K212728",
          "K960415",
          "K854215",
          "K820420",
          "K022112",
          "K042572",
          "K083685",
          "K022648",
          "K842174",
          "K110804",
          "K122653",
          "K113351",
          "K871003",
          "K800507",
          "K921478",
          "K833621",
          "K820441",
          "K063118",
          "K923983",
          "K142816",
          "K973891",
          "K830987",
          "K873441",
          "K925854",
          "K791257",
          "K810573",
          "K261890",
          "K971757",
          "K133251",
          "K972102",
          "K894857",
          "K041987",
          "K180725",
          "K943240",
          "K852966",
          "K822631",
          "K830794",
          "K893209",
          "K072652",
          "K833947",
          "K851975",
          "K172556",
          "K190977",
          "K984310",
          "K191885",
          "K163002",
          "K915171",
          "K922667",
          "K926581",
          "K883336",
          "K011341",
          "K222846",
          "K990389",
          "K821231",
          "K092628",
          "K844203",
          "K902684",
          "K860285",
          "K895366",
          "K813585",
          "K234145",
          "K112860",
          "K841986",
          "K822228",
          "K180026",
          "K982500",
          "K921697",
          "K031492",
          "K131183",
          "K020005",
          "K842841",
          "K844222",
          "K120272",
          "K902729",
          "K121571",
          "K984628",
          "K864182",
          "K200646",
          "K021034",
          "K822004",
          "K890430",
          "K903325",
          "K960677",
          "K123555",
          "K923002",
          "K760708",
          "K854660",
          "K880232",
          "K800510",
          "K974806",
          "K202748",
          "K033562",
          "K840656",
          "K220588",
          "K100365",
          "K851672",
          "K864218",
          "K102712",
          "K921370",
          "K891371",
          "K820442",
          "K020120",
          "K945012",
          "K030413",
          "K832501",
          "K130483",
          "K110005",
          "K083770",
          "K082654",
          "K991668",
          "K810392",
          "K811968",
          "K923046",
          "K833492",
          "K955181",
          "K841039",
          "K831974",
          "K885220",
          "K913784",
          "K831868",
          "K900624",
          "K874154",
          "K810996",
          "K011340",
          "K011077",
          "K103109",
          "K970890",
          "K872096",
          "K972025",
          "K832966",
          "K901130",
          "K112342",
          "K822879",
          "K980305",
          "K943685",
          "K912047",
          "K013144",
          "K822875",
          "K990863",
          "K802482",
          "K853353",
          "K872097",
          "K051551",
          "K864683",
          "K102548",
          "K935512",
          "K183336",
          "K903662",
          "K123716",
          "K890424",
          "K851539",
          "K801511",
          "K910332",
          "K841330",
          "K832859",
          "K080340",
          "K212444",
          "K883319",
          "K160787",
          "K014297",
          "K902099",
          "K871848",
          "K943675",
          "K831968",
          "K991918",
          "K791269",
          "K921026",
          "K141631",
          "K861862"
        ],
        "registration_number": [
          "",
          "3005533818",
          "2182291",
          "2029015",
          "3010594263",
          "3010280860",
          "3015443199",
          "3005099803",
          "3025772870",
          "3010331226",
          "3013298431",
          "8040607",
          "3036773433",
          "2431014",
          "2029275",
          "8020785",
          "1423507",
          "1417592",
          "1825146",
          "1713468",
          "3007014520",
          "3005609751",
          "3008044552",
          "3009247326",
          "8030107",
          "1047429",
          "3015479927",
          "2025851",
          "3007285848",
          "1423662",
          "3020417125",
          "3006271008",
          "3010757606",
          "1828132",
          "2518417",
          "9610849",
          "2080225",
          "9610694",
          "3014421917",
          "3011707784",
          "3011088743",
          "3014637079",
          "2320961",
          "3002807314",
          "3006161756",
          "3006538822",
          "1911916",
          "3034669683",
          "3010761399",
          "3017081712",
          "3006260740",
          "3006673317",
          "3009291251",
          "1223004",
          "1221019",
          "2011171",
          "1214422",
          "3013234390",
          "1526012",
          "3009746061",
          "1000121056",
          "3010155661",
          "1119193",
          "2024311",
          "2518902",
          "1718873",
          "3022964615",
          "3018094310",
          "1647149",
          "2016945",
          "2032112",
          "3045857302",
          "2245270",
          "3043226252",
          "3004111573",
          "3009698186",
          "3009558349",
          "1048735",
          "3012568936",
          "3004365956",
          "3018507299",
          "3026671998",
          "3008311916",
          "3012481535",
          "3031289255",
          "9681572",
          "3010958377",
          "3010972758",
          "1649518",
          "9616088",
          "2024024",
          "3004550973",
          "1314417",
          "3018167761",
          "3044994889",
          "1220477",
          "1024404",
          "3010041511",
          "1450662",
          "9680143",
          "9611665",
          "3008810357",
          "1320894",
          "3015237508",
          "3004122598",
          "1221765",
          "1623486",
          "1921846",
          "3009265194",
          "1219930",
          "3010452421",
          "3012881740",
          "8010042",
          "3005210579",
          "3002808467",
          "3012494290",
          "3009211636",
          "3014656749",
          "3012421607",
          "3018366470",
          "1419937",
          "3008808123",
          "3001124136",
          "3005674464",
          "3018116884",
          "3017961114",
          "1313046",
          "3004904811",
          "9710641",
          "3006606901",
          "1422634",
          "3014663754",
          "3011270181",
          "9680425",
          "3013508647",
          "3005669815",
          "3010400865",
          "3042225166",
          "1423537",
          "3006606984",
          "1043214",
          "1293141",
          "9617594",
          "3003627840",
          "3014926389",
          "3004194523",
          "2133810",
          "3010131137",
          "8044169",
          "3011237770",
          "3014656668",
          "8022978",
          "1424263",
          "1480288",
          "8022210",
          "9611594",
          "9611590",
          "1066607",
          "3031190571",
          "2126666",
          "3015173212",
          "3013557562",
          "1058397",
          "3011324393",
          "1223925",
          "3012169247",
          "3011299837",
          "1222168",
          "1820334",
          "1526854",
          "3011987967",
          "3017059666",
          "3030447506",
          "9680934",
          "2243072",
          "3005670221",
          "3016616327",
          "3017140456",
          "3013188547",
          "1018233",
          "1055236",
          "3007697249",
          "3005987240",
          "3008806809",
          "3005060859",
          "2032098",
          "1643817",
          "3009888344",
          "1030451",
          "1000317571",
          "1037905",
          "9612030",
          "3005580113",
          "2183673",
          "3006082230",
          "1213809",
          "3008570882",
          "3009632672",
          "3005483737",
          "9618003",
          "3000268902",
          "3007766601",
          "3006157842",
          "1928237",
          "3006131984",
          "2183744",
          "1319639",
          "9681477",
          "1221485",
          "3017607339",
          "3035642068",
          "3011802696",
          "3010273872",
          "3030446844",
          "3003915875",
          "3012441529",
          "1320599",
          "2030624",
          "1054241",
          "3003903031",
          "9681260",
          "3013524793",
          "3030708724",
          "1056553",
          "3005012805",
          "3011137372",
          "3006017180",
          "2528981",
          "3014579161",
          "3016785268",
          "3006795797",
          "3016171914"
        ],
        "fei_number": [
          "",
          "3005533818",
          "1000138054",
          "2182291",
          "3002902735",
          "3005747797",
          "3003565456",
          "3010594263",
          "3010280860",
          "3015443199",
          "3005099803",
          "3025772870",
          "3001236905",
          "3003263857",
          "3010331226",
          "3013298431",
          "3036773433",
          "2431014",
          "1417592",
          "1825146",
          "1713468",
          "3007014520",
          "3005609751",
          "3008044552",
          "3009247326",
          "1047429",
          "3015479927",
          "2025851",
          "3007285848",
          "3009751352",
          "3002807391",
          "1000514656",
          "3020417125",
          "3006271008",
          "3010757606",
          "3005596530",
          "1828132",
          "3002806591",
          "2518417",
          "2080225",
          "3014421917",
          "3011707784",
          "3000206435",
          "3011088743",
          "3014637079",
          "3003172243",
          "3002807314",
          "3006538822",
          "2937446",
          "1911916",
          "3034669683",
          "3010761399",
          "3017081712",
          "3006260740",
          "3009291251",
          "3003282547",
          "1221019",
          "2011171",
          "1214422",
          "3013234390",
          "3003076831",
          "3009746061",
          "1000121056",
          "3010155661",
          "1119193",
          "3001236277",
          "3007596547",
          "1526012",
          "1718873",
          "3022964615",
          "3018094310",
          "2518902",
          "1000120555",
          "2016945",
          "3045857302",
          "2245270",
          "1000158532",
          "1000306431",
          "3043226252",
          "3004111573",
          "3009698186",
          "3009558349",
          "1048735",
          "3012568936",
          "3004365956",
          "3018507299",
          "3026671998",
          "3008311916",
          "3012481535",
          "3002806565",
          "3031289255",
          "1000305720",
          "3010958377",
          "7682",
          "3010972758",
          "2024024",
          "3004550973",
          "1314417",
          "3018167761",
          "3044994889",
          "1220477",
          "1024404",
          "3002808128",
          "3002807095",
          "3010041511",
          "1450662",
          "3003058448",
          "3008810357",
          "3015237508",
          "3004122598",
          "1221765",
          "1623486",
          "1921846",
          "3009265194",
          "1219930",
          "3010452421",
          "1000160458",
          "3003264198",
          "3008002452",
          "3012881740",
          "3005210579",
          "3002808467",
          "3012494290",
          "3009211636",
          "3014656749",
          "3012421607",
          "3018366470",
          "1419937",
          "3008808123",
          "3001124136",
          "3005674464",
          "3018116884",
          "3017961114",
          "3002807308",
          "3004904811",
          "1313046",
          "3006606901",
          "3014663754",
          "3011270181",
          "3013508647",
          "3005669815",
          "2024311",
          "3010400865",
          "3042225166",
          "3006606984",
          "1043214",
          "1293141",
          "1000112968",
          "3003627840",
          "3014926389",
          "1000132370",
          "3004194523",
          "3010131137",
          "3011237770",
          "3014656668",
          "1480288",
          "3000215273",
          "3031190571",
          "2126666",
          "3015173212",
          "1000283452",
          "3002806934",
          "3013557562",
          "3011324393",
          "3006993165",
          "3002806731",
          "3012169247",
          "1222168",
          "1820334",
          "1526854",
          "3000717944",
          "3011987967",
          "3017059666",
          "3030447506",
          "1000237915",
          "2243072",
          "3005670221",
          "3016616327",
          "3017140456",
          "3013188547",
          "3000203066",
          "1018233",
          "1000112877",
          "3007697249",
          "3005987240",
          "3008806809",
          "3005060859",
          "3006795797",
          "1643817",
          "3009888344",
          "1030451",
          "1000317571",
          "1037905",
          "3005580113",
          "2183673",
          "3006082230",
          "3002807439",
          "1213809",
          "3009632672",
          "3005483737",
          "3000268902",
          "3007766601",
          "3932",
          "3006157842",
          "3003015552",
          "1928237",
          "2183744",
          "3006131984",
          "1000306139",
          "1221485",
          "3002808127",
          "3017607339",
          "3035642068",
          "3011802696",
          "3010273872",
          "3030446844",
          "3003915875",
          "3012441529",
          "3003903031",
          "3030708724",
          "3013524793",
          "3006582718",
          "3005012805",
          "3006017180",
          "2528981",
          "3014579161",
          "3003807466",
          "3016785268",
          "3001451571",
          "3002806436",
          "3016171914"
        ],
        "device_name": "Tubes, Gastrointestinal (And Accessories)",
        "medical_specialty_description": "Gastroenterology, Urology",
        "regulation_number": "876.5980",
        "device_class": "2"
      }
    }
  ]
}