  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-08-27",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "30827",
      "product_res_number": "Z-0509-04",
      "event_date_initiated": "2003-12-05",
      "event_date_posted": "2004-02-04",
      "recall_status": "Terminated",
      "event_date_terminated": "2005-04-25",
      "res_event_number": "27971",
      "product_code": "DKJ--",
      "k_numbers": [
        "K844070"
      ],
      "product_description": "TDx/TDxFLx Salicylate, list 9533-60; Abbott Laboratories, Diagnostics Division, Abbott Park, IL 60064",
      "code_info": "All lots with insert versions 69-6373/R4, 69-6003/R3 or 66-6296/R2.  The corrected insert is version 34-3033/R5.",
      "firm_fei_number": "1415939",
      "recalling_firm": "Abbott Laboratories HPD/ADD/GPRD",
      "address_1": "100/200 Abbott Park Road",
      "city": "Abbott Park",
      "state": "IL",
      "postal_code": "60064",
      "additional_info_contact": "Abbott Customer Support\n877-422-2688",
      "reason_for_recall": "The reagent package inserts for the products contain incorrect information in the Specific Performance Characteristics section.  % Recovery incorrect.",
      "root_cause_description": "Other",
      "action": "Field Correction recall by Important Product Information letter dated 12/5/03, sent via Federal Express Priority on the same date to all customers receiving the affected reagents. The accounts were informed of the incorrect values for the Specific Performance Characteristics listed in the package inserts of the reagent kits. The letters contained a table listing the incorrect values and the correct values for each of the products, and the users were instructed to replace their product inserts with  the enclosed revised package inserts for all product specific reagent kits in their laboratory.",
      "product_quantity": "1,999 kits",
      "distribution_pattern": "Nationwide and internationally through Abbott subsidiaries in Mexico, Guatemala, Colombia, Chile, Argentina, Brazil, Puerto Rico, San Salvador, Canada, England, Japan, Singapore, Hong Kong, Taiwan, Thailand, Korea, Australia, New Zealand, Honduras, Costa Rica, Bahamas, Panama, Dominican Republic and China.",
      "openfda": {}
    }
  ]
}