  "results": [
    {
      "cfres_id": "30609",
      "product_res_number": "Z-0293-04",
      "event_date_initiated": "2003-11-23",
      "event_date_posted": "2004-01-06",
      "recall_status": "Terminated",
      "event_date_terminated": "2004-07-22",
      "res_event_number": "27835",
      "product_code": "CGN--",
      "product_description": "IMx Folate Reagent Pack, list 2220-88 (100 tests - reagents packaged with Ion Capture Reaction Cells) and list 2220-20 (100 tests - reagents alone); Abbott Laboratories, Abbott Park, IL 60064 USA",
      "code_info": "all lots",
      "firm_fei_number": "1415939",
      "recalling_firm": "Abbott Laboratories HPD/ADD/GPRD",
      "address_1": "100/200 Abbott Park Road",
      "city": "Abbott Park",
      "state": "IL",
      "postal_code": "60064",
      "reason_for_recall": "Chemotherapeutic concentrations of methotrexate in patient samples may carryover to subsequent samples, which would result in falsely elevated folate values, when performing IMx Folate testing.",
      "root_cause_description": "Other",
      "action": "Abbott notified their international affiliates via e-mail on 11/23/03.  Important Product Information letters dated 11/24/03 will be sent to the end users by the affiliates, informing them of the possible methotrexate carryover to subsequent samples and falsely elevated folate values.",
      "product_quantity": "3,475 packs",
      "distribution_pattern": "There was no U.S. distribution.  The product was internationally distributed through Abbott subsidiaries in Colombia, Chile, Uruguay, Argentina, Brazil, Canada, England and Australia.",
      "openfda": {}
    }
  ]
}