      "cfres_id": "29645",
      "product_res_number": "Z-0044-04",
      "event_date_initiated": "2003-09-23",
      "event_date_posted": "2003-10-15",
      "recall_status": "Terminated",
      "event_date_terminated": "2004-01-21",
      "res_event_number": "27301",
      "product_description": "Pin Driver Coupler, component of MicroAire Drill\\Reamer Model 7500. Driver couplers are reuseable medical devices, non-sterile, packed one per box and labeled in part ***MicroAire Surgical Instruments, 1641 Edlich Drive, Charlottesville, VA 22911***",
      "code_info": "Serial numbers: 1007, 1008, 1009, 1010, 1011, 1012, 1013, 1014, 1016, 1021, 1022, 1023, 1024, 1025, 1026, 1027, 1028, 1030, 1031, 1032, 1034, 1035, 1036, 1038, 1039, 1040, 1041, 1042, 1043, 1046, 1047, 1048, 1049, 1051, 1052, 1053, 1054, 1057, 1058, 1059, 1060, 1061, 1062, 1065, 1066, 1067, 1068, 1069, 1070, 1071, 1072, 1073, 1074, 1075, 1076, 1077, 1078, 1079, 1080, 1081, 1082, 1083, 1084, 1085, 1086, 1087, 1088, 1089, 1090, 1091, 1093, 1095, 1096, 1097, 1098, 1099, 1100, 1101, 1102, 1103, 1104, 1105, 1106, 1107, 1108, 1109, 1110, 1111, 1113, 1114, 1115, 1116, 1119, 1120, 1121, 1122, 1123, 1124, 1125, 1126, 1127, 1128, 1129, 1130, 1131, 1132, 1133, 1134, 1135, 1136, 1137, 1138, 1139, 1140, 1141, 1142, 1143, 1144, 1145, 1146, 1149, 1150, 1151, 1152, 1153, 1155, 1156, 1157, 1158.",
      "firm_fei_number": "2020601",
      "recalling_firm": "Microaire Surgical Instruments",
      "address_1": "1641 Edlich Dr",
      "city": "Charlottesville",
      "state": "VA",
      "postal_code": "22911",
      "additional_info_contact": "Chris Spofford\n800-538-5561 Ext. 344",
      "reason_for_recall": "Wiring pin locking mechanism defective, potential to come apart in surgical site.",
      "root_cause_description": "Other",
      "action": "The firm notified consignees by certified letter 9/23/03. The recall notifications were serial number specific and advised consignees of the potential defect with the wirning pin/locking mechanism. Consignees were instructed to cease use and distribution and obtain a return authorization for repair. Distributors were requested to extend the recall and recover and return devices for repair.",
      "product_quantity": "129",
      "distribution_pattern": "Product was distributed to direct accounts around the United States and to distributors around the world.",