{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-08-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "29644",
      "product_res_number": "Z-0043-04",
      "event_date_initiated": "2003-09-23",
      "event_date_posted": "2003-10-15",
      "recall_status": "Terminated",
      "event_date_terminated": "2004-01-21",
      "res_event_number": "27301",
      "product_description": "Wire Driver Coupler, Model 7100-045, a component of MicroAire Drill/Reamer Model 7500. Driver couplers are reuseable medical devices, non-sterile, packed one per box and labeled in part ***MicroAire Surgical Instruments, 1641 Edlich Drive, Charlottesville, VA 22911***.",
      "code_info": "Serial numbers: 1007, 1008, 1009, 1010, 1011, 1012, 1013, 1014, 1015, 1016, 1017, 1018, 1019, 1020, 1021, 1022, 1024, 1025, 1026, 1027, 1028, 1029, 1030, 1031, 1032, 1033, 1034, 1035, 1036, 1037, 1038, 1039, 1042, 1043, 1044, 1045, 1047, 1049, 1050, 1051, 1052, 1053, 1054, 1055, 1056, 1057, 1058, 1059, 1060, 1061, 1062, 1064, 1065, 1066, 1068, 1069, 1070, 1071, 1072, 1073, 1074, 1075, 1076, 1079, 1082, 1083, 1084, 1087, 1088, 1089, 1090, 1094, 1095, 1096, 1097, 1099, 1100, 1101, 1102, 1103, 1104, 1105, 1106, 1107, 1108, 1109, 1110, 1112, 1113, 1114, 1115, 1117, 1118, 1119, 1120, 1121, 1122, 1124, 1126, 1127, 1128, 1129, 1130, 1132, 1133, 1134, 1135, 1136, 1137, 1138, 1139, 1140, 1141, 1142, 1143, 1144, 1145, 1146, 1162, 1163, 1164, 1165, 1166, 1167, 1168, 1169, 1170, 1171, 1172, 1173, 1175, 1176, 1177, 1178, 1179.",
      "firm_fei_number": "2020601",
      "recalling_firm": "Microaire Surgical Instruments",
      "address_1": "1641 Edlich Dr",
      "city": "Charlottesville",
      "state": "VA",
      "postal_code": "22911",
      "additional_info_contact": "Chris Spofford\n800-538-5561 Ext. 344",
      "reason_for_recall": "Wiring pin locking mechanism defective, potential to come apart in surgical site.",
      "root_cause_description": "Other",
      "action": "The firm notified consignees by certified letter 9/23/03. The recall notifications were serial number specific and advised consignees of the potential defect with the wirning pin/locking mechanism. Consignees were instructed to cease use and distribution and obtain a return authorization for repair. Distributors were requested to extend the recall and recover and return devices for repair.",
      "product_quantity": "135",
      "distribution_pattern": "Product was distributed to direct accounts around the United States and to distributors around the world.",
      "openfda": {}
    }
  ]
}