  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-08-20",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "26689",
      "product_res_number": "Z-0902-03",
      "event_date_initiated": "2003-04-24",
      "event_date_posted": "2003-05-29",
      "recall_status": "Terminated",
      "event_date_terminated": "2003-09-25",
      "res_event_number": "25899",
      "product_code": "KNC",
      "k_numbers": [
        "K915779"
      ],
      "product_description": "Hill-Rom brand Affinity three birthing bed",
      "code_info": "All units distributed between January 1, 1999 and July 1, 2002.",
      "firm_fei_number": "1824206",
      "recalling_firm": "Hill-Rom, Inc.",
      "address_1": "125 E Pearl St",
      "city": "Batesville",
      "state": "IN",
      "postal_code": "47006",
      "additional_info_contact": "812-934-7777",
      "reason_for_recall": "Possible electrical shock hazard, as the power cord grounding pin may break off or become detached.",
      "root_cause_description": "Other",
      "action": "Letters dated April 24, 2003 were sent to each consignee instructing them to inspect the beds for power cords with non-flared ground pins, to cut plug off any such cords when found, and replace these plugs with new plugs  with the flared ground pin.  [The type of plug on the unit can be seen by the naked eye, as the plug is transparent plastic molding.]",
      "distribution_pattern": "United States, Bahamas, Bermuda, Canada, Colombia, Costa Rica, Dominican Republic, Guam, Japan, Mexico, Panama, Saudi Arabia, South Korea, Taiwan and Venezuela",
      "openfda": {
        "k_number": [
          "K914544",
          "K831309",
          "K852171",
          "K861030",
          "K842962",
          "K862931",
          "K001261",
          "K924420",
          "K932035",
          "K903973",
          "K844183",
          "K870745",
          "K790950",
          "K820175",
          "K852174",
          "K830873",
          "K833321",
          "K792568",
          "K811902",
          "K770437",
          "K812141",
          "K831998",
          "K915779",
          "K864194",
          "K801622",
          "K884868"
        ],
        "device_name": "Table, Obstetric (And Accessories)",
        "medical_specialty_description": "Obstetrics/Gynecology",
        "regulation_number": "884.4900",
        "device_class": "2",
        "registration_number": [
          "1824206",
          "3007031002",
          "3007123908",
          "3040337865",
          "3033433849",
          "3006697241"
        ],
        "fei_number": [
          "1824206",
          "3007031002",
          "3007123908",
          "3040337865",
          "3033433849",
          "3006697241"
        ]
      }
    }
  ]
}