  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-08-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "26374",
      "product_res_number": "Z-0695-03",
      "event_date_initiated": "2003-02-14",
      "event_date_posted": "2003-03-29",
      "recall_status": "Terminated",
      "event_date_terminated": "2004-09-27",
      "res_event_number": "25725",
      "product_code": "6885",
      "k_numbers": [
        "K981054"
      ],
      "product_description": "Cidex Plus Solution Test Strips",
      "code_info": "Cidex Plus, product code 2924, lots: 8423  8560 9229   Cidex Plus, Product code 2926, lots: 8312 8658 9120",
      "firm_fei_number": "3003662624",
      "recalling_firm": "Advanced Sterilization Products",
      "address_1": "33 Technology Dr",
      "city": "Irvine",
      "state": "CA",
      "postal_code": "926182346",
      "additional_info_contact": "Nevine Erian\n949-453-6422",
      "reason_for_recall": "Test strips which are used for QC of sterilant solution fail due to ingress of moisture.",
      "root_cause_description": "Other",
      "action": "Several emails and letters were sent between 2/14/03 and 2/27/03 to customers requesting return of product and also advising them that should return not be a viable option due to availability of replacements, a suggested temporary QC procedure could be used to separate good lots of strips from affected lots.",
      "product_quantity": "10,047 cases, (2 bottles per case)",
      "distribution_pattern": "Nationwide. Australia, Brazil, Canada, Puerto Rico, Colombia, Japan, Korea, South Africa, Thailand, England, Uruguay, Venezuela, Republic of Panama.",
      "openfda": {}
    }
  ]
}