      "cfres_id": "25821",
      "product_res_number": "Z-0597-03",
      "event_date_initiated": "2003-01-02",
      "event_date_posted": "2003-02-28",
      "recall_status": "Terminated",
      "event_date_terminated": "2006-02-15",
      "res_event_number": "25430",
      "k_numbers": [
        "K011093"
      ],
      "product_description": "Intellivue Information Center\nModel: M3150BU#CO1",
      "code_info": "4225A00169 4225A00171 4225A00181 4225A00182 4213A13456 4131A13457",
      "firm_fei_number": "1218950",
      "recalling_firm": "Philips Medical Systems, Inc. Cardiac & Monitoring Systems",
      "address_1": "3000 Minuteman Road",
      "city": "Andover",
      "state": "MA",
      "postal_code": "01810",
      "additional_info_contact": "Brenda Getchell\n978-659-2134",
      "reason_for_recall": "Central Station Monitoring Information Centers may experience sreeen freeze, i.e. loss of display,  keyboard or mouse lock up",
      "root_cause_description": "Other",
      "action": "Philips Medical Systems notified users by Certified Mail on/about 2/19/03 that the firm was initiating  a device field correction (recall) to conduct an upgrade to units witha HP VL400 or VL420 computer which is cconfigured for dual display by replacing the video and network communication card. Previous to this notification, Philips sent letters on 1/2/03 addressed to Biomedical Engineering Manager to advise users\nto reboot the system in the event of a system freeze.",
      "product_quantity": "7 units",
      "distribution_pattern": "Nationwide\nCanada, Europe, Asia, Latin America",
      "openfda": {}
    }
  ]
}