  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-08-20",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "217158",
      "product_res_number": "Z-0882-2026",
      "event_date_initiated": "2025-10-29",
      "event_date_posted": "2025-12-03",
      "recall_status": "Open, Classified",
      "res_event_number": "97919",
      "product_code": "JAK",
      "k_numbers": [
        "K232491"
      ],
      "product_description": "CT 5300;\nSoftware Version Number: 4.5, 5.0, 5.1;",
      "code_info": "1) Model Number: 728285;\nUDI-DI: 00884838113237;\nSerial Numbers: 35071, 800007, 800002, 800001, 800004, 800003;",
      "firm_fei_number": "3016618143",
      "recalling_firm": "Philips North America Llc",
      "address_1": "222 Jacobs St",
      "city": "Cambridge",
      "state": "MA",
      "postal_code": "02141-2289",
      "reason_for_recall": "If the screws of tube heat exchanger on the rotating scanner are not tightened well after replacing. This component may become detached and make contact with other components located within the Incisive CT systems during rotation. Other components could be damaged due to the contact with the tube heat exchanger.",
      "root_cause_description": "Under Investigation by firm",
      "action": "On October 23, 2025 URGENT Medical Device Correction letters were sent to customers. Actions to be taken: 1. Please continue to use your system in accordance with its intended use and do not stay in the scan room when scanning.\n2. Circulate this URGENT Medical Device Correction letter to all users of this device so that they are aware of the issue.\n3.  Place this URGENT Medical Device Correction with your system documentation and ensure the letter is in a place likely to be seen/viewed.\n4. Please complete and return the attached response form to Philips promptly and no later than 30 days from receipt. Completing this form confirms receipt of the URGENT Medical Device Correction letter, understanding of the issue, and required actions to be taken. Philips will contact you to schedule a time for a Field Service Engineer (FSE) to visit your site to inspect the tube heat exchanger screws and correct the loosen screw if necessary (reference FCO72800842).\nPlease be assured that maintaining a high level of safety and quality is our highest priority. If you need any further information or support concerning this issue, please contact your local Philips Customer Care Solutions center: For North America, contact the Customer Care Solutions Center {1-800-722-9377, 8AM - 8PM EST, Monday-Friday).",
      "product_quantity": "6 units (OUS only)",
      "distribution_pattern": "Worldwide distribution - US Nationwide and the countries of Albania, Algeria, Angola, Argentina, Australia, Austria, Bahamas, Bahrain, Bangladesh, Belgium, Bhutan, Bolivia, Bosnia-Herz., Brazil, Bulgaria, Cambodia, Canada, Chile, China, Colombia, Costa Rica, C¿te D'Ivoire, Croatia, Cyprus, Czech Republic, Denmark, Dominican Rep, Ecuador, Egypt, El Salvador, Ethiopia, France, FrenchPolynesia, Georgia, Germany, Greece, Guadeloupe, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kosovo, Kuwait, Lao, Latvia, Lebanon, Lesotho, Libya, Lithuania, Macedonia, Malaysia, Mauritius, Mexico, Moldova, Montenegro, Montserrat, Morocco, Mozambique, Myanmar, Nepal, Netherlands, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, Palestine, Panama, Papua Nw Guinea, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, R¿union, Romania, Russian Fed., Rwanda, Saint Kitts, Saint Lucia, Saudi Arabia, Serbia, Sint Maarten (D, Slovakia, South Africa, South Korea, Spain, St.Pier,Miquel., Sweden, Switzerland, Syria, Taiwan, Tanzania, Thailand, Trinidad,Tobago, Tunisia, T¿rkiye, Turkmenistan, Ukraine, United Kingdom, Utd.Arab.Emir., Uzbekistan, Venezuela, Vietnam, Yemen.",
      "openfda": {
        "k_number": [
          "K080271",
          "K884268",
          "K163643",
          "K033502",
          "K221116",
          "K881305",
          "K790975",
          "K201745",
          "K073003",
          "K160350",
          "K163289",
          "K200626",
          "K211168",
          "K250370",
          "K932848",
          "K052216",
          "K243240",
          "K024149",
          "K023943",
          "K090220",
          "K233731",
          "K212067",
          "K875370",
          "K973168",
          "K042438",
          "K051833",
          "K040902",
          "K121072",
          "K063324",
          "K082104",
          "K203699",
          "K172983",
          "K042328",
          "K803181",
          "K061310",
          "K071806",
          "K192590",
          "K012106",
          "K113732",
          "K032536",
          "K130645",
          "K161009",
          "K032243",
          "K110227",
          "K072574",
          "K811227",
          "K013561",
          "K242874",
          "K191332",
          "K151376",
          "K191468",
          "K123500",
          "K150595",
          "K022326",
          "K260166",
          "K151752",
          "K023785",
          "K120579",
          "K260169",
          "K183143",
          "K151765",
          "K081482",
          "K101888",
          "K760688",
          "K982787",
          "K151673",
          "K202526",
          "K022919",
          "K012836",
          "K142465",
          "K031871",
          "K013422",
          "K911500",
          "K982060",
          "K990426",
          "K003152",
          "K062673",
          "K041267",
          "K203187",
          "K894542",
          "K150565",
          "K022050",
          "K223804",
          "K761330",
          "K950972",
          "K053046",
          "K061187",
          "K173607",
          "K935454",
          "K062351",
          "K173468",
          "K151533",
          "K243398",
          "K183202",
          "K253560",
          "K244008",
          "K250766",
          "K940606",
          "K982349",
          "K790134",
          "K260167",
          "K182716",
          "K192956",
          "K073138",
          "K092747",
          "K182901",
          "K233750",
          "K082960",
          "K812509",
          "K091673",
          "K133640",
          "K241166",
          "K944131",
          "K192828",
          "K944326",
          "K063241",
          "K023687",
          "K010813",
          "K120300",
          "K171183",
          "K160587",
          "K203411",
          "K203617",
          "K232227",
          "K840860",
          "K923106",
          "K163297",
          "K902557",
          "K913029",
          "K133646",
          "K253770",
          "K791163",
          "K211828",
          "K163284",
          "K191136",
          "K771082",
          "K874767",
          "K972399",
          "K134025",
          "K130552",
          "K133674",
          "K863935",
          "K150757",
          "K242403",
          "K060779",
          "K033326",
          "K984634",
          "K132410",
          "K231281",
          "K982536",
          "K052839",
          "K790342",
          "K141741",
          "K945512",
          "K020796",
          "K160126",
          "K955268",
          "K884555",
          "K150785",
          "K781431",
          "K133424",
          "K122909",
          "K211645",
          "K232835",
          "K131773",
          "K093995",
          "K010042",
          "K790395",
          "K991600",
          "K915738",
          "K123584",
          "K772125",
          "K193289",
          "K924985",
          "K011160",
          "K230220",
          "K090173",
          "K991764",
          "K153429",
          "K160743",
          "K231574",
          "K770855",
          "K201053",
          "K133324",
          "K252992",
          "K091529",
          "K812241",
          "K882022",
          "K191777",
          "K211180",
          "K192832",
          "K241200",
          "K233583",
          "K924810",
          "K201188",
          "K110812",
          "K181646",
          "K081379",
          "K002978",
          "K780928",
          "K214019",
          "K020929",
          "K844926",
          "K153331",
          "K181414",
          "K033832",
          "K882906",
          "K193281",
          "K884788",
          "K083293",
          "K180225",
          "K031206",
          "K921192",
          "K193267",
          "K201586",
          "K891931",
          "K093146",
          "K761332",
          "K120710",
          "K841997",
          "K960635",
          "K250822",
          "K192763",
          "K923109",
          "K201898",
          "K082220",
          "K010817",
          "K170221",
          "K082761",
          "K022013",
          "K974455",
          "K143294",
          "K002005",
          "K980169",
          "K112989",
          "K960985",
          "K092013",
          "K150745",
          "K182147",
          "K200524",
          "K200515",
          "K031261",
          "K080624",
          "K060937",
          "K900660",
          "K000819",
          "K911501",
          "K070658",
          "K161157",
          "K173625",
          "K223424",
          "K251805",
          "K970999",
          "K111633",
          "K884667",
          "K253649",
          "K193073",
          "K131433",
          "K923111",
          "K171677",
          "K240844",
          "K200990",
          "K241295",
          "K871859",
          "K002258",
          "K801407",
          "K123541",
          "K203720",
          "K965004",
          "K941745",
          "K180901",
          "K911764",
          "K190578",
          "K230807",
          "K121418",
          "K872150",
          "K972903",
          "K970980",
          "K163528",
          "K202213",
          "K253596",
          "K200498",
          "K880965",
          "K161196",
          "K073238",
          "K052260",
          "K212441",
          "K854340",
          "K913485",
          "K013146",
          "K250648",
          "K896456",
          "K143401",
          "K183046",
          "K190841",
          "K760882",
          "K100637",
          "K913491",
          "K930765",
          "K052029",
          "K023760",
          "K153444",
          "K013999",
          "K864046",
          "K163213",
          "K183291",
          "K213649",
          "K153004",
          "K091934",
          "K201836",
          "K761033",
          "K063762",
          "K133901",
          "K173630",
          "K992740",
          "K923108",
          "K060052",
          "K161066",
          "K011447",
          "K000300",
          "K780326",
          "K780115",
          "K030450",
          "K123540",
          "K072606",
          "K781040",
          "K093810",
          "K173632",
          "K251061",
          "K850027",
          "K884671",
          "K150713",
          "K161748",
          "K183548",
          "K223447",
          "K023332",
          "K062386",
          "K073695",
          "K173076",
          "K011180",
          "K013381",
          "K983140",
          "K191394",
          "K133497",
          "K171013",
          "K250928",
          "K120331",
          "K123596",
          "K182223",
          "K233334",
          "K183161",
          "K013521",
          "K921432",
          "K173821",
          "K903701",
          "K961464",
          "K230112",
          "K142584",
          "K163281",
          "K073373",
          "K233657",
          "K980837",
          "K943623",
          "K192686",
          "K081022",
          "K213715",
          "K970606",
          "K240369",
          "K912454",
          "K963896",
          "K862317",
          "K253686",
          "K934832",
          "K973908",
          "K020385",
          "K923114",
          "K792603",
          "K083498",
          "K913352",
          "K132222",
          "K132813",
          "K220814",
          "K080540",
          "K875088",
          "K781936",
          "K232206",
          "K780480",
          "K984156",
          "K813162",
          "K222361",
          "K120546",
          "K234030",
          "K172188",
          "K050458",
          "K974513",
          "K003230",
          "K253450",
          "K153263",
          "K955255",
          "K812410",
          "K012238",
          "K133373",
          "K063189",
          "K230130",
          "K121827",
          "K843927",
          "K093891",
          "K935487",
          "K110366",
          "K151283",
          "K211591",
          "K020913",
          "K852865",
          "K780737",
          "K904971",
          "K072693",
          "K973329",
          "K032915",
          "K252217",
          "K040782",
          "K050297",
          "K110066",
          "K012009",
          "K772289",
          "K003639",
          "K171022",
          "K183268",
          "K061625",
          "K102102",
          "K902848",
          "K091195",
          "K923524",
          "K153360",
          "K083282",
          "K092742",
          "K230223",
          "K243617",
          "K982583",
          "K132523",
          "K213504",
          "K243229",
          "K130960",
          "K973327",
          "K910375",
          "K021491",
          "K031469",
          "K210085",
          "K022354",
          "K191550",
          "K872615",
          "K211373",
          "K032823",
          "K211711",
          "K221219",
          "K870513",
          "K250650",
          "K233875",
          "K872526",
          "K060903",
          "K260078",
          "K972879",
          "K170019",
          "K152473",
          "K153220",
          "K213725",
          "K223261",
          "K924082",
          "K110093",
          "K213944",
          "K252249",
          "K203042",
          "K241440",
          "K113421",
          "K151919",
          "K760201",
          "K090504",
          "K033357",
          "K121792",
          "K190505",
          "K092365",
          "K071308",
          "K203020",
          "K003736",
          "K040372",
          "K072841",
          "K131576",
          "K243523",
          "K201092",
          "K811358",
          "K802045",
          "K013214",
          "K121498",
          "K061817",
          "K162302",
          "K163711",
          "K220961",
          "K942040",
          "K991278",
          "K172539",
          "K132902",
          "K143400",
          "K193170",
          "K041542",
          "K133705",
          "K230421",
          "K102259",
          "K080211",
          "K040665",
          "K030420",
          "K123576",
          "K220450",
          "K043086",
          "K772291",
          "K173478",
          "K232155",
          "K150104",
          "K053210",
          "K250941",
          "K083227",
          "K903763",
          "K203514",
          "K913996",
          "K772198",
          "K913252",
          "K212889",
          "K232045",
          "K193293",
          "K232392",
          "K183204",
          "K091103",
          "K142955",
          "K150003",
          "K250662",
          "K143223",
          "K192427",
          "K001682",
          "K033677",
          "K130224",
          "K113715",
          "K122471",
          "K881674",
          "K132524",
          "K772305",
          "K132108",
          "K053024",
          "K163296",
          "K936272",
          "K861698",
          "K833228",
          "K063184",
          "K912752",
          "K133589",
          "K151749",
          "K140912",
          "K012313",
          "K133648",
          "K993792",
          "K243672",
          "K090462",
          "K982010",
          "K133677",
          "K251484",
          "K882378",
          "K203169",
          "K103127",
          "K051980",
          "K991716",
          "K113466",
          "K011141",
          "K203225",
          "K971717",
          "K191891",
          "K180015",
          "K894887",
          "K772317",
          "K120833",
          "K071378",
          "K974307",
          "K052855",
          "K121553",
          "K123509",
          "K061370",
          "K811723",
          "K974344",
          "K122109",
          "K802206",
          "K923110",
          "K042293",
          "K012385",
          "K090991",
          "K161285",
          "K253026",
          "K061449",
          "K213713",
          "K961055",
          "K771990",
          "K971054",
          "K974715",
          "K181797",
          "K242188",
          "K213759",
          "K900170",
          "K081105",
          "K113342",
          "K220910",
          "K233753",
          "K061834",
          "K052852",
          "K152036",
          "K220986",
          "K163341",
          "K141069",
          "K203216",
          "K994431",
          "K964318",
          "K062315",
          "K791195",
          "K102677",
          "K223726",
          "K233650",
          "K233749",
          "K894888",
          "K941546",
          "K130901",
          "K982265",
          "K900233",
          "K113548",
          "K151372",
          "K222819",
          "K213096",
          "K071310",
          "K233080",
          "K212494",
          "K003408",
          "K183448",
          "K062451",
          "K172058",
          "K151414",
          "K223311",
          "K880822",
          "K220664",
          "K140232",
          "K853444",
          "K003014",
          "K894269",
          "K112020",
          "K071513",
          "K771978",
          "K955253",
          "K172135",
          "K893820",
          "K801446",
          "K103424",
          "K193454",
          "K151833",
          "K173637",
          "K884670",
          "K131431",
          "K210760",
          "K791153",
          "K073153",
          "K232412",
          "K013522",
          "K772123",
          "K170789",
          "K050867",
          "K072737",
          "K132087",
          "K162314",
          "K982631",
          "K993645",
          "K113483",
          "K133590",
          "K991553",
          "K230162",
          "K861708",
          "K142697",
          "K123585",
          "K083524",
          "K902643",
          "K052587",
          "K170273",
          "K010641",
          "K180727",
          "K903622",
          "K980176",
          "K093787",
          "K133649",
          "K170238",
          "K233698",
          "K192402",
          "K033418",
          "K162838",
          "K192061",
          "K250970",
          "K970819",
          "K113287",
          "K170177",
          "K113063",
          "K063507",
          "K242411",
          "K041521",
          "K241757",
          "K113341",
          "K964747",
          "K133603",
          "K173244",
          "K964583",
          "K171201",
          "K052157",
          "K812393",
          "K954255",
          "K213999",
          "K022239",
          "K223028",
          "K183271",
          "K251645",
          "K053148",
          "K232136",
          "K150665",
          "K770573",
          "K230370",
          "K911387",
          "K102604",
          "K081847",
          "K082816",
          "K063029",
          "K171198",
          "K133643",
          "K232491",
          "K193277",
          "K915623",
          "K093581",
          "K923077",
          "K760808",
          "K790942",
          "K043111",
          "K932508",
          "K212875",
          "K051765",
          "K944983",
          "K771811",
          "K053610",
          "K033774",
          "K032475",
          "K944013",
          "K001088",
          "K052447",
          "K143409",
          "K160723",
          "K180393",
          "K201034",
          "K043087",
          "K242811",
          "K040577",
          "K171738",
          "K884668",
          "K082458",
          "K103489",
          "K201231",
          "K964746",
          "K152584",
          "K014240",
          "K171850",
          "K990134",
          "K993791",
          "K232022",
          "K882581",
          "K213927",
          "K241079",
          "K231482",
          "K253574",
          "K143416",
          "K033102",
          "K964890",
          "K120910",
          "K122106",
          "K200016",
          "K000359",
          "K884669",
          "K022794",
          "K955583",
          "K241713",
          "K812400",
          "K820688",
          "K151383",
          "K781752",
          "K213465",
          "K223758",
          "K955254",
          "K930090",
          "K780930",
          "K171261",
          "K170023",
          "K253173",
          "K040224",
          "K982004",
          "K885322",
          "K080581",
          "K213938",
          "K024171",
          "K043469",
          "K222360",
          "K914305",
          "K780769",
          "K991766",
          "K231631",
          "K042229",
          "K073346",
          "K213829",
          "K203448",
          "K221727",
          "K120796",
          "K813196",
          "K253520",
          "K170716",
          "K233767",
          "K910859",
          "K192188",
          "K160611",
          "K142222",
          "K223514"
        ],
        "registration_number": [
          "",
          "3033527283",
          "2134243",
          "8043933",
          "3026047529",
          "3007225492",
          "3015537315",
          "3013847349",
          "3009529630",
          "3029521953",
          "3010789868",
          "3027338963",
          "3007233544",
          "3011825597",
          "3012085143",
          "3002808157",
          "3004091527",
          "3003678543",
          "3005930550",
          "3012044688",
          "1047843",
          "2134213",
          "3019216",
          "3011050570",
          "3036666444",
          "3016965352",
          "3004056159",
          "3008114965",
          "3009963139",
          "3010123525",
          "3006487372",
          "1066019",
          "3003967489",
          "3009543051",
          "3010642685",
          "3009077524",
          "3011015597",
          "3033527241",
          "3003202425",
          "3018596740",
          "9612283",
          "3017425975",
          "3007797756",
          "3016007746",
          "3002807636",
          "2011171",
          "3014246639",
          "2520313",
          "2134850",
          "3018094310",
          "3011802799",
          "3008355361",
          "3031503890",
          "2032112",
          "3011797532",
          "3004962788",
          "3005156333",
          "3011642792",
          "1423253",
          "3004938766",
          "3017500061",
          "2247992",
          "3006610845",
          "3015169361",
          "3011416394",
          "3017426576",
          "3018423337",
          "3010281549",
          "3008007310",
          "1000246715",
          "3013035022",
          "3008632256",
          "9613299",
          "3015276088",
          "3008284470",
          "3013875850",
          "3023140882",
          "3012127758",
          "3023003446",
          "1317307",
          "3010601176",
          "3032109181",
          "3012938564",
          "3030982377",
          "1834003",
          "3012957335",
          "3013679558",
          "3010326005",
          "3003575350",
          "3010703925",
          "3027339477",
          "3012453401",
          "3010398867",
          "3017210169",
          "3017375271",
          "3006209933",
          "3027245002",
          "3000126629",
          "3002907620",
          "3027373189",
          "8043728",
          "3043512431",
          "3033524251",
          "9613445",
          "3007562060",
          "3005810333",
          "3015777306",
          "3014134519",
          "3015060224",
          "3004198450",
          "3009154005",
          "3014938447",
          "3008496839",
          "3009196578",
          "2126677",
          "3014653800",
          "3004977335",
          "3011396687",
          "1034973",
          "3004753774",
          "9614698",
          "3011950073",
          "3004083736",
          "3015232217",
          "1054713",
          "3017894900",
          "3010151377",
          "3039211310",
          "3012761544",
          "3005877899",
          "1811755",
          "8022257",
          "3001722928",
          "3015763655",
          "3042175844",
          "3021059265",
          "3035950510",
          "3032109225",
          "9617978",
          "9616567",
          "3016619943",
          "3006621332",
          "3014343100",
          "3014809646",
          "3004857110",
          "3033527282",
          "3003477135",
          "3006894636",
          "1123169",
          "2183744",
          "1226188",
          "3003094912",
          "3042172882",
          "3026970882",
          "3012939903",
          "2030624",
          "9611343",
          "3004993756",
          "3015223550",
          "3026391023",
          "3007545404",
          "3008769756",
          "3023460764",
          "3011786147",
          "3003768277"
        ],
        "fei_number": [
          "",
          "3033527283",
          "3026047529",
          "3007225492",
          "3015537315",
          "3013847349",
          "3009529630",
          "3029521953",
          "3010789868",
          "3027338963",
          "3007233544",
          "3011825597",
          "3012085143",
          "3002808157",
          "3004091527",
          "3003678543",
          "3005930550",
          "3012044688",
          "1047843",
          "3003079232",
          "3001654007",
          "3019216",
          "3011050570",
          "3036666444",
          "3016965352",
          "3004056159",
          "3008114965",
          "3009963139",
          "3010123525",
          "3006487372",
          "3003967489",
          "3009543051",
          "3010642685",
          "3000206435",
          "1000139922",
          "3009077524",
          "3011015597",
          "3033527241",
          "3003202425",
          "3018596740",
          "3004571294",
          "3007797756",
          "3016007746",
          "1000114111",
          "3014447948",
          "3002807636",
          "2011171",
          "3003076831",
          "3014246639",
          "2520313",
          "3018094310",
          "3011802799",
          "3002619595",
          "3008355361",
          "3031503890",
          "3003757840",
          "3011797532",
          "3002807903",
          "3004962788",
          "3005156333",
          "3011642792",
          "1000284313",
          "3004938766",
          "3017500061",
          "3006610845",
          "3001726808",
          "3015169361",
          "3011416394",
          "3017426576",
          "3018423337",
          "3010281549",
          "3008007310",
          "1000246715",
          "3013035022",
          "3008632256",
          "3015276088",
          "3008284470",
          "3013875850",
          "3023140882",
          "3012127758",
          "3023003446",
          "1317307",
          "3010601176",
          "3032109181",
          "3012938564",
          "3030982377",
          "3012957335",
          "3013679558",
          "3010326005",
          "3003575350",
          "3010703925",
          "1000514697",
          "3027339477",
          "3012453401",
          "3010398867",
          "3017210169",
          "3017375271",
          "3006209933",
          "3027245002",
          "3000126629",
          "3002907620",
          "3027373189",
          "3043512431",
          "3002963732",
          "3033524251",
          "3007562060",
          "3004198450",
          "3005810333",
          "3015777306",
          "3014134519",
          "3015060224",
          "3009154005",
          "3014938447",
          "3008496839",
          "3009196578",
          "3008342029",
          "2126677",
          "3014653800",
          "3004977335",
          "3011396687",
          "1034973",
          "3004753774",
          "3011950073",
          "3004083736",
          "3015232217",
          "3017894900",
          "3010151377",
          "3039211310",
          "3012761544",
          "3005877899",
          "1811755",
          "3003007876",
          "1000123143",
          "3002808529",
          "1000208827",
          "3001722928",
          "3015763655",
          "3042175844",
          "3021059265",
          "3035950510",
          "3032109225",
          "3016619943",
          "3006621332",
          "3002807880",
          "3014343100",
          "3014809646",
          "3004857110",
          "3033527282",
          "3003477135",
          "3006894636",
          "2183744",
          "3003094912",
          "3042172882",
          "3002175190",
          "3026970882",
          "3012939903",
          "3008581796",
          "3004993756",
          "3015223550",
          "3026391023",
          "3007545404",
          "3002808382",
          "3008769756",
          "3023460764",
          "3011786147"
        ],
        "device_name": "System, X-Ray, Tomography, Computed",
        "medical_specialty_description": "Radiology",
        "regulation_number": "892.1750",
        "device_class": "2"
      }
    }
  ]
}