  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-08-20",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "217054",
      "product_res_number": "Z-0779-2026",
      "event_date_initiated": "2025-11-03",
      "event_date_posted": "2025-12-01",
      "recall_status": "Open, Classified",
      "res_event_number": "97896",
      "product_code": "QPO",
      "product_description": "Gut Microbiome Test Medium",
      "code_info": "EAN: 616612785886;\nSKU: 2A111;\nUDI-DI: None;\nLot/Serial Number: All Lots;",
      "firm_fei_number": "3031995890",
      "recalling_firm": "GET TESTED INTERNATIONAL AB",
      "address_1": "59334 Vastervik 18 vastervik lysingsvagen",
      "city": "Vastervik",
      "postal_code": "59334",
      "country": "Sweden",
      "reason_for_recall": "Distribution without premarket approval/clearance.",
      "root_cause_description": "No Marketing Application",
      "action": "On November 3, 2025, MEDICAL DEVICE RECALL letters were sent to customers. Actions to be taken: If you still have an unused test, it should not be used. You should render the kit unusable and discard the defaced components in accordance with ordinary household disposal practices and applicable local regulations. If your kit has already been used: You do not need to return used components. Please do not base medical decisions solely on that prior result and do not use any remaining units. If you have questions about your results or your health, consult your healthcare provider, who may recommend retesting using an FDA-authorized test that is appropriate for your situation. You may still contact us through the form for support or goodwill assistance regarding your order. Forwarding requirement for further distribution: If you further distributed the affected product(s) to other locations or customers, please immediately forward this notice to all downstream recipients and ensure they stop use/stop distribution, quarantine, and follow the disposition instructions.",
      "product_quantity": "3 units",
      "distribution_pattern": "US Nationwide distribution.",
      "openfda": {
        "k_number": [
          "DEN200040"
        ],
        "registration_number": [
          "3022474803",
          "3004142665",
          "3003742580"
        ],
        "fei_number": [
          "3022474803",
          "3004142665",
          "3003742580"
        ],
        "device_name": "Device To Preserve And Stabilize Relative Abundances Of Microbial Nucleic Acids In Clinical Samples",
        "medical_specialty_description": "Microbiology",
        "regulation_number": "866.2952",
        "device_class": "2"
      }
    }
  ]
}