  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-08-13",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "216190",
      "product_res_number": "Z-0063-2026",
      "event_date_initiated": "2025-09-11",
      "event_date_posted": "2025-10-16",
      "recall_status": "Open, Classified",
      "res_event_number": "97441",
      "product_code": "EOQ",
      "k_numbers": [
        "K121959"
      ],
      "product_description": "EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-Q190",
      "code_info": "Model No. BF-Q190; UDI: 4953170434792, 4953170335198; All Serial No.",
      "firm_fei_number": "2429304",
      "recalling_firm": "Olympus Corporation of the Americas",
      "address_1": "3500 Corporate Pkwy",
      "city": "Center Valley",
      "state": "PA",
      "postal_code": "18034-8229",
      "additional_info_contact": "Cynthia Ow\n647-9993203",
      "reason_for_recall": "Additional IFU updates to provide further clarification on safe and effective use of bronchoscopes when used in conjunction with laser, argon plasma coagulation, and high-frequency therapy equipment.",
      "root_cause_description": "Device Design",
      "action": "An URGENT Medical Device Correction notification dated 9/11/25 was mailed to consignees to notify them of this recall action. Consignees are asked to carefully read the recall notification and identify any affected devices in their inventory. Consignees are to ensure all personnel are knowledgeable and aware of the updated IFU. Consignees are to acknowledge receipt of the recall notification through Olympus' web portal at https://olympusamerica.com/recall. Forward the provided recall notification if product was further distributed. Consignees with any questions can contact Cynthia Ow by phone at 647-999-3203 or by email at Cynthia.Ow@olympus.com.",
      "product_quantity": "916 units US; 1,402 units OUS",
      "distribution_pattern": "US Nationwide Distribution.",
      "openfda": {
        "k_number": [
          "K952081",
          "K990547",
          "K244017",
          "K822257",
          "K201300",
          "K221631",
          "K090853",
          "K895415",
          "K152466",
          "K182188",
          "K844834",
          "K252045",
          "K911412",
          "K223144",
          "K233671",
          "K954707",
          "K923847",
          "K071530",
          "K130845",
          "K233779",
          "K040940",
          "K220399",
          "K252341",
          "K250862",
          "K121959",
          "K242070",
          "K933433",
          "K926430",
          "K910423",
          "K024165",
          "K152319",
          "K945268",
          "K150280",
          "K832492",
          "K962378",
          "K832010",
          "K251731",
          "K143727",
          "K944649",
          "K121035",
          "K770500",
          "K211947",
          "K141366",
          "K963033",
          "K960791",
          "K833483",
          "K941264",
          "K060243",
          "K892687",
          "K223836",
          "K142909",
          "K211493",
          "K934920",
          "K023376",
          "K221340",
          "K142839",
          "K210379",
          "K941161",
          "K250863",
          "K223782",
          "K963252",
          "K914181",
          "K944976",
          "K082720",
          "K122535",
          "K950925",
          "K954797",
          "K950636",
          "K983124",
          "K980046",
          "K213235",
          "K240075",
          "K252528",
          "K220957",
          "K183419",
          "K822844",
          "K251823",
          "K942388",
          "K240289",
          "K250118",
          "K202370",
          "K240135",
          "K061313",
          "K822136",
          "K962736",
          "K792342",
          "K033225",
          "K162611",
          "K931442",
          "K832449",
          "K211030",
          "K961178",
          "K200342",
          "K151315",
          "K950786",
          "K081518",
          "K911922",
          "K131028",
          "K161656",
          "K951051",
          "K962555",
          "K250232",
          "K870063",
          "K071473",
          "K040018",
          "K201758",
          "K183256",
          "K091250",
          "K962533",
          "K200702",
          "K192367",
          "K232038",
          "K013731",
          "K792267",
          "K260565",
          "K022446",
          "K191948",
          "K962367",
          "K231970",
          "K210928",
          "K171291",
          "K022546",
          "K241612",
          "K896146",
          "K163248",
          "K230948",
          "K231521",
          "K893006",
          "K200397",
          "K211169",
          "K960880",
          "K023984",
          "K921393",
          "K242793",
          "K951199",
          "K232582",
          "K962380",
          "K950965",
          "K163377",
          "K223299",
          "K193488",
          "K213635",
          "K202987",
          "K171492",
          "K242324",
          "K231107",
          "K944650",
          "K250210",
          "K172955",
          "K222861",
          "K222014",
          "K243541",
          "K212667",
          "K822180",
          "K212656",
          "K050220",
          "K173727",
          "K142934",
          "K982983",
          "K854581",
          "K212048",
          "K922937",
          "K862835",
          "K202180",
          "K895948",
          "K131946",
          "K854414",
          "K251752",
          "K222910",
          "K031966",
          "K760171",
          "K250957",
          "K963795",
          "K981543",
          "K213782",
          "K954125",
          "K172726",
          "K183085",
          "K242416",
          "K232860",
          "K970107",
          "K201146",
          "K940521",
          "K232984",
          "K260420",
          "K173760",
          "K012073",
          "K253137",
          "K984157",
          "K091768",
          "K220606",
          "K182253",
          "K974642",
          "K221302",
          "K013111",
          "K910536",
          "K183654",
          "K200678",
          "K210883",
          "K922204",
          "K193534",
          "K983878",
          "K234113",
          "K983122",
          "K811218",
          "K173554",
          "K050907",
          "K220299",
          "K131234",
          "K953596",
          "K192164",
          "K251894",
          "K183607",
          "K915816",
          "K962485",
          "K920025",
          "K021344",
          "K260382",
          "K183516",
          "K993900",
          "K951826",
          "K232601",
          "K131401",
          "K163537",
          "K242010",
          "K230428",
          "K943700",
          "K941158",
          "K811059",
          "K962901",
          "K243497",
          "K931154",
          "K011267",
          "K831240",
          "K811181",
          "K182825",
          "K254037",
          "K212886",
          "K202346",
          "K243219",
          "K191828"
        ],
        "registration_number": [
          "",
          "3004837686",
          "3012322232",
          "3012494290",
          "1220477",
          "3013666218",
          "3013992977",
          "3018094310",
          "2244478",
          "3004065885",
          "3014246639",
          "9613846",
          "3015136743",
          "3029929134",
          "2243757",
          "3008386005",
          "9611680",
          "1223925",
          "3009039068",
          "3012448339",
          "2032112",
          "3010684280",
          "9681260",
          "3015895145",
          "3011063460",
          "2030624",
          "3007487007",
          "9610877",
          "9611997",
          "3010157088",
          "3005099803",
          "9681147",
          "2029015",
          "3015142721",
          "3017616393",
          "3025749544",
          "3012797688",
          "3007695959",
          "1000484830",
          "3013557562",
          "3012627460",
          "3023102955",
          "9680515",
          "9610691",
          "3004962788",
          "3016555232",
          "3009719225",
          "3016832577",
          "3014663814",
          "3009171251",
          "3023460764",
          "3033509898",
          "3014635169",
          "1221485",
          "3004111573",
          "3013077751",
          "9710524",
          "3006984716",
          "9611102",
          "3015316279",
          "1650907",
          "3012843412",
          "3022631687",
          "2011171",
          "3026787544",
          "8010273",
          "3023809956",
          "2124215",
          "1314417",
          "3015139943",
          "3016607092",
          "3020746799",
          "3010396739",
          "3027205723",
          "1000393132",
          "3005465238",
          "3042172882",
          "1721686",
          "3005369776",
          "8030673",
          "3038613760",
          "2242464",
          "3009420598",
          "3020331900",
          "1220592",
          "3014769442",
          "3007738736",
          "3014342096",
          "3007867778",
          "3016045249",
          "3019807891",
          "3006621103",
          "3003183807",
          "1526611",
          "1037905",
          "3001675293",
          "8010047",
          "3017820152",
          "3004644065",
          "3003761012",
          "3038182369",
          "3010383847",
          "3022472",
          "8040233",
          "3010398867",
          "3003782610",
          "3013283620",
          "3006174295",
          "3042225166",
          "3004040397",
          "3010432890",
          "1000121056",
          "3015153469",
          "3002808148",
          "3010273872",
          "9610595",
          "3015970743",
          "3020369549",
          "3004142795",
          "3026606306",
          "2020550",
          "3010041511",
          "3010885402",
          "3015224777",
          "3030504644",
          "3017425975",
          "3010202439",
          "3017859830",
          "3015612252",
          "3017140456",
          "3024021261",
          "3010220595",
          "3014491269",
          "2183744",
          "3006546082",
          "2246552",
          "3008309062",
          "3007305485",
          "1320894",
          "3013440400",
          "3017367014",
          "3010280860",
          "3015925378",
          "3025349506",
          "3035710892",
          "3017619652",
          "3010273867",
          "1828132",
          "3021325287",
          "3011050570",
          "3005382983",
          "3021031769",
          "3010421104",
          "3016096100",
          "3043226252",
          "3005987240",
          "3014247261",
          "3042940623",
          "3013528571",
          "3011137372",
          "3013868968",
          "3010493171",
          "1037955",
          "3023214418",
          "9615393",
          "3033628477",
          "3006786907",
          "1220850",
          "2938071",
          "3008439484",
          "3033536312",
          "3008040402",
          "3007084728",
          "3013298431",
          "3003637092",
          "3010707607",
          "3006161756",
          "9614641",
          "3016678045",
          "1221826",
          "3025603301",
          "3013665681",
          "3036756245",
          "9610617",
          "1319639",
          "3004083736",
          "3014605202",
          "3040391040",
          "3008729892",
          "3003678543",
          "1833700",
          "3011416394"
        ],
        "fei_number": [
          "",
          "3004837686",
          "3012322232",
          "3012494290",
          "3003076831",
          "1220477",
          "3013666218",
          "1000160458",
          "3013992977",
          "3001451571",
          "3018094310",
          "3004065885",
          "3014246639",
          "3015136743",
          "3029929134",
          "2243757",
          "3008386005",
          "3009039068",
          "3012448339",
          "3010684280",
          "3015895145",
          "3011063460",
          "3007487007",
          "3010157088",
          "3005099803",
          "3014447948",
          "3015142721",
          "3017616393",
          "3025749544",
          "3004213967",
          "3012797688",
          "3007695959",
          "1000484830",
          "3013557562",
          "3005747797",
          "3002806565",
          "3012627460",
          "3023102955",
          "3004962788",
          "3003944071",
          "3016555232",
          "3009719225",
          "3016832577",
          "3014663814",
          "3023460764",
          "3033509898",
          "3014635169",
          "3002808268",
          "1221485",
          "3004111573",
          "3013077751",
          "3006984716",
          "1000491840",
          "1000138054",
          "3015316279",
          "3012843412",
          "3022631687",
          "2011171",
          "3026787544",
          "3023809956",
          "3006493107",
          "2124215",
          "1314417",
          "3015139943",
          "3016607092",
          "3020746799",
          "3010396739",
          "3027205723",
          "1000393132",
          "3005465238",
          "3042172882",
          "1721686",
          "3005369776",
          "3006759904",
          "3038613760",
          "2242464",
          "3009420598",
          "3020331900",
          "1220592",
          "3014769442",
          "3007738736",
          "3014342096",
          "3007867778",
          "3016045249",
          "3019807891",
          "3006621103",
          "3003183807",
          "1526611",
          "1037905",
          "3002807969",
          "3001675293",
          "3017820152",
          "3004644065",
          "3003761012",
          "3038182369",
          "3010383847",
          "3022472",
          "3010398867",
          "3003782610",
          "3013283620",
          "3006174295",
          "3042225166",
          "3004040397",
          "3010432890",
          "1000121056",
          "3015153469",
          "3003499455",
          "3007475226",
          "3010273872",
          "3002808148",
          "3003995201",
          "3002681132",
          "3002808001",
          "3015970743",
          "3020369549",
          "3004142795",
          "3026606306",
          "2020550",
          "3010041511",
          "3010885402",
          "3015224777",
          "3030504644",
          "3017859830",
          "3015612252",
          "3017140456",
          "3024021261",
          "3010220595",
          "3014491269",
          "2183744",
          "3006546082",
          "2246552",
          "3008309062",
          "3007305485",
          "3008185830",
          "3013440400",
          "3017367014",
          "3010280860",
          "3015925378",
          "3025349506",
          "1000115586",
          "3035710892",
          "3017619652",
          "3010273867",
          "1828132",
          "3021325287",
          "3011050570",
          "3005969827",
          "3002807769",
          "3005382983",
          "3002807555",
          "1000305720",
          "3021031769",
          "3016096100",
          "3043226252",
          "3010421104",
          "3005987240",
          "4486",
          "3014247261",
          "3042940623",
          "3013528571",
          "3013868968",
          "3010493171",
          "1037955",
          "3005947796",
          "3023214418",
          "3033628477",
          "3006786907",
          "1220850",
          "2938071",
          "3008439484",
          "3033536312",
          "3008040402",
          "3007084728",
          "3002806359",
          "3002807795",
          "3013298431",
          "3003637092",
          "3002964398",
          "3010707607",
          "1221826",
          "3016678045",
          "3025603301",
          "3013665681",
          "3036756245",
          "3004083736",
          "3014605202",
          "3040391040",
          "1000136269",
          "3003678543",
          "3002808127",
          "3011416394",
          "3000206435"
        ],
        "device_name": "Bronchoscope (Flexible Or Rigid)",
        "medical_specialty_description": "Ear, Nose, Throat",
        "regulation_number": "874.4680",
        "device_class": "2"
      }
    }
  ]
}