{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-08-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "211576",
      "product_res_number": "Z-0842-2025",
      "event_date_initiated": "2022-03-14",
      "event_date_posted": "2025-01-10",
      "recall_status": "Completed",
      "res_event_number": "95913",
      "product_code": "DWJ",
      "product_description": "Augustine Surgical Inc. HOT DOG PATIENT WARMING, Temperature Management Controller, Single Port, Models:  \n1) WC71V;    \n2) WC71V-DEMO;   \n3) WC71V-R; \nused in conjunction with Warming Blankets and Mattresses, veterinary use.",
      "code_info": "1) UDI/DI 00850011479247, Serial Numbers:  20090015, 20090016, 20090017, 20090018, 20090020, 20090021, 20090022, 20090023, 20090024, 20090025, 20090026, 20090027, 20090028, 20090029, 20090030, 20090031, 20090032, 20090033, 20090034, 20090035, 20090036, 20090037, 20090038, 20090039, 20090040, 20090041, 20090042, 20090043, 20090044, 20100030, 20100031, 20100032, 20100033, 20100034, 20100035, 20100036, 20100037, 20100038, 20100040, 20100041, 20100042, 20100043, 20100044, 20100045, 20100046, 20100047, 20100048, 20100049, 20100050, 20100051, 20100052, 20100053, 20100054, 20100055, 20100056, 20100057, 20100058, 20100059, 20100091, 20100092, 20100093, 20100094, 20100095, 20100096, 20100097, 20100098, 20100099, 20100100, 20100101, 20100103, 20110000, 20110001, 20110002, 20110003, 20110004, 20110005, 20110006, 20110006, 20110007, 20110008, 20110009, 20110010, 20110011, 20110012, 20110013, 20110014, 20110015, 20110016, 20110017, 20110018, 20110019, 20110020, 20110021, 20110022, 20110023, 20110024, 20110025, 20110026, 20110027, 20110028, 20110029, 20110030, 20110031, 20110032, 20110033, 20110034, 20110035, 20110036, 20110037, 20110038, 20110039, 20110040, 20110041, 20110042, 20110043, 20110044, 20110061, 20110062, 20110063, 20110064, 20110065, 20110066, 20110067, 20110069, 20110070, 20110071, 20110072, 20110073, 20110074, 20110075, 20110077, 20110078, 20110079, 20110080, 20110081, 20110082, 20110083, 20110084, 20110085, 20110086, 20110087, 20110088, 20110089, 20110094, 20110105, 20110106, 20110109, 20110110, 20110111, 20110112, 20110113, 20110115, 20110116, 20110118, 20110119, 20110120, 20110121, 20110123, 20110124, 20110125, 20110126, 20110127, 20110128, 20110129, 20110131, 20110132, 20110133, 20110134, 20110135, 20110136, 20110137, 20110138, 20110139, 20110140, 20110141, 20110142, 20110143, 20110145, 20110146, 20110147, 20110148, 20110149, 20120045, 20120046, 20120047, 20120048, 20120049, 20120050, 20120051, 20120052, 20120053, 20120054, 20120055, 20120056, 20120057, 20120058, 20120059, 20120060, 20120061, 20120062, 20120063, 20120064, 20120066, 20120067, 20120068, 20120069, 20120070, 20120071, 20120073, 20120074, 20120075, 20120076, 20120077, 20120078, 20120079, 20120080, 20120081, 20120082, 20120083, 20120084, 20120086, 20120087, 20120089, 21010000, 21010001, 21010002, 21010003, 21010004, 21010005, 21010006, 21010007, 21010008, 21010009, 21010010, 21010011, 21010012, 21010013, 21010014, 21010015, 21010016, 21010017, 21010018, 21010019, 21010020, 21010021, 21010022, 21010023, 21010024, 21010025, 21010026, 21010027, 21010028, 21010029, 21010060, 21010061, 21010062, 21010063, 21010064, 21010065, 21010066, 21010067, 21010068, 21010069, 21010070, 21010071, 21010072, 21010073, 21010074, 21010075, 21010076, 21010078, 21010079, 21010080, 21010081, 21010082, 21010083, 21010084, 21010085, 21010087, 21010088, 21010089, 21010090, 21010092, 21010093, 21010094, 21010095, 21010096, 21010097, 21010098, 21010099, 21010100, 21010102, 21010103, 21010104, 21020060, 21020061, 21020062, 21020063, 21020064, 21020065, 21020066, 21020067, 21020068, 21020069, 21020071, 21020072, 21020073, 21020074, 21020075, 21020076, 21020077, 21020078, 21020079, 21020080, 21020081, 21020082, 21020083, 21020084, 21020085, 21020086, 21020087, 21020088, 21020089, 21030000, 21030001, 21030002, 21030003, 21030004, 21030005, 21030006, 21030007, 21030008, 21030009, 21030010, 21030011, 21030012, 21030013, 21030014, 21030015, 21030016, 21030017, 21030018, 21030019, 21030020, 21030022, 21030023, 21030024, 21030025, 21030026, 21030027, 21030028, 21030029, 21030060, 21030061, 21030062, 21030063, 21030064, 21030065, 21030066, 21030067, 21030068, 21030069, 21030071, 21030072, 21030073, 21030074, 21030075, 21030076, 21030077, 21030078, 21030079, 21030080, 21030081, 21030082, 21030084, 21030085, 21030086, 21030087, 21030088, 21030089, 21040030, 21040031, 21040032, 21040033, 21040034, 21040035, 21040036, 21040037, 21040038, 21040039, 21040040, 21040041, 21040042, 21040043, 21040044, 21040045, 21040046, 21040047, 21040048, 21040049, 21040050, 21040051, 21040052, 21040053, 21040054, 21040055, 21040056, 21040057, 21040058, 21040059, 21040105, 21040106, 21040107, 21040108, 21040109, 21040110, 21040111, 21040112, 21040113, 21040114, 21040115, 21040116, 21040117, 21040118, 21040119, 21040120, 21040121, 21040122, 21040123, 21040125, 21040126, 21040127, 21040128, 21040129, 21040130, 21040131, 21040132, 21040133, 21040134, 21040135, 21040136, 21040138, 21040139, 21040140, 21040141, 21040142, 21040143, 21040144, 21040145, 21040146, 21040147, 21040148, 21040149, 21040150, 21040152, 21040153, 21040154, 21040155, 21040156, 21040157, 21040158, 21040159, 21040160, 21040161, 21040162, 21040164, 21060000, 21060001, 21060002, 21060003, 21060004, 21060006, 21060007, 21060008, 21060010, 21060011, 21060013, 21060014, 21060015, 21060016, 21060017, 21060018, 21060019, 21060020, 21060022, 21060023, 21060024, 21060025, 21060027, 21060028, 21060029, 21060030, 21060031, 21060032, 21060033, 21060034, 21060035, 21060036, 21060037, 21060038, 21060039, 21060040, 21060041, 21060042, 21060043, 21060044, 21060135, 21060136, 21060137, 21060138, 21060139, 21060140, 21060141, 21060142, 21060143, 21060144, 21060145, 21060146, 21060147, 21060149, 21060150, 21060151, 21060152, 21060153, 21060154, 21060155, 21060156, 21060157, 21060158, 21060159, 21060160, 21060161, 21060162, 21060163, 21060164, 21060165, 21060166, 21060167, 21060168, 21060169, 21060170, 21060171, 21060172, 21060174, 21060175, 21060176, 21060177, 21070123, 21070124, 21070125, 21070126, 21070127, 21070128, 21070129, 21070130, 21070131, 21070132, 21070134, 21070136, 21070137, 21070138, 21070139, 21070140, 21070141, 21070142, 21070143, 21070144, 21070145, 21070146, 21070147, 21070148, 21070149, 21070150, 21070151, 21070152, 21070153, 21070154, 21070155, 21070156, 21070157, 21070158, 21070159, 21070160, 21070161, 21070162, 21070163, 21070164, 21070165, 21070166, 21070167, 21070168, 21070170, 21070171, 21070172, 21070173, 21070174, 21070176, 21070177, 21070178, 21070179, 21090000, 21090001, 21090002, 21090003, 21090004, 21090005, 21090006, 21090007, 21090008, 21090009, 21090010, 21090011, 21090012, 21090013, 21090014, 21090015, 21090016, 21090017, 21090018, 21090019, 21090020, 21090021, 21090024, 21090025, 21090027, 21090028, 21090029, 21090030, 21090031, 21090032, 21090033, 21090034, 21090035, 21090036, 21090037, 21090038, 21090039, 21090040, 21090041, 21090042, 21090043, 21090044, 21100000, 21100001, 21100002, 21100003, 21100004, 21100005, 21100006, 21100007, 21100008, 21100009, 21100010, 21100011, 21100012, 21100013, 21100014, 21100015, 21100016, 21100017, 21100018, 21100019, 21100020, 21100021, 21100022, 21100023, 21100024, 21100025, 21100026, 21100027, 21100028, 21100029, 21100030, 21100031, 21100032, 21100033, 21100034, 21100035, 21100036, 21100037, 21100038, 21100039, 21100040, 21100041, 21100042, 21100043, 21100044, 21100045, 21100046, 21100047, 21100048, 21100050, 21100051, 21100052, 21100053, 21100054, 21100055, 21100057, 21100058, 21100059, 21110065, 21110067, 21110068, 21110069, 21110070, 21110071, 21110072, 21110073, 21110074, 21110075, 21110077, 21110078, 21110079, 21110080, 21110081, 21110083, 21110084, 21110085, 21110086, 21110087, 21110088, 21110089, 21110090, 21110091, 21110092, 21110093, 21110095, 21110096, 21110097, 21110098, 21110099, 21110100, 21110101, 21110102, 21110103, 21110104, 21110105, 21110106, 21110107, 21110108, 21110109, 21110110, 21110111, 21110112, 21110113, 21110114, 21110115, 21110117, 21110118, 21110120, 21110121, 21110122, 21110123, 21110124, 21120000, 21120001, 21120002, 21120003, 21120004, 21120005, 21120006, 21120007, 21120008, 21120009, 21120010, 21120011, 21120012, 21120013, 21120014, 21120015, 21120016, 21120017, 21120018, 21120019, 21120020, 21120021, 21120022, 21120023, 21120024, 21120025, 21120026, 21120027, 21120028, 21120029, 21120030, 21120031, 21120032, 21120033, 21120034, 21120035, 21120036, 21120037, 21120038, 21120039, 21120040, 21120041, 21120042, 21120043, 21120044, 21120045, 21120046, 21120047, 21120048, 21120049, 21120050, 21120051, 21120052, 21120053, 21120054, 21120055, 21120056, 21120057, 21120058, 21120059, 21120060, 21120061, 21120062, 21120063, 21120064, 21120065, 21120066, 21120067, 21120068, 21120069, 21120070, 21120071, 21120072, 21120073, 21120074, 22010000, 22010001, 22010002, 22010003, 22010004, 22010005, 22010006, 22010007, 22010008, 22010009, 22010010, 22010011, 22010012, 22010013, 22010014, 22020000, 22020001, 22020002, 22020003, 22020004, 22020005, 22020006, 22020007, 22020008, 22020009, 22020010, 22020011, 22020012, 22020013, 22020014, 22020015, 22020016, 22020017, 22020018, 22020019, 22020020, 22020021, 22020022, 22020023, 22020024, 22020025, 22020026, 22020027, 22020028, 22020029, 22020030, 22020031, 22020032, 22020033, 22020034, 22020035, 22020036, 22020037, 22020038, 22020039, 22020040, 22020041, 22020042, 22020043, 22020044, 22020045, 22020046, 22020047, 22020049, 22020050, 22020051, 22020052, 22020053, 22020054, 22020055, 22020056, 22020057, 22020058, 22020059, 22030090, 22030091, 22030092, 22030093, 22030094, 22030095, 22030096, 22030097, 22030098, 22030099, 22030100, 22030101, 22030102, 22030103, 22030104, 22030105, 22030106, 22030107, 22030108, 22030109, 22030110, 22030112, 22030113, 22030114, 22030115, 22030116, 22030117, 22030118, 22030119, 22030120, 22030121, 22030122, 22030123, 22030124, 22030125, 22030126, 22030127, 22030128, 22030129, 22030130, 22030131, 22030132, 22030133, 22030134, 22030135, 22030136, 22030137, 22030138, 22030139, 22030140, 22030141, 22030142, 22030143, 22030144, 22030145, 22030146, 22030147, 22030148, 22030149, 22040090, 22040091, 22040092, 22040093, 22040094, 22040095, 22040096, 22040097, 22040098, 22040099, 22040100, 22040101, 22040103, 22040104, 22040105, 22040106, 22040107, 22040108, 22040109, 22040110, 22040111, 22040112, 22040113, 22040114, 22040115, 22040116, 22040117, 22040118, 22040119, 22040120, 22040121, 22040122, 22040123, 22040124, 22040125, 22040126, 22040127, 22040128, 22040129, 22040130, 22040131, 22040132, 22040133, 22040134, 22040135, 22040136, 22040137, 22040138, 22040139, 22040140, 22040141, 22040142, 22040143, 22040144, 22040145, 22040146, 22040147, 22040148, 22040149, 22040150, 22040151, 22040152, 22040153, 22040154, 22040155, 22040156, 22040157, 22040158, 22040159, 22040160, 22040161, 22040162, 22040163, 22040164, 22040165, 22040166, 22040167, 22040168, 22040169, 22040170, 22040171, 22040172, 22040173, 22040174, 22040175, 22040177, 22040178, 22040179, 22040180, 22040181, 22040182, 22040183, 22040184, 22040185, 22040186, 22040187, 22040188, 22040189, 22040190, 22040191, 22040193, 22040194, 22040195, 22040196, 22040197, 22040198, 22040199, 22040200, 22040201, 22040202, 22040203, 22040204, 22040205, 22040206, 22040207, 22040208, 22040209, 22050060, 22050061, 22050062, 22050063, 22050064, 22050065, 22050066, 22050067, 22050068, 22050069, 22050070, 22050071, 22050072, 22050073, 22050074, 22050075, 22050076, 22050077, 22050078, 22050079, 22050080, 22050081, 22050082, 22050083, 22050084, 22050085, 22050086, 22050087, 22050088, 22050089, 22070030, 22070031, 22070032, 22070033, 22070034, 22070035, 22070036, 22070037, 22070038, 22070039, 22070040, 22070041, 22070042, 22070043, 22070044, 22080000, 22080001, 22080002, 22080004, 22080005, 22080006, 22080007, 22080008, 22080009, 22080010, 22080011, 22080012, 22080013, 22080014, 22080015, 22080016, 22080017, 22080018, 22080019, 22080020, 22080021, 22080022, 22080023, 22080024, 22080025, 22080026, 22080027, 22080028, 22080029, 22080030, 22080031, 22080032, 22080033, 22080034, 22080035, 22080036, 22080037, 22080038, 22080039, 22080040, 22080041, 22080043, 22080044, 22080045, 22080046, 22080047, 22080048, 22080049, 22080050, 22080051, 22080052, 22080053, 22080054, 22080055, 22080057, 22080058, 22080059, 22080120, 22080121, 22080122, 22080123, 22080124, 22080125, 22080126, 22080127, 22080128, 22080129, 22080130, 22080131, 22080132, 22080133, 22080134, 22080135, 22080136, 22080137, 22080138, 22080139, 22080140, 22080141, 22080142, 22080143, 22080144, 22080145, 22080146, 22080147, 22080148, 22080149, 22080150, 22080151, 22080152, 22080153, 22080154, 22080155, 22080156, 22080157, 22080158, 22080159, 22080160, 22080161, 22080162, 22080163, 22080164, 22090110, 22090111, 22090112, 22090113, 22090114, 22090115, 22090116, 22090117, 22090118, 22090119, 22090120, 22090121, 22090122, 22090123, 22090124, 22090125, 22090126, 22090127, 22090128, 22090129, 22090130, 22090131, 22090132, 22090133, 22090134, 22090135, 22090136, 22090137, 22090138, 22090139, 22090140, 22090141, 22090142, 22090143, 22090144, 22090145, 22090146, 22090147, 22090148, 22090149, 22090150, 22090151, 22090152, 22090153, 22090154, 22090155, 22090156, 22090157, 22090158, 22090159, 22090160, 22090161, 22090162, 22090163, 22090164, 22090165, 22090166, 22090167, 22090168, 22090169, 22090171, 22090172, 22090173, 22090174, 22090175, 22090176, 22090177, 22090178, 22090179, 22090180, 22090181, 22090182, 22090183, 22090184, 22090185, 22090186, 22090187, 22090188, 22090189, 22090190, 22090191, 22090192, 22090193, 22090194, 22090195, 22090196, 22090197, 22090198, 22090199, 22110000, 22110001, 22110002, 22110003, 22110004, 22110005, 22110006, 22110007, 22110008, 22110009, 22110010, 22110011, 22110012, 22110013, 22110014, 22110015, 22110016, 22110017, 22110018, 22110019, 22110020, 22110021, 22110022, 22110023, 22110024, 22110025, 22110026, 22110027, 22110028, 22110029, 22110030, 22110031, 22110032, 22110033, 22110034, 22110035, 22110036, 22110037, 22110038, 22110039, 22110040, 22110041, 22110042, 22110043, 22110044, 22110045, 22110046, 22110047, 22110048, 22110049, 22110050, 22110051, 22110052, 22110053, 22110054, 22110055, 22110057, 22110058, 22110059, 22110060, 22110061, 22110062, 22110063, 22110064, 22110065, 22110066, 22110067, 22110068, 22110069, 22110070, 22110072, 22110073, 22110074, 22110075, 22110076, 22110077, 22110078, 22110079, 22110080, 22110081, 22110082, 22110083, 22110084, 22110085, 22110086, 22110087, 22110088, 22110089, 22110090, 22110091, 22110092, 22110093, 22110095, 22110096, 22110097, 22110098, 22110099, 22110100, 22110101, 22110102, 22110103, 22110104, 22110105, 22110106, 22110107, 22110108, 22110109, 22110110, 22110111, 22110112, 22110113, 22110114, 22110115, 22110116, 22110117, 22110118, 22110119, 23010091, 23010092, 23010093, 23010095, 23010096, 23010097, 23010098, 23010099, 23010100, 23010101, 23010102, 23010103, 23010104, 23010105, 23010106, 23010107, 23010108, 23010109, 23010110, 23010111, 23010112, 23010113, 23010115, 23010116, 23010117, 23010118, 23010119, 23010120, 23010121, 23010122, 23010123, 23010124, 23010125, 23010126, 23010127, 23010128, 23010129, 23010130, 23010131, 23010132, 23010133, 23010134, 23010135, 23010136, 23010137, 23010138, 23010139, 23010140, 23010141, 23010142, 23010143, 23010144, 23010145, 23010146, 23010147, 23010148, 23010149, 23010150, 23010151, 23010152, 23010153, 23010154, 23010155, 23010156, 23010157, 23010158, 23010159, 23010160, 23010161, 23010162, 23010163, 23010164, 23010165, 23030001, 23030002, 23030003, 23030004, 23030005, 23030006, 23030007, 23030008, 23030009, 23030010, 23030011, 23030012, 23030013, 23030014, 23030015, 23030016, 23030017, 23030018, 23030019, 23030020, 23030021, 23030022, 23030023, 23030024, 23030025, 23030026, 23030027, 23030028, 23030029, 23030030, 23030031, 23030032, 23030033, 23030034, 23030035, 23030036, 23030037, 23030038, 23030039, 23030040, 23030041, 23030042, 23030043, 23030044, 23030045, 23040001, 23040002, 23040003, 23040004, 23040005, 23040006, 23040007, 23040008, 23040009, 23040010, 23040011, 23040012, 23040013, 23040014, 23040015, 23040016, 23040017, 23040018, 23040019, 23040020, 23040021, 23040022, 23040023, 23040024, 23040025, 23040026, 23040027, 23040028, 23040029, 23040030, 23040031, 23040032, 23040033, 23040034, 23040035, 23040036, 23040037, 23040038, 23040039, 23040040, 23040041, 23040042, 23040043, 23040044, 23040045, 23060076, 23060077, 23060078, 23060079, 23060080, 23060081, 23060082, 23060083, 23060084, 23060085, 23060086, 23060087, 23060088, 23060089, 23060090, 23060091, 23060092, 23060093, 23060094, 23060095, 23060096, 23060097, 23060098, 23060099, 23060100, 23060101, 23060102, 23060103, 23060104, 23060105, 23060106, 23060107, 23060108, 23060109, 23060110, 23060112, 23060113, 23060114, 23060115, 23060116, 23060117, 23060118, 23060119, 23060120, 23060121, 23060122, 23060123, 23060124, 23060125, 23060126, 23060127, 23060128, 23060129, 23060130, 23060131, 23060132, 23060133, 23060134, 23060135, 23070061, 23070062, 23070063, 23070064, 23070065, 23070066, 23070067, 23070068, 23070069, 23070070, 23070072, 23070073, 23070074, 23070075, 23070076, 23070077, 23070078, 23070079, 23070080, 23070081, 23070082, 23070083, 23070084, 23070085, 23070086, 23070087, 23070088, 23070089, 23070090, 23070091, 23070092, 23070093, 23070094, 23070095, 23070096, 23070097, 23070098, 23070099, 23070100, 23070101, 23070102, 23070103, 23070104, 23070105, 23070106, 23070107, 23070108, 23070109, 23070111, 23070112, 23070113, 23070114, 23070115, 23070116, 23070117, 23070118, 23070119, 23070120, 23090001, 23090002, 23090003, 23090004, 23090005, 23090006, 23090007, 23090008, 23090009, 23090010, 23090011, 23090012, 23090013, 23090014, 23090015, 23090016, 23090017, 23090018, 23090019, 23090020, 23090021, 23090022, 23090023, 23090024, 23090025, 23090026, 23090027, 23090028, 23090029, 23090030, 23090031, 23090032, 23090033, 23090034, 23090035, 23090036, 23090037, 23090038, 23090039, 23090040, 23090041, 23090042, 23090043, 23090044, 23090045, 23090046, 23090047, 23090048, 23090049, 23090050, 23090051, 23090052, 23090053, 23090054, 23090055, 23090056, 23090057, 23090058, 23090059, 23090060, 23110001, 23110002, 23110003, 23110004, 23110005, 23110006, 23110007, 23110008, 23110009, 23110010, 23110011, 23110012, 23110013, 23110014, 23110015, 23110016, 23110017, 23110018, 23110019, 23110020, 23110021, 23110022, 23110023, 23110024, 23110026, 23110027, 23110028, 23110029, 23110030, 23110031, 23110032, 23110033, 23110034, 23110035, 23110036, 23110037, 23110038, 23110039, 23110040, 23110041, 23110042, 23110043, 23110044, 23110045, 23110046, 23110047, 23110048, 23110049, 23110050, 23110051, 23110052, 23110053, 23110054, 23110055, 23110056, 23110057, 23110058, 23110059, 23110060, 23110181, 23110182, 23110183, 23110184, 23110185, 23110186, 23110187, 23110188, 23110189, 23110190, 23110191, 23110192, 23110193, 23110194, 23110195, 23110196, 23110197, 23110198, 23110199, 23110200, 23110201, 23110202, 23110203, 23110204, 23110205, 23110206, 23110207, 23110208, 23110209, 23110210, 23110211, 23110212, 23110213, 23110214, 23110215, 23110216, 23110217, 23110218, 23110219, 23110220, 23110221, 23110222, 23110223, 23110224, 23110225, 23110226, 23110227, 23110228, 23110229, 23110230, 23110231, 23110232, 23110233, 23110234, 23110235, 23110236, 23110237, 23110238, 23110239, 23110240, 23110241, 23110242, 23110243, 23110244, 23110245, 23110246, 23110247, 23110248, 23110249, 23110250, 23110251, 23110252, 23110253, 23110254, 23110255, 23110256, 23110257, 23110258, 23110259, 23110260, 23110262, 23110263, 23110264, 23110265, 23110266, 23110267, 23110268, 23110269, 23110270, 23120121, 23120122, 23120123, 23120124, 23120125, 23120126, 23120127, 23120128, 23120129, 23120130, 23120131, 23120132, 23120133, 23120134, 23120135, 23120136, 23120137, 23120138, 23120139, 23120140, 23120141, 23120142, 23120143, 23120144, 23120145, 23120146, 23120147, 23120148, 23120149, 23120150, 23120151, 23120152, 23120153, 23120154, 23120155, 23120156, 23120158, 23120159, 23120160, 23120161, 23120162, 23120163, 23120164, 23120165, 24010047, 24010048, 24010049, 24010050, 24010051, 24010052, 24010053, 24010055, 24010056, 24010057, 24010058, 24010059, 24010060, 24010061, 24010062, 24010063, 24010064, 24010065, 24010066, 24010067, 24010068, 24010069, 24010070, 24010071, 24010072, 24010073, 24010074, 24010075, 24010076, 24010077, 24010078, 24010079, 24010080, 24010081, 24010082, 24010083, 24010084, 24010085, 24010086, 24010087, 24010088, 24010089, 24010090, 24010091, 24010092, 24010093, 24010094, 24010095, 24010096, 24010097, 24010098, 24010099, 24010100, 24010101, 24010102, 24010103, 24010104, 24010105, 24030031, 24030032, 24030033, 24030034, 24030035, 24030036, 24030037, 24030038, 24030039, 24030040, 24030041, 24030042, 24030043, 24030044, 24030045, 24030046, 24030047, 24030048, 24030049, 24030050, 24030051, 24030052, 24030053, 24030054, 24030055, 24030056, 24030057, 24030058, 24030059, 24030060, 24030061, 24030062, 24030063, 24030064, 24030065, 24030066, 24030067, 24030068, 24030069, 24030070, 24030071, 24030072, 24030073, 24030074, 24030075, 24030076, 24030077, 24030078, 24030079, 24030080, 24030081, 24030082, 24030083, 24030084, 24030085, 24030086, 24030087, 24030088, 24030089, 24030090, 24030091, 24030092, 24030093, 24030094, 24030095, 24030096, 24030097, 24030098, 24030099, 24030100, 24030101, 24030102, 24030103, 24030104, 24030106, 24030107, 24030108, 24030109, 24030110, 24030111, 24030112, 24030113, 24030114, 24030115, 24030116, 24030117, 24030118, 24030119, 24030120; \n2) UDI/DI 00850011479247, Serial Numbers:  20100090, 20110076, 21030021; \n3) UDI/DI 00850011479247, Serial Numbers:  22030111, 22040102, 22040192, 22110056, 23010094, 23070071, 23070110, 23110261, 24010046.",
      "firm_fei_number": "3005857264",
      "recalling_firm": "Augustine Temperature Management, LLC",
      "address_1": "15305 Minnetonka Blvd",
      "city": "Minnetonka",
      "state": "MN",
      "postal_code": "55345-1512",
      "additional_info_contact": "Nic Raph\n952-465-3526",
      "reason_for_recall": "HotDog Temperature Management Controllers Models WC71, WC77, and Temperature Management Controller MP Model 2083516 may experience a Power-on self-test error code \"EA POST\".",
      "root_cause_description": "Device Design",
      "action": "Augustine Temperature Management, LLC began correcting (installing ferrites and software updates) the affected devices February, 2022 and completed the actions in August 2023.  The firm did not perform any notification to its consignees that corrections would be implemented.",
      "product_quantity": "2271 units",
      "distribution_pattern": "Worldwide",
      "openfda": {
        "k_number": [
          "K914383",
          "K920265",
          "K093244",
          "K001149",
          "K051419",
          "K812955",
          "K202197",
          "K903360",
          "K902650",
          "K930345",
          "K942599",
          "K232508",
          "K913734",
          "K772015",
          "K140635",
          "K051336",
          "K014210",
          "K811742",
          "K943276",
          "K881322",
          "K051278",
          "K123083",
          "K942790",
          "K001546",
          "K053645",
          "K024128",
          "K073712",
          "K171373",
          "K950416",
          "K921165",
          "K921920",
          "K180834",
          "K093694",
          "K051097",
          "K142702",
          "K920432",
          "K231596",
          "K080763",
          "K152266",
          "K040632",
          "K251974",
          "K002577",
          "K910327",
          "K100585",
          "K071918",
          "K141686",
          "K003349",
          "K232638",
          "K210605",
          "K180375",
          "K171149",
          "K210603",
          "K011859",
          "K200446",
          "K123946",
          "K102112",
          "K820814",
          "K061513",
          "K121720",
          "K252861",
          "K923104",
          "K200225",
          "K232502",
          "K022903",
          "K190221",
          "K221669",
          "K946196",
          "K033021",
          "K070112",
          "K882442",
          "K072513",
          "K021473",
          "K040911",
          "K770839",
          "K003368",
          "K231211",
          "K112488",
          "K800812",
          "K242964",
          "K132048",
          "K113705",
          "K161602",
          "K981093",
          "K140078",
          "K230866",
          "K952329",
          "K991684",
          "K854856",
          "K023934",
          "K933764",
          "K964377",
          "K101092",
          "K261437",
          "K092807",
          "K031728",
          "K873745",
          "K914382",
          "K083662",
          "K102856",
          "K181699",
          "K973705",
          "K011907",
          "K162562",
          "K882035",
          "K171234",
          "K771537",
          "K094056",
          "K232627",
          "K243942",
          "K873546",
          "K952864",
          "K940171",
          "K120849",
          "K904690",
          "K913016",
          "K101705",
          "K110104",
          "K061902",
          "K964673",
          "K162679",
          "K890575",
          "K801253",
          "K020604",
          "K101148",
          "K041686",
          "K203085",
          "K060865",
          "K260278",
          "K131361",
          "K121669",
          "K960167",
          "K132454",
          "K761087",
          "K952086",
          "K992386",
          "K071341",
          "K912051",
          "K181671",
          "K811743",
          "K914805",
          "K201779",
          "K014121",
          "K220941",
          "K900186",
          "K083336",
          "K083135",
          "K960511",
          "K902756",
          "K942881",
          "K260903",
          "K120081",
          "K062794",
          "K101589",
          "K963731",
          "K040440",
          "K232844",
          "K970367",
          "K894980",
          "K052392",
          "K010338"
        ],
        "registration_number": [
          "2133810",
          "3014426939",
          "8022590",
          "3005997949",
          "3009746061",
          "3012421607",
          "3021073836",
          "3043226252",
          "3013557562",
          "3010380518",
          "3029871637",
          "3030597679",
          "2011171",
          "3044156918",
          "2029015",
          "3038241934",
          "3016844398",
          "2025816",
          "3013188547",
          "3005857264",
          "3010317211",
          "3004491421",
          "3010168054",
          "9611590",
          "9617473",
          "1516825",
          "3041276890",
          "3035108304",
          "3008879776",
          "1055823",
          "3012689738",
          "3003857828",
          "1219702",
          "1066427",
          "1725056",
          "3005941719",
          "3005238200",
          "3012494290",
          "3004362278",
          "3031841433",
          "1043572",
          "3018094310",
          "3036108847",
          "3012972101",
          "9616096",
          "9680284",
          "9614306",
          "3010476117",
          "3013319212",
          "9616567",
          "1722070",
          "1000122786",
          "3042272797",
          "1043214",
          "3013410562",
          "3014421917",
          "3009093866",
          "3003542872",
          "3008352384",
          "3009609358",
          "1018233",
          "1060680",
          "1417592",
          "3011059079",
          "3010617000",
          "1055236",
          "3012315097",
          "1836518",
          "3007309299",
          "3006240053",
          "2936999",
          "2530138",
          "3008727520",
          "1220351",
          "3014263632",
          "2432320",
          "3032911850",
          "1000484830",
          "2183744",
          "1054241",
          "1043644",
          "2320762",
          "2133641",
          "9617604",
          "2032112",
          "3016507522",
          "3034393266",
          "3013041594",
          "3007616088",
          "9610849",
          "2183613"
        ],
        "fei_number": [
          "1000208624",
          "3014426939",
          "3005997949",
          "3009746061",
          "3012421607",
          "3021073836",
          "3043226252",
          "3003793491",
          "3013557562",
          "3010380518",
          "3029871637",
          "1000221205",
          "3002807111",
          "1000111140",
          "1000160458",
          "1000117260",
          "3030597679",
          "2011171",
          "3044156918",
          "3038241934",
          "3016844398",
          "2025816",
          "3000203466",
          "3003807323",
          "3013188547",
          "3005857264",
          "3010317211",
          "3004491421",
          "3010168054",
          "3001733054",
          "1516825",
          "3041276890",
          "3035108304",
          "3003076831",
          "3007513035",
          "3008879776",
          "1000237915",
          "3002558692",
          "1000112877",
          "3012689738",
          "3003857828",
          "1219702",
          "3003040625",
          "3005238200",
          "3012494290",
          "3004362278",
          "3002809016",
          "3003261020",
          "3031841433",
          "1043572",
          "3018094310",
          "3036108847",
          "3012972101",
          "3010476117",
          "3013319212",
          "1000122786",
          "3042272797",
          "1000112968",
          "1043214",
          "3002807391",
          "3013410562",
          "3495",
          "3014421917",
          "3009093866",
          "3003542872",
          "3008352384",
          "3009609358",
          "1018233",
          "3003329224",
          "1417592",
          "3011059079",
          "3003753496",
          "3012315097",
          "3007309299",
          "3006240053",
          "1000521377",
          "2936999",
          "3008727520",
          "1220351",
          "3014263632",
          "3032911850",
          "1000484830",
          "2183744",
          "1043644",
          "3016507522",
          "3034393266",
          "3000203066",
          "3002963732",
          "3013041594",
          "3007616088",
          "2183613"
        ],
        "device_name": "System, Thermal Regulating",
        "medical_specialty_description": "Cardiovascular",
        "regulation_number": "870.5900",
        "device_class": "2"
      }
    }
  ]
}