{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-08-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "210858",
      "product_res_number": "Z-0593-2025",
      "event_date_initiated": "2024-10-31",
      "event_date_posted": "2024-12-03",
      "recall_status": "Open, Classified",
      "res_event_number": "95636",
      "product_code": "MDB",
      "k_numbers": [
        "K915796"
      ],
      "product_description": "BD BACTEC FX, Instrument Top, Packaged\nVersion or Model: 441385\nCatalog Number: 441385",
      "code_info": "Catalog No: 441385; UDI-DI: 00382904413859; Serial Numbers: FT9750, FT8660, FT9839, FT5241, FT9840, FT9841, T10574, FT5251, FT9796, FT1325, FT9880, FT9881, T10575, FT9884, FT7169, FT7177, T10563, T10519, FT9734, FT7161, FT7181, T10536, FT7756, FT0538, FT8653, T10008, FT7203, FT7197, FT7714, T10821, FT7192, FT7201, T10037, T10042, T10817, T10615, T10629, T10746, T12324, T10716, FT7215, T10011, FT7222, T10834, T12452, T12455, FT7204, T10845, FT7209, T10945, FT5382, FT0043, T10758, FT0055, T10589, FT0071, T10289, T10293, T12380, T10547, FT0059, T10550, FT0106, T12372, FT0110, FT7232, FT7680, FT7220, FT0116, FT0084, FT7211, FT7480, FT7489, FT0120, FT7219, T10836, T10683, FT7213, T10791, FT7245, T12357, FT7504, FT7619, T12511, FT7675, FT0282, FT7679, T12379, FT7715, FT7246, T12195, T12442, T11036, T12215, FT0383, T12200, T12418, T12212, FT7256, FT0373, FT5317, FT0341, T12202, FT0615, T12294, T12209, T12308, FT5261, FT7978, T12206, T12482, FT5263, T12213, FT7270, T12229, FT0384, FT0385, FT5266, T12128, T12233, FT0436, FT0666, FT0671, FT7281, FT0560, T12003, FT7276, T12225, FT0118, FT5386, FT5388, FT0434, FT7307, FT5389, FT0374, T12205, FT0519, T12134, FT7303, T12222, FT0521, T12646, T12226, T12228, FT0007, T12789, T12254, T12234, FT7311, FT0500, T12885, T12124, FT0414, FT0631, T12901, FT0894, T12615, T12232, FT0966, FT0659, T12637, FT0437, FT7358, T12259, FT0785, FT7366, FT0748, T12498, T12325, T13006, FT7360, FT0784, FT0808, T13184, FT0799, FT3882, T13182, T12349, FT5429, T20193, FT0653, T20198, FT7380, T12327, FT0951, T12456, FT0893, FT0881, T10797, FT0701, T20230, FT2230, FT7390, T12047, FT0923, T20232, FT0878, T20300, FT0921, T12048, FT1022, T12046, T10963, T10957, T10695, T10798, FT1303, FT5141, FT1177, T10691, FT4841, T10696, FT1219, T10702, FT1053, T10728, FT0996, FT7785, FT1711, FT1069, FT7787, FT7788, FT3603, FT1714, FT7789, FT8201, T10792, FT1296, FT7452, FT5665, T10793, FT1145, FT5670, FT1241, T10701, T10969, FT1759, FT5439, FT0058, FT4270, FT7793, FT2848, FT5441, FT2849, FT5447, FT2852, FT2853, FT1629, FT3347, FT1626, FT3348, FT1096, FT3895, FT1277, FT4037, FT4274, FT4583, FT7405, FT2823, FT7407, FT1323, FT7406, FT2762, T10855, FT1607, FT1006, T10858, T12135, FT0906, FT0406, FT7822, FT2234, FT1489, FT3594, FT8691, FT1528, FT1537, FT1552, T12179, T10865, FT1591, T12183, FT1091, T10873, FT1593, FT1696, FT8497, T10860, FT1089, FT1646, FT1592, T12533, FT7839, FT1632, FT1609, FT1170, FT0930, FT0788, FT1719, FT7837, FT1747, FT2583, FT1728, T10331, T10449, T12139, T12163, T12473, FT1754, FT1087, T12170, T10867, FT2683, T12028, FT7849, FT1917, FT7854, T12164, FT7827, T10321, FT6277, FT7866, T10337, T10333, FT1367, FT1780, FT7838, FT2501, FT7841, FB1310, T12168, T12042, FT1748, FT1850, FT1101, T12905, FT7860, FT1169, T10953, FT1864, FT3551, FT7900, T10753, T12238, FT1994, FT2042, FT7883, FT2019, FT7894, FT7907, FT7911, T12766, FT1171, T12041, FT7930, FT2176, FT7908, T10979, T12040, FT2206, FT2038, FT1337, T12060, FT0424, FT7906, T12044, FT2044, FT7909, T12059, FT8780, FT2342, FT2311, FT8814, FT8739, T12074, FT2393, FT8743, FT7913, FT3359, T12072, T13060, T13005, FT2430, FT1710, FT1583, T13017, T13065, FT2942, FT2534, FT9158, FT2535, FT9148, FT2496, FT2464, FT2537, FT5762, T12105, FT2490, FT2491, T13082, FT2426, FT2647, FT8268, FT2545, FT8253, T10973, T11030, FT7919, FT2722, T13191, FT1023 BFX#2, T12187, FT2544, T12201, FT2574, FT2580, FT2991, T12197, FT1020 BFX#4, T12198, FT2694, FT2715, FT2723, FT2718, FT2709, T11021, FT2802, FT1816, T12110, FT2803, T13125, FT7439A, FT7959, FT2805, FT7962, FT7965, T10992, T10999, T13255, FT1344, T11024, FT2755, FT2800, FT0171, T12339, FT1385, T12512, FT0716, FT2763, FT2819, T12378, FT2954, T12381, T12387, FT7979, FT7602, FT1717, FT6050, FT2956, FT2717, T12501, FT2829, FT1824, FT1471, FT1470, FT7929, FT2851, FT2892, FT7954, T12568, FT7956, FT2894, FT2923, FT7971, FT7981, FT1588, FT2367, FT2133, FT2870, FT1476, T12363, FT0933, FT4523, T12514, FT2921, FT7336, FT3057, T12565, T12567, FT7509, T12569, FT2992, FT7994, T12584, FT2004, T12090, T12585, FT2728, T12607, FT2757, FT4363, FT2203, FT3060, FT7992, FT7782, FT7997, FT7447, FT2972, FT7244, FT8049, FT5046, FT8502, FT3054, FT3605, FT2990, FT2519, FT3058, FT0935, FT9231, FT3247, FT4225, FT8335, FT8003, FT1098, FT1181, FT3324, FT3111, FT5564, FT3333, FT8288, FT3338, FT7991, FT8538, T12601, FT3246, FT3239, FT7543, T10174, T10176, FT2764, T10398, T12596, FT8882, T10395, T12606, FT3386, T10507, T12133, FT3334, FT8958, FT3350, T12488, FT3372, FT9068, FT8358, FT3373, FT9383, FT3371, T12486, FT3345, FT1484, FT7704, FT2940, T12839, T12633, FT3410, T12640, FT3409, T13185, T12671, FT9254, FT3388, T10160, T12655, FT3441, FT9250, FT1113, T13206, T10684, T12459, T12597, FT3428, FT2383, FT3440, FT8507, FT2387, FT8010, FT2482, FT0070, FT3462, FT1360, FT2329, FT9813, FT8008, FT2124, T12659, FT2327, T12617, FT3532, FT0321, T12618, FT1358, FT3533, FT8004, T12652, FT3482, FT8014, FT1362, FT3534, FT3483, FT2385, FT3539, FT3541, FT3543, FT5757, FT4657, FT3546, T12628, FT2123, T12629, FT5218, FT5225, FT8039, FT8040, T10517, FT3589, FT8001, FT8012, FT3586, FT8019, T12644, FT8017, FT4957, FT8027, FT8029, FT3300, FT8047, T10784, FT3194, FT1779, FT1483, FT3306, FT3540, FT1485, FT8052, T12740, FT3621, T12718, FT3645, FT1242, FT3893, FT8058, FT3899, FT8050, FT1254, FT4105, FT1104, FT8053, FT8056, FT8057, FT1460, FT1791, FT3885, FT3876, FT1255, FT3934, FT0859, T12974, T13027, FT3930, T13095, T13099, FT8064, FT1331, FT3984, FT7242, FT3996, FT3299, FT7248, FT3997, FT7296, FT3197, FT7362, FT7435, FT0099, FT1838, FT3977, FT7432, FT0308, FT1322   MIB Rev 11, FT7440, FT4035, FT7371, FT4102, FT4196, FT0036, FT8136, FT0440, FT7445, FT4131, FT0439, FT4759, FT4130, FT1321, FT0075, FT8051, FT0392, FT8066, FT4263, FT3186, FT0685, FT0872, FT1739, FT4763, FT0772, FT4284, FT0897, FT6061, FT4291, FT6134, FT4148, FT1010, FT0898, FT1099, FT3008, FT1017, FT4155, FT1139, FT1140, FT1076, FT1212, FT1431, FT6147, FT1777, FT1430, FT8121, FT1429, FT0714, FT1477, FT1570, FT7623, FT8113, FT1590, FT1809, FT4220, FT1589, FT4343, FT2400, FT2355, FT2590, FT8119, FT2948, FT1925, FT4428, FT8120, FT4346, FT2059, FT2893, FT2120, FT2871, FT2697, FT8144, FT7635, FT4147, FT8148, FT2112, FT7689, FT7690, FT4179, FT4206, FT2381, FT4360, FT7703, FT2565, FT4413, FT4446, FT7832, FT3464, FT3308, FT3443, FT7924, FT3467, FT4194, FT0034, FT4510, FT8246, FT0322, FT7801, FT3613, FT0561, FT1810, FT4525, FT0743, FT3999, FT4498, FT0773, FT4541, FT0858, FT8073, FT4453, FT0938, FT0838, FT4251, FT1078, FT1118, FT7008, FT4545, FT4254, FT1106, FT4521, FT5613, FT7043, FT3810, FT1279, FT7089, FT5468, FT1280, FT4543, FT1329, FT7110, FT4504, FT5470, FT1481, FT4567, FT5633, FT1794, FT7146, FT5113, FT1795, FT4655, FT1812, FT7162, FT8225, FT5638, FT8363, FT1922, FT3009, FT1814, FT2739, FT4626, FT2860, FT1815, FT7239, FT4618, FT2924, FT5639, FT4621, FT5640, FT4632, FT2858, FT7572, FT4628, FT3032, FT1954, FT4640, FT2854, FT1951, FT5646, FT4606, FT2009, FT7579, FT8337, FT5477, FT2113, FT4633, FT8359, FT2010, FT4607, FT2122, FT7581, FT5253, FT2255, FT4663, FT4561, FT8366, FT2224, FT3445, FT2319, FT2316, FT2384, FT2204, FT7764, FT2483, FT5459, FT3729, FT2484, FT4680, FT5475, FT2527, FT3769, FT4490, FT3622, FT4139, FT7823, FT3641, FT3639, FT7828, FT4697, FT5560, FT4076, FT4806, FT4736, FT7833, FT4505, FT7773, FT8385, FT4095, FT4786, FT5466, FT4802, FT4813, FT1861, FT8071, FT2657, FT7836, FT4776, FT4785, FT5084, FT3527, FT3733, FT4779, FT8105, FT4897, FT4905, FT4781, FT3592, FT4855, FT9426, FT3593, FT3513, FT4078, FT3720, FT9032, FT3597, FT3600, FT4073, FT9087, FT3815, FT9268, FT4074, FT4079, FT3816, FT8183, FT3736, FT3820, FT8191, FT3744, FT3821, FT3624, FT3739, FT8108, FT9471, FT4083, FT3831, FT9468, FT3740, FT2767, FT9479, FT2916, FT6016, FT3542, FT9486, FT8188, FT4864, FT9870, FT4885, FT8394, FT4907, FT4908, FT4909, T12399, FT4921, T12397, FT3651, FT4923, FT5498, FT4925, FT9489, FT8297, FT4935, FT9492, FT4939, FT4952, FT4961, FT4967, FT4996, T12188, FT5030, FT5502, FT8312, FT9550, T12383, FT2071, FT4200, FT8055, FT8382, FT8432, FT8559, FT4338, FT8384, FT5109, FT4330, T12143, FT5029, FT2795, FT8514, FT8395, FT5069, FT8400, FT8402, FT8424, FT5076, FT8429, FT5079, FT5068, FT8709, T10373, T10381, FT8722, FT8688, FT8721, FT8389, FT5170, FT8391, FT8682, FT5204, FT7334, FT8692, FT5200, FT7339, T12765, FT3864, T10254, FT8694, T13014, T13015, T13069, FT5224, FT3878, FT8695, FT5173, FT8697, T13132, FT2564, FT3135, FT5222, FT5181, FT7471, FT8746, FT8494, FT4177, FT8747, FT5158, FT8476, FT5167, FT8473, FT5228, FT5700, FT5175, FT5215, FT8763, FT7419, FT8773, T10340, FT5587, FT8755, FT8479, FT8808, FT5322, FT8478, FT8485, FT8819, FT8487, FT5595, FT7603, FT5318, FT6254, FT5584, FT5586, FT6246, FT8751, FT5591, FT5361, FT5589, FT6249, FT5597, FT5364, FT6251, FT5316, FT8804, FT8813, FT5372, FT7069, FT7088, FT8821, FT5572, FT8519, FT8077, FT5339, FT5189, FT8522, FT8528, FT8948, FT8530, FT8825, FT8534, FT8537, FT5676, FT5341, FT5472, FT8953, FT7574, FT7776, FT5698, FT8712, FT8085, FT8087, FT4565, FT8166, FT8338, FT8291, FT8025, FT8552, FT0886, FT5385, FT8941, FT8233, FT8554, FT1744, FT8943, FT5479, FT8232, FT8208, FT5320, FT8209, FT1038 BFX#6, FT8217, FT8293, FT8328, FT8295, FT9036, FT5430, FT8221, FT9309, FT5376, FT8223, FT8318, FT8320, FT9004, FT5480, FT8433, FT8438, FT8321, FT8421, FT8330, FT8423, T12023, FT8417, FT8326, FT8418, FT0889, FT8442, FT0970, FT5467, FT8436, FT9156, FT5478, FT8444, FT8331, FT9157, FT8448, FT8345, FT8419, FT9163, FT5327, FT0896, FT8759, FT5321, FT8587, FT9172, FT8555, FT0974, FT5319, FT0976, FT8450, FT8913, FT9166, FT5530, FT8607, FT9244, FT1036, FT5594, FT5526, FT9243, FT5583, FT8980, FT8979, FT5378, FT5377, FT9113, FT9247, FT5505, FT8598, FT0900, FT9131, FT5580, FT9344, FT9248, FT5582, FT9449, FT9249, FT5592, FT8594, FT9450, FT5611, FT8616, FT9251, FT8619, FT9253, FT9025, FT5454, FT9028, FT9027, FT9704, FT9026, FT9705, FT9029, FT1040, FT9115, FT5683, FT9090, FT1052, FT9733, FT9749, FT8706, FT5108, FT9751, FT9084, FT1278, FT1148, FT9204, FT1373, FT9323, FT5702, FT9759, FT9885, FT5723, FT9214, FT9893, FT5695, FT1906, FT5706, FT9490, FT9727, FT9319, FT9728, FT9729, FT9730, FT7201A, FT8665, FT8680, FT9539, T10112, FT9538, FT8664, FT5769, FT1372, FT5758, FT5727, T10118, FT8667, T10088, FT5795, T10203, FT9541, FT1638, T10650, T12313, FT9544, T12314, FT7227, FT5849, FT1377, T12312, FT7294, FT1519, FT9534, FT0628, FT8659, FT5880, FT8677, FT8689, FT7121, FT8690, FT7437, FT1295, FT7427, FT7469, T12319, FT5821, FT1379, T12588, FT1434, FT9535, FT9533, FT7549, T12589, FT1516, FT8683, T12590, FT1661, FT6086, FT5836, FT1752, FT1854, FT5840, FT8699, FT1928, FT8705, FT1386, T12032, FT5847, FT2211, FT2212, T12039, FT5894, T12146, FT7693, T12178, FT0255, FT5900, T12276, FT5813, T12281, FT5824, T12282, FT1527, T12285, FT8700, T12286, FT9604, FT9605, FT2642, T12289, FT2644, FT3326, FT3360, FT0917, FT2281, FT2285, FT2571, FT1090, T10001, FT2568, FT3927, FT3923, FT2166, FT3941, FT2168, FT5919, FT3924, FT2167, FT3925, FT5930, FT4441, T10097, FT5927, FT3955, FT0176, FT0192, FT0195, FT5600, FT3991, T10089, FT3786, FT8752, FT0132, T12621, FT9631, FT8754, FT4242, FT5983, FT4342, FT1745, T12681, T10108, FT5984, FT4359, FT8720, T12721, T10109, FT4435, FT8757, FT4436, FT8758, T12768, T10096, FT4433, T12829, FT4455, FT1070, T12956, FT1388, FT2284, FT4844, FT1387, FT4828, FT1389, FT4770, FT2282, T13176, FT4878, T13152, T10005, FT5250, FT1034, FT6026, FT4936, T13173, FT4934, FT6023, FT6020, FT6022, T13177, FT6021, FT2873, FT1247, FT6005, T13178, FT5994, FT4637, FT4713, FT4727, FT5697, FT0801, FT5991, FT5992, FT7557, FT0916, FT4747, FT7701, FT6095, FT6171, FT4405, FT7716, FT6056, FT6039, FT2175, FT2180, FT2646, FT6204, FT7561, FT0752, FT0750, FT6258, FT0749, FT0140, FT2003, FT8793, FT0148, FT0884, FT1050, FT5335, FT4126, FT4116, FT5445, FT4108, FT5675, FT0188, FT0181, FT5344, FT5345, FT6079, FT5351, FT6096, FT0305, FT1871, FT5563, T10208, FT1870, FT0459, FT5645, FT0558, FT1874, FT5312, FT3505, T10511, FT5647, FT6102, FT3318, FT5649, FT3120, FT0153, FT1877, FT6132, FT3041, FT1539, FT0637, FT8794, FT5735, FT1825, T10628, FT8796, FT5756, FT1829, T10709, FT4133, T10698, FT0866, FT4109, FT6025, FT3806, FT6015, FT2618, FT2949, FT3367, FT3702, FT6030, FT3379, FT4052, FT1532, FT2515, FT3804, FT0465, FT1231, FT5740, FT0081, FT1882, FT1940, FT0101, FT3108, FT3210, FT2511, T12125, FT1509, FT1137, FT6223, FT1533, FT1162, FT6237, FT2233, FT1265, FT1266, T12440, FT1352, T10976, FT5780, FT1449, FT8837, FT1766, FT6127, FT3887, FT1475, FT6219, FT1763, FT5782, FT2566, T12336, FT0693, FT2158, FT3212, FT5484, FT3688, FT5922, FT4891, FT1989, FT5923, FT1541, FT4887, FT3380, FT2839, FT3183, FT1507, FT4906, FT0815, FT6232, FT4919, FT3378, FT6077, FT2461, FT2637, T12950, FT4110, FT3493, FT6284, FT0956, FT2838, FT0873, FT4463, FT0678, FT4464, FT1136, FT2216, FT5815, FT1447, FT5820, FT1351, FT5123, FT1762, FT5098, FT3381, T10998, FT4694, FT1941, FT4954, FT0768, FT5009, FT1291, FT6261, FT4722, FT1938, FT5575, FT3670, FT0687, FT0718, FT6199, FT0352, FT5958, FT8820, FT0865, FT8849, FT3892, FT0869, FT4161, FT6273, FT1370, FT6272, FT8846, FT2267, FT0719, FT4013, FT3812, FT3832, FT0722, FT4195, T10283, FT1365, T10291, FT2963, FT3167, T12919, FT2891, T12920, FT3165, T10308, FT8749, FT8817, FT4773, FT8840, T12723, T12729, T12192, FT7717, FT5476, T12745, FT7175, FT5885, FT5659, FT5660, FT1740, FT5701, FT5952, FT5711, FT5717, FT1741, FT5982, T10760, FT8878, FT6040, FT7273, FT2159, FT4654, FT7331, FT1904, FT7381, FT6146, T13128, FT7369, FT7361, FT7449, FT8873, FT8870, FT8900, T10894, FT8906, FT1175, FT1639, FT7298, FT7462, FT7516, T10892, FT7147, FT7607, FT3817, FT7540, FT1640, FT1952, FT3909, FT7610, FT1134, FT8895, FT6160, FT8907, FT2962, FT1956, T10905, FT3763, T10907, FT6169, FT8904, FT3525, T10878, FT7172, FT7165, FT6229, T10384, FT8905, FT3772, FT5458, FT7041, FT1290, FT8465, FT7050, FT8897, FT7086, FT2615, FT7099, FT2809, FT2584, FT1840, FT8923, FT2997, T10403, FT1227, FT1653, T10396, FT3773, FT3192, FT7351, FT4674, FT3792, T10440, FT8931, FT2994, T10417, FT5411, T10369, FT3115, FT2070, FT5827, FT3833, FT5829, T10445, FT3830, FT3834, FT2898, FT5964, FT3954, FT8932, FT7713, FT6206, FT3105, FT1289, FT0622, FT8198, FT0354, FT8199, FT0692, FT8205, FT1474, FT8212, FT4054, FT8224, FT3319, FT8306, FT1803, FT4134, FT4041, FT8197, FT1423, T10604, FT8353, FT0158, FT1767, FT3958, FT3746, FT3314, FT0050, FT8741, FT3680, FT8275, FT0769, FT4394, FT3745, FT8732, FT0082, FT8748, FT1206, FT3747, FT0419, FT1990, FT3695, FT3813, FT3748, FT3215, FT3896, FT9080, FT2585, FT3919, FT0125, FT3752, FT9003, FT1215, FT3902, FT9055, FT2215, FT3956, FT1203, FT3753, FT1768, FT3706, FT1217, FT3756, FT9754, FT0623, FT1263, FT9745, FT0850, FT4114, FT9756, FT2265, T10382, FT3907, FT1506, FT4125, FT3681, FT1764, FT3910, FT2910, T10595, FT3911, FT1712, T10603, FT3762, FT0767, FT9743, FT0156, FT3964, FT9746, FT1448, T10294, FT3764, FT9747, FT3365, FT2446, FT3765, FT0086, FT3917, T10034, FT0092, FT4032, T10593, FT9017, T10594, FT5089, T10576, FT9022, T10609, FT5062, T10614, FT5169, FT4122, T10572, FT5092, FT7151, FT4028, T10082, T10083, T10098, T10719, FT9041, T10138, FT9046, FT4117, FT8048, T10707, FT3484, FT9044, T10134, T10779, T10182, FT5037, T10752, FT9999, T10197, T10651, T10193, T10762, FT8500, T10721, T10187, T10898, FT8884, FT0770, FT9209, FT9219, FT5857, FT8642, FT5974, FT3770, FT5490, FT5150, FT3775, FT3777, FT5048, T12154, FT9077, FT3778, FT8733, FT3793, FT6099, FT6110, FT9097, T12114, FT6074, FT9414, T10205, T10320, T10161, FT9128, T10026, FT9130, FT9133, T10311, FT9138, FT3562, T11003, FT5051, FT3650, T11004, FT5050, T11011, T12012, FT5116, FT9161, T12415, FT3019, FT3678, T10136, FT3683, T12734, FT4959, T12769, FT4949, T12656, T12776, T12792, FT9187, FT5473, T12831, T12833, FT5627, T12097, FT3526, T12834, FT5053, FT6145, T12838, T12841, T12843, FT4596, FT5117, FT5426, FT5349, FT4918, FT5497, FT5063, FT4162, FT5065, FT4418, FT9228, FT5632, FT5687, FT5622, T13168, FT5599, FT5606, FT9990, FT9235, FT3040, FT5066, FT4524, FT4675, FT9245, FT0100, FT0988, FT5435, FT5356, FT4164, FT5115, FT5745, FT1074, T12242, FT5004, FT0641, FT9276, FT9280, FT4417, FT6031, T12283, FT9277, FT4818, FT4948, T12306, FT5624, FT9281, FT5970, FT4883, FT5996, FT1725, T10343, FT4214, FT7052, FT5214, T10380, FT4420, FT4275, FT4769, FT5212, FT2182, FT9301, FT9304, FT9299, FT4940, FT4422, FT4211, FT4733, T12078, FT5240, FT2392, FT7743, FT4425, FT1157, FT5119, T12092, FT2422, FT1706, T12120, FT2467, FT5055, FT4656, FT5362, FT9284, T12030, FT9607, FT5768, T12093, FT5773, FT8742, FT9545, FT1003, T10852, FT1314, FT9549, FT1807, T12331, FT5775, T12252, FT5776, FT9542, FT0160, FT5777, FT9566, T10191, T12247, FT7269, FT5792, FT5801, FT7122, FT9579, FT5811, FT9580, FT9582, FT9588, FT5889, FT9592, FT5890, T10704, FT5981, T12361, FT7760, FT1428, T12394, FT9614, FT9676, FT7628, FT5946, FT9618, FT9620, FT9613, FT4020, T10183, FT6032, FT9617, FT7818, FT9621, FT7352, T12844, FT9624, FT2702, FT9636, FT7098, T11038, FT7057, T10019, FT9634, FT1442, FT7665, T10090, FT9635, T10152, FT1147, FT7143, FT8045, T10154, FT4900, FT3947, FT5790, FT0658, FT8129, FT1687, T10741, T10742, FT7934, T10738, FT4945, T10740, T10744, FT0112, FT7932, FT9668, FT1311, T10745, FT9675, FT5986, FT8100, FT9660, T10748, FT9991, FT5636, FT9685, FT1197, T10020, T10982, FT9679, FT5653, FT9692, FT5485, T10676, FT8103, T10271, FT8134, T11054, FT7936, T11050, FT9686, T10977, FT9696, FT7108, FT3867, FT9640, FT4341, T10406, FT5026, T12101, FT9691, FT9700, FT0399, T10400, T10408, FT2548, T12126, FT1204, FT1205, FT5941, FT9220, T12095, FT9332, FT4207, T10402, T10444, FT9330, FT9364, FT5022, FT5360, FT3901, FT9375, FT9341, FT5333, FT1260, FT3697, FT9374, FT1259, FT5494, FT3707, FT4103, T10428, FT5032, FT9322, FT3714, FT2586, T12180, FT1310, FT2587, T10451, FT3715, FT3716, FT5615, T10454, FT9393, FT5612, FT4121, FT9410, FT1410, FT4144, FT9394, T10393, FT9415, T10475, FT1411, T10437, T10631, FT5078, T12175, FT5219, FT5384, T10474, T12493, T10470, T10471, FT1262, T10482, FT9738, FT9744, FT5402, T10637, FT9670, FT9694, FT4157, T10849, FT9703, FT3920, FT4960, FT3922, FT1806, FT9695, FT5407, T12304, T12367, T12371, FT7470, FT4975, T12373, FT4974, FT4983, T12303, FT4984, T12374, T12287, T12302, FT8095, FT9765, FT0431, FT4386, T12422, FT7056, FT9761, FT0688, FT4391, FT9764, FT5144, FT4390, T10014, FT9767, FT7261, FT7235, FT9763, FT1564, T12496, FT4383, T12388, T10994, FT9789, FT5917, T12031, FT8273, FT9786, FT9791, T12098, FT9221, FT3723, FT9419, FT4472, FT4478, T12106, T12107, FT9799, FT4480, FT8301, FT3726, FT8460, T10770, FT8896, FT4483, T12112, FT9817, FT9814, FT9824, FT4175, T12145, FT9826, FT1151, FT4400, FT4184, FT5781, FT4176, FT4181, T12151, T12155, FT9416, FT4187, FT9871, FT1152, T10725, FT9882, T12087, FT4601, T12117, T12158, FT4605, T12802, FT5120, FT9199, FT5035, T10800, FT9891, FT9455, FT1563, FT1708, FT5436, T10827, FT9716, FT9895, FT5437, T10876, FT9883, FT1790, FT4486, FT9907, FT4497, FT3767, FT5041, FT9909, FT5043, FT5157, FT5163, FT9931, FT9906, FT1155, FT4993, FT9927, FT3785, T10956, FT9935, T10961, FT5254, FT5257, T10923, FT9941, FT9938, FT9945, T12301, FT9946, FT9957, T12424, FT8022, T12342, FT9971, T12344, T12425, FT4245, FT5295, FT5493, FT5296, FT2453, FT9979, FT5297, FT8284, FT4985, FT3990, FT9649, FT9666, FT4990, T10122, FT2749, T10158, FT3801, FT3803, FT4989, T10281, T12706, FT2824, FT4661, FT4140, FT4002, FT4004, FT4262, T12709, FT0128, FT4264, FT1580, FT4671, T12748, FT4005, T12712, FT3698, FT2492, FT2826, FT2980, T10044, FT4368, T12923, FT3059, T12929, FT3106, FT5023, FT3133, FT4930, T10047, FT4690, T10049, FT3989, FT3530, FT5021, FT5289, FT3265, T10045, FT3980, FT5315, T10046, FT3583, T12271, FT3628, FT4696, FT3332, FT3982, T12857, T12265, FT5766, FT4700, T12273, FT3251, FT3985, FT3998, FT4701, T12268, FT3531, T12266, FT4703, T12428, FT9762, FT4009, T12474, T10413, T10415, FT4296, FT3949, FT4299, FT3782, FT3952, FT4305, T12476, FT3881, FT4306, T10955, FT4014, FT3844, FT4022, T12480, FT3701, FT0612, FT5102, FT3916, T10073, FT5099, T12479, FT0819, FT4997, FT4023, FT4317, FT3627, FT4025, FT4321, T12481, FT4332, FT3993, FT4610, FT4334, T12613, T12747, FT5798, FT1518, T12922, FT4050, T12691, T12924, T10104, T10107, FT3938, FT3703, FT4840, T12612, FT3705, FT5679, FT2293, T12174, T12275, T12863, FT5655, FT4051, T10141, FT3936, FT4053, T12877, T10137, FT5499, FT4578, FT5303, T10375, T12879, T12906, FT3942, T12909, FT5159, FT5164, T10145, FT3976, T12986, T10146, T12988, T12983, T12751, T10151, FT5203, T13026, FT3843, FT5346, T13054, FT5414, FT5352, FT5420, FT5353, T13032, FT5355, T13033, FT3837, FT5357, FT3848, FT5358, T13035, T13059, T13057, T10172, FT5462, FT5708, FT3227, FT5755, T10180, T10181, FT3851, FT5765, T12959, T12980, FT5522, T12744, T12984, T12743, FT3180, FT5942, FT5523, FT6003, T10251, T12987, T10278, FT4099, T13087, FT4065, T13201, T13122, FT4067, FT2791, FT2834, FT4068, FT9863, T10302, T10305, T20110, T10215, FT2734, T12464, T12953, T10705, T12491, T20111, T12739, T12872, T10660, FT2778, T10708, T20114, T20289, T20141, T12405, T20143, T20202, T20213, T12391, T20215, T10962, T20233, FT5661, FT3232, T12982, FT7364, T12782, T12783, T10334, T12788, T20235, T13012, T20239, T10407, FT5509, T12780, FT7464, FT7637, FT6314, FT2412, FT7564, FT7990, FT9093, FT8082, T12926, T12960, FT5528, FT5529, FT9053, FT9061, FT9063, FT2249, FT7844, FT5540, T13058, FT5539, FT3425, FT5542, FT9485, FT9227, FT2605, FT9239, FT5567, FT5571, FT2000, FT8289, FT9325, FT8414, FT1092, FT8415, FT4623, FT1907, FT2074, T13016, FT9687, FT8539, T13023, FT1911, FT9849, FT8548, FT2021, FT9856, FT8586, FT2606, FT6071, FT8593, FT1910, T13107, FT2664, FT8860, T13024, T13096, FT2612, FT2011, FT2607, FT2475, FT1436, FT2635, FT0219, FT8859, FT8862, FT0407, FT4619, FT0594, FT2161, FT0736, FT0737, FT1441, FT0913, FT2608, FT4615, FT1677, FT4622, FT4627, FT2280, FT9825, FT2283, FT9822, FT2641, FT9615, FT9830, FT2164, FT9829, FT2911, FT2914, FT4634, FT2666, FT2951, FT2198, FT3658, FT3666, FT2595, FT2772, FT2064, FT9989, T10016, FT1501, FT2600, T10092, FT2783, FT2737, FT2076, T10640, FT1560, FT2732, FT2677, FT2669, FT1643, T10495, FT2735, FT1623, T10494, T10496, FT2909, FT0642, T13129, FT1999, FT1198, FT2077, T10601, T10659, FT0318, FT4636, T10633, T10688, FT3068, FT2193, FT4647, FT2092, FT2856, FT3297, FT1028, FT1073, FT3454, T13146, FT4707, T13133, FT1716, T12996, T10790, FT5287, FT5326, FT2634, FT5369, FT1981, FT5678, T13165, FT1210, FT7779, FT5856, FT3387, T12000, FT4706, T12144, T10318, T12545, FT5677, FT4704, T10209, T10452, FT2633, FT2733, T10632, T10638, T13197, T13194, FT3555, T13116, FT2601, T10712, T12594, T10666, T10713, T10731, FT5681, T12934, FT1072, T12635, T12966, FT2907, FT2687, FT4970, FT2688, T20113, FT1394, FT1402, FT2863, FT1392, T12104, FT2292, FT7873, FT1440, FT1553, FT3580, FT7870, FT4928, FT2097, FT4971, FT1439, T11027, FT1625, T11037, FT7865, FT1945, FT1556, FT1396, T12099, T13103, FT3422, FT1451, T11017, FT2219, FT2796, T13193, FT1868, T13180, FT4969, FT5047, FT2793, FT2619, FT3302, FT2981, T13021, FT1108, FT1645, FT1673, FT1107, FT0604, FT4124, FT1948, FT2620, FT1949, FT2073, FT2890, FT2101, FT2785, FT0372, FT1888, FT1617, FT1996, FT2884, FT1933, FT1993, FT2900, FT3088, FT2107, FT2792, FT8370, FT2147, FT1936, FT8186, FT1934, FT5019, FT8240, FT2102, FT2184, FT2136, T12918, FT8192, FT3087, FT2324, FT3015, FT1935, T13203, T13263, FT2699, T13267, FT2105, FT2627, FT2196, FT8369, FT3274, FT2103, FT2626, FT7843, FT2189, FT2625, FT2604, FT2259, FT2773, FT2134, FT8377, FT2690, FT2186, FT1834, T20251, FT2082, T12735, FT2099, FT1060, FT9109, FT2100, FT1061, T13304, FT8508, T13307, FT8702, T13293, FT8510, FT1937, FT8701, FT2438, FT3091, FT2396, FT2260, FT2323, FT7506, FT2151, FT1058, FT2094, FT7952, FT1963, FT2726, FT7948, FT1944, FT8381, FT3908, FT2701, FT8698, FT2091, FT2006, FT2046, FT8412, FT6152, FT6298, FT4771, FT3553, FT8805, FT1946, FT2090, FT7176, FT2110, FT8830, FT8035, FT8826, FT9054, FT8985, FT7634, FT4614, FT9049, FT4599, FT7344, FT4602, FT9064, T13154, FT2083, T13181, FT4854, FT4740, FT3424, FT7547, FT2225, T13259, FT8949, FT2662, FT2093, T13135, FT9048, FT7231, FT7285, FT8955, T10227, T10298, T10301, T10304, FT8652, FT9654, FT2248, FT9936, FT3554, T10815, FT8740, FT9326, FT9321, FT4830, FT4837, FT3038, FT4777, FT9312, FT2408, FT4796, FT4847, FT2081, FT2295, FT4851, FT9308, FT4670, FT3556, FT1463, FT9501, FT2454, FT2428, FT2576, FT2645, FT9521, FT2937, FT2929, FT4782, FT3005, FT4760, FT3152, FT2286, FT9606, FT3228, FT3242, FT3243, FT9735, FT4314, FT3459, FT4810, FT9851, FT3477, FT2301, FT3478, FT5010, FT3434, FT3479, T10086, FT3548, FT3549, FT3608, FT2294, FT2297, FT2227, FT4021, FT8639, FT8944, FT2228, FT4695, FT4530, FT9260, FT4825, FT9665, FT4798, FT9664, FT2360, FT6106, FT9663, FT4799, FT9671, FT3576, FT2361, FT2296, FT9659, FT1624, FT2906, FT4784, FT2904, FT4715, FT2302, FT2672, FT3089, T10958, FT2308, FT4795, FT9837, FT3498, FT1631, FT2067, FT9916, FT9912, FT2857, FT3267, FT5097, FT2577, FT4823, FT2480, FT4812, FT2080, FT7330, FT9947, FT4310, FT4829, FT4827, FT9940, FT5995, FT5106, FT4724, FT9943, FT4712, FT5128, FT2681, T10013, FT4709, FT5139, FT7079, FT1662, FT4728, FT1492, FT4723, FT4859, FT2592, FT2049, FT5001, T10913, FT5011, FT2236, T10952, FT1786, FT5787, FT2673, FT2936, FT1494, FT1496, FT5761, FT2596, FT2885, FT2339, FT7722, FT2115, FT7758, T12066, FT1498, FT2338, FT4471, FT3575, FT7966, T12337, FT1704, FT5772, T12338, FT5791, FT1785, FT7967, T12355, FT7969, FT1188, FT2671, T12185, FT1491, T12186, FT1666, FT8302, T12495, FT1701, FT8489, ft2027, FT5734, FT5803, FT1047, T12803, FT1048, FT5336, FT1049, FT1667, T12867, FT1679, FT8990, T12870, FT8992, FT4080, FT8995, FT9001, FT2140, FT5830, FT9007, FT9009, FT8966, T12730, FT9010, FT2444, FT9011, FT4113, FT1681, FT4170, FT5816, T12991, FT9092, FT3233, FT1860, T13104, FT2306, FT3578, FT9601, FT2522, T13153, FT2567, FT1493, FT1190, T13031, FT2524, FT5846, T20283, FT3419, FT2404, FT6088, T10163, FT1613, T10165, FT1515, T10167, FT1783, T10170, FT1512, FT6091, FT4558, FT1186, FT3420, FT2403, FT6104, FT6105, FT6111, FT6115, FT2410, FT2409, FT1224, FT5865, FT2035, FT5869, FB1576, FT2181, FT2238, FT2417, FT2653, FT2655, FT8982, FT2740, FT2751, FT2953, FT3535, FT9173, FT2436, FT2862, FT4024, FT3959, FT2540, FT3273, FT4574, FT5719, FT2087, FT4534, FT4536, FT4540, T10179, FT5875, FT5876, FT4563, FT5205, FT5206, FT5882, FT2303, FT5213, FT5884, FT5284, T10214, FT1833, FT2507, FT5913, FT5895, T10268, FT5887, FT2458, FT5910, FT2084, FT2901, FT6121, T10222, FT6122, FT4732, T10260, T10419, T10234, FT5951, FT2553, FT6001, T10218, FT6010, FT6013, FT3285, T10263, FT6046, FT6052, T10217, FT2443, FT6055, FT1301, T10258, FT3504, T12506, FT3471, FT0145, FT2442, T12507, FT6067, T12532, FT3507, FT2440, FT3508, T12540, T10261, T10229, T10253, T10255, T10264, T10262, FT9910, FT5929, T12460, T12461, T12227, T12231, FT5937, T12416, T12240, FT2557, T12288, T12295, FT5950, T12296, T12524, T12527, FT7221, T12435, T12439, FT1663, T10426, T10431, T10420, FT5968, FT7237, FT3514, T12753, T12754, T11013, FT5987, T12967, FT3523, T12955, T12972, T11015, T12948, T10906, T12994, T10259, T12551, T20166, T12566, T12445, T12449, T12466, T12470, T12472, T12761, FT6193, FT7876, FT6208, T20245, T13288, T13216, T12893, T13034, T13043, FT3500, FT5999, T13044, FT2813, T13037, FT7255, T13041, FT0146, FT7478, T12902, FT7483, FT2810, FT7012, FT7476, T13045, FT7487, T13050, FT0135, FT9170, FT9174, FT7021, FT9175, T13010, FT9176, T13049, T13051, FT9477, FT3455, FT9474, FT6042, FT9493, FT9495, FT9553, FT9555, T13052, FT9481, T13053, FT9547, T20097, T20100, FT0239, FT7490, FT9808, T13048, FT9806, FT6087, FT6063, T20155, FT7797, FT2794, FT6073, T20151, T20088, T20142, FT5948, T20103, FT6233, FT8856, T20117, FT2814, FT5953, T20121, FT7902, FT9159, FT9160, FT7443, FT3869, FT9167, FT2416, FT9467, FT0173, FT2727, FT9594, FT9591, FT7554, FT9590, FT7544, FT7555, FT6128, FT0174, FT6241, FT6136, FT6268, FT0162, FT6133, FT5980, FT6135, FT7585, FT6138, FT7590, FT7333, FT7125, FT7878, FT6141, FT6158, FT7596, FT7950, FT7951, FT7354, FT7616, FT7355, FT6178, FT6183, FT8116, FT8744, FT8818, FT6182, FT7627, FT6186, FT7398, FT7624, FT0194, FT0234, T10799, FT8524, FT8525, FT8535, FT7430, T10282, FT7082, FT8155, FT8185, FT7441, FT2987, FT5871, FT8563, FT7429, FT5872, FT8566, FT5873, FT0599, FT0278, FT7081, FT6017, FT7315, FT7326, FT9002, FT1297, FT6012, FT1413, FT2952, FT7083, FT1435, FT7413, FT6190, FT7343, FT6191, FT7373, FT7676, FT7399, FT6185, FT7467, FT7466, FT6192, FT7681, FT2494, FT2866, FT0201, FT2986, FT8656, FT6200, FT6044, FT6201, FT8767, FT2970, FT3044, FT8835, FT3023, FT2072, FT3055, FT3107, FT8832, FT7696, FT6240, FT8829, FT5034, FT7054, FT6058, FT6062, FT7055, FT7045, FT6234, FT6084, FT0143, FT6236, FT8877, FT7060, FT7092, FT0238, FT7105, FT7133, FT0240, FT7141, FT6116, FT7101, FT6118, FT6250, FT6256, FT6120, FT7170, FT0241, FT6253, T10370, FT6271, FT2840, FT2841, FT9407, FT7282, FT9548, FT7324, FT9241, FT3635, FT3637, FT3642, FT3644, FT9256, FT3649, FT8090, FT8170, FT9748, FT8299, FT9351, FT0344, FT9352, FT0245, FT6300, FT3043, FT6305, FT8588, FT8592, FT0343, FT8848, FT8971, FT8970, T10435, T10169, FT9098, FT2830, FT9074, FT3653, FT3654, FT3655, FT3660, FT3662, T10127, FT3687, T10132, FT3685, FT3690, FT3693, T10310, FT2998, FT2999, FT7335, FT0209, FT7377, T10484, T10488, FT9234, T10491, FT9201, FT9646, FT9269, FT0332, FT0324, T10520, FT9741, FT9667, FT9709, FT0014, T10508, FT1523, FT7131, FT9724, FT0191, T10339, FT7423, T10341, T10344, T10499, T10501, FT7080, FT7420, FT0030, FT0142, FT1599, FT7385, FT3000, FT2352, FT2843, FT2362, T10503, FT0019, FT0029, FT0021, FT0217, T10487, FT0024, FT0025, FT7600, FT1194, FT0026, FT0977, FT0090, T10513, FT0910, FT0448, FT7528, FT0186, T10841, FT0598, FT0911, FT5196, FT5197, FT0982, FT7508, FT1120, FT1863, FT2882, FT7541, FT1230, FT7542, FT1407, FT1408, FT2016, T10844, FT7515, FT1526, FT7606, FT1229, FT2334, FT2350, T10904, FT2336, FT1578, FT2876, FT7492, FT2353, FT2017, FT7378, FT2363, FT2351, FT7327, FT2374, FT2512, FT7493, FT2706, FT7546, FT2375, FT2752, FT2704, FT0303, FT4764, FT6299, T10918, FT2088, T10936, T10972, FT6295, FT6282, FT6312, FT3049, FT6306, FT5944, FT3050, T10922, FT5233, FT3181, FT2930, T10974, FT3017, T10985, FT3018, T10130, FT4753, T10990, FT7068, FT7065, FT3026, FT3027 - MIB Rev 9, FT7035, FT7095, T10350, T10354, FT7569, T10390, FT4898, FT3028 - MIB Rev 9, FT7604, FT2931, FT7560, FT2932, T10588, T10358, FT5237, FT3754, FT5171, FT5229, FT7499, FT4120, FT0360, FT4185, FT0357, T10533, FT0582, FT3156, FT7568, FT7567, FT8041, T10546, FT8026, FT0111, FT3045, T10555, FT0107, FT0361, T10559, FT8086, FT4595, FT0363, FT5332, FT3046, FT5395, FT3155, FT0425, FT5500, FT8034, FT5715, FT1452, FT8099, FT0611, FT0371, T10551, FT0681, FT5733, FT3158, FT0581, FT0926, FT3126, FT1462, T10528, FT7109, FT0964, FT7115, FT3606, FT1243, T10596, FT3137, T10584, FT7134, T10552, FT3430, FT1721, T10635, FT2705, FT7662, FT1921, T10612, FT4129, FT3322, FT2043, FT7063, FT7135, FT3835, T10583, FT2079, FT2543, T10558, FT2541, FT6029, T12089, FT7198, T10587, FT0564, FT2651, T10582, T10591, FT6275, FT6270, FT7249, T10656, T10670, FT0811, FT7152, FT7154, FT7692, FT0379, FT7711, FT0510, T10703, FT7683, FT7328, FT3937, T10602, FT7805, FT0401, T10621, FT7835, FT7320, FT5764, FT7322, FT4861, T10598, FT9780, FT7742, FT7157, FT7740, FT7313, FT7829, FT7291, FT4351, T10690, FT7754, FT7193, FT4349, T10347, FT7734, FT5546, FT2201, FT2202, FT7735, FT1927, FT7769, FT0427, FT0669, FT0566, FT7342, FT7260, FT1465, FT0989, FT0492, T10657, T10652, FT7458, FT3082, T10653, FT4552, T10655, FT0260, FT4515, FT7345 No8, FT7456, FT4547, FT4544, FT0644, FT0600, FT7612, FT4577, FT4581, FT3066, T10726, FT5556, T10730, FT5550, FT5578, FT5558, FT5025, FT5553, FT5547, FT2845, FT4652, FT5557, FT3201, FT5549, FT5548, FT7649, FT4714, T10733, FT5561, FT5545, FT4894, FT7641, FT3981, FT7647, T10727, FT0309, FT7721, FT0283, FT0133, FT7501, FT7719, FT2271, FT2037, FT0486, FT7806, FT7642, FT7765, FT7589, FT0487, T10747, FT7810, FT3860, FT4554, FT0488, FT4916, FT7803, FT4943, T10761, FT5180, FT0552, FT5183, FT5188, FT3097, FT5255, T10757, T10774, FT3099, FT5288, FT7977, T10739, FT5375, FT3079, FT5491, FT0484, FT5555, T10736, FT7780, FT7800, T10824, FT7804, FT7976, FT5707, FT5754, FT8043, FT5747, FT5753, FT0589, FT0525, FT6057, T10778, FT6035, FT0621, FT7975, FT7921, FT7914, FT0485, FT6109, FT0527, FT8031, FT6112, FT8265, FT0624, T13002, FT8270, FT7912, FT8267, FT7986, FT8269, T12434, FT6221, FT8023, FT6070, T10807, FT7984, FT5940, FT8280, T10737, FT3114, T10820, FT8271, T10771, FT8229, FT8171, FT8283, FT7751, T10788, T10805, FT8228, FT4687, FT3172, T10590, T10828, FT8276, FT3122, FT9211, FT9384, T10838, FT9272, T10823, T10840, FT9389, T10813, FT0543, FT7988, T10848, FT0469, FT0546, FT7985, T10853, T10830, FT9391, FT9169, FT5413, FT9215, FT9255, FT0408, T12129, FT8054, T12332, T12329, FT0529, FT8152, T10874, FT8154, T10877, FT8149, FT8158, FT8230, FT8163, T12887, FT0547, FT0715, T12351, FT0548, T12392, FT8091, FT8298, T12558, FT7937, FT8128, FT8150, FT8226, FT3203, FT8307, FT8157, FT8292, FT8809, FT5371, FT7770, FT8319, T11035, T12549, FT7514, FT8361, FT8371, T12096, FT0462, T12548, FT8262, FT0502, FT8408, T12555, FT8406, FT0475, T12796, FT8407, T12797, T20071, FT9066, T12799, FT9165, FT0474, T12552, FT9303, FT0634, FT9295, FT8495, FT9371, T20126, T12563, FT0478, FT8428, T12572, FT8403, FT4439, T20135, T20138, FT4282, T12577, FT8413, FT8258, T10542, FT0813, FT8247, T13292, FT8161, FT3223, T10540, FT8260, FT8542, FT3224, T20312, T20324, T13063, T12631, FT0481, T13160, FT7448, FT8261, T12638, FT7438, FT7400, FT6093, T10539, T10565, FT0962, FT0445, FT8546, T10379, FT8547, FT8575, FT7426, FT0635, FT7459, FT8348, FT8603, FT8350, FT8604, FT7491, FT4389, FT3154, FT8638, FT7553, FT7576, FT8585, FT8259, FT0330, FT8349, FT8249, FT6309, FT7498, T13167, FT8589, FT8641, FT0645, T12116, FT8728, FT7030, FT0647, FT8250, FT8351, FT0646, FT8124, FT2835, FT8126, FT8336, FT8648, FT5535, FT6301, FT3321, FT8679, FT5354, FT8853, FT0411, T12305, T12323, FT8422, FT7577, FT2313, FT4276, FT8544, FT7586, FT5732, FT7323, FT5738, FT3824, FT8684, FT7595, FT8867, T12375, FT5779, FT7583, FT7643, T12386, FT8876, FT7762, FT8872, FT3377, T12413, FT4470, FT8852, FT3235, FT3254, FT5837, FT5838, FT8917, FT4986, T12421, FT5265, FT8898, FT5008, FT0809, FT5818, FT8214, FT5052, FT8092, FT3351, FT8093, FT8175, FT3258, FT8902, FT3259, FT5343, FT0821, FT3256, FT9357, FT0812, FT8954, FT8957, FT5460, FT0820, FT8640, FT3528, FT8977, FT3558, FT8964, T12471, FT8987, FT8461, FT0829, FT8463, FT7626, FT5630, FT8988, FT8446, FT8466, FT0226, FT8634, T12483, T12484, FT9013, FT7638, FT0540, FT8464, FT8637, FT8579, FT9035, FT8571, FT8573, FT7661, FT3402, FT9040, FT9042, FT7664, FT8644, FT7655, FT7663, T12509, FT9045, FT2589, FT7453, FT0698, T12508, FT3395, FT4978, FT8397, FT9108, FT9058, FT0690, FT2172, FT3404, FT3323, FT2922, T12523, FT2919, FT2759, T12525, FT0697, FT0853, FT9072, FT0832, FT9073, FT7685, FT3084, FT8658, FT7617, T12559, FT9101, FT9102, FT9103, FT8455, FT7699, FT8625, FT9137, FT7700, FT9142, FT3414, FT3450, FT0836, T12393, FT2111, FT9452, T12402, FT0840, T12407, FT0839, FT8533, T12530, FT9599, T10204, T12537, T12539, FT7214, T12544, FT9641, FT9656, FT0864, FT9143, FT7726, FT9149, FT7731, FT7228, FT9150, FT3439, FT9134, FT9140, FT7723, FT0860, FT7724, FT7252, FT8576, FT0861, FT7253, FT9785, FT8620, FT9350, FT9788, FT6307, FT8568, FT7317, FT9475, FT7319, FT9418, FT8613, FT9402, FT8623, FT9359, FT3456, FT4222, FT7325, FT9811, FT6117, FT9812, T11014, FT7348, FT4240, FT9862, FT9859, FT7374, FT7379, FT9353, FT9873, FT4232, FT7396, FT8624, FT7768, FT7397, FT9349, FT7418, FT9327, FT6155, FT9368, FT6163, FT9378, FT6167, FT4267, T11022, FT7454, FT8627, FT8662, FT9878, FT8710, FT4230, FT4237, FT8714, T11029, FT9412, FT8708, FT9803, FT8772, T11032, FT8574, FT6262, FT9874, FT8786, FT6156, FT6274, FT8784, FT6320, T11034, FT6220, FT8785, FT6216, T10135, FT9432, T11044, T10024, FT7005, FT4287, FT4290, T11052, T10143, FT1421, FT4293, FT7457, FT4294, FT7477, FT2477, FT4295, FT8838, FT8843, FT4300, FT7087, FT4302, T11049, FT4303, FT7014, FT8777, FT7455, FT8778, FT7015, T10277, FT0539, FT7006, FT7888, FT2780, FT7011, FT9358, FT7026, FT8831, FT8836, T12022, FT7016, FT7023, FT4255, FT8195, FT9409, FT9422, FT8845, FT4277, FT8309, T12014, T12025, FT8868, FT9208, FT7053, FT8244, FT8864, FT8378, FT9210, T12015, FT8810, FT8300, FT8802, FT9237, T12034, FT8474, FT9212, FT7040, T12027, FT9229, FT8596, FT0441, FT4344, T12036, FT4347, FT8390, FT9267, FT9107, FT0113, FT4366, FT8776, FT4362, T12049, FT8560, FT0630, FT4365, FT0573, FT8924, FT4367, FT0954, FT4381, FT8920, FT9470, FT0927, T12053, FT8922, FT8929, FT1406, FT4375, FT8915, T12052, FT1458, FT4370, T12068, FT8834, FT4380, FT4382, FT1608, FT9644, FT4312, FT7127, FT4313, T12070, T10272, FT1461, FT4319, T10275, T10009, FT8938, FT9367, FT0056, FT7142, FT8939, T12077, FT9311, T12081, FT9376, FT7853, FT9380, FT7852, FT9524, FT8769, T12648, FT9526, T12651, FT2432, FT4407, FT4416, FT9420, FT4410, FT4507, FT9473, FT9177, FT8975, FT9779, FT3250, FT9180, T12580, FT9182, T12582, FT3294, FT8967, FT3020, FT4586, FT3022, FT3407, FT8959, FT8986, FT9673, FT8961, FT9206, FT3873, FT8976, FT4458, T12678, T12683, FT8974, FT9804, FT9805, FT9783, FT4495, FT4427, T12665, T12749, T12675, T12700, T12759, FT4927, T12562, T12673, FT7889, T12677, T12679, T12680, T12689, T12693, T12694, FT9794, FT5192, FT7897, FT3561, FT4518, T12798, T10059, FT7904, FT4520, T12806, FT5688, FT9070, T12807, FT5703, T12570, T12809, FT5737, T10808, T12701, FT9289, FT9950, FT9952, FT9955, FT8266, T12813, T12816, FT9379, T12817, FT9294, T12818, FT9328, T12823, FT4553, T12825, T12830, T12836, T10022, T10012, T12848, T10015, FT6288, T12860, T12837, T12847, FT9995, FT9996, FT4496, T12849, FT9331, T10035, T12868, FT9336, FT0795, FT4550, T12992, T20033, FT4542, T20038, T20041, FT9387, FT8176, T20025, FT0794, FT9282, T12998, FT9342, FT4444, T12881, T12884, T13004, FT8241, FT4447, T12888, FT8243, FT4450, FT8248, T10079, T12999, T12871, T20036, FT4549, T12875, T20043, FT8242, FT4548, T12876, FT8256, T12883, T12889, FT8340, FT5601, FT4556, T20057, T10252, FT4562, T12890, FT9340, T12894, FT4566, FT8304, FT9496, T12892, T12895, FT9442, T12898, FT9397, T10140, FT9427, T10280, FT9443, T10206, T10196, T12930, T20050, FT5057, T20054, T12928, FT7137, FT7139, T12933, T20045, FT7140, T20049, T20053, T20056, T12882, FT9083, FT9076, FT5060, T12937, T20059, FT3094, T20060, FT3096, FT5165, T20062, FT9225, FT5172, FT1830, FT9121, FT5154, FT9125, T12945, FT9127, T20065, FT9226, FT2776, T20066, FT9124, T20068, T20070, T20018, FT5095, FT5136, T20008, FT9132, FT5083, T20013, T20023, T20083, FT9123, FT2983, FT1503, FT9120, FT9114, FT5177, FT3432, T20075, T20092, FT9192, T20079, FT5110, T20080, T20007, T20082, T20026, T20084, T20086, FT2775, FT5100, T20089, FT5300, FT5103, T20019, FT0804, FT5208, T20028, T20085, FT2603, T20020, FT9194, T12714, T20091, T20105, FT5221, T20077, FT9454, FT5368, FT5235, FT5275, FT5227, T20101, FT9444, T20115, T20125, T12710, T20123, FT9445, FT7673, T20124, T20129, T20130, T20139, FT9437, FT7709, T20131, T20150, FT9447, FT7816, FT9448, FT9400, FT9399, T20145, FT9512, T20149, T20153, FT3230, T20163, FT9505, T20168, T20190, FT2597, T20183, FT7784, T20179, FT9502, T20192, T20195, T20199, FT9503, FT9567, T20200, T20158, T20225, T20219, FT3278, T20180, FT9506, FT9509, T20216, FT9513, T20127, T20223, T20234, FT9515, T20226, T20242, T20247, T20250, T13273, T20252, T13271, FT9569, FT9435, FT7885, FT9712, T20265, T20291, T20270, T20286, T20295, FT9725, FT4790, T20258, FT9772, T13311, FT9776, T20276, FT7629, FT7688, FT9510, FT7748, FT7750, FT1889, FT7830, FT7529, FT9855, FT7720, FT7707, FT7733, FT9774, FT9845, FT9848, T13300, T10054, FT9844, FT9896, FT3686, FT9852, FT5554, FT5551, FT5559, FT5552, FT5574, FT5570, FT5543, FT5525, FT5533, FT4058, T20316, FT9905, FT7811, FT7729, FT7736, FT7808, FT7745, FT9903, FT7749, T20311, FT7807, FT7812, FT7706, T20320, FT7737, FT7738, FT7813, T20329, T10058, FT9908, FT9963, FT9970, FT9961, FT9755, FT9967, FT3790, FT7814, FT7802, FT9973, FT7915, FT1832, FT9969, FT7918, FT8998, FT9974, T10053, FT7987, T10050, FT9964, FT9992, FT9988, FT4010, FT7815, FT7809, FT9968, FT0978, FT7922, FT9914, FT9918, FT7920, FT9307, FT7916, FT9972, FT7983, T10063, FT9584, FT9091, FT8766, FT2579, FT9094, TOP0010, FT2578, FT9000, FT2290, FT7939, FT1750, T10066, FT3423, T10055, FT2761, FT0907, FT9921, T10051, FT8164, T10064, FT9183, FT9196, T10100, FT2026, FT9271, FT9302, FT9300, T10106, FT2145, T10239, FT9188, T10123, FT8245, FT7931, FT8160, T10312, T10124, T10317, T10313, FT3159, FT8257, FT3136, T10314, FT8411, FT3476, T10819, T10887, FT8341, T10884, T10881, FT8362, T10944, FT8360, FT8354, T10472, FT8255, FT7536, T10242, T10244, FT8342, FT8329, FT8343, T10119, FT8405, T10327, FT8332, T10117, T10235, FT9986, T10368, FT8339, T10722, T12384, T10366, T10490, FT8399, T10367, T12732, T10120, T10247, FT8346, T10362, FT8324, T10300, FT8333, T10404, FT8310, FT8357, FT8515, T10296, T10303, FT8823, T12430, T10457, T10461, T10459, T10460, FT9760, FT8471, T10492, T10530, FT2784, T10627, FT3161, T10315, T10626, FT8655, FT5045, FT5486, T10456, FT5489, FT5501, T10465, FT5908, T10610, FT4485, T10932, T10937, T10553, T10920, FT8572, T12270, FT8467, T10714, FT8458, T10514, FT8608, T10526, T12261, FT8628, T10606, T10523, T10608, FT8610, T10605, FT8606, FT8622, T12084, T12269, T10522, T13042, FT8724, FT7300, FT8151, FT7286, FT0745, FT8771, FT7421, FT8774, FT8779, FT8782, FT7314, FT7465, FT7505, FT8713, FT7510, FT8717, FT7525, T12593, FT9436, FT7533, FT7534, FT8850, FT8803, FT7548, FT5090, T12267, FT7584, FT7580, T12258, FT7539, FT7538, FT8869, FT7494, FT8783, FT6293, FT6264, FT8787, FT6265, FT8921, FT6303, FT6308, FT8916, FT8926, FT8928, FT6242, FT8918, FT8925, FT6153, T12485, FT8911, FT8909, FT7038, FT7039, FT6292, FT8914, FT7032, FT6245, FT8781, FT7009, T13000, FT7096, FT9373, FT7559, FT8983, FT7562, FT7563, T13147, T13151, FT7611, FT9469, FT9106, FT9451, FT7502, FT8969, FT8968, FT9572, FT7431, FT8984, FT8981, FT8098, FT8074, FT8097, FT8936, FT8935, FT8973, FT9111, FT8094, FT8096, FT7064, T10068, FT7061, FT9622, FT7160, FT7174, FT9288, FT9297, FT9291, FT5122, FT7144, FT6238, T10224, T10216, T10220, FT7194, FT7182, FT4340, FT7196, FT9293, FT7120, FT9388, FT9329, FT7171, FT7230, FT7247, FT4443, FT7266, FT7234, FT7274, FT7257, FT3473, FT9338, FT7267, FT7290, FT9333, FT7293, FT9339, FT9335, FT6225, FT3354, FT7292, FT7264, FT4977, FT9334, FT3356, FT9395, FT9498, FT7517, FT9398, FT4273, FT9396, FT3355, FT9441, T10541, FT7556, FT0571, FT9112, T12085, FT9116, FT2707, FT0570, FT1983, T10846, FT7609, FT7648, FT9078, FT9081, FT1986, FT2837, FT7640, FT7645, FT7646, FT7799, T12835, FT7856, FT9223, FT4438, FT0498, FT5442, FT5242, FT8122, FT9086, FT9129, FT5667, FT8131, FT8146, FT5238, FT9218, FT5657, FT5568, FT9122, FT5565, FT9126, FT8046, FT8037, FT8084, FT8177, FT9190, FT8173, FT8174, FT8279, FT8282, FT9230, FT8285, FT4266, FT9200, FT8775, FT5730, FT8234, FT9197, FT5146, FT0831, FT9185, FT8042, FT9203, FT8227, FT2254, FT8278, FT5186, FT8277, FT9577, FT5804, FT4901, FT2419, FT5210, FT9440, FT9217, T10392, FT5826, FT9259, FT9354, FT5247, FT9355, FT9439, FT5839, FT5841, FT3738, FT9446, FT5049, FT3749, FT8235, FT5248, FT5270, FT4755, FT9438, T12502, FT9516, FT8308, FT8311, FT8290, T12278, FT8454, FT9576, FT8365, FT9504, FT0533, FT9497, FT8367, FT8404, FT1272, T12636, FT0535, FT8426, FT8236, FT8238, FT1698, FT1799, FT8451, FT0175, FT8431, FT0177, T12861, FT9575, FT8401, FT1163, FT8504, FT8506, FT2714, FT8540, FT8493, FT5692, FT2806, FT9571, FT3342, FT8581, T12858, FT9568, FT8612, FT8632, FT5972, FT8645, FT7205, FT8647, FT9508, FT3275, FT7241, FT8590, FT9551, FT8676, FT8651, FT7265, FT9511, FT7000, FT9574, FT8591, FT9578, FT8595, FT9714, FT8633, FT7321, T10497, FT9706, FT9711, FT9722, T10856, FT9570, FT9573, FT7340, FT7022, FT8729, FT7018, FT9713, FT9720, FT8649, FT8490, FT7388, FT7712, FT8491, FT7393, FT6159, FT8602, FT9777, FT7513, FT3537, FT4448, FT9552, FT8737, FT8600, FT8601, FT3538, FT9770, FT8861, FT9854, FT8858, FT7126, FT7128, FT9838, FT9847, FT8881, FT6286, FT9499, FT9773, FT8912, FT8919, FT5573, FT7488, FT9775, FT7481, FT8725, FT9846, FT7497, FT8894, FT9850, FT9853, FT7482, FT7370, FT7674, FT7861, FT7225, FT7864, FT8854, T10057, FT8543, FT7444, FT9842, FT7882, FT9843, FT7884, FT8952, FT8956, FT3486, FT8962, FT7887, T10060, FT8855, FT7895, FT8059, FT8063, FT9904, FT8978, FT8143, FT8147, FT8153, FT8159, T10004, FT9018, FT9184, FT9019, FT9020, FT8162, T10002, FT8165, FT8189, FT9015, FT8210, FT9898, FT9047, FT9095, FT9901, FT8565, FT8788, FT9900, FT8798, FT9110, FT9060, FT9057, FT9056, FT9100, FT8927, FT8933, FT8934, FT9962, FT9966, FT9105, FT9965, T10061, FT9136, T10056, FT9982, FT9993, FT9519, FT9525, FT9959, FT9153, T10230, FT8999, FT9154, T10309, FT9179, FT7881, FT9181, FT9144, FT9917, FT9139, FT9960, T10048, FT9707, T10062, FT9629, FT9392, T10052, FT9717, FT9117, FT9361, FT9771, FT9926, FT0010, FT9823, FT9543, FT9382, FT9886, FT7702, FT9977, FT9981, FT9924, T10238, T10023, FT9923, T10065, FT0057, FT4585, FT9919, FT0422, FT2451, FT2698, FT2855, T10399, FT3177, FT3178, FT3327, FT0137, FT3330, FT9430, FT3349, FT4027, T10941, FT4278, FT4280, FT4492, FT4493, FT4494, FT5007, T10111, FT7409, T10115, FT7408, T10121, T10549, FT9424, T10531, T10818, T10795, T10883, T10885, T12554, FT9434, T12561, T10947, FT9429, FT9155, T10125, T12443, FT0925, FT8498, FT0340, FT0924, T10447, FT1517, T10332, FT9205, T12141, T10323, T12165, T10243, T12172, T12171, FT9213, FT9224, T12148, T12291, FT9263, T12311, T10325, FT9264, T12298, T10297, FT9265, T10335, FT9202, T10299, T12123, T10116, FT9266, FT2799, T10246, T12006, T11023, T12026, T10360, T12033, T10306, T10363, T10233, T10245, T10348, T12682, T12820, FT4192, T10240, FT9612, T10295, T10211, T12688, T10361, T12684, T10365, T12779, FT9643, T10329, FT9472, T12968, FT9198, T10458, T10455, FT9610, T10464, FT9790, FT2915, FT3013, FT9781, T13281, FT9782, FT9784, FT3147, T10462, FT5126, FT5142, T10624, FT5145, T10527, FT4401, FT9736, T10556, FT9800, FT4756, FT1482, FT4087, T10607, FT9857, FT9831, FT1746, T10717, FT9833, FT3146, T10554, FT9797, FT6255, T10521, T10525, T10524, FT3190, FT3392, FT9462, FT0616, FT9956, FT6170, FT2815, T10270, FT8693, T10578, T10570, FT3217, T10032, PM4423, T10031, T10567, T10579, T12731, T10569, FT7112, T10564, FT8536, T10751, T10634, FT8738, T10822, FT9147, T12454, FT5751, T10157, T10248, T12320, T10091, T10094, FT4660, T10087, FT9189, FT9195, T12356, FT7524, FT7503, FT7173, FT9565, FT8356, T12315, T12358, T10105, T10078, T10984, T11048, T12203, T12208, T10150, T10636, FT5191, T12207, FT2366, FT1308, T12214, T10236, T12217, FT7338, T12218, T12235, T12221, FT4942, T20236, FT4955, T12236, T10871, FT5626, T12220, FT5628, T10889, FT5881, T10806, FT3496, T10924, FT6176, FT3596, FT6278, FT6279, T12255, FT8483, T12250, T10171, T10178, T10184, T12237, T12244, T12253, T12245, T12297, T12300, FT7163, FT7608, T12248, T10466, T12453, FT3109, T12457, FT2946, T12088, T11016, FT2966, T12166, FT2661, T13011, T10664, T13008, T10689, T10692, T13174, T10693, T10697, T13175, FT0651, FT2257, T10687, FT0854, FT2317, T10706, FT0890, FT2256, FT1550, FT1551, T10682, FT1565, FT1926, T10942, FT2321, FT2380, T10699, FT2486, FT2485, T10754, FT2332, FT1792, T10837, FT2913, FT0045, FT0044, T10869, FT4664, FT4966, FT3014, FT0781, FT0782, T10868, FT0074, FT2741, FT3249, FT2865, FT0704, T10929, FT0189, FT5281, FT2941, FT3301, T10866, FT3011, FT0127, FT3012, FT1115, FT0119, FT3304, FT0130, FT3305, FT1011, FT5290, FT3307, FT0281, FT1024, T12045, T10718, FT3828, FT0109, FT3734, FT0258, FT4138, FT0285, FT0375, FT4142, FT0306, FT4217, FT4202, FT0170, FT3518, FT3850, FT3519, FT0342, T12043, FT4869, FT5234, FT5231, FT4741, FT5249, FT5415, FT0418, FT0435, FT0609, FT5405, FT5383, FT3325, FT0461, FT5614, T10965, FT5610, FT0742, FT0096, FT2241, FT2055, FT2781, FT0348, FT1334, FT0235, FT3196, T10861, FT0605, FT0608, FT3595, FT0524, FT0520, T12211, FT0638, FT3599, FT0639, FT3814, FT0700, FT0251, FT0443, FT3823, FT0563, FT3825, FT0672, FT0941, FT0725, FT1332, T11025, FT3003, FT1573, FT0746, FT1574, T12071, FT1480, T12073, FT0783, T12196, FT2318, FT1029, FT2525, FT0990, FT2121, FT4411, FT0995, FT2958, FT3195, FT0952, T10996, FT0807, FT3303, FT4748, FT4212, FT3309, FT0063, FT4219, FT0252, FT0955, FT0320, FT0391, FT0741, FT3565, T12513, FT0914, FT0607, FT0751, T12256, FT0895, FT0936, T12321, FT4720, FT1012, T12322, FT0857, FT1077, FT1240, FT1105, T12377, FT1249, FT1116, FT1119, T12346, FT0994, FT1459, FT0993, FT1568, FT1572, T12382, FT1051, FT1793, T12500, FT1302, FT1796, FT1333, FT1335, T12499, FT1571, T12503, FT1178, FT1808, T12510, T10376, FT5634, T12517, FT5635, T12518, FT1813, T12520, FT5637, T12522, FT1811, FT1211, FT0169, FT1924, FT1257, FT1950, FT1923, FT1273, FT1239, T12318, FT1953, FT2057, FT1955, FT2058, FT1305, FT2060, FT1304, FT2054, FT3001, FT2912, FT2183, FT1276, FT2330, FT3117, FT1275, FT2382, FT2388, FT2386, T12586, FT1317, FT2528, FT2526, T10266, FT1357, FT1274, FT1695, FT1605, FT1606, FT2573, T10864, FT1715, FT1697, FT1720, T12599, FT1753, FT1776, FT1773, FT2756, FT1757, FT1758, FT3453, FT1848, FT1918, FT1100, FT1856, FT1857, FT1858, T12639, T12642, FT1847, T12662, FT6024, T12686, FT1849, T12591, T12595, FT2005, T12603, FT0108, T12605, FT1992, FT2007, FT6000, FT6007, T12627, FT2154, T12619, T12625, T12634, T12021, FT2200, FT2531, FT2493, FT2274, FT7268, FT2310, FT2315, FT7301, FT2343, FT2536, T12643, FT2273, FT2431, FT7412, FT2277, FT2423, FT2424, FT7416, FT2425, T12720, FT2533, T12742, FT2532, FT2489, FT7507, T13162, FT2470, FT7565, FT7570, T12990, T12997, FT2495, T13097, T13101, FT2575, FT2572, FT2696, FT2521, FT7295, FT2056, FT2719, FT7367, FT0117, FT7375, FT7474, FT2804, FT2695, FT2743, FT6043, FT6045, FT5963, FT6069, FT7578, FT6089, FT0183, FT7650, FT6119, FT0184, FT6137, FT0727, FT7651, FT2827, FT6139, FT7653, FT2808, FT9305, FT6142, FT2817, FT7656, FT7657, FT2844, FT8169, FT8207, FT2779, FT2874, FT8327, FT8219, FT7186, FT7411, FT2920, FT8322, FT7632, FT7415, FT8323, FT7297, FT8325, FT7434, FT7527, FT2950, FT7694, FT7356, FT8760, FT7752, FT7341, FT2967, FT2957, FT0726, FT8807, FT7926, FT0280, FT8811, FT8065, FT3035, FT8899, FT2768, FT8901, FT3034, FT8910, FT7090, FT7027, FT3252, FT7532, FT0163, FT7033, FT3236, FT7070, FT7652, FT3248, FT9118, FT7084, FT7728, FT3132, FT7031, FT9296, FT7116, FT3328, FT7119, FT3391, FT3394, FT9682, FT7150, FT3003/D, FT9690, FT9697, FT3237, FT3269, FT9689, FT3150, FT7168, FT3245, FT3238, FT2643, FT2502, FT7185, FT3293, FT0915, FT3292, FT9752, FT3287, FT7190, FT3077, FT7486, FT3343, FT7571, FT7573, FT9887, FT4431, FT7668, FT3346, FT7757, FT0397, FT3344, FT7761, FT9867, FT4537, FT2117, FT2562, FT3413, FT2569, FT3364, T10783, FT2570, FT7774, T10842, FT7821, FT3487, T10925, FT7826, FT3390, FT9899, FT7786, T10000, FT3408, FT7772, FT3411, T10110, FT3436, FT7775, FT7790, FT7796, FT7795, FT1675, FT3437, FT 3449, FT7798, FT3463, T10649, FT3452, FT7834, T10654, T12317, FT8072, T12309, FT3490, T12310, FT3485, FT8076, FT3489, FT4598, FT8106, FT3488, FT8133, FT8135, T12587, FT8140, FT3545, T12583, FT8137, FT3569, FT8937, FT3571, FT9031, T12604, FT3572, FT9151, FT8110, FT3579, FT3587, FT8141, FT3588, FT8145, FT5337, FT5340, FT3604, T12038, FT8107, FT1778, T12204, FT8200, FT1727, FT8211, FT5347, FT3615, FT8220, FT5350, FT3570, FT2777, FT3611, T12280, FT5604, FT3610, T12284, FT8109, FT8231, FT5656, FT3617, FT8251, FT3620, FT8254, FT3884, T12293, FT5704, FT8396, FT3877, FT8193, T11055, FT5888, FT3933, FT8305, FT5936, FT3840, T10911, T10981, T10987, T11033, FT8557, FT3894, T11042, T11046, FT8364, FT8372, FT3987, FT0053, T12005, FT8457, T12024, T12002, T12977, FT3891, T12608, FT8513, FT5305, FT4031, FT4047, T12610, FT8379, FT5133, T12826, FT4115, FT4128, FT8518, T12672, FT8516, FT4261, FT8520, FT8521, T12767, T12722, FT4260, FT2721, T12770, FT8523, FT4265, FT8715, FT4271, T12962, FT7299, T12965, FT7363, FT8716, T12961, FT8719, FT4143, FT4030, FT7372, FT4292, T13131, FT4152, FT8727, T13172, FT7392, FT7394, FT7484, FT4297, FT7485, FT4304, T13183, FT4322, FT7771, FT8696, FT7877, FT8726, FT8745, FT4426, FT8138, FT8753, FT7479, FT4345, FT4616, FT7460, FT4403, FT8686, FT4355, FT0381, FT5947, FT0417, FT4468, FT4404, FT7279, FT7368, FT4501, FT4487, FT8723, FT4395, FT4469, FT4467, FT8770, FT7383, FT4388, FT8822, FT4449, FT4499, FT8946, FT7537, FT4488, FT7618, FT8950, FT8996, FT8997, FT9005, FT4833, FT6157, FT4662, FT6164, FT6165, FT4639, FT6166, FT4604, FT4402, FT7500, FT4432, FT4609, FT8945, FT2693, FT9065, FT6239, FT2221, FT2229, FT4679, FT2205, FT2371, FT4669, FT4676, FT4904, FT9162, FT3625, FT2370, FT9164, FT3634, FT7046, FT3626, FT4738, FT5809, FT4775, FT5814, FT9517, FT7106, FT4473, FT9236, FT9232, FT4834, FT3636, FT4711, FT4964, FT9240, FT4686, FT4509, FT9238, FT3665, FT4717, FT3682, FT3460, FT4814, FT9246, FT7705, FT4839, FT4681, FT4870, FT8947, FT4871, FT4826, FT5461, FT4872, FT4084, FT5390, FT5391, FT9024, FT5262, FT5412, FT4791, FT4876, FT4797, FT5832, FT4865, FT4874, FT3643, FT5833, FT4780, FT5867, FT4750, FT5710, FT0601, FT5721, FT5965, FT5726, FT4858, FT5967, FT6187, FT9278, FT6154, FT9310, FT9316, FT9491, FT1652, FT9487, FT3552, FT9527, FT1732, FT4766, FT4817, FT9529, FT4862, FT4866, FT0777, FT4881, FT4882, FT4912, FT4922, FT4933, FT4941, FT4994, FT5002, FT5070, FT3193, FT1324, FT2369, FT5094, FT2692, FT9531, FT9532, FT1522, FT1577, FT2372, FT2995, FT9520, FT0557, FT1133, FT5088, FT9528, FT5202, FT3393, FT0423, FT5199, FT5174, FT9530, FT9537, FT2801, FT9583, FT5166, FT5168, FT9433, FT9625, FT5256, FT6033, T10103, FT5182, FT8452, FT5147, FT5140, FT9626, FT8202, FT8203, FT8204, FT8213, FT9630, FT8218, FT9633, FT9674, FT5348, FT8352, T10006, FT8482, FT7615, T10018, FT8718, T10021, FT5380, T10029, FT8274, T10038, FT5408, T10039, FT5190, FT8801, FT5201, T10188, FT9929, FT5379, FT9559, FT9171, FT9739, T10003, FT9742, FT9753, FT5392, FT5425, FT5424, FT5373, FT5153, FT5125, T10081, T10085, FT5456, T10177, T10075, FT5451, T10077, FT6100, T10126, FT6064, FT6037, FT6140, FT6075, FT6150, FT6149, FT5518, FT5532, FT9872, FT5534, FT5576, FT5577, FT5581, T10480, FT5608, T10510, FT5609, T10568, FT5617, T10515, T10732, FT0848, FT1424, FT5396, FT2514, FT0593, FT5596, T10192, FT0213, FT5511, FT1353, FT3316, FT1769, FT2513, FT2214, FT3315, FT2517, FT5590, FT2964, FT4118, TOP0022, T12149, FT4111, FT5673, FT3808, T12153, FT4210, FT0159, T12260, FT0114, T12264, FT2407, FT0294, FT0851, FT1350, FT3119, FT4209, FT2552, FT1080, FT5690, FT2889, FT2679, FT5075, FT0674, FT0761, FT1294, FT5072, FT3211, FT5705, FT1988, FT3164, FT0814, FT1376, FT1213, FT5729, FT4151, FT3104, FT5728, FT0721, FT5714, FT3168, FT3311, FT5709, FT2268, FT5724, FT2680, FT2867, FT5736, FT3805, FT1214, T10978, FT2945, T10997, FT2881, T11000, FT3317, FT5771, T11012, FT3103, T12004, FT1285, FT1288, FT1939, FT1422, FT3366, FT1991, FT1971, FT0466, FT2266, FT2462, FT0052, FT0353, FT1081, FT1425, T12429, FT3313, FT5749, FT2639, FT1504, T12728, FT1366, T12791, FT3148, FT3382, FT5794, FT1987, FT5096, FT5848, T12840, T12842, FT3573, FT5960, FT3438, FT5825, FT5843, FT5817, FT5905, FT1384, FT5904, FT5860, FT5861, FT3618, FT5770, FT5858, FT3298, FT5915, FT5906, FT4835, FT5831, FT4678, FT5973, FT7062, FT5720, FT5859, FT3271, FT9287, FT3412, FT5978, FT7048, FT6028, FT5990, T10857, FT6230, FT2648, FT0891, FT0892, FT0998, FT3619, FT3609, FT4112, FT2226, FT6078, T10917, FT6097, FT7974, FT6060, FT7630, FT6175, FT7755, FT8481, FT8734, FT1267, FT0398, FT2868, FT6107, FT3312, FT7051, FT4137, FT7075, FT6123, T10194, FT2947, FT0356, FT6227, FT3368, FT1268, FT0874, FT3370, FT0155, FT5788, FT7730, FT0876, FT1444, FT5862, FT0694, FT5886, FT1473, FT0355, FT2516, FT5901, FT3209, FT3166, FT0293, FT4046, FT9740, FT2965, FT5985, FT2406, FT2880, T10129, FT3121, FT2405, FT6283, FT0875, FT6285, T10195, FT4107, FT4072, FT4017, FT4203, FT6194, FT3667, FT3208, FT2269, T12094, FT2617, FT4208, FT5641, FT3696, FT5452, FT6287, FT0464, FT1957, FT6263, FT0087, FT0296, FT0093, FT0650, FT2640, FT2213, FT1264, FT3169, FT3030, FT0847, FT2616, FT3149, FT0292, FT1368, FT5496, FT5495, FT4594, FT3768, FT5211, T10276, FT7941, FT7927, FT7943, FT7942, FT4754, FT7938, FT7945, FT7946, FT7947, FT5431, T10669, FT5621, FT8102, FT8112, FT5005, FT8101, FT4951, FT8104, FT5330, T10794, FT4991, FT7944, FT8125, FT8127, FT5293, FT4533, FT4950, FT4500, T10835, FT3862, FT5654, FT3863, T10909, FT3871, FT5648, FT5155, FT5359, T10378, FT5367, FT5370, FT4850, FT3699, T10385, FT3700, FT4100, FT4106, FT5325, T10397, FT3717, FT4223, FT5722, FT3718, FT4442, FT5924, FT4580, FT4592, FT5607, FT0039, FT4995, FT5743, FT5328, FT0115, T10416, T10200, FT5148, FT5195, FT4903, FT0049, T10409, FT4867, T10538, FT5401, FT5400, FT4156, FT4944, FT4165, FT5298, FT3928, FT5112, FT4166, FT5404, FT3944, FT5406, FT4169, T10599, FT5503, FT5409, FT4535, T10537, FT4328, FT5410, FT5684, FT4981, FT4982, FT5217, FT5156, FT4283, FT4384, FT4281, T10592, FT4393, FT4734, FT4399, FT5020, T10611, FT5279, FT5474, FT4937, T10571, FT4457, FT4856, FT4774, FT5625, T10663, FT3721, T10665, FT3722, FT5209, FT5286, FT4475, FT4406, FT3724, FT4843, FT4421, FT4149, FT4481, FT5111, FT5085, FT5579, FT4180, T10759, FT4173, FT4182, FT4186, FT4188, FT4191, FT4423, T10763, FT4197, T12401, T12130, FT4597, FT7551, FT4745, FT5036, FT5438, FT4800, FT5440, FT5038, FT5434, T12419, FT5039, FT0572, FT5718, FT9998, FT5744, FT5752, FT5042, FT5759, FT0797, T12462, FT5778, FT5453, FT2765, T12490, FT2989, T12492, FT5786, FT5845, T10364, FT4915, T10162, FT5294, FT5299, T10292, FT5365, FT5422, FT3774, FT3776, FT5891, FT5907, FT5767, FT5911, FT6034, FT6294, FT5956, T10414, FT5267, T12348, T10007, T12137, T12542, T10017, T12395, FT4239, FT4241, T12194, T10274, T12189, FT5292, FT9642, FT9652, FT1404, T12173, T13200, FT4392, FT3151, FT4250, FT4000, T10750, T10755, T10680, T12111, T10765, FT6162, FT4668, T10677, FT4258, T12113, FT4259, T10155, FT4007, T10164, T12978, FT4008, T13094, T10405, T10273, T11001, FT4684, FT4688, FT4691, FT4692, FT4693, T12241, T10410, FT4702, T10422, FT9877, T10436, FT4011, T10442, FT5114, FT4038, T12279, T10425, FT4324, FT4327, T13118, FT4335, FT4326, FT4336, T10427, FT4339, FT4333, T10429, FT7389, T10430, T10432, FT5799, FT4931, T10433, T10434, T10453, T10473, T10468, FT4055, T10476, T10477, T12037, T10643, T10645, FT8252, T10647, T10734, T12246, T12251, T12243, FT9052, FT2774, FT9119, FT9514, T12376, FT3557, T12423, FT4093, FT3281, FT4094, T12390, FT4098, FT4082, T10991, T10995, FT9688, FT4096, FT4097, T12013, T12019, T10671, T10673, T12051, FT5138, T12103, T12108, T10099, T10101, T10102, FT7289, T12115, T12160, T10921, T11010, FT0041, T10975, FT0393, T12167, T12156, FT1676, T10674, FT1678, T11053, FT2899, FT3470, FT4308, FT2982, T10803, T10825, T12100, T10114, T10724, T10802, T12029, T10888, FT3222, T10394, FT5618, T10893, T10940, T10948, T10324, T12009, T10959, T12016, T10960, T10675, T12131, T10685, T10980, T10983, T10694, FT3231, T12182, T12181, T12426, T12140, T12620, T12263, FT2978, T12708, FT0076, FT9801, T12489, T12828, T12845, T10686, T12853, T12711, FT1195, FT5619, T12427, T12746, T12963, T12475, T12477, T13088, FT5620, T12609, FT1403, FT1393, T12611, T12614, FT2652, T12409, T12692, T12696, T12878, T12873, FT1500, T20099, T12907, FT4811, FT2191, FT3897, T12935, FT2300, FT0920, T13036, FT0919, T13038, FT1642, FT1622, FT4808, T12936, FT2861, T13039, T13056, FT2973, T13071, T12793, T12790, FT2481, FT1931, T12958, FT0918, FT2014, FT1616, FT1618, FT2859, T12973, T12975, FT2836, FT3286, FT5920, FT1867, FT5931, T12985, FT2138, T13072, FT5232, T13073, FT1094, T13077, FT1138, FT2738, FT2251, FT1196, T12827, FT5672, FT1209, FT2289, T13083, FT5674, T20072, T13204, FT5562, FT3926, FT1286, FT5680, FT0559, FT2299, FT1057, T20108, T13202, FT2599, FT2668, T20144, FT1131, FT2905, FT1450, T20228, FT2731, FT3039, FT1401, T20237, FT8942, FT9033, FT9051, FT1391, FT9062, FT7582, FT2908, FT1557, FT1395, FT1620, FT1619, FT7037, FT1644, FT1554, T10041, FT3417, T10804, FT2975, T10768, FT3067, T12804, FT1438, FT7961, T12852, FT1869, FT3416, FT1670, FT3071, FT2935, FT3037, FT2938, FT2663, FT1891, FT1348, FT1892, FT8453, FT1895, FT1901, FT2510, FT8409, T12854, FT2979, FT4016, FT4361, T12874, FT3093, T12917, FT3070, FT8499, T12911, FT8496, FT8215, T12921, FT8577, FT8578, FT1930, FT2013, FT1559, FT3415, FT5810, FT2137, FT8650, FT2903, FT8730, FT2135, TOP0018, FT1671, FT8965, FT2886, FT3234, FT1932, FT9146, FT1898, FT9145, FT3229, FT2015, FT9401, FT1890, FT9421, FT3092, FT4315, FT3226, FT4320, FT2089, FT2790, FT5135, T10080, FT2075, T10142, FT2086, FT3090, FT4415, FT4408, T12737, FT3866, FT1896, T12741, FT1558, FT3069, FT4624, FT4873, FT2085, FT4879, FT2078, FT3781, FT5185, FT4649, FT4803, FT1899, FT5244, FT4683, FT5310, FT5748, FT5852, FT8083, T12957, T10809, FT3086, FT4953, FT2291, FT3282, T10573, FT0640, FT3435, FT3279, FT2250, FT4729, T12912, FT3689, T12307, FT2247, T12976, FT2192, T12964, FT2304, FT7159, T13068, FT3742, T10966, T10988, FT2288, FT2307, T11040, FT5449, FT3427, FT4123, FT5428, FT2476, T13022, T13025, FT2479, FT2547, FT2602, FT2305, FT3073, FT2682, T13121, FT2667, FT2287, T13119, FT2598, FT2674, FT2593, FT7034, T13117, T13112, FT3277, T13018, FT8410, FT3272, T20094, T13126, T13127, T13079, T13189, T13030, T13186, FT3095, FT2887, FT1900, T13136, T12995, FT3426, T13258, T13179, FT7404, FT8960, FT2902, FT9014, FT2155, FT2675, FT2505, FT2518, FT2455, FT3036, FT1893, FT2649, FT2926, FT2977, FT3182, FT3244, FT3457, FT3461, FT3536, FT3550, FT3566, FT3547, FT4006, T10780, T13219, T10785, T10789, T13190, T10617, T12223, T12350, T13286, FT7010, FT8016, T13093, T13212, T13261, FT7518, FT7519, T20290, FT7520, FT7521, FT7522, T13192, FT6125, T20284, T20172, FT7566, FT7739, FT6126, FT7963, FT7511, FT7964, FT7512, FT7970, FT7972, FT7973, FT8456, FT8459, FT8486, FT8631, FT8989, FT8991, FT8993, FT8551, FT8994, FT9006, FT9012, FT8580, FT8605, FT7218, FT8654, FT9600, FT9603, FT9609, FT7238, FT9731, FT9866, T10166, FT6181, FT1529, FT2034, FT6210, FT7254, FT7258, FT2730, FT7251, FT2820, FT2821, FT2822, FT2720, FT7463, FT2750, FT3331, FT7020, FT3939, FT3932, FT1855, FT7275, FT2261, FT2869, FT3994, FT3567, FT4506, FT4538, FT7024, FT4546, FT4233, FT4587, FT4588, FT5285, FT7495, FT7496, FT5306, FT5304, FT4568, FT4570, FT4819, T10290, FT5949, FT5954, FT4559, FT5957, FT5961, T10383, FT6004, FT6047, FT7531, FT7552, FT6267, FT6053, FT5977, FT6276, FT7028, FT5979, FT7042, T10934, FT7044, FT7558, FT7047, FT7587, T12341, T12408, FT7329, T12458, T12463, T12528, T12529, T12531, T12556, T12557, FT7350, FT7599, FT7353, FT7359, T12758, FT5998, T12954, T12979, T12981, T13108, FT5993, T13085, FT7625, FT6011, FT6014, FT7077, FT7085, FT9141, FT7678, FT9135, FT5874, FT9152, FT9178, FT9556, FT5893, FT9500, FT5897, FT9478, FT7682, FT9598, FT5898, FT6048, FT7691, FT9463, FT7695, FT9809, FT7710, FT7747, FT7753, FT6049, FT6081, FT6082, FT7091, FT7136, FT7153, FT7158, FT6101, FT6103, FT3398, FT3638, FT3646, FT3647, FT3648, FT5939, FT5938, FT0730, FT1021, FT3652, FT3656, FT3664, FT3691, FT3692, FT1630, FT2968, FT2969, T10198, FT9589, FT9585, FT9645, FT9723, FT5160, FT4557, FT9875, FT9879, FT8427, T10084, FT9168, FT9356, FT9457, FT5194, FT9458, FT9597, FT9596, FT9595, FT9456, T12018, T12020, FT2497, FT0168, FT0287, T10257, FT0257, FT0256, T10377, FT0415, T10372, FT0775, T13106, FT1018, FT1478, FT7401, FT1315, FT7402, FT7403, FT1685, FT7468, FT2298, FT2264, FT1713, FT2488, FT2542, FT3296, FT3363, FT3859, FT3900, FT8067, FT3940, FT4012, FT7071, FT3979, FT7094, FT1756, FT0702, FT3544, FT2716, FT4414, FT7103, FT7149, FT7187, FT4516, FT4445, FT7188, FT7148, FT7284, FT4576, FT4590, FT4682, FT4689, FT8021, FT4519, FT8011, FT4886, FT0310, FT3329, FT8070, FT3607, FT2309, FT1025, FT8168, FT1316, FT4893, FT8287, FT4920, FT8317, FT4946, FT4947, FT4973, FT5006, FT8439, FT5024, FT8880, FT5301, FT8643, FT5488, FT8972, FT9099, FT9089, FT9075, FT6059, FT6065, FT5851, FT5902, FT5903, FT6068, FT3972, FT5236, FT6161, FT6202, FT6222, FT5239, FT4419, T10832, FT4409, T10391, FT4452, FT4456, FT4589, FT4564, FT5739, T12127, T12335, FT5883, T12352, T12353, T12354, FT4666, T12385, FT4801, FT0098, T12400, T10964, T11031, T12122, T12102, FT3262, FT2650, T12795, T10749, T10880, FT6203, T12993, T13066, T13067, T13070, FT3480, FT4917, T10872, T12431, T12411, FT1774, FT8018, FT2263, T10859, FT7111, FT7114, T10927, T10943, T12152, FT7316, FT7318, FT7337, T12333, FT7766, FT7767, FT7667, FT8030, FT8172, T12886, FT8286, FT8206, T12938, T12940, FT8239, FT8303, T12784, T12859, FT8387, FT8570, T20147, T20161, FT8582, T20203, FT8583, T20210, T20222, FT8584, T20306, T20310, FT8736, T20212, T20326, FT7451, FT5909   , FT7442, FT7631, FT7621, FT7550, FT7472, FT7475, FT7461, FT7473, FT9408, FT7545, FT7526, FT9650, FT9732, FT7593, FT9787, FT7591, FT7592, FT9816, FT9861, FT9860, FT9869, FT9423, FT7594, FT7601, FT7597, FT7622, FT7620, FT9949, T10040, T10025, T10139, T10156, FT7614, FT7671, FT7658, FT7670, FT0105, FT7669, FT7672, FT7686, FT0712, FT7697, FT1146, FT7684, FT7698, FT1355, FT1456, FT7727, FT7725, FT2066, FT2178, FT1143, FT3295, FT7744, FT7746, FT7763, FT3872, FT4059, FT7783, FT4466, FT4555, FT4888, FT5272, FT6321, FT6315, FT6319, FT5510, FT7001, FT5746, FT7013, FT7004, FT5934, FT5014, FT3031, T10903, T10908, FT7049, FT7072, FT7066, T10618, FT7093, FT7113, T10767, T10796, T12403, T12406, FT7794, FT7859, FT7855, FT7848, FT7863, FT7869, T12150, FT7867, FT7845, FT7868, FT7879, FT7880, FT7893, FT7890, T12292, T12132, T12119, FT7896, FT7891, FT7886, FT7892, FT7898, FT7899, T12345, T12343, FT8263, FT8264, T10839, T10915, FT8115, FT8187, FT8190, T10847, FT8194, T10967, FT8178, FT8180, FT8179, FT8196, T12118, FT8222, T12176, T12184, T12534, T12142, FT7138, FT7156, FT7155, T12161, FT7167, T12738, FT7180, FT7178, FT7179, FT7206, FT7216, T13064, FT7195, FT7199, FT7200, FT7207, FT7202, FT7208, FT7217, T13155, T13158, FT7212, FT7223, T13124, FT7229, FT7226, FT7233, FT7243, FT7250, T13092, FT7271, FT7424, FT7425, FT7263, FT7302, FT4999, FT7280, FT7283, FT5044, FT7287, FT7003, FT7304, FT7073, FT7306, FT7074, FT7308, FT7059, FT7310, FT7588, FT7718, FT7312, FT7792, FT8015, FT8080, FT5854, FT7349, FT7382, FT7376, FT7384, FT9096, FT7395, FT7391, FT9363, FT9365, T10028, FT7417, FT7422, FT7791, FT5013, FT7819, FT7820, FT7824, FT0800, FT7831, FT0779, FT0870, FT0871, FT7842, FT1414, FT7846, FT7847, FT7850, FT1536, FT7858, FT2354, FT2530, FT2498, FT7905, FT3268, FT3466, FT7917, FT7923, FT3529, FT7928, FT3612, FT7933, FT7935, FT7910, FT4060, FT7901, FT4249, FT4269, FT4268, FT4364, FT4385, FT4397, FT4699, FT7903, FT4591, FT5091, FT5086, FT5245, FT5087, FT5259, FT4318, FT5334, FT5338, FT5308, FT5366, FT5448, FT5457, FT5566, FT3890, FT5682, FT7955, FT5741, FT7957, FT6038, FT1544, FT6041, FT5912, T10667, T10772, FT7980, T12410, FT7940, FT7953, FT7960, FT7982, T12136, T12328, FT7995, T10769, FT7989, FT7996, FT7998, T12465, T12494, FT7993, FT8006, FT7999, T13262, FT7968, T13272, T20164, T20173, FT8007, FT8009, FT8000, FT8013, FT6247, FT8024, FT8036, FT7067, FT7076, FT7078, FT8005, FT7636, FT8020, FT8028, FT5643, FT8033, FT5650, FT5651, FT8075, FT7164, FT8061, FT8060, FT7191, FT7240, FT7925, FT7777, FT7778, FT8062, FT8078, FT8079, FT8081, FT8088, FT8294, FT9347, FT8216, FT8313, FT8314, FT8315, FT8316, FT8434, FT8435, FT8437, FT8398, FT8425, FT3021, FT8416, FT8132, FT8068, FT8069, FT8545, FT8447, FT8440, FT8443, FT8445, FT8449, FT8420, FT8567, FT8549, FT8553, FT8123, FT8118, FT9216, FT9207, FT9222, FT7633, FT8334, FT9810, FT8344, FT8347, FT9858, FT4132, FT1223, FT8374, FT8375, FT8386, FT1722, FT0060, FT2020, FT2143, FT1853, FT9431, FT2847, FT2872, FT9286, FT9459, FT3950, FT9460, FT3861, FT9465, FT3969, FT3988, FT4352, FT4348, FT4356, FT4429, FT4434, FT9488, FT1611, FT4527, FT1102, FT9480, FT1579, FT9494, FT0438, FT1610, FT4852, FT9546, FT4932, FT4938, FT9313, FT9554, FT8388, FT8393, FT8470, FT6281, FT3033, FT8441, FT4902, FT8472, FT8480, T10529, FT8469, FT8477, FT8281, FT8468, FT8484, FT8488, FT8492, FT8517, FT8526, FT8527, FT5483, FT8531, FT8532, T11019, FT8609, FT8562, FT7189, FT7184A, FT8615, FT8617, T12908, FT8618, T12952, T12725, T12727, FT8621, FT8681, T13120, FT8626, FT8629, FT8635, FT8657, FT8673, T13257, T20256, FT8675, FT8685, FT8646, FT8666, FT8670, FT8671, FT8687, FT9611, FT8703, FT8704, FT8707, FT8735, FT8761, FT8756, FT8764, FT2278, FT3614, FT8614, FT8765, FT8791, FT8797, FT5455, FT8789, FT7029, FT8790, FT8792, FT8799, FT7433, FT8795, FT8501, FT8541, FT8842, FT8430, FT8800, FT8839, FT8844, FT8847, FT8841, FT9411, FT8857, FT8833, FT8871, FT8875, T10093, FT8908, FT8885, FT8893, FT8903, FT8874, FT8475, FT8462, FT1648, FT8930, FT2116, FT2499, FT8951, FT9021, FT9023, FT9030, FT3880, FT9079, FT9088, FT9067, FT9082, FT9071, FT9104, FT9186, T10389, T10386, T10438, T10441, T10212, FT9242, T12177, FT9290, FT9292, FT9270, FT9274, FT9279, FT9283, FT9275, T10349, FT6318, T10423, FT9285, FT7058, FT9518, FT7259, FT7262, FT9586, FT9523, FT9522, FT9558, FT9562, FT9564, FT8368, FT9560, FT9581, FT9587, FT9593, FT9602, FT9672, FT9623, FT9619, FT9453, FT9632, FT9627, FT9637, FT9639, FT9628, FT9647, FT9651, FT9658, FT9681, FT9684, FT9680, FT9683, FT9482, FT9678, FT9669, FT2222, FT9698, FT9638, FT2223, FT2504, FT9677, FT9715, FT9315, FT2520, FT2654, FT9317, FT9320, FT2818, FT9404, FT2976, FT3176, FT9345, FT3110, FT3264, FT3336, FT9343, FT9346, FT3616, FT9366, FT3783, FT9369, FT3951, FT3883, FT9476, FT4437, FT9193, FT3481, FT4820, FT9737, FT5162, FT9693, FT9699, FT9701, FT9702, FT9721, FT9719, FT5416, FT9718, FT5417, FT5419, FT5421, FT9757, FT9758, FT4914, FT9766, FT5067, FT5853, FT5943, FT9793, FT9778, FT9792, FT6224, FT6177, FT9795, FT9802, FT9798, FT9807, FT9815, FT9820, T10879, FT9821, T10886, T10891, T10775, FT9710, FT9834, FT9835, FT9864, FT9865, T10446, FT9876, FT9889, FT9868, FT9888, T12404, FT9892, FT9897, T12396, FT9902, FT9911, FT9934, FT9922, T10912, FT9932, FT9925, FT9933, T12781, T12787, FT9939, T12855, T12785, T12925, FT9944, T12927, FT9948, T12947, T12949, FT9953, T13061, FT9976, T12969, FT9983, FT9975, FT9978, FT9980, FT9984, FT9985, FT9987, T10010, T10027, T10033, T10043, FT8888, FT8891, FT8863, FT8865, FT8866, T10072, T10074, T10067, T10071, FT9818, FT9827, FT9828, T10070, T10095, T10069, T10113, T10144, T10147, T10148, T10149, T10159, T10175, T10210, T10185, T10207, T10213, T10223, T10711, T10715, T10241, FT2472, T10250, T10516, FT8032, T10279, T10288, T10285, T10286, T10287, FT7874, FT7872, FT8142, FT8181, FT8597, FT9381, FT9324, T10322, T10357, FT9348, T10352, FT9306, FT8182, FT8373, T10401, FT8237, FT0243, FT5524, FT5531, FT5537, FT5541, FT8272, FT5569, FT8380, FT8550, FT8512, FT8529, FT8509, FT1969, FT8503, FT8636, FT1970, FT8812, FT8630, FT1912, FT2012, FT2061, FT1980, FT8890, FT2052, FT8505, FT8879, FT2199, FT9043, T10173, T10219, T10221, FT9318, FT2197, FT9298, FT8883, FT8886, FT9540, FT9370, FT9372, FT9608, FT9257, FT9262, FT8678, FT9252, FT8887, FT9258, FT9662, FT9661, FT9648, FT9657, FT9836, FT9832, FT2246, FT9768, FT9915, FT9958, FT2095, FT9920, FT9954, FT2169, FT2096, FT9951, FT2106, T10168, FT9034, FT1784, FT1782, FT2218, FT1497, FT2660, FT2149, FT1997, FT1185, FT2217, FT1184, FT1703, FT1187, FT1787, FT1995, FT2240, FT1700, FT1045, FT1191, FT1046, FT2188, FT2242, FT1998, FT1495, FT2104, FT2148, FT0908, FT1859, FT2658, FT1702, FT2187, FT2237, FT1680, FT1788, FT2185, FT1183, FT1862, FT1189, FT1614, FT1615, FT1514, FT2109, FT2141, FT2239, FT2045, FT2243, FT1612, FT1682, FT2270, FT2335, FT2244, FT3006, FT3007, FT2276, T10231, T10232, T10256, T10418, T10336, T10267, T10412, T10265, T12366, FT2359, FT2358, T10816, T12504, FT2001, T10237, FT9928, T10226, FT2190, FT9942, FT2048, T12061, T12239, FT2150, T12290, FT2050, T12368, T12436, T12441, T12437, T12438, T10424, T10954, T12536, T12550, FT2235, T12560, T12448, T12444, T12446, T12447, T12468, T12763, T12469, T12478, T12760, T12762, T12505, FT2156, FT2160, T12864, T12900, T12903, T13040, T13046, T20156, T13047, FT2529, T20157, FT2429, FT2523, FT2401, FT3263, FT2402, FT2465, FT7862, FT7875, FT7857, FT0048, FT2418, FT7309, FT7387, FT2441, FT7840, FT2435, FT2459, FT8130, FT2437, FT8184, FT2397, FT2398, FT9377, FT2506, FT8561, FT8558, FT8564, FT8569, FT8661, FT2399, FT8663, FT2457, FT2508, FT8599, FT8762, FT2509, FT8824, FT8828, FT8851, FT2473, FT2554, FT2555, FT2556, FT2558, FT2559, FT9314, FT9483, FT9484, FT9536, FT9726, FT3291, FT3506, FT3520, FT3521, FT3522, FT3199, FT3204, FT5538, FT3189, FT5544, FT3185, FT0037, FT0017, FT2624, FT0134, FT0141, FT0015, FT0013, FT0016, FT0020, FT0023, FT0027, FT0064, FT0062, FT0088, FT0066, FT0091, FT0178, FT0089, FT0179, FT0449, FT0595, FT0824, FT0909, FT0912, FT1525, FT1535, FT1534, FT1524, FT1690, FT2210, FT1691, FT1749, FT2364, FT2744, FT2724, FT0167, FT0172, FT3260, FT3051, FT3341, FT0277, FT4822, FT4311, FT5207, FT0151, FT0202, FT0193, FT0262, FT2984, FT2770, FT2789, FT0203, FT0237, FT2897, FT2816, FT2798, FT2864, FT0242, FT0334, FT2877, FT0144, FT2842, FT0244, FT0247, FT0246, FT2878, FT0346, FT2988, FT0216, FT2928, FT0215, FT3016, FT0336, FT0335, FT0333, FT0460, FT0364, FT0225, FT3025, FT2879, FT0338, FT3127, FT0362, FT3064, FT0359, FT0358, FT3157, FT3128, FT0579, FT3061, FT3063, FT3131, FT3138, FT3062, FT3048, FT3047, FT0446, FT3139, FT0396, FT3141, FT0447, FT3144, FT3145, FT0509, FT3101, FT0380, FT0511, FT0508, FT0565, FT0404, FT0620, FT3200, FT0507, FT0619, FT0537, FT0451, FT3170, FT0551, FT0550, FT3083, FT3065, FT0388, FT0514, FT3098, FT3102, FT0516, FT0387, FT3078, FT3130, FT0386, FT3289, FT0545, FT0389, FT0586, FT0544, FT0625, FT0526, FT0588, FT0587, FT3123, FT0549, FT0528, FT3220, FT0473, FT3221, FT0472, FT1086, FT0633, FT0501, FT0790, FT0412, FT0664, FT3288, FT0717, FT3310, FT3253, FT3320, FT3290, FT3276, FT0810, FT3257, FT3255, FT0735, FT0822, FT0823, FT3385, FT4301, FT3429, FT0691, FT0787, FT3405, FT0696, FT0862, FT0835, FT0833, FT0834, FT0841, FT0843, FT0856, FT0855, FT0883, FT0877, FT0879, FT4227, FT4213, FT1420, FT4221, FT4243, FT2833, FT4307, FT0673, FT4256, FT4257, FT4279, FT4353, FT4350, FT4357, FT4358, FT4376, FT4371, FT4373, FT4374, FT4379, FT4396, FT4337, FT4412, FT4477, FT4503, FT4511, FT4484, FT4526, FT4474, FT4476, FT4461, FT4459, FT4522, FT4569, FT4454, FT4451, FT5605, FT4593, FT4575, FT5058, FT5054, FT5161, FT5149, FT5151, FT5152, FT5124, FT5127, FT5129, FT5131, FT5137, FT5143, FT5187, FT5178, FT5268, FT5105, FT5271, FT5302, FT5309, FT5311, FT5223, FT5269, FT5363, FT5230, FT5329, FT5276, FT5277, FT5243, FT5278, FT5246, FT5280, FT5283, FT5331, FT5374, FT5381, FT5403, FT5399, FT5397, FT5323, FT5314, FT5291, FT5264, FT5387, FT5393, FT5427, FT0905, FT1174, FT1009, FT1167, FT1005, FT1088, FT1168, FT1818, FT1097, FT1326, FT1817, FT1582, FT1336, FT1733, FT1709, FT1466, FT1467, FT1014  BFX#3, FT1819, FT1019 BFX#1, FT1082, FT4085, FT1820, FT1584, FT1585, FT1821, FT1823, FT1718, FT0934, FT0932, FT1345, FT1729, FT1179, FT1730, FT1180, FT1182, FT1731, FT3446, FT1132, FT1361, FT1598, FT1595, FT1594, FT3444, FT1487, FT1486, FT1827, FT1828, FT1839, FT1734, FT1735, FT1737, FT1166, FT4204, FT4193, FT4183, FT1841, FT1842, FT4190, FT4198, FT3512, FT4205, FT3807, FT5464, FT3809, FT5469, FT2725, FT3725, FT3727, FT3728, FT3730, FT3640, FT3732, FT3339, FT4086, FT4089, FT3632, FT3560, FT3629, FT3631, FT3623, FT3633, FT4171, FT3868, FT3874, FT3875, FT3879, FT3888, FT3889, FT4172, FT4174, FT5585, FT5669, FT5685, FT5694, FT5691, FT5699, FT0901, FT0971, FT0972, FT0975, FT0922, FT1041, FT1044 BFX#10, FT1043, FT1375, FT1540, FT1637, FT1378, FT1390, FT1876, FT1664, FT3757, FT1872, FT1751, FT1875, FT1873, FT1761, FT1218, FT1238, FT1232, FT1760, FT1415, FT1665, FT1538, FT1233, FT1772, FT1246, FT1789, FT1887, FT1412, FT1771, FT1635, FT1742, FT1903, FT1905, FT1908, FT3760, FT0771, FT3694, FT3906, FT3818, FT3819, FT3511, FT3827, FT3829, FT2047, FT2069, FT3581, FT3761, FT3758, FT3912, FT3913, FT3375, FT3442, FT3750, FT3914, FT3898, FT3751, FT3755, FT3903, FT3904, FT3780, FT3953, FT3759, FT3957, FT4029, FT3961, FT3965, FT4018, FT3915, FT3918, FT3766, FT4033, FT4039, FT4034, FT1312, FT3771, FT2253, FT5482, FT3574, FT3779, FT3563, FT4895, FT3787, FT3789, FT3564, FT3676, FT3671, FT5631, FT1141, FT5603, FT5602, FT5629, FT5623, FT1427, FT1085, FT1075, FT1071, FT0903, FT1299, FT1443, FT1165, FT1426, FT1007, FT1261, FT1546, FT1307, FT1054, FT1569, FT1688, FT1313, FT1699, FT1693, FT1418, FT1269, FT1270, FT1705, FT1707, FT1283, FT1419, FT3709, FT5463, FT3517, FT4153, FT4992, FT4979, FT5017, FT4998, FT4145, FT3986, FT4158, FT4163, FT4167, FT3516, FT3584, FT3515, FT3585, FT3503, FT3838, FT3852, FT3704, FT3502, FT3711, FT3921, FT3929, FT4892, FT3945, FT4868, FT3948, FT3849, FT3847, FT4875, FT4880, FT4890, FT3854, FT3853, FT4043, FT3856, FT3855, FT4040, FT4045, FT4069, FT5516, FT5666, FT5514, FT5515, FT5519, FT5520, FT5521, FT4630, FT2613, FT4631, FT2628, FT4617, FT4620, FT4629, FT2684, FT2678, FT2636 BFX#8, FT4635, FT4641, FT4644, FT4642, FT4645, FT4646, FT4648, FT4653, FT4705, FT2632, FT4685, FT2686, FT4913, FT2685, FT2689, FT4962, FT4926, FT4963, FT5027, FT5028, FT5031, FT5040, FT0078, FT2700, FT0131, FT2621, FT2622, FT0080, FT2703, FT2691, FT4744, FT4611, FT4608, FT4613, FT4746, FT4743, FT4742, FT4749, FT4838, FT4842, FT4787, FT4853, FT4793, FT4768, FT4718, FT4809, FT4821, FT4807, FT4710, FT4794, FT4815, FT4816, FT4832, FT4716, FT4719, FT4721, FT5016, FT4725, FT4731, FT5000, FT5015, FT5725, FT2883, FT2895, FT2896, FT5789, FT5784, FT5797, FT3501, FT5774, FT5802, FT5800, FT5808, FT5793, FT5822, FT5812, FT5834, FT5835, FT5842, FT5844, FT5850, FT6076, FT6085, FT6080, FT6092, FT6090, FT6094, FT6098, FT6108, FT6114, FT5866, FT5863, FT5864, FT5868, FT5870, FT5716, FT5877, FT5878, FT5892, FT5896, FT4049, FT5914, FT5713, FT5750, FT5926, FT5928, FT5933, FT5935, FT5955, FT5962, FT0676, FT0675, FT5969, FT5988, FT5989, FT6006, FT6009, FT6008, FT5932, FT6018, FT6019, FT6036, FT6054, FT6051, FT3524, FT6066, FT6072, FT0207, FT6124, FT6113, FT6130, FT6131, FT6129, FT6143, FT6184, FT6174, FT6179, FT6211, FT6189, FT6217, FT6213, FT6228, FT6209, FT6188, FT6226, FT6197, FT6198, FT6205, FT6235, FT6218, FT6196, FT6243, FT6231, FT6244, FT6252, FT6257, FT6259, FT6289, FT6266, FT6280, FT3335, FT6297, FT6304, FT6302, FT6311, FT6310, FT6313, FT6316, T10448, T10479, T10483, T10485, T10489, T10518, T10345, T10502, T10500, T10505, T10509, T10875, T10882, T10895, T10928, T10968, T10946, T10950, T10914, T10989, T10993, T11007, T10356, T10355, T10316, T10353, T10543, T10545, T10544, T10548, T10581, T10557, T10622, T10577, T10560, T10580, T10585, T10620, T10625, T10630, T10613, T10616, T10639, T10642, T10648, T10623, T10586, T10668, T10678, T10681, T10662, T10710, T10641, T10644, T10661, T10658, T10720, T10532, T10729, T10773, T10781, T10777, T10764, T10776, T10787, T10786, T10782, T10814, T10812, T10811, T10810, T10833, T10843, T10831, T10851, T10870, T10850, T10863, T10890, T10900, T10919, T11005, T12564, T12334, T12553, T12581, T12573, T12575, T12574, T12632, T12578, T10493, T10469, T12082, T12079, T12091, T12109, T12080, T12272, T12262, T12316, T12326, T12412, T12414, T12420, T12432, T12433, T12487, T12526, T12519, T12535, T12398, T12538, T12541, T12543, T12546, T12547, T11058, T12001, T12010, T11006, T11041, T11047, T11051, T11056, T12007, T12008, T12017, T12011, T12035, T12050, T12054, T12055, T12056, T12057, T12064, T12069, T12647, T12649, T12650, T12663, T12579, T12669, T12698, T12674, T12676, T12685, T12695, T12707, T12724, T12697, T12703, T12705, T12699, T12704, T12702, T12856, T12862, T12865, T12880, T12866, T12869, T12891, T12896, T12897, T12899, T12913, T12910, T12914, T12915, T12916, T12931, T12932, T12939, T12941, T12942, T12944, T12943, T12946, T12951, T20021, T20029, T20005, T20009, T20027, T20012, T20011, T20015, T12274, T12715, T12719, T12750, T12752, T12755, T12757, T12764, T12773, T12774, T12775, T12794, T12726, T12805, T12810, T12812, T12808, T12811, T12815, T12814, T12821, T12824, T12822, T12832, T12846, T12850, T12851, T20006, T20010, T20014, T20022, T20024, T20017, T20035, T20042, T20031, T20051, T20039, T20040, T20046, T20055, T20058, T20067, T20061, T20034, T20044, T20047, T20063, T20037, T20048, T20064, T20069, T20073, T20076, T20078, T20074, T20081, T20087, T20107, T20090, T20093, T20095, T20096, T20098, T20106, T20104, T20109, T20116, T20118, T20112, T20120, T20136, T20140, T20133, T20134, T20146, T20184, T20160, T20171, T20186, T20197, T20187, T20188, T20189, T20181, T20185, T20194, T20196, T20205, T20208, T20209, T20214, T20204, T20207, T20218, T20221, T20201, T20217, T20191, T20220, T20227, T20254, T13274, T13264, T13282, T13294, T13297, T13275, T13285, T13312, T20292, T20294, T20285, T20305, T13284, T13313, T13328, T20332, T13298, T13302, T13303, T20229, T20322, T20268, T20333, T20349, FT6083, FT7949, FT8117, FT8511, FT8674, FT1633, FT3075, FT3002, FT3116, FT3160, FT3134, FT3219, FT3509, FT3510, FT6248, T10330, FT8806, FT8816, FT8827, FT9708, FT9819, T10284, FT3943, FT5443, FT5492, T10926, T10931, T10933, T10935, T10899, T10901, T10916, T12219, T12277, FT9417, FT9464, FT9466, FT8750, FT9930, T10225, FT3796, FT2959, FT2807, FT3184, FT3162, FT4228, FT5446, FT5668, FT5819, T13113.",
      "firm_fei_number": "1119779",
      "recalling_firm": "Becton Dickinson & Co.",
      "address_1": "7 Loveton Cir",
      "city": "Sparks",
      "state": "MD",
      "postal_code": "21152-9212",
      "additional_info_contact": "North American Regional Complaint Center\n410-316-4000",
      "reason_for_recall": "BD confirmed that product service credentials used by some BD technical support teams to access certain BD products were accessed by an unauthorized actor. Until these product service credentials are updated, there is a risk of unauthorized access that may impact the confidentiality, integrity and/or availability of the relevant products and associated data.",
      "root_cause_description": "Software design",
      "action": "On October 31, 2024 URGENT\" MEDICAL DEVICE PRODUCT CORRECTION letters were sent to consignees. IT, Safety and Security Actions to be Taken:\nBD encourages customers to follow best practices for maintaining strong security measures to protect hospital networks and medical devices including:\n\"Ensure access to potentially vulnerable devices is limited to authorized personnel\n\"Inform authorized users of issue, and ensure all relevant passwords are tightly controlled\n\"Monitor and log network traffic attempting to reach medical device managementenvironments for suspicious activity\n\"Where possible, isolate affected devices in a secure VLAN or behind firewalls with restrictedaccess that only permits communication with trusted hosts in other networks when needed\n\"Impacted devices do not require use of RDP ports and these should be disabled or blockedif enabled\n\"Ensure permissions on file shares are appropriately established and enforced, and monitorand log access for evidence of suspicious activity\n\"Disconnect devices from the network if connectivity is not necessary\nNext Steps to be Taken by Customer:\n1.Ensure the contents of this notification are read and understood.\n2.Share and post this customer letter within your facility network including IT and safety andsecurity department. Forward to any customers to whom you may have distributed theproduct(s) to ensure awareness.\n3.Complete the attached Customer Response Form and return it to the BD contact noted on theForm so that BD may acknowledge your receipt of this notification per FDA requirements.\n4.Report any adverse health consequences experienced with the use of these products to BD.Events may also be reported to the FDA's MedWatch Adverse Event Reporting program via:\nWeb: MedWatch website at www.fda.gov/medwatch Phone: 1-800-FDA-1088 (1-800-332-1088)\nMail: MedWatch, HF-2, FDA, 5600 Fisher s Lane, Rockville, MD 20852-9787\nIf you require further assistance, please contact: Technical Service\n+1 888 451 5514\nTechnical Questi",
      "product_quantity": "10,129 units (2,836 US, 7,293 OUS)",
      "distribution_pattern": "Worldwide distribution - US Nationwide and the countries of Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Benin, Bhutan, Bolivia, Bosnia and Herzegovina, Botswana, Brazil, Brunei, Bulgaria, Burkina Faso, Burundi, Cambodia, Cameroon, Canada, Chile, China, Colombia, Costa Rica, Cote d'Ivoire, Croatia, Cyprus, Czech Republic, Democratic Republic of the Congo, Denmark, DOM, Dominica, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Ethiopia, Faroe Islands, Fiji, Finland, France, French Guiana, Gambia, Georgia, Germany, Ghana, Greece, Guadeloupe, Guam, Guatemala, Guinea, Guyana, Haiti, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Jamaica Japan, Jordan, Kazakhstan, Kenya, Kosovo, Kuwait, Kyrgyzstan, Laos, Latvia, Lebanon Liberia, Libyan Arab Jamahiriya, Liechtenstein, Lithuania, Luxembourg, Macau, Madagascar Malawi, Malaysia, Maldives, Mali, Malta, Martinique, Mauritania, Mauritius, Mayotte Mexico, MKD, Moldova, Mongolia, Morocco, Mozambique, Namibia, Nepal, Netherlands Antilles, New Caledonia, New Zealand, Nicaragua, Nigeria, Northern Mariana Islands, Norway, Oman, Pakistan, Panama, Papua new Guinea, Paraguay, Peru, Philippines PNG, Poland, Portugal, Puerto Rico, Qatar, Reunion, Romania, Russia, Rwanda, Saint Helena San Marino, Saudi Arabia, Senegal, Serbia, Seychelles, Sierra Leone, Singapore, Slovakia Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sudan, Swaziland, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Tajikistan, Tanzania, United Republic of Thailand, Timor-Leste, Tonga, Trinidad And Tobago, Tunisia, Turkey, Turkmenistan Uganda, Ukraine, United Arab Emirates, United Kingdom, Uruguay Uzbekistan, Vietnam, Yemen, Zambia, Zimbabwe.",
      "openfda": {
        "k_number": [
          "K954921",
          "K925693",
          "K121455",
          "K222591",
          "K151866",
          "K161816",
          "K973325",
          "K020813",
          "K962210",
          "K993576",
          "K970333",
          "K120994",
          "K981736",
          "K021123",
          "K973305",
          "K954932",
          "K032306",
          "K121461",
          "K972756",
          "K983772",
          "K992400",
          "K993421",
          "K031737",
          "K915796",
          "K992401",
          "K946268",
          "K812315",
          "K173873",
          "K954468",
          "K982263",
          "K935711",
          "K123903",
          "K922475",
          "K920933",
          "K934593",
          "K161810",
          "K260213",
          "K810613",
          "K994343",
          "K222559",
          "K020923",
          "K993423",
          "K020815",
          "K113558",
          "K954927",
          "K190405",
          "K141810",
          "K992432",
          "K161306",
          "K944594",
          "K183166",
          "K974883",
          "K954925",
          "K921133",
          "K903505",
          "K926304",
          "K953129",
          "K920249",
          "K020931",
          "K954924",
          "K954930",
          "K833888",
          "K083572",
          "K000378",
          "K121446",
          "K972758",
          "K921637",
          "K970512"
        ],
        "registration_number": [
          "3013188547",
          "3002769706",
          "3013718871",
          "3012421607",
          "1530126",
          "2647876",
          "3031977134",
          "3003030793",
          "3013557562",
          "2124274",
          "3009637528",
          "3016678045",
          "1950204",
          "1119779",
          "1111096",
          "1025402",
          "3003542872",
          "1423662",
          "1924669"
        ],
        "fei_number": [
          "3013188547",
          "1000515493",
          "3002769706",
          "3013718871",
          "3012421607",
          "2647876",
          "3031977134",
          "3003030793",
          "3013557562",
          "3000215273",
          "2124274",
          "3009637528",
          "3016678045",
          "1950204",
          "1119779",
          "1111096",
          "1025402",
          "3003542872",
          "1924669"
        ],
        "device_name": "System, Blood Culturing",
        "medical_specialty_description": "Microbiology",
        "regulation_number": "866.2560",
        "device_class": "1"
      }
    }
  ]
}