{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-08-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "210816",
      "product_res_number": "Z-0649-2025",
      "event_date_initiated": "2024-10-09",
      "event_date_posted": "2024-12-05",
      "recall_status": "Open, Classified",
      "res_event_number": "95621",
      "product_code": "OJH",
      "product_description": "MEDLINE convenience kits labeled as:  \n1) ORTHO LAMINECTOMY-LF, REF CDS780112J; \n2) NEURO CRANIOTOMY CDS #36-RF, REF CDS780119V; \n3) BASIC BACK CDS, REF CDS780147N; \n4) INSTRUMENTED BACK CDS, REF CDS780148Q; \n5) INSTRUMENTED BACK CDS, REF CDS780148R; \n6) TOTAL HIP, REF CDS920027W; \n7) TOTAL HIP, REF CDS920027X; \n8) TOTAL HIP, REF CDS920027Y; \n9) TOTAL KNEE CDS-LF, REF CDS940047AF; \n10) TOTAL KNEE CDS-LF, REF CDS940047AG; \n11) TOTAL HIP CDS-LF, REF CDS940048AF; \n12) TOTAL HIP CDS-LF, REF CDS940048AG; \n13) TOTAL HIP CDS-LF, REF CDS940048AI; \n14) TOTAL KNEE, REF CDS940072W; \n15) TOTAL KNEE, REF CDS940072Y; \n16) TOTAL KNEE REPLACEMENT CDS, REF CDS940823AB; \n17) TOTAL KNEE CDS, REF CDS940842Y; \n18) TOTAL HIP CDS, REF CDS940843AA; \n19) FRACTURE TABLE-LF, REF CDS940857K; \n20) FRACTURE TABLE-LF, REF CDS940857L; \n21) EXTREMITY-LF, REF CDS940860P; \n22) EXTREMITY-LF, REF CDS940860Q; \n23) KNEE/SHOULDER, REF CDS940911Q; \n24) TOTAL HIP CDS, REF CDS940963Y; \n25) SHOULDER CDS, REF CDS941065L; \n26) TOTAL JOINT-LF, REF CDS941167M; \n27) TOTAL JOINT-LF, REF CDS941167N; \n28) LOWER EXTREMITY CDS-LF, REF CDS980666I; \n29) LOWER EXTREMITY CDS-LF, REF CDS980666J; \n30) TOTAL HIP-LF, REF CDS980788O; \n31) TOTAL HIP-LF, REF CDS980788P; \n32) EXTREMITY KIT, REF CDS981314R; \n33) EXTREMITY KIT, REF CDS981314S; \n34) ORIF HIP KIT, REF CDS981318T; \n35) ORIF HIP KIT, REF CDS981318U; \n36) TOTAL HIP KIT, REF CDS981320V; \n37) TOTAL HIP KIT, REF CDS981320W; \n38) TOTAL KNEE KIT, REF CDS981323AA; \n39) TOTAL KNEE KIT, REF CDS981323Y; \n40) BI LATERAL KNEE KIT, REF CDS981324T; \n41) UPPER EXTREMITY CDS-LF, REF CDS981564G; \n42) UPPER EXTREMITY CDS-LF, REF CDS981564I; \n43) LAMI, REF CDS981886X; \n44) CRANIOTOMY CDS, REF CDS981888W; \n45) CRANIOTOMY CDS, REF CDS981888X; \n46) TOTAL KNEE CDS, REF CDS981926D; \n47) TOTAL KNEE PROCEDURE-LF, REF CDS982082F; \n48) BHMC SYSTEM HIP KIT, REF CDS982396N; \n49) BAPTIST NORTH TOTAL KNEE CDS, REF CDS982404M; \n50) BAP NORTH SHOULDER CDS, REF CDS982406M; \n51) BAP NORTH UNIV EXTREMITY CDS, REF CDS982408K; \n52) TOTAL KNEE CDS, REF CDS982823Q; \n53) LOWER EXTREMITY CDS-LF, REF CDS982927M; \n54) LUMBAR LAMINECTOMY CDS-LF, REF CDS982928O; \n55) TOTAL HIP REPLACEMENT CDS-LF, REF CDS982929K; \n56) TOTAL KNEE REPLACEMENT CDS-LF, REF CDS982931L; \n57) UPPER EXTREMITY-LF, REF CDS982932K; \n58) LOWER EXTREMITY CDS, REF CDS982943Q; \n59) LOWER EXTREMITY CDS, REF CDS982943R; \n60) LOWER EXTREMITY CDS, REF CDS982943S; \n61) TOTAL KNEE, REF CDS982986T; \n62) SPECIALIZED HIP CDS, REF CDS983030V; \n63) EXTREMITY CDS, REF CDS983031Q; \n64) ZALE SPINE CDS, REF CDS983182J; \n65) ZALE SPINE CDS, REF CDS983182K; \n66) ZALE CRANIOTOMY CDS, REF CDS983188K; \n67) DR GILL'S CDS, REF CDS983189G; \n68) ZALE TOTAL KNEE CDS, REF CDS983194I; \n69) ZALE TOTAL KNEE CDS, REF CDS983194J; \n70) ZALE TOTAL KNEE CDS, REF CDS983194K; \n71) ZALE TOTAL KNEE CDS, REF CDS983194L; \n72) ZALE TOTAL HIP CDS, REF CDS983195I; \n73) ZALE TOTAL HIP CDS, REF CDS983195J; \n74) ZALE TOTAL HIP CDS, REF CDS983195K; \n75) ZALE EXTREMITY CDS, REF CDS983199K; \n76) ZALE TOTAL SHOULDER CDS, REF CDS983203J; \n77) ZALE TOTAL SHOULDER CDS, REF CDS983203K; \n78) ZALE TOTAL SHOULDER CDS, REF CDS983203L; \n79) ORTHO EXTREMITY CDS, REF CDS983408G; \n80) ORTHO-CV CDS, REF CDS983415G; \n81) PODIATRY CDS, REF CDS983416D; \n82) PODIATRY CDS, REF CDS983416G; \n83) ORTHO EXTREMITY, REF CDS983468O; \n84) LUMBAR LAMINECTOMY NEURO CDS, REF CDS983476J; \n85) SITTING CERVICAL CDS, REF CDS983480G; \n86) ORTHO SPINE, REF CDS983482L; \n87) ARTHROSCOPY SHOULDER, REF CDS983485L; \n88) EXTREMITY-LF, REF CDS983487K; \n89) ACL-LF, REF CDS983490M; \n90) PODIATRY CDS, REF CDS983610L; \n91) TOTAL JOINT-LF, REF CDS983863F; \n92) UPPER EXTREMITY CDS, REF CDS983907K; \n93) UPPER EXTREMITY CDS, REF CDS983907L; \n94) LOWER EXTREMITY CDS, REF CDS983909O; \n95) LOWER EXTREMITY CDS, REF CDS983909P; \n96) GENERAL ORTHO CDS, REF CDS983915L; \n97) GENERAL ORTHO CDS, REF CDS983915M; \n98) MAJOR ORTHO, REF CDS984253L; \n99) LOWER EXTREMTIY, REF CDS984254K; \n100) LOWER EXTREMTIY, REF CDS98425",
      "code_info": "86OJF\n1) REF CDS780112J, UDI/DI 10195327009663 (each) 40195327009664 (case), Lot Numbers:  22FBF455, 23CBH389, 23FBO500; \n2) REF CDS780119V, UDI/DI 10195327114916 (each) 40195327114917 (case), Lot Numbers:  22KBN442, 22HBO015, 23ABF941, 23DBE182, 23BBQ326; \n3) REF CDS780147N, UDI/DI 10195327268770 (each) 40195327268771 (case), Lot Numbers:  23DBT165, 22JBU233, 23BBJ852; \n4) REF CDS780148Q, UDI/DI 10193489202939 (each) 40193489202930 (case), Lot Numbers:  22JBL294; \n5) REF CDS780148R, UDI/DI 10195327268787 (each) 40195327268788 (case), Lot Numbers:  23CBE921, 22LBL862, 23EBA185, 23EBU780; \n6) REF CDS920027W, UDI/DI 10193489946956 (each) 40193489946957 (case), Lot Numbers:  22GBL731; \n7) REF CDS920027X, UDI/DI 10195327241742 (each) 40195327241743 (case), Lot Numbers:  22JBF953; \n8) REF CDS920027Y, UDI/DI 10195327298531 (each) 40195327298532 (case), Lot Numbers:  23CBI608, 23CBF425; \n9) REF CDS940047AF, UDI/DI 10195327243890 (each) 40195327243891 (case), Lot Numbers:  23DDC114, 23FDA378, 23IDA521, 23HDB417; \n10) REF CDS940047AG, UDI/DI 10195327516710 (each) 40195327516711 (case), Lot Numbers:  23KDA158; \n11) REF CDS940048AF, UDI/DI 10195327150259 (each) 40195327150250 (case), Lot Numbers:  22EDB749; \n12) REF CDS940048AG, UDI/DI 10195327241896 (each) 40195327241897 (case), Lot Numbers:  22IDA258, 23CDB146, 23IDA944, 23HDA534; \n13) REF CDS940048AI, UDI/DI 10195327516703 (each) 40195327516704 (case), Lot Numbers:  23KDA886; \n14) REF CDS940072W, UDI/DI 10193489288544 (each) 40193489288545 (case), Lot Numbers:  22DBN494; \n15) REF CDS940072Y, UDI/DI 10195327300845 (each) 40195327300846 (case), Lot Numbers:  23EBG408, 23EBE750; \n16) REF CDS940823AB, UDI/DI 10193489883282 (each) 40193489883283 (case), Lot Numbers:  22FBP685, 22GBM982, 22FBR756, 23CBS044, 23ABJ584, 22KBN648, 22JBL399, 23EBG438; \n17) REF CDS940842Y, UDI/DI 10193489845129 (each) 40193489845120 (case), Lot Numbers:  23EBV624, 23DBT379; \n18) REF CDS940843AA, UDI/DI 10195327151409 (each) 40195327151400 (case), Lot Numbers:  22EBD486, 22KBW061, 23DBH163, 23CBV755, 23CBC381, 23EBN012; \n19) REF CDS940857K, UDI/DI 10195327247553 (each) 40195327247554 (case), Lot Numbers:  22JBD506, 23ABC389, 23FBM896, 23EBM731; \n20) REF CDS940857L, UDI/DI 10195327533311 (each) 40195327533312 (case), Lot Numbers:  23KBF380; \n21) REF CDS940860P, UDI/DI 10195327009403 (each) 40195327009404 (case), Lot Numbers:  22FBS102, 22GBM011, 23CBE976, 23CBU516, 23JBH990; \n22) REF CDS940860Q, UDI/DI 10195327533304 (each) 40195327533305 (case), Lot Numbers:  23KBX215; \n23) REF CDS940911Q, UDI/DI 10195327349073 (each) 40195327349074 (case), Lot Numbers: 23DBD900, 23GBJ502, 23IBB454, 23CBV672, 23GBU420, 23FBR727, 23JBM716, 23JBN033, 23IBV466;  \n24) REF CDS940963Y, UDI/DI 10193489883671 (each) 40193489883672 (case), Lot Numbers:  22EBN044, 22EBN226, 22GBX355, 22GBR212, 23CBO762, 23BBG535, 22KBE790,\n25) REF CDS941065L, UDI/DI 10193489445350 (each) 40193489445351 (case), Lot Numbers:  22KBD585, 23BBN941, 23EBL605;\n26) REF CDS941167M, UDI/DI 10195327010454 (each) 40195327010455 (case), Lot Numbers:  22EBM271, 22EBL380, 23ABF515, 23ABD332, 22KBF510, 22IBT122, 23CBU256, 23JBY809, 23JBN695, 23IBC605;\n27) REF CDS941167N, UDI/DI 10195327535735 (each) 40195327535736 (case), Lot Numbers:  23KBV806; \n28) REF CDS980666I, UDI/DI 10193489485233 (each) 40193489485234 (case), Lot Numbers:  22FME365, 22FMC206, 22JMA036, 23GMB747, 23HMH273, 23KME164,\n29) REF CDS980666J, UDI/DI 10195327568566 (each) 40195327568567 (case), Lot Numbers:  24AMF689, 24CME853, 24BMJ472, 24AMJ370, 24DMD132; \n30) REF CDS980788O, UDI/DI 10193489233285 (each) 40193489233286 (case), Lot Numbers:  22JLA593, 22LLA050, 23ELA147, 23BLA703; \n31) REF CDS980788P, UDI/DI 10195327384722 (each) 40195327384723 (case), Lot Numbers:  24ALA706, 23HLA176, 23GLA079, 23FLA091, 23ELA727, 23LLA549, 23LLB042, 24CLA125, 24DLA497; \n32) REF CDS981314R, UDI/DI 10195327080457 (each) 40195327080458 (case), Lot Numbers:  22GMF503, 22KMC777, 22FMF044, 22FMF045, 23HME905, 23GMC775; \n33) REF CDS981314S, UDI/DI 10195327539269 (each) 40195327539260 (case), Lot Numbers:  24AMD136; \n34) REF CDS981318T, UDI/DI 10195327080174 (each) 40195327080175 (case), Lot Numbers:  23AMD897, 23JMC381; \n35) REF CDS981318U, UDI/DI 10195327539245 (each) 40195327539246 (case), Lot Numbers:  23LMA922; \n36) REF CDS981320V, UDI/DI 10195327080105 (each) 40195327080106 (case), Lot Numbers:  22GMC259, 22FMI088, 23EMD881, 23EMB534, 23AMC198, 23AME227, 23GME223, 23HMB373, 23KME353; \n37) REF CDS981320W, UDI/DI 10195327539306 (each) 40195327539307 (case), Lot Numbers:  23LMG443; \n38) REF CDS981323AA, UDI/DI 10195327539313 (each) 40195327539314 (case), Lot Numbers:  24AMC457, 23LMD523; \n39) REF CDS981323Y, UDI/DI 10195327080112 (each) 40195327080113 (case), Lot Numbers:  23DME426, 23EMI771, 23JME028, 23JMG482; \n40) REF CDS981324T, UDI/DI 10195327080426 (each) 40195327080427 (case), Lot Numbers:  23CMC158, 23CME830; \n41) REF CDS981564G, UDI/DI 10193489485226 (each) 40193489485227 (case), Lot Numbers:  22HMA210, 22KMH608, 23AME249, 22LMF194, 23GMB938, 23FME185, 23IMH127, 23IMA748, 23KMI894; \n42) REF CDS981564I, UDI/DI 10195327568672 (each) 40195327568673 (case), Lot Numbers:  24BMJ063, 24EMC702, 24DMB943; \n43) REF CDS981886X, UDI/DI 10195327145002 (each) 40195327145003 (case), Lot Numbers:  22EBO274, 22JBE914, 22LBF066, 23ABJ993, 23EBS630, 23EBV032, 23JBA693, 23LBB198, 24BBE467, 24CBA388, 24DBK454; \n44) REF CDS981888W, UDI/DI 10195327055806 (each) 40195327055807 (case), Lot Numbers:  22DBK058; \n45) REF CDS981888X, UDI/DI 10195327144760 (each) 40195327144761 (case), Lot Numbers:  22JBX645, 23DBA432, 23HBT873, 23JBR359, 24DBI356, 24CBD698, 23JBG798, 23JBG925; \n46) REF CDS981926D, UDI/DI 10195327141172 (each) 40195327141173 (case), Lot Numbers:  22HBF223, 22GBW207, 22KBO460, 23CBC519, 23BBA923, 23HBP475, 23HBN990, 23EBU765, 23KBF411, 24BBB303, 24EBC197, 24CBK503, 24CBU581; \n47) REF CDS982082F, UDI/DI 10193489748765 (each) 40193489748766 (case), Lot Numbers:  22EDB581, 22LDB741, 23CDC093, 23FDB555, 24ADB287, 24DDB001; \n48) REF CDS982396N, UDI/DI 10193489985245 (each) 40193489985246 (case), Lot Numbers:  23EBJ391, 23ABH250, 23BBS724, 23GBG962, 23LBF687, 23KBF527, 24BBP516, 24ABE656, 24EBP535, 24EBA091; \n49) REF CDS982404M, UDI/DI 10195327010362 (each) 40195327010363 (case), Lot Numbers:  22LBL557, 22JBY563, 23GBG866, 23EBH327, 23JBR515, 23IBG369, 23KBM706, 24ABA228, 24CBA071, 24CBA243, 24DBD789, 24EBF079; \n50) REF CDS982406M, UDI/DI 10195327010379 (each) 40195327010370 (case), Lot Numbers:  23ABQ862, 23ABP846, 23EBR980, 23HBX924, 23IBR566, 23IBQ656, 23KBS511, 24BBS092, 24ABV912, 24DBI838; \n51) REF CDS982408K, UDI/DI 10195327029159 (each) 40195327029150 (case), Lot Numbers:  22GBU424, 22FBI473, 22FBG543, 23ABC232, 22LBJ951, 22KBU540, 22KBA291, 23HBH304, 23DBD824, 23IBR319, 23EBS632, 24ABV379, 23LBK663, 23KBL500, 24CBB696, 24DBI834, 24EBO787; \n52) REF CDS982823Q, UDI/DI 10195327238339 (each) 40195327238330 (case), Lot Numbers:  23ABT038, 22KBT004, 23BBK666; \n53) REF CDS982927M, UDI/DI 10193489879155 (each) 40193489879156 (case), Lot Numbers:  22FBX307, 22GBU507, 22JBQ096; \n54) REF CDS982928O, UDI/DI 10193489879162 (each) 40193489879163 (case), Lot Numbers:  22GBF591, 22JBI126; \n55) REF CDS982929K, UDI/DI 10193489879179 (each) 40193489879170 (case), Lot Numbers:  22GBT440, 22JBH490; \n56) REF CDS982931L, UDI/DI 10193489879377 (each) 40193489879378 (case), Lot Numbers:  22GBU520, 22FBG546, 22FBG547, 22GBF485, 22JBH489; \n57) REF CDS982932K, UDI/DI 10193489879384 (each) 40193489879385 (case), Lot Numbers:  22GBD664, 22FBU390, 22FBG515; \n58) REF CDS982943Q, UDI/DI 10193489363258 (each) 40193489363259 (case), Lot Numbers:  22FBN424; \n59) REF CDS982943R, UDI/DI 10195327185268 (each) 40195327185269 (case), Lot Numbers:  23DBR765, 23EBQ503; \n60) REF CDS982943S, UDI/DI 10195327432393 (each) 40195327432394 (case), Lot Numbers:  23GBJ647; \n61) REF CDS982986T, UDI/DI 10195327175283 (each) 40195327175284 (case), Lot Numbers:  23DBA273, 23ABN818, 23ABJ008, 23FBF632, 23EBG702; \n62) REF CDS983030V, UDI/DI 10195327151416 (each) 40195327151417 (case), Lot Numbers:  22GBT329, 22HBL176, 22KBS029; \n63) REF CDS983031Q, UDI/DI 10193489848359 (each) 40193489848350 (case), Lot Numbers:  22HBS390, 22JBQ016, 23ABC508, 23EBK971; \n64) REF CDS983182J, UDI/DI 10195327147266 (each) 40195327147267 (case), Lot Numbers:  22HBO659, 22GBK869, 22GBK872, 22IBE783; \n65) REF CDS983182K, UDI/DI 10195327297503 (each) 40195327297504 (case), Lot Numbers:  23ABB680, 23ABB679, 23DBL294, 23DBL131, 23KBM508, 23JBO877; \n66) REF CDS983188K, UDI/DI 10195327105600 (each) 40195327105601 (case), Lot Numbers:  22LBP537, 22JBE497, 22GBZ339, 22GBK865, 23DBF027, 23BBJ482, 22LBS194, 23HBJ188, 23DBG097, 23FBA995, 23IBE032; \n67) REF CDS983189G, UDI/DI 10195327164485 (each) 40195327164486 (case), Lot Numbers:  23ABR014, 23JBJ120, 23JBF561, 23HBF458, 23LBQ693; \n68) REF CDS983194I, UDI/DI 10193489368017 (each) 40193489368018 (case), Lot Numbers:  22EBG718; \n69) REF CDS983194J, UDI/DI 10195327164614 (each) 40195327164615 (case), Lot Numbers:  22FBI875, 22GBO380; \n70) REF CDS983194K, UDI/DI 10195327260828 (each) 40195327260829 (case), Lot Numbers:  22LBM727; \n71) REF CDS983194L, UDI/DI 10195327379490 (each) 40195327379491 (case), Lot Numbers:  23GBF702; \n72) REF CDS983195I, UDI/DI 10193489368024 (each) 40193489368025 (case), Lot Numbers:  22EBI460; \n73) REF CDS983195J, UDI/DI 10195327164577 (each) 40195327164578 (case), Lot Numbers:  22GBZ464, 22FBV949, 22GBO379; \n74) REF CDS983195K, UDI/DI 10195327297466 (each) 40195327297467 (case), Lot Numbers:  22LBL260; \n75) REF CDS983199K, UDI/DI 10195327490140 (each) 40195327490141 (case), Lot Numbers:  23KBF439; \n76) REF CDS983203J, UDI/DI 10195327164584 (each) 40195327164585 (case), Lot Numbers:  22GBZ347, 22HBA033, 22GBV925; \n77) REF CDS983203K, UDI/DI 10195327260835 (each) 40195327260836 (case), Lot Numbers:  23BBG462, 23DBF658; \n78) REF CDS983203L, UDI/DI 10195327379575 (each) 40195327379576 (case), Lot Numbers:  23JBO016, 23JBF453; \n79) REF CDS983408G, UDI/DI 10195327150266 (each) 40195327150267 (case), Lot Numbers:  23EDA279, 23CDC647, 23JDC508, 23FDB193; \n80) REF CDS983415G, UDI/DI 10195327150181 (each) 40195327150182 (case), Lot Numbers:  22KDC167, 23JDB387, 23JDA841, 23IDA942; \n81) REF CDS983416D, UDI/DI 10195327150198 (each) 40195327150199 (case), Lot Numbers:  22GDA312; \n82) REF CDS983416G, UDI/DI 10195327279783 (each) 40195327279784 (case), Lot Numbers:  23CDA511; \n83) REF CDS983468O, UDI/DI 10195327560959 (each) 40195327560950 (case), Lot Numbers:  24DDA929, 24ADA428, 24EDA794; \n84) REF CDS983476J, UDI/DI 10195327497262 (each) 40195327497263 (case), Lot Numbers:  24BDA356, 24ADA401, 23JDA222, 23KDC060, 24DDA670; \n85) REF CDS983480G, UDI/DI 10195327497842 (each) 40195327497843 (case), Lot Numbers:  24ADA524, 24CDB716; \n86) REF CDS983482L, UDI/DI 10195327497491 (each) 40195327497492 (case), Lot Numbers:  24DBD083, 24BBG190, 24DBD506, 23LBO753; \n87) REF CDS983485L, UDI/DI 10195327493073 (each) 40195327493074 (case), Lot Numbers:  23JDB449, 23IDB676, 24ADC157, 23LDB088, 23LDA537, 24CDA450; \n88) REF CDS983487K, UDI/DI 10195327493080 (each) 40195327493081 (case), Lot Numbers:  24DDC105, 24ADA398; \n89) REF CDS983490M, UDI/DI 10195327493110 (each) 40195327493111 (case), Lot Numbers:  23KDA049, 23IDB775, 24BDA104; \n90) REF CDS983610L, UDI/DI 10195327210151 (each) 40195327210152 (case), Lot Numbers:  23HBC818, 23DBE011, 22JBP692, 22KBL228, 23ABS787, 23ABK600; \n91) REF CDS983863F, UDI/DI 10195327457600 (each) 40195327457601 (case), Lot Numbers:  23JDC333; \n92) REF CDS983907K, UDI/DI 10193489402766 (each) 40193489402767 (case), Lot Numbers:  22GBG415, 22JBQ095; \n93) REF CDS983907L, UDI/DI 10195327238346 (each) 40195327238347 (case), Lot Numbers:  23DBN421, 23FBA959, 23FBA960, 23ABJ737, 23ABN875, 23ABR460, 23GBV529, 23FBU350; \n94) REF CDS983909O, UDI/DI 10193489402780 (each) 40193489402781 (case), Lot Numbers:  22GBG991, 22GBG385; \n95) REF CDS983909P, UDI/DI 10195327238384 (each) 40195327238385 (case), Lot Numbers:  22JBB183, 23BBS793, 22LBL297, 23FBG201, 23EBD620, 23DBD683, 23CBT465, 23GBF419; \n96) REF CDS983915L, UDI/DI 10193489402704 (each) 40193489402705 (case), Lot Numbers:  22DBL676, 22FBU344; \n97) REF CDS983915M, UDI/DI 10195327238353 (each) 40195327238354 (case), Lot Numbers:  22KBI139, 23BBG644, 22LBT311, 23ABT196, 23DBR773; \n98) REF CDS984253L, UDI/DI 10193489288513 (each) 40193489288514 (case), Lot Numbers:  22GBE919, 22DBE741, 22OBM146, 22JBR832, 23CBI600, 23DBD705; \n99) REF CDS984254K, UDI/DI 10193489289091 (each) 40193489289092 (case), Lot Numbers:  22GBO260, 22KBP500; \n100) REF CDS984254L, UDI/DI 10195327298159 (each) 40195327298150 (case), Lot Numbers:  23CBR791, 23EBJ573, 23ABK592, 23DBD812; \n101) REF CDS984279N, UDI/DI 10195327438784 (each) 40195327438785 (case), Lot Numbers:  23GLA089; \n102) REF CDS984281K, UDI/DI 10195327291525 (each) 40195327291526 (case), Lot Numbers:  23DMH792; \n103) REF CDS984346Q, UDI/DI 10195327404383 (each) 40195327404384 (case), Lot Numbers:  23FBI845; \n104) REF CDS984459N, UDI/DI 10195327029142 (each) 40195327029143 (case), Lot Numbers:  22IBJ488, 22GBV039, 23JBF444, 23CBE952, 23KBI538, 24ABN206, 23ABQ886, 23ABI080, 23ABI356, 23GBW460; \n105) REF CDS984462M, UDI/DI 10195327029135 (each) 40195327029136 (case), Lot Numbers:  23HBA114, 23FBM912, 23DBB265, 23ABS777, 23ABH406, 22HBQ525, 23FBC858, 22KBU534, 22KBU490, 23HBQ503, 24ABB913, 24GBJ266, 24BBU602, 23EBI540, 24DBI839, 23HBG580; \n106) REF CDS984586P, UDI/DI 10195327103378 (each) 40195327103379 (case), Lot Numbers:  22JBN244, 22GBN110, 22JBR373, 23ABL092; \n107) REF CDS984597F, UDI/DI 10193489486001 (each) 40193489486002 (case), Lot Numbers:  22KMG311, 22KMD511, 23CMF681, 22LMH126, 23EMD601, 23CMA025, 23DMH793, 23FMI884, 23JMG377, 23KMJ524; \n108) REF CDS984597G, UDI/DI 10195327569655 (each) 40195327569656 (case), Lot Numbers:  24DMF470, 24BMJ471, 24BMC871, 24AMH721, 24FMC508, 24DMF900; \n109) REF CDS984598G, UDI/DI 10193489486131 (each) 40193489486132 (case), Lot Numbers:  22EMH742, 22JMA043, 22FMG243, 22HME734, 23CMB375, 23CMB376, 23GMF053, 23FMC757, 23KMJ523, 23IMH761, 23IMC032; \n110) REF CDS984598I, UDI/DI 10195327569662 (each) 40195327569663 (case), Lot Numbers:  24AMD497, 24EMF652; \n111) REF CDS984639D, UDI/DI 10193489699722 (each) 40193489699723 (case), Lot Numbers:  22EBL309, 22FBN472, 23DBD903, 23ABR103, 23CBS010; \n112) REF CDS984641D, UDI/DI 10193489699739 (each) 40193489699730 (case), Lot Numbers:  22FBT396, 22FBD663, 22KBR850, 23DBN261; \n113) REF CDS984641F, UDI/DI 10195327435813 (each) 40195327435814 (case), Lot Numbers:  23GBD768, 23LBO224; \n114) REF CDS984694D, UDI/DI 10195327461621 (each) 40195327461622 (case), Lot Numbers:  23HMH236, 23HMF875, 23IMG658, 23KME902, 23JMG082, 23JMB329; \n115) REF CDS984896I, UDI/DI 10193489983920 (each) 40193489983921 (case), Lot Numbers:  23KBD732, 23KBO191, 24BBF369; \n116) REF CDS984896J, UDI/DI 10195327621032 (each) 40195327621033 (case), Lot Numbers:  24EBK430; \n117) REF CDS984898J, UDI/DI 10193489981261 (each) 40193489981262 (case), Lot Numbers:  22GBZ344; \n118) REF CDS984898K, UDI/DI 10195327265687 (each) 40195327265688 (case), Lot Numbers:  23KBD086, 23ABE795, 24ABH472, 24ABD440, 23HBU517; \n119) REF CDS984898L, UDI/DI 10195327621049 (each) 40195327621040 (case), Lot Numbers:  24CBD710; \n120) REF CDS985234P, UDI/DI 10195327128401 (each) 40195327128402 (case), Lot Numbers:  23ABS175; \n121) REF CDS985235N, UDI/DI 10195327103385 (each) 40195327103386 (case), Lot Numbers:  22GBI200; \n122) REF CDS985236O, UDI/DI 10195327103392 (each) 40195327103393 (case), Lot Numbers:  22IBI035, 22FBI859, 22FBK151, 22FBI858, 23ABM166; \n123) REF CDS985237L, UDI/DI 10195327173920 (each) 40195327173921 (case), Lot Numbers:  22KBP663, 23ABC074; \n124) REF CDS985279K, UDI/DI 10195327497460 (each) 40195327497461 (case), Lot Numbers:  23LDA420, 24CDA763; \n125) REF CDS985319I, UDI/DI 10195327091545 (each) 40195327091546 (case), Lot Numbers:  22EBJ454, 22GBF348, 22GBF350, 22EBK321, 22EBI070, 23ABB768, 22JBL431, 22JBH315, 23CBJ711; \n126) REF CDS985322O, UDI/DI 10195327091484 (each) 40195327091485 (case), Lot Numbers:  22HBB447, 22FBT914, 23EBE522, 23ABN848, 22LBQ609, 22LBF840, 22JBJ445, 23CBP708; \n127) REF CDS985345G, UDI/DI 10195327243586 (each) 40195327243587 (case), Lot Numbers:  22JDA453, 22LDA845, 23CDC650; \n128) REF CDS985345I, UDI/DI 10195327428754 (each) 40195327428755 (case), Lot Numbers:  23HDB317, 23JDC127, 23LDA893, 23KDA304, 24BDA211, 24EDA295; \n129) REF CDS985346I, UDI/DI 10195327243579 (each) 40195327243570 (case), Lot Numbers:  22LDB527, 23BDB709, 23CDB801, 23FDB094; \n130) REF CDS985346K, UDI/DI 10195327452230 (each) 40195327452231 (case), Lot Numbers:  23HDB810, 23KDB011, 24CDC028; \n131) REF CDS985347F, UDI/DI 10195327243616 (each) 40195327243617 (case), Lot Numbers:  22KDC138, 22JDB124, 23IDB172, 23KDA093, 23LDB766, 23KDB661, 23CDC499, 23DDB392, 23EDB196, 24EDA310; \n132) REF CDS985428L, UDI/DI 10195327470722 (each) 40195327470723 (case), Lot Numbers:  23HBU628, 23HBQ472; \n133) REF CDS985428M, UDI/DI 10195327527259 (each) 40195327527250 (case), Lot Numbers:  23KBE466, 24ABB045, 24BBK956, 24ABV395, 24ABS126, 24DBO092; \n134) REF CDS985432N, UDI/DI 10195327336127 (each) 40195327336128 (case), Lot Numbers:  23BBE587; \n135) REF CDS985432O, UDI/DI 10195327382025 (each) 40195327382026 (case), Lot Numbers:  23EBP900, 23EBE298; \n136) REF CDS985432Q, UDI/DI 10195327527198 (each) 40195327527199 (case), Lot Numbers:  23KBS506, 24GBJ137, 24CBN817, 24CBB250, 24DBD900; \n137) REF CDS985433M, UDI/DI 10195327336134 (each) 40195327336135 (case), Lot Numbers:  23BBL088, 23EBE899, 23EBC985; \n138) REF CDS985433N, UDI/DI 10195327398255 (each) 40195327398256 (case), Lot Numbers:  23FBU308, 23FBO225; \n139) REF CDS985433P, UDI/DI 10195327527204 (each) 40195327527205 (case), Lot Numbers:  24DBP406; \n140) REF CDS985443J, UDI/DI 10195327228101 (each) 40195327228102 (case), Lot Numbers:  22JBO938, 22JBL358, 22JBJ683, 22IBT393, 22IBJ799, 22JBU319, 23ABM747; \n141) REF CDSOPM834W, UDI/DI 10193489748758 (each) 40193489748759 (case), Lot Numbers:  22FDB552, 22JDB343, 23ADB631, 23CDB828, 23JDB638, 24ADA064; \n142) REF CDSOPP508AF, UDI/DI 10193489401783 (each) 40193489401784 (case), Lot Numbers:  22DBT540, 22FBN433; \n143) REF CDSOPP508AG, UDI/DI 10195327145149 (each) 40195327145140 (case), Lot Numbers:  22HBO323; \n144) REF DYNJ00284L, UDI/DI 10193489962970 (each) 40193489962971 (case), Lot Numbers:  22JMD793, 22HMC141, 22GMB675, 23BMB762; \n145) REF DYNJ00284M, UDI/DI 10195327412036 (each) 40195327412037 (case), Lot Numbers:  23FMB532; \n146) REF DYNJ0101051L, UDI/DI 10195327339067 (each) 40195327339068 (case), Lot Numbers:  23DDA067, 23IDA268, 23IDC014; \n147) REF DYNJ0101067N, UDI/DI 10195327104146 (each) 40195327104147 (case), Lot Numbers:  22EMG580, 22GMD049, 23BMA253, 23DMD611, 23AMG496, 23IME763, 23JMB621; \n148) REF DYNJ0101282F, UDI/DI 10195327288679 (each) 40195327288670 (case), Lot Numbers:  23ADB994, 23DDB459; \n149) REF DYNJ0101282G, UDI/DI 10195327390921 (each) 40195327390922 (case), Lot Numbers:  23IDA957, 23HDA132; \n150) REF DYNJ0101283F, UDI/DI 10195327288662 (each) 40195327288663 (case), Lot Numbers:  23DDB252; \n151) REF DYNJ0101283G, UDI/DI 10195327391393 (each) 40195327391394 (case), Lot Numbers:  23GDA746, 23JDA056; \n152) REF DYNJ0101287J, UDI/DI 10195327286484 (each) 40195327286485 (case), Lot Numbers:  23EDA927; \n153) REF DYNJ0101292F, UDI/DI 10193489434040 (each) 40193489434041 (case), Lot Numbers:  22EDB214, 22EDB043; \n154) REF DYNJ0101292G, UDI/DI 10195327286255 (each) 40195327286256 (case), Lot Numbers:  23JDA057, 23HDB829, 23DDA637; \n155) REF DYNJ0101293F, UDI/DI 10195327286514 (each) 40195327286515 (case), Lot Numbers:  23BDB632, 23EDA019, 23JDA719; \n156) REF DYNJ0101300G, UDI/DI 10195327288686 (each) 40195327288687 (case), Lot Numbers:  23BDB321, 23KDC177, 23HDC663, 24BDB589; \n157) REF DYNJ0101312K, UDI/DI 10195327333102 (each) 40195327333103 (case), Lot Numbers:  23HBP291, 23HBM321, 23GBJ788; \n158) REF DYNJ0101328K, UDI/DI 10195327295943 (each) 40195327295944 (case), Lot Numbers:  23AMA479; \n159) REF DYNJ0101328L, UDI/DI 10195327332907 (each) 40195327332908 (case), Lot Numbers:  23JBW902; \n160) REF DYNJ0101349M, UDI/DI 10195327295950 (each) 40195327295951 (case), Lot Numbers:  22LMD920; \n161) REF DYNJ0101349N, UDI/DI 10195327332914 (each) 40195327332915 (case), Lot Numbers:  23IBT537, 23HMB680; \n162) REF DYNJ0105943T, UDI/DI 10195327120672 (each) 40195327120673 (case), Lot Numbers:  22DMD765; \n163) REF DYNJ0106852G, UDI/DI 10889942854172 (each) 40889942854173 (case), Lot Numbers:  22KBO320, 23LBH415, 23GBF928; \n164) REF DYNJ0111383L, UDI/DI 10193489315714 (each) 40193489315715 (case), Lot Numbers:  23AMC168, 22EMC657, 23DME952, 23FMI326, 23HMD592; \n165) REF DYNJ0122428I, UDI/DI 10889942854196 (each) 40889942854197 (case), Lot Numbers:  24DBR554; \n166) REF DYNJ0160354C, UDI/DI 10195327320072 (each) 40195327320073 (case), Lot Numbers:  23BDA901; \n167) REF DYNJ0190795O, UDI/DI 10193489442618 (each) 40193489442619 (case), Lot Numbers:  23BMI341, 23GMH800, 23JMJ190, 23IMH749, 23FMF312, 23LMH163, 24FME100; \n168) REF DYNJ0190827M, UDI/DI 10193489427677 (each) 40193489427678 (case), Lot Numbers:  22EBJ731, 22GBR015, 23ABT829, 23ABS484, 23ABO254, 23JBV636, 23JBB839, 23IBE648, 23LBS788, 24DBA509; \n169) REF DYNJ01927K, UDI/DI 10195327491567 (each) 40195327491568 (case), Lot Numbers:  23JBQ414; \n170) REF DYNJ02464K, UDI/DI 10195327291129 (each) 40195327291120 (case), Lot Numbers:  22LMC143; \n171) REF DYNJ02465J, UDI/DI 10195327291136 (each) 40195327291137 (case), Lot Numbers:  23AMA990, 23KMI177, 23IMF419; \n172) REF DYNJ02466K, UDI/DI 10195327291037 (each) 40195327291038 (case), Lot Numbers:  23EMJ733, 23HMD136; \n173) REF DYNJ02468K, UDI/DI 10195327291044 (each) 40195327291045 (case), Lot Numbers:  23AMH491, 23HME303; \n174) REF DYNJ0282079T, UDI/DI 10195327088590 (each) 40195327088591 (case), Lot Numbers:  22GBT971, 22EBS078, 22KBL646, 23BBD836, 23EBV356, 23HBR885, 23FBR451, 23JBL864; \n175) REF DYNJ0340302I, UDI/DI 10195327104191 (each) 40195327104192 (case), Lot Numbers:  22CDC058, 22EDB821, 23EDB166, 23FDA825; \n176) REF DYNJ0340302J, UDI/DI 10195327449971 (each) 40195327449972 (case), Lot Numbers:  23GDA731; \n177) REF DYNJ0341231M, UDI/DI 10193489429060 (each) 40193489429061 (case), Lot Numbers:  22EMA406, 22GMH120, 22IMA199, 23EMK175, 23BMI342, 23FMF374, 23JMJ187, 23IMH944, 23KME561, 23LMG321, 24EMB392, 24BME816, 24CMB075, 24CMI977; \n178) REF DYNJ0366324O, UDI/DI 10195327510688 (each) 40195327510689 (case), Lot Numbers:  23JMH693; \n179) REF DYNJ0368519U, UDI/DI 10193489772326 (each) 40193489772327 (case), Lot Numbers:  22FBM508, 22OBC596, 22DBT321; \n180) REF DYNJ0368519V, UDI/DI 10195327214302 (each) 40195327214303 (case), Lot Numbers:  22HBK670, 22JBB941; \n181) REF DYNJ0368519W, UDI/DI 10195327268190 (each) 40195327268191 (case), Lot Numbers:  22LBG638, 22LBG641, 23DBB518, 23EBE443; \n182) REF DYNJ0368519X, UDI/DI 10195327418038 (each) 40195327418039 (case), Lot Numbers:  23FBJ313; \n183) REF DYNJ0378245N, UDI/DI 10195327500672 (each) 40195327500673 (case), Lot Numbers:  23IBP545, 23JBG896; \n184) REF DYNJ0378266O, UDI/DI 10195327500450 (each) 40195327500451 (case), Lot Numbers:  23JBO727; \n185) REF DYNJ0378293N, UDI/DI 10195327500689 (each) 40195327500680 (case), Lot Numbers:  23IBN337, 23JBV569; \n186) REF DYNJ0421428O, UDI/DI 10195327089344 (each) 40195327089345 (case), Lot Numbers:  22DMI806; \n187) REF DYNJ0421449X, UDI/DI 10195327085025 (each) 40195327085026 (case), Lot Numbers:  22GME045, 22FMB786, 22EMB558, 22GMH968, 23AME130, 23CMA591, 23FMJ595; \n188) REF DYNJ0421476T, UDI/DI 10195327083588 (each) 40195327083589 (case), Lot Numbers:  22GMH371, 22EMB325; \n189) REF DYNJ0421476U, UDI/DI 10195327249687 (each) 40195327249688 (case), Lot Numbers:  23GMD427, 23BME541, 22LMD501, 23DMB982; \n190) REF DYNJ0451527N, UDI/DI 10195327046040 (each) 40195327046041 (case), Lot Numbers: 22FBJ134, 22JBA082;  \n191) REF DYNJ0451527O, UDI/DI 10195327306083 (each) 40195327306084 (case), Lot Numbers:  23GBL779; \n192) REF DYNJ0451527P, UDI/DI 10195327504243 (each) 40195327504244 (case), Lot Numbers:  23IBI357; \n193) REF DYNJ0475125K, UDI/DI 10195327498313 (each) 40195327498314 (case), Lot Numbers:  23IBQ973; \n194) REF DYNJ0484186O, UDI/DI 10195327505103 (each) 40195327505104 (case), Lot Numbers:  23JBC704; \n195) REF DYNJ0487967R, UDI/DI 10889942854387 (each) 40889942854388 (case), Lot Numbers:  23LBA367, 23DBI922, 23ABN926, 23FBT419, 24CBU495; \n196) REF DYNJ0488567O, UDI/DI 10195327133986 (each) 40195327133987 (case), Lot Numbers:  22EBQ674; \n197) REF DYNJ0488567P, UDI/DI 10195327212926 (each) 40195327212927 (case), Lot Numbers:  23BBH294, 23CBG444, 23EBS558, 23HBP870, 23JBC892, 23KBF853; \n198) REF DYNJ0488588P, UDI/DI 10195327133979 (each) 40195327133970 (case), Lot Numbers:  22EBQ413, 22EBP051, 22EBL705, 22EBC278, 22DBR608; \n199) REF DYNJ05137F, UDI/DI 10193489298567 (each) 40193489298568 (case), Lot Numbers:  22GDA791, 22FDB899, 22HDA197; \n200) REF DYNJ05137G, UDI/DI 10195327224042 (each) 40195327224043 (case), Lot Numbers:  22IDC103, 22IDC163, 22KDB600, 23CDB185, 23ADB946, 23HDB483, 23FDC237; \n201) REF DYNJ0519273N, UDI/DI 10195327083472 (each) 40195327083473 (case), Lot Numbers: 22GMH141, 22EMI446, 22LMI802, 22JMG306, 23CMD008;  \n202) REF DYNJ0530895AO, UDI/DI 10193489782790 (each) 40193489782791 (case), Lot Numbers:  22CMA594, 22DME722, 22GME323, 22IMF976, 23BMD594, 23BMD312; \n203) REF DYNJ0532846S, UDI/DI 10195327356781 (each) 40195327356782 (case), Lot Numbers:  23DMA332; \n204) REF DYNJ0532852Q, UDI/DI 10195327356705 (each) 40195327356706 (case), Lot Numbers:  23DMA333, 23FMI190; \n205) REF DYNJ0536892U, UDI/DI 10195327206239 (each) 40195327206230 (case), Lot Numbers:  23BMC358, 22HMB123, 22KME303, 23CMF965, 23EMC256; \n206) REF DYNJ0536903AL, UDI/DI 10193489782783 (each) 40193489782784 (case), Lot Numbers:  22DBI585, 22DBM797, 23CBK347, 22DBH512, 22KBM157, 23DBP819, 23FBE647; \n207) REF DYNJ0537284AA, UDI/DI 10195327268084 (each) 40195327268085 (case), Lot Numbers:  23ABJ706, 23DBA241; \n208) REF DYNJ0537284AB, UDI/DI 10195327418151 (each) 40195327418152 (case), Lot Numbers:  23EBX477; \n209) REF DYNJ0537284X, UDI/DI 10193489681062 (each) 40193489681063 (case), Lot Numbers:  22DBG361, 22DBH399, 22GBD136; \n210) REF DYNJ0537284Y, UDI/DI 10195327214111 (each) 40195327214112 (case), Lot Numbers:  22JBB218, 22JBB808; \n211) REF DYNJ0587115Q, UDI/DI 10889942796847 (each) 40889942796848 (case), Lot Numbers:  22CMG299, 22FMF669, 22KMH033, 23BMC363, 23BMC362; \n212) REF DYNJ0618354V, UDI/DI 10889942860760 (each) 40889942860761 (case), Lot Numbers:  23BMD622; \n213) REF DYNJ0646272S, UDI/DI 10889942860920 (each) 40889942860921 (case), Lot Numbers:  22JMG686, 22LMG294; \n214) REF DYNJ0774643J, UDI/DI 10195327045982 (each) 40195327045983 (case), Lot Numbers:  22HBJ605, 22EBF304, 22EBF853, 23HBK421; \n215) REF DYNJ0826358T, UDI/DI 10193489384314 (each) 40193489384315 (case), Lot Numbers:  23CBV754; \n216) REF DYNJ0826379O, UDI/DI 10193489300932 (each) 40193489300933 (case), Lot Numbers:  22EMB415; \n217) REF DYNJ0826390X, UDI/DI 10193489265668 (each) 40193489265669 (case), Lot Numbers:  23BMB094, 23BMD322, 22GMC365, 22EMB564, 22CMI377, 23IMH180, 23EMK004; \n218) REF DYNJ0826417T, UDI/DI 10193489265675 (each) 40193489265676 (case), Lot Numbers:  22FMH746, 23BMC016, 22JMF901, 23CMD855, 23FMB448, 23IME089, 23FMI373, 24AMB092; \n219) REF DYNJ0826438T, UDI/DI 10193489265651 (each) 40193489265652 (case), Lot Numbers:  22JMF791, 22DMH603, 22FMH747, 22LMF278, 23DMB109, 23AMG343, 23KMD955, 23IMH965; \n220) REF DYNJ0826492Q, UDI/DI 10193489300949 (each) 40193489300940 (case), Lot Numbers:  22DMF131, 23DMH876, 22FMH748, 23HMG639, 23IMH966, 23JMH219; \n221) REF DYNJ0842910G, UDI/DI 10195327286637 (each) 40195327286638 (case), Lot Numbers:  23EDB905, 23FDA509; \n222) REF DYNJ0843009F, UDI/DI 10195327288709 (each) 40195327288700 (case), Lot Numbers:  23JDA475; \n223) REF DYNJ0843185G, UDI/DI 10195327288631 (each) 40195327288632 (case), Lot Numbers:  23GDA827; \n224) REF DYNJ0843201I, UDI/DI 10195327286545 (each) 40195327286546 (case), Lot Numbers:  23DDA017, 23DDA929, 23BDB798, 23JDA011, 23GDA968; \n225) REF DYNJ0843339M, UDI/DI 10193489964615 (each) 40193489964616 (case), Lot Numbers:  22EDB876, 22KDB701, 22DDA738; \n226) REF DYNJ0843339N, UDI/DI 10195327286552 (each) 40195327286553 (case), Lot Numbers:  23ADA251, 23JDA990, 23IDB329, 23LDA449; \n227) REF DYNJ0843648L, UDI/DI 10195327342982 (each) 40195327342983 (case), Lot Numbers:  23EDB923; \n228) REF DYNJ0843648M, UDI/DI 10195327491512 (each) 40195327491513 (case), Lot Numbers:  23IDA183; \n229) REF DYNJ0843840I, UDI/DI 10195327156534 (each) 40195327156535 (case), Lot Numbers:  22EDA885, 22FDA471, 22GDB104, 23ADB822; \n230) REF DYNJ0843840J, UDI/DI 10195327343262 (each) 40195327343263 (case), Lot Numbers:  23FDC326, 23DDA006; \n231) REF DYNJ0843840K, UDI/DI 10195327491536 (each) 40195327491537 (case), Lot Numbers:  23IDC068; \n232) REF DYNJ0867555K, UDI/DI 10195327026004 (each) 40195327026005 (case), Lot Numbers:  22EMD876; \n233) REF DYNJ0885263F, UDI/DI 10195327219635 (each) 40195327219636 (case), Lot Numbers:  23BMH237, 23BMB097, 22IMC316, 23FMG984; \n234) REF DYNJ0949139J, UDI/DI 10195327500443 (each) 40195327500444 (case), Lot Numbers:  23IBR863; \n235) REF DYNJ0952375N, UDI/DI 10195327156565 (each) 40195327156566 (case), Lot Numbers:  22EDB511, 22GDA482; \n236) REF DYNJ0952375O, UDI/DI 10195327343071 (each) 40195327343072 (case), Lot Numbers:  23EDC229, 23GDC003; \n237) REF DYNJ09693N, UDI/DI 10195327349196 (each) 40195327349197 (case), Lot Numbers:  23EBQ825, 23CBA158, 23GBK653, 23IBQ939, 23KBG590; \n238) REF DYNJ16154G, UDI/DI 10193489486124 (each) 40193489486125 (case), Lot Numbers:  22KMB030, 23IMH968, 23AMG897, 23AMG917, 23FME243; \n239) REF DYNJ17330M, UDI/DI 10195327227104 (each) 40195327227105 (case), Lot Numbers:  23FDA066, 23HDB271; \n240) REF DYNJ17332M, UDI/DI 10195327227128 (each) 40195327227129 (case), Lot Numbers:  23FDC214, 23HDA042, 23JDA194; \n241) REF DYNJ17451P, UDI/DI 10195327491598 (each) 40195327491599 (case), Lot Numbers:  23IBT526, 23KBH396; \n242) REF DYNJ21826R, UDI/DI 10195327104412 (each) 40195327104413 (case), Lot Numbers:  22JBB065, 22JBG214, 23BBE709, 23GBA464; \n243) REF DYNJ24631O, UDI/DI 10195327059385 (each) 40195327059386 (case), Lot Numbers:  23BBC196; \n244) REF DYNJ24636I, UDI/DI 10193489982145 (each) 40193489982146 (case), Lot Numbers:  22FBF313, 23CBT957, 23HBP876, 23KBO981, 23JBJ725, 24DBV407; \n245) REF DYNJ24638I, UDI/DI 10193489989953 (each) 40193489989954 (case), Lot Numbers:  23ABJ283, 22JBQ318, 23HBI011, 23DBB295, 23FBL170, 23JBJ640, 24DBV525, 24DBI498; \n246) REF DYNJ24639J, UDI/DI 10193489989960 (each) 40193489989961 (case), Lot Numbers:  23EBI295, 23DBB296, 23JBJ726, 24BBK339; \n247) REF DYNJ24715L, UDI/DI 10195327165772 (each) 40195327165773 (case), Lot Numbers:  22HMC683, 23GMI336, 24BMA798, 24CMI024, 23LMD460; \n248) REF DYNJ26425L, UDI/DI 10195327156985 (each) 40195327156986 (case), Lot Numbers:  23ADA252, 23CDA966, 24BDA884, 23IDB380, 23JDB892, 24BDA961; \n249) REF DYNJ28262K, UDI/DI 10195327157012 (each) 40195327157013 (case), Lot Numbers:  22GMD387, 23FMA009, 23AMA810, 23HMC742, 24CMI248; \n250) REF DYNJ30224M, UDI/DI 10195327155032 (each) 40195327155033 (case), Lot Numbers:  22EBK924, 22JBH673, 22KBI794, 22LBN714, 23EBU149; \n251) REF DYNJ30227K, UDI/DI 10195327155018 (each) 40195327155019 (case), Lot Numbers:  23BBL575, 22LBN755, 22KBH659, 23HBN310; \n252) REF DYNJ30802B, UDI/DI 10193489917932 (each) 40193489917933 (case), Lot Numbers:  23DBD087; \n253) REF DYNJ31296I, UDI/DI 10195327337681 (each) 40195327337682 (case), Lot Numbers:  23GBP372, 23KBJ324, 23HBW677, 23JBI597, 24BBS906; \n254) REF DYNJ31297I, UDI/DI 10195327011932 (each) 40195327011933 (case), Lot Numbers:  22LBN785; \n255) REF DYNJ31297J, UDI/DI 10195327337674 (each) 40195327337675 (case), Lot Numbers:  23HBT707, 23JBS439, 24DBH633; \n256) REF DYNJ31382D, UDI/DI 10195327141189 (each) 40195327141180 (case), Lot Numbers:  22GMF907, 22FMF594, 23EMF680, 23KBE540, 23JBL300, 23FMG919, 24CBF002; \n257) REF DYNJ32629K, UDI/DI 10195327527037 (each) 40195327527038 (case), Lot Numbers:  23KMF693; \n258) REF DYNJ33594G, UDI/DI 10195327469573 (each) 40195327469574 (case), Lot Numbers:  23IDC076; \n259) REF DYNJ33671G, UDI/DI 10195327469603 (each) 40195327469604 (case), Lot Numbers:  23JDB275; \n260) REF DYNJ34420L, UDI/DI 10193489969290 (each) 40193489969291 (case), Lot Numbers:  22FBP867, 22CBX861, 22DBE974; \n261) REF DYNJ34420M, UDI/DI 10195327213787 (each) 40195327213788 (case), Lot Numbers:  23ABJ177, 22KBS965, 22IBJ525, 22GBT442, 23CBC266, 23DBC094; \n262) REF DYNJ34420N, UDI/DI 10195327384388 (each) 40195327384389 (case), Lot Numbers:  23EBE625; \n263) REF DYNJ34428N, UDI/DI 10193489969306 (each) 40193489969307 (case), Lot Numbers:  22KMA354, 22FME407, 22DMG673, 22CMG297, 23CMD488, 23AMI797; \n264) REF DYNJ34428O, UDI/DI 10195327384333 (each) 40195327384334 (case), Lot Numbers:  23FMB966; \n265) REF DYNJ35432G, UDI/DI 10195327469610 (each) 40195327469611 (case), Lot Numbers:  23GDB898; \n266) REF DYNJ35546I, UDI/DI 10195327100346 (each) 40195327100347 (case), Lot Numbers:  22GMD907, 22FMF899, 23IMG957, 23IMB627, 23JMB935; \n267) REF DYNJ35835F, UDI/DI 10193489970593 (each) 40193489970594 (case), Lot Numbers:  22LMC292, 22IMA391, 23AMD112, 23DME157, 23BMI107; \n268) REF DYNJ35835G, UDI/DI 10195327412005 (each) 40195327412006 (case), Lot Numbers: 23FMA390;  \n269) REF DYNJ37617I, UDI/DI 10195327497484 (each) 40195327497485 (case), Lot Numbers:  24ADA010, 23JDC007, 23IDB240; \n270) REF DYNJ38234F, UDI/DI 10195327227920 (each) 40195327227921 (case), Lot Numbers:  22KBS839, 22KBI198, 22JBF056; \n271) REF DYNJ39434J, UDI/DI 10195327360139 (each) 40195327360130 (case), Lot Numbers:  23GMH694, 23JMB538, 23CMG472 23FMG021; \n272) REF DYNJ39434K, UDI/DI 10195327526658 (each) 40195327526659 (case), Lot Numbers:  24EMB145, 24AME049; \n273) REF DYNJ39437J, UDI/DI 10195327360221 (each) 40195327360222 (case), Lot Numbers:  23CMG203, 23GMG246, 23EMK016, 23HMD212, 23IMF367; \n274) REF DYNJ39437K, UDI/DI 10195327526603 (each) 40195327526604 (case), Lot Numbers:  23KMI284, 24BMD691, 24AMD988; \n275) REF DYNJ39438I, UDI/DI 10195327360153 (each) 40195327360154 (case), Lot Numbers:  23JMB446, 23CME727, 23EMA386, 23HMC487, 23IMD476, 23EMK017; \n276) REF DYNJ39438J, UDI/DI 10195327526627 (each) 40195327526628 (case), Lot Numbers:  24EME404, 23KMG648, 24BMD113, 24CMA721; \n277) REF DYNJ40330J, UDI/DI 10195327109141 (each) 40195327109142 (case), Lot Numbers:  22FBQ740, 22FBB791, 22EBB074, 22HBG874, 23CBV711; \n278) REF DYNJ40605D, UDI/DI 10193489699746 (each) 40193489699747 (case), Lot Numbers:  22JBG481, 23CBH487, 23EBT505; \n279) REF DYNJ40605F, UDI/DI 10195327435837 (each) 40195327435838 (case), Lot Numbers:  23JBR578; \n280) REF DYNJ40606D, UDI/DI 10193489699647 (each) 40193489699648 (case), Lot Numbers:  22DBS696, 22FBD687, 23CBU826, 23EBS919; \n281) REF DYNJ40611F, UDI/DI 10195327431624 (each) 40195327431625 (case), Lot Numbers:  23IBU427; \n282) REF DYNJ40982D, UDI/DI 10889942564002 (each) 40889942564003 (case), Lot Numbers:  22KBP840, 22OBJ515, 23ABM257, 22IBC523, 22FBM880, 22JBO605, 23CBC833, 23EBE098; \n283) REF DYNJ40995B, UDI/DI 10889942431236 (each) 40889942431237 (case), Lot Numbers:  22DBQ804, 22DBP068, 22FBC847, 22KBS967, 22LBN471, 23DBM132, 23CBX018; \n284) REF DYNJ41030K, UDI/DI 10195327156763 (each) 40195327156764 (case), Lot Numbers:  22IMH451, 22HME079, 22IMG229, 22KMF345, 22GMA588, 22JMH238, 23EMA396, 23CMG600, 23HME100, 23KMD613, 23JMB121, 23IMH192, 24FMI644, 24BME267, 23FME845, 24DMF361; \n285) REF DYNJ41031J, UDI/DI 10195327156770 (each) 40195327156771 (case), Lot Numbers:  22IMG230, 22IMA227, 22JMA832, 22KMF346, 22GMI022, 23CME629, 22JMG797, 23IMF440, 23FMF122, 23EMI560, 23IMH981, 24AME178A, 24BMA779, 24AME178, 24BMA020, 24BMI693; \n286) REF DYNJ41416G, UDI/DI 10195327220976 (each) 40195327220977 (case), Lot Numbers:  22LBF253, 22LBD615, 23CBF856, 23BBB377, 23DBR926; \n287) REF DYNJ42734B, UDI/DI 10193489877311 (each) 40193489877312 (case), Lot Numbers:  22KMA648, 22IME750; \n288) REF DYNJ42793D, UDI/DI 10193489427547 (each) 40193489427548 (case), Lot Numbers:  22DMH374, 22CMF261, 23BMD338, 22IMI373, 23AMF364, 23EMB857; \n289) REF DYNJ43142D, UDI/DI 10195327431617 (each) 40195327431618 (case), Lot Numbers:  23IBM984, 23FBL003; \n290) REF DYNJ43208G, UDI/DI 10193489953053 (each) 40193489953054 (case), Lot Numbers:  22EME372, 22KMH677, 23BMA133; \n291) REF DYNJ43215I, UDI/DI 10193489953084 (each) 40193489953085 (case), Lot Numbers:  22EMH454, 22GMB007, 23AMD774; \n292) REF DYNJ43216J, UDI/DI 10193489953091 (each) 40193489953092 (case), Lot Numbers:  22FMF377, 22GMG333, 22GMG494, 23AMJ532; \n293) REF DYNJ43218F, UDI/DI 10193489953107 (each) 40193489953108 (case), Lot Numbers:  23AMH390, 22LME703, 22IMD857, 23FMA260; \n294) REF DYNJ43220G, UDI/DI 10193489953114 (each) 40193489953115 (case), Lot Numbers:  22HMA07023CMF759; \n295) REF DYNJ43225G, UDI/DI 10193489958874 (each) 40193489958875 (case), Lot Numbers:  22FMH990, 22EMF784, 22JMI242, 23AME669, 22HMG986, 23CMB185, 23FMI388; \n296) REF DYNJ43226I, UDI/DI 10193489953039 (each) 40193489953030 (case), Lot Numbers:  22EMA090, 22GME084, 23AMI071, 22JMG344, 23CMB129, 23CMF288, 23EME963; \n297) REF DYNJ43768G, UDI/DI 10193489981797 (each) 40193489981798 (case), Lot Numbers:  22FBJ135, 22GBI865; \n298) REF DYNJ43768I, UDI/DI 10195327249175 (each) 40195327249176 (case), Lot Numbers:  22IDB663, 23JDB276, 23KDC250, 23HDB344, 24ADB738, 24CDA125; \n299) REF DYNJ43780F, UDI/DI 10193489990140 (each) 40193489990141 (case), Lot Numbers:  23ABL238, 23CBH008, 22JBT939, 23KBE894, 23JBJ733, 23DBP912, 23KBV860, 23HBX578, 24BBJ402, 24CBA721, 24DBQ766; \n300) REF DYNJ43781I, UDI/DI 10193489989687 (each) 40193489989688 (case), Lot Numbers:  22DDB534, 23GDB334, 23JDA572, 23HDC670, 23CDA576, 23ADB464, 23ADB505, 24BDA892, 23CDC640; \n301) REF DYNJ43785I, UDI/DI 10193489989908 (each) 40193489989909 (case), Lot Numbers:  22GBF626, 22JBA976, 22LBK921, 23ABV318, 22JBK138;\n302) REF DYNJ43785J, UDI/DI 10195327325732 (each) 40195327325733 (case), Lot Numbers:  23CBH105, 23CBW026, 23IBP045, 23GBH237, 23JBT551, 23HBP233, 24BBG167, 23LBE280, 24ABJ751, 24CBK842, 24DBV432; \n303) REF DYNJ43803B, UDI/DI 10195327243067 (each) 40195327243068 (case), Lot Numbers:  22IMF487, 23IMH444, 23LMC448, 23IME778, 23EMJ046; \n304) REF DYNJ43852F, UDI/DI 10195327200541 (each) 40195327200542 (case), Lot Numbers:  23CBU837, 22LBP910, 22LBK345, 23FBK894, 23EBQ613, 23IBT917, 23JBG171; \n305) REF DYNJ44001C, UDI/DI 10193489326406 (each) 40193489326407 (case), Lot Numbers:  22FBI040, 22KBL649, 23DBM945, 23JBW572A, 23KBE407, 23JBW572, 23LBL113, 24CBJ890; \n306) REF DYNJ44050F, UDI/DI 10193489989915 (each) 40193489989916 (case), Lot Numbers:  22IBV694, 22IBC526, 22KBJ970, 23ABG949, 23DBB309, 23DBM689, 23HBS968, 23GBB813, 23KBG864, 23FBS727, 24ABR380, 24BBK330; \n307) REF DYNJ44152I, UDI/DI 10195327412616 (each) 40195327412617 (case), Lot Numbers:  23IBU434, 23GBO640, 23FBL402, 23EBN414; \n308) REF DYNJ44166I, UDI/DI 10195327412999 (each) 40195327412990 (case), Lot Numbers:  23IBP289, 23HBR348; \n309) REF DYNJ44168G, UDI/DI 10195327412975 (each) 40195327412976 (case), Lot Numbers:  23JBG033, 23HBK346, 23KBK728; \n310) REF DYNJ44171I, UDI/DI 10195327413057 (each) 40195327413058 (case), Lot Numbers:  23IBW172, 23FBA570; \n311) REF DYNJ44411G, UDI/DI 10195327412678 (each) 40195327412679 (case), Lot Numbers:  23KBG483; \n312) REF DYNJ44681J, UDI/DI 10195327130886 (each) 40195327130887 (case), Lot Numbers:  22LDA584, 23BDA484; \n313) REF DYNJ44685L, UDI/DI 10195327127404 (each) 40195327127405 (case), Lot Numbers:  22EBU372, 23BBD629, 23EBA774; \n314) REF DYNJ44687L, UDI/DI 10195327127411 (each) 40195327127412 (case), Lot Numbers:  23BMC549, 23FMB860; \n315) REF DYNJ44688I, UDI/DI 10195327127374 (each) 40195327127375 (case), Lot Numbers:  22DBQ406, 22KBH121, 23BBO707, 23CBS467; \n316) REF DYNJ44692F, UDI/DI 10193489931099 (each) 40193489931090 (case), Lot Numbers:  22HDB559; \n317) REF DYNJ44692G, UDI/DI 10195327277574 (each) 40195327277575 (case), Lot Numbers:  22LDB435, 22KDA921, 23ADA839; \n318) REF DYNJ44765G, UDI/DI 10195327041595 (each) 40195327041596 (case), Lot Numbers:  22FBT048, 22LBK333, 23BBT899; \n319) REF DYNJ44849F, UDI/DI 10195327127572 (each) 40195327127573 (case), Lot Numbers:  22KDA717, 23CDA837; \n320) REF DYNJ44860I, UDI/DI 10195327127343 (each) 40195327127344 (case), Lot Numbers:  22KDB967, 22KDB783, 23DDA157; \n321) REF DYNJ44878D, UDI/DI 10195327127701 (each) 40195327127702 (case), Lot Numbers:  22IDA448, 22GDB722, 22FDA133, 22KDA718, 23ADB011, 23CDC301, 23BDA570; \n322) REF DYNJ44884F, UDI/DI 10195327127732 (each) 40195327127733 (case), Lot Numbers:  22FDA348, 22KDB140, 22IDA333, 22HDB242, 22GDB723; \n323) REF DYNJ44884G, UDI/DI 10195327298425 (each) 40195327298426 (case), Lot Numbers:  22LDC049, 23BDA921, 23DDA265; \n324) REF DYNJ44915M, UDI/DI 10195327298555 (each) 40195327298556 (case), Lot Numbers:  23BDB644; \n325) REF DYNJ44915O, UDI/DI 10195327406561 (each) 40195327406562 (case), Lot Numbers:  23EDC394, 23HDA181, 23IDC127; \n326) REF DYNJ44972N, UDI/DI 10195327381059 (each) 40195327381050 (case), Lot Numbers:  23HBG262, 23GBT305, 23EBP155, 23EBA093, 23JBY309, 23IBO662; \n327) REF DYNJ45332G, UDI/DI 10195327242534 (each) 40195327242535 (case), Lot Numbers:  22KMA935, 22IMF489, 23DMB597, 23BMI199, 23CME732, 23EMF684; \n328) REF DYNJ45579B, UDI/DI 10195327059170 (each) 40195327059171 (case), Lot Numbers:  22FBD961, 23DBJ428, 23DBB520, 22LBR094, 23BBD390, 23GBF698, 23EBQ227, 23EBQ405, 23JBQ936, 23LBB064, 24BBN755, 23IBA342, 24EBC381; \n329) REF DYNJ46360O, UDI/DI 10195327110215 (each) 40195327110216 (case), Lot Numbers:  23AMA827; \n330) REF DYNJ46380N, UDI/DI 10195327110222 (each) 40195327110223 (case), Lot Numbers:  22FMI359, 22EMF968, 22LMG147, 23DMC396; \n331) REF DYNJ46380O, UDI/DI 10195327441609 (each) 40195327441600 (case), Lot Numbers:  23GMD060, 23HME464, 23HMG224; \n332) REF DYNJ46676F, UDI/DI 10193489887617 (each) 40193489887618 (case), Lot Numbers:  23DDA493, 23FDA402, 22FDC031, 22JDC182, 22LDA136, 23CDB470, 23BDB476; \n333) REF DYNJ46689J, UDI/DI 10193489887600 (each) 40193489887601 (case), Lot Numbers:  22CDB871, 22DDC332, 22FDA846, 22GDC191; \n334) REF DYNJ46689K, UDI/DI 10195327224059 (each) 40195327224050 (case), Lot Numbers:  23CDC023, 23BDA845, 22JDB062, 22LDB236, 23EDB229, 23FDC371; \n335) REF DYNJ46698G, UDI/DI 10195327224066 (each) 40195327224067 (case), Lot Numbers:  22IDB648, 22KDA680, 23CDB321, 23BDA774, 22LDC233, 23DDA884; \n336) REF DYNJ47012O, UDI/DI 10195327072209 (each) 40195327072200 (case), Lot Numbers:  23CBW027, 23CBF864; \n337) REF DYNJ47176L, UDI/DI 10193489909289 (each) 40193489909280 (case), Lot Numbers:  22JBT414, 23JBK941; \n338) REF DYNJ47701I, UDI/DI 10195327357566 (each) 40195327357567 (case), Lot Numbers:  23DBK180, 23EBB152; \n339) REF DYNJ47752G, UDI/DI 10195327269968 (each) 40195327269969 (case), Lot Numbers:  23EMI287; \n340) REF DYNJ48084D, UDI/DI 10195327019419 (each) 40195327019410 (case), Lot Numbers:  22DMB406, 23AMJ103, 22LME957, 22LMI847, 22JMG004, 23AMH881; \n341) REF DYNJ48090F, UDI/DI 10195327019686 (each) 40195327019687 (case), Lot Numbers:  23CMC256, 22DMH446, 22CMH255, 22LMG790, 23AMG372, 23BMC408; \n342) REF DYNJ48091F, UDI/DI 10195327019693 (each) 40195327019694 (case), Lot Numbers:  22JMH937, 22IMB869, 23BMD482, 23FMA051; \n343) REF DYNJ48130J, UDI/DI 10195327209568 (each) 40195327209569 (case), Lot Numbers:  22KBR701; \n344) REF DYNJ48130K, UDI/DI 10195327349448 (each) 40195327349449 (case), Lot Numbers:  23CBE710; \n345) REF DYNJ48131J, UDI/DI 10193489848069 (each) 40193489848060 (case), Lot Numbers:  22EBR840, 22HBV293; \n346) REF DYNJ48259D, UDI/DI 10193489425338 (each) 40193489425339 (case), Lot Numbers:  22JMF417, 22LMD282, 23HMB682, 23EMF966, 23IMG826, 24BBA279; \n347) REF DYNJ49315G, UDI/DI 10195327495718 (each) 40195327495719 (case), Lot Numbers:  23IDC243, 23KDC461, 24CDA382; \n348) REF DYNJ49667K, UDI/DI 10195327059422 (each) 40195327059423 (case), Lot Numbers:  22FBX139, 22LBL766, 23CBH006, 23BBL426, 23CBU843, 22JBQ877; \n349) REF DYNJ50062M, UDI/DI 10195327059361 (each) 40195327059362 (case), Lot Numbers: 22HBC896, 23ABN437;  \n350) REF DYNJ50072F, UDI/DI 10193489340921 (each) 40193489340922 (case), Lot Numbers:  22KMC281, 23BMH293, 23CMC263, 23KMB691, 23EMB984, 23HMG455, 24AMI546; \n351) REF DYNJ50127C, UDI/DI 10889942916993 (each) 40889942916994 (case), Lot Numbers:  22DBD577; \n352) REF DYNJ50621K, UDI/DI 10195327002756 (each) 40195327002757 (case), Lot Numbers:  22CME958, 23BMG900, 22EMF160, 23AME835, 22HMB251, 22JMH260, 22JMG614, 22JMI402, 23BMG421, 23HMG764, 23FMG924; \n353) REF DYNJ50622J, UDI/DI 10193489922943 (each) 40193489922944 (case), Lot Numbers:  22GMF307, 22CMD580, 23DMI638, 22JMI258, 22IMC803, 22HMH756, 23DMA353, 23HMH918; \n354) REF DYNJ50665N, UDI/DI 10195327542733 (each) 40195327542734 (case), Lot Numbers: 23KBE170, 23LBI160, 24ABK459;  \n355) REF DYNJ50666M, UDI/DI 10195327600983 (each) 40195327600984 (case), Lot Numbers:  24BBN662, 24EBE931; \n356) REF DYNJ50884C, UDI/DI 10195327179519 (each) 40195327179510 (case), Lot Numbers:  23AMF855, 23CMC475, 23BMA394, 23EMJ710, 23EME118, 24DMD645; \n357) REF DYNJ50994I, UDI/DI 10195327088637 (each) 40195327088638 (case), Lot Numbers:  23ABR305, 22EBG897, 22HBE747, 22GBV082, 22KBD063, 23HBC086, 23JBB233, 23LBE846, 23FBR628; \n358) REF DYNJ51149R, UDI/DI 10195327127015 (each) 40195327127016 (case), Lot Numbers:  23CBA582; \n359) REF DYNJ51149S, UDI/DI 10195327510374 (each) 40195327510375 (case), Lot Numbers:  23KBF450; \n360) REF DYNJ51150O, UDI/DI 10195327127022 (each) 40195327127023 (case), Lot Numbers:  22FBY642, 22FBF024, 23ABL412, 23ABL413, 23GBL833, 23FBC091, 23HBQ398; \n361) REF DYNJ51150P, UDI/DI 10195327510404 (each) 40195327510405 (case), Lot Numbers:  23JBE656, 23JBW291; \n362) REF DYNJ51320L, UDI/DI 10193489841374 (each) 40193489841375 (case), Lot Numbers:  22FBW403, 22DBT710, 22JBO825, 23KBF216; \n363) REF DYNJ51802J, UDI/DI 10195327475079 (each) 40195327475070 (case), Lot Numbers:  23GDC154, 23IDA304; \n364) REF DYNJ51803I, UDI/DI 10195327475062 (each) 40195327475063 (case), Lot Numbers:  23GDC144, 23HDA736; \n365) REF DYNJ52007C, UDI/DI 10889942796878 (each) 40889942796879 (case), Lot Numbers:  22CMG176, 23EMA071, 22DMF070, 22JME038, 22JME069, 23CMF980, 22LMI347, 22LMI350, 23BMC588; \n366) REF DYNJ52331F, UDI/DI 10195327237691 (each) 40195327237692 (case), Lot Numbers:  23EBI999, 22LBE306, 22KBQ085, 22IBP287, 23DBK756, 23FBJ664; \n367) REF DYNJ52332G, UDI/DI 10195327237707 (each) 40195327237708 (case), Lot Numbers:  22IBE036, 22HBV289, 22JBQ833, 23CBH602; \n368) REF DYNJ52333D, UDI/DI 10195327237714 (each) 40195327237715 (case), Lot Numbers:  22HBV378, 22JBP714, 22JBP717, 22LBD328, 23BBH346, 23ABO549, 23CBA576; \n369) REF DYNJ52398N, UDI/DI 10195327339104 (each) 40195327339105 (case), Lot Numbers:  23FDA686, 23JDA047, 23HDA446; \n370) REF DYNJ52405I, UDI/DI 10195327222628 (each) 40195327222629 (case), Lot Numbers:  22JDC201; \n371) REF DYNJ52405J, UDI/DI 10195327339166 (each) 40195327339167 (case), Lot Numbers:  23GDB229, 23HDB575, 23DDA852, 23FDA445, 23JDA254; \n372) REF DYNJ52415L, UDI/DI 10195327339234 (each) 40195327339235 (case), Lot Numbers:  23GDA970, 23EDC509; \n373) REF DYNJ52418J, UDI/DI 10195327222697 (each) 40195327222698 (case), Lot Numbers:  22KDA708; \n374) REF DYNJ52418K, UDI/DI 10195327339258 (each) 40195327339259 (case), Lot Numbers:  23DDA689, 23IDC320, 23HDB839; \n375) REF DYNJ52420J, UDI/DI 10195327222710 (each) 40195327222711 (case), Lot Numbers:  22KDA303; \n376) REF DYNJ52420K, UDI/DI 10195327339272 (each) 40195327339273 (case), Lot Numbers:  23CDA077, 23CDC516, 23HDC167, 23JDA243; \n377) REF DYNJ52422F, UDI/DI 10195327222734 (each) 40195327222735 (case), Lot Numbers:  22JDC148; \n378) REF DYNJ52422G, UDI/DI 10195327339296 (each) 40195327339297 (case), Lot Numbers:  23HDA050; \n379) REF DYNJ52423K, UDI/DI 10195327339302 (each) 40195327339303 (case), Lot Numbers:  23GDA262, 23FDA281, 23IDC072; \n380) REF DYNJ52427J, UDI/DI 10195327339340 (each) 40195327339341 (case), Lot Numbers:  23DDA942, 23IDA396; \n381) REF DYNJ52430K, UDI/DI 10195327339593 (each) 40195327339594 (case), Lot Numbers:  23DDB533, 23GDA993, 23JDA099; \n382) REF DYNJ52431K, UDI/DI 10195327222802 (each) 40195327222803 (case), Lot Numbers:  22KDA296; \n383) REF DYNJ52431L, UDI/DI 10195327339609 (each) 40195327339600 (case), Lot Numbers:  23BDA248, 23DDA589, 23EDB819, 23GDA329, 24CDA929, 23IDA827, 23JDA949; \n384) REF DYNJ52436K, UDI/DI 10195327222826 (each) 40195327222827 (case), Lot Numbers:  22HDA788; \n385) REF DYNJ52436L, UDI/DI 10195327339647 (each) 40195327339648 (case), Lot Numbers:  23CDC601, 23GDC227; \n386) REF DYNJ52628B, UDI/DI 10889942872060 (each) 40889942872061 (case), Lot Numbers:  22JDA509, 22GDB458, 23EDA896, 23ADA462; \n387) REF DYNJ52638C, UDI/DI 10195327088644 (each) 40195327088645 (case), Lot Numbers:  22KBL650, 23HBF875, 23KBK761, 23KBB235, 23EBE717, 23EBV500, 23FBM239; \n388) REF DYNJ53006C, UDI/DI 10193489782745 (each) 40193489782746 (case), Lot Numbers:  22CMD559, 22DMF577, 23BMB746, 23AMJ124, 22HME683, 22GMH417, 22IMI055, 22IMI057; \n389) REF DYNJ53322G, UDI/DI 10195327360214 (each) 40195327360215 (case), Lot Numbers:  23CMH443, 23IME654, 23CME925, 23EMG576, 23FMG819; \n390) REF DYNJ53322I, UDI/DI 10195327526573 (each) 40195327526574 (case), Lot Numbers:  23KMD822, 23JME049, 24CMI637, 24BMI474, 24AME052; \n391) REF DYNJ53555F, UDI/DI 10195327410056 (each) 40195327410057 (case), Lot Numbers:  23EBI165, 23FBL406; \n392) REF DYNJ53652C, UDI/DI 10195327012137 (each) 40195327012138 (case), Lot Numbers:  22ELA688, 23ALA032, 22KLA300, 22JLA010; \n393) REF DYNJ53652D, UDI/DI 10195327337650 (each) 40195327337651 (case), Lot Numbers:  23CLA577, 23BLB064, 23BLA460, 23KLA826, 23DLA361, 23ELA054, 23JLB077, 24BLA326, 24DLA706, 24DLA103; \n394) REF DYNJ53682B, UDI/DI 10193489883084 (each) 40193489883085 (case), Lot Numbers:  23EDB818; \n395) REF DYNJ53687B, UDI/DI 10193489886511 (each) 40193489886512 (case), Lot Numbers:  22HDA058; \n396) REF DYNJ54149D, UDI/DI 10195327107284 (each) 40195327107285 (case), Lot Numbers:  22DMF213; \n397) REF DYNJ54149F, UDI/DI 10195327292614 (each) 40195327292615 (case), Lot Numbers:  22LMD587, 23JMA188, 23IMD234, 23KMI574; \n398) REF DYNJ54462C, UDI/DI 10195327075040 (each) 40195327075041 (case), Lot Numbers:  22EDB554, 23DDB397, 23JDA699, 23JDA083;\n399) REF DYNJ54463F, UDI/DI 10195327155704 (each) 40195327155705 (case), Lot Numbers:  23CME386, 22EMD530, 22DMI914, 23JMH897, 23HMF144; \n400) REF DYNJ54464G, UDI/DI 10195327155711 (each) 40195327155712 (case), Lot Numbers:  22FMB906, 22FMB839, 22EMA876, 23DMD917; \n401) REF DYNJ54934B, UDI/DI 10195327344399 (each) 40195327344390 (case), Lot Numbers:  23BDA904; \n402) REF DYNJ55049D, UDI/DI 10195327145705 (each) 40195327145706 (case), Lot Numbers:  22JMC226, 22EMD458; \n403) REF DYNJ55667D, UDI/DI 10195327281939 (each) 40195327281930 (case), Lot Numbers:  23BBP785, 23ABN999, 23CBL665, 23IBN555, 23HBQ079, 23JBY358, 23JBY654, 23JBD701; \n404) REF DYNJ55733G, UDI/DI 10193489454659 (each) 40193489454650 (case), Lot Numbers:  22JBU639, 23GBJ474, 23CBX532, 23BBE511, 22HBY185; \n405) REF DYNJ55851A, UDI/DI 10193489985030 (each) 40193489985031 (case), Lot Numbers:  22CMF878, 22JMA855, 22EMI149; \n406) REF DYNJ55852B, UDI/DI 10193489985016 (each) 40193489985017 (case), Lot Numbers:  22IMB760; \n407) REF DYNJ55852C, UDI/DI 10195327267575 (each) 40195327267576 (case), Lot Numbers:  23BMH431, 23EMI576; \n408) REF DYNJ56070D, UDI/DI 10193489773026 (each) 40193489773027 (case), Lot Numbers:  22EMG162, 22JMB598, 23AMA086, 23BMA752, 23KMD985, 23FMF245, 23EMI181, 23LBV444, 23IMH211, 24BBG378, 24DBA267; \n409) REF DYNJ56952C, UDI/DI 10195327485917 (each) 40195327485918 (case), Lot Numbers:  23JMH674, 23LME884, 24EMB237; \n410) REF DYNJ57113C, UDI/DI 10195327074630 (each) 40195327074631 (case), Lot Numbers:  22HMI210, 23IBU927, 23HBP819; \n411) REF DYNJ57546B, UDI/DI 10195327124663 (each) 40195327124664 (case), Lot Numbers:  22FMD947; \n412) REF DYNJ57546C, UDI/DI 10195327216580 (each) 40195327216581 (case), Lot Numbers:  22LMB359, 22KME770, 22HMG755, 23CMD289, 23BMC613; \n413) REF DYNJ57648D, UDI/DI 10195327027391 (each) 40195327027392 (case), Lot Numbers:  22EMA228, 22FMI369, 23HMG727, 23FMB585, 23AMH213, 22JMA245, 23FMJ099; \n414) REF DYNJ57841C, UDI/DI 10195327488581 (each) 40195327488582 (case), Lot Numbers:  23HBR517, 23IBL726; \n415) REF DYNJ57883C, UDI/DI 10195327046019 (each) 40195327046010 (case), Lot Numbers:  22DBS083; \n416) REF DYNJ57883D, UDI/DI 10195327204143 (each) 40195327204144 (case), Lot Numbers:  22HBE753, 24ABT515, 23BBM998, 23FBL672, 23CBJ367, 23JBW817, 23IBQ129; \n417) REF DYNJ57971B, UDI/DI 10195327049041 (each) 40195327049042 (case), Lot Numbers:  22GBS099; \n418) REF DYNJ57971C, UDI/DI 10195327225209 (each) 40195327225200 (case), Lot Numbers:  22KBM557, 22KBM558, 23ABF124, 23JBY165, 23LBD844, 23IBB085, 24ABW581, 24DBD421; \n419) REF DYNJ57975D, UDI/DI 10195327225124 (each) 40195327225125 (case), Lot Numbers:  23CMH203, 23AMA728, 22KMD412, 22JMB348, 23BMD721, 23FMC375, 23HMG228, 23JMG688, 23IMF776, 23EMC530, 24AMF292; \n420) REF DYNJ57975F, UDI/DI 10195327659431 (each) 40195327659432 (case), Lot Numbers:  24EMD224, 24CMH221, 24HMF645; \n421) REF DYNJ58053D, UDI/DI 10195327049058 (each) 40195327049059 (case), Lot Numbers:  22GMH676, 22EMH464; \n422) REF DYNJ58053F, UDI/DI 10195327225131 (each) 40195327225132 (case), Lot Numbers:  23GMI449, 23EMB074, 23FMH212, 24BMI409, 24CMB013; \n423) REF DYNJ58557C, UDI/DI 10193489972276 (each) 40193489972277 (case), Lot Numbers:  22JMF230, 23AME789, 23FMC763, 23FMI232; \n424) REF DYNJ58563D, UDI/DI 10193489972320 (each) 40193489972321 (case), Lot Numbers:  22CMA822, 22CMA823, 22KMH220, 23GMA044, 23FMH095, 22LMG325; \n425) REF DYNJ58564D, UDI/DI 10193489972337 (each) 40193489972338 (case), Lot Numbers:  22CMC588, 22KMG860, 23AME526, 23JMB659, 23IMD727, 23KMD387; \n426) REF DYNJ58631G, UDI/DI 10195327026585 (each) 40195327026586 (case), Lot Numbers:  22EDB868, 22FDB176, 23CDA513, 23HDA582, 23KDB279, 24ADB364; \n427) REF DYNJ58633D, UDI/DI 10195327026592 (each) 40195327026593 (case), Lot Numbers:  22FDB011, 22JDB863, 22KDB382, 23CDA345, 23CDC561, 23DDC160, 23JDB393, 23IDB456, 23IDA932, 23KDC183, 24BDA964; \n428) REF DYNJ58665F, UDI/DI 10195327529741 (each) 40195327529742 (case), Lot Numbers:  23KMB334; \n429) REF DYNJ59011G, UDI/DI 10195327332648 (each) 40195327332649 (case), Lot Numbers:  23IDC046, 23GDB612; \n430) REF DYNJ59025D, UDI/DI 10195327519636 (each) 40195327519637 (case), Lot Numbers:  23JBB388; \n431) REF DYNJ59080C, UDI/DI 10195327200299 (each) 40195327200290 (case), Lot Numbers:  22JBM549, 23BBI615, 22LBM723, 23DBB343, 23FBK899, 23HBL384, 23HBG733, 23JBN309; \n432) REF DYNJ59106C, UDI/DI 10195327519643 (each) 40195327519644 (case), Lot Numbers:  23JMB434; \n433) REF DYNJ59581D, UDI/DI 10195327165765 (each) 40195327165766 (case), Lot Numbers:  22JME822, 23DMI271, 23EMH407, 23EME554, 24AMG444, 24DMB376, 24BMD084, 24BMA718; \n434) REF DYNJ59819A, UDI/DI 10195327331597 (each) 40195327331598 (case), Lot Numbers:  23CBF431; \n435) REF DYNJ60049B, UDI/DI 10193489782776 (each) 40193489782777 (case), Lot Numbers:  22DBH397, 22JBY793, 22LBK547, 23CBX671, 23FBU291; \n436) REF DYNJ60243G, UDI/DI 10195327332884 (each) 40195327332885 (case), Lot Numbers:  23BMD070, 23IMC653; \n437) REF DYNJ60546B, UDI/DI 10195327361754 (each) 40195327361755 (case), Lot Numbers:  23CBX535; \n438) REF DYNJ60550B, UDI/DI 10195327361853 (each) 40195327361854 (case), Lot Numbers:  23CBQ320; \n439) REF DYNJ60553B, UDI/DI 10195327361822 (each) 40195327361823 (case), Lot Numbers:  23CBX537; \n440) REF DYNJ60626B, UDI/DI 10193489961904 (each) 40193489961905 (case), Lot Numbers:  22FBS949, 23ABR523, 23DBA353, 23IBS559; \n441) REF DYNJ60818B, UDI/DI 10195327454661 (each) 40195327454662 (case), Lot Numbers:  23HBJ977, 23GBS227, 23IBV407; \n442) REF DYNJ60818C, UDI/DI 10195327521325 (each) 40195327521326 (case), Lot Numbers:  23JBT400, 23KBJ738, 24ABA318; \n443) REF DYNJ60824B, UDI/DI 10195327521288 (each) 40195327521289 (case), Lot Numbers:  23JBT425; \n444) REF DYNJ60896B, UDI/DI 10195327500429 (each) 40195327500420 (case), Lot Numbers:  23IBS475, 23IBH842; \n445) REF DYNJ60906A, UDI/DI 10193489835960 (each) 40193489835961 (case), Lot Numbers:  22EMH358, 22EME827, 22CMB326, 22FMC970, 22HME390, 22KBS938, 22KBM814, 23EBC437, 23DBC688; \n446) REF DYNJ60906B, UDI/DI 10195327500351 (each) 40195327500352 (case), Lot Numbers:  23IBS476; \n447) REF DYNJ60921C, UDI/DI 10195327470395 (each) 40195327470396 (case), Lot Numbers:  23GBT246, 23KBG585, 23JBF899; \n448) REF DYNJ60930C, UDI/DI 10195327470340 (each) 40195327470341 (case), Lot Numbers:  23HBL078, 23JBL127; \n449) REF DYNJ61137F, UDI/DI 10195327470449 (each) 40195327470440 (case), Lot Numbers:  23HBL077, 23IBN862, 23JBV920; \n450) REF DYNJ61143F, UDI/DI 10195327470425 (each) 40195327470426 (case), Lot Numbers:  23HBQ168; \n451) REF DYNJ61146F, UDI/DI 10195327470418 (each) 40195327470419 (case), Lot Numbers:  23HBM016, 23JBC547; \n452) REF DYNJ61152D, UDI/DI 10195327470210 (each) 40195327470211 (case), Lot Numbers:  23GBT618, 23JBY125, 23JBE190; \n453) REF DYNJ61155B, UDI/DI 10195327080846 (each) 40195327080847 (case), Lot Numbers:  22EBK970, 23CBU121, 23FBC089, 23GBL975; \n454) REF DYNJ61155C, UDI/DI 10195327470487 (each) 40195327470488 (case), Lot Numbers:  23HBL129, 23KBA990; \n455) REF DYNJ61156C, UDI/DI 10195327470227 (each) 40195327470228 (case), Lot Numbers:  23JBL221; \n456) REF DYNJ61306A, UDI/DI 10195327072261 (each) 40195327072262 (case), Lot Numbers:  22EMA261, 23AMI588, 23EMB090, 23JBW540; \n457) REF DYNJ61340A, UDI/DI 10193489873801 (each) 40193489873802 (case), Lot Numbers:  22KBO231, 22HBY295, 22HBE528, 22JBO026, 23BBM091; \n458) REF DYNJ61414B, UDI/DI 10193489873900 (each) 40193489873901 (case), Lot Numbers:  22DBU513, 22HBX138, 22HBK088, 22GBA319, 22IBR399, 22LBP338, 22LBO290, 23CBE238, 23BBU334, 23DBC271; \n459) REF DYNJ61500, UDI/DI 10193489574098 (each) 40193489574099 (case), Lot Numbers:  22KME391, 22JMI283, 22IMH886, 22GMH683; \n460) REF DYNJ61582C, UDI/DI 10195327152161 (each) 40195327152162 (case), Lot Numbers:  22GME126; \n461) REF DYNJ61651I, UDI/DI 10195327327125 (each) 40195327327126 (case), Lot Numbers:  23BMC068, 23CMD237; \n462) REF DYNJ61663A, UDI/DI 10193489929652 (each) 40193489929653 (case), Lot Numbers:  22IBF170, 23CBF409, 22LBN360, 22LBA795; \n463) REF DYNJ61844B, UDI/DI 10195327495954 (each) 40195327495955 (case), Lot Numbers:  23IDB452, 23KDA607, 24ADA592, 24BDC034; \n464) REF DYNJ61845A, UDI/DI 10195327496050 (each) 40195327496051 (case), Lot Numbers:  23LDA497, 23JDB429, 23IDB697; \n465) REF DYNJ61846A, UDI/DI 10195327496487 (each) 40195327496488 (case), Lot Numbers:  23IDB696; \n466) REF DYNJ61849D, UDI/DI 10195327495558 (each) 40195327495559 (case), Lot Numbers:  23KDC327, 23JDA309, 24ADA958, 24CDB779, 24DDB411; \n467) REF DYNJ61908C, UDI/DI 10195327281915 (each) 40195327281916 (case), Lot Numbers:  23JBT556, 23GBT165, 23GBM755, 23EBX530, 23HBQ082, 23HBL263; \n468) REF DYNJ62031B, UDI/DI 10195327058272 (each) 40195327058273 (case), Lot Numbers:  22DBR051, 22JBW801, 23FBT974, 22LBT054, 23BBC855, 22LBS076, 23CBU872, 23IBV506, 23GBN446; \n469) REF DYNJ62034C, UDI/DI 10195327627157 (each) 40195327627158 (case), Lot Numbers:  24CBR453; \n470) REF DYNJ62035C, UDI/DI 10195327058258 (each) 40195327058259 (case), Lot Numbers:  22FBL674, 23DBB035, 23JBX179, 23GBP806; \n471) REF DYNJ62395C, UDI/DI 10195327316617 (each) 40195327316618 (case), Lot Numbers:  23ABN929, 23EBU099, 23KBQ093; \n472) REF DYNJ62395D, UDI/DI 10195327611378 (each) 40195327611379 (case), Lot Numbers:  24BBN072, 24DBT910; \n473) REF DYNJ62650B, UDI/DI 10195327121846 (each) 40195327121847 (case), Lot Numbers:  22GBF339, 23BBD864, 22HBM930, 22IBK987, 23CBO687, 23EBE814, 23FBE980; \n474) REF DYNJ62668C, UDI/DI 10195327546342 (each) 40195327546343 (case), Lot Numbers:  23KMC875; \n475) REF DYNJ62777D, UDI/DI 10195327068028 (each) 40195327068029 (case), Lot Numbers:  22EDA852; \n476) REF DYNJ63082B, UDI/DI 10195327102135 (each) 40195327102136 (case), Lot Numbers:  22CMD192, 23EMI678, 22IMA803, 22IMA805, 22LMJ344, 23CMI070; \n477) REF DYNJ63090A, UDI/DI 10195327358952 (each) 40195327358953 (case), Lot Numbers:  23FBL996, 23JBX772; \n478) REF DYNJ63095A, UDI/DI 10195327358990 (each) 40195327358991 (case), Lot Numbers:  23EBP508; \n479) REF DYNJ63300C, UDI/DI 10195327041038 (each) 40195327041039 (case), Lot Numbers:  22FBY595, 22FBX540; \n480) REF DYNJ63300D, UDI/DI 10195327275716 (each) 40195327275717 (case), Lot Numbers:  23DMD465, 23IMA358, 23IMG385, 24DMH488; \n481) REF DYNJ63301B, UDI/DI 10193489896961 (each) 40193489896962 (case), Lot Numbers:  22FBC835, 22DBN079; 22GBF112; \n482) REF DYNJ63301C, UDI/DI 10195327273644 (each) 40195327273645 (case), Lot Numbers:  23EMI318, 23IME153; \n483) REF DYNJ63302C, UDI/DI 10195327040987 (each) 40195327040988 (case), Lot Numbers:  22EBC472, 22FBD800; \n484) REF DYNJ63302D, UDI/DI 10195327275686 (each) 40195327275687 (case), Lot Numbers:  23FMI709, 23IMA802, 23IMA941; \n485) REF DYNJ63303C, UDI/DI 10195327040994 (each) 40195327040995 (case), Lot Numbers:  22HMG296, 22HMG297; \n486) REF DYNJ63303D, UDI/DI 10195327275723 (each) 40195327275724 (case), Lot Numbers:  23BMD048, 23IMH305, 23HMI184, 23KMB839; \n487) REF DYNJ63305B, UDI/DI 10195327041052 (each) 40195327041053 (case), Lot Numbers:  22FME428; \n488) REF DYNJ63306B, UDI/DI 10195327270049 (each) 40195327270040 (case), Lot Numbers:  23JMG735, 23HME655, 23GMD024, 23DMC975, 23BMC104; \n489) REF DYNJ63310C, UDI/DI 10195327041168 (each) 40195327041169 (case), Lot Numbers:  22FBU999; \n490) REF DYNJ63310D, UDI/DI 10195327270636 (each) 40195327270637 (case), Lot Numbers:  23DMB224, 23IMD732, 23JMG130; \n491) REF DYNJ63311B, UDI/DI 10193489896978 (each) 40193489896979 (case), Lot Numbers:  22EBG812, 22EBG200; \n492) REF DYNJ63311C, UDI/DI 10195327273637 (each) 40195327273638 (case), Lot Numbers:  23FMG395, 23HME647, 23IME154; \n493) REF DYNJ63312C, UDI/DI 10195327270667 (each) 40195327270668 (case), Lot Numbers:  23BMA186, 23AMI101; \n494) REF DYNJ63312D, UDI/DI 10195327478476 (each) 40195327478477 (case), Lot Numbers:  23IME155; \n495) REF DYNJ63316D, UDI/DI 10195327273545 (each) 40195327273546 (case), Lot Numbers:  23IMA324, 22LMH495, 22LMH499; \n496) REF DYNJ63318A, UDI/DI 10193489989441 (each) 40193489989442 (case), Lot Numbers:  22GBH439, 22GBB669; \n497) REF DYNJ63318B, UDI/DI 10195327270063 (each) 40195327270064 (case), Lot Numbers:  23AMH442, 23GMC956, 23IMA362, 23JMF710; \n498) REF DYNJ63319B, UDI/DI 10193489949391 (each) 40193489949392 (case), Lot Numbers:  22GBJ839; \n499) REF DYNJ63319C, UDI/DI 10195327273651 (each) 40195327273652 (case), Lot Numbers:  23AMG001; \n500) REF DYNJ63324B, UDI/DI 10193489958140 (each) 40193489958141 (case), Lot Numbers:  22EBI790, 22EBI793, 22IMG131; \n501) REF DYNJ63324C, UDI/DI 10195327275761 (each) 40195327275762 (case), Lot Numbers:  23DMC973, 23DMA796; \n502) REF DYNJ63324D, UDI/DI 10195327402143 (each) 40195327402144 (case), Lot Numbers:  23IMA813, 23JMG201, 23KMI808; \n503) REF DYNJ63339D, UDI/DI 10195327273606 (each) 40195327273607 (case), Lot Numbers:  23HBE292; \n504) REF DYNJ63349C, UDI/DI 10193489983746 (each) 40193489983747 (case), Lot Numbers:  22EBF338, 22GMA096, 22FMB860; \n505) REF DYNJ63373D, UDI/DI 10195327086152 (each) 40195327086153 (case), Lot Numbers:  22FBY598; \n506) REF DYNJ63373F, UDI/DI 10195327270780 (each) 40195327270781 (case), Lot Numbers:  22LBI728, 22LBI729, 23CBT768, 23FBG742, 23IBO340; \n507) REF DYNJ63436A, UDI/DI 10193489252972 (each) 40193489252973 (case), Lot Numbers:  22GBJ224, 22GBP404; \n508) REF DYNJ63436B, UDI/DI 10195327282240 (each) 40195327282241 (case), Lot Numbers:  22KBQ089; \n509) REF DYNJ63436C, UDI/DI 10195327319380 (each) 40195327319381 (case), Lot Numbers:  23FBB921, 23HBJ572; \n510) REF DYNJ63473C, UDI/DI 10195327510923 (each) 40195327510924 (case), Lot Numbers:  23IMF358, 24BMA722; \n511) REF DYNJ63552B, UDI/DI 10193489900378 (each) 40193489900379 (case), Lot Numbers:  22FBF321, 23BBC141, 22LBJ155, 23BBJ095, 23EBV888, 23GBU067, 23KBI857, 23JBO055, 23IBO111; \n512) REF DYNJ63557B, UDI/DI 10193489900392 (each) 40193489900393 (case), Lot Numbers:  22DBN108, 23CBM182, 23BBH750, 23HBP781; \n513) REF DYNJ63558B, UDI/DI 10193489900415 (each) 40193489900416 (case), Lot Numbers:  23ABU703, 22LBN733, 23DBJ042, 23GBU068, 23IBQ589; \n514) REF DYNJ63559B, UDI/DI 10193489900422 (each) 40193489900423 (case), Lot Numbers:  22DBS937, 22FBS409, 22IBG382, 23ABO512, 23IBK484, 23HBP711, 23GBU069, 23KBI346; \n515) REF DYNJ63583B, UDI/DI 10195327122898 (each) 40195327122899 (case), Lot Numbers:  23GBL805; \n516) REF DYNJ63604B, UDI/DI 10193489900231 (each) 40193489900232 (case), Lot Numbers:  22FBO473, 22KBH234, 22KBH617, 22KBH240, 23CBF429, 23GBP853, 23JBE719; \n517) REF DYNJ63607B, UDI/DI 10193489900255 (each) 40193489900256 (case), Lot Numbers:  22FBV007, 23JBP047, 23JBA953, 23IBA206, 23GBQ019, 23EBQ718, 23CBS745; \n518) REF DYNJ63609C, UDI/DI 10195327350284 (each) 40195327350285 (case), Lot Numbers:  23JBQ193, 23HBI039; \n519) REF DYNJ63614B, UDI/DI 10195327350543 (each) 40195327350544 (case), Lot Numbers:  23FBE255, 23GBT158, 23CBS818, 23IBA158; \n520) REF DYNJ63616B, UDI/DI 10195327350369 (each) 40195327350360 (case), Lot Numbers:  23EBT358, 23DBE423, 23GBO792, 23JBR539; \n521) REF DYNJ63620B, UDI/DI 10193489900279 (each) 40193489900270 (case), Lot Numbers:  22DBS615, 22DBS700, 22DBS701; \n522) REF DYNJ63639B, UDI/DI 10193489900323 (each) 40193489900324 (case), Lot Numbers:  22FBK829, 22LBP339; \n523) REF DYNJ63641B, UDI/DI 10195327122805 (each) 40195327122806 (case), Lot Numbers:  22JBJ076, 22FBV009, 23EBW585, 23IBN177, 23JBC776; \n524) REF DYNJ63646B, UDI/DI 10195327350352 (each) 40195327350353 (case), Lot Numbers:  23CBT773, 23JBR021; \n525) REF DYNJ63653C, UDI/DI 10195327350468 (each) 40195327350469 (case), Lot Numbers:  23DBB425, 23FBF856, 23IBA351, 23GBS392, 23JBV605; \n526) REF DYNJ63654A, UDI/DI 10195327350512 (each) 40195327350513 (case), Lot Numbers:  23CBC038, 23FBD023, 23GBJ801, 24ABW300, 23JBC777; \n527) REF DYNJ63831C, UDI/DI 10195327041007 (each) 40195327041008 (case), Lot Numbers:  22FBB477, 22GBD802, 22FBD777; \n528) REF DYNJ63831D, UDI/DI 10195327275754 (each) 40195327275755 (case), Lot Numbers:  23BMD043, 23KMF707; \n529) REF DYNJ63886A, UDI/DI 10195327221836 (each) 40195327221837 (case), Lot Numbers:  23GDB254, 23KDB405, 23FDB375, 24CDB332; \n530) REF DYNJ63939A, UDI/DI 10193489270006 (each) 40193489270007 (case), Lot Numbers:  22EDA758; \n531) REF DYNJ63939B, UDI/DI 10195327221850 (each) 40195327221851 (case), Lot Numbers:  22IDC124, 23GDB082, 23JDA267, 23KDB911, 23KDA705, 23CDB169, 24BDA519, 23DDB707, 24CDB509; \n532) REF DYNJ63942A, UDI/DI 10193489270013 (each) 40193489270014 (case), Lot Numbers:  22FDA747; \n533) REF DYNJ63942B, UDI/DI 10195327221867 (each) 40195327221868 (case), Lot Numbers:  23BDA783; \n534) REF DYNJ63942C, UDI/DI 10195327347192 (each) 40195327347193 (case), Lot Numbers:  23JDA396, 23CDB849, 23HDC482, 23GDA461, 23KDB369, 24ADB754; \n535) REF DYNJ63957B, UDI/DI 10195327364939 (each) 40195327364930 (case), Lot Numbers:  23DDC317; \n536) REF DYNJ63957C, UDI/DI 10195327428860 (each) 40195327428861 (case), Lot Numbers:  23GDB209, 23IDB634, 23KDB938, 23FDB374; \n537) REF DYNJ63957D, UDI/DI 10195327580285 (each) 40195327580286 (case), Lot Numbers:  23LDB149, 24ADA221, 24ADB760; \n538) REF DYNJ64004D, UDI/DI 10195327118679 (each) 40195327118670 (case), Lot Numbers:  22EDB550, 22HDA301, 23FDA090, 23GDB263, 23JDA746; \n539) REF DYNJ64005D, UDI/DI 10195327118662 (each) 40195327118663 (case), Lot Numbers:  22DDA783, 22HDB620, 23CDB599, 23DDC345, 23ADC123, 22FDA748, 23HDB656; \n540) REF DYNJ64008C, UDI/DI 10195327118587 (each) 40195327118588 (case), Lot Numbers:  22GDC267, 22EDB024, 23BDA201, 23CDA348, 23IDC065; \n541) REF DYNJ64011D, UDI/DI 10195327118594 (each) 40195327118595 (case), Lot Numbers:  23BDA197; \n542) REF DYNJ64184C, UDI/DI 10195327278793 (each) 40195327278794 (case), Lot Numbers:  23ABB562, 23FBL254, 23HBM054; \n543) REF DYNJ64233B, UDI/DI 10193489382648 (each) 40193489382649 (case), Lot Numbers:  22EMF960, 22KMF038, 22IMG521, 22IMG699, 23BMG941; \n544) REF DYNJ64240A, UDI/DI 10193489382631 (each) 40193489382632 (case), Lot Numbers:  22EMA915, 22GMA838, 22JMB039, 22KME606, 22IMF876, 23DMB656, 23CMG377; \n545) REF DYNJ64276C, UDI/DI 10195327091095 (each) 40195327091096 (case), Lot Numbers:  22DBB451, 23EBG970, 22IBF538, 22FBX958, 22FBX960, 23ABL863, 23BBQ435; \n546) REF DYNJ64344A, UDI/DI 10195327157425 (each) 40195327157426 (case), Lot Numbers:  22FBB836; \n547) REF DYNJ64344B, UDI/DI 10195327190293 (each) 40195327190294 (case), Lot Numbers:  22KBJ534, 22LBM629; \n548) REF DYNJ64672A, UDI/DI 10193489496963 (each) 40193489496964 (case), Lot Numbers:  22GMA040; \n549) REF DYNJ64755C, UDI/DI 10195327665968 (each) 40195327665969 (case), Lot Numbers:  24CDB511; \n550) REF DYNJ64796C, UDI/DI 10195327217945 (each) 40195327217946 (case), Lot Numbers:  23DDC094, 23GDB171, 23EDC564; \n551) REF DYNJ64907A, UDI/DI 10195327053734 (each) 40195327053735 (case), Lot Numbers:  22EBI609, 22EBP072, 22JBM618, 22LBP896, 22KBD336, 22JBS427; \n552) REF DYNJ64907B, UDI/DI 10195327489632 (each) 40195327489633 (case), Lot Numbers:  23IBH849; \n553) REF DYNJ64955A, UDI/DI 10195327110949 (each) 40195327110940 (case), Lot Numbers:  22JMA953, 22HMA793, 22FMG704, 23EMB887, 23BMD644, 23BMD059, 23EMG035, 23IMH097, 23JMJ933; \n554) REF DYNJ64981F, UDI/DI 10195327528133 (each) 40195327528134 (case), Lot Numbers:  23KMA711; \n555) REF DYNJ65065A, UDI/DI 10193489317244 (each) 40193489317245 (case), Lot Numbers:  23AMG500, 22FME986, 23BMC082, 23HMC959, 24DMJ424, 23KMA049, 23LMH308, 24BMJ166, 24AMG861; \n556) REF DYNJ65066B, UDI/DI 10193489317183 (each) 40193489317184 (case), Lot Numbers:  23AMG583, 22KME040, 22GMH242, 22GMA626, 22DMF180, 23DMA376, 23BMB169, 24BMA894, 23LMF624, 23KME638, 23FMD960, 23EMI651, 24AMG151, 24FMI648; \n557) REF DYNJ65069A, UDI/DI 10193489317237 (each) 40193489317238 (case), Lot Numbers:  22LMD079, 23HMH939, 23KMJ410, 23KME112, 23BME898, 23AMG551, 23BMA302, 23AMB608, 24CMB410; \n558) REF DYNJ65240A, UDI/DI 10193489416978 (each) 40193489416979 (case), Lot Numbers:  22DBT346; \n559) REF DYNJ65240B, UDI/DI 10195327163778 (each) 40195327163779 (case), Lot Numbers:  22FBX940, 22LBL503, 23DBH093, 23ABT638, 23FBT717; \n560) REF DYNJ65246, UDI/DI 10193489257595 (each) 40193489257596 (case), Lot Numbers:  23AMG386, 23FMB173, 23FMI413, 23JMJ638; \n561) REF DYNJ65386C, UDI/DI 10195327410087 (each) 40195327410088 (case), Lot Numbers:  23FMC511; \n562) REF DYNJ65484A, UDI/DI 10195327303501 (each) 40195327303502 (case), Lot Numbers:  23HMD135, 23CMF388; \n563) REF DYNJ65825A, UDI/DI 10195327187729 (each) 40195327187720 (case), Lot Numbers:  23HMB027, 23IMD579; \n564) REF DYNJ65855B, UDI/DI 10195327232146 (each) 40195327232147 (case), Lot Numbers:  22HBW485, 22LBD735, 23BBQ528; \n565) REF DYNJ66236C, UDI/DI 10193489897272 (each) 40193489897273 (case), Lot Numbers:  22IMA620; \n566) REF DYNJ66236D, UDI/DI 10195327275730 (each) 40195327275731 (case), Lot Numbers:  23AMH457, 23GMI240, 23HMI160; \n567) REF DYNJ66237C, UDI/DI 10193489897289 (each) 40193489897280 (case), Lot Numbers:  22FBU863; \n568) REF DYNJ66237D, UDI/DI 10195327275747 (each) 40195327275748 (case), Lot Numbers:  23FMI227, 23DMD426, 23EMB393; \n569) REF DYNJ66357C, UDI/DI 10195327227975 (each) 40195327227976 (case), Lot Numbers:  22IBG338, 22HBX757, 22JBV661, 22JBW838, 23CBI755, 23DBB811; \n570) REF DYNJ66649D, UDI/DI 10195327374570 (each) 40195327374571 (case), Lot Numbers:  23DBP833; \n571) REF DYNJ66730A, UDI/DI 10195327504465 (each) 40195327504466 (case), Lot Numbers:  23LBD988, 24ABN196, 24BBC802, 24BBC859; \n572) REF DYNJ66738A, UDI/DI 10195327504946 (each) 40195327504947 (case), Lot Numbers:  23KBN307, 24ABE488, 24BBF455, 24CBI276, 24EBL622; \n573) REF DYNJ66968A, UDI/DI 10195327267117 (each) 40195327267118 (case), Lot Numbers:  23IBJ698; \n574) REF DYNJ66969A, UDI/DI 10195327267520 (each) 40195327267521 (case), Lot Numbers:  22LBD280, 23FBO304, 23HBL009, 23IBQ506; \n575) REF DYNJ66976A, UDI/DI 10193489879490 (each) 40193489879491 (case), Lot Numbers:  22JBG282, 23KBD050, 22LBD165; \n576) REF DYNJ66981A, UDI/DI 10193489879537 (each) 40193489879538 (case), Lot Numbers:  23CBH243, 22DBK542, 22KBL344, 22IBR511, 23HBD638, 23KBC458, 23JBM179, 23KBQ338, 23KBQ341, 23ABK662, 23FBO864, 23DBJ273; \n577) REF DYNJ67195C, UDI/DI 10195327299705 (each) 40195327299706 (case), Lot Numbers:  23BBI892; \n578) REF DYNJ67196B, UDI/DI 10195327299712 (each) 40195327299713 (case), Lot Numbers:  23CBW673, 22LBO927; \n579) REF DYNJ67212A, UDI/DI 10195327150549 (each) 40195327150540 (case), Lot Numbers:  23JBI789, 23KBU521, 22LBD041, 22LBD045, 23BBC539, 23DBM715, 23GBA079; \n580) REF DYNJ67212B, UDI/DI 10195327577940 (each) 40195327577941 (case), Lot Numbers:  24ABB293; \n581) REF DYNJ67212C, UDI/DI 10195327621735 (each) 40195327621736 (case), Lot Numbers:  24BBU528; \n582) REF DYNJ67212D, UDI/DI 10195327668358 (each) 40195327668359 (case), Lot Numbers:  24DBQ160; \n583) REF DYNJ67213A, UDI/DI 10195327150556 (each) 40195327150557 (case), Lot Numbers:  22EBP199, 23BBB397, 23BBB201, 23CBA507;\n584) REF DYNJ67264A, UDI/DI 10195327039530 (each) 40195327039531 (case), Lot Numbers:  22EBM706, 22JBX213, 22JBG223, 23IBC366, 23HBY717, 23HBB411, 23JBF392, 23KBS927, 23DBK428, 23GBI625; \n585) REF DYNJ67264B, UDI/DI 10195327578527 (each) 40195327578528 (case), Lot Numbers:  24ABU946; \n586) REF DYNJ67264C, UDI/DI 10195327670153 (each) 40195327670154 (case), Lot Numbers:  24GBW785, 24DBO685; \n587) REF DYNJ67265B, UDI/DI 10195327150921 (each) 40195327150922 (case), Lot Numbers:  22GBF576, 22EBM303, 22JBI919, 23JBC924, 23GBQ718, 23KBC712, 23LBC682, 22LBD113, 23ABJ042; \n588) REF DYNJ67520G, UDI/DI 10195327334833 (each) 40195327334834 (case), Lot Numbers:  23BBC223, 23ABQ621, 23JBQ768, 23JBF278, 23GBH254, 23KBR697, 24ABF738, 23IBF489, 24CBO857; \n589) REF DYNJ67521C, UDI/DI 10195327397913 (each) 40195327397914 (case), Lot Numbers:  23GDC211, 23KDA461, 24BDA015, 23IDA321, 23DDB944; \n590) REF DYNJ67522D, UDI/DI 10195327332556 (each) 40195327332557 (case), Lot Numbers:  23KBG850, 23LBJ239, 23LBK732, 23ABN233, 23GBB678; \n591) REF DYNJ67574I, UDI/DI 10195327406578 (each) 40195327406579 (case), Lot Numbers:  23FBM267; \n592) REF DYNJ67596A, UDI/DI 10193489841060 (each) 40193489841061 (case), Lot Numbers:  22FBU865; \n593) REF DYNJ67596B, UDI/DI 10195327217532 (each) 40195327217533 (case), Lot Numbers:  22JBL753, 23CBC289, 23IBR961, 23FBM685, 23KBB954; \n594) REF DYNJ67677A, UDI/DI 10195327504779 (each) 40195327504770 (case), Lot Numbers:  23KBM911; \n595) REF DYNJ67801B, UDI/DI 10195327228026 (each) 40195327228027 (case), Lot Numbers:  22IBR468, 22IBA999, 22HBX760, 22LBJ174, 23CBB332, 23BBH671, 23DBC354; \n596) REF DYNJ67802B, UDI/DI 10195327227951 (each) 40195327227952 (case), Lot Numbers:  22IBB139, 22IBA731, 23ABJ302; \n597) REF DYNJ67886B, UDI/DI 10193489980523 (each) 40193489980524 (case), Lot Numbers:  22ILA134, 22HLA375; \n598) REF DYNJ67886C, UDI/DI 10195327265359 (each) 40195327265350 (case), Lot Numbers:  22JLB007, 22KLA316, 23CLA643, 23DLA767; \n599) REF DYNJ67887B, UDI/DI 10193489970616 (each) 40193489970617 (case), Lot Numbers:  22IMF688; \n600) REF DYNJ67959, UDI/DI 10193489459425 (each) 40193489459426 (case), Lot Numbers:  22DBT576, 22DBF011, 22FBO580, 22JBI285, 22IBO190, 23CBA566, 23ABA923; \n601) REF DYNJ67983A, UDI/DI 10195327104429 (each) 40195327104420 (case), Lot Numbers:  22EBF444, 22EBF445; \n602) REF DYNJ68204A, UDI/DI 10195327026356 (each) 40195327026357 (case), Lot Numbers:  22EMA922; \n603) REF DYNJ68204B, UDI/DI 10195327352684 (each) 40195327352685 (case), Lot Numbers:  23EMB404, 23HMF322, 23FMD314, 23EMI706; \n604) REF DYNJ68204C, UDI/DI 10195327512095 (each) 40195327512096 (case), Lot Numbers:  23JMG139, 23IMG918, 24BMA772, 24AMC705, 24AMA748, 24AMH321, 24CMC554; \n605) REF DYNJ68206A, UDI/DI 10195327026226 (each) 40195327026227 (case), Lot Numbers:  22ELA483, 22KLA629, 22JLB227, 23ALB033; \n606) REF DYNJ68206B, UDI/DI 10195327352615 (each) 40195327352616 (case), Lot Numbers:  23HLB045, 23GLA095, 23DLA972, 23CLA562; \n607) REF DYNJ68206C, UDI/DI 10195327512118 (each) 40195327512119 (case), Lot Numbers:  23KLA363, 24ALB145; \n608) REF DYNJ68553A, UDI/DI 10193489994889 (each) 40193489994880 (case), Lot Numbers:  23CBL534, 22IBD248, 23HBG150, 23BBC646, 23ABI373, 23DBR643, 23GBC957; \n609) REF DYNJ69170A, UDI/DI 10195327156497 (each) 40195327156498 (case), Lot Numbers:  22EMG183, 22JMD107; \n610) REF DYNJ69170B, UDI/DI 10195327342975 (each) 40195327342976 (case), Lot Numbers:  23GMC991, 23FMB245; \n611) REF DYNJ69170C, UDI/DI 10195327491352 (each) 40195327491353 (case), Lot Numbers:  23LMH595, 23JMI162; \n612) REF DYNJ69207A, UDI/DI 10195327187712 (each) 40195327187713 (case), Lot Numbers:  23DMC276, 23FMH470; \n613) REF DYNJ69207B, UDI/DI 10195327516239 (each) 40195327516230 (case), Lot Numbers:  23IMH529; \n614) REF DYNJ69487B, UDI/DI 10195327227159 (each) 40195327227150 (case), Lot Numbers:  23GDB343, 23CDA273, 23BDB852, 23JDA197; \n615) REF DYNJ69577C, UDI/DI 10195327137205 (each) 40195327137206 (case), Lot Numbers:  22EBH426, 22FBF033; \n616) REF DYNJ69577F, UDI/DI 10195327212896 (each) 40195327212897 (case), Lot Numbers:  22JBX725; \n617) REF DYNJ69577G, UDI/DI 10195327302337 (each) 40195327302338 (case), Lot Numbers:  23KBG592, 24ABI922, 23LBL002, 23DBI943, 24ABU087, 23EBW816, 23IBJ299, 23JBL134; \n618) REF DYNJ69578B, UDI/DI 10195327130435 (each) 40195327130436 (case), Lot Numbers:  22EME040; \n619) REF DYNJ69578D, UDI/DI 10195327302320 (each) 40195327302321 (case), Lot Numbers:  23LMH283, 23KMJ024, 24AMG093, 23KMA437, 23FMA685, 23AMJ500, 23AMB592; \n620) REF DYNJ80110A, UDI/DI 10195327098964 (each) 40195327098965 (case), Lot Numbers:  22IBH937, 23CME933, 23KME620; \n621) REF DYNJ80130, UDI/DI 10195327010577 (each) 40195327010578 (case), Lot Numbers:  22IMF536; \n622) REF DYNJ80140A, UDI/DI 10195327148058 (each) 40195327148059 (case), Lot Numbers:  22IMA277; \n623) REF DYNJ80245A, UDI/DI 10195327195076 (each) 40195327195077 (case), Lot Numbers:  22FBW650; \n624) REF DYNJ80284, UDI/DI 10195327020750 (each) 40195327020751 (case), Lot Numbers:  22GMA635, 22GMA293, 22DMH685, 22LMG202, 23DMB252, 24DMH294, 23IMA947; \n625) REF DYNJ80298, UDI/DI 10195327021979 (each) 40195327021970 (case), Lot Numbers:  22GME140, 22IMB430; \n626) REF DYNJ80424A, UDI/DI 10195327234119 (each) 40195327234110 (case), Lot Numbers:  22IBL761, 23BBE371; \n627) REF DYNJ80435A, UDI/DI 10195327234102 (each) 40195327234103 (case), Lot Numbers:  22IBL761, 23BBE371, 23DBH684; \n628) REF DYNJ80450A, UDI/DI 10195327168919 (each) 40195327168910 (case), Lot Numbers:  22FLB145, 22FLA590, 22GLA921; \n629) REF DYNJ80450B, UDI/DI 10195327328542 (each) 40195327328543 (case), Lot Numbers:  23ALB121, 23CLA158, 23HLB046; \n630) REF DYNJ80451, UDI/DI 10195327038304 (each) 40195327038305 (case), Lot Numbers:  22ILA015, 22JLA682; \n631) REF DYNJ80503A, UDI/DI 10195327225483 (each) 40195327225484 (case), Lot Numbers:  22IMF538, 22KMB095, 23EMI342, 23EMB067, 23FMH680; \n632) REF DYNJ80505, UDI/DI 10195327049195 (each) 40195327049196 (case), Lot Numbers:  22GMH017; \n633) REF DYNJ80696, UDI/DI 10195327066482 (each) 40195327066483 (case), Lot Numbers:  22EBE965; \n634) REF DYNJ80696A, UDI/DI 10195327228415 (each) 40195327228416 (case), Lot Numbers:  22JBC276; \n635) REF DYNJ80696B, UDI/DI 10195327274115 (each) 40195327274116 (case), Lot Numbers:  23BBD025, 22LBR313; \n636) REF DYNJ80697A, UDI/DI 10195327228521 (each) 40195327228522 (case), Lot Numbers:  22IBL693; \n637) REF DYNJ80697B, UDI/DI 10195327274108 (each) 40195327274109 (case), Lot Numbers:  23ABR781, 22JBS491, 23BBO609; \n638) REF DYNJ81176, UDI/DI 10195327089443 (each) 40195327089444 (case), Lot Numbers:  22EBM753, 22JBS368, 23ABG787, 23ABG437, 23HBB483, 23CBV612, 23JBN909, 23LBF379, 24GBV429, 24EBE649; \n639) REF DYNJ81405A, UDI/DI 10195327470159 (each) 40195327470150 (case), Lot Numbers:  23IBN860; \n640) REF DYNJ81598, UDI/DI 10195327124014 (each) 40195327124015 (case), Lot Numbers:  22HBO968, 22EBJ445, 22IBU035, 22HBX019; \n641) REF DYNJ81598A, UDI/DI 10195327314934 (each) 40195327314935 (case), Lot Numbers:  22LBP952, 22LBN846, 23DBB923; \n642) REF DYNJ81608, UDI/DI 10195327123925 (each) 40195327123926 (case), Lot Numbers:  22IBF053, 22EBJ747, 22JBS356; \n643) REF DYNJ81608A, UDI/DI 10195327314927 (each) 40195327314928 (case), Lot Numbers:  23ABA190, 22LBN901, 23DBP300; \n644) REF DYNJ81611, UDI/DI 10195327123963 (each) 40195327123964 (case), Lot Numbers:  22JBD143; \n645) REF DYNJ81612, UDI/DI 10195327123987 (each) 40195327123988 (case), Lot Numbers:  22EBJ448, 22IBR520, 22HBY228; \n646) REF DYNJ81612A, UDI/DI 10195327315061 (each) 40195327315062 (case), Lot Numbers:  22LBN205, 23CBA405, 23DBM719; \n647) REF DYNJ81783A, UDI/DI 10195327515874 (each) 40195327515875 (case), Lot Numbers:  23IBT721; \n648) REF DYNJ81783B, UDI/DI 10195327532215 (each) 40195327532216 (case), Lot Numbers:  23KBT832; \n649) REF DYNJ81786C, UDI/DI 10195327525514 (each) 40195327525515 (case), Lot Numbers:  23JBI949; \n650) REF DYNJ81786D, UDI/DI 10195327532222 (each) 40195327532223 (case), Lot Numbers:  23KBT803, 24BBC051, 24CBR666, 24EBD427; \n651) REF DYNJ81812B, UDI/DI 10195327681517 (each) 40195327681518 (case), Lot Numbers:  24DDB407; \n652) REF DYNJ81840, UDI/DI 10195327149543 (each) 40195327149544 (case), Lot Numbers:  22FMF269; \n653) REF DYNJ82178, UDI/DI 10195327182694 (each) 40195327182695 (case), Lot Numbers:  23GDB510, 23CDA038, 23FDA552, 23IDC200, 23KMD170; \n654) REF DYNJ82203, UDI/DI 10195327183691 (each) 40195327183692 (case), Lot Numbers:  23HMI536; \n655) REF DYNJ82204, UDI/DI 10195327183714 (each) 40195327183715 (case), Lot Numbers:  23EDB094, 23HDA590, 23FDC215; \n656) REF DYNJ82204A, UDI/DI 10195327520069 (each) 40195327520060 (case), Lot Numbers:  23JDA223; \n657) REF DYNJ82205, UDI/DI 10195327183707 (each) 40195327183708 (case), Lot Numbers:  23EDB908, 23IDB080, 23CDA306; \n658) REF DYNJ82258, UDI/DI 10195327181703 (each) 40195327181704 (case), Lot Numbers:  23FDA696, 23DDA374; \n659) REF DYNJ82287, UDI/DI 10195327183974 (each) 40195327183975 (case), Lot Numbers:  23FDA654, 23GDB511; \n660) REF DYNJ82293, UDI/DI 10195327184315 (each) 40195327184316 (case), Lot Numbers:  23BDA115; \n661) REF DYNJ82293A, UDI/DI 10195327503512 (each) 40195327503513 (case), Lot Numbers:  23IDB110, 23JMH999; \n662) REF DYNJ82299, UDI/DI 10195327184322 (each) 40195327184323 (case), Lot Numbers:  23HDB220, 23HDA189; \n663) REF DYNJ82299A, UDI/DI 10195327503543 (each) 40195327503544 (case), Lot Numbers:  23JMH016, 23JMB780; \n664) REF DYNJ82311, UDI/DI 10195327184346 (each) 40195327184347 (case), Lot Numbers:  23FDC334; \n665) REF DYNJ82311A, UDI/DI 10195327503475 (each) 40195327503476 (case), Lot Numbers:  23IDB027; \n666) REF DYNJ82313, UDI/DI 10195327184339 (each) 40195327184330 (case), Lot Numbers:  23FDA667, 23HDB650, 23IMA325; \n667) REF DYNJ82315, UDI/DI 10195327184353 (each) 40195327184354 (case), Lot Numbers:  23CDB089; \n668) REF DYNJ82315A, UDI/DI 10195327503505 (each) 40195327503506 (case), Lot Numbers:  23JMH046; \n669) REF DYNJ82317, UDI/DI 10195327185008 (each) 40195327185009 (case), Lot Numbers:  23DDA012, 23GDB650, 23GDA239; \n670) REF DYNJ82331, UDI/DI 10195327185046 (each) 40195327185047 (case), Lot Numbers:  23ADB798, 23FDA073; \n671) REF DYNJ82334, UDI/DI 10195327184988 (each) 40195327184989 (case), Lot Numbers:  23BDA016, 23FDA729; \n672) REF DYNJ82340, UDI/DI 10195327185015 (each) 40195327185016 (case), Lot Numbers:  23HDA579, 23IDA276, 23GDA240; \n673) REF DYNJ82341, UDI/DI 10195327185084 (each) 40195327185085 (case), Lot Numbers:  23ADC134, 23IMA893; \n674) REF DYNJ82343, UDI/DI 10195327185091 (each) 40195327185092 (case), Lot Numbers:  23EDB409, 23IDB878; \n675) REF DYNJ82345A, UDI/DI 10195327316556 (each) 40195327316557 (case), Lot Numbers:  23ABE332, 23GBR353, 23CBV824, 23LBT387, 23GBK085, 23GBD178, 23FBE485, 24ABT128; \n676) REF DYNJ82365, UDI/DI 10195327186715 (each) 40195327186716 (case), Lot Numbers:  23HDA097, 23JMB452; \n677) REF DYNJ82367A, UDI/DI 10195327316563 (each) 40195327316564 (case), Lot Numbers:  23ABI222, 23HBC300, 23LBD137, 24ABT015, 24ABE893, 23GBK112, 23FBU725; \n678) REF DYNJ82669, UDI/DI 10195327202811 (each) 40195327202812 (case), Lot Numbers:  23HMF026, 23GMI179, 23HMI734; \n679) REF DYNJ82673, UDI/DI 10195327203047 (each) 40195327203048 (case), Lot Numbers:  23DMI297, 23GMD089, 23IMI418, 23HMC831, 23JMJ228, 23LME450; \n680) REF DYNJ82675, UDI/DI 10195327203061 (each) 40195327203062 (case), Lot Numbers:  23GMD792, 23HME628; \n681) REF DYNJ82685, UDI/DI 10195327202873 (each) 40195327202874 (case), Lot Numbers:  23FMF892, 23GMD848, 23KME727, 23IMG386, 23HMF087; \n682) REF DYNJ82689, UDI/DI 10195327202910 (each) 40195327202911 (case), Lot Numbers:  23JMH128; \n683) REF DYNJ82693, UDI/DI 10195327202958 (each) 40195327202959 (case), Lot Numbers:  23GMC986, 23IMA575, 23KMF187; \n684) REF DYNJ82950, UDI/DI 10195327218966 (each) 40195327218967 (case), Lot Numbers:  22JBD140, 23BBS169; \n685) REF DYNJ82953, UDI/DI 10195327218997 (each) 40195327218998 (case), Lot Numbers:  22JBD999, 23DBO510, 23CBJ924; \n686) REF DYNJ83005, UDI/DI 10195327225896 (each) 40195327225897 (case), Lot Numbers:  22JMH340; \n687) REF DYNJ83005A, UDI/DI 10195327377083 (each) 40195327377084 (case), Lot Numbers:  23HME353, 23KMD629, 23FME324, 24AMD974, 24DME585, 24CMB225; \n688) REF DYNJ83006, UDI/DI 10195327225902 (each) 40195327225903 (case), Lot Numbers:  23AMG948, 22JME292; \n689) REF DYNJ83006A, UDI/DI 10195327377090 (each) 40195327377091 (case), Lot Numbers:  23LMA658, 23KMB885, 24FMI201, 24AMD927, 24CMC319; \n690) REF DYNJ83010, UDI/DI 10195327226251 (each) 40195327226252 (case), Lot Numbers:  22KMA653, 23CMF803; \n691) REF DYNJ83010A, UDI/DI 10195327377106 (each) 40195327377107 (case), Lot Numbers:  23HMA251, 23FME163; \n692) REF DYNJ83010B, UDI/DI 10195327500733 (each) 40195327500734 (case), Lot Numbers:  23IMF327, 23JMI712, 23LMA550, 23KME568, 24AMG159, 24CMC390; \n693) REF DYNJ83107, UDI/DI 10195327232221 (each) 40195327232222 (case), Lot Numbers:  22LMF005; \n694) REF DYNJ83107A, UDI/DI 10195327381950 (each) 40195327381951 (case), Lot Numbers:  23DMF950; \n695) REF DYNJ83186A, UDI/DI 10195327282745 (each) 40195327282746 (case), Lot Numbers:  22KBM902; \n696) REF DYNJ83186B, UDI/DI 10195327364625 (each) 40195327364626 (case), Lot Numbers:  23DBB562, 24ABI748, 23JBE810, 23KBW468, 23KBA804, 23FBE338, 24ABN430, 24BBG127, 23LBP872; \n697) REF DYNJ83234, UDI/DI 10195327247782 (each) 40195327247783 (case), Lot Numbers:  23EBC088, 22KBI755, 24FBO388, 24EBA239, 24CBD652; \n698) REF DYNJ83420, UDI/DI 10195327265380 (each) 40195327265381 (case), Lot Numbers:  22LBP312, 22LBN377, 23CBH126, 23EBT725, 23HBO639, 23KBL617, 23JBW081, 24ABK394, 24CBA696, 24DBV489; \n699) REF DYNJ83495, UDI/DI 10195327271800 (each) 40195327271801 (case), Lot Numbers:  23ABV177, 23IBK256; \n700) REF DYNJ83591, UDI/DI 10195327283704 (each) 40195327283705 (case), Lot Numbers:  23ABV178; \n701) REF DYNJ83592, UDI/DI 10195327283803 (each) 40195327283804 (case), Lot Numbers:  23BMG540; \n702) REF DYNJ83628, UDI/DI 10195327286897 (each) 40195327286898 (case), Lot Numbers:  23HMA558, 23FMB991, 23EMD373, 23IMC902, 23KMJ347, 24BMB018; \n703) REF DYNJ83628A, UDI/DI 10195327396817 (each) 40195327396818 (case), Lot Numbers:  23CBC376; \n704) REF DYNJ83660, UDI/DI 10195327292430 (each) 40195327292431 (case), Lot Numbers:  23CBS225; \n705) REF DYNJ83664, UDI/DI 10195327292553 (each) 40195327292554 (case), Lot Numbers:  23BBP135; \n706) REF DYNJ83760, UDI/DI 10195327301743 (each) 40195327301744 (case), Lot Numbers:  23BBP136, 23BBP535; \n707) REF DYNJ83761, UDI/DI 10195327301750 (each) 40195327301751 (case), Lot Numbers:  23CBV738; \n708) REF DYNJ84057A, UDI/DI 10195327358020 (each) 40195327358021 (case), Lot Numbers:  23CBV739; \n709) REF DYNJ84060A, UDI/DI 10195327357993 (each) 40195327357994 (case), Lot Numbers:  23HMA973, 23IMD249; \n710) REF DYNJ84185, UDI/DI 10195327340575 (each) 40195327340576 (case), Lot Numbers:  23HME837, 23IMF690; \n711) REF DYNJ84186, UDI/DI 10195327340568 (each) 40195327340569 (case), Lot Numbers:  23FBN190; \n712) REF DYNJ84288, UDI/DI 10195327348182 (each) 40195327348183 (case), Lot Numbers:  23KME091; \n713) REF DYNJ84967, UDI/DI 10195327441807 (each) 40195327441808 (case), Lot Numbers:  24DBB343; \n714) REF DYNJ85243, UDI/DI 10195327482756 (each) 40195327482757 (case), Lot Numbers:  22GBM717, 22EBA964; \n715) REF DYNJ85467A, UDI/DI 10195327637941 (each) 40195327637942 (case), Lot Numbers:  22JBS880, 22JBU383, 22KBQ468, 23CBN666, 23HBC933, 23EBA463, 24ABJ696, 24BBT055, 23KBO627; \n716) REF DYNJ900023R, UDI/DI 10193489841381 (each) 40193489841382 (case), Lot Numbers:  23CBJ714, 23GBA115; \n717) REF DYNJ900023S, UDI/DI 10195327213640 (each) 40195327213641 (case), Lot Numbers:  22GBV188, 22FBF101, 22EBO298, 22EBB375, 23ABF776, 22LBM184, 23EBT067; \n718) REF DYNJ900268G, UDI/DI 10195327348168 (each) 40195327348169 (case), Lot Numbers:  22GBR251, 22GBR253, 23GBC525, 23ABM532, 23DBN404, 23EBG707, 23EBS012; \n719) REF DYNJ900283K, UDI/DI 10193489915976 (each) 40193489915977 (case), Lot Numbers:  22FBF090, 22JBP203; \n720) REF DYNJ900288G, UDI/DI 10193489915105 (each) 40193489915106 (case), Lot Numbers:  23BBU941, 23EBI554, 23GBL756; \n721) REF DYNJ900302G, UDI/DI 10193489915488 (each) 40193489915489 (case), Lot Numbers:  22FBN690, 22FBN691, 22IBM297, 22GBH609, 22KBD779, 23ABS791, 23ABQ840, 23ABK496, 22LBC403, 23ABM639, 23ABK504, 23FBG452; \n722) REF DYNJ900303M, UDI/DI 10195327175269 (each) 40195327175260 (case), Lot Numbers:  23EBQ520, 23BBS235, 23LBV758, 23KBL260, 24ABR230, 24EBF033, 24EBP422; \n723) REF DYNJ900304N, UDI/DI 10195327175276 (each) 40195327175277 (case), Lot Numbers:  22FBN690, 22FBN691, 22IBM297, 22GBH609, 22KBD779, 23ABS791, 23ABQ840, 23ABK496, 22LBC403, 23ABM639, 23ABK504, 23FBG452; \n724) REF DYNJ900414K, UDI/DI 10195327029432 (each) 40195327029433 (case), Lot Numbers:  23EBQ520, 23BBS235, 23LBV758, 23KBL260, 24ABR230, 24EBF033, 24EBP422; \n725) REF DYNJ900419P, UDI/DI 10195327043773 (each) 40195327043774 (case), Lot Numbers:  22FBY504, 22KBG527, 23ABB690, 23CBE025, 23CBD656, 23JBB780, 23GBG691, 23HBT096, 24ABE589, 23LBF707, 24BBQ926, 24BBB082, 24EBA349, 24DBD222, 24GBL525; \n726) REF DYNJ900565I, UDI/DI 10193489595369 (each) 40193489595360 (case), Lot Numbers:  23ABC040, 22KBJ801, 22LBL182, 23BBL703, 23DBH249; \n727) REF DYNJ900570I, UDI/DI 10193489848670 (each) 40193489848671 (case), Lot Numbers:  23ALA606, 23CLA120, 24DLA275, 23ILA275, 23GLA212, 23LLA791, 23HLA377, 23GLA531, 24FLA281, 24ELA550; \n728) REF DYNJ900606F, UDI/DI 10193489463750 (each) 40193489463751 (case), Lot Numbers:  22FDA713, 22FDA433, 22EDB172; \n729) REF DYNJ900606G, UDI/DI 10195327224097 (each) 40195327224098 (case), Lot Numbers:  22IDB441, 22KDA284, 23BDB270, 23ADB286, 23DDA543; \n730) REF DYNJ900624F, UDI/DI 10193489463743 (each) 40193489463744 (case), Lot Numbers:  22EDB592, 22FDB481, 22FDC080; \n731) REF DYNJ900624G, UDI/DI 10195327224134 (each) 40195327224135 (case), Lot Numbers:  22HDB968, 22KDA295, 23EDA423, 23BDC197, 23ADA511; \n732) REF DYNJ900651C, UDI/DI 10193489436716 (each) 40193489436717 (case), Lot Numbers:  22JBL369, 22IBV324, 22LBG981, 23EBK655, 23CBO965; \n733) REF DYNJ900658B, UDI/DI 10193489436709 (each) 40193489436700 (case), Lot Numbers:  22IBU638, 22DBN187, 22DBN529, 23ABP621, 23EBJ341, 23BBF724; \n734) REF DYNJ900713D, UDI/DI 10193489951318 (each) 40193489951319 (case), Lot Numbers:  22GBS111, 22FBU842, 23DBJ848, 23CBJ961, 23ABC194, 22FBS164, 22LBS197; \n735) REF DYNJ900718C, UDI/DI 10193489951578 (each) 40193489951579 (case), Lot Numbers:  22EBM260, 22HBF175, 23CBS226, 23CBS227, 22KBF981, 23ABM644, 23DBO409; \n736) REF DYNJ900719B, UDI/DI 10193489951585 (each) 40193489951586 (case), Lot Numbers:  23ABQ858, 23ABD345; \n737) REF DYNJ900720D, UDI/DI 10193489951592 (each) 40193489951593 (case), Lot Numbers:  22FBO545, 22JBF689, 22IBN802, 22GBO891, 23ABP512, 22JBY062, 22LBC649, 23DBD541, 23EBJ422; \n738) REF DYNJ900722C, UDI/DI 10193489951615 (each) 40193489951616 (case), Lot Numbers:  22GBS256, 23DBK565, 23CBV883, 23CBS228, 23CBS229, 23CBJ715, 22LBD478, 22KBF311, 23ABP483; \n739) REF DYNJ900745D, UDI/DI 10193489526059 (each) 40193489526050 (case), Lot Numbers:  23FBK849, 23ABR066, 22KBR848, 23EBG717, 23DBD756, 24DBL753, 24ABY468, 24ABS951; \n740) REF DYNJ900747G, UDI/DI 10889942365067 (each) 40889942365068 (case), Lot Numbers:  22JBK492, 23DBD481, 23BBI448, 23ABM357, 23ABD436, 22LBF812, 22KBS011, 23FBV621, 23IBB489, 23EBL072, 24GBX528, 23LBH747, 24ABX872, 24DBM828; \n741) REF DYNJ900748K, UDI/DI 10193489655247 (each) 40193489655248 (case), Lot Numbers:  22JBP204, 22KBS040, 23ABK742, 23EBF972, 23BBI402, 23FBA664, 23EBV896, 23IBD644, 24ABJ166, 23KBH147, 23KBC019, 23GBK063, 24CBU901, 24CBM212, 24EBM861; \n742) REF DYNJ901064F, UDI/DI 10193489944396 (each) 40193489944397 (case), Lot Numbers:  22HMH812, 22HMF271, 22HME031; \n743) REF DYNJ901064I, UDI/DI 10195327232597 (each) 40195327232598 (case), Lot Numbers:  22KMC858, 22JMC315, 23CME948, 23HMD823, 23GMB741, 23FMI200, 23FMC572, 23BMF414, 23BMF424, 23DMF243; \n744) REF DYNJ901069K, UDI/DI 10195327229948 (each) 40195327229949 (case), Lot Numbers:  23AMF005, 23AMA681, 22LMI497, 22LMF749, 22LMB479, 23CMA535, 23BMI246, 23DMB761; \n745) REF DYNJ901074G, UDI/DI 10193489944624 (each) 40193489944625 (case), Lot Numbers:  22HMF139; \n746) REF DYNJ901074J, UDI/DI 10195327230036 (each) 40195327230037 (case), Lot Numbers:  23DMI015; \n747) REF DYNJ901111C, UDI/DI 10195327127503 (each) 40195327127504 (case), Lot Numbers:  22KBI772, 22GBR320, 23CBR809, 23ABM465, 24BBO660, 23JBT819, 23EBL425; \n748) REF DYNJ901748I, UDI/DI 10195327495725 (each) 40195327495726 (case), Lot Numbers:  23LDB126, 23JDA078, 24ADB909, 24DDB346; \n749) REF DYNJ901751Q, UDI/DI 10195327496067 (each) 40195327496068 (case), Lot Numbers:  24ADB894, 24CDB979; \n750) REF DYNJ901752Q, UDI/DI 10195327496173 (each) 40195327496174 (case), Lot Numbers:  24BDB918, 24ADA253, 23JDA606, 23LDA958, 23LDA660, 23JDB757, 24DDA210, 24JDA090, 24FDC008, 24EDA615; \n751) REF DYNJ901755Q, UDI/DI 10195327582418 (each) 40195327582419 (case), Lot Numbers:  24AMD405; \n752) REF DYNJ901764O, UDI/DI 10195327582401 (each) 40195327582402 (case), Lot Numbers:  24AMF214; \n753) REF DYNJ901808I, UDI/DI 10195327496043 (each) 40195327496044 (case), Lot Numbers:  23KDA545, 23JDB826, 24ADA149, 24CDA340, 24DDB799; \n754) REF DYNJ901816I, UDI/DI 10195327495701 (each) 40195327495702 (case), Lot Numbers:  24DDB344, 23LDB492; \n755) REF DYNJ902058C, UDI/DI 10193489359459 (each) 40193489359450 (case), Lot Numbers:  23ABK598, 22JBV808, 23CBC488; \n756) REF DYNJ902058D, UDI/DI 10195327422479 (each) 40195327422470 (case), Lot Numbers:  23KBR583, 23KBG149; \n757) REF DYNJ902058F, UDI/DI 10195327572389 (each) 40195327572380 (case), Lot Numbers:  24CBA332, 24BBK015, 24CBS468; \n758) REF DYNJ902061G, UDI/DI 10195327086756 (each) 40195327086757 (case), Lot Numbers:  22HBD345, 22JBR422, 22KBC742, 23EBF045, 23EBE314, 23BBN957, 23BBB178; \n759) REF DYNJ902061I, UDI/DI 10195327422349 (each) 40195327422340 (case), Lot Numbers:  23KBB124, 23JBL348; \n760) REF DYNJ902061J, UDI/DI 10195327572402 (each) 40195327572403 (case), Lot Numbers:  24DBI718, 24CBK603, 24CBK416; \n761) REF DYNJ902133K, UDI/DI 10195327497248 (each) 40195327497249 (case), Lot Numbers:  24ADB559, 23LDA150, 23JDB901; \n762) REF DYNJ902282I, UDI/DI 10193489983890 (each) 40193489983891 (case), Lot Numbers:  22FDA613, 22DDC497, 22IDB317, 22HDA055; \n763) REF DYNJ902282J, UDI/DI 10195327313319 (each) 40195327313310 (case), Lot Numbers:  23HDA883, 23FDB844, 23EDB115, 23CDB987, 23ADB634, 22LDB600, 23IDB843, 23KDB385; \n764) REF DYNJ902282K, UDI/DI 10195327585174 (each) 40195327585175 (case), Lot Numbers:  24EDA308; \n765) REF DYNJ902313G, UDI/DI 10195327493943 (each) 40195327493944 (case), Lot Numbers:  23KDA051, 23IDB388, 24BDB273, 23LDB168, 24CDB486; \n766) REF DYNJ902365L, UDI/DI 10195327144753 (each) 40195327144754 (case), Lot Numbers:  23ABD120, 23IBH853, 23GBJ208, 23IBP814, 23KBA934; \n767) REF DYNJ902395D, UDI/DI 10195327457556 (each) 40195327457557 (case), Lot Numbers:  23JDB263, 23IDB602; \n768) REF DYNJ902396G, UDI/DI 10195327457570 (each) 40195327457571 (case), Lot Numbers:  23HDB038; \n769) REF DYNJ902699P, UDI/DI 10193489841015 (each) 40193489841016 (case), Lot Numbers:  22EBO528; \n770) REF DYNJ902699Q, UDI/DI 10195327213657 (each) 40195327213658 (case), Lot Numbers:  23HBF074; \n771) REF DYNJ902763G, UDI/DI 10195327461713 (each) 40195327461714 (case), Lot Numbers:  23KBD941, 23KBE318, 24BBG303, 24ABP224; \n772) REF DYNJ902776F, UDI/DI 10195327462116 (each) 40195327462117 (case), Lot Numbers:  24ABW975, 24ABH021; \n773) REF DYNJ902782G, UDI/DI 10195327461690 (each) 40195327461691 (case), Lot Numbers:  23JBB093, 24ABP601, 24BBH057; \n774) REF DYNJ902974N, UDI/DI 10195327147501 (each) 40195327147502 (case), Lot Numbers:  22IDA129; \n775) REF DYNJ902977N, UDI/DI 10195327206673 (each) 40195327206674 (case), Lot Numbers:  22LDB526, 22KDA429; \n776) REF DYNJ902978L, UDI/DI 10195327147525 (each) 40195327147526 (case), Lot Numbers:  23CDA033, 23GDB554; \n777) REF DYNJ902979O, UDI/DI 10195327144340 (each) 40195327144341 (case), Lot Numbers:  22LDA303; \n778) REF DYNJ903053F, UDI/DI 10195327461614 (each) 40195327461615 (case), Lot Numbers:  23IMA506, 23IMH666; \n779) REF DYNJ903053G, UDI/DI 10195327555115 (each) 40195327555116 (case), Lot Numbers:  24AMA676, 24AMF097; \n780) REF DYNJ903053I, UDI/DI 10195327612986 (each) 40195327612987 (case), Lot Numbers:  24CMG800; \n781) REF DYNJ903230F, UDI/DI 10195327461607 (each) 40195327461608 (case), Lot Numbers:  23IMA463, 23JMA077; \n782) REF DYNJ903230G, UDI/DI 10195327555122 (each) 40195327555123 (case), Lot Numbers:  24BMD681, 24AMG423; \n783) REF DYNJ903534D, UDI/DI 10195327461638 (each) 40195327461639 (case), Lot Numbers:  23IMH904, 23HMG998, 23IMF613; \n784) REF DYNJ903561C, UDI/DI 10195327348212 (each) 40195327348213 (case), Lot Numbers:  23HBN852; \n785) REF DYNJ903606C, UDI/DI 10193489959413 (each) 40193489959414 (case), Lot Numbers:  22GBT643; \n786) REF DYNJ903606D, UDI/DI 10195327237592 (each) 40195327237593 (case), Lot Numbers:  22LBH046, 22KBP516, 22KBJ512, 22JBP030; \n787) REF DYNJ903606F, UDI/DI 10195327368425 (each) 40195327368426 (case), Lot Numbers:  23EBH942, 23HBM044, 23GBN295, 24ABX857, 23JBU718, 24BBQ150, 23FBS614; \n788) REF DYNJ903606G, UDI/DI 10195327658205 (each) 40195327658206 (case), Lot Numbers:  24DBK732; \n789) REF DYNJ903607G, UDI/DI 10195327373238 (each) 40195327373239 (case), Lot Numbers:  23IBW024, 24DBN828, 23LBD696; \n790) REF DYNJ903608C, UDI/DI 10195327237615 (each) 40195327237616 (case), Lot Numbers:  23ABE687, 23ABC300; \n791) REF DYNJ903608D, UDI/DI 10195327368401 (each) 40195327368402 (case), Lot Numbers:  23JBN509, 24ABT610, 24BBD902; \n792) REF DYNJ903610C, UDI/DI 10195327237479 (each) 40195327237470 (case), Lot Numbers:  22KBC036, 22KBA216, 22LBS792, 23BBB781; \n793) REF DYNJ903610F, UDI/DI 10195327364441 (each) 40195327364442 (case), Lot Numbers:  23DBR784; \n794) REF DYNJ903610G, UDI/DI 10195327400927 (each) 40195327400928 (case), Lot Numbers:  24CBG839, 23JBE453, 23JBD275, 23HBU311, \n795) REF DYNJ903611B, UDI/DI 10888277761667 (each) 40888277761668 (case), Lot Numbers:  22GBP110; \n796) REF DYNJ903611C, UDI/DI 10195327237486 (each) 40195327237487 (case), Lot Numbers:  23BBO734, 22KBL171, 22JBO970, 23ABA728; \n797) REF DYNJ903611D, UDI/DI 10195327344733 (each) 40195327344734 (case), Lot Numbers:  23CBD751; \n798) REF DYNJ903611G, UDI/DI 10195327400842 (each) 40195327400843 (case), Lot Numbers:  23KBJ499, 24CBF351, 23IBK710, 23IBK768, 24ABS861; \n799) REF DYNJ903611I, UDI/DI 10195327648237 (each) 40195327648238 (case), Lot Numbers:  24DBR209, 24DBR261; \n800) REF DYNJ903613G, UDI/DI 10195327400835 (each) 40195327400836 (case), Lot Numbers:  23HBT724, 24BBQ921, 24EBI139, 24EBI392, 24BBC917; \n801) REF DYNJ903620F, UDI/DI 10195327323639 (each) 40195327323630 (case), Lot Numbers:  23BBQ238; \n802) REF DYNJ903620G, UDI/DI 10195327364410 (each) 40195327364411 (case), Lot Numbers:  24ABS999, 23IBQ541; \n803) REF DYNJ903621F, UDI/DI 10195327237585 (each) 40195327237586 (case), Lot Numbers:  23BBN895, 22JBR475; \n804) REF DYNJ903621G, UDI/DI 10195327373092 (each) 40195327373093 (case), Lot Numbers:  23EBH910, 23IBO529, 23HBL684, 24ABR045; \n805) REF DYNJ903621I, UDI/DI 10195327648244 (each) 40195327648245 (case), Lot Numbers:  24DBC507; \n806) REF DYNJ903763K, UDI/DI 10195327217655 (each) 40195327217656 (case), Lot Numbers: 22KBP183;  \n807) REF DYNJ903763L, UDI/DI 10195327334406 (each) 40195327334407 (case), Lot Numbers:  23BBS389; \n808) REF DYNJ903763N, UDI/DI 10195327467463 (each) 40195327467464 (case), Lot Numbers:  23IBF008, 23HBH072, 23KBW125, 23JBH997, 24BBI400, 24ABO355, 24DBF357; \n809) REF DYNJ903763O, UDI/DI 10195327675837 (each) 40195327675838 (case), Lot Numbers:  24DBQ302, 24DBQ917; \n810) REF DYNJ903764J, UDI/DI 10195327217518 (each) 40195327217519 (case), Lot Numbers:  23ABL498, 23BBB953; \n811) REF DYNJ903766I, UDI/DI 10195327217686 (each) 40195327217687 (case), Lot Numbers:  23CBG888, 23BBG549, 23ABN709, 23ABN187, 23EBQ535, 24BBD230, 24CBM973; \n812) REF DYNJ903766J, UDI/DI 10195327675943 (each) 40195327675944 (case), Lot Numbers:  24DBQ916; \n813) REF DYNJ903800I, UDI/DI 10193489274479 (each) 40193489274470 (case), Lot Numbers:  22LBF915, 23CBI550, 22LBB553, 22LBB549, 23CBA567, 23HBT725, 24ABB989, 23KBT454, 23JBT594, 23GBF923, 24BBG975, 24ABT474, 24CBA746, 24BBK582; \n814) REF DYNJ903800J, UDI/DI 10195327640682 (each) 40195327640683 (case), Lot Numbers:  24DBE535; \n815) REF DYNJ903801D, UDI/DI 10193489274974 (each) 40193489274975 (case), Lot Numbers:  22FBG658, 22HBS607, 23HBN657, 23BBH118, 23CBE017, 23IBG494, 23JBQ332, 24BBK583, 23LBN311; \n816) REF DYNJ903801F, UDI/DI 10195327640675 (each) 40195327640676 (case), Lot Numbers:  24DBC481, 24EBE301; \n817) REF DYNJ903802D, UDI/DI 10193489275582 (each) 40193489275583 (case), Lot Numbers:  23DBE028, 22JBW593, 23ABB749, 23FBV467, 23EBS008, 24ABL803, 23LBF498; \n818) REF DYNJ903802F, UDI/DI 10195327640668 (each) 40195327640669 (case), Lot Numbers:  24CBP664; \n819) REF DYNJ903803C, UDI/DI 10193489274981 (each) 40193489274982 (case), Lot Numbers:  22GBH541, 22GBU423, 22LBS208, 22KBE702, 23ABQ937, 22JBW280; \n820) REF DYNJ903863C, UDI/DI 10193489232974 (each) 40193489232975 (case), Lot Numbers:  22IMF361, 22IMC641, 22DMA800; \n821) REF DYNJ903863D, UDI/DI 10195327236632 (each) 40195327236633 (case), Lot Numbers:  22LMD818, 23CMF085, 23CMF358, 23CMB373, 23GMC259, 23HMA303; \n822) REF DYNJ903870D, UDI/DI 10193489415971 (each) 40193489415972 (case), Lot Numbers:  22IBB541, 22EBL327, 22HBL344, 22HBM583; \n823) REF DYNJ904084F, UDI/DI 10193489282849 (each) 40193489282840 (case), Lot Numbers:  22HDB932, 22JDB404, 23ADA882, 22KDB333, 22LDA036; \n824) REF DYNJ904109F, UDI/DI 10193489841404 (each) 40193489841405 (case), Lot Numbers:  23GBP508, 23EBC803, 23DBG140, 23KBG241; \n825) REF DYNJ904113L, UDI/DI 10193489841367 (each) 40193489841368 (case), Lot Numbers:  22GBI998, 22GBI999; \n826) REF DYNJ904113N, UDI/DI 10195327272043 (each) 40195327272044 (case), Lot Numbers:  23FBM599, 23FBF556, 22LBL190; \n827) REF DYNJ904205G, UDI/DI 10195327450816 (each) 40195327450817 (case), Lot Numbers:  23HDC122; \n828) REF DYNJ904206G, UDI/DI 10195327450823 (each) 40195327450824 (case), Lot Numbers:  23HDC360, 23IDA118; \n829) REF DYNJ904209I, UDI/DI 10195327450830 (each) 40195327450831 (case), Lot Numbers:  23HDC084, 23KDA682, 23JDA024; \n830) REF DYNJ904210I, UDI/DI 10195327450847 (each) 40195327450848 (case), Lot Numbers:  23GDC141, 23IDA584; \n831) REF DYNJ904399F, UDI/DI 10195327545536 (each) 40195327545537 (case), Lot Numbers:  24AMD530; \n832) REF DYNJ904475K, UDI/DI 10195327186302 (each) 40195327186303 (case), Lot Numbers:  23ABL097; \n833) REF DYNJ904480K, UDI/DI 10195327186456 (each) 40195327186457 (case), Lot Numbers:  22KBC062, 23ABG264, 22LBT100; \n834) REF DYNJ904481J, UDI/DI 10195327186364 (each) 40195327186365 (case), Lot Numbers:  22GBG217, 22HBO530, 22HBD285; \n835) REF DYNJ904482F, UDI/DI 10195327165888 (each) 40195327165889 (case), Lot Numbers:  22GBA664; \n836) REF DYNJ904482G, UDI/DI 10195327186500 (each) 40195327186501 (case), Lot Numbers:  22FBU445, 22GBZ137; \n837) REF DYNJ904487J, UDI/DI 10195327186395 (each) 40195327186396 (case), Lot Numbers:  22KBS411; \n838) REF DYNJ904489I, UDI/DI 10195327166212 (each) 40195327166213 (case), Lot Numbers:  22FBO367; \n839) REF DYNJ904497J, UDI/DI 10195327186227 (each) 40195327186228 (case), Lot Numbers:  22FBY372, 22HBD286; \n840) REF DYNJ904498K, UDI/DI 10195327186234 (each) 40195327186235 (case), Lot Numbers:  23ABH742, 22GBT218; \n841) REF DYNJ904589F, UDI/DI 10195327497361 (each) 40195327497362 (case), Lot Numbers:  24ADB251; \n842) REF DYNJ904648F, UDI/DI 10193489931228 (each) 40193489931229 (case), Lot Numbers:  22EBJ527, 22FBL680, 22JBB373; \n843) REF DYNJ904648G, UDI/DI 10195327277833 (each) 40195327277834 (case), Lot Numbers:  22KBP499, 23ABE040, 23CBJ375; \n844) REF DYNJ904698F, UDI/DI 10195327169282 (each) 40195327169283 (case), Lot Numbers:  23CDC646, 22IDB991, 23EDA530; \n845) REF DYNJ904702G, UDI/DI 10195327169312 (each) 40195327169313 (case), Lot Numbers:  22GDB187, 22HDA865, 23ADA273, 22LDA441, 23DDC167, 23CDB936, 23BDB141; \n846) REF DYNJ904703F, UDI/DI 10195327169268 (each) 40195327169269 (case), Lot Numbers:  22GDB058, 22FDA909, 22IDA556, 22JDB907, 23DDB953, 23CDB132, 23CDA593, 23BDA147; \n847) REF DYNJ904704D, UDI/DI 10195327126247 (each) 40195327126248 (case), Lot Numbers:  22GDB156, 22LDA718, 23DDA759; \n848) REF DYNJ904765B, UDI/DI 10195327010485 (each) 40195327010486 (case), Lot Numbers:  22GBQ785, 22GBK512, 23CBC561, 23JBO143, 23IBT377; \n849) REF DYNJ904765C, UDI/DI 10195327535742 (each) 40195327535743 (case), Lot Numbers:  23KBV823; \n850) REF DYNJ904860C, UDI/DI 10195327124793 (each) 40195327124794 (case), Lot Numbers:  22HMD644; \n851) REF DYNJ904860D, UDI/DI 10195327222536 (each) 40195327222537 (case), Lot Numbers:  22JMC904, 22LMA059, 22KMG895, 22KMF847, 23HMH067, 23FMI126, 23CMF389, 23CMF009, 23CMD418, 23IMD317, 24AMA086, 23KMI863, 23KME314, 24AMJ020; \n852) REF DYNJ904906G, UDI/DI 10195327065164 (each) 40195327065165 (case), Lot Numbers:  22FBN681, 22FBN682, 22FBS643, 22IBA018; \n853) REF DYNJ904906I, UDI/DI 10195327247805 (each) 40195327247806 (case), Lot Numbers:  22KBT768, 23DBD885, 23FBQ980, 23FBP020, 23HBX774; \n854) REF DYNJ904993L, UDI/DI 10195327186272 (each) 40195327186273 (case), Lot Numbers:  22IBO970; \n855) REF DYNJ904995G, UDI/DI 10195327165680 (each) 40195327165681 (case), Lot Numbers:  22FDA708; \n856) REF DYNJ904996I, UDI/DI 10195327190286 (each) 40195327190287 (case), Lot Numbers:  23ABI076; \n857) REF DYNJ905075F, UDI/DI 10195327045302 (each) 40195327045303 (case), Lot Numbers:  22FBE945, 22GBE231, 22KBM325, 22JBH247; \n858) REF DYNJ905075G, UDI/DI 10195327233839 (each) 40195327233830 (case), Lot Numbers:  22KBI003, 22KBI004, 22LBH034, 23DBD746, 23DBA929, 23FBG302; \n859) REF DYNJ905078F, UDI/DI 10193489437768 (each) 40193489437769 (case), Lot Numbers:  22NBF134, 22GBB090, 22KBH368, 22JBY286; \n860) REF DYNJ905080G, UDI/DI 10195327046552 (each) 40195327046553 (case), Lot Numbers:  23ABA878, 23EBS459; \n861) REF DYNJ905112B, UDI/DI 10193489887624 (each) 40193489887625 (case), Lot Numbers:  22EDB438, 22FDB733; \n862) REF DYNJ905112C, UDI/DI 10195327224158 (each) 40195327224159 (case), Lot Numbers:  22JDA094, 23ADA503, 22KDA336, 23BDB721; \n863) REF DYNJ905196I, UDI/DI 10193489466744 (each) 40193489466745 (case), Lot Numbers:  22HME748, 22GMF184, 22FMA081; \n864) REF DYNJ905196J, UDI/DI 10195327230821 (each) 40195327230822 (case), Lot Numbers:  22JMD693, 23AMI844, 23AME521, 22JMF861, 23DMB374, 22LMB547, 23CMD438; \n865) REF DYNJ905334C, UDI/DI 10195327638016 (each) 40195327638017 (case), Lot Numbers:  24EBG630, 24CBO936, 24HBO616, 24GBH746; \n866) REF DYNJ905443C, UDI/DI 10193489883688 (each) 40193489883689 (case), Lot Numbers:  22JBB831, 22IBN307, 22GBY782, 22GBK088, 22FBP816, 22DBO430, 23BBA885, 22KBV220, 23EBE319, 23CBO783;\n867) REF DYNJ905464G, UDI/DI 10193489841022 (each) 40193489841023 (case), Lot Numbers:  22GBM716, 22FBV950; \n868) REF DYNJ905464I, UDI/DI 10195327217549 (each) 40195327217540 (case), Lot Numbers:  23DBK508, 23CBP055, 23GBU897, 23EBN913, 23IBW052, 23IBQ658, 23HBJ896, 23JBU586; \n869) REF DYNJ905469C, UDI/DI 10195327200282 (each) 40195327200283 (case), Lot Numbers:  22KBR546, 22KBR551, 23CBV484, 23EBB489, 23JBI044, 23KBR699, 23KBR556, 24BBT857; \n870) REF DYNJ905471B, UDI/DI 10195327200510 (each) 40195327200511 (case), Lot Numbers:  23DBI548, 22KBR568, 22KBR569, 23HBY031, 24BBK748, 23LBN256, 23JBP241, 23JBF443; \n871) REF DYNJ905604F, UDI/DI 10195327066550 (each) 40195327066551 (case), Lot Numbers:  22HBK220, 22GBW114, 22GBW115, 22KBV697; \n872) REF DYNJ905604G, UDI/DI 10195327329662 (each) 40195327329663 (case), Lot Numbers:  23EBF189, 23CBP671; \n873) REF DYNJ905605F, UDI/DI 10195327066567 (each) 40195327066568 (case), Lot Numbers:  22EBG716, 22EBI074, 22KBO825, 23BBA688; \n874) REF DYNJ905605G, UDI/DI 10195327329679 (each) 40195327329670 (case), Lot Numbers:  23FBF791; \n875) REF DYNJ905617F, UDI/DI 10195327675851 (each) 40195327675852 (case), Lot Numbers:  24EBG331; \n876) REF DYNJ905813D, UDI/DI 10195327094188 (each) 40195327094189 (case), Lot Numbers:  23CDC412, 23ADC095, 23IDB681, 23HDB420, 23LDA261, 23KDA402, 24ADB902; \n877) REF DYNJ905814F, UDI/DI 10195327129606 (each) 40195327129607 (case), Lot Numbers:  22JDB123, 22GDA019, 22KDC152, 23DDB941, 23BDB252, 23GDA260, 23IDB391, 23HDB799, 24BDB864, 24ADA055; \n878) REF DYNJ905974A, UDI/DI 10195327037727 (each) 40195327037728 (case), Lot Numbers:  22DBU738; \n879) REF DYNJ905974B, UDI/DI 10195327311315 (each) 40195327311316 (case), Lot Numbers:  23GBO163, 24BBE649, 23KBG150, 23IBL206; \n880) REF DYNJ905978A, UDI/DI 10195327030629 (each) 40195327030620 (case), Lot Numbers:  23ABS782, 23ABQ824, 22GBO887, 23ABM782, 23FBR258, 23EBN920; \n881) REF DYNJ905992D, UDI/DI 10195327451301 (each) 40195327451302 (case), Lot Numbers:  23GBS854, 23HBX882; \n882) REF DYNJ905992F, UDI/DI 10195327527211 (each) 40195327527212 (case), Lot Numbers:  24BBB309, 24ABV739, 24DBN864; \n883) REF DYNJ906053D, UDI/DI 10193489984576 (each) 40193489984577 (case), Lot Numbers:  22DDC498; \n884) REF DYNJ906053G, UDI/DI 10195327293062 (each) 40195327293063 (case), Lot Numbers:  22KDB741; \n885) REF DYNJ906053I, UDI/DI 10195327313333 (each) 40195327313334 (case), Lot Numbers:  23CDA692, 23EDB002, 23ADA201; \n886) REF DYNJ906053J, UDI/DI 10195327518233 (each) 40195327518234 (case), Lot Numbers:  23KDC354, 24ADA141, 24BDA491; \n887) REF DYNJ906161A, UDI/DI 10193489233278 (each) 40193489233279 (case), Lot Numbers:  22JLA637, 23FLA345, 22LLA948, 22LLA588, 23CLB249, 23BLA914, 23ILA274, 23GLA749, 23LLA802, 23KLA293, 23KLA415, 23JLA817, 23LLA905, 24ALA873, 24CLA142, 24ELA833; \n888) REF DYNJ906163B, UDI/DI 10193489819410 (each) 40193489819411 (case), Lot Numbers:  22FBV613, 22HBS633, 22KBM272, 23BBE573, 23LBA471, 23FBN758, 23EBL829, 23KBK823, 23GBE754, 24CBA354, 24BBF086; \n889) REF DYNJ906166A, UDI/DI 10195327036607 (each) 40195327036608 (case), Lot Numbers:  22KBT481, 23EBT073, 23BBQ295, 23BBG458, 23DBM963, 23IBK707, 24CBP673, 24ABQ967, 24EBC230; \n890) REF DYNJ906168C, UDI/DI 10195327120658 (each) 40195327120659 (case), Lot Numbers:  23FBD132, 23BBP552, 23IBM223; \n891) REF DYNJ906223B, UDI/DI 10195327344405 (each) 40195327344406 (case), Lot Numbers:  23CBX460, 23CBI971; \n892) REF DYNJ906261C, UDI/DI 10195327344436 (each) 40195327344437 (case), Lot Numbers:  23DBE791; \n893) REF DYNJ906262C, UDI/DI 10195327344368 (each) 40195327344369 (case), Lot Numbers:  23FBR888; \n894) REF DYNJ906317G, UDI/DI 10195327370909 (each) 40195327370900 (case), Lot Numbers:  23JBT885, 23IBP592, 23GBK296, 23GBA476; \n895) REF DYNJ906409A, UDI/DI 10193489874853 (each) 40193489874854 (case), Lot Numbers:  22KBN513, 22FBS708; \n896) REF DYNJ906409C, UDI/DI 10195327319809 (each) 40195327319800 (case), Lot Numbers:  23KBC002; \n897) REF DYNJ906448C, UDI/DI 10195327167578 (each) 40195327167579 (case), Lot Numbers:  23BBR525, 22GBD044, 22GBD045, 23ABQ954, 22LBL552; \n898) REF DYNJ906448F, UDI/DI 10195327436421 (each) 40195327436422 (case), Lot Numbers:  23JBH738, 23KBC876; \n899) REF DYNJ906450C, UDI/DI 10193489900439 (each) 40193489900430 (case), Lot Numbers:  22EBB644, 22EBC235; \n900) REF DYNJ906450D, UDI/DI 10195327170660 (each) 40195327170661 (case), Lot Numbers:  23BBO821; \n901) REF DYNJ906450F, UDI/DI 10195327348656 (each) 40195327348657 (case), Lot Numbers:  23CBP017, 23HBM407, 23KBA715; \n902) REF DYNJ906451C, UDI/DI 10195327167448 (each) 40195327167449 (case), Lot Numbers:  22LBQ816; \n903) REF DYNJ906451D, UDI/DI 10195327348649 (each) 40195327348640 (case), Lot Numbers:  23FBV262, 23JBA209; \n904) REF DYNJ906457C, UDI/DI 10195327167561 (each) 40195327167562 (case), Lot Numbers:  22KBV019, 23ABR027, 23CBW474; \n905) REF DYNJ906457D, UDI/DI 10195327436650 (each) 40195327436651 (case), Lot Numbers:  23IBD447; \n906) REF DYNJ906458C, UDI/DI 10193489900200 (each) 40193489900201 (case), Lot Numbers:  22EBF057, 22FBQ883; \n907) REF DYNJ906458D, UDI/DI 10195327167462 (each) 40195327167463 (case), Lot Numbers:  22GBN080, 22FBV564, 23ABQ884, 22KBW787, 23BBQ294; \n908) REF DYNJ906458F, UDI/DI 10195327355883 (each) 40195327355884 (case), Lot Numbers:  23GBU813, 23KBW142, 23KBW056, 23IBF302, 24BBP102, 24GBR743, 24CBO485; \n909) REF DYNJ906469D, UDI/DI 10195327041731 (each) 40195327041732 (case), Lot Numbers:  22DDA747, 22FDA670; \n910) REF DYNJ906469F, UDI/DI 10195327210939 (each) 40195327210930 (case), Lot Numbers:  22HDC339, 23IDC208; \n911) REF DYNJ906479C, UDI/DI 10195327024307 (each) 40195327024308 (case), Lot Numbers:  22HDB405, 22GDA236; \n912) REF DYNJ906479D, UDI/DI 10195327202682 (each) 40195327202683 (case), Lot Numbers:  23DDB574, 23FDB867, 23FDB142, 23HDC215, 23JDB994, 23JDB288; \n913) REF DYNJ906485D, UDI/DI 10195327025250 (each) 40195327025251 (case), Lot Numbers:  22GDB445; \n914) REF DYNJ906485F, UDI/DI 10195327202767 (each) 40195327202768 (case), Lot Numbers:  23EDA580, 23CDB072, 23JDB048; \n915) REF DYNJ906495F, UDI/DI 10195327210977 (each) 40195327210978 (case), Lot Numbers:  23GDA783, 23FDB527; \n916) REF DYNJ906496F, UDI/DI 10195327370954 (each) 40195327370955 (case), Lot Numbers:  23GLA582, 23FLA898, 23JLA675, 23ILA204, 23ILA505, 23ILA840; \n917) REF DYNJ906539F, UDI/DI 10195327427412 (each) 40195327427413 (case), Lot Numbers:  23GBE724; \n918) REF DYNJ906542A, UDI/DI 10193489270020 (each) 40193489270021 (case), Lot Numbers:  22HDB047; \n919) REF DYNJ906542B, UDI/DI 10195327221874 (each) 40195327221875 (case), Lot Numbers:  23FDA462, 24GDA418; \n920) REF DYNJ906602C, UDI/DI 10193489466768 (each) 40193489466769 (case), Lot Numbers:  22GME256, 22FMD475; \n921) REF DYNJ906602D, UDI/DI 10195327230814 (each) 40195327230815 (case), Lot Numbers:  23DME802, 23BMG010, 23BMB203, 22LMB267, 23GMG878; \n922) REF DYNJ906614B, UDI/DI 10193489345322 (each) 40193489345323 (case), Lot Numbers:  22FBS243, 23ABD346, 23ABQ836, 23LBH022, 23IBV173, 23IBV010, 23GBC569, 23GBD902, 23KBP690, 23KBP510; \n923) REF DYNJ906651F, UDI/DI 10193489992342 (each) 40193489992343 (case), Lot Numbers:  22LDB126, 22KDC114; \n924) REF DYNJ906658F, UDI/DI 10195327000899 (each) 40195327000890 (case), Lot Numbers:  22JDA648, 22KDA427, 22LDA989; \n925) REF DYNJ906660G, UDI/DI 10195327070038 (each) 40195327070039 (case), Lot Numbers:  22IDB435, 22GDB475, 22JDB999, 22KDC189, 23CDC432; \n926) REF DYNJ906664G, UDI/DI 10195327070175 (each) 40195327070176 (case), Lot Numbers:  22LDB485, 23BDA224, 23GDA475, 23FDA867; \n927) REF DYNJ906665G, UDI/DI 10195327092481 (each) 40195327092482 (case), Lot Numbers:  22EDB767, 22EDA992, 22CDC258, 22FDA817, 22LDA149, 22IDA259, 22GDB555, 23ADB845; \n928) REF DYNJ906666J, UDI/DI 10195327129590 (each) 40195327129591 (case), Lot Numbers:  22GDA384, 22FDA952, 22FDA056, 22HDA321, 23ABR423, 23ABB003, 22KBC065, 23CBA140, 23ABR422, 22GDB274; \n929) REF DYNJ906723A, UDI/DI 10195327059866 (each) 40195327059867 (case), Lot Numbers:  22DDC232, 22FDA614; \n930) REF DYNJ906890B, UDI/DI 10195327303587 (each) 40195327303588 (case), Lot Numbers:  23FBO221, 23KBT243, 23KBE431, 23CBP900; \n931) REF DYNJ906891C, UDI/DI 10195327303570 (each) 40195327303571 (case), Lot Numbers:  23CBP901, 24DBS776, 24CBN224; \n932) REF DYNJ906892A, UDI/DI 10195327303426 (each) 40195327303427 (case), Lot Numbers:  23DBQ690, 23HBN943, 23KBG736, 23GBV420, 23KBB202; \n933) REF DYNJ906893D, UDI/DI 10195327303372 (each) 40195327303373 (case), Lot Numbers:  23IMC078; \n934) REF DYNJ906894A, UDI/DI 10195327303365 (each) 40195327303366 (case), Lot Numbers:  23HLB075, 23ELA244, 23CLA420, 23BLA206; \n935) REF DYNJ906911A, UDI/DI 10195327303594 (each) 40195327303595 (case), Lot Numbers:  23CBH442, 23ABC522, 23FBV559, 23FBC212, 23HBS918; \n936) REF DYNJ906950B, UDI/DI 10195327381011 (each) 40195327381012 (case), Lot Numbers:  23HBB790, 23EBK134, 23KBR567, 23KBI260, 23LBB806; \n937) REF DYNJ906950C, UDI/DI 10195327566807 (each) 40195327566808 (case), Lot Numbers:  24ABW981, 24DBD940, 24ABL051; \n938) REF DYNJ906955D, UDI/DI 10195327381141 (each) 40195327381142 (case), Lot Numbers:  23HBB476, 23KBQ769, 23KBP472, 23LBB805, 23FBC881, 23IBP953; \n939) REF DYNJ906956C, UDI/DI 10195327381158 (each) 40195327381159 (case), Lot Numbers:  23HBC932, 23EBG857, 23EBC449, 23KBP686, 23KBK837, 23LBP373, 23FBI132; \n940) REF DYNJ906956D, UDI/DI 10195327567125 (each) 40195327567126 (case), Lot Numbers:  24ABH035, 24BBC929, 24DBP220; \n941) REF DYNJ906957B, UDI/DI 10195327381165 (each) 40195327381166 (case), Lot Numbers:  23FBI110, 23KBO184, 23KBB084, 23EBM815; \n942) REF DYNJ906957C, UDI/DI 10195327567132 (each) 40195327567133 (case), Lot Numbers:  24CBM172, 24BBJ355, 24ABH098, 24DBN816; \n943) REF DYNJ906974F, UDI/DI 10195327502881 (each) 40195327502882 (case), Lot Numbers:  23KBB168, 23JBU563, 23LBS348; \n944) REF DYNJ906974I, UDI/DI 10195327621414 (each) 40195327621415 (case), Lot Numbers:  24GBH723, 24CBM366, 24DBG769; \n945) REF DYNJ906975D, UDI/DI 10195327502874 (each) 40195327502875 (case), Lot Numbers:  23LLA276, 23KLA951; \n946) REF DYNJ907022C, UDI/DI 10193489846577 (each) 40193489846578 (case), Lot Numbers:  22GBM076, 22HBS352, 22EBA303, 23BBL462, 23ABI384, 23EBB693, 23DBH227; \n947) REF DYNJ907038D, UDI/DI 10195327228163 (each) 40195327228164 (case), Lot Numbers:  22IBG109, 22IBG110, 23ABL541, 23ABB862, 22KBI775, 22JBX889; \n948) REF DYNJ907038F, UDI/DI 10195327372682 (each) 40195327372683 (case), Lot Numbers:  23EBJ311, 23CBX587, 23CBQ477, 23FBO835; \n949) REF DYNJ907222B, UDI/DI 10195327324278 (each) 40195327324279 (case), Lot Numbers:  23EMD890, 23EMB435; \n950) REF DYNJ907227C, UDI/DI 10195327145378 (each) 40195327145379 (case), Lot Numbers:  22FBO226, 22FBI627; \n951) REF DYNJ907227D, UDI/DI 10195327247430 (each) 40195327247431 (case), Lot Numbers:  22JBR773, 23FBF615, 23ABC457, 23KBF400, 23KBB212, 23LBO474, 24FMI517; \n952) REF DYNJ907249B, UDI/DI 10195327138585 (each) 40195327138586 (case), Lot Numbers:  22KME146, 22JMI139, 23CMG071, 23BMD847, 23BMG030, 23BMF499, 23GMC765, 23CMC866; \n953) REF DYNJ907249C, UDI/DI 10195327486037 (each) 40195327486038 (case), Lot Numbers:  23LMH533, 23KMD710, 24AMH722, 24CMD208, 24DMB386, 24EMG375; \n954) REF DYNJ907252A, UDI/DI 10193489872484 (each) 40193489872485 (case), Lot Numbers:  22GBS302, 22DBH380, 22DBJ406, 22JBR774, 22LBC149, 23DBM247, 23EBP777, 23KBG206, 23HBF251, 23LBG693; \n955) REF DYNJ907253A, UDI/DI 10193489879506 (each) 40193489879507 (case), Lot Numbers:  22GBV712, 22GBR355, 22GBK981, 22DBG367, 22DBF377, 22LBP264, 23DBL869, 23EBQ837, 23EBR132, 23GBG859, 23LBM201, 23LBG921, 23KBF383; \n956) REF DYNJ907330F, UDI/DI 10195327345471 (each) 40195327345472 (case), Lot Numbers:  23JBL940, 23IBQ659, 23KBT044; \n957) REF DYNJ907331I, UDI/DI 10195327345518 (each) 40195327345519 (case), Lot Numbers:  23CBH324, 23FBF070, 23JBH942; \n958) REF DYNJ907394B, UDI/DI 10195327359225 (each) 40195327359226 (case), Lot Numbers:  23DMH233, 23EMD892; \n959) REF DYNJ907398A, UDI/DI 10195327324384 (each) 40195327324385 (case), Lot Numbers:  23EME758; \n960) REF DYNJ907403A, UDI/DI 10195327359218 (each) 40195327359219 (case), Lot Numbers:  23GMH562, 23EMC989; \n961) REF DYNJ907412A, UDI/DI 10195327335731 (each) 40195327335732 (case), Lot Numbers:  23DMF107, 23EME405; \n962) REF DYNJ907419A, UDI/DI 10195327358754 (each) 40195327358755 (case), Lot Numbers:  23FMD834, 23FMA440, 23DME759; \n963) REF DYNJ907420A, UDI/DI 10195327358761 (each) 40195327358762 (case), Lot Numbers:  23DMF108, 23FME339, 23EMK083, 23HMA290; \n964) REF DYNJ907487F, UDI/DI 10195327515560 (each) 40195327515561 (case), Lot Numbers:  23KBW901; \n965) REF DYNJ907516G, UDI/DI 10195327345532 (each) 40195327345533 (case), Lot Numbers:  23DBD216, 23JBE155; \n966) REF DYNJ907516I, UDI/DI 10198459059469 (each) 40198459059460 (case), Lot Numbers:  24HBU187, 24GBR656; \n967) REF DYNJ907526F, UDI/DI 10195327345501 (each) 40195327345502 (case), Lot Numbers:  23JBA742; \n968) REF DYNJ907527G, UDI/DI 10195327345327 (each) 40195327345328 (case), Lot Numbers:  23DBN688, 23IBA261; \n969) REF DYNJ907528F, UDI/DI 10195327345334 (each) 40195327345335 (case), Lot Numbers:  23EBA831, 23GBI252, 23HBX983, 23JBS791; \n970) REF DYNJ907581A, UDI/DI 10195327045296 (each) 40195327045297 (case), Lot Numbers:  22GBD237, 22IBU694, 23CBM613; \n971) REF DYNJ907588I, UDI/DI 10195327345341 (each) 40195327345342 (case), Lot Numbers:  23CBJ928, 23DBP536, 23FBN592, 23FBN449, 23IBB003, 23KBG198; \n972) REF DYNJ907788C, UDI/DI 10195327141059 (each) 40195327141050 (case), Lot Numbers:  22HBA124; \n973) REF DYNJ907788F, UDI/DI 10195327273057 (each) 40195327273058 (case), Lot Numbers:  23CBV889, 23DBD546; \n974) REF DYNJ907790B, UDI/DI 10193489999273 (each) 40193489999274 (case), Lot Numbers:  22GBA671; \n975) REF DYNJ907790C, UDI/DI 10195327185305 (each) 40195327185306 (case), Lot Numbers:  22HBS804, 23EBT254, 23CBX589; \n976) REF DYNJ907790D, UDI/DI 10195327431655 (each) 40195327431656 (case), Lot Numbers:  23GBH340; \n977) REF DYNJ908005, UDI/DI 10193489960372 (each) 40193489960373 (case), Lot Numbers:  22GBX610, 22HBS815, 23ABM610, 23DBE071; \n978) REF DYNJ908009A, UDI/DI 10195327113674 (each) 40195327113675 (case), Lot Numbers:  22FMD128, 22FMG824; \n979) REF DYNJ908009B, UDI/DI 10195327257347 (each) 40195327257348 (case), Lot Numbers:  23FMH537, 23DMC858, 23DMB913, 23JMD035; \n980) REF DYNJ908063A, UDI/DI 10195327125554 (each) 40195327125555 (case), Lot Numbers:  23DBD691, 23EBA988, 23EBN881, 23EBC773; \n981) REF DYNJ908067A, UDI/DI 10195327125721 (each) 40195327125722 (case), Lot Numbers:  22HBH011, 22FBW369, 22FBW370, 22FBO644, 22EBO471, 23DBE108; \n982) REF DYNJ908070A, UDI/DI 10195327125752 (each) 40195327125753 (case), Lot Numbers:  22EBM394, 22FBG080, 22EBU045, 22EBN035, 23ABM724; \n983) REF DYNJ908071A, UDI/DI 10195327125769 (each) 40195327125760 (case), Lot Numbers:  22EBJ761, 22EBJ762, 23DLA327, 23ELA964, 23HLA995, 23FLA953; \n984) REF DYNJ908073, UDI/DI 10193489967869 (each) 40193489967860 (case), Lot Numbers:  22KBE178, 22KBP162, 23CBM743, 23ABJ520, 23GBH156; \n985) REF DYNJ908173C, UDI/DI 10195327345457 (each) 40195327345458 (case), Lot Numbers:  23CBC564, 23FBV454, 23FBN722, 23HBY069, 23JBG923; \n986) REF DYNJ908256, UDI/DI 10195327016159 (each) 40195327016150 (case), Lot Numbers:  22KBG059, 22KBF265, 23ABM691; \n987) REF DYNJ908290A, UDI/DI 10195327216610 (each) 40195327216611 (case), Lot Numbers:  22JMB449; \n988) REF DYNJ908292, UDI/DI 10195327029289 (each) 40195327029280 (case), Lot Numbers:  23BBS142, 23ABP750, 22JBU899, 23FBF303, 23EBF157, 23EBC430, 24ABG432, 23KBJ003, 23JBV879, 24BBJ154, 23LBK688, 24CBU510; \n989) REF DYNJ908295, UDI/DI 10195327029579 (each) 40195327029570 (case), Lot Numbers:  22FBP832, 22JBQ931, 23EBL980, 23ABH904, 23CBI554, 23BBO814, 23KBT462, 23KBC620, 23LBU359, 24BBC925, 24CBF715, 24DBN038; \n990) REF DYNJ908296, UDI/DI 10195327033392 (each) 40195327033393 (case), Lot Numbers:  22FBS143, 22FBS144; \n991) REF DYNJ908299, UDI/DI 10195327030285 (each) 40195327030286 (case), Lot Numbers:  23ABQ876, 23DBE301, 23DBD644, 23ABM209, 23ABF344, 23BBR968, 23EBO972; \n992) REF DYNJ908331B, UDI/DI 10195327449636 (each) 40195327449637 (case), Lot Numbers:  23HBG005; \n993) REF DYNJ908332B, UDI/DI 10195327474416 (each) 40195327474417 (case), Lot Numbers:  23JBM281; \n994) REF DYNJ908345, UDI/DI 10195327048365 (each) 40195327048366 (case), Lot Numbers:  22LME372; \n995) REF DYNJ908346, UDI/DI 10195327048372 (each) 40195327048373 (case), Lot Numbers:  22GMB091, 22FMF893, 22HMF816; \n996) REF DYNJ908348, UDI/DI 10195327048501 (each) 40195327048502 (case), Lot Numbers:  22GMD710, 22HMF303, 22FMA020, 22LMH351, 22LMF421, 22JMF319, 22KMB101; \n997) REF DYNJ908363, UDI/DI 10195327058319 (each) 40195327058310 (case), Lot Numbers:  22DDB949; \n998) REF DYNJ908363A, UDI/DI 10195327163082 (each) 40195327163083 (case), Lot Numbers:  22FDA839, 22GDB474; \n999) REF DYNJ908363D, UDI/DI 10195327369453 (each) 40195327369454 (case), Lot Numbers:  23HDA539, 23GDA374, 23KDA772; \n1000) REF DYNJ908363F, UDI/DI 10195327576332 (each) 40195327576333 (case), Lot Numbers:  24ADB827, 24BDA921, 24EDA369, 24CDB360, 24FDA234; \n1001) REF DYNJ908374A, UDI/DI 10195327175696 (each) 40195327175697 (case), Lot Numbers:  22FDB256, 22FDB975, 22HDB892, 22HDB095; \n1002) REF DYNJ908374D, UDI/DI 10195327369460 (each) 40195327369461 (case), Lot Numbers: 23JDA117, 23FDA796, 23IDA784;  \n1003) REF DYNJ908519A, UDI/DI 10195327470401 (each) 40195327470402 (case), Lot Numbers:  23KBS441; \n1004) REF DYNJ908520, UDI/DI 10195327082178 (each) 40195327082179 (case), Lot Numbers:  22IBV995, 22EBG946, 22LBF448, 23BBF957; \n1005) REF DYNJ908520A, UDI/DI 10195327342302 (each) 40195327342303 (case), Lot Numbers:  23CBO976, 23BBP231, 23GBG150; \n1006) REF DYNJ908520B, UDI/DI 10195327470470 (each) 40195327470471 (case), Lot Numbers:  23JBY177, 23LBB383, 23HBT110; \n1007) REF DYNJ908521, UDI/DI 10195327082161 (each) 40195327082162 (case), Lot Numbers:  22FBU636, 22FBK410, 22IBH122, 23BBK665; \n1008) REF DYNJ908521A, UDI/DI 10195327342296 (each) 40195327342297 (case), Lot Numbers:  23DBB093, 23FBB333; \n1009) REF DYNJ908521B, UDI/DI 10195327470371 (each) 40195327470372 (case), Lot Numbers:  23IBC633; \n1010) REF DYNJ908587, UDI/DI 10195327108304 (each) 40195327108305 (case), Lot Numbers:  22GDB889; \n1011) REF DYNJ908587A, UDI/DI 10195327298975 (each) 40195327298976 (case), Lot Numbers:  23ADB763, 23ADA639; \n1012) REF DYNJ908587B, UDI/DI 10195327369477 (each) 40195327369478 (case), Lot Numbers:  23GDA286, 23HDB750; \n1013) REF DYNJ908587C, UDI/DI 10195327576219 (each) 40195327576210 (case), Lot Numbers:  24BDA532, 24DDB345; \n1014) REF DYNJ908638, UDI/DI 10195327122287 (each) 40195327122288 (case), Lot Numbers:  22EBR654, 22EBL550, 22KBW220, 22KBP678, 23BBN934, 23ABQ841, 23ABO551, 23DBE144, 23DBB106, 23CBP300; \n1015) REF DYNJ908642A, UDI/DI 10195327168407 (each) 40195327168408 (case), Lot Numbers:  22FBR725, 22JBL299; \n1016) REF DYNJ908642B, UDI/DI 10195327243432 (each) 40195327243433 (case), Lot Numbers:  22KBS080, 22KBA822, 22JBE869; \n1017) REF DYNJ908642C, UDI/DI 10195327315009 (each) 40195327315000 (case), Lot Numbers:  23ABP725, 23CBQ351, 23DBH519; \n1018) REF DYNJ908644A, UDI/DI 10195327168650 (each) 40195327168651 (case), Lot Numbers:  22FBP241; \n1019) REF DYNJ908644B, UDI/DI 10195327243449 (each) 40195327243440 (case), Lot Numbers:  22IBL137, 22KBS065; \n1020) REF DYNJ908644C, UDI/DI 10195327315016 (each) 40195327315017 (case), Lot Numbers:  23BBL706, 23CBT963; \n1021) REF DYNJ908647A, UDI/DI 10195327168438 (each) 40195327168439 (case), Lot Numbers:  22IBE709, 22JBH372, 23ABM544, 22JBR824, 23EBG073; \n1022) REF DYNJ908674, UDI/DI 10195327136581 (each) 40195327136582 (case), Lot Numbers:  22GDA786; \n1023) REF DYNJ908674A, UDI/DI 10195327251741 (each) 40195327251742 (case), Lot Numbers:  22LDA348; \n1024) REF DYNJ908674B, UDI/DI 10195327334000 (each) 40195327334001 (case), Lot Numbers:  23DDC352; \n1025) REF DYNJ908674C, UDI/DI 10195327428952 (each) 40195327428953 (case), Lot Numbers:  23JDA143, 23IDA761, 23HDC130, 23GDB834, 23GDA042; \n1026) REF DYNJ908676B, UDI/DI 10195327434861 (each) 40195327434862 (case), Lot Numbers:  23HDB965, 23JDC407; \n1027) REF DYNJ908681, UDI/DI 10195327139681 (each) 40195327139682 (case), Lot Numbers:  22GDB816, 22LDA354, 23DDA403, 23JDC215, 23KDB420, 23IDA575, 24ADC245; \n1028) REF DYNJ908682B, UDI/DI 10195327334031 (each) 40195327334032 (case), Lot Numbers:  23JDC268; \n1029) REF DYNJ908683, UDI/DI 10195327139698 (each) 40195327139699 (case), Lot Numbers:  22HDC171; \n1030) REF DYNJ908683A, UDI/DI 10195327251772 (each) 40195327251773 (case), Lot Numbers:  22KDC147; \n1031) REF DYNJ908683B, UDI/DI 10195327396435 (each) 40195327396436 (case), Lot Numbers:  23IDB009; \n1032) REF DYNJ908780, UDI/DI 10195327183356 (each) 40195327183357 (case), Lot Numbers:  23CDA809, 23FDA816, 23HDB677, 23DDA731; \n1033) REF DYNJ908782, UDI/DI 10195327183325 (each) 40195327183326 (case), Lot Numbers:  23FDA586, 23IDB890; \n1034) REF DYNJ908786, UDI/DI 10195327183332 (each) 40195327183333 (case), Lot Numbers:  23CDB081, 23JME836, 23IDA376; \n1035) REF DYNJ908798, UDI/DI 10195327183349 (each) 40195327183340 (case), Lot Numbers:  23CDB381, 23FDA714, 23HDC011; \n1036) REF DYNJ908799, UDI/DI 10195327183370 (each) 40195327183371 (case), Lot Numbers:  23DDB167, 23FDA791, 23IDC106, 23HDA802; \n1037) REF DYNJ908800, UDI/DI 10195327183387 (each) 40195327183388 (case), Lot Numbers:  23FDB403, 23HDC467; \n1038) REF DYNJ908800A, UDI/DI 10195327520816 (each) 40195327520817 (case), Lot Numbers:  23KBO196, 23KBD414, 23KBW418; \n1039) REF DYNJ908801, UDI/DI 10195327183394 (each) 40195327183395 (case), Lot Numbers:  22IDA928, 23IDA359, 23GDB160; \n1040) REF DYNJ908801A, UDI/DI 10195327520809 (each) 40195327520800 (case), Lot Numbers:  23LBN623, 23KBE475, 23JBY024; \n1041) REF DYNJ908802, UDI/DI 10195327183363 (each) 40195327183364 (case), Lot Numbers:  23DDB298, 23FDA792, 23HDB652; \n1042) REF DYNJ908802A, UDI/DI 10195327520793 (each) 40195327520794 (case), Lot Numbers:  23LBT966, 23KBT242; \n1043) REF DYNJ908811A, UDI/DI 10195327525422 (each) 40195327525423 (case), Lot Numbers:  23KMI866; \n1044) REF DYNJ908816, UDI/DI 10195327186531 (each) 40195327186532 (case), Lot Numbers:  22KDB737; \n1045) REF DYNJ908817, UDI/DI 10195327186548 (each) 40195327186549 (case), Lot Numbers:  23JMB238, 23DDC016, 23JMI125, 23IDA754; \n1046) REF DYNJ908819, UDI/DI 10195327185626 (each) 40195327185627 (case), Lot Numbers:  23EDA729, 23DDB170, 23IDA168, 23LDA179, 23JDA966; \n1047) REF DYNJ908828, UDI/DI 10195327185725 (each) 40195327185726 (case), Lot Numbers:  23FDA575, 23KMD570, 23IDB082; \n1048) REF DYNJ908831, UDI/DI 10195327185640 (each) 40195327185641 (case), Lot Numbers:  23CDB151, 23IDA920; \n1049) REF DYNJ908832, UDI/DI 10195327185657 (each) 40195327185658 (case), Lot Numbers:  23EDA173, 23FDA820, 23IMH088, 23IDB313, 23JMI448; \n1050) REF DYNJ908865C, UDI/DI 10195327481988 (each) 40195327481989 (case), Lot Numbers:  23ILA619, 23JLA397; \n1051) REF DYNJ908979, UDI/DI 10195327238995 (each) 40195327238996 (case), Lot Numbers:  22KBB139; \n1052) REF DYNJ908998, UDI/DI 10195327237783 (each) 40195327237784 (case), Lot Numbers:  22KDB214; \n1053) REF DYNJ909000A, UDI/DI 10195327428327 (each) 40195327428328 (case), Lot Numbers:  23HBQ467, 23GBK254, 23JBH758; \n1054) REF DYNJ909002, UDI/DI 10195327238148 (each) 40195327238149 (case), Lot Numbers:  23EBI750, 23EBT346, 23FBG449; \n1055) REF DYNJ909002A, UDI/DI 10195327428334 (each) 40195327428335 (case), Lot Numbers:  23HBQ478, 23KBK334; \n1056) REF DYNJ909004, UDI/DI 10195327238971 (each) 40195327238972 (case), Lot Numbers:  22KDB988; \n1057) REF DYNJ909004A, UDI/DI 10195327444129 (each) 40195327444120 (case), Lot Numbers:  23GDB166; \n1058) REF DYNJ909005, UDI/DI 10195327238988 (each) 40195327238989 (case), Lot Numbers:  22KDB989, 23CDC590, 23GDB399, 23FDB461; \n1059) REF DYNJ909019, UDI/DI 10195327247799 (each) 40195327247790 (case), Lot Numbers:  22LBP962; \n1060) REF DYNJ909019A, UDI/DI 10195327596842 (each) 40195327596843 (case), Lot Numbers:  24CBK591; \n1061) REF DYNJ909143B, UDI/DI 10195327498436 (each) 40195327498437 (case), Lot Numbers:  23JMD034; \n1062) REF DYNJ909161, UDI/DI 10195327281854 (each) 40195327281855 (case), Lot Numbers:  23ABL819, 23EBD646; \n1063) REF DYNJ909162, UDI/DI 10195327281861 (each) 40195327281862 (case), Lot Numbers:  23ABL818; \n1064) REF DYNJ909380, UDI/DI 10195327323134 (each) 40195327323135 (case), Lot Numbers:  23FDB613; \n1065) REF DYNJ909400, UDI/DI 10195327336103 (each) 40195327336104 (case), Lot Numbers:  23BBR478, 23FBS605, 23FBK753; \n1066) REF DYNJ909405, UDI/DI 10195327338435 (each) 40195327338436 (case), Lot Numbers:  23CMB928; \n1067) REF DYNJ909412, UDI/DI 10195327339814 (each) 40195327339815 (case), Lot Numbers:  23EMA940, 23KMI102, 23KMB366, 23HMC618, 23JMC097, 23JMB263, 23IMB861, 23HMI892, 23JMI583, 23JMI722, 23JMC920, 24BMF068, 24AMA070; \n1068) REF DYNJ909413, UDI/DI 10195327339845 (each) 40195327339846 (case), Lot Numbers:  23EME763; \n1069) REF DYNJ909414, UDI/DI 10195327339944 (each) 40195327339945 (case), Lot Numbers:  23DMB749. 23HMB359; \n1070) REF DYNJ909415, UDI/DI 10195327340063 (each) 40195327340064 (case), Lot Numbers:  23FMG135, 23EMG876, 23IMG783, 23IMA715, 23JMH846, 23JMB339, 24AME242; \n1071) REF DYNJ909422, UDI/DI 10195327347741 (each) 40195327347742 (case), Lot Numbers:  23DBD577; \n1072) REF DYNJ909423, UDI/DI 10195327347727 (each) 40195327347728 (case), Lot Numbers:  23DBF657; \n1073) REF DYNJ909424, UDI/DI 10195327347765 (each) 40195327347766 (case), Lot Numbers:  23DBF804; \n1074) REF DYNJ909462, UDI/DI 10195327361563 (each) 40195327361564 (case), Lot Numbers:  23DBM644, 23GBH101, 23FBB537, 23JBD213; \n1075) REF DYNJ909463, UDI/DI 10195327361570 (each) 40195327361571 (case), Lot Numbers:  23GBT173, 23EBN778, 23EBM890, 23DBR999; \n1076) REF DYNJ909464, UDI/DI 10195327361594 (each) 40195327361595 (case), Lot Numbers:  23DBN690; \n1077) REF DYNJ909494, UDI/DI 10195327374907 (each) 40195327374908 (case), Lot Numbers:  23HBE924; \n1078) REF DYNJ909494A, UDI/DI 10195327528836 (each) 40195327528837 (case), Lot Numbers:  23KBN713, 23KBN908; \n1079) REF DYNJ909500, UDI/DI 10195327384951 (each) 40195327384952 (case), Lot Numbers:  23KBN713; \n1080) REF DYNJ909502, UDI/DI 10195327384968 (each) 40195327384969 (case), Lot Numbers:  23KBN908; \n1081) REF DYNJ909505, UDI/DI 10195327386030 (each) 40195327386031 (case), Lot Numbers:  23FMD557; \n1082) REF DYNJ909510, UDI/DI 10195327387235 (each) 40195327387236 (case), Lot Numbers:  23GLA881, 23GLA007; \n1083) REF DYNJ909579, UDI/DI 10195327409715 (each) 40195327409716 (case), Lot Numbers:  23HBM873, 23JBR241; \n1084) REF DYNJ909582, UDI/DI 10195327413590 (each) 40195327413591 (case), Lot Numbers:  23JBT141, 23FBP595; \n1085) REF DYNJ909590A, UDI/DI 10195327501105 (each) 40195327501106 (case), Lot Numbers:  23JBQ279, 23IBS510; \n1086) REF DYNJ909593, UDI/DI 10195327413644 (each) 40195327413645 (case), Lot Numbers:  23FBT925; \n1087) REF DYNJ909595, UDI/DI 10195327413583 (each) 40195327413584 (case), Lot Numbers:  23JBT858, 23JBL547; \n1088) REF DYNJ909625, UDI/DI 10195327415143 (each) 40195327415144 (case), Lot Numbers:  23IBL744; \n1089) REF DYNJ909657, UDI/DI 10195327431532 (each) 40195327431533 (case), Lot Numbers:  23KBF228, 23FBQ974, 24BBJ335, 24ABR146, 24CBA927, 24EBD307; \n1090) REF DYNJ909658, UDI/DI 10195327431549 (each) 40195327431540 (case), Lot Numbers:  24ABO335; \n1091) REF DYNJ909687, UDI/DI 10195327442293 (each) 40195327442294 (case), Lot Numbers:  23IBE284; \n1092) REF DYNJ909795, UDI/DI 10195327480363 (each) 40195327480364 (case), Lot Numbers:  24ABK091; \n1093) REF DYNJ909796A, UDI/DI 10195327639266 (each) 40195327639267 (case), Lot Numbers:  24JBB147, 24HBD833, 24GBL455, 24DBK383; \n1094) REF DYNJ909797, UDI/DI 10195327481513 (each) 40195327481514 (case), Lot Numbers:  23LLA172; \n1095) REF DYNJ909930B, UDI/DI 10195327638023 (each) 40195327638024 (case), Lot Numbers:  24DBJ984, 24GBD808; \n1096) REF DYNJ909970, UDI/DI 10195327522179 (each) 40195327522170 (case), Lot Numbers:  24ALA572, 24ALA502; \n1097) REF DYNJ910010, UDI/DI 10195327542443 (each) 40195327542444 (case), Lot Numbers:  24ADC084; \n1098) REF DYNJ910068, UDI/DI 10195327567163 (each) 40195327567164 (case), Lot Numbers:  24BBN397; \n1099) REF DYNJ910071, UDI/DI 10195327567194 (each) 40195327567195 (case), Lot Numbers:  24BBN398; \n1100) REF DYNJ910085, UDI/DI 10195327569105 (each) 40195327569106 (case), Lot Numbers:  23LBO017, 24CBM480, 24BBR419, 24BBD693; \n1101) REF DYNJ910132, UDI/DI 10195327597214 (each) 40195327597215 (case), Lot Numbers:  24DBI622; \n1102) REF DYNJ910418, UDI/DI 10195327681531 (each) 40195327681532 (case), Lot Numbers:  24EBH428; \n1103) REF DYNJ910419, UDI/DI 10195327681555 (each) 40195327681556 (case), Lot Numbers:  24EBM474; \n1104) REF DYNJC9270K, UDI/DI 10195327109172 (each) 40195327109173 (case), Lot Numbers:  22HDA634, 23CDA802, 23GDC010; \n1105) REF DYNJC9270L, UDI/DI 10195327476236 (each) 40195327476237 (case), Lot Numbers:  23HDB757; \n1106) REF DYNJC9271N, UDI/DI 10195327109189 (each) 40195327109180 (case), Lot Numbers:  23ABD952, 22LBF230, 23BBU052, 23ABR018, 23HBP102; \n1107) REF DYNJC9271O, UDI/DI 10195327476007 (each) 40195327476008 (case), Lot Numbers:  23KBG170; \n1108) REF DYNJC9272O, UDI/DI 10195327109097 (each) 40195327109098 (case), Lot Numbers:  22FDB204, 23FDA025, 23HDB305, 23HDA828; \n1109) REF DYNJC9272P, UDI/DI 10195327476229 (each) 40195327476220 (case), Lot Numbers:  23IDC228; \n1110) REF DYNJC9309K, UDI/DI 10195327109103 (each) 40195327109104 (case), Lot Numbers:  22GBN804, 23ABC988, 23DBM210; \n1111) REF DYNJC9320I, UDI/DI 10193489989939 (each) 40193489989930 (case), Lot Numbers:  23HDA687; \n1112) REF DYNJC9323K, UDI/DI 10195327027490 (each) 40195327027491 (case), Lot Numbers:  23FBG142, 23LBT898, 24DBC169, 24EBC455, 24DBR463; \n1113) REF DYNJC9324G, UDI/DI 10193489989946 (each) 40193489989947 (case), Lot Numbers:  22IBE723, 22FBP755, 22JBQ158, 22HBD446, 22GBX113, 22HBH857; \n1114) REF DYNJV0277J, UDI/DI 10193489972146 (each) 40193489972147 (case), Lot Numbers:  22GMA861, 22JMG291, 23AMF743, 22KMD625, 23EMH516; \n1115) REF DYNJV0277K, UDI/DI 10195327441074 (each) 40195327441075 (case), Lot Numbers:  23FMI731; \n1116) REF DYNJV0279I, UDI/DI 10193489971989 (each) 40193489971980 (case), Lot Numbers:  22CMB321; \n1117) REF DYNJV0279J, UDI/DI 10195327161798 (each) 40195327161799 (case), Lot Numbers:  22EME514, 23BMA404, 22JME083, 22JME084, 23DMC089, 23EMJ121; \n1118) REF DYNJVA900018B, UDI/DI 10193489379518 (each) 40193489379519 (case), Lot Numbers:  23BLB028; \n1119) REF DYNJVB1276C, UDI/DI 10195327295158 (each) 40195327295159 (case), Lot Numbers:  23ALA976, 23GLA480, 23GLA099, 23DLB133, 23CLA044, 22LLA333, 24CLA095, 24ALA960, 23KLB076, 23HLB285, 24DLA755, 24DLA325; \n1120) REF DYNJVB1277D, UDI/DI 10195327482107 (each) 40195327482108 (case), Lot Numbers:  23HLA529; \n1121) REF DYNJVB91006, UDI/DI 10193489801552 (each) 40193489801553 (case), Lot Numbers:  22ILB119, 22KLA992, 23DLA379, 23ALB020, 23ALA419, 23GLA235, 23ELB243; \n1122) REF DYNJVB91038B, UDI/DI 10195327171759 (each) 40195327171750 (case), Lot Numbers:  23KDA679, 23HDB476, 23HDA877, 24BDB627, 23BDB258, 23BDA097, 24BDB591, 23CDC501, 23EDB105; \n1123) REF DYNJVB91047C, UDI/DI 10195327171797 (each) 40195327171798 (case), Lot Numbers:  22JDA678; \n1124) REF DYNJVB91047D, UDI/DI 10195327312299 (each) 40195327312290 (case), Lot Numbers:  23GDB483, 23EDB468, 23IDA442; \n1125) REF DYNJVB91048D, UDI/DI 10195327171704 (each) 40195327171705 (case), Lot Numbers:  22JDA345, 22GDA981, 22KDB953; \n1126) REF DYNJVB91048F, UDI/DI 10195327312305 (each) 40195327312306 (case), Lot Numbers:  23BDA154, 23EDB601, 23KDA631, 23JDB496, 23IDC212, 24ADA251, 24ADC222, 24DDA319, 24CDC026, 24DDC083; \n1127) REF DYNJVB91059, UDI/DI 10193489801187 (each) 40193489801188 (case), Lot Numbers:  22HLA783, 22ELB050, 22FLA771; \n1128) REF DYNJVB91059B, UDI/DI 10195327550158 (each) 40195327550159 (case), Lot Numbers:  24ALA957, 24CLA534, 24DLA595; \n1129) REF DYNJVB91128B, UDI/DI 10193489986600 (each) 40193489986601 (case), Lot Numbers:  22ILA838; \n1130) REF DYNJVB91128C, UDI/DI 10195327295165 (each) 40195327295166 (case), Lot Numbers:  23CLA774, 22LLA614; \n1131) REF DYNJVB91128D, UDI/DI 10195327367916 (each) 40195327367917 (case), Lot Numbers:  23ILA486, 23HLA869, 23HLA337, 23GLA693, 23FLA076, 23DLA842, 24CLA151, 23JLA241, 24DLB010, 24CLA896; \n1132) REF PHS102077N, UDI/DI 10195327353759 (each) 40195327353750 (case), Lot Numbers:  23DDA851, 23FDA520, 23EDB088, 23IDC335, 23HDA278",
      "firm_fei_number": "1417592",
      "recalling_firm": "MEDLINE INDUSTRIES, LP - Northfield",
      "address_1": "3 Lakes Dr",
      "city": "Northfield",
      "state": "IL",
      "postal_code": "60093-2753",
      "additional_info_contact": "Haley Barclay\n866-359-1704",
      "reason_for_recall": "Medline is recalling medical convenience manufactured using certain lots of SafeAir Smoke Evacuation Pencil which were subsequently recalled by Stryker.",
      "root_cause_description": "Under Investigation by firm",
      "action": "Medline issued an URGENT MEDICAL DEVICE RECALL notice to its consignees on 10/11/2024 via mail and email.  The notice explained the problem with the Stryker component, potential risk, requested the kits be quarantined pending additional labeling and o respond through their portal:  https://recalls.medline.com.  Once the response is received, Medline will provide additional labeling to be applied to the affected kits which states the following:  \"Attention: R-24-219, This kit contains a recalled item, STRYKER SafeAir Smoke Evacuation Pencil, VPN: 0703-047-000, 0703-047-001, 0703-047-002, 0703-046-000, 0703-046-001, 0703-046-002.  DO NOT USE THIS COMPONENT.  The rest of the kit may be used.  Please remove this item only and replace with product from supply.  *Do not cover or remove this sticker*\".\nThose further distributing the product were directed to notify their customers.",
      "product_quantity": "252162 units",
      "distribution_pattern": "US Nationwide distribution.",
      "openfda": {
        "registration_number": [
          "2029015",
          "1043534",
          "3008544874",
          "3012120772",
          "3015543569",
          "3005595283",
          "3018077422",
          "9616088",
          "1221053",
          "3045816907",
          "3008395366",
          "3002807315",
          "3006981798",
          "1450662",
          "1424263",
          "3005012805",
          "1928237",
          "3035642068",
          "1423537",
          "3045816902",
          "1123137",
          "3007289408",
          "3017307654",
          "1061124",
          "3008864603",
          "3014170143",
          "3015131017",
          "8044102",
          "1825034",
          "9710629",
          "8031020",
          "3004911384",
          "1020279",
          "3004122598",
          "3007663067",
          "3003477135",
          "3010667733",
          "3013552417",
          "1030451",
          "3014554088",
          "3007329020",
          "3004111573",
          "3011187779",
          "1923569",
          "3009203853",
          "3017823294",
          "3011430871",
          "1835444",
          "1417592",
          "1423395",
          "3014680744",
          "3010052483",
          "3009888344",
          "1047429",
          "3008111820",
          "1643817",
          "3015173212",
          "3008534770"
        ],
        "fei_number": [
          "1043534",
          "3008544874",
          "3012120772",
          "3015543569",
          "1000512168",
          "3005595283",
          "3018077422",
          "3013299874",
          "1221053",
          "3045816907",
          "3008395366",
          "3002807315",
          "3006981798",
          "1450662",
          "3005012805",
          "3003015552",
          "1643817",
          "1928237",
          "3035642068",
          "3045816902",
          "3007289408",
          "3017307654",
          "3000717687",
          "3014170143",
          "3008864603",
          "3015131017",
          "1825034",
          "3002807830",
          "3004911384",
          "1020279",
          "3004122598",
          "3007663067",
          "3003477135",
          "3001236905",
          "3010667733",
          "3002902735",
          "3013552417",
          "1030451",
          "3014554088",
          "3007329020",
          "1000160458",
          "3004111573",
          "3011187779",
          "1923569",
          "3009203853",
          "3017823294",
          "3011430871",
          "1417592",
          "3000717925",
          "3014680744",
          "3010052483",
          "1000123180",
          "3009888344",
          "1047429",
          "3003407235",
          "3008111820",
          "3015173212",
          "3008534770"
        ],
        "device_name": "Orthopedic Tray",
        "medical_specialty_description": "General, Plastic Surgery",
        "regulation_number": "878.4800",
        "device_class": "1"
      }
    }
  ]
}