{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-08-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "209779",
      "product_res_number": "Z-0040-2025",
      "event_date_initiated": "2024-07-31",
      "event_date_posted": "2024-10-09",
      "recall_status": "Open, Classified",
      "res_event_number": "95230",
      "product_code": "MNQ",
      "product_description": "Inspire Model 4340 Respiratory Sensing Lead, component of the Inspire Upper Airway Stimulation (UAS) system.",
      "code_info": "Model 4340. UDI-DI: 10855728005946. Part Number: 900-014-00. Serial Numbers: D59544, D59550, D59543, D59541, D59542, D59528, D59526, D59534, D59521, D59477, D59519, D59515, D59516, D59489, D59514, D59512, D59510, D59509, D59547, D59549, D59505, D59506, D59504, D59503, D59502, D59529, D59497, D59494, D59490, D59485, D59486, D56077, D56072, D56064, D56060, D56059, D56055, D56038, D56082, D56035, D56044, D56031, D56032, D56033, D56034, D56040, D56046, D56030, D56036, D56053, D56054, D56058, D55284, D48681, D48668, D48661, D48655, D48650, D48410, D46477, D46478, D43687, D43688, D43700, D48014, D48007, D47994, D47991, D47983, D47964, D47963, D47958, D47940, D45763, D47939, D45768, D47913, D47228, D47903, D47896, D47897, D47909, D47910, D47951, D45706, D45721, D45724, D47877, D47872, D47879, D47880, D47691, D47670, D47654, D47243, D47240, D47220, D47187, D47181, D47176, D47177, D47191, D47192, D47175, D47169, D47163, D47164, D47142, D47131, D47926, D47102, D47013, D47006, D46990, D46966, D46955, D46854, D46855, D46847, D46846, D46835, D46804, D46797, D46784, D47202, D46699, D46700, D46664, D46566, D46567, D46561, D46535, D46532, D46527, D46528, D46529, D46525, D46526, D46463, D46451, D45962, D45949, D45947, D45777, D45773, D45745, D45708, D44687, D44685, D44677, D44675, D26997.",
      "firm_fei_number": "3007666314",
      "recalling_firm": "Inspire Medical Systems Inc.",
      "address_1": "5500 Wayzata Blvd Ste 1600 5500 Wayzata Blvd",
      "address_2": "Ste 1600",
      "city": "Golden Valley",
      "state": "MN",
      "postal_code": "55416-1237",
      "additional_info_contact": "Andrea Rasmussen\n763-205-7974",
      "reason_for_recall": "Labeling error with a mismatch between the shelf box serial number label and the serial number of the lead provided in the box.",
      "root_cause_description": "Process control",
      "action": "Inspire notified consignees on 07/31/2024, via emailed letter.  Consignees were instructed to confirm the serial number on the lead is the serial number used to register the patient and to contact your Inspire representative if any labeling inconsistency is identified for complaint reporting.",
      "product_quantity": "152 units",
      "distribution_pattern": "Worldwide distribution - US Nationwide and the countries of Belgium, Switzerland, Germany, United Kingdom, Netherlands, Singapore.",
      "openfda": {
        "pma_number": [
          "P240024",
          "P130008",
          "P250013"
        ],
        "registration_number": [
          "3017191859",
          "3036471728",
          "3030499299",
          "2183744",
          "3030384719",
          "2183613",
          "3003857828",
          "3009183442",
          "3008851924",
          "1047843",
          "3024996452",
          "3007666314",
          "3005672861",
          "3030384402",
          "2183787",
          "3010376846"
        ],
        "fei_number": [
          "3017191859",
          "3036471728",
          "3030499299",
          "2183744",
          "3030384719",
          "2183613",
          "3003857828",
          "3009183442",
          "3008851924",
          "1047843",
          "3024996452",
          "3007666314",
          "3005672861",
          "3030384402",
          "2183787",
          "3010376846"
        ],
        "device_name": "Stimulator, Hypoglossal Nerve, Implanted, Apnea",
        "medical_specialty_description": "Unknown",
        "regulation_number": "",
        "device_class": "3"
      }
    }
  ]
}