  "results": [
    {
      "cfres_id": "204604",
      "product_res_number": "Z-0575-2024",
      "event_date_initiated": "2023-10-24",
      "event_date_posted": "2023-12-18",
      "recall_status": "Open, Classified",
      "res_event_number": "93328",
      "product_code": "CCW",
      "product_description": "Injection Needle, curved 23 cm, REF: 8598A; Injection Needle, straight, 23 cm, REF: 8598B; Injection Needle, distal 45-degrees, 23 cm, REF: 8598D",
      "code_info": "REF/UDI-DI: 8598A/04048551213278, 8598B/04048551001462, 8598D/04048551213285\n\nAll lot numbers.",
      "firm_fei_number": "3007475226",
      "recalling_firm": "Karl Storz Endoscopy",
      "address_1": "2151 E Grand Ave",
      "city": "El Segundo",
      "state": "CA",
      "postal_code": "90245-5017",
      "additional_info_contact": "Kathy Strange\n774-284-1736",
      "reason_for_recall": "The efficacy of the manual cleaning process cannot be assured for the affected injection needles intended for general surgery and visceral surgery, which if used, may expose the patient to a higher risk of infection.",
      "root_cause_description": "Device Design",
      "action": "On 10/24/23, recall notices were mailed to representatives for medical product safety, users, operators, distributors who were asked to do the following:\n\n1) Discontinue use of all affected Injection Needles.\n2) Pass on the Recall Notice to all users of the affected products and all other persons who need to be aware within your organization.\n3) If you have or may have distributed the products listed, please identify and\npromptly notify those recipients, or provide the recalling firm with a list of customers\nwho received/may have received the products listed.\n4) Contact Customer Support to arrange for the return of affected devices by calling 800-421-0837.\n5) Return the completed Customer Reply Form via email to karlstorz5190@sedgwick.com\n\nFor all other questions regarding this communication, please send an email to KSEAProductLabeling@karlstorz.com",
      "product_quantity": "4,878",
      "distribution_pattern": "US: MI, NH, SC, PA, FL, NY, NC, MS, MO, CA, NJ, IL, NV, OK, CO, OH, WA, WI, VA, IN, IA, AR, TX, WV, MD, MN, GA, MA, NM, VT, AK, OR, ID, LA, KS, DC, KY, HI\n\nOUS: Mexica, Canada, Germany",
      "openfda": {
        "k_number": [
          "K163412",
          "K883296",
          "K831981",
          "K833055",
          "K922019",
          "K861026",
          "K954446",
          "K945901",
          "K971377",
          "K062230",
          "K944686",
          "K851705",
          "K830047",
          "K884073",
          "K832650",
          "K062231",
          "K945425",
          "K801452",
          "K883830",
          "K041852",
          "K961832",
          "K915161",
          "K063477",
          "K921263",
          "K872480",
          "K984101",
          "K882930",
          "K951804",
          "K945269",
          "K954438",
          "K926158",
          "K905273",
          "K922275",
          "K833133",
          "K953179",
          "K882433",
          "K191602",
          "K930804",
          "K944428",
          "K901680",
          "K844776",
          "K912905",
          "K945301",
          "K832984",
          "K915804",
          "K970619",
          "K954787",
          "K883819",
          "K951147",
          "K062523",
          "K883414",
          "K841778",
          "K895954",
          "K882659",
          "K914168",
          "K921166",
          "K140951",
          "K941319",
          "K901164",
          "K904918",
          "K901489",
          "K914741",
          "K121378",
          "K945590"
        ],
        "registration_number": [
          "",
          "8010374",
          "3021031769",
          "8021788",
          "3010272829",
          "3007481893",
          "3013557562",
          "3005273623",
          "3010676356",
          "3013358456",
          "3009512503",
          "3011499367",
          "8010099",
          "3004168759",
          "3012843412",
          "3014680906",
          "3013528571",
          "3009340886",
          "3003431869",
          "9614093",
          "3012507533",
          "1928237",
          "3012002722",
          "3010317211",
          "3008770252",
          "2183650",
          "3010149039",
          "3000279801",
          "3004748541",
          "3002858762",
          "3007789561",
          "3015139770",
          "2030745",
          "3010611926",
          "3007123908",
          "3009255580",
          "9615393",
          "3016195092",
          "3010402691",
          "3018094310",
          "3006154292",
          "3012322232",
          "3010396739",
          "3009669249",
          "3015398570",
          "3008388245",
          "3014656749",
          "3022320321",
          "3006455037",
          "3014129616",
          "3021477808",
          "1225968",
          "9611102",
          "9611283",
          "9680837",
          "8040884",
          "3002807115",
          "3003898360",
          "3003924186",
          "3006389770",
          "3010707607",
          "2936999",
          "1836161",
          "3005579585",
          "3043585862",
          "3007817051",
          "1720747",
          "3030516433",
          "3004365956",
          "3033010858",
          "3012728206",
          "3026201231",
          "9613926",
          "8010482",
          "9617248",
          "3007421149",
          "3042158672",
          "1926681",
          "2410251",
          "1314417",
          "3006891796",
          "9610691",
          "3003790304",
          "1036781",
          "9610531",
          "2025816",
          "3013901429",
          "9680411",
          "3010202439",
          "3013784566",
          "3003674698",
          "9681622",
          "3030733800",
          "9614084",
          "3014023545",
          "3006546082",
          "8040263",
          "3005418651",
          "3011120183",
          "8010704",
          "3003807268",
          "3009560891",
          "3019530907",
          "3003902679",
          "3010564249",
          "9680143",
          "3005067367",
          "1519132",
          "2243757",
          "3006059835",
          "3010829427",
          "3005092287",
          "3007137643",
          "1000122786",
          "3003759558",
          "1643116",
          "3008936260",
          "3029754081",
          "3009420598",
          "2182318",
          "3015216264",
          "3008520150",
          "9613563",
          "3017166758",
          "9613910",
          "3007751820",
          "9611665",
          "8010418",
          "3003877407",
          "9611062",
          "3022862651",
          "3005813597",
          "9681246",
          "3003761012",
          "3032747418",
          "3008264111",
          "3032391",
          "3002808270",
          "3009513193",
          "9611112",
          "3002807267",
          "2432320",
          "3005216736",
          "2030624",
          "3017202422",
          "3011543740",
          "3014579161",
          "3004863697",
          "2518436",
          "3019784580",
          "2080783",
          "3012797688",
          "9610861",
          "3009703496",
          "3015173212",
          "3002769835",
          "3015895045",
          "3008693658",
          "1000379039",
          "3013503739",
          "3015351976",
          "3009756153",
          "9616250",
          "3003435550",
          "8020893",
          "3000126234",
          "3009217531",
          "9680244",
          "3003418325",
          "3016734145",
          "3010806653",
          "3002761766",
          "1316463",
          "1421101",
          "8040278",
          "3027556548",
          "1054713",
          "3011371465",
          "3038087139",
          "3012445994",
          "3011497859",
          "3008808049",
          "3010353847",
          "3031564283",
          "3015512299",
          "1057358",
          "3014317364",
          "9680168",
          "3002090132",
          "3010220595",
          "3029082594",
          "3004552850",
          "1066427",
          "3004111573",
          "3014908171",
          "3013247477",
          "3012494290",
          "3016634868",
          "3007647244",
          "9613079",
          "3006713849",
          "9610617",
          "3007597038",
          "9616354",
          "3024026674",
          "9680721",
          "1722070",
          "8043316",
          "3042888274",
          "3003322138",
          "3003334541",
          "3036978610",
          "9610773",
          "3010455030",
          "3012104670",
          "3004571672",
          "1417592",
          "3006513362",
          "3003644849",
          "3003032798",
          "3006787036",
          "1125612",
          "3009429841",
          "3001620590",
          "3016965929",
          "1123137",
          "3004001706",
          "3007583964",
          "3045058049",
          "3005698350",
          "3013831261",
          "3006087789",
          "3015997711",
          "3006061749",
          "3004765453",
          "3004530184",
          "8030673",
          "8010939",
          "3025349182",
          "3013501110",
          "9613073",
          "2434008",
          "3009538961",
          "3030446844",
          "3027655341",
          "8022032",
          "1221826",
          "3005536872",
          "3004215117",
          "3007532396",
          "3020984207",
          "3009747810",
          "3006786907",
          "3003882387",
          "3016678045",
          "3002834291",
          "3010042047",
          "3003800159",
          "3005528784",
          "3003213883",
          "3004598675",
          "2085947",
          "3005450700",
          "9610612",
          "3020460367",
          "3001614464",
          "9710567",
          "9614069",
          "2244478",
          "8040510",
          "3011600579",
          "3008338766",
          "9681384",
          "3003851906",
          "3027847726",
          "9681540",
          "1222773",
          "2022648",
          "3011050570",
          "3014233600",
          "3010683054",
          "3012542015",
          "3019404837",
          "2431166",
          "3011137372",
          "3045010519",
          "9612074",
          "3027187888",
          "3014498758",
          "1412854",
          "3030839824",
          "3008305513",
          "3011307234",
          "3022472",
          "2529846",
          "3004608878",
          "8010298",
          "3005034064",
          "3013735264",
          "8021995",
          "3013515223",
          "3002675176",
          "2020550",
          "3039414397",
          "9680002",
          "3014342096",
          "3023657851",
          "3001650535",
          "8043903",
          "9680518",
          "3015143057",
          "3010726901",
          "3021666757",
          "3014279668",
          "3013524663",
          "8010427",
          "3017178358"
        ],
        "fei_number": [
          "",
          "2523865",
          "3021031769",
          "3010272829",
          "3002807308",
          "3002808268",
          "3007481893",
          "3013557562",
          "3005273623",
          "3010676356",
          "3013358456",
          "3003772581",
          "3009512503",
          "3011499367",
          "3004168759",
          "3012843412",
          "3002750084",
          "1000117260",
          "3014680906",
          "5870",
          "3013528571",
          "3009340886",
          "3003431869",
          "3012507533",
          "1928237",
          "3012002722",
          "3010317211",
          "3008770252",
          "2183650",
          "3010149039",
          "3008268830",
          "1000136191",
          "3000279801",
          "3004748541",
          "3002858762",
          "3002807850",
          "3002807836",
          "3007789561",
          "3015139770",
          "3010611926",
          "3007123908",
          "3009255580",
          "3008782585",
          "3016195092",
          "3010402691",
          "3018094310",
          "1000208608",
          "3006154292",
          "3002807214",
          "3012322232",
          "3010396739",
          "3009669249",
          "3015398570",
          "3008388245",
          "3014656749",
          "3005947796",
          "3002807095",
          "3022320321",
          "3006455037",
          "3014129616",
          "3021477808",
          "3002807115",
          "3003898360",
          "3003924186",
          "3006389770",
          "1000521377",
          "3010707607",
          "3002807626",
          "2936999",
          "3005579585",
          "3043585862",
          "3000206435",
          "3007817051",
          "1720747",
          "3030516433",
          "3008205331",
          "3004216008",
          "3004365956",
          "3000204983",
          "3002807682",
          "3033010858",
          "3012728206",
          "3026201231",
          "3002808325",
          "3002808440",
          "3007421149",
          "3042158672",
          "1926681",
          "2410251",
          "1314417",
          "3006891796",
          "3003790304",
          "2025816",
          "3013901429",
          "3013784566",
          "3003674698",
          "3546",
          "3030733800",
          "3014023545",
          "3006546082",
          "3005418651",
          "3011120183",
          "3003807268",
          "3009560891",
          "3019530907",
          "3003902679",
          "3010564249",
          "1000123180",
          "3005693456",
          "1000295034",
          "1519132",
          "2243757",
          "3006059835",
          "3010829427",
          "3005092287",
          "3007137643",
          "1000122786",
          "3002807907",
          "3003759558",
          "1643116",
          "3008936260",
          "3029754081",
          "3009420598",
          "2182318",
          "3015216264",
          "3008520150",
          "3017166758",
          "1000402972",
          "3007751820",
          "3003877407",
          "3022862651",
          "3006363476",
          "3001717360",
          "3005813597",
          "3003761012",
          "3032747418",
          "3008264111",
          "3002931663",
          "3002808270",
          "3009513193",
          "3002807267",
          "3005216736",
          "3000321209",
          "3017202422",
          "3011543740",
          "3001319476",
          "3014579161",
          "3004863697",
          "2518436",
          "3019784580",
          "2080783",
          "3002807183",
          "3012797688",
          "3009703496",
          "3015173212",
          "3002769835",
          "3015895045",
          "3002807754",
          "3008693658",
          "1000379039",
          "3013503739",
          "3015351976",
          "1000244733",
          "3009756153",
          "3003435550",
          "3000126234",
          "3009217531",
          "3003448211",
          "3003418325",
          "3016734145",
          "3010806653",
          "3002761766",
          "1316463",
          "1421101",
          "3027556548",
          "3002802903",
          "3011371465",
          "3038087139",
          "3012445994",
          "3011497859",
          "3008808049",
          "3010353847",
          "3031564283",
          "3015512299",
          "3014317364",
          "3002090132",
          "5682",
          "3010220595",
          "3029082594",
          "3004552850",
          "3004111573",
          "3014908171",
          "6034",
          "3013247477",
          "3002806359",
          "1000111918",
          "3012494290",
          "3016634868",
          "3007647244",
          "3006713849",
          "1000555370",
          "1000114111",
          "3024026674",
          "3004213967",
          "3042888274",
          "3002807522",
          "3003322138",
          "3003334541",
          "3036978610",
          "1000120768",
          "3010455030",
          "3012104670",
          "3004571672",
          "3007130719",
          "1417592",
          "3002648735",
          "3006513362",
          "3003644849",
          "3003032798",
          "3006787036",
          "3009429841",
          "3001620590",
          "3016965929",
          "3004001706",
          "3002806315",
          "3007583964",
          "1000136269",
          "3045058049",
          "3005698350",
          "3013831261",
          "3006087789",
          "1000292048",
          "3015997711",
          "3003927064",
          "3006061749",
          "3004765453",
          "3004530184",
          "3025349182",
          "3013501110",
          "2434008",
          "3009538961",
          "3030446844",
          "3027655341",
          "3003256355",
          "1221826",
          "3005536872",
          "3004215117",
          "3007532396",
          "3020984207",
          "3009747810",
          "3005435924",
          "3006786907",
          "3003882387",
          "3016678045",
          "3002834291",
          "3010042047",
          "3003800159",
          "3005528784",
          "3003213883",
          "3007475226",
          "3004598675",
          "3003249645",
          "3005450700",
          "3020460367",
          "3001614464",
          "3002299608",
          "3003319165",
          "3011600579",
          "3007513035",
          "3002807798",
          "3008338766",
          "3374",
          "3003851906",
          "3027847726",
          "3004574120",
          "2022648",
          "1000347621",
          "3011050570",
          "3014233600",
          "3010683054",
          "3012542015",
          "3001620388",
          "3003418980",
          "3019404837",
          "3002807072",
          "3045010519",
          "3027187888",
          "3014498758",
          "1412854",
          "3030839824",
          "3008305513",
          "3011307234",
          "3022472",
          "3002806816",
          "3004608878",
          "3005034064",
          "3002808001",
          "3000204191",
          "3013735264",
          "3013515223",
          "3002675176",
          "2020550",
          "3039414397",
          "3014342096",
          "1000508235",
          "3023657851",
          "3001650535",
          "3015143057",
          "3010726901",
          "3021666757",
          "3005747797",
          "3005304066",
          "3014279668",
          "3013524663",
          "3002480801",
          "3002808088",
          "3017178358"
        ],
        "device_name": "Laryngoscope, Rigid",
        "medical_specialty_description": "Anesthesiology",
        "regulation_number": "868.5540",
        "device_class": "1"
      }
    }
  ]
}