{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-08-27",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "204269",
      "product_res_number": "Z-0478-2024",
      "event_date_initiated": "2023-10-31",
      "event_date_posted": "2023-12-02",
      "recall_status": "Open, Classified",
      "res_event_number": "93405",
      "product_code": "DSQ",
      "product_description": "PUMP 1103 HVAD (HeartWare Ventricular Assist Device)  IMPLANT KIT, Model Number 1103",
      "code_info": "GTIN none, Serial Numbers:  HW10991, HW11011, HW11012, HW11013, HW11017, HW11019, HW11020, HW11022, HW11023, HW11024, HW11026, HW11030, HW11032, HW11033, HW11034, HW11035, HW11036, HW11037, HW11038, HW11040, HW11042, HW11043, HW11044, HW11046, HW11047, HW11050, HW11051, HW11052, HW11053, HW11054, HW11057, HW11060, HW11061, HW11062, HW11063, HW11064, HW11065, HW11067, HW11070, HW11071, HW11072, HW11074, HW11075, HW11076, HW11077, HW11078, HW11082, HW11083, HW11085, HW11086, HW11087, HW11088, HW11089, HW11091, HW11092, HW11093, HW11094, HW11096, HW11100, HW11101, HW11103, HW11105, HW11106, HW11107, HW11109, HW11111, HW11112, HW11113, HW11114, HW11116, HW11117, HW11118, HW11119, HW11120, HW11121, HW11122, HW11123, HW11125, HW11126, HW11129, HW11130, HW11133, HW11134, HW11137, HW11138, HW11139, HW11140, HW11141, HW11142, HW11146, HW11147, HW11148, HW11149, HW11150, HW11151, HW11152, HW11153, HW11155, HW11157, HW11160, HW11161, HW11164, HW11165, HW11166, HW11168, HW11171, HW11172, HW11173, HW11174, HW11176, HW11177, HW11178, HW11180, HW11181, HW11183, HW11184, HW11185, HW11186, HW11187, HW11188, HW11190, HW11191, HW11192, HW11193, HW11194, HW11195, HW11196, HW11197, HW11198, HW11199, HW11201, HW11202, HW11204, HW11205, HW11209, HW11210, HW11216, HW11217, HW11218, HW11219, HW11221, HW11222, HW11223, HW11224, HW11225, HW11227, HW11228, HW11229, HW11230, HW11231, HW11232, HW11233, HW11234, HW11235, HW11236, HW11239, HW11240, HW11241, HW11242, HW11244, HW11246, HW11247, HW11248, HW11250, HW11251, HW11252, HW11255, HW11256, HW11257, HW11258, HW11259, HW11260, HW11262, HW11263, HW11264, HW11266, HW11267, HW11269, HW11271, HW11272, HW11274, HW11275, HW11276, HW11277, HW11279, HW11280, HW11281, HW11283, HW11284, HW11286, HW11287, HW11288, HW11289, HW11290, HW11291, HW11292, HW11293, HW11294, HW11295, hw11296, HW11298, HW11299, HW11302, HW11303, HW11305, HW11306, HW11307, HW11309, HW11310, HW11311, HW11312, HW11313, HW11314, HW11316, HW11317, HW11318, HW11319, HW11320, HW11321, HW11324, HW11326, HW11327, HW11328, HW11329, HW11330, HW11331, HW11332, HW11333, HW11334, HW11335, HW11336, HW11337, HW11338, HW11339, HW11340, HW11342, HW11343, HW11347, HW11348, HW11349, HW11350, HW11351, HW11352, HW11353, HW11354, HW11355, HW11356, HW11357, HW11358, HW11359, HW11360, HW11362, HW11364, HW11365, HW11366, HW11368, HW11369, HW11370, HW11372, HW11373, HW11374, HW11377, HW11378, HW11379, HW11380, HW11383, HW11384, HW11385, HW11386, HW11387, HW11388, HW11389, HW11391, HW11392, HW11393, HW11394, HW11396, HW11398, HW11399, HW11401, HW11402, HW11403, HW11404, HW11405, HW11408, HW11410, HW11411, HW11412, HW11413, HW11415, HW11418, HW11419, HW11421, HW11422, HW11423, HW11424, HW11426, HW11427, HW11430, HW11432, HW11433, HW11434, HW11435, HW11436, HW11437, HW11438, HW11440, HW11441, HW11442, HW11443, HW11444, HW11445, HW11446, HW11448, HW11449, HW11450, HW11452, HW11453, HW11454, HW11456, HW11458, HW11459, HW11460, HW11463, HW11465, HW11466, HW11467, HW11468, HW11469, HW11470, HW11471, HW11472, HW11473, HW11474, HW11476, HW11477, HW11478, HW11480, HW11481, HW11482, HW11485, HW11487, HW11488, HW11489, HW11490, HW11492, HW11493, HW11494, HW11495, HW11497, HW11498, HW11499, HW11500, HW11501, HW11503, HW11504, HW11505, HW11506, HW11509, HW11510, HW11514, HW11515, HW11516, HW11517, HW11518, HW11519, HW11520, HW11521, HW11522, HW11523, HW11524, HW11525, HW11526, HW11527, HW11528, HW11529, HW11530, HW11532, HW11533, HW11534, HW11535, HW11536, HW11537, HW11538, HW11539, HW11540, HW11541, HW11542, HW11543, HW11545, HW11547, HW11548, HW11549, HW11554, HW11555, HW11557, HW11558, HW11559, HW11561, HW11563, HW11564, HW11565, HW11569, HW11572, HW11574, HW11575, HW11576, HW11577, HW11579, HW11580, HW11581, HW11582, HW11585, HW11586, HW11587, HW11588, HW11589, HW11591, HW11593, HW11594, HW11595, HW11596, HW11597, HW11598, HW11599, HW11600, HW11601, HW11602, HW11604, HW11605, HW11606, HW11607, HW11609, HW11610, HW11611, HW11612, HW11613, HW11614, HW11615, HW11616, HW11619, HW11620, HW11622, HW11624, HW11625, HW11626, HW11627, HW11628, HW11630, HW11631, HW11632, HW11633, HW11634, HW11636, HW11637, HW11638, HW11639, HW11640, HW11641, HW11643, HW11645, HW11646, HW11647, HW11648, HW11650, HW11651, HW11653, HW11654, HW11655, HW11657, HW11658, HW11659, HW11660, HW11662, HW11664, HW11665, HW11666, HW11668, HW11669, HW11670, HW11671, HW11672, HW11674, HW11675, HW11677, HW11678, HW11679, HW11680, HW11681, HW11682, HW11683, HW11684, HW11685, HW11686, HW11687, HW11688, HW11689, HW11690, HW11691, HW11692, HW11693, HW11694, HW11695, HW11699, HW11700, HW11701, HW11702, HW11703, HW11705, HW11707, HW11708, HW11709, HW11710, HW11711, HW11713, HW11714, HW11716, HW11717, HW11718, HW11719, HW11721, HW11722, HW11724, HW11725, HW11726, HW11727, HW11728, HW11729, HW11730, HW11731, HW11732, HW11734, HW11735, HW11736, HW11737, HW11738, HW11740, HW11742, HW11743, HW11744, HW11745, HW11746, HW11747, HW11748, HW11749, HW11750, HW11751, HW11752, HW11753, HW11754, HW11755, HW11756, HW11757, HW11758, HW11760, HW11762, HW11765, HW11766, HW11767, HW11768, HW11769, HW11770, HW11771, HW11772, HW11773, HW11774, HW11775, HW11776, HW11777, HW11778, HW11779, HW11780, HW11783, HW11785, HW11787, HW11788, HW11789, HW11790, HW11791, HW11792, HW11793, HW11795, HW11796, HW11797, HW11798, HW11799, HW11800, HW11801, HW11802, HW11804, HW11805, HW11807, HW11809, HW11810, HW11811, HW11812, HW11813, HW11814, HW11815, HW11816, HW11817, HW11820, HW11822, HW11823, HW11824, HW11825, HW11826, HW11827, HW11828, HW11830, HW11831, HW11832, HW11834, HW11836, HW11837, HW11838, HW11846, HW11847, HW11848, HW11849, HW11850, HW11851, HW11853, HW11854, HW11855, HW11857, HW11858, HW11859, HW11860, HW11863, HW11864, HW11865, HW11866, HW11868, HW11870, HW11872, HW11873, HW11874, HW11875, HW11876, HW11877, HW11878, HW11879, HW11880, HW11881, HW11882, HW11883, HW11884, HW11886, HW11887, HW11888, HW11889, HW11890, HW11891, HW11892, HW11893, HW11894, HW11895, HW11896, HW11897, HW11898, HW11899, HW11900, HW11901, HW11902, HW11903, HW11904, HW11905, HW11907, HW11908, HW11909, HW11911, HW11912, HW11913, HW11914, HW11915, HW11916, HW11917, HW11919, HW11920, HW11921, HW11922, HW11923, HW11924, HW11925, HW11926, HW11927, HW11928, HW11930, HW11931, HW11932, HW11934, HW11935, HW11938, HW11939, HW11941, HW11942, HW11943, HW11944, HW11945, HW11946, HW11947, HW11948, HW11949, HW11950, HW11951, HW11952, HW11953, HW11954, HW11955, HW11956, HW11957, HW11959, HW11961, HW11962, HW11963, HW11964, HW11965, HW11966, HW11967, HW11968, HW11969, HW11970, HW11972, HW11973, HW11975, HW11976, HW11977, HW11980, HW11981, HW11982, HW11985, HW11986, HW11987, HW11988, HW11989, HW11991, HW11993, HW11994, HW11995, HW11996, HW11997, HW11998, HW12000, HW12001, HW12002, HW12004, HW12005, HW12006, HW12007, HW12008, HW12009, HW12010, HW12011, HW12012, HW12013, HW12014, HW12015, HW12016, HW12017, HW12018, HW12021, HW12023, HW12024, HW12025, HW12026, HW12027, HW12029, HW12030, HW12031, HW12033, HW12036, HW12038, HW12039, HW12040, HW12041, HW12042, HW12043, HW12046, HW12047, HW12048, HW12049, HW12050, HW12052, HW12053, HW12054, HW12056, HW12057, HW12058, HW12060, HW12061, HW12062, HW12064, HW12065, HW12066, HW12067, HW12069, HW12070, HW12071, HW12072, HW12073, HW12074, HW12075, HW12076, HW12077, HW12078, HW12079, HW12080, HW12081, HW12082, HW12084, HW12085, HW12086, HW12087, HW12088, HW12089, HW12090, HW12091, HW12092, HW12093, HW12094, HW12095, HW12096, HW12097, HW12098, HW12099, HW12101, HW12102, HW12105, HW12106, HW12107, HW12108, HW12110, HW12111, HW12113, HW12114, HW12116, HW12117, HW12118, HW12119, HW12121, HW12122, HW12123, HW12124, HW12125, HW12126, HW12127, HW12128, HW12129, HW12130, HW12131, HW12132, HW12133, HW12134, HW12136, HW12137, HW12138, HW12140, HW12141, HW12142, HW12143, HW12144, HW12145, HW12146, HW12147, HW12148, HW12149, HW12150, HW12151, HW12152, HW12153, HW12155, HW12156, HW12157, HW12158, HW12159, HW12160, HW12161, HW12162, HW12163, HW12165, HW12167, HW12169, HW12170, HW12171, HW12172, HW12174, HW12175, HW12176, HW12177, HW12180, HW12181, HW12183, HW12184, HW12185, HW12186, HW12188, HW12190, HW12191, HW12192, HW12194, HW12195, HW12196, HW12197, HW12198, HW12200, HW12202, HW12203, HW12204, HW12206, HW12207, HW12208, HW12209, HW12210, HW12211, HW12213, HW12214, HW12216, HW12217, HW12218, HW12219, HW12220, HW12221, HW12223, HW12224, HW12225, HW12226, HW12227, HW12228, HW12229, HW12230, HW12231, HW12232, HW12233, HW12234, HW12235, HW12236, HW12237, HW12238, HW12239, HW12240, HW12241, HW12242, HW12243, HW12244, HW12246, HW12247, HW12249, HW12250, HW12252, HW12253, HW12254, HW12256, HW12257, HW12258, HW12259, HW12260, HW12261, HW12262, HW12264, HW12265, HW12266, HW12267, HW12268, HW12269, HW12270, HW12271, HW12274, HW12275, HW12276, HW12277, HW12278, HW12279, HW12280, HW12282, HW12283, HW12284, HW12285, HW12286, HW12287, HW12288, HW12289, HW12290, HW12292, HW12293, HW12294, HW12295, HW12296, HW12297, HW12298, HW12299, HW12300, HW12301, HW12302, HW12303, HW12304, HW12305, HW12306, HW12307, HW12308, HW12309, HW12310, HW12311, HW12312, HW12313, HW12314, HW12315, HW12317, HW12318, HW12320, HW12321, HW12322, HW12323, HW12324, HW12325, HW12326, HW12327, HW12328, HW12330, HW12331, HW12332, HW12333, HW12334, HW12335, HW12336, HW12337, HW12338, HW12339, HW12340, HW12342, HW12343, HW12344, HW12345, HW12346, HW12347, HW12349, HW12350, HW12351, HW12353, HW12354, HW12355, HW12356, HW12357, HW12358, HW12359, HW12361, HW12362, HW12363, HW12364, HW12366, HW12367, HW12368, HW12369, HW12370, HW12371, HW12372, HW12373, HW12374, HW12376, HW12377, HW12378, HW12380, HW12382, HW12383, HW12384, HW12385, HW12386, HW12387, HW12388, HW12389, HW12390, HW12392, HW12393, HW12394, HW12395, HW12397, HW12398, HW12399, HW12400, HW12401, HW12402, HW12403, HW12404, HW12405, HW12406, HW12407, HW12408, HW12409, HW1375, HW1378, HW1379, HW1381, HW1390, HW1872, HW1873, HW1886, HW1887, HW1890, HW1893, HW1904, HW1998, HW1999, HW2000, HW2001, HW2004, HW2005, HW2006, HW2007, HW2008, HW2010, HW2011, HW2012, HW2014, HW2015, HW2016, HW2017, HW2034, HW2046, HW2048, HW2049, HW2050, HW2051, HW2052, HW2053, HW2054, HW2056, HW2057, HW2059, HW2060, HW2063, HW2065, HW2068, HW2070, HW2071, HW2072, HW2074, HW2076, HW2077, HW2078, HW2080, HW2081, HW2082, HW2083, HW2088, HW2090, HW2091, HW2093, HW2094, HW2095, HW2106, HW2107, HW2108, HW2109, HW2116, HW2117, HW2118, HW2119, HW2120, HW2121, HW2123, HW2126, HW2127, HW2128, HW2129, HW2130, HW2131, HW2132, HW2133, HW2134, HW2137, HW21370, HW2138, HW2139, HW2141, HW2142, HW2143, HW21447, HW2146, HW2147, HW21488, HW2149, HW2150, HW21507, HW2151, HW21527, HW21528, HW21529, HW21530, HW21532, HW21533, HW21534, HW21535, HW2154, HW21541, HW21542, HW21544, HW21548, HW21549, HW2155, HW21550, HW21551, HW21552, HW21553, HW21556, HW21559, HW21560, HW21562, HW21564, HW21565, HW21566, HW21567, HW2157, HW21570, HW21573, HW21575, HW21576, HW21577, HW21578, HW21579, HW21580, HW21581, HW21582, HW21585, HW21586, HW21589, HW21590, HW21592, HW21599, HW2160, HW21608, HW21609, HW2161, HW21610, HW21611, HW21612, HW21614, HW21615, HW21617, HW21618, HW21619, HW21621, HW21624, HW21625, HW21626, HW21627, HW21628, HW21629, HW21630, HW21631, HW21632, HW21633, HW21634, HW21635, HW21638, HW21662, HW21663, HW21673, HW21674, HW21678, HW21679, HW21680, HW21685, HW21686, HW21687, HW21688, HW21691, HW21694, HW21695, HW21696, HW21699, HW21701, HW21702, HW21703, HW21704, HW21705, HW21706, HW21707, HW21708, HW21711, HW21712, HW21713, HW21714, HW21716, HW21717, HW21718, HW21719, HW21720, HW21721, HW21726, HW21730, HW21731, HW21732, HW21733, HW21735, HW21736, HW21737, HW21738, HW21739, HW21741, HW21743, HW21744, HW21745, HW2177, HW2178, HW2180, HW2181, HW2182, HW2184, HW2185, HW2186, HW2187, HW2188, HW2189, HW21942, HW21943, HW21946, HW21948, HW21949, HW21950, HW21951, HW21952, HW21953, HW21954, HW21956, HW21957, HW21958, HW21959, HW21960, HW21961, HW21962, HW21963, HW21965, HW21966, HW21967, HW21969, HW21970, HW21971, HW21972, HW21973, HW21975, HW21976, HW21977, HW21979, HW21980, HW21981, HW21983, HW21984, HW21985, HW21986, HW21987, HW21988, HW21989, HW21990, HW21991, HW21993, HW21994, HW21995, HW21996, HW21997, HW21999, HW22000, HW22001, HW22002, HW22004, HW22005, HW22006, HW22007, HW22008, HW22009, HW22010, HW22011, HW22012, HW22013, HW22014, HW22015, HW22016, HW22017, HW22018, HW22019, HW22020, HW22021, HW22022, HW22023, HW22024, HW22025, HW22026, HW22027, HW22028, HW2203, HW22030, HW22031, HW22033, HW22034, HW22035, HW22037, HW22038, HW22039, HW22040, HW22041, HW22042, HW22043, HW22044, HW22045, HW22046, HW22047, HW22048, HW22049, HW2205, HW22050, HW22051, HW22052, HW22053, HW22054, HW22055, HW22057, HW22059, HW22060, HW22061, HW22062, HW22063, HW22064, HW22065, HW22067, HW22069, HW22070, HW22071, HW22072, HW22074, HW22075, HW22076, HW22077, HW22079, HW2208, HW22080, HW22081, HW22083, HW22084, HW22085, HW22086, HW22087, HW22088, HW22089, HW2209, HW22090, HW22091, HW22092, HW22093, HW22095, HW22097, HW22098, HW2210, HW22100, HW22101, HW22102, HW22103, HW22104, HW22106, HW22107, HW22108, HW22109, HW2211, HW22110, HW22111, HW22112, HW22114, HW22116, HW22117, HW22118, HW22119, HW2212, HW22120, HW22121, HW22122, HW22124, HW22125, HW22126, HW22127, HW22128, HW22129, HW22130, HW22131, HW22132, HW22134, HW22135, HW22136, HW22138, HW22141, HW22142, HW22143, HW22146, HW22147, HW22148, HW22149, HW22151, HW22152, HW22153, HW22154, HW22155, HW22156, HW22157, HW22158, HW22160, HW22161, HW22162, HW22163, HW22164, HW22165, HW22166, HW22167, HW22169, HW22170, HW22171, HW22172, HW22173, HW22174, HW22175, HW22176, HW22177, HW22178, HW22184, HW22202, HW22204, HW22205, HW22206, HW22207, HW22210, HW22211, HW22212, HW22213, HW22214, HW22215, HW22216, HW22217, HW22220, HW22223, HW22224, HW22228, HW22246, HW2225, HW22261, HW22262, HW22263, HW22265, HW22266, HW22268, HW22269, HW22270, HW22271, HW22272, HW22273, HW22275, HW22277, HW22278, HW22279, HW22281, HW22282, HW22283, HW22284, HW22285, HW22286, HW22287, HW22289, HW22290, HW22291, HW22292, HW22293, HW2230, HW22318, HW22319, HW22320, HW22323, HW22324, HW22326, HW22337, HW22339, HW2234, HW22340, HW22341, HW22342, HW22343, HW22344, HW22345, HW22346, HW22349, HW2235, HW22350, HW22351, HW22352, HW22353, HW22354, HW22355, HW22358, HW22359, HW2236, HW22360, HW22388, HW22393, HW22394, HW2240, HW22405, HW22409, HW2241, HW2242, HW22428, HW2243, HW22449, HW2245, HW22452, HW22453, HW22454, HW22455, HW22456, HW22457, HW22459, HW22460, HW22461, HW22462, HW22463, HW22466, HW22467, HW22468, HW22470, HW22472, HW22473, HW22474, HW22476, HW22477, HW22478, HW22479, HW22480, HW22481, HW22482, HW22484, HW22485, HW22487, HW22488, HW22489, HW22490, HW22491, HW22492, HW22493, HW22494, HW22497, HW22514, HW22528, HW22531, HW22535, HW22536, HW22538, HW22539, HW22540, HW22546, HW22547, HW22552, HW22553, HW22554, HW22556, HW22557, HW22558, HW22559, HW2256, HW22560, HW22561, HW22562, HW22563, HW22565, HW22566, HW22567, HW22568, HW22569, HW22570, HW22571, HW22572, HW22574, HW22575, HW22576, HW22577, HW22578, HW22579, HW2258, HW22584, HW22585, HW2259, HW2260, HW22606, HW22607, HW2261, HW22610, HW22611, HW22612, HW22613, HW22614, HW22615, HW22617, HW22619, HW22620, HW22621, HW22622, HW22623, HW22624, HW22626, HW22629, HW2263, HW22637, HW22639, HW2264, HW22643, HW22644, HW2265, HW22653, HW22655, HW22656, HW22657, HW22658, HW2266, HW22660, HW22661, HW22662, HW22663, HW22664, HW22666, HW22667, HW22668, HW22669, HW22670, HW22673, HW22674, HW22675, HW2268, HW2269, HW2270, HW22704, HW22706, HW22711, HW22715, HW22716, HW22717, HW22720, HW22723, HW22725, HW22726, HW22727, HW22728, HW22730, HW22732, HW22733, HW22734, HW22735, HW22736, HW22737, HW22738, HW22739, HW22740, HW22741, HW22742, HW22743, HW22746, HW22748, HW22749, HW22750, HW22751, HW22752, HW22753, HW22754, HW22755, HW22757, HW22759, HW22760, HW22762, HW22763, HW22768, HW22769, HW22770, HW22771, HW22772, HW22773, HW22774, HW22778, HW22793, HW22795, HW22796, HW22798, HW22799, HW22800, HW22803, HW22804, HW22805, HW22806, HW22807, HW22809, HW2281, HW22813, HW22814, HW22816, HW22817, HW22818, HW22819, HW22820, HW22821, HW22822, HW22823, HW2283, HW2286, HW2287, HW22873, HW22874, HW22875, HW22876, HW22877, HW22878, HW22879, HW2288, HW22880, HW22881, HW22882, HW22883, HW22885, HW22887, HW22891, HW22894, HW22895, HW22896, HW22897, HW22898, HW2291, HW2292, HW22924, HW22926, HW22927, HW22928, HW22932, HW22933, HW22934, HW22936, HW22937, HW22938, HW22951, HW2296, HW22972, HW22973, HW22974, HW22976, HW22978, HW2298, HW22980, HW22981, HW22982, HW22984, HW22985, HW22986, HW22989, HW22990, HW22991, HW22992, HW22993, HW22994, HW22995, HW22996, HW22997, HW22998, HW22999, HW2300, HW23000, HW23001, HW23002, HW23003, HW23004, HW23005, HW23008, HW23010, HW23014, HW2306, HW23063, HW23065, HW23066, HW23069, HW23071, HW23072, HW23073, HW23074, HW23075, HW23076, HW23079, HW23080, HW23082, HW23083, HW23084, HW23086, HW23087, HW23088, HW23089, HW23090, HW23091, HW23092, HW23095, HW23096, HW23097, HW23098, HW23099, HW23100, HW23102, HW23103, HW23104, HW23107, HW23108, HW23110, HW23113, HW23114, HW23115, HW23118, HW23119, HW2312, HW23120, HW23121, HW23122, HW23123, HW23139, HW23148, HW23149, HW23150, HW23151, HW23152, HW23153, HW23154, HW23156, HW23157, HW23158, HW23159, HW23161, HW23163, HW23164, HW23165, HW23167, HW23169, HW23170, HW23171, HW23172, HW23173, HW23175, HW23177, HW23178, HW23179, HW2318, HW23181, HW23182, HW23183, HW23184, HW23185, HW23186, HW2319, HW23193, HW2320, HW23207, HW2322, HW23221, HW23222, HW23224, HW23225, HW23227, HW23229, HW2323, HW23230, HW23231, HW23232, HW23233, HW23234, HW23235, HW23236, HW23237, HW23240, HW23241, HW23242, HW23244, HW23245, HW23246, HW23247, HW23248, HW23249, HW2325, HW23251, HW23252, HW23290, HW23291, HW23292, HW23293, HW23294, HW23295, HW23299, HW23300, HW23302, HW23303, HW23304, HW23305, HW23306, HW23307, HW23308, HW23310, HW23312, HW23313, HW23314, HW23316, HW23318, HW23319, HW23320, HW23321, HW23364, HW23383, HW23384, HW23385, HW23386, HW23389, HW2339, HW23390, HW23391, HW23392, HW23393, HW23394, HW23395, HW23397, HW23398, HW23399, HW23407, HW23408, HW23409, HW23410, HW23411, HW23412, HW23413, HW23414, HW23416, HW23417, HW23418, HW23419, HW23420, HW23421, HW23422, HW23423, HW23424, HW23425, HW23426, HW23427, HW23428, HW23429, HW23430, HW23431, HW23432, HW23434, HW23436, HW23437, HW23441, HW23448, HW2345, HW23459, HW23461, HW23462, HW23463, HW23465, HW23467, HW23468, HW23469, HW23470, HW23471, HW23472, HW23473, HW23474, HW23476, HW23477, HW23478, HW23479, HW23480, HW23481, HW23482, HW23483, HW23484, HW23486, HW23487, HW23488, HW23489, HW23490, HW23491, HW23493, HW23494, HW23495, HW23496, HW23497, HW23498, HW23499, HW23500, HW23501, HW23502, HW23503, HW23504, HW23505, HW2351, HW23528, HW23529, HW23530, HW23531, HW23532, HW23533, HW23534, HW23535, HW23536, HW23537, HW23538, HW2354, HW23540, HW23543, HW23544, HW23545, HW23546, HW23547, HW2355, HW23554, HW23555, HW23556, HW23557, HW23558, HW23559, HW2356, HW23560, HW23561, HW23562, HW23563, HW23564, HW23566, HW23567, HW23568, HW23570, HW23571, HW23572, HW23597, HW23599, HW23600, HW23601, HW23602, HW23603, HW23605, HW23606, HW2361, HW23618, HW23620, HW23621, HW23622, HW23623, HW23624, HW23626, HW23627, HW23628, HW23629, HW2363, HW23630, HW23631, HW23632, HW23633, HW23635, HW23637, HW23641, HW2366, HW23675, HW23678, HW23685, HW23686, HW23689, HW23691, HW23693, HW23694, HW23696, HW23697, HW23698, HW23700, HW23701, HW23703, HW23704, HW23708, HW23709, HW23723, HW23725, HW23726, HW23727, HW23729, HW23730, HW23732, HW23746, HW23755, HW23756, HW23757, HW23758, HW23759, HW23761, HW23762, HW23763, HW23764, HW23765, HW23768, HW23769, HW23770, HW23771, HW23772, HW23773, HW23774, HW23775, HW23777, HW23779, HW23780, HW23781, HW23782, HW23783, HW23784, HW23786, HW23787, HW23789, HW23790, HW23792, HW23794, HW23795, HW23798, HW23799, HW23800, HW23801, HW23802, HW23804, HW23805, HW23807, HW23808, HW23809, HW23810, HW23812, HW23813, HW23815, HW23816, HW23817, HW23821, HW23823, HW23824, HW23826, HW23827, HW23828, HW23829, HW23830, HW23831, HW23833, HW23834, HW23836, HW23845, HW23846, HW23847, HW23850, HW23851, HW23853, HW23854, HW23855, HW23857, HW23858, HW23859, HW23860, HW23861, HW23877, HW23879, HW2388, HW23880, HW23881, HW23882, HW23884, HW23885, HW23886, HW23887, HW23888, HW23889, HW23890, HW23891, HW23892, HW23893, HW23894, HW23895, HW23896, HW23897, HW23898, HW23900, HW23901, HW23903, HW23904, HW23908, HW23909, HW23933, HW23939, HW23940, HW23941, HW23942, HW23943, HW23944, HW23945, HW23946, HW23947, HW23948, HW23950, HW23951, HW23952, HW23954, HW23955, HW23956, HW23957, HW23958, HW23959, HW23961, HW23962, HW23964, HW23965, HW23988, HW23989, HW23990, HW23991, HW23992, HW23993, HW23995, HW23996, HW23998, HW23999, HW24000, HW24001, HW24002, HW24003, HW24005, HW24006, HW24007, HW24008, HW24009, HW24010, HW24011, HW24012, HW24013, HW24034, HW24045, HW24048, HW24049, HW24050, HW24051, HW24052, HW24053, HW24054, HW24055, HW24057, HW24058, HW24059, HW24061, HW24062, HW24063, HW24064, HW24066, HW24067, HW24070, HW24071, HW24072, HW24073, HW24075, HW24076, HW24077, HW24105, HW24106, HW24107, HW24108, HW24109, HW24111, HW24112, HW24113, HW24116, HW24117, HW24118, HW24119, HW24120, HW24121, HW24123, HW24127, HW24128, HW24130, HW24131, HW24132, HW24134, HW24144, HW2415, HW24169, HW24172, HW24179, HW24194, HW24195, HW24197, HW24199, HW24200, HW24201, HW24202, HW24203, HW24204, HW24205, HW24206, HW24207, HW24208, HW24210, HW24212, HW24214, HW24215, HW24216, HW24217, HW24218, HW24244, HW24245, HW24246, HW24247, HW24248, HW24250, HW24251, HW24253, HW24254, HW24255, HW24256, HW24257, HW24260, HW24262, HW24263, HW24294, HW2430, HW24308, HW24309, HW24310, HW24312, HW24313, HW24314, HW24315, HW24316, HW24317, HW24318, HW24326, HW24327, HW24328, HW24331, HW24333, HW24336, HW24337, HW24338, HW24339, HW2434, HW24341, HW24342, HW24343, HW24344, HW24345, HW24347, HW24348, HW24349, HW24352, HW24353, HW2438, HW2443, HW24434, HW2444, HW2446, HW2447, HW24496, HW2450, HW24500, HW24501, HW24502, HW24503, HW24504, HW24505, HW24507, HW24508, HW2451, HW24510, HW24511, HW24515, HW24516, HW24517, HW24518, HW24519, HW2452, HW24520, HW24521, HW24522, HW24523, HW24529, HW24530, HW24531, HW24532, HW24534, HW24535, HW2454, HW2455, HW2456, HW24594, HW24595, HW24596, HW24597, HW24599, HW24600, HW24601, HW24602, HW24603, HW24605, HW24608, HW24609, HW24611, HW24612, HW24613, HW24615, HW24618, HW2466, HW24662, HW24678, HW24683, HW24685, HW24686, HW24690, HW24700, HW24711, HW24724, HW24725, HW24726, HW24727, HW24728, HW24730, HW24731, HW24732, HW24734, HW24735, HW24737, HW24738, HW24739, HW24740, HW24742, HW24743, HW24744, HW24745, HW24746, HW24747, HW24748, HW24750, HW24751, HW24752, HW24753, HW24754, HW24764, HW24765, HW24766, HW24767, HW24768, HW24769, HW24770, HW24771, HW24773, HW24775, HW24776, HW24779, HW24780, HW24781, HW24782, HW24783, HW24785, HW24787, HW24790, HW24791, HW24806, HW2481, HW24813, HW24814, HW24815, HW24816, HW24817, HW24818, HW24819, HW24820, HW24821, HW24822, HW24823, HW24824, HW24825, HW24827, HW24828, HW24829, HW24830, HW24832, HW24837, HW24846, HW24847, HW24850, HW24852, HW24853, HW24854, HW24855, HW24858, HW24859, HW24860, HW24861, HW24862, HW24863, HW24864, HW24866, HW24869, HW24870, HW24871, HW24872, HW24873, HW24874, HW24876, HW24877, HW24879, HW24880, HW24881, HW24882, HW24884, HW24885, HW24887, HW24888, HW24892, HW24893, HW24894, HW24896, HW24898, HW24899, HW24902, HW24903, HW24914, HW24915, HW24916, HW24917, HW24918, HW24919, HW24920, HW24921, HW24922, HW24923, HW24943, HW24944, HW24946, HW24947, HW24948, HW24949, HW24950, HW24952, HW24953, HW24954, HW24956, HW24960, HW24961, HW24962, HW2497, HW24974, HW24976, HW24977, HW24978, HW24979, HW24980, HW24981, HW24982, HW24983, HW24987, HW24993, HW25000, HW25003, HW25006, HW25007, HW25008, HW25009, HW25010, HW25011, HW25012, HW25013, HW25015, HW25017, HW25018, HW25019, HW2502, HW25020, HW25022, HW25023, HW25024, HW25025, HW25026, HW25027, HW25028, HW25029, HW2503, HW25031, HW25032, HW25033.HW25034, HW25035, HW25037, HW25038, HW25039, HW2504, HW25040, HW25041, HW25042, HW25043, HW25045, HW25046, HW25047, HW25048, HW25049, HW25050, HW25051, HW25052, HW25053, HW25054, HW25055, HW25056, HW25057, HW25058, HW25060, HW25063, HW25064, HW25066, HW25068, HW25069, HW25070, HW25071, HW25072, HW25073, HW25074, HW25075, HW25077, HW25078, HW2508, HW25098, HW25099, HW25100, HW25117, HW25118, HW25119, HW25120, HW25122, HW25123, HW25125, HW25127, HW25128, HW25129, HW25130, HW25132, HW25134, HW25135, HW25136, HW25149, HW25151, HW25152, HW25153, HW25156, HW25157, HW25164, HW2517, HW25171, HW25176, HW25178, HW25179, HW25181, HW25182, HW25183, HW25184, HW25186, HW25187, HW25188, HW25189, HW25190, HW25191, HW25192, HW25194, HW25196, HW25197, HW25198, HW25199, HW25200, HW25202, HW25203, HW25204, HW25205, HW25206, HW25207, HW25208, HW2521, HW25211, HW25213, HW25214, HW25216, HW25217, HW25219, HW2522, HW25220, HW25223, HW25224, HW25226, HW25227, HW25229, HW25231, HW25239, HW2524, HW25240, HW25242, HW25253, HW25254, HW25256, HW25263, HW25264, HW25265, HW25267, HW25268, HW25273, HW25275, HW25276, HW25279, HW25281, HW25282, HW25284, HW25285, HW25286, HW25287, HW25288, HW25289, HW25290, HW25291, HW25292, HW25314, HW25316, HW25318, HW2532, HW25320, HW25321, HW25322, HW25323, HW25324, HW25326, HW25327, HW25328, HW25330, HW25331, HW25333, HW2534, HW25342, HW25349, HW2535, HW25350, HW25351, HW2538, HW25403, HW25408, HW2541, HW25410, HW25412, HW25413, HW25414, HW25415, HW25416, HW25417, HW25418, HW2542, HW25420, HW25422, HW25423, HW25424, HW25425, HW25426, HW25428, HW25430, HW25432, HW25434, HW25435, HW25436, HW25437, HW25438, HW2544, HW25440, HW25441, HW25442, HW25443, HW25444, HW25445, HW25446, HW25447, HW25448, HW25449, HW25456, HW25458, HW25459, HW2546, HW25460, HW25463, HW2547, HW25477, HW25479, HW2548, HW25480, HW25482, HW25484, HW25485, HW2549, HW25494, HW25496, HW25497, HW25498, HW25499, HW25500, HW25501, HW25503, HW25504, HW25506, HW25507, HW25508, HW25509, HW2551, HW25511, HW25512, HW25515, HW2553, HW25532, HW25541, HW25543, HW2555, HW2556, HW25645, HW25646, HW25647, HW25648, HW25649, HW25650, HW25652, HW25654, HW25656, HW25658, HW25659, HW25660, HW25663, HW25670, HW25679, HW25683, HW25684, HW25693, HW25696, HW25707, HW25710, HW25711, HW25713, HW25714, HW25716, HW25717, HW25719, HW25720, HW25722, HW25723, HW25724, HW25725, HW25726, HW25737, HW25754, HW25756, HW25760, HW25773, HW25774, HW25777, HW25778, HW25779, HW25781, HW25782, HW25783, HW25784, HW25785, HW25786, HW25787, HW2581, HW25810, HW25815, HW25824, HW2583, HW25838, HW25839, HW25848, HW25854, HW25855, HW25856, HW25857, HW25858, HW25860, HW25863, HW25865, HW25866, HW25868, HW25869, HW25870, HW25874, HW25894, HW2590, HW25909, HW2591, HW25910, HW25912, HW25913, HW25915, HW25916, HW25923, HW25936, HW25937, HW25939, HW25940, HW25941, HW25942, HW25943, HW25945, HW25946, HW25947, HW25948, HW25949, HW25950, HW25952, HW25953, HW25955, HW25956, HW25957, HW25958, HW25959, HW2596, HW25961, HW25962, HW25963, HW25964, HW25965, HW25966, HW25967, HW25969, HW25970, HW25971, HW25972, HW25973, HW25974, HW25975, HW25976, HW25978, HW25979, HW2598, HW25981, HW25982, HW25983, HW25984, HW25985, HW25986, HW25987, HW25988, HW2599, HW25991, HW25992, HW25993, HW25995, HW25997, HW25998, HW25999, HW26001, HW26003, HW26004, HW26005, HW26007, HW26008, HW26009, HW26010, HW26013, HW26014, HW26016, HW26017, HW26018, HW26019, HW26020, HW26022, HW26023, HW26024, HW26025, HW26027, HW26028, HW26029, HW26030, HW26033, HW26035, HW26036, HW26038, HW26039, HW26042, HW26043, HW26052, HW26054, HW26056, HW26057, HW26061, HW26062, HW26063, HW26064, HW26065, HW26066, HW26068, HW26071, HW26072, HW26075, HW26081, HW26082, HW26083, HW26084, HW26085, HW26087, HW26088, HW26089, HW26090, HW26091, HW26092, HW26107, HW26117, HW26118, HW26123, HW26127, HW26129, HW26130, HW26131, HW26132, HW26133, HW26134, HW26135, HW26136, HW26137, HW26139, HW26141, HW26142, HW26144, HW26152, HW26154, HW26161, HW2617, HW26174, HW26175, HW26176, HW26177, HW26178, HW26179, HW26181, HW26182, HW26183, HW26184, HW26185, HW26186, HW26187, HW26188, HW26190, HW26191, HW26192, HW26193, HW26194, HW26195, HW26197, HW26198, HW2622, HW26247, HW2625, HW2627, HW2629, HW2630, HW26300, HW26311, HW26312, HW26313, HW26314, HW26316, HW26317, HW26318, HW26319, HW26320, HW26322, HW26323, HW26324, HW26326, HW26327, HW26328, HW2633, HW26338, HW26339, HW26340, HW26341, HW26342, HW26343, HW26345, HW26346, HW26347, HW26348, HW26349, HW2635, HW26350, HW26351, HW26352, HW26353, HW26354, HW26355, HW26356, HW26357, HW26358, HW26360, HW26361, HW26364, HW26365, HW26367, HW26368, HW2637, HW26370, HW26372, HW26373, HW26374, HW26375, HW26376, HW26377, HW26378, HW26401, HW26402, HW26404, HW26405, HW26406, HW26407, HW26408, HW2641, HW26410, HW26411, HW26412, HW26413, HW26414, HW26415, HW26416, HW26417, HW26418, HW26420, HW26422, HW26423, HW26424, HW26425, HW26427, HW26429, HW26431, HW26432, HW26433, HW26434, HW26435, HW2646, HW26460, HW26461, HW26463, HW26465, HW26466, HW26468, HW26469, HW26470, HW26471, HW26472, HW26473, HW26474, HW26475, HW26476, HW26477, HW26478, HW26479, HW2648, HW26480, HW26481, HW26482, HW26483, HW26485, HW26486, HW26487, HW26488, HW26489, HW26492, HW26493, HW26496, HW26497, HW26508, HW26516, HW26518, HW26519, HW26520, HW26521, HW26522, HW26523, HW26524, HW26525, HW26526, HW26528, HW26529, HW2653, HW26530, HW26531, HW26532, HW26534, HW26535, HW26536, HW26537, HW26538, HW26539, HW2654, HW26540, HW26541, HW26542, HW26543, HW26544, HW26545, HW26546, HW26548, HW2655, HW26564, HW26576, HW26577, HW26578, HW26579, HW26580, HW26581, HW26583, HW26584, HW26585, HW26586, HW26587, HW26589, HW2659, HW26590, HW26591, HW26592, HW26593, HW26594, HW26596, HW26597, HW26598, HW26599, HW2660, HW26601, HW26603, HW26604, HW26605, HW26606, HW26607, HW26609, HW26625, HW26626, HW26627, HW26628, HW26629, HW26630, HW26631, HW26632, HW26633, HW26634, HW26635, HW26636, HW26637, HW26638, HW26639, HW26640, HW26641, HW26642, HW26643, HW26644, HW26645, HW26646, HW26654, HW26666, HW26667, HW26668, HW2667, HW26670, HW26672, HW26673, HW26674, HW26675, HW26677, HW26678, HW26679, HW26680, HW26681, HW26682, HW26683, HW26685, HW26686, HW26687, HW26688, HW26693, HW26694, HW26695, HW26705, HW26710, HW26711, HW26712, HW26713, HW26716, HW26717, HW26718, HW26720, HW26722, HW26723, HW26725, HW26726, HW26727, HW26728, HW26729, HW26737, HW26738, HW26739, HW26745, HW2675, HW26752, HW26756, HW26757, HW26758, HW26759, HW26760, HW26764, HW26767, HW26768, HW26770, HW26771, HW26772, HW26774, HW26775, HW26776, HW26777, HW26779, HW26782, HW26783, HW26784, HW26785, HW26786, HW26799, HW26800, HW26801, HW26802, HW26803, HW26804, HW26805, HW26806, HW26807, HW26808, HW26809, HW26810, HW26811, HW26812, HW26814, HW26816, HW26817, HW26818, HW26819, HW26820, HW26822, HW26823, HW26824, HW26826, HW26827, HW26828, HW26829, HW2683, HW26830, HW26831, HW26832, HW26833, HW26834, HW26837, HW26840, HW26842, HW26845, HW26848, HW26851, HW26852, HW26853, HW26854, HW26855, HW26862, HW26863, HW26864, HW26865, HW26866, HW26867, HW26868, HW26869, HW26870, HW26872, HW26873, HW26874, HW26875, HW26876, HW26877, HW26878, HW26879, HW26880, HW26881, HW26882, HW26883, HW26885, HW26886, HW26887, HW26889, HW26890, HW26891, HW26892, HW26893, HW26894, HW26896, HW26897, HW26898, HW26899, HW2690, HW26902, HW26903, HW26904, HW26905, HW26907, HW26909, HW26910, HW26912, HW26913, HW26914, HW26915, HW26916, HW26917, HW26918, HW26919, HW26920, HW26921, HW26922, HW26923, HW26924, HW26927, HW26928, HW26929, HW26930, HW26931, HW26932, HW26934, HW26937, HW26942, HW26943, HW26944, HW26945, HW26946, HW26947, HW26948, HW26949, HW2695, HW26950, HW26951, HW26954, HW26955, HW26956, HW26957, HW26958, HW26959, HW26960, HW26961, HW26962, HW26963, HW26965, HW26966, HW26967, HW2697, HW26970, HW26972, HW26973, HW26974, HW26975, HW26978, HW2698, HW26986, HW26987, HW26988, HW26989, HW26990, HW26991, HW26992, HW26993, HW26994, HW26995, HW26996, HW26997, HW27004, HW27006, HW27007, HW27008, HW27009, HW27010, HW27011, HW27013, HW27014, HW27015, HW27016, HW27017, HW27018, HW27019, HW2702, HW27020, HW27022, HW27023, HW2704, HW2705, HW27056, HW27057, HW27058, HW2706, HW2707, HW27081, HW27082, HW27083, HW27084, HW27085, HW27086, HW27087, HW27089, HW27090, HW27091, HW27092, HW27094, HW27096, HW27097, HW27098, HW2710, HW27100, HW27102, HW27103, HW27119, HW2714, HW27141, HW27149, HW27152, HW27153, HW27155, HW27156, HW27158, HW27160, HW2718, HW27209, HW27218, HW27221, HW27225, HW27228, HW27229, HW2723, HW27237, HW27239, HW27240, HW27241, HW27242, HW27243, HW27244, HW27245, HW27248, HW27249, HW27250, HW27251, HW27252, HW27253, HW27254, HW27255, HW27256, HW27257, HW27258, HW27259, HW2726, HW27260, HW27261, HW27262, HW27263, HW27264, HW27265, HW27266, HW27267, HW27269, HW2727, HW27270, HW27271, HW27273, HW27274, HW27275, HW27276, HW27277, HW27278, HW27279, HW27280, HW27281, HW27282, HW27283, HW27284, HW27285, HW27286, HW27288, HW27289, HW27290, HW27291, HW27292, HW27294, HW27295, HW27296, HW27297, HW27298, HW27299, HW27302, HW27303, HW27304, HW27305, HW27306, HW27308, HW27309, HW27310, HW27312, HW27313, HW27314, HW27315, HW27316, HW27317, HW27319, HW27321, HW27322, HW27323, HW27324, HW27325, HW27326, HW27327, HW27328, HW27330, HW27331, HW27332, HW27333, HW27334, HW27335, HW27336, HW27338, HW27339, HW27341, HW27342, HW27343, HW27344, HW27345, HW27346, HW27347, HW27348, HW27352, HW27353, HW27358, HW27363, HW27365, HW27366, HW27368, HW27369, HW27370, HW27371, HW27373, HW27374, HW27376, HW27377, HW27378, HW27379, HW27380, HW27381, HW27382, HW27385, HW27386, HW27394, HW2740, HW27405, HW2741, HW2742, HW27420, HW27421, HW27422, HW27423, HW27424, HW27425, HW27426, HW27427, HW27428, HW27429, HW27430, HW27431, HW27432, HW27433, HW27435, HW27436, HW27437, HW27439, HW27440, HW27441, HW27447, HW27450, HW27455, HW27456, HW27457, HW27459, HW2746, HW27460, HW27461, HW27462, HW27464, HW27465, HW27466, HW27467, HW27469, HW27470, HW27471, HW27473, HW27474, HW27475, HW27477, HW27478, HW27479, HW27480, HW27482, HW27484, HW27486, HW27488, HW27490, HW27491, HW27493, HW27494, HW27495, HW27496, HW27497, HW27498, HW27500, HW27501, HW27502, HW27503, HW27504, HW27505, HW27506, HW27507, HW27508, HW27510, HW27511, HW27512, HW27513, HW27514, HW27515, HW27516, HW27517, HW27518, HW27519, HW27520, HW27521, HW27522, HW27523, HW27524, HW27525, HW27526, HW27539, HW27540, HW27541, HW27543, HW27544, HW27545, HW27546, HW27547, HW27549, HW27550, HW27551, HW27552, HW27554, HW27555, HW27556, HW27557, HW27558, HW27559, HW27561, HW27562, HW27565, HW27566, HW27567, HW27569, HW27572, HW27574, HW27585, HW27597, HW27603, HW27606, HW27607, HW27608, HW27609, HW27613, HW27614, HW27615, HW27616, HW27617, HW27619, HW2762, HW27620, HW27621, HW27622, HW27623, HW27624, HW27625, HW27626, HW27627, HW27629, HW27630, HW27633, HW27634, HW27635, HW27637, HW27638, HW27640, HW27641, HW27642, HW27643, HW27645, HW27647, HW27648, HW27649, HW27653, HW27654, HW27655, HW27656, HW27657, HW27658, HW27659, HW27689, HW2769, HW27691, HW27702, HW27703, HW27704, HW27706, HW27709, HW2771, HW27710, HW27712, HW27713, HW27716, HW27717, HW27719, HW27720, HW27721, HW27722, HW27723, HW27724, HW27725, HW27726, HW27727, HW27749, HW2775, HW27754, HW27755, HW27756, HW27757, HW27758, HW27760, HW27762, HW27764, HW27765, HW27768, HW27769, HW27770, HW27772, HW27774, HW27793, HW27794, HW27795, HW27796, HW27797, HW27798, HW27801, HW27802, HW27803, HW27804, HW27810, HW27812, HW27813, HW27815, HW27816, HW27817, HW27818, HW27819, HW27820, HW27821, HW27822, HW27824, HW27825, HW27827, HW27828, HW27829, HW27830, HW27831, HW27832, HW27833, HW27834, HW27835, HW27836, HW27838, HW27839, HW27842, HW27843, HW27844, HW27845, HW27855, HW27870, HW27889, HW27891, HW27892, HW27893, HW27894, HW27895, HW27896, HW27901, HW27905, HW27906, HW27907, HW27911, HW27912, HW27914, HW27917, HW27918, HW27919, HW2792, HW27920, HW27921, HW27922, HW27924, HW27925, HW27927, HW27928, HW27929, HW27930, HW27931, HW27932, HW27933, HW27934, HW27936, HW27938, HW27939, HW27940, HW27941, HW27942, HW27943, HW27944, HW27945, HW27946, HW27947, HW27948, HW27949, HW27950, HW27951, HW27952, HW27953, HW27954, HW27968, HW2797, HW2798, HW28005, HW28006, HW28011, HW28012, HW28016, HW28017, HW28018, HW28019, HW28020, HW28027, HW2804, HW28040, HW2805, HW28074, HW28084, HW28085, HW28086, HW28087, HW28088, HW28089, HW28090, HW28091, HW28093, HW28094, HW28095, HW28096, HW28098, HW28099, HW28100, HW28101, HW28102, HW28103, HW28104, HW28105, HW28107, HW28108, HW28110, HW28111, HW28112, HW28113, HW28116, HW28118, HW28119, HW2812, HW28121, HW28122, HW28123, HW28124, HW28126, HW28128, HW28129, HW28130, HW28131, HW28132, HW28133, HW28134, HW28135, HW28136, HW28138, HW28140, HW28141, HW28142, HW28143, HW2815, HW28171, HW28172, HW28174, HW28178, HW28179, HW28182, HW28183, HW28184, HW28185, HW28186, HW28187, HW28188, HW28189, HW2819, HW28190, HW28191, HW28192, HW28193, HW28213, HW28219, HW28225, HW28226, HW28233, HW28235, HW28240, HW28241, HW28242, HW28243, HW28244, HW28245, HW28246, HW28249, HW28250, HW28251, HW28252, HW28253, HW28254, HW28255, HW28256, HW28257, HW28259, HW28260, HW28261, HW28262, HW28263, HW28264, HW28265, HW28267, HW28268, HW28269, HW2827, HW28270, HW28271, HW28273, HW28274, HW28275, HW28276, HW28278, HW28281, HW28282, HW28287, HW28288, HW28289, HW28293, HW28294, HW28295, HW28297, HW28298, HW28300, HW28301, HW28302, HW28304, HW28305, HW28306, HW28308, HW28318, HW28320, HW28321, HW28322, HW28323, HW28324, HW28325, HW28326, HW28327, HW28329, HW28330, HW28332, HW28333, HW28334, HW28335, HW28357, HW28358, HW28359, HW28360, HW28361, HW28363, HW28365, HW28366, HW28367, HW28368, HW2837, HW2838, HW28381, HW28388, HW28389, HW2839, HW28391, HW28392, HW28393, HW28394, HW28395, HW28396, HW28397, HW28398, HW28399, HW2840, HW28400, HW28401, HW28402, HW28403, HW28404, HW28405, HW28406, HW28408, HW28409, HW28421, HW28424, HW28425, HW28426, HW28427, HW28428, HW28429, HW28430, HW28431, HW28432, HW28433, HW28434, HW28435, HW28436, HW28437, HW28438, HW28440, HW28441, HW28442, HW28443, HW28448, HW28449, HW28450, HW28451, HW28452, HW28453, HW28454, HW28455, HW28456, HW28457, HW28459, HW28460, HW28461, HW28462, HW28463, HW28464, HW28465, HW28466, HW28467, HW28470, HW28476, HW28486, HW28487, HW28488, HW28492, HW28493, HW28495, HW28496, HW28497, HW28498, HW28499, HW28502, HW28503, HW28504, HW28505, HW28507, HW28508, HW28509, HW28510, HW28511, HW28512, HW28513, HW28514, HW28515, HW28516, HW28517, HW28518, HW28519, HW28520, HW28521, HW28522, HW28523, HW28524, HW28525, HW28526, HW28527, HW28528, HW28530, HW28532, HW28533, HW28534, HW28535, HW28565, HW28591, HW28592, HW28593, HW28594, HW28595, HW28596, HW28597, HW28599, HW28600, HW28601, HW28603, HW28604, HW28605, HW28606, HW28607, HW28608, HW28609, HW28610, HW28612, HW28613, HW28614, HW28615, HW28641, HW28642, HW28643, HW28644, HW28645, HW28646, HW28647, HW28648, HW28649, HW2865, HW28650, HW28651, HW28652, HW28653, HW28654, HW28655, HW28656, HW28657, HW28661, HW28662, HW28663, HW28664, HW28682, HW28684, HW28686, HW28687, HW28688, HW28689, HW2869, HW28690, HW28692, HW28693, HW28694, HW28695, HW28696, HW28698, HW28699, HW2870, HW28700, HW28702, HW28703, HW28704, HW28706, HW28707, HW28708, HW28709, HW2871, HW28710, HW28711, HW28712, HW28713, HW28714, HW28715, HW28716, HW28717, HW28718, HW28719, HW28720, HW28721, HW28722, HW28724, HW28725, HW28727, HW28728, HW28729, HW2873, HW28730, HW28731, HW28732, HW28733, HW28734, HW28735, HW28736, HW28738, HW28739, HW28742, HW28743, HW28746, HW28749, HW28758, HW28760, HW28762, HW28763, HW28765, HW28768, HW28769, HW28771, HW28772, HW28773, HW28775, HW28778, HW28779, HW28781, HW28784, HW28785, HW28786, HW28787, HW28788, HW28789, HW28790, HW28791, HW28793, HW28794, HW28795, HW28797, HW28798, HW28799, HW28800, HW28801, HW28802, HW28803, HW28804, HW28806, HW28807, HW28810, HW28811, HW28812, HW28813, HW28814, HW28815, HW28816, HW28818, HW28819, HW28820, HW28822, HW28832, HW28836, HW28840, HW28841, HW28843, HW28857, HW28859, HW28869, HW28870, HW28871, HW28872, HW28873, HW28874, HW28875, HW28876, HW28877, HW28878, HW28879, HW28880, HW28881, HW28882, HW28883, HW28884, HW28889, HW28893, HW28898, HW28899, HW28900, HW28901, HW28902, HW28903, HW28904, HW28905, HW28906, HW28907, HW28908, HW28909, HW28910, HW28911, HW28912, HW28913, HW28914, HW28915, HW28917, HW28918, HW28919, HW28920, HW28929, HW28936, HW28937, HW28938, HW28939, HW28940, HW28941, HW28942, HW28943, HW28944, HW28945, HW28946, HW28948, HW28949, HW28950, HW28952, HW28953, HW28954, HW28956, HW28957, HW28961, HW28965, HW28966, HW28969, HW28970, HW28972, HW28974, HW28975, HW28976, HW28979, HW28980, HW28983, HW28984, HW28985, HW28986, HW28988, HW28989, HW28990, HW28991, HW28992, HW28993, HW28994, HW28995, HW28996, HW28997, HW28998, HW28999, HW29003, HW29004, HW29005, HW29010, HW29012, HW29013, HW29026, HW29030, HW29031, HW29032, HW29033, HW29034, HW29035, HW29036, HW29037, HW29038, HW29039, HW29040, HW29041, HW29042, HW29043, HW29044, HW29045, HW29046, HW29047, HW29048, HW29049, HW29050, HW29051, HW29052, HW29053, HW29054, HW29055, HW29056, HW29057, HW29058, HW29059, HW29060, HW29061, HW29062, HW29063, HW29064, HW29065, HW29066, HW29077, HW29079, HW29081, HW29082, HW29083, HW29085, HW29086, HW29087, HW29088, HW29089, HW29090, HW29091, HW29096, HW29103, HW29104, HW29105, HW29106, HW29108, HW29110, HW29111, HW29112, HW29113, HW29114, HW29115, HW29116, HW29117, HW29118, HW29119, HW29120, HW29121, HW29122, HW29123, HW29124, HW29125, HW29126, HW29127, HW29128, HW29137, HW29138, HW29139, HW29140, HW29141, HW29142, HW29143, HW29144, HW29145, HW29146, HW29147, HW29148, HW29149, HW29150, HW29151, HW29152, HW29153, HW29155, HW29156, HW29157, HW29158, HW29159, HW29160, HW29161, hw29162, HW29163, HW29164, HW29165, HW29166, HW29168, HW29169, HW29171, HW29172, HW29173, HW29174, HW29175, HW29177, HW29179, HW29180, HW29181, HW29182, HW29183, HW29184, HW29185, HW29186, HW29187, HW29188, HW29189, HW29190, HW29191, HW29193, HW29194, HW29195, HW29196, HW29197, HW29198, HW29200, HW29201, HW29202, HW29204, HW29205, HW29206, HW29210, HW29215, HW29221, HW29230, HW29253, HW29255, HW29257, HW29258, HW29259, HW29260, HW29261, HW29262, HW29263, HW29267, HW29269, HW29270, HW29272, HW29277, HW29278, HW29280, HW29281, HW29289, HW29291, HW29294, HW29295, HW29296, HW29300, HW29301, HW29304, HW29311, HW29315, HW29316, HW29317, HW29319, HW29322, HW29323, HW29324, HW29326, HW29327, HW29328, HW29330, HW29331, HW29332, HW29334, HW29335, HW29336, HW29337, HW29338, HW29339, HW29340, HW29341, HW29342, HW29343, HW29345, HW29346, HW29347, HW29348, HW29349, HW29350, HW29358, HW29364, HW29372, HW29373, HW29374, HW29375, HW29376, HW29377, HW29378, HW29379, HW29380, HW29381, HW29382, HW29383, HW29385, HW29386, HW29389, HW29390, HW29391, HW29392, HW29393, HW29394, HW29395, HW29397, HW29398, HW29399, HW29400, HW29401, HW29402, HW29403, HW29404, HW29405, HW29407, HW29408, HW29409, HW29411, HW29412, HW29413, HW29414, HW29416, HW29417, HW29418, HW29419, HW29420, HW29422, HW29423, HW29431, HW29434, HW29435, HW29436, HW29437, HW29443, HW29445, HW29448, HW29449, HW29450, HW29452, HW29453, HW29455, HW29456, HW29457, HW29458, HW29459, HW29460, HW29461, HW29466, HW29467, HW29468, HW29469, HW29479, HW29481, HW29484, HW29488, HW29489, HW29490, HW29491, HW29492, HW29493, HW29494, HW29495, HW29496, HW29497, HW29498, HW29499, HW29500, HW29501, HW29504, HW29506, HW29507, HW29508, HW29509, HW29510, HW29511, HW29513, HW29514, HW29527, HW29528, HW29529, HW29530, HW29531, HW29544, HW29545, HW29553, HW29554, HW29555, HW29556, HW29557, HW29558, HW29559, HW2956, HW29560, HW29565, HW2957, HW29577, HW29584, HW2959, HW29596, HW29598, HW29599, HW2960, HW29605, HW29606, HW29607, HW29608, HW29609, HW29610, hw29611, hw29613, HW29616, HW29619, HW29621, HW29622, HW29623, HW29624, HW29625, HW29626, HW29627, HW29628, HW2963, HW29630, HW29631, HW29632, HW29634, HW29636, HW29637, HW29638, HW29639, HW2964, HW29640, HW29641, HW29642, hw29643, HW29644, HW29646, hw29647, HW29648, HW29649, HW2965, HW29651, HW29652, HW29654, HW29655, HW29656, HW29657, HW2966, HW29662, HW29663, HW2967, HW29678, HW2968, HW29681, HW2969, HW29695, HW29703, HW29704, hw29706, HW29707, hw29709, HW29710, HW29711, HW29712, HW29715, HW29716, HW29718, HW29719, HW29720, HW29721, HW29722, HW29723, HW29724, HW29725, HW29726, HW29728, HW29730, HW29732, HW29733, HW29735, HW29736, HW29739, HW29740, HW29742, HW29744, HW29745, HW29751, HW29753, HW29756, HW29772, HW29773, HW29775, HW29777, HW29778, hw29785, HW29792, HW29799, HW29801, HW29803, HW29804, HW29805, HW29806, HW29807, HW29808, HW29810, HW29812, HW29813, HW29814, HW29816, HW29817, HW29818, HW29819, HW29821, HW29822, HW29823, HW29826, hw29828, HW29829, HW29835, HW29838, HW29848, HW29849, hw29851, hw29852, HW29853, HW29854, HW29855, HW29856, HW29857, HW29858, HW29860, HW29861, HW29862, HW29863, HW29865, HW29868, HW2987, HW29871, HW29872, HW29878, HW29879, HW29881, HW29882, HW29883, HW29884, HW29885, HW29886, HW29888, HW29889, HW2989, HW29890, HW29891, HW29892, HW29897, HW29898, HW29899, HW2990, HW29902, HW29905, HW29906, HW29907, HW29908, HW29910, HW29912, HW29913, HW29914, HW29919, HW2992, HW29920, HW29924, HW29925, HW29929, HW29930, HW29934, HW29935, HW29937, HW29938, HW29939, HW29941, HW29942, HW29943, HW29944, HW29945, HW29948, HW2995, HW29951, HW29952, HW29953, HW29954, HW29955, HW29956, HW29957, HW29959, HW29960, HW29962, HW29966, HW29968, HW29969, HW2997, HW29971, HW29973, HW29977, HW29979, HW2998, HW29980, HW29981, HW29982, HW29984, HW29986, HW29987, HW2999, HW29998, HW3000, HW30000, HW30001, HW30002, HW30004, HW30005, HW30006, HW30007, HW30008, HW30009, HW3001, HW30010, HW30012, HW30013, HW30014, HW30016, HW30017, HW30018, HW30019, HW3002, HW30020, HW30021, HW30023, HW30025, HW30026, HW30027, HW30028, HW30029, HW3003, HW30030, HW30031, HW30032, HW30034, HW30037, HW3004, HW30048, HW30049, HW3005, HW30050, HW30051, HW30052, HW30053, HW30054, HW30055, HW30056, HW30057, HW30058, HW30059, HW30060, HW30061, HW30062, HW30063, HW30064, HW30065, HW30066, HW30067, HW30068, HW30069, HW3007, HW30070, HW30071, HW30072, HW30073, HW30075, HW30076, HW30077, HW30078, HW3008, HW3009, HW30097, HW3010, HW30119, HW3012, HW30120, HW30122, HW30123, HW30124, HW30125, HW30126, HW30127, HW30128, HW30129, HW3013, HW30130, HW30131, HW30132, HW30133, HW30134, HW3014, HW30159, HW3016, HW30160, HW30161, HW30162, HW30163, HW30164, HW30165, HW30166, HW30168, HW30169, HW3017, HW30170, HW30171, HW30172, HW30173, HW30174, HW30176, HW30177, HW30179, HW3018, HW30180, HW30181, HW30182, HW30183, HW30184, HW30185, HW30186, HW30187, HW30188, HW30189, HW3019, HW30190, HW30191, HW30192, HW30203, HW30206, HW3021, HW30215, HW30216, HW30217, HW30218, HW30219, HW3022, HW30220, HW30221, HW30228, HW3023, HW30230, HW30232, HW30233, HW30234, HW30235, HW30236, HW30237, HW30238, HW30240, HW30241, HW30242, HW30243, HW30244, HW30245, HW30246, HW30247, HW30248, HW30249, HW3025, HW30250, HW30256, HW30257, HW30258, HW30259, HW30260, HW30261, HW30262, HW30263, HW30264, HW30265, HW30266, HW30268, HW30269, HW3027, HW30270, HW30271, HW30272, HW30273, HW30274, HW30276, HW30277, HW30278, HW30279, HW30280, HW30281, HW30282, HW30283, HW30284, HW30285, HW30286, HW30287, HW30288, HW30289, HW30290, HW30291, HW30292, HW30293, HW30294, HW30295, HW30296, HW30298, HW30299, HW30300, HW30301, HW30302, HW30303, HW30304, HW30305, HW30306, HW30307, HW30308, HW30309, HW30310, HW30311, HW30312, HW30313, HW30314, HW30315, HW30316, HW30317, HW30318, HW30321, HW30323, HW30325, HW30326, HW30327, HW30328, HW30329, HW30330, HW30331, HW30332, HW30333, HW30334, HW30335, HW30336, HW30337, HW30338, HW30339, HW30340, HW30341, HW30343, HW30344, HW30346, HW30347, HW30348, HW30349, HW3035, HW30350, HW30351, HW30356, HW30357, HW3036, HW3037, HW30377, HW30379, HW30388, HW3040, HW30403, HW30416, HW30417, HW30418, HW30419, HW30420, HW30422, HW30423, HW30424, HW30426, HW30427, HW30428, HW30429, HW30431, HW30432, HW30433, HW30434, HW30435, HW30436, HW30437, HW30438, HW30439, HW30440, HW30441, HW30442, HW30443, HW30444, HW30446, HW30447, HW30448, HW30450, HW30451, HW30452, HW30454, HW30456, HW30457, HW30458, HW30459, HW30460, HW30464, HW30465, HW30466, HW30470, HW30474, HW30476, HW30478, HW30479, HW30480, HW30481, HW30482, HW30487, HW30488, HW30489, HW30490, HW30492, HW30493, HW30495, HW30496, HW30507, HW30519, HW30527, HW30530, HW30531, HW30532, HW30533, HW30534, HW30535, HW30536, HW30537, HW30538, HW30539, HW30540, HW30541, HW30542, HW30543, HW30544, HW30545, HW30546, HW30547, HW30548, HW30550, HW30551, HW30552, HW30553, HW30554, HW30555, HW30556, HW30557, HW30558, HW30559, HW30560, HW30561, HW30562, HW30563, HW30564, HW30565, HW30566, HW30567, HW30568, HW30569, HW30570, HW30571, HW30572, HW30573, HW30574, HW30575, HW30576, HW30577, HW30578, HW30579, HW30580, HW30581, HW30582, HW30583, HW30584, HW30585, HW30586, HW30587, HW30588, HW30592, HW30609, HW30610, HW30611, HW30612, HW30615, HW30616, HW30617, HW30618, HW30620, HW30621, HW30622, HW30623, HW30624, HW30625, HW30626, HW30627, HW30629, HW30630, HW30631, HW30636, HW30637, HW30640, HW30641, HW30642, HW30644, HW30647, HW30648, HW30649, HW30653, HW30657, HW30659, HW30665, HW30668, HW30669, HW30670, HW30671, HW30672, HW30673, HW30674, HW30675, HW30676, HW30678, HW30679, HW30680, HW30681, HW30682, HW30683, HW30684, HW30685, HW30686, HW30687, HW30688, HW30689, HW30690, HW30691, HW30692, HW30693, HW30694, HW30695, HW30696, HW30697, HW30698, HW30699, HW30700, HW30701, HW30702, HW30704, HW30707, HW30708, HW30709, HW30710, HW30712, HW30713, HW30714, HW30715, HW30716, HW30717, HW30718, HW30719, HW3072, HW30720, HW30721, HW30722, HW30723, HW30724, HW30725, HW30727, HW30729, HW30730, HW30731, HW30732, HW3074, HW30742, HW30744, HW30745, HW30746, HW30747, HW30748, HW30749, HW3075, HW30750, HW30751, HW30752, HW30753, HW30754, HW30755, HW30756, HW30757, HW30758, HW30759, HW3076, HW30760, HW30761, HW30762, HW30772, HW30773, HW30774, HW30775, HW30777, HW30778, HW30779, HW3078, HW30780, HW30781, HW30782, HW30784, HW30785, HW30786, HW30787, HW30788, HW30789, HW30790, HW30792, HW3080, HW3081, HW3082, HW3083, HW3085, HW3086, HW3087, HW3088, HW30882, HW30883, HW30884, HW30885, HW30886, HW30888, HW30889, HW30890, HW30891, HW30892, HW30893, HW30894, HW30895, HW30896, HW30897, HW30898, HW30899, HW30900, HW30901, HW30902, HW30903, HW30904, HW30905, HW30906, HW30907, HW30908, HW30909, HW30910, HW30911, HW30912, HW30913, HW30914, HW30915, HW30917, HW30918, HW30920, HW30921, HW30922, HW30923, HW30924, HW30925, hw30926, hw30927, hw30928, hw30929, HW3093, hw30930, hw30931, hw30932, hw30933, hw30935, hw30936, hw30937, hw30938, hw30940, hw30941, hw30942, hw30943, hw30944, hw30945, hw30946, hw30947, HW3096, HW30965, HW30969, HW30975, HW30977, hw30987, hw30989, hw30990, hw30991, hw30992, HW30993, hw30994, hw30995, HW30997, hw30998, hw30999, HW3100, hw31000, HW31001, HW31002, hw31005, HW31006, HW31007, hw31008, HW31010, hw31011, hw31012, HW31017, hw31019, hw31022, hw31023, hw31024, hw31025, hw31027, hw31028, HW31029, HW3103, HW31030, hw31031, hw31033, HW31034, HW31039, HW3104, hw31040, HW31041, HW31042, HW31043, hw31044, HW31045, HW31047, HW31049, HW31050, HW31051, hw31052, HW31054, HW31055, hw31056, hw31057, hw31059, HW31061, HW31062, HW31064, HW31065, HW31066, HW31067, HW31068, HW31069, HW3107, hw31073, hw31078, HW31097, hw31102, hw31109, HW31116, HW31118, HW31122, HW31126, HW31127, HW31131, HW31135, HW31139, HW31142, HW31148, HW31151, HW31152, HW31153, HW31159, HW31160, HW31172, HW31182, HW31184, HW31185, HW31186, HW31188, HW31190, HW31191, hw31193, hw31195, hw31197, HW31198, HW31199, hw31200, HW31201, HW31203, hw31204, HW31205, HW31206, HW31207, HW31209, HW3121, hw31210, hw31211, HW31212, HW31213, HW31214, HW31218, HW31220, HW31221, hw31222, hw31224, HW31225, HW31227, hw31229, HW31230, HW31231, hw31236, hw31237, HW31238, HW31239, HW31240, HW31241, HW31243, HW31245, HW31256, HW31259, HW3126, HW31269, HW31270, HW31272, hw31274, hw31292, HW31298, HW31300, HW31304, HW31306, HW31308, HW31311, HW31312, HW31315, HW31316, HW31318, HW31320, HW31321, HW31323, HW31324, HW31325, HW31326, HW31327, HW31329, HW31330, HW31331, HW31333, HW31334, HW31335, HW31336, HW31337, HW31339, HW31344, HW31345, HW31367, HW31368, HW31392, HW31395, HW31396, HW31397, HW31400, HW31403, HW31404, HW31405, HW31407, HW31459, HW31460, HW31461, HW31462, HW31463, HW31464, HW31465, HW31466, HW31467, HW31468, HW31469, HW31470, HW31474, HW31475, HW31476, HW31477, HW31478, HW31479, HW31481, HW31482, HW31484, HW31490, HW31501, HW31522, HW31528, HW31529, HW31536, HW31537, HW31540, HW3276, HW3277, HW3278, HW3279, HW3280, HW3281, HW3285, HW3287, HW3288, HW3290, HW3291, HW3293, HW3294, HW3295, HW3296, HW3297, HW3298, HW3299, HW3300, HW3303, HW3304, HW3305, HW3306, HW3309, HW3310, HW3311, HW3313, HW3315, HW3316, HW3318, HW3319, HW3320, HW3321, HW3322, HW3323, HW3326, HW3327, HW3328, HW3329, HW3330, HW3332, HW3334, HW3335, HW3336, HW3337, HW3340, HW3341, HW3342, HW3343, HW3346, HW3349, HW3350, HW3352, HW3353, HW3355, HW3356, HW3361, HW3362, HW3364, HW3365, HW3368, HW3369, HW3370, HW3371, HW3372, HW3373, HW3375, HW3376, HW3377, HW3379, HW3380, HW3381, HW3382, HW3383, HW3385, HW3386, HW3388, HW3392, HW3394, HW3455, HW3456, HW3458, HW3459, HW3460, HW3461, HW3462, HW3463, HW3464, HW3465, HW3466, HW3468, HW3469, HW3471, HW3499, HW3500, HW3501, HW3502, HW3503, HW3504, HW3505, HW3506, HW3508, HW3509, HW3510, HW3511, HW3512, HW3515, HW3516, HW3517, HW3518, HW3519, HW3522, HW3523, HW3524, HW3525, HW3527, HW3529, HW3532, HW3536, HW3537, HW3538, HW3539, HW3540, HW3542, HW3543, HW3545, HW3546, HW3547, HW3550, HW3552, HW3554, HW3556, HW3557, HW3560, HW3561, HW3563, HW3665, HW3667, HW3668, HW3672, HW3674, HW3676, HW3678, HW3679, HW3681, HW3682, HW3683, HW3684, HW3685, HW3687, HW3688, HW3690, HW3692, HW3694, HW3695, HW3696, HW3697, HW3698, HW3699, HW3700, HW3701, HW3702, HW3703, HW3705, HW3706, HW3707, HW3710, HW3712, HW3713, HW3715, HW3716, HW3718, HW3719, HW3722, HW3726, HW3727, HW3728, HW3730, HW3735, HW3736, HW3737, HW3739, HW3741, HW3742, HW3744, HW3748, HW3751, HW3752, HW3758, HW3763, HW3765, HW3767, HW3768, HW3769, HW3770, HW3771, HW3772, HW3773, HW3774, HW3775, HW3776, HW3777, HW3778, HW3779, HW3780, HW3781, HW3782, HW3783, HW3784, HW3785, HW3786, HW3787, HW3788, HW3789, HW3791, HW3793, HW3794, HW3795, HW3796, HW3797, HW3799, HW3801, HW3802, HW3803, HW3804, HW3805, HW3807, HW3809, HW3810, HW3811, HW3812, HW3813, HW3814, HW3815, HW3817, HW3818, HW3821, HW3822, HW3823, HW3825, HW3828, HW3829, HW3831, HW3833, HW3836, HW3841, HW3842, HW3845, HW3846, HW3847, HW3848, HW3849, HW3850, HW3854, HW3855, HW3856, HW3859, HW3860, HW3861, HW3862, HW3863, HW3864, HW3865, HW3866, HW3868, HW3869, HW3870, HW3871, HW3872, HW3873, HW3874, HW3876, HW3877, HW3879, HW3880, HW3881, HW3882, HW3883, HW3884, HW3885, HW3886, HW3888, HW3890, HW3891, HW3892, HW3893, HW3894, HW3895, HW3898, HW3899, HW3900, HW3901, HW3902, HW3903, HW3904, HW3905, HW3906, HW3907, HW3908, HW3909, HW3910, HW3911, HW3912, HW3913, HW3914, HW3915, HW3917, HW3918, HW3919, HW3920, HW3921, HW3922, HW3923, HW3924, HW3925, HW3926, HW3929, HW3930, HW3931, HW3932, HW3934, HW3937, HW3938, HW3939, HW3941, HW3943, HW3944, HW3945, HW3946, HW3947, HW3948, HW3949, HW3950, HW3951, HW3952, HW3953, HW3954, HW3955, HW3956, HW3957, HW3958, HW3959, HW3960, HW3962, HW3963, HW3964, HW3965, HW3966, HW3967, HW3968, HW3969, HW3970, HW3971, HW3972, HW3973, HW3975, HW3980, HW3981, HW3982, HW3983, HW3984, HW3985, HW3986, HW3987, HW3988, HW3989, HW3990, HW3991, HW3992, HW3994, HW3995, HW3996, HW3997, HW3998, HW3999, HW4000, HW4001, HW4003, HW4005, HW4007, HW4008, HW4009, HW4010, HW4011, HW4012, HW4013, HW4014, HW4015, HW4016, HW4017, HW4018, HW4020, HW4021, HW4024, HW4025, HW4028, HW4029, HW4030, HW4031, HW4034, HW4035, HW4036, HW4038, HW4040, HW4041, HW4042, HW4043, HW4044, HW4045, HW4046, HW4047, HW4049, HW4050, HW4053, HW4054, HW4055, HW4056, HW4057, HW4058, HW4059, HW4060, HW4061, HW4063, HW4064, HW4065, HW4066, HW4068, HW4069, HW4070, HW4071, HW4072, HW4073, HW4074, HW4075, HW4076, HW4077, HW4078, HW4079, HW4080, HW4081, HW4082, HW4084, HW4085, HW4086, HW4087, HW4088, HW4093, HW4095, HW4096, HW4097, HW4098, HW4099, HW4100, HW4103, HW4105, HW4107, HW4108, HW4109, HW4110, HW4112, HW4114, HW4115, HW4117, HW4118, HW4121, HW4124, HW4125, HW4126, HW4127, HW4128, HW4129, HW4130, HW4131, HW4133, HW4134, HW4135, HW4136, HW4137, HW4138, HW4139, HW4140, HW4142, HW4143, HW4144, HW4145, HW4149, HW4150, HW4152, HW4155, HW4156, HW4157, HW4158, HW4162, HW4163, HW4164, HW4165, HW4168, HW4169, HW4170, HW4171, HW4172, HW4173, HW4175, HW4176, HW4177, HW4178, HW4180, HW4181, HW4182, HW4185, HW4186, HW4187, HW4190, HW4192, HW4193, HW4194, HW4195, HW4197, HW4198, HW4199, HW4201, HW4202, HW4203, HW4204, HW4207, HW4208, HW4209, HW4210, HW4211, HW4212, HW4214, HW4215, HW4216, HW4217, HW4218, HW4219, HW4220, HW4221, HW4222, HW4223, HW4224, hw4225, HW4226, HW4229, HW4231, HW4232, HW4235, HW4236, HW4237, HW4238, HW4241, HW4243, HW4246, HW4247, HW4249, HW4251, HW4254, HW4255, HW4258, HW4259, HW4260, HW4261, HW4268, HW4272, HW4280, HW4282, HW4285, HW4294, HW4296, HW4298, HW4299, HW4301, HW4307, HW4308, HW4309, HW4310, HW4316, HW4317, HW4319, HW4321, HW4322, HW4325, HW4327, HW4328, HW4329, HW4330, HW4331, HW4332, HW4334, HW4336, HW4337, HW4338, HW4339, HW4340, HW4341, HW4342, HW4343, HW4344, HW4345, HW4346, HW4347, HW4348, HW4349, HW4350, HW4352, HW4353, HW4354, HW4355, HW4356, HW4357, HW4361, HW4362, HW4363, HW4364, HW4365, HW4366, HW4368, HW4369, HW4370, HW4373, HW4374, HW4376, HW4377, HW4378, HW4381, HW4382, HW4383, HW4384, HW4385, HW4387, HW4388, HW4389, HW4390, HW4391, HW4392, HW4393, HW4394, HW4395, HW4398, HW4399, HW4400, HW4401, HW4402, HW4403, HW4404, HW4406, HW4408, HW4409, HW4410, HW4412, HW4413, HW4414, HW4415, HW4416, HW4418, HW4419, HW4421, HW4423, HW4424, HW4425, HW4428, HW4429, HW4430, HW4431, HW4432, HW4433, HW4434, HW4435, HW4436, HW4437, HW4438, HW4439, HW4440, HW4442, HW4444, HW4445, HW4446, HW4447, HW4449, HW4451, HW4453, HW4454, HW4455, HW4456, HW4457, HW4458, HW4459, HW4460, HW4461, HW4462, HW4463, HW4464, HW4467, HW4472, HW4473, HW4474, HW4475, HW4476, HW4477, HW4478, HW4480, HW4481, HW4484, HW4486, HW4488, HW4489, HW4492, HW4493, HW4494, HW24514FSCA; \nGTIN 00888707002639, Serial Numbers:  , HW34164, HW32491, HW34485, HW35471, HW31401, HW35287, HW35450, HW34832, HW35192, HW35415, HW34690, HW35190, HW35183, HW29946, HW35452, HW35425, HW34322, HW34741, HW35473, HW34874, HW35008, HW34970, HW35076, HW34599, HW34714, HW32323, HW35135, HW34865, HW35438, HW34831, HW35263, HW34798, HW35462, HW35249, HW34407, HW34676, HW35098, HW35461, HW34654, HW34735, HW35255, HW35433, HW34736, HW35082, HW35026, HW31281, HW32361, HW34635, HW35264, HW32425, HW31804, HW35299, HW35003, HW35168, HW34952, HW35193, HW35129, HW35235, HW35221, HW34883, HW34653, HW34453, HW32486, HW35182, HW35219, HW34153, HW35243, HW35180, HW34984, HW35236, HW35434, HW35066, HW35089, HW32535, HW34976, HW35111, HW34823, HW34230, HW35077, HW35110, HW35045, HW34937, HW35015, HW34315, HW34987, HW34299, HW34828, HW34814, HW35010, HW34921, HW35011, HW34580, HW34483, HW34787, HW34872, HW32405, HW31174, HW34709, HW34472, HW34683, HW34613, HW34576, HW34333, HW34525, HW34397, HW34795, HW31776, HW30015, HW32409, HW34537, HW31181, HW31871, HW34167, HW32376, HW32345, HW32487, HW32178, HW31862, HW34353, HW34650, HW34699, HW34506, HW34679, HW34298, HW34218, HW34188, HW34486, HW32463, HW34671, HW34657, HW32527, HW34394, HW34496, HW34589, HW34477, HW34572, HW34602, HW34499, HW32451, HW34492, HW34451, HW34323, HW34406, HW32363, HW31783, HW34413, HW32449, HW34080, HW34416, HW32375, HW34441, HW34174, HW31613, HW34258, HW32550, HW34306, HW34064, HW31705, HW31652, HW34304, HW34194, HW34014, HW34264, HW31677, HW34212, HW31829, HW34127, HW32312, HW32267, HW32442, HW29831, HW31785, HW32439, HW34083, HW31918, HW32249, HW32318, HW32188, HW34116, HW31647, HW32424, HW32390, HW32417, HW32381, HW32140, HW32126, HW32334, HW32283, HW32379, HW32382, HW32352, HW31923.HW32313, HW32087, HW31798, HW31720, HW32339, HW32151, HW31914, HW31864, HW32145, HW31608, HW31480, HW32003, HW31870, HW31279, HW32300, HW31765, HW32136, HW31971, HW31764, HW32069, HW31866, HW31898, HW31738, HW31719, HW31617, HW31671, HW31730, HW31741, HW31762, HW31740, HW31302, HW31606, HW31612, HW31535, HW31692, HW31624, HW31643, HW31289, HW31595, HW31363, HW31575, HW31576, HW31177, HW31313, HW31099, HW34158, HW35232, HW34081, HW32089, HW32533, HW34826, HW35437, HW35295, HW34763, HW35058, HW35130, HW34728, HW34729, HW35448, HW35449, HW35115, HW35116, HW35456, HW35442, HW35436, HW35431, HW34505, HW35413, HW35403, HW35020, HW35294, HW35293, HW35280, HW35238, HW35226, HW35195, HW35169, HW35176, HW35166, HW35167, HW35072, HW35051, HW35048, HW35014, HW34967, HW34896, HW34856, HW34860, HW34219, HW32327, HW32351, HW34775, HW34781, HW34770, HW31246, HW34715, HW34155, HW34156, HW34660, HW34662, HW34658, HW34645, HW34118, HW34585, HW34601, HW34566, HW34574, HW34575, HW32083, HW34549, HW34562, HW34143, HW34466, HW34471, HW34473, HW34399, HW34391, HW31689, HW34311, HW34314, HW34282, HW34263, HW34265, HW34245, HW34240, HW34221, HW34216, HW34195, HW34183, HW34171, HW34175, HW34163, HW34166, HW34137, HW34105, HW34119, HW34016, HW32469, HW34067, HW32532, HW32500, HW32370, HW32480, HW32485, HW32456, HW32422, HW32251, HW32271, HW32399, HW32380, HW31648, HW32325, HW32280, HW32146, HW32117, HW32107, HW31880, HW31897, HW32074, HW32063, HW32045, HW32044, HW31951, HW31869, HW31567, HW31609, HW31782, HW31775, HW31695, HW31634, HW31690, HW31716, HW31649, HW31599, HW31189, HW31215, HW31591, HW29766, HW29809, HW29874, HW31498, HW31369, HW31389, HW31361, HW31347, HW31349, HW31278, HW31169, HW35260, HW32346; \nGTIN  00888707003261, Serial Numbers:  HW40857, HW41573, HW42283, HW42324, HW42804, HW42323, HW41797, HW41478, HW41588, HW42279, HW41756, HW42819, HW42121, HW42210, HW41186, HW41688, HW41584, HW42810, HW42389, HW42736, HW41664, HW42768, HW41880, HW41807, HW42277, HW41598, HW41795, HW42125, HW40928, HW41077, HW40977, HW42142, HW41489, HW41823, HW41626, HW41821, HW41820, HW42281, HW41731, HW41227, HW42343, HW42144, HW41844, HW41608, HW41873, HW41605, HW40783, HW41451, HW41657, HW41431, HW42242, HW42134, HW42129, HW41735, HW41719, HW42220, HW41630, HW41867, HW41148, HW40337, HW41712, HW41179, HW40384, HW41590, HW41586, HW41692, HW41836, HW41841, HW41240, HW41708, HW41084, HW41812, HW41631, HW41481, HW41642, HW41424, HW41816, HW41435, HW41456, HW40025, HW41668, HW41419, HW41659, HW41385, HW41382, HW41438, HW41681, HW41585, HW41278, HW41679, HW41537, HW41228, HW40326, HW41703, HW41662, HW41410, HW41334, HW40805, HW41425, HW40998, HW41596, HW41627, HW41421, HW41289, HW41603, HW41508, HW41146, HW40690, HW40443, HW41529, HW41525, HW41432, HW41288, HW41614, HW41551, HW41497, HW41578, HW40778, HW41616, HW41388, HW41100, HW41423, HW41464, HW41441, HW41412, HW41548, HW41250, HW41111, HW41059, HW41064, HW41394, HW40927, HW41531, HW41427, HW40905, HW40874, HW41519, HW41577, HW41470, HW40865, HW41453, HW41321, HW40488, HW41247, HW41463, HW41504, HW41154, HW41001, HW40863, HW41461, HW40317, HW41360, HW41361, HW41503, HW41404, HW41234, HW40862, HW41459, HW40968, HW41104, HW41188, HW40723, HW41124, HW41331, HW41157, HW40962, HW41468, HW41237, HW41400, HW41195, HW40006, HW41225, HW41069, HW41310, HW41248, HW41232, HW41038, HW41158, HW41167, HW40939, HW41275, HW41137, HW40832, HW41229, HW41081, HW40894, HW40583, HW40884, HW40916, HW41345, HW41341, HW41098, HW40813, HW41066, HW40568, HW40200, HW41327, HW40499, HW41208, HW40585, HW40745, HW40700, HW40882, HW41092, HW41194, HW41076, HW40635, HW41060, HW40800, HW40801, HW41072, HW41057, HW40935, HW40870, HW40046, HW40405, HW40683, HW40199, HW40829, HW40460, HW40902, HW40694, HW40746, HW40703, HW40807, HW40941, HW40054, HW41007, HW40991, HW40875, HW40687, HW40769, HW40230, HW40917, HW40616, HW40925, HW40903, HW40606, HW40359, HW40908, HW40861, HW40089, HW40696, HW40753, HW40819, HW40810, HW40767, HW40510, HW40672, HW40891.HW40772, HW40621, HW40627, HW40501, HW40780, HW40466, HW40529, HW40868, HW40739, HW40815, HW40750, HW40732, HW40719, HW40489, HW40755, HW40762, HW40558, HW40481, HW40528, HW40440, HW40631, HW40294, HW40644, HW40433, HW40737, HW40572, HW40705, HW40590, HW40515, HW40192, HW40633, HW40448, HW40049, HW40684, HW40399, HW40061, HW40017, HW40245, HW40281, HW40305, HW40593, HW40592, HW40397, HW40527, HW40445, HW40168, HW40201, HW40491, HW40444, HW40390, HW40240, HW40524, HW40299, HW40402, HW40324, HW40143, HW40381, HW40223, HW40004, HW40259, HW40073, HW40301, HW40329, HW40293, HW40087, HW40209, HW40263, HW40002, HW40112, HW40272, HW40152, HW40189, HW40210, HW40169, HW40109, HW40099, HW40176, HW40009, HW40202, HW40147, HW40121, HW40190, HW40153, HW40177, HW40166, HW40048, HW40120, HW40987, HW41330, HW41276, HW41733, HW41524, HW42727, HW41866, HW41861, HW41787, HW41814, HW41772, HW41749, HW41747, HW41729, HW41707, HW41636, HW41666, HW41589, HW41457, HW40841, HW41182, HW41480, HW41479, HW41165, HW41389, HW41295, HW41291, HW41283, HW41298, HW41300, HW40986, HW41351, HW41338, HW41280, HW40298, HW41185, HW40776, HW41176, HW41180, HW41115, HW41107, HW41073, HW41054, HW41035, HW41022, HW41017, HW41008, HW41012, HW41005, HW40995, HW40636, HW40976, HW40967, HW40948, HW40944, HW40930, HW40922, HW40918, HW40913, HW40883, HW40622, HW40821, HW40831, HW40876, HW40678, HW40834, HW40838, HW40717, HW40759, HW40757, HW40749, HW40744, HW40534, HW40728, HW40718, HW40708, HW40620, HW40605, HW40157, HW40483, HW40682, HW40307, HW40567, HW40569, HW40546, HW40533, HW40512, HW40504, HW40334, HW40292, HW40482, HW40520, HW40530, HW40509, HW40503, HW40490, HW40453, HW40436, HW40395, HW40392, HW40376, HW40373, HW40322, HW40302, HW40303, HW40138, HW40295, HW40288, HW40226, HW40273, HW40051, HW40270, HW40271, HW40247, HW40238, HW40232, HW40206, HW40212, HW40186, HW40178, HW40067, HW40154, HW40149, HW40139, HW40122, HW40068, HW40100, HW40079, HW40085, HW40090, HW40094, HW40097, HW40045, HW40074, HW40075, HW42770",
      "firm_fei_number": "3007042319",
      "recalling_firm": "Heartware, Inc.",
      "address_1": "14400 Nw 60th Ave",
      "city": "Miami Lakes",
      "state": "FL",
      "postal_code": "33014-2807",
      "additional_info_contact": "MCS Customer Service\n877-367-4823",
      "reason_for_recall": "Updates to the HVAD system instructions for use (IFU) and patient manual (PM) will be provided. These updates will clarify (1) the conditions under which an HVAD [Controller Fault] alarm may sound and the recommended troubleshooting actions and (2) instructions relating to the useful life of the HVAD system components",
      "root_cause_description": "Labeling design",
      "action": "Heartware issued an URGENT:  MEDICAL DEVICE CORRECTION notice to its consignees on 10/31/2023 via letter (UPS).  The notice explained the issue and informed them of the updates to the HVAD IFU and PM (patient manual).",
      "product_quantity": "7438 units",
      "distribution_pattern": "Worldwide",
      "openfda": {
        "device_name": "Ventricular (Assist) Bypass",
        "medical_specialty_description": "Unknown",
        "regulation_number": "",
        "device_class": "3",
        "pma_number": [
          "P170038",
          "P160054",
          "P060040",
          "P870072",
          "P980012",
          "P900023",
          "P160035",
          "P100047",
          "P920014"
        ],
        "registration_number": [
          "3005228872",
          "3008792657",
          "2916596",
          "3004582654",
          "1721676",
          "1319639",
          "2134470",
          "3002807090",
          "1927395",
          "2528981",
          "3012931014",
          "3007042319",
          "2029275",
          "1218444",
          "2133810",
          "3019807891",
          "3003306248",
          "2182208",
          "3008454189",
          "3011823740",
          "2031044",
          "9612515",
          "3010163695"
        ],
        "fei_number": [
          "3005228872",
          "1000306431",
          "3008792657",
          "2916596",
          "3004582654",
          "3001236763",
          "1721676",
          "3002462840",
          "1000138054",
          "1927395",
          "3006003617",
          "2528981",
          "3012931014",
          "3007042319",
          "1218444",
          "3019807891",
          "3002808437",
          "3003306248",
          "2182208",
          "3008454189",
          "3011823740",
          "3000203066",
          "3010163695"
        ]
      }
    }
  ]
}