  "results": [
    {
      "cfres_id": "197235",
      "product_res_number": "Z-0459-2023",
      "event_date_initiated": "2022-11-22",
      "event_date_posted": "2022-12-08",
      "recall_status": "Open, Classified",
      "res_event_number": "91195",
      "product_code": "LNH",
      "k_numbers": [
        "K173356",
        "K173451",
        "K183063",
        "K193215",
        "K213516"
      ],
      "product_description": "Philips Ingenia Elition S MR System, Model Numbers 781357, 782106, 782137",
      "code_info": "Model 781357:\nUDI-DI: 00884838088108;\nSerial Numbers: 46002\n46014\n46025\n46065\n46077\n46088\n46097\n46109\n46071;\n\nModel 782106:\nUDI-DI: 00884838098329;\nSerial Numbers: 46224\n46225\n46265\n46269\n46311;\n\nModel Number 782137:\nUDI-DI: 00884838108615;\nSerial Numbers: Included in Letter but No Devices Distributed in the US.",
      "firm_fei_number": "3016618143",
      "recalling_firm": "Philips North America Llc",
      "address_1": "222 Jacobs St",
      "city": "Cambridge",
      "state": "MA",
      "postal_code": "02141-2289",
      "reason_for_recall": "Gradient Coil may act as a heat source with a potential to produce smoke and/or fire.",
      "root_cause_description": "Under Investigation by firm",
      "action": "An URGENT Medical Device Correction Notice dated 11/22/22 was sent to customers.\n\nActions that should be taken by the customer / user in order to prevent risks for patients or users\n\n1. Systems that do not yet have the SmokeDetector Interlock feature installed are identified in Appendix A. If your system is listed in Appendix A immediately discontinue use of your\nimpacted MR system(s) until Philips has installed the SmokeDetector Interlock.\n\" Post a  Do not use  notice on or near the impacted MR system(s)\n\" Post this notice near the affected MR system(s) for ease of reference.\n\" Circulate this notice to all users of this device so that they are aware of the product issue and associated hazard/harm until this issue has been resolved.\n\" Please complete and return the attached customer response form to Philips promptly and no later than 30 days from receipt of this letter.\n\" Note: Once your system has the SmokeDetector Interlock installed, please follow the advice in steps 2 through 5.\n\n2. Philips has installed a SmokeDetector Interlock feature on the systems listed in Appendix B.\nImmediately confirm if your system has the SmokeDetector Interlock installed by confirming your serial number is listed in Appendix B.\n\" Post this notice near the affected MR system(s) for ease of reference.\n\" Circulate this notice to all users of this device so that they are aware of the product issue and associated hazard/harm until this issue has been resolved.\n\" Please complete and return the attached customer response form to Philips promptly and no later than 30 days from receipt of this letter.\n\n3. The SmokeDetector Interlock feature is integrated with the MR System and powers off the relevant part of the system (Gradient Amplifier) when smoke is detected in order to prevent further development of smoke and/or fire. If the SmokeDetector Interlock has detected smoke and has powered off the Gradient Amplifier:\n\" Remove patient from the system according to the instructions for use because",
      "product_quantity": "566 total",
      "distribution_pattern": "US Nationwide.",
      "openfda": {
        "k_number": [
          "K253808",
          "K241331",
          "K915395",
          "K033168",
          "K893881",
          "K230187",
          "K903527",
          "K921046",
          "K943189",
          "K935015",
          "K943440",
          "K230972",
          "K982959",
          "K072237",
          "K934514",
          "K202966",
          "K902310",
          "K915112",
          "K032678",
          "K162183",
          "K934485",
          "K903048",
          "K092469",
          "K910817",
          "K900286",
          "K011485",
          "K230355",
          "K143128",
          "K243667",
          "K040877",
          "K232535",
          "K964753",
          "K183274",
          "K901242",
          "K172685",
          "K010057",
          "K896508",
          "K961969",
          "K902316",
          "K970037",
          "K041476",
          "K052078",
          "K041111",
          "K192851",
          "K171778",
          "K893675",
          "K223439",
          "K250436",
          "K911993",
          "K903770",
          "K923944",
          "K113281",
          "K930709",
          "K143251",
          "K945875",
          "K043147",
          "K242524",
          "K193282",
          "K091536",
          "K934719",
          "K900728",
          "K193116",
          "K932552",
          "K944979",
          "K945517",
          "K043262",
          "K980645",
          "K925858",
          "K923823",
          "K964531",
          "K984224",
          "K132951",
          "K191157",
          "K201039",
          "K923881",
          "K923581",
          "K193097",
          "K915567",
          "K012970",
          "K223442",
          "K953918",
          "K192690",
          "K111242",
          "K202170",
          "K260852",
          "K002997",
          "K974530",
          "K925845",
          "K173749",
          "K211037",
          "K900284",
          "K003623",
          "K052305",
          "K884369",
          "K974498",
          "K933297",
          "K042236",
          "K110802",
          "K093984",
          "K993803",
          "K232322",
          "K894016",
          "K182756",
          "K130692",
          "K122613",
          "K202238",
          "K022390",
          "K221923",
          "K934499",
          "K905834",
          "K940874",
          "K911975",
          "K032232",
          "K202014",
          "K191128",
          "K903768",
          "K884062",
          "K905578",
          "K220332",
          "K080515",
          "K212056",
          "K946244",
          "K142421",
          "K910220",
          "K192506",
          "K981190",
          "K151466",
          "K143253",
          "K233629",
          "K193218",
          "K920441",
          "K063361",
          "K883648",
          "K912539",
          "K934318",
          "K933645",
          "K921941",
          "K231617",
          "K922470",
          "K921720",
          "K903319",
          "K964370",
          "K052164",
          "K031230",
          "K992225",
          "K924653",
          "K091444",
          "K905259",
          "K990335",
          "K932644",
          "K193102",
          "K895328",
          "K252838",
          "K983793",
          "K253779",
          "K120778",
          "K202767",
          "K912963",
          "K932344",
          "K001773",
          "K884816",
          "K965068",
          "K931392",
          "K920215",
          "K183254",
          "K253413",
          "K131241",
          "K053309",
          "K251276",
          "K120701",
          "K231133",
          "K192574",
          "K213805",
          "K970990",
          "K073081",
          "K942560",
          "K992892",
          "K932375",
          "K890666",
          "K221025",
          "K925985",
          "K001987",
          "K200024",
          "K233673",
          "K243397",
          "K041020",
          "K961092",
          "K162940",
          "K901241",
          "K250379",
          "K990826",
          "K082485",
          "K033315",
          "K911800",
          "K002420",
          "K920578",
          "K903753",
          "K885223",
          "K090080",
          "K964306",
          "K954646",
          "K935752",
          "K890394",
          "K994351",
          "K924808",
          "K891541",
          "K142361",
          "K081028",
          "K991943",
          "K935671",
          "K935627",
          "K942609",
          "K924489",
          "K980114",
          "K253499",
          "K130262",
          "K061132",
          "K172110",
          "K900285",
          "K001334",
          "K153740",
          "K010691",
          "K191662",
          "K925987",
          "K963429",
          "K905603",
          "K080762",
          "K260265",
          "K092237",
          "K020268",
          "K890815",
          "K994283",
          "K922006",
          "K935369",
          "K901602",
          "K140666",
          "K190461",
          "K073099",
          "K101295",
          "K082427",
          "K931544",
          "K082964",
          "K962117",
          "K940602",
          "K982761",
          "K890789",
          "K924605",
          "K132831",
          "K911959",
          "K903318",
          "K983327",
          "K992120",
          "K202210",
          "K043290",
          "K893750",
          "K211406",
          "K960973",
          "K883482",
          "K220335",
          "K133848",
          "K950973",
          "K962933",
          "K915873",
          "K900283",
          "K911642",
          "K251029",
          "K942874",
          "K043271",
          "K944662",
          "K032795",
          "K121239",
          "K896789",
          "K935691",
          "K930960",
          "K972704",
          "K905615",
          "K071099",
          "K100352",
          "K052013",
          "K053009",
          "K013858",
          "K153547",
          "K231587",
          "K082331",
          "K033485",
          "K041602",
          "K933296",
          "K072279",
          "K110151",
          "K944070",
          "K163192",
          "K203443",
          "K914209",
          "K181433",
          "K925009",
          "K943320",
          "K884484",
          "K120638",
          "K921098",
          "K940796",
          "K883650",
          "K073425",
          "K901639",
          "K973799",
          "K171128",
          "K971380",
          "K192496",
          "K014011",
          "K915601",
          "K895966",
          "K935664",
          "K251822",
          "K904528",
          "K240944",
          "K941864",
          "K923984",
          "K153003",
          "K971884",
          "K002531",
          "K210611",
          "K220192",
          "K891598",
          "K901620",
          "K905442",
          "K931022",
          "K920159",
          "K884604",
          "K052023",
          "K930811",
          "K092308",
          "K033507",
          "K201616",
          "K944773",
          "K083533",
          "K895629",
          "K911806",
          "K932403",
          "K012875",
          "K162931",
          "K232482",
          "K180100",
          "K232030",
          "K941996",
          "K920126",
          "K923423",
          "K913770",
          "K092925",
          "K133526",
          "K891543",
          "K971279",
          "K071154",
          "K141945",
          "K161397",
          "K926397",
          "K222387",
          "K900727",
          "K920227",
          "K946085",
          "K231560",
          "K932292",
          "K121160",
          "K891769",
          "K903204",
          "K921413",
          "K013150",
          "K973658",
          "K915260",
          "K915598",
          "K223426",
          "K211633",
          "K963990",
          "K122006",
          "K223875",
          "K203372",
          "K930674",
          "K253648",
          "K981475",
          "K932338",
          "K160621",
          "K020991",
          "K943800",
          "K030918",
          "K220939",
          "K173592",
          "K190138",
          "K912410",
          "K905104",
          "K062454",
          "K981398",
          "K103411",
          "K250443",
          "K251808",
          "K030874",
          "K122026",
          "K924479",
          "K954353",
          "K173356",
          "K961174",
          "K181371",
          "K895794",
          "K240608",
          "K892959",
          "K220151",
          "K895221",
          "K935661",
          "K001894",
          "K013344",
          "K963262",
          "K073156",
          "K893640",
          "K183244",
          "K924756",
          "K945778",
          "K963345",
          "K921653",
          "K883622",
          "K943799",
          "K982834",
          "K101347",
          "K924154",
          "K935224",
          "K023511",
          "K883756",
          "K923010",
          "K220425",
          "K933658",
          "K111803",
          "K103330",
          "K173558",
          "K141977",
          "K013636",
          "K123068",
          "K242573",
          "K935084",
          "K221785",
          "K090433",
          "K143008",
          "K003628",
          "K123522",
          "K964608",
          "K941812",
          "K000315",
          "K920097",
          "K944331",
          "K212367",
          "K050602",
          "K032428",
          "K941408",
          "K050132",
          "K133692",
          "K080098",
          "K990260",
          "K261331",
          "K092899",
          "K202030",
          "K151015",
          "K223203",
          "K203323",
          "K942669",
          "K920714",
          "K032541",
          "K050237",
          "K152283",
          "K061429",
          "K931656",
          "K132119",
          "K233687",
          "K031396",
          "K992533",
          "K930779",
          "K251397",
          "K092253",
          "K162722",
          "K041112",
          "K190757",
          "K213516",
          "K961121",
          "K914535",
          "K994117",
          "K163116",
          "K940364",
          "K951649",
          "K212419",
          "K191050",
          "K150427",
          "K212271",
          "K220575",
          "K971668",
          "K974212",
          "K893509",
          "K962138",
          "K900889",
          "K910513",
          "K905833",
          "K031927",
          "K940514",
          "K920094",
          "K011416",
          "K894974",
          "K123938",
          "K173079",
          "K930544",
          "K902933",
          "K151668",
          "K201540",
          "K210843",
          "K000832",
          "K925597",
          "K232346",
          "K123510",
          "K920789",
          "K905835",
          "K964778",
          "K002179",
          "K170412",
          "K030520",
          "K920098",
          "K980831",
          "K181613",
          "K210415",
          "K943021",
          "K913534",
          "K914114",
          "K962927",
          "K022157",
          "K961876",
          "K111340",
          "K962048",
          "K232644",
          "K892235",
          "K213583",
          "K962618",
          "K951650",
          "K133435",
          "K912095",
          "K172920",
          "K913325",
          "K962892",
          "K920543",
          "K012495",
          "K905070",
          "K061916",
          "K974614",
          "K212673",
          "K193200",
          "K190524",
          "K901818",
          "K121434",
          "K960181",
          "K073457",
          "K201389",
          "K251399",
          "K905832",
          "K924349",
          "K203451",
          "K905366",
          "K202518",
          "K913563",
          "K905441",
          "K173409",
          "K012202",
          "K911746",
          "K990153",
          "K163331",
          "K203553",
          "K923870",
          "K945945",
          "K191032",
          "K920791",
          "K080968",
          "K901730",
          "K161567",
          "K904327",
          "K951681",
          "K934987",
          "K962249",
          "K052978",
          "K905512",
          "K024066",
          "K090038",
          "K982932",
          "K173382",
          "K203280",
          "K192426",
          "K920125",
          "K121650",
          "K884607",
          "K883737",
          "K994316",
          "K003947",
          "K001796",
          "K894855",
          "K932374",
          "K163211",
          "K000802",
          "K083421",
          "K930045",
          "K971055",
          "K922798",
          "K252371",
          "K173617",
          "K943321",
          "K222259",
          "K893744",
          "K924426",
          "K163312",
          "K100164",
          "K113394",
          "K110238",
          "K911892",
          "K934538",
          "K931460",
          "K010850",
          "K253862",
          "K241607",
          "K896647",
          "K945827",
          "K141472",
          "K132376",
          "K041577",
          "K991652",
          "K945798",
          "K914536",
          "K901864",
          "K896309",
          "K970852",
          "K002611",
          "K210340",
          "K905095",
          "K213960",
          "K962677",
          "K932091",
          "K935450",
          "K934212",
          "K902663",
          "K915754",
          "K961624",
          "K902217",
          "K920199",
          "K925029",
          "K240571",
          "K260113",
          "K955389",
          "K932271",
          "K180592",
          "K992135",
          "K903203",
          "K915657",
          "K926001",
          "K230152",
          "K945779",
          "K961366",
          "K213603",
          "K073362",
          "K012200",
          "K950837",
          "K926275",
          "K213693",
          "K240238",
          "K883547",
          "K151074",
          "K260746",
          "K121676",
          "K113337",
          "K964752",
          "K103507",
          "K222968",
          "K093336",
          "K122034",
          "K170978",
          "K043030",
          "K944112",
          "K012154",
          "K241925",
          "K990620",
          "K900064",
          "K930991",
          "K183222",
          "K211983",
          "K885047",
          "K042948",
          "K992746",
          "K902044",
          "K080048",
          "K092230",
          "K232981",
          "K930122",
          "K170840",
          "K003573",
          "K232494",
          "K103238",
          "K250246",
          "K893739",
          "K113093",
          "K232680",
          "K990968",
          "K971684",
          "K896336",
          "K913673",
          "K994282",
          "K221733",
          "K922677",
          "K231459",
          "K161973",
          "K942776",
          "K040444",
          "K883621",
          "K983110",
          "K890092",
          "K884743",
          "K011418",
          "K113511",
          "K980417",
          "K894290",
          "K113490",
          "K181370",
          "K161795",
          "K221619",
          "K132160",
          "K093738",
          "K061930",
          "K242825",
          "K153447",
          "K201141",
          "K151579",
          "K884606",
          "K920096",
          "K220589",
          "K901888",
          "K903028",
          "K980306",
          "K052315",
          "K251901",
          "K210164",
          "K003651",
          "K944724",
          "K953262",
          "K925107",
          "K924472",
          "K915552",
          "K223079",
          "K233728",
          "K212233",
          "K193215",
          "K230758",
          "K071925",
          "K931298",
          "K200327",
          "K123417",
          "K961842",
          "K233733",
          "K961238",
          "K024042",
          "K910839",
          "K931008",
          "K932645",
          "K091669",
          "K895131",
          "K242006",
          "K211118",
          "K243122",
          "K901305",
          "K921881",
          "K884603",
          "K211980",
          "K912881",
          "K932244",
          "K080038",
          "K002242",
          "K020164",
          "K173451",
          "K923948",
          "K192852",
          "K072850",
          "K251682",
          "K203279",
          "K052698",
          "K003899",
          "K050200",
          "K063559",
          "K101813",
          "K934396",
          "K060956",
          "K983902",
          "K140253",
          "K984280",
          "K220018",
          "K041286",
          "K911495",
          "K930193",
          "K990329",
          "K252239",
          "K920835",
          "K923615",
          "K191036",
          "K942873",
          "K110942",
          "K020573",
          "K883994",
          "K910269",
          "K911872",
          "K890855",
          "K923025",
          "K102344",
          "K133670",
          "K223343",
          "K171597",
          "K023130",
          "K143345",
          "K142734",
          "K190017",
          "K013586",
          "K253489",
          "K926171",
          "K010570",
          "K933659",
          "K935570",
          "K941622",
          "K161238",
          "K883649",
          "K002415",
          "K143606",
          "K153324",
          "K213717",
          "K091580",
          "K933018",
          "K930708",
          "K893518",
          "K900026",
          "K162403",
          "K981410",
          "K063207",
          "K173224",
          "K900289",
          "K160152",
          "K925286",
          "K143020",
          "K900288",
          "K032082",
          "K111226",
          "K993731",
          "K060003",
          "K883872",
          "K240744",
          "K924921",
          "K003516",
          "K942355",
          "K033867",
          "K914012",
          "K960901",
          "K062860",
          "K180926",
          "K230208",
          "K895151",
          "K031815",
          "K253625",
          "K203053",
          "K212456",
          "K112003",
          "K990550",
          "K952235",
          "K043112",
          "K883516",
          "K242258",
          "K121997",
          "K910498",
          "K223523",
          "K162102",
          "K914263",
          "K203653",
          "K911348",
          "K914019",
          "K923140",
          "K902314",
          "K935670",
          "K172878",
          "K992412",
          "K162943",
          "K933295",
          "K911667",
          "K901952",
          "K903822",
          "K963953",
          "K243335",
          "K072473",
          "K240540",
          "K033504",
          "K022397",
          "K933356",
          "K130885",
          "K142997",
          "K903755",
          "K902315",
          "K132381",
          "K942875",
          "K052423",
          "K913869",
          "K232760",
          "K061368",
          "K193021",
          "K241429",
          "K993682",
          "K170396",
          "K142515",
          "K223458",
          "K131996",
          "K131737",
          "K222755",
          "K023686",
          "K173600",
          "K180925",
          "K090873",
          "K183685",
          "K243547",
          "K943798",
          "K953248",
          "K910514",
          "K183231",
          "K913463",
          "K183657",
          "K083147",
          "K182129",
          "K250236",
          "K972990",
          "K913441",
          "K895869",
          "K891760",
          "K935549",
          "K052293",
          "K921940",
          "K964626",
          "K181593",
          "K002490",
          "K160185",
          "K002225",
          "K183137",
          "K183221",
          "K223247",
          "K965044",
          "K945791",
          "K133490",
          "K962061",
          "K182282",
          "K903661",
          "K922000",
          "K052987",
          "K900287",
          "K920254",
          "K073167",
          "K241242",
          "K942953",
          "K984274",
          "K940772",
          "K954355",
          "K905634",
          "K050199",
          "K142085",
          "K901092",
          "K153343",
          "K982157",
          "K963179",
          "K955811",
          "K031731",
          "K032121",
          "K911991",
          "K893619",
          "K120487",
          "K960294",
          "K883378",
          "K213305",
          "K894714",
          "K900725",
          "K003192",
          "K902630",
          "K902799",
          "K944469",
          "K954039",
          "K911276",
          "K941666",
          "K081916",
          "K023207",
          "K232765",
          "K920204",
          "K923352",
          "K031350",
          "K041695",
          "K900919",
          "K943415",
          "K173507",
          "K061255",
          "K181322",
          "K945397",
          "K945386",
          "K153039",
          "K944228",
          "K923022",
          "K954354",
          "K910715",
          "K981898",
          "K883521",
          "K883520",
          "K002964",
          "K945828",
          "K915579",
          "K010773",
          "K183063",
          "K926416",
          "K952530",
          "K923264",
          "K120748",
          "K192650",
          "K893689",
          "K130115",
          "K032490",
          "K250901",
          "K160632",
          "K003853",
          "K920095",
          "K113145",
          "K130658",
          "K945730",
          "K991243",
          "K921609",
          "K972826",
          "K963186",
          "K211818",
          "K923897",
          "K131896",
          "K072998",
          "K102489",
          "K883519",
          "K903754",
          "K952232",
          "K941582",
          "K970573",
          "K974844",
          "K170090",
          "K182299",
          "K102155",
          "K932693",
          "K221393",
          "K192002",
          "K913475",
          "K121545",
          "K883699",
          "K193176",
          "K240157",
          "K905858",
          "K070968",
          "K122035",
          "K083137",
          "K251386",
          "K082100",
          "K181479",
          "K922846",
          "K241496",
          "K181093",
          "K052579",
          "K213668",
          "K192924",
          "K052571",
          "K960497",
          "K940894",
          "K160618",
          "K940534",
          "K052172",
          "K921891",
          "K910802",
          "K201722",
          "K011296",
          "K251937",
          "K984221",
          "K941025",
          "K183621",
          "K931454",
          "K020289",
          "K241133",
          "K152371",
          "K003453",
          "K140956",
          "K193324",
          "K201615",
          "K011466",
          "K252379",
          "K254277",
          "K993574",
          "K883898",
          "K901980",
          "K901384",
          "K940541",
          "K232361",
          "K253780",
          "K926571",
          "K945155",
          "K896419",
          "K030981",
          "K942604",
          "K994287",
          "K103327",
          "K945665",
          "K933835",
          "K091166"
        ],
        "pma_number": [
          "P860010",
          "P860062",
          "P830081",
          "P840063",
          "P830052",
          "P850043",
          "P830051",
          "P830074",
          "P840007",
          "P830076",
          "P830053"
        ],
        "registration_number": [
          "",
          "3014310698",
          "3026047529",
          "3006073271",
          "8043485",
          "1058584",
          "3029521953",
          "3004994584",
          "9617277",
          "3019595609",
          "3006369484",
          "3008156140",
          "9616047",
          "3012085143",
          "3002808157",
          "3005450718",
          "2134213",
          "3036666444",
          "1826555",
          "3008447096",
          "3009963139",
          "1055916",
          "3007734888",
          "3014635027",
          "3009077524",
          "3011015597",
          "3006262888",
          "1721113",
          "1529041",
          "3005326970",
          "3009882464",
          "2011171",
          "3014111222",
          "2520313",
          "3020162557",
          "3018094310",
          "3011767965",
          "3008355361",
          "3031503890",
          "2183553",
          "3014232555",
          "3015431711",
          "3021593788",
          "3004754211",
          "3004531751",
          "3008986341",
          "3009755250",
          "3007546534",
          "3023385454",
          "1835830",
          "2247992",
          "9617016",
          "3011416394",
          "3018423337",
          "3008632256",
          "3018004899",
          "3015276088",
          "3008284470",
          "9613936",
          "3013875850",
          "3006015847",
          "3042177665",
          "3012127758",
          "3023003446",
          "3009715136",
          "3007350713",
          "3032109181",
          "3017297314",
          "1038814",
          "3030982377",
          "3012354318",
          "1526534",
          "3013679558",
          "3010326005",
          "3010703925",
          "3027339477",
          "3010398867",
          "3014680716",
          "3019844073",
          "3003494057",
          "3017375271",
          "3010949642",
          "3033536331",
          "1063285",
          "3012169560",
          "3000126629",
          "3027373189",
          "3017383732",
          "3012075008",
          "3011106828",
          "3007562060",
          "3005810333",
          "1221108",
          "3009154005",
          "3008496839",
          "3009196578",
          "3006738448",
          "2126677",
          "3005049692",
          "3013695852",
          "3004929307",
          "3004977335",
          "3034286426",
          "9614698",
          "3004859018",
          "3011950073",
          "2029299",
          "3015232217",
          "1054713",
          "3012439308",
          "1650907",
          "3010009632",
          "3033333119",
          "3006239787",
          "3012761544",
          "3005877899",
          "3015329423",
          "3009714754",
          "3005976071",
          "3016591365",
          "2432211",
          "3010701021",
          "3006849754",
          "1723170",
          "3021059265",
          "3035950510",
          "3032109225",
          "3016727312",
          "3003661577",
          "3006621332",
          "3014343100",
          "3014809646",
          "3015425142",
          "1319639",
          "1932738",
          "3027325776",
          "3005517190",
          "3011657336",
          "3012939903",
          "9611343",
          "3009504230",
          "3016851825",
          "3015223550",
          "3026391023",
          "3007545404",
          "3033527281",
          "3003768277",
          "3004526608"
        ],
        "fei_number": [
          "",
          "1000138054",
          "3014310698",
          "1000219637",
          "1000143606",
          "3026047529",
          "3006073271",
          "3029521953",
          "3004994584",
          "3019595609",
          "3006369484",
          "3008156140",
          "3001814065",
          "3012085143",
          "3002808157",
          "3005450718",
          "3001654007",
          "3036666444",
          "1826555",
          "3008447096",
          "3009963139",
          "3007734888",
          "3014635027",
          "3009077524",
          "3011015597",
          "3002806900",
          "3006262888",
          "1721113",
          "3005326970",
          "3009882464",
          "1000571342",
          "3003190524",
          "1000114111",
          "2011171",
          "3014111222",
          "2520313",
          "3020162557",
          "3018094310",
          "3011767965",
          "1000113348",
          "3008355361",
          "3031503890",
          "2183553",
          "3014232555",
          "3015431711",
          "3021593788",
          "3004754211",
          "3004531751",
          "3008986341",
          "1000284313",
          "3009755250",
          "3007546534",
          "1937223",
          "3023385454",
          "3011416394",
          "3018423337",
          "3008632256",
          "3018004899",
          "3015276088",
          "3008284470",
          "3013875850",
          "3006015847",
          "3002681132",
          "3042177665",
          "3012127758",
          "3023003446",
          "3009715136",
          "3007350713",
          "3032109181",
          "3017297314",
          "1038814",
          "3030982377",
          "3012354318",
          "1526534",
          "3013679558",
          "3010326005",
          "3010703925",
          "3027339477",
          "3010398867",
          "3014680716",
          "3019844073",
          "3003494057",
          "3017375271",
          "3010949642",
          "3033536331",
          "3012169560",
          "3000126629",
          "3027373189",
          "3017383732",
          "3012075008",
          "3011106828",
          "3007562060",
          "3005810333",
          "1221108",
          "3009154005",
          "3008496839",
          "3009196578",
          "3006738448",
          "1000517638",
          "2126677",
          "3005049692",
          "3013695852",
          "3004929307",
          "3003377038",
          "3004977335",
          "3004859018",
          "3011950073",
          "3015232217",
          "3012439308",
          "3033333119",
          "3010009632",
          "3006239787",
          "3012761544",
          "3005877899",
          "3015329423",
          "1000208827",
          "3009714754",
          "3005976071",
          "3016591365",
          "2432211",
          "3005214683",
          "3010701021",
          "3006849754",
          "3021059265",
          "3035950510",
          "3032109225",
          "3016727312",
          "3003661577",
          "3006621332",
          "3002807880",
          "3014343100",
          "3014809646",
          "3015425142",
          "3002806926",
          "1000403400",
          "3027325776",
          "3005517190",
          "3011657336",
          "3012939903",
          "3009504230",
          "3016851825",
          "3015223550",
          "1000306474",
          "3026391023",
          "3007545404",
          "3002808382",
          "3033527281",
          "3004526608"
        ],
        "device_name": "System, Nuclear Magnetic Resonance Imaging",
        "medical_specialty_description": "Radiology",
        "regulation_number": "892.1000",
        "device_class": "2"
      }
    }
  ]
}