  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-08-20",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "197211",
      "product_res_number": "Z-0472-2023",
      "event_date_initiated": "2022-11-09",
      "event_date_posted": "2022-12-09",
      "recall_status": "Open, Classified",
      "res_event_number": "91182",
      "product_code": "IYE",
      "k_numbers": [
        "K142434"
      ],
      "product_description": "ARTISTE, ONCOR and  PRIMUS with syngo RT Therapist with software version\n4.3.1 MR3.\n\nMaterial Number: 8162815\nIntended Use:  Family of linear accelerator systems is to deliver X-Ray photon and electron radiation for the therapeutic treatment of cancer.",
      "code_info": "UDI: N/A\nS/N: 10736, 10737, 10822, 10850, 10886,10843, 10865,10025\nSoftware version 4.3.1 MR3",
      "firm_fei_number": "2240869",
      "recalling_firm": "Siemens Medical Solutions USA, Inc",
      "address_1": "40 Liberty Blvd",
      "city": "Malvern",
      "state": "PA",
      "postal_code": "19355-1418",
      "additional_info_contact": "SAME\n610-219-4834",
      "reason_for_recall": "When selecting the site fraction group for treatment, the User may select the wrong site for treatment, and potentially deliver the dose to wrong isocenter",
      "root_cause_description": "Software design",
      "action": "Siemens issued Urgent Medical Device Correction Letter CSAN TH001/22/S to all affected customers on 11/09/2022. Letter states reason for recall, health risk and action to take:\nSiemens is developing the new syngo RT Therapist (RTT) software version RTT 4.3.1 MR4. With the new software version, the User will be able to load only one site at a time and avoid confusion of multiple displayed sites. If the User tries to download multiple sites from ARIA to syngo RT Therapist (RTT), the RTT will reject the complete patient load. If this occurs, User should immediately change their workflow by preparing one scheduler per planned site in ARIA. Availability of the new syngo RT Therapist (RTT) software version RTT 4.3.1 MR4 is currently unknown. Confirm acknowledgement of the information and instructions outlined in this Urgent Safety Advisory Notice by signing this document.\nRequest that you promptly notify and instruct accordingly all the staff at your\norganization who need to be aware of this notice and will comply with the ecommendations therein. Please ensure that this safety advisory is retained in your system Instructions for Use. In addition, if you may have further distributed this product,\nplease identify your customers, and notify them at once of this product recall. If the device has been sold and is therefore no longer in your possession, please forward this notice to the new owner. We would also request you to inform us of the identity of the device's new owner where possible. Please forward this safety information to any other organizations that could be affected by this measure.",
      "product_quantity": "8 units US",
      "distribution_pattern": "US Nationwide distribution in the states of CA, IA, LA, NE, NV, OH, SC.",
      "openfda": {
        "k_number": [
          "K972804",
          "K102671",
          "K930182",
          "K962080",
          "K152321",
          "K130678",
          "K052920",
          "K202747",
          "K020320",
          "K926008",
          "K943464",
          "K090044",
          "K932598",
          "K913612",
          "K223229",
          "K924495",
          "K051932",
          "K960510",
          "K210894",
          "K031634",
          "K223839",
          "K020294",
          "K963451",
          "K092845",
          "K173926",
          "K970586",
          "K951808",
          "K162862",
          "K922276",
          "K961170",
          "K003330",
          "K033343",
          "K122888",
          "K112286",
          "K970137",
          "K941118",
          "K182138",
          "K935411",
          "K140823",
          "K950866",
          "K250392",
          "K220539",
          "K871214",
          "K933614",
          "K181218",
          "K982463",
          "K253962",
          "K973889",
          "K960834",
          "K873398",
          "K100264",
          "K092041",
          "K071531",
          "K991319",
          "K981597",
          "K900434",
          "K032762",
          "K120233",
          "K970291",
          "K926321",
          "K102237",
          "K991133",
          "K142617",
          "K160209",
          "K142431",
          "K942092",
          "K832216",
          "K142294",
          "K200652",
          "K203387",
          "K123371",
          "K152867",
          "K070920",
          "K063839",
          "K925714",
          "K121663",
          "K200940",
          "K904364",
          "K160405",
          "K961484",
          "K153014",
          "K232031",
          "K250429",
          "K243142",
          "K940663",
          "K911750",
          "K162393",
          "K951142",
          "K972275",
          "K082122",
          "K971893",
          "K954564",
          "K172961",
          "K213439",
          "K193243",
          "K981535",
          "K200907",
          "K072485",
          "K150767",
          "K170345",
          "K192102",
          "K781174",
          "K102583",
          "K131466",
          "K902849",
          "K092424",
          "K100890",
          "K161984",
          "K210835",
          "K990383",
          "K223209",
          "K252937",
          "K790388",
          "K121295",
          "K192196",
          "K924632",
          "K864910",
          "K010225",
          "K052682",
          "K072079",
          "K153270",
          "K931927",
          "K942282",
          "K170307",
          "K062611",
          "K945321",
          "K150873",
          "K010975",
          "K190155",
          "K912630",
          "K943224",
          "K132705",
          "K973540",
          "K970236",
          "K954679",
          "K963258",
          "K250099",
          "K962335",
          "K891292",
          "K961400",
          "K990997",
          "K903139",
          "K981346",
          "K030175",
          "K232870",
          "K123230",
          "K162472",
          "K052220",
          "K031764",
          "K100115",
          "K092479",
          "K122137",
          "K073273",
          "K231317",
          "K930246",
          "K233622",
          "K161807",
          "K944315",
          "K060912",
          "K962950",
          "K133227",
          "K060226",
          "K973963",
          "K926449",
          "K082582",
          "K112210",
          "K162629",
          "K213053",
          "K162265",
          "K993960",
          "K010449",
          "K990085",
          "K150848",
          "K982502",
          "K970564",
          "K984241",
          "K151180",
          "K024284",
          "K120067",
          "K990209",
          "K182076",
          "K081760",
          "K230312",
          "K220583",
          "K253012",
          "K993594",
          "K062817",
          "K110840",
          "K071701",
          "K081188",
          "K100691",
          "K170788",
          "K952457",
          "K141572",
          "K981911",
          "K935300",
          "K800759",
          "K032156",
          "K050479",
          "K040192",
          "K112692",
          "K023449",
          "K212583",
          "K920879",
          "K894230",
          "K070305",
          "K211001",
          "K974700",
          "K112460",
          "K222762",
          "K081859",
          "K952108",
          "K182050",
          "K072916",
          "K964840",
          "K162577",
          "K050888",
          "K943714",
          "K032100",
          "K183606",
          "K180021",
          "K090683",
          "K930279",
          "K952729",
          "K092019",
          "K181404",
          "K151426",
          "K962155",
          "K113758",
          "K010062",
          "K071992",
          "K202050",
          "K003636",
          "K000209",
          "K131862",
          "K963860",
          "K101503",
          "K802808",
          "K072046",
          "K132908",
          "K243301",
          "K072374",
          "K010464",
          "K964785",
          "K060895",
          "K181032",
          "K140528",
          "K883101",
          "K954165",
          "K213526",
          "K963624",
          "K060737",
          "K002051",
          "K063274",
          "K121530",
          "K800790",
          "K965010",
          "K882231",
          "K213370",
          "K072454",
          "K202412",
          "K121845",
          "K173838",
          "K883019",
          "K942346",
          "K121657",
          "K170570",
          "K946074",
          "K800792",
          "K040709",
          "K231052",
          "K771809",
          "K122450",
          "K161724",
          "K222977",
          "K252948",
          "K863705",
          "K050701",
          "K050442",
          "K201119",
          "K191158",
          "K211663",
          "K003779",
          "K131795",
          "K232293",
          "K832502",
          "K231273",
          "K141787",
          "K002444",
          "K223234",
          "K123788",
          "K252258",
          "K143560",
          "K041315",
          "K092145",
          "K071360",
          "K212114",
          "K213977",
          "K032725",
          "K061140",
          "K832732",
          "K934523",
          "K071673",
          "K123291",
          "K250911",
          "K021463",
          "K142420",
          "K203722",
          "K240431",
          "K192405",
          "K081218",
          "K790747",
          "K070611",
          "K232113",
          "K001643",
          "K231852",
          "K781872",
          "K880200",
          "K212606",
          "K171352",
          "K261856",
          "K122451",
          "K121328",
          "K183374",
          "K111862",
          "K982513",
          "K131313",
          "K081935",
          "K882326",
          "K991107",
          "K210921",
          "K061713",
          "K103616",
          "K991336",
          "K942083",
          "K060633",
          "K171734",
          "K953508",
          "K250696",
          "K191646",
          "K122616",
          "K071614",
          "K141230",
          "K171733",
          "K894123",
          "K972816",
          "K771260",
          "K013279",
          "K000246",
          "K771902",
          "K980750",
          "K183034",
          "K892860",
          "K991966",
          "K231185",
          "K011763",
          "K014081",
          "K192482",
          "K014022",
          "K981313",
          "K232738",
          "K142268",
          "K182374",
          "K940660",
          "K980629",
          "K980379",
          "K133572",
          "K863358",
          "K090349",
          "K053500",
          "K021124",
          "K202068",
          "K954634",
          "K972617",
          "K170751",
          "K974310",
          "K771358",
          "K874558",
          "K072171",
          "K103606",
          "K030439",
          "K120668",
          "K232923",
          "K131807",
          "K971367",
          "K993359",
          "K013391",
          "K930506",
          "K193431",
          "K901932",
          "K111332",
          "K191267",
          "K791967",
          "K951143",
          "K070094",
          "K980402",
          "K940001",
          "K082773",
          "K152761",
          "K223834",
          "K103394",
          "K851041",
          "K895500",
          "K230576",
          "K191761",
          "K982713",
          "K862645",
          "K213137",
          "K953046",
          "K032248",
          "K042794",
          "K080726",
          "K920041",
          "K023519",
          "K102765",
          "K171837",
          "K800761",
          "K003285",
          "K120335",
          "K190894",
          "K840136",
          "K162355",
          "K883264",
          "K232489",
          "K153026",
          "K982549",
          "K091602",
          "K905007",
          "K960293",
          "K895733",
          "K041519",
          "K100625",
          "K180586",
          "K063270",
          "K895697",
          "K210500",
          "K993959",
          "K961214",
          "K981055",
          "K012227",
          "K193207",
          "K080585",
          "K892425",
          "K103193",
          "K992689",
          "K180106",
          "K252919",
          "K092285",
          "K790688",
          "K020860",
          "K042875",
          "K242957",
          "K254010",
          "K911612",
          "K191944",
          "K052391",
          "K080072",
          "K113276",
          "K933227",
          "K200435",
          "K173251",
          "K183698",
          "K964206",
          "K241937",
          "K050372",
          "K151097",
          "K101131",
          "K142142",
          "K991546",
          "K003778",
          "K133240",
          "K110772",
          "K192377",
          "K142434",
          "K952314",
          "K980904",
          "K800791",
          "K932501",
          "K862065",
          "K162476",
          "K023293",
          "K072504",
          "K112778",
          "K982624",
          "K101992",
          "K150265",
          "K940145",
          "K961208",
          "K253004",
          "K171133",
          "K920455",
          "K990734",
          "K091132",
          "K150706",
          "K211939",
          "K132935",
          "K031975",
          "K252174",
          "K993742",
          "K092871",
          "K972709",
          "K230699",
          "K010916",
          "K935412",
          "K110971",
          "K951717",
          "K032594",
          "K944040",
          "K141855",
          "K974703",
          "K100096",
          "K994355",
          "K060906",
          "K011845",
          "K950061",
          "K181989",
          "K011332",
          "K040585",
          "K042720",
          "K034051",
          "K131704",
          "K820310",
          "K102650",
          "K173848",
          "K004022",
          "K934303",
          "K921740",
          "K172013",
          "K041557",
          "K123808",
          "K132124",
          "K203669",
          "K192968",
          "K902339",
          "K913119",
          "K072506",
          "K192242",
          "K151946",
          "K161146",
          "K121934",
          "K140660",
          "K892845",
          "K960338",
          "K883067",
          "K091645",
          "K934566",
          "K933117",
          "K112613",
          "K150382",
          "K212294",
          "K964167",
          "K874893",
          "K170817",
          "K893978",
          "K981790",
          "K171086",
          "K914698",
          "K201117",
          "K032996",
          "K211760",
          "K100801",
          "K170058",
          "K002608",
          "K112841",
          "K970464",
          "K220338",
          "K990833",
          "K953482",
          "K933558",
          "K973936",
          "K111106",
          "K182687",
          "K132605",
          "K212794",
          "K934657",
          "K820448",
          "K940237",
          "K092939",
          "K161946",
          "K951947",
          "K182189",
          "K101751",
          "K951764",
          "K073018",
          "K981424",
          "K152994",
          "K954705",
          "K042956",
          "K021338",
          "K933493",
          "K121284",
          "K780027",
          "K935928",
          "K191797",
          "K802809",
          "K041937",
          "K022708",
          "K091209",
          "K010063",
          "K151207",
          "K952189",
          "K943406",
          "K172534",
          "K012289",
          "K950075",
          "K921991",
          "K030051",
          "K203172",
          "K040687",
          "K200600",
          "K993425",
          "K162002",
          "K991523",
          "K001052",
          "K203173",
          "K961467",
          "K032171",
          "K143224",
          "K072529",
          "K930513",
          "K011024",
          "K130992",
          "K964517",
          "K092550",
          "K924240",
          "K951581",
          "K180595",
          "K151882",
          "K160057",
          "K161144",
          "K051921",
          "K010174",
          "K833941",
          "K945903",
          "K211590",
          "K212958",
          "K063021",
          "K123812",
          "K771808",
          "K052212",
          "K201276",
          "K190137",
          "K190978",
          "K093527",
          "K082909",
          "K142560",
          "K913174",
          "K252188",
          "K935067",
          "K964485",
          "K252977",
          "K171804",
          "K160627",
          "K926187",
          "K935615",
          "K221448",
          "K102200",
          "K924685",
          "K911973",
          "K033347",
          "K171350",
          "K182079",
          "K043409",
          "K143485",
          "K091999",
          "K933613",
          "K023636",
          "K031631",
          "K852609",
          "K980826",
          "K131965",
          "K894912",
          "K913674",
          "K010472",
          "K072047",
          "K221408",
          "K260792",
          "K962487",
          "K935435",
          "K180697",
          "K935021",
          "K923522",
          "K033287",
          "K090802",
          "K223159",
          "K972451",
          "K837347",
          "K252988",
          "K870525",
          "K781057",
          "K972905",
          "K162468",
          "K062618",
          "K943199",
          "K180027",
          "K052705",
          "K082775",
          "K953570",
          "K102024",
          "K112839",
          "K123137",
          "K894722",
          "K222941",
          "K010065",
          "K081036",
          "K904908",
          "K132064",
          "K962792",
          "K040773",
          "K080742",
          "K212909",
          "K955774",
          "K004023",
          "K120789",
          "K093967",
          "K141800",
          "K192424",
          "K161400",
          "K971382",
          "K894706",
          "K004015",
          "K153011",
          "K951679",
          "K832503",
          "K812032",
          "K100267",
          "K981153",
          "K001833",
          "K231573",
          "K970384",
          "K882729",
          "K092018",
          "K171467",
          "K862339",
          "K954861"
        ],
        "registration_number": [
          "8043933",
          "3004753785",
          "3006073271",
          "3005099803",
          "3030077501",
          "9613183",
          "3007165183",
          "3003418325",
          "3002808157",
          "3038289344",
          "3022973841",
          "3012631246",
          "3009302839",
          "3010123525",
          "1000162206",
          "9681124",
          "3009649442",
          "3006946288",
          "3013012623",
          "3002807314",
          "3009077524",
          "3014113695",
          "1721113",
          "1219183",
          "3012039133",
          "3014404697",
          "3016686005",
          "1720929",
          "3018094310",
          "3004637983",
          "1647149",
          "3011716550",
          "3004962788",
          "9616684",
          "3027080417",
          "2436865",
          "3004363352",
          "3006621300",
          "3007546534",
          "2247992",
          "9617016",
          "3013410552",
          "3015620182",
          "3015496236",
          "3004608878",
          "3018004899",
          "1220477",
          "3026961165",
          "2939248",
          "1450662",
          "3012205931",
          "2182762",
          "3007350713",
          "3010625741",
          "3015119628",
          "2155",
          "1038814",
          "9614952",
          "3004832819",
          "3009140718",
          "1000523114",
          "2134494",
          "3031194398",
          "3003803149",
          "1651102",
          "3003586672",
          "2184007",
          "3042022819",
          "3006579682",
          "3007037519",
          "1722661",
          "3022719374",
          "2431392",
          "3008191247",
          "3009394603",
          "3016602874",
          "1055485",
          "3015232217",
          "3007615597",
          "1526854",
          "3009039068",
          "3009607881",
          "3019552217",
          "3008706707",
          "3015621321",
          "2030598",
          "3001722928",
          "3003873069",
          "3015548663",
          "9681134",
          "3016591365",
          "9710332",
          "3010701021",
          "1530561",
          "3006981798",
          "3007791541",
          "3003793371",
          "3009894282",
          "3006405073",
          "3025765259",
          "3003477135",
          "3006894636",
          "1123169",
          "2183744",
          "3003094912",
          "1932738",
          "3004580128",
          "2916710",
          "9614584",
          "3009754495",
          "3008317830",
          "3010875109",
          "3014428919",
          "3033484057",
          "3003768277"
        ],
        "fei_number": [
          "1000143606",
          "3004753785",
          "3006073271",
          "3005099803",
          "3030077501",
          "3003062786",
          "3003418325",
          "3007165183",
          "3002808157",
          "3038289344",
          "3022973841",
          "3012631246",
          "3009302839",
          "3010123525",
          "1000162206",
          "3009649442",
          "3006946288",
          "3007360740",
          "3013012623",
          "3009077524",
          "3002807314",
          "1000113153",
          "3002806900",
          "3014113695",
          "1721113",
          "1219183",
          "1000160359",
          "3012039133",
          "3014404697",
          "3016686005",
          "1720929",
          "3018094310",
          "3002619595",
          "3004637983",
          "3011716550",
          "3004962788",
          "3027080417",
          "3004363352",
          "3006621300",
          "3007546534",
          "3013410552",
          "3015620182",
          "3015496236",
          "3004608878",
          "3018004899",
          "1220477",
          "3026961165",
          "2939248",
          "1000518731",
          "1450662",
          "3012205931",
          "2182762",
          "3007350713",
          "1000115748",
          "3010625741",
          "3015119628",
          "2155",
          "1038814",
          "3000253133",
          "3004832819",
          "3009140718",
          "3006653576",
          "1000523114",
          "3031194398",
          "3003803149",
          "3003586672",
          "3002913962",
          "3042022819",
          "3006579682",
          "3002802883",
          "3007037519",
          "3022719374",
          "2431392",
          "3008191247",
          "3009394603",
          "3016602874",
          "3015232217",
          "3007615597",
          "1526854",
          "3009039068",
          "3009607881",
          "3019552217",
          "3008706707",
          "3015621321",
          "1000123143",
          "1000208827",
          "3001722928",
          "3003873069",
          "3015548663",
          "3016591365",
          "3010701021",
          "1000521490",
          "3006981798",
          "3007791541",
          "3002619738",
          "3003793371",
          "3009894282",
          "3006405073",
          "3002807880",
          "3001236129",
          "3025765259",
          "3003477135",
          "3006894636",
          "3003251135",
          "2183744",
          "1000306139",
          "1000227792",
          "3003094912",
          "3004580128",
          "2916710",
          "3009754495",
          "3008317830",
          "3010875109",
          "3014428919",
          "3033484057"
        ],
        "device_name": "Accelerator, Linear, Medical",
        "medical_specialty_description": "Radiology",
        "regulation_number": "892.5050",
        "device_class": "2"
      }
    }
  ]
}