  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-08-20",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "197095",
      "product_res_number": "Z-0815-2023",
      "event_date_initiated": "2022-11-09",
      "event_date_posted": "2023-01-04",
      "recall_status": "Open, Classified",
      "res_event_number": "91148",
      "product_code": "PEX",
      "product_description": "VITEK MS CLINICAL REAGENT VITEK MS-CHCA, CATALOG 411071",
      "code_info": "UDI/DI 03573026362836, Batch Numbers: 1009391190, 1009434380, 1009472100",
      "firm_fei_number": "1950204",
      "recalling_firm": "Biomerieux Inc",
      "address_1": "595 Anglum Rd",
      "city": "Hazelwood",
      "state": "MO",
      "postal_code": "63042-2320",
      "additional_info_contact": "Clinical Customer Support \n800-682-2666",
      "reason_for_recall": "As the temperature and time out of range was exceeded, product performances cannot be guaranteed",
      "root_cause_description": "Storage",
      "action": "Biomerieux sent a FIELD SAFETY CORRECTIVE ACTION notice to the US subsidiary on 11/09/2022 by email.  The notice requested the subsidiary identify their customers and issue an URGENT FIELD SAFETY NOTICE which explains the problem and request the following actions:  \n\"1. Confirm this letter has been distributed to and reviewed by all appropriate personnel within your organization.\n2. Check your inventory for any of the impacted products (reference Table 1) associated with local shipments.\n3. Stop using and scrap any remaining impacted products in your inventory.\na. Request product replacement or credit for the destroyed product.\nb. Document the quantity destroyed in the Product Information section of the attached Acknowledgement\nForm.\n4. Please store this letter with your bioMrieux documentation.\n5. Complete the attached Acknowledgement Form and return it to your local bioMrieux representative as soon\nas possible.\"\n\nThe customer notices were sent by FedEx Overnight Letter on 18 NOV 2022.",
      "product_quantity": "46 units",
      "distribution_pattern": "US Nationwide distribution.",
      "openfda": {
        "k_number": [
          "K142677",
          "K130831",
          "K162950",
          "DEN130013",
          "K163536"
        ],
        "device_name": "System, Mass Spectrometry, Maldi Tof, Microorganism Identification, Cultured Isolates",
        "medical_specialty_description": "Microbiology",
        "regulation_number": "866.3361",
        "device_class": "2",
        "registration_number": [
          "3009171251",
          "3009637528",
          "3010191749",
          "3010099434",
          "3013718871",
          "9615754",
          "3018426298",
          "3003961381"
        ],
        "fei_number": [
          "3009637528",
          "3008237827",
          "3006759904",
          "3010191749",
          "3010099434",
          "3013735140",
          "3013718871",
          "3003961381"
        ]
      }
    }
  ]
}