  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-08-20",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "196806",
      "product_res_number": "Z-0505-2023",
      "event_date_initiated": "2022-11-02",
      "event_date_posted": "2022-12-14",
      "recall_status": "Open, Classified",
      "res_event_number": "91124",
      "product_code": "PKE",
      "product_description": "HENRY SCHEIN, SURGICAL PAIN TRAY, Item No.570-1482",
      "code_info": "UDI/DI (case) M75257014821, UDI/DI (kit)M75257014820, Serial/Lot Numbers:  22235489736",
      "firm_fei_number": "1000111025",
      "recalling_firm": "Stradis Medical, LLC dba Stradis Healthcare",
      "address_1": "3025 Northwoods Pkwy",
      "city": "Peachtree Corners",
      "state": "GA",
      "postal_code": "30071-1524",
      "additional_info_contact": "800-886-7257",
      "reason_for_recall": "Medical and surgical procedure trays and kits underwent certain manufacturing conditions that may result in the outer bag being incompletely sealed, which may result in a breach in the sterility of the kit.",
      "root_cause_description": "Packaging process control",
      "action": "Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email.  The notice explained the problem, risk, and requested the following action be taken:  \n   Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute.   \nStradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com.   \nIn the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly.  \n  Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter.   \nStradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.",
      "product_quantity": "20 units",
      "distribution_pattern": "US nationwide distribution, and Canada.",
      "openfda": {
        "registration_number": [
          "3009888344",
          "1061124",
          "3004111573",
          "9616088",
          "1423537",
          "3007329020",
          "1643817"
        ],
        "fei_number": [
          "3009888344",
          "3002902735",
          "3004111573",
          "3001236905",
          "1000512168",
          "3007329020",
          "1643817"
        ],
        "device_name": "Laparoscopic Cholecystectomy Kit",
        "medical_specialty_description": "General, Plastic Surgery",
        "regulation_number": "878.4800",
        "device_class": "1"
      }
    }
  ]
}