  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-08-13",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "196056",
      "product_res_number": "Z-0075-2023",
      "event_date_initiated": "2022-08-29",
      "event_date_posted": "2022-10-14",
      "recall_status": "Open, Classified",
      "res_event_number": "90816",
      "product_code": "CAH",
      "k_numbers": [
        "K965016"
      ],
      "product_description": "Iso-Gard Filter S, with Expandi-Flex, Tethered Cap, REF: a) 19261T, OUS only; b) 19262T, OUS only; c) 19272T, OUS only;  Microbial medical gas filter, single-use,",
      "code_info": "a) 19261T, UDI:    (01)04026704400492(17)250528(10)20FT51, (01)04026704400492(17)251028(10)KMH20L0409, (01)04026704400492(17)251228(10)KMH21A0086, (01)04026704400492(17)251228(10)KMH21A0190, (01)04026704400492(17)260128(10)KMH21B0163, (01)04026704400492(17)260128(10)KMH21B0343, (01)04026704400492(17)260128(10)KMH21B0355, (01)04026704400492(17)260228(10)KMH21C0177, (01)04026704400492(17)260228(10)KMH21C0387, (01)04026704400492(17)260528(10)KMH21F0156, (01)04026704400492(17)260528(10)KMH21F0337, (01)04026704400492(17)260628(10)KMH21G0426, (01)04026704400492(17)260828(10)KMH21J0020, (01)04026704400492(17)260828(10)KMH21J0245, (01)04026704400492(17)260928(10)KMH21K0142, (01)04026704400492(17)261128(10)KMH21M0182, (01)04026704400492(17)270128(10)KMH22B0151, (01)04026704400492(17)270128(10)KMH22B0248;  \n\nb) 19262T, UDI: (01)04026704400508(17)250528(10)202025, (01)04026704400508(17)250528(10)202026, (01)04026704400508(17)250628(10)202030, (01)04026704400508(17)250828(10)KMZ20J0163, (01)04026704400508(17)250828(10)KMZ20J0171, (01)04026704400508(17)250828(10)KMZ20J0582, (01)04026704400508(17)250828(10)KMZ20J0586, (01)04026704400508(17)250828(10)KMZ20J0587, (01)04026704400508(17)250928(10)KMZ20K0681, (01)04026704400508(17)251028(10)KMZ20L0968, (01)04026704400508(17)251028(10)KMZ20L0977, (01)04026704400508(17)251028(10)KMZ20L1275, (01)04026704400508(17)251128(10)KMZ20M0544, (01)04026704400508(17)251128(10)KMZ20M0661, (01)04026704400508(17)251228(10)KMZ21A0175, (01)04026704400508(17)251228(10)KMZ21A0186, (01)04026704400508(17)251228(10)KMZ21A0187, (01)04026704400508(17)251228(10)KMZ21A0208, (01)04026704400508(17)260128(10)KMZ21B0483, (01)04026704400508(17)260128(10)KMZ21B0485, (01)04026704400508(17)260128(10)KMZ21B0495, (01)04026704400508(17)260128(10)KMZ21B0811, (01)04026704400508(17)260128(10)KMZ21B0918, (01)04026704400508(17)260128(10)KMZ21B1172, (01)04026704400508(17)260228(10)KMZ21C0884, (01)04026704400508(17)260228(10)KMZ21C0910, (01)04026704400508(17)260228(10)KMZ21C0929, (01)04026704400508(17)260228(10)KMZ21C0931, (01)04026704400508(17)260328(10)KMZ21D0419, (01)04026704400508(17)260528(10)KMZ21F0223, (01)04026704400508(17)260528(10)KMZ21F0260, (01)04026704400508(17)260528(10)KMZ21F0528, (01)04026704400508(17)260628(10)KMZ21G0728, (01)04026704400508(17)260628(10)KMZ21G0735, (01)04026704400508(17)260628(10)KMZ21G0739, (01)04026704400508(17)260728(10)KMZ21H0055, (01)04026704400508(17)260828(10)KMZ21J0006, (01)04026704400508(17)260828(10)KMZ21J0007, (01)04026704400508(17)260828(10)KMZ21J0008, (01)04026704400508(17)260928(10)KMZ21K0510, (01)04026704400508(17)260928(10)KMZ21K0524, (01)04026704400508(17)260928(10)KMZ21K0539, (01)04026704400508(17)260928(10)KMZ21K0572, (01)04026704400508(17)260928(10)KMZ21K0578, (01)04026704400508(17)260928(10)KMZ21K0592, (01)04026704400508(17)260928(10)KMZ21K0952, (01)04026704400508(17)261028(10)KMZ21L0218, (01)04026704400508(17)261128(10)KMZ21M0674, (01)04026704400508(17)270128(10)KMZ22B0125, (01)04026704400508(17)270128(10)KMZ22B0345, (01)04026704400508(17)270128(10)KMZ22B0809, (01)04026704400508(17)270128(10)KMZ22B0828, (01)04026704400508(17)270128(10)KMZ22B0838, (01)04026704400508(17)270128(10)KMZ22B0869, (01)04026704400508(17)270228(10)KMZ22C0659;  \n\nc) 19272T, UDI:  (01)04026704400515(17)250528(10)202025, (01)04026704400515(17)250528(10)202026, (01)04026704400515(17)250628(10)202027, (01)04026704400515(17)250628(10)202028, (01)04026704400515(17)250628(10)202029, (01)04026704400515(17)250628(10)202030, (01)04026704400515(17)250628(10)202031, (01)04026704400515(17)250628(10)KMZ20G0005, (01)04026704400515(17)250628(10)KMZ20G0026, (01)04026704400515(17)250628(10)KMZ20G0105, (01)04026704400515(17)250628(10)KMZ20G0112, (01)04026704400515(17)250728(10)KMZ20H0081, (01)04026704400515(17)250728(10)KMZ20H0096, (01)04026704400515(17)250728(10)KMZ20H0158, (01)04026704400515(17)250728(10)KMZ20H0379, (01)04026704400515(17)250728(10)KMZ20H0530, (01)04026704400515(17)250728(10)KMZ20H0540, (01)04026704400515(17)250728(10)KMZ20H0564, (01)04026704400515(17)250728(10)KMZ20H0812, (01)04026704400515(17)250728(10)KMZ20H0841, (01)04026704400515(17)250728(10)KMZ20H1021, (01)04026704400515(17)250728(10)KMZ20H1024, (01)04026704400515(17)250728(10)KMZ20H1031, (01)04026704400515(17)250828(10)KMZ20J0092, (01)04026704400515(17)250828(10)KMZ20J0102, (01)04026704400515(17)250828(10)KMZ20J0227, (01)04026704400515(17)250828(10)KMZ20J0441, (01)04026704400515(17)250828(10)KMZ20J0535, (01)04026704400515(17)250828(10)KMZ20J0551, (01)04026704400515(17)250828(10)KMZ20J0566, (01)04026704400515(17)250828(10)KMZ20J0640, (01)04026704400515(17)250828(10)KMZ20J0654, (01)04026704400515(17)250828(10)KMZ20J0663, (01)04026704400515(17)250828(10)KMZ20J0893, (01)04026704400515(17)250828(10)KMZ20J0958, (01)04026704400515(17)250928(10)KMZ20K0034, (01)04026704400515(17)250928(10)KMZ20K0196, (01)04026704400515(17)250928(10)KMZ20K0236, (01)04026704400515(17)250928(10)KMZ20K0430, (01)04026704400515(17)250928(10)KMZ20K0659, (01)04026704400515(17)250928(10)KMZ20K0662, (01)04026704400515(17)250928(10)KMZ20K0838, (01)04026704400515(17)250928(10)KMZ20K0946, (01)04026704400515(17)250928(10)KMZ20K1260, (01)04026704400515(17)250928(10)KMZ20K1289, (01)04026704400515(17)250928(10)KMZ20K1294, (01)04026704400515(17)250928(10)KMZ20K1310, (01)04026704400515(17)250928(10)KMZ20K1339, (01)04026704400515(17)250928(10)KMZ20K1340, (01)04026704400515(17)251028(10)KMZ20L0003, (01)04026704400515(17)251028(10)KMZ20L0030, (01)04026704400515(17)251028(10)KMZ20L0031, (01)04026704400515(17)251028(10)KMZ20L0117, (01)04026704400515(17)251028(10)KMZ20L0214, (01)04026704400515(17)251028(10)KMZ20L0351, (01)04026704400515(17)251028(10)KMZ20L0359, (01)04026704400515(17)251028(10)KMZ20L0499, (01)04026704400515(17)251028(10)KMZ20L0537, (01)04026704400515(17)251028(10)KMZ20L0706, (01)04026704400515(17)251028(10)KMZ20L0724, (01)04026704400515(17)251028(10)KMZ20L0885, (01)04026704400515(17)251028(10)KMZ20L0886, (01)04026704400515(17)251028(10)KMZ20L0933, (01)04026704400515(17)251028(10)KMZ20L0945, (01)04026704400515(17)251028(10)KMZ20L1056, (01)04026704400515(17)251028(10)KMZ20L1091, (01)04026704400515(17)251028(10)KMZ20L1103, (01)04026704400515(17)251028(10)KMZ20L1104, (01)04026704400515(17)251128(10)KMZ20M0387, (01)04026704400515(17)251128(10)KMZ20M0777, (01)04026704400515(17)251128(10)KMZ20M0778, (01)04026704400515(17)251128(10)KMZ20M0787, (01)04026704400515(17)251128(10)KMZ20M0822, (01)04026704400515(17)251128(10)KMZ20M0842, (01)04026704400515(17)251128(10)KMZ20M0938, (01)04026704400515(17)251128(10)KMZ20M0967, (01)04026704400515(17)251128(10)KMZ20M0990, (01)04026704400515(17)251128(10)KMZ20M1249, (01)04026704400515(17)251128(10)KMZ20M1250, (01)04026704400515(17)251128(10)KMZ20M1257, (01)04026704400515(17)251128(10)KMZ20M1267, (01)04026704400515(17)251228(10)KMZ21A0012, (01)04026704400515(17)251228(10)KMZ21A0024, (01)04026704400515(17)251228(10)KMZ21A0117, (01)04026704400515(17)251228(10)KMZ21A0126, (01)04026704400515(17)251228(10)KMZ21A0155, (01)04026704400515(17)251228(10)KMZ21A0162, (01)04026704400515(17)251228(10)KMZ21A0164, (01)04026704400515(17)251228(10)KMZ21A0203, (01)04026704400515(17)251228(10)KMZ21A0218, (01)04026704400515(17)251228(10)KMZ21A0226, (01)04026704400515(17)251228(10)KMZ21A0368, (01)04026704400515(17)251228(10)KMZ21A0618, (01)04026704400515(17)251228(10)KMZ21A0635, (01)04026704400515(17)251228(10)KMZ21A0693, (01)04026704400515(17)251228(10)KMZ21A0706, (01)04026704400515(17)251228(10)KMZ21A0714, (01)04026704400515(17)251228(10)KMZ21A0725, (01)04026704400515(17)251228(10)KMZ21A0734, (01)04026704400515(17)251228(10)KMZ21A0950, (01)04026704400515(17)260128(10)KMZ21B0101, (01)04026704400515(17)260128(10)KMZ21B0163, (01)04026704400515(17)260128(10)KMZ21B0181, (01)04026704400515(17)260128(10)KMZ21B0183, (01)04026704400515(17)260128(10)KMZ21B0185, (01)04026704400515(17)260128(10)KMZ21B0189, (01)04026704400515(17)260128(10)KMZ21B0227, (01)04026704400515(17)260128(10)KMZ21B0234, (01)04026704400515(17)260128(10)KMZ21B0305, (01)04026704400515(17)260128(10)KMZ21B0320, (01)04026704400515(17)260128(10)KMZ21B0417, (01)04026704400515(17)260128(10)KMZ21B0421, (01)04026704400515(17)260128(10)KMZ21B0435, (01)04026704400515(17)260128(10)KMZ21B0465, (01)04026704400515(17)260128(10)KMZ21B0471, (01)04026704400515(17)260128(10)KMZ21B0479, (01)04026704400515(17)260128(10)KMZ21B0533, (01)04026704400515(17)260128(10)KMZ21B0689, (01)04026704400515(17)260128(10)KMZ21B1171, (01)04026704400515(17)260228(10)KMZ21C0270, (01)04026704400515(17)260228(10)KMZ21C0388, (01)04026704400515(17)260228(10)KMZ21C0398, (01)04026704400515(17)260228(10)KMZ21C0883, (01)04026704400515(17)260228(10)KMZ21C1159, (01)04026704400515(17)260228(10)KMZ21C1160, (01)04026704400515(17)260228(10)KMZ21C1358, (01)04026704400515(17)260328(10)KMZ21D0420, (01)04026704400515(17)260528(10)KMZ21F0227, (01)04026704400515(17)260528(10)KMZ21F0265, (01)04026704400515(17)260528(10)KMZ21F0296, (01)04026704400515(17)260528(10)KMZ21F0335, (01)04026704400515(17)260528(10)KMZ21F0375, (01)04026704400515(17)260528(10)KMZ21F0430, (01)04026704400515(17)260528(10)KMZ21F0467, (01)04026704400515(17)260528(10)KMZ21F0468, (01)04026704400515(17)260528(10)KMZ21F0478, (01)04026704400515(17)260528(10)KMZ21F0501, (01)04026704400515(17)260528(10)KMZ21F0517, (01)04026704400515(17)260528(10)KMZ21F0534, (01)04026704400515(17)260528(10)KMZ21F0550, (01)04026704400515(17)260528(10)KMZ21F0551, (01)04026704400515(17)260528(10)KMZ21F0552, (01)04026704400515(17)260528(10)KMZ21F0557, (01)04026704400515(17)260528(10)KMZ21F0567, (01)04026704400515(17)260528(10)KMZ21F0568, (01)04026704400515(17)260528(10)KMZ21F0588, (01)04026704400515(17)260528(10)KMZ21F0594, (01)04026704400515(17)260528(10)KMZ21F0742, (01)04026704400515(17)260528(10)KMZ21F0777, (01)04026704400515(17)260628(10)KMZ21G0399, (01)04026704400515(17)260628(10)KMZ21G0403, (01)04026704400515(17)260628(10)KMZ21G0404, (01)04026704400515(17)260628(10)KMZ21G0751, (01)04026704400515(17)260628(10)KMZ21G0765, (01)04026704400515(17)260628(10)KMZ21G0770, (01)04026704400515(17)260628(10)KMZ21G0778, (01)04026704400515(17)260628(10)KMZ21G0786, (01)04026704400515(17)260628(10)KMZ21G0800, (01)04026704400515(17)260628(10)KMZ21G0806, (01)04026704400515(17)260628(10)KMZ21G0838, (01)04026704400515(17)260628(10)KMZ21G0911, (01)04026704400515(17)260628(10)KMZ21G0912, (01)04026704400515(17)260628(10)KMZ21G0917, (01)04026704400515(17)260628(10)KMZ21G0954, (01)04026704400515(17)260628(10)KMZ21G0955, (01)04026704400515(17)260628(10)KMZ21G0956, (01)04026704400515(17)260628(10)KMZ21G0964, (01)04026704400515(17)260628(10)KMZ21G0971, (01)04026704400515(17)260628(10)KMZ21G0982, (01)04026704400515(17)260628(10)KMZ21G0986, (01)04026704400515(17)260628(10)KMZ21G0992, (01)04026704400515(17)260628(10)KMZ21G0993, (01)04026704400515(17)260628(10)KMZ21G0995, (01)04026704400515(17)260628(10)KMZ21G1001, (01)04026704400515(17)260628(10)KMZ21G1008, (01)04026704400515(17)260628(10)KMZ21G1017, (01)04026704400515(17)260628(10)KMZ21G1037, (01)04026704400515(17)260628(10)KMZ21G1038, (01)04026704400515(17)260728(10)KMZ21H0115, (01)04026704400515(17)260728(10)KMZ21H0257, (01)04026704400515(17)260728(10)KMZ21H0299, (01)04026704400515(17)260728(10)KMZ21H0405, (01)04026704400515(17)260728(10)KMZ21H0429, (01)04026704400515(17)260728(10)KMZ21H0431, (01)04026704400515(17)260728(10)KMZ21H0433, (01)04026704400515(17)260728(10)KMZ21H0435, (01)04026704400515(17)260728(10)KMZ21H0504, (01)04026704400515(17)260728(10)KMZ21H0508, (01)04026704400515(17)260728(10)KMZ21H0526, (01)04026704400515(17)260728(10)KMZ21H0560, (01)04026704400515(17)260728(10)KMZ21H0561, (01)04026704400515(17)260828(10)KMZ21J0050, (01)04026704400515(17)260828(10)KMZ21J0077, (01)04026704400515(17)260828(10)KMZ21J0088, (01)04026704400515(17)260828(10)KMZ21J0089, (01)04026704400515(17)260828(10)KMZ21J0446, (01)04026704400515(17)260828(10)KMZ21J0447, (01)04026704400515(17)260828(10)KMZ21J0448, (01)04026704400515(17)260828(10)KMZ21J0484, (01)04026704400515(17)260828(10)KMZ21J0485, (01)04026704400515(17)260828(10)KMZ21J0534, (01)04026704400515(17)260828(10)KMZ21J0546, (01)04026704400515(17)260828(10)KMZ21J0554, (01)04026704400515(17)260828(10)KMZ21J0555, (01)04026704400515(17)260828(10)KMZ21J0570, (01)04026704400515(17)260828(10)KMZ21J0571, (01)04026704400515(17)260828(10)KMZ21J0579, (01)04026704400515(17)260828(10)KMZ21J0580, (01)04026704400515(17)260928(10)KMZ21K0045, (01)04026704400515(17)260928(10)KMZ21K0114, (01)04026704400515(17)260928(10)KMZ21K0163, (01)04026704400515(17)260928(10)KMZ21K0164, (01)04026704400515(17)260928(10)KMZ21K0266, (01)04026704400515(17)260928(10)KMZ21K0274, (01)04026704400515(17)260928(10)KMZ21K0295, (01)04026704400515(17)260928(10)KMZ21K0299, (01)04026704400515(17)260928(10)KMZ21K0308, (01)04026704400515(17)260928(10)KMZ21K0314, (01)04026704400515(17)260928(10)KMZ21K0318, (01)04026704400515(17)260928(10)KMZ21K0323, (01)04026704400515(17)260928(10)KMZ21K0452, (01)04026704400515(17)260928(10)KMZ21K0453, (01)04026704400515(17)260928(10)KMZ21K0509, (01)04026704400515(17)260928(10)KMZ21K0926, (01)04026704400515(17)260928(10)KMZ21K0953, (01)04026704400515(17)260928(10)KMZ21K0954, (01)04026704400515(17)261028(10)KMZ21L0047, (01)04026704400515(17)261028(10)KMZ21L0070, (01)04026704400515(17)261028(10)KMZ21L0074, (01)04026704400515(17)261028(10)KMZ21L0077, (01)04026704400515(17)261028(10)KMZ21L0079, (01)04026704400515(17)261028(10)KMZ21L0219, (01)04026704400515(17)261028(10)KMZ21L0220, (01)04026704400515(17)261028(10)KMZ21L0238, (01)04026704400515(17)261028(10)KMZ21L0239, (01)04026704400515(17)261028(10)KMZ21L0244, (01)04026704400515(17)261028(10)KMZ21L0295, (01)04026704400515(17)261028(10)KMZ21L0296, (01)04026704400515(17)261028(10)KMZ21L0307, (01)04026704400515(17)261028(10)KMZ21L0308, (01)04026704400515(17)261028(10)KMZ21L0316, (01)04026704400515(17)261028(10)KMZ21L0333, (01)04026704400515(17)261028(10)KMZ21L0334, (01)04026704400515(17)261028(10)KMZ21L0337, (01)04026704400515(17)261028(10)KMZ21L0344, (01)04026704400515(17)261028(10)KMZ21L0367, (01)04026704400515(17)261028(10)KMZ21L0368, (01)04026704400515(17)261028(10)KMZ21L0382, (01)04026704400515(17)261028(10)KMZ21L0389, (01)04026704400515(17)261028(10)KMZ21L0541, (01)04026704400515(17)261028(10)KMZ21L0798, (01)04026704400515(17)261028(10)KMZ21L0878, (01)04026704400515(17)261028(10)KMZ21L0880, (01)04026704400515(17)261028(10)KMZ21L0901, (01)04026704400515(17)261028(10)KMZ21L0924, (01)04026704400515(17)261128(10)KMZ21M0050, (01)04026704400515(17)261128(10)KMZ21M0690, (01)04026704400515(17)261128(10)KMZ21M0692, (01)04026704400515(17)261128(10)KMZ21M0719, (01)04026704400515(17)261128(10)KMZ21M0756, (01)04026704400515(17)261128(10)KMZ21M0781, (01)04026704400515(17)261128(10)KMZ21M0786, (01)04026704400515(17)261128(10)KMZ21M0790, (01)04026704400515(17)261128(10)KMZ21M0805, (01)04026704400515(17)261228(10)KMZ22A0635, (01)04026704400515(17)261228(10)KMZ22A0636, (01)04026704400515(17)261228(10)KMZ22A0650, (01)04026704400515(17)261228(10)KMZ22A0871, (01)04026704400515(17)270128(10)KMZ22B0130, (01)04026704400515(17)270128(10)KMZ22B0131, (01)04026704400515(17)270128(10)KMZ22B0192, (01)04026704400515(17)270128(10)KMZ22B0509, (01)04026704400515(17)270128(10)KMZ22B0524, (01)04026704400515(17)270128(10)KMZ22B0525, (01)04026704400515(17)270128(10)KMZ22B0556, (01)04026704400515(17)270128(10)KMZ22B0557, (01)04026704400515(17)270128(10)KMZ22B0784, (01)04026704400515(17)270128(10)KMZ22B0785, (01)04026704400515(17)270228(10)KMZ22C0395, (01)04026704400515(17)270228(10)KMZ22C0620, (01)04026704400515(17)270228(10)KMZ22C0802, (01)04026704400515(17)270228(10)KMZ22C0803.",
      "firm_fei_number": "3005747797",
      "recalling_firm": "TELEFLEX LLC",
      "address_1": "3015 Carrington Mill Blvd",
      "city": "Morrisville",
      "state": "NC",
      "postal_code": "27560-5437",
      "reason_for_recall": "Incidents of device splitting or detaching during use",
      "root_cause_description": "Under Investigation by firm",
      "action": "Teleflex, Inc. sent an URGENT MEDICAL DEVICE RECALL  notice by Fedex 2-day mail on 08/29/2022.  the notice described the problem, listed the complaints and requested the following actions be taken:\nMedical facilities: \n1. Immediately check your inventory for product within the scope of this\nrecall. Cease use and distribution of affected product and immediately\nquarantine the affected product.\n2. If you have impacted product, mark the applicable checkbox on the Acknowledgement Form and contact Teleflex Customer Service by calling the phone number provided.\n Teleflex Customer Service will issue a return goods authorization (RGA) number to\nyou. Write the RGA number in the Acknowledgement Form and\nreturn form by email to Recalls@teleflex.com.\n3. If you do not have impacted product, mark the applicable checkbox on the Acknowledgement Form and return the form to Teleflex at Recalls@teleflex.com.\n4. Teleflex (or your local dealer) will issue a credit note upon receipt of the returned affected product.   \nDistributors:  \n1. Provide this recall notice to all customers who have received impacted product. Each of your customers must complete the Acknowledgement Form and return it to you.\n2. Immediately check your inventory for impacted product. Cease use and distribution of impacted product and immediately quarantine the affected product. You may return all product in scope.\n3. As a distributor, you must confirm to Teleflex that you have completed the\nfield activity outlined in actions 1 and 2 of this Action List Number 2. Upon completion of your actions, please forward the completed Acknowledgement Form to Teleflex Customer Service. \nImportant - Please ensure you only list batch numbers in scope of this recall notice when\ncompleting this form.\n4. If you have further distributed product outside of your country, please notify Teleflex\nCustomer Service by email to Recalls@teleflex.com \n\nOn October 28, 2022, the firm sent an updated recall notice with the expansion to additional",
      "product_quantity": "355 units",
      "distribution_pattern": "US nationwide distribution",
      "openfda": {
        "k_number": [
          "K770136",
          "K801803",
          "K954828",
          "K934540",
          "K013089",
          "K850327",
          "K770392",
          "K954490",
          "K091363",
          "K133666",
          "K953687",
          "K190022",
          "K223258",
          "K925388",
          "K842580",
          "K222917",
          "K221836",
          "K883693",
          "K081536",
          "K980662",
          "K252643",
          "K092409",
          "K002506",
          "K013122",
          "K221472",
          "K901476",
          "K930760",
          "K903056",
          "K952428",
          "K923661",
          "K823257",
          "K182862",
          "K890362",
          "K781696",
          "K780535",
          "K063526",
          "K202459",
          "K951047",
          "K760611",
          "K221250",
          "K931217",
          "K000654",
          "K132709",
          "K121382",
          "K962045",
          "K892230",
          "K890346",
          "K791307",
          "K961838",
          "K911738",
          "K880958",
          "K915750",
          "K792097",
          "K163300",
          "K963053",
          "K833066",
          "K812573",
          "K760041",
          "K061426",
          "K900990",
          "K791773",
          "K012485",
          "K211286",
          "K925462",
          "K896116",
          "K892520",
          "K092451",
          "K941536",
          "K952062",
          "K192713",
          "K010040",
          "K840960",
          "K772047",
          "K812098",
          "K072002",
          "K802914",
          "K925217",
          "K091185",
          "K831856",
          "K121026",
          "K982127",
          "K920885",
          "K890035",
          "K973314",
          "K964979",
          "K770376",
          "K895233",
          "K965020",
          "K241339",
          "K090738",
          "K961914",
          "K864103",
          "K821310",
          "K911651",
          "K862786",
          "K030990",
          "K904830",
          "K964382",
          "K934475",
          "K043148",
          "K112958",
          "K014282",
          "K820577",
          "K853577",
          "K862790",
          "K023641",
          "K002201",
          "K934272",
          "K132679",
          "K862789",
          "K841581",
          "K830618",
          "K862787",
          "K911427",
          "K951410",
          "K030071",
          "K802763",
          "K021020",
          "K251583",
          "K983610",
          "K013200",
          "K833267",
          "K092218",
          "K912577",
          "K941381",
          "K952015",
          "K033008",
          "K983817",
          "K820576",
          "K941676",
          "K893030",
          "K844492",
          "K102483",
          "K110246",
          "K973797",
          "K944124",
          "K030271",
          "K974630",
          "K844491",
          "K760567",
          "K011132",
          "K832935",
          "K951484",
          "K964204",
          "K840205",
          "K894978",
          "K013093",
          "K930816",
          "K922492",
          "K032322",
          "K851070",
          "K031653",
          "K151498",
          "K934509",
          "K050953",
          "K862785",
          "K945543",
          "K833269",
          "K973238",
          "K012958",
          "K021265",
          "K934189",
          "K024270",
          "K884465",
          "K944189",
          "K032878",
          "K014281",
          "K063125",
          "K970102",
          "K210352",
          "K111408",
          "K130664",
          "K982239",
          "K101629",
          "K880681",
          "K770375",
          "K990949",
          "K934132",
          "K052087",
          "K931842",
          "K043120",
          "K812075",
          "K050927",
          "K801804",
          "K191909",
          "K943874",
          "K790896",
          "K002839",
          "K032716",
          "K896832",
          "K230483",
          "K131626",
          "K901816",
          "K811546",
          "K093426",
          "K850265",
          "K841236",
          "K760081",
          "K965016",
          "K011212",
          "K141087",
          "K871366"
        ],
        "registration_number": [
          "3003898360",
          "",
          "1000604079",
          "3012494290",
          "3012821221",
          "3005012805",
          "3004428458",
          "3018094310",
          "2024841",
          "3017202422",
          "2648727",
          "1824226",
          "8043316",
          "3006534321",
          "3015134079",
          "1526863",
          "8021995",
          "1054241",
          "2529776",
          "3013530359",
          "3004874794",
          "2518443",
          "8022265",
          "2320762",
          "1722070",
          "3013557562",
          "3006061749",
          "3004127293",
          "3005998497",
          "3031369651",
          "1928237",
          "3009171251",
          "3006254047",
          "3008567544",
          "3005587132",
          "3010381606",
          "3003697063",
          "3004050971",
          "2085602",
          "2024500",
          "3027242374",
          "1314417",
          "1024404",
          "3010834934",
          "8030673",
          "2936999",
          "3001313714",
          "3013242701",
          "3010131137",
          "2025816",
          "3010439239",
          "3007709321",
          "3020777588",
          "2028807",
          "3006657376",
          "3009887154",
          "1832562",
          "9611451",
          "3006783791",
          "3012316249",
          "3008412416",
          "3011088743",
          "9680661",
          "3001124136",
          "3015191173",
          "2132517",
          "3010082909",
          "3010402691",
          "3006419626",
          "3010889742",
          "3008806809",
          "3014936217",
          "3009896423",
          "3012365569",
          "8044169",
          "1924066",
          "3015537026",
          "9616567",
          "3001104093",
          "1043214",
          "3033959233",
          "3014421917",
          "3017987980",
          "3007289408",
          "9681848",
          "3030446844",
          "3005855942",
          "3012329712",
          "3004116032",
          "3010877292",
          "9710644",
          "9614485",
          "3006446479",
          "8040412",
          "8010042",
          "2433214",
          "2183585",
          "3045816907",
          "3021284283",
          "3008264065",
          "3017174574",
          "3004476631",
          "3006157842",
          "3013187142",
          "3006606984",
          "1220253",
          "3005519819",
          "3027977504",
          "9611500",
          "3017520736",
          "3005515211",
          "3044994756",
          "3009747810",
          "3011707784",
          "3003332988",
          "3003862188",
          "3015168843",
          "8020893",
          "3021244152",
          "3013095415",
          "3007337939",
          "3011987967",
          "1055236",
          "3006673317",
          "3011137372",
          "9617604",
          "3018116884",
          "3004122598",
          "9615102",
          "3010042047",
          "3010838917",
          "3020614",
          "3023294726",
          "3011503227",
          "3009746061",
          "3004365956",
          "9710567",
          "3045816902",
          "3000268902",
          "9680602",
          "3003399703",
          "3008693658",
          "3008311916",
          "3002807267"
        ],
        "fei_number": [
          "3003898360",
          "",
          "1000604079",
          "3012494290",
          "3012821221",
          "3005012805",
          "3004428458",
          "3018094310",
          "2024841",
          "3017202422",
          "2648727",
          "1824226",
          "3006534321",
          "3002807293",
          "3015134079",
          "1526863",
          "3013530359",
          "3004874794",
          "2518443",
          "3004213967",
          "3013557562",
          "3006061749",
          "3005747797",
          "3004127293",
          "3005998497",
          "3031369651",
          "1000121196",
          "1928237",
          "3002807850",
          "3006254047",
          "3000213164",
          "3008567544",
          "3005587132",
          "1000221012",
          "3010381606",
          "3003697063",
          "3004050971",
          "2024500",
          "3027242374",
          "1314417",
          "1024404",
          "3010834934",
          "1000374126",
          "2936999",
          "3006759904",
          "3003040625",
          "3013242701",
          "3010131137",
          "1000271197",
          "3003319165",
          "3010439239",
          "2025816",
          "3007709321",
          "3020777588",
          "3006657376",
          "1000112877",
          "3009887154",
          "1832562",
          "3000210616",
          "3012316249",
          "3006783791",
          "3006363476",
          "3008412416",
          "3011088743",
          "3001124136",
          "3015191173",
          "1000283452",
          "3005596530",
          "3010082909",
          "3010402691",
          "1000117260",
          "3010889742",
          "3008806809",
          "3006419626",
          "3014936217",
          "3009896423",
          "3012365569",
          "1924066",
          "3015537026",
          "3001104093",
          "1043214",
          "3033959233",
          "3014421917",
          "3017987980",
          "3007289408",
          "3003256355",
          "3002861325",
          "3030446844",
          "1000268645",
          "3005855942",
          "3012329712",
          "3004116032",
          "3010877292",
          "3023530182",
          "3006446479",
          "2433214",
          "2183585",
          "3045816907",
          "3021284283",
          "3008264065",
          "3017174574",
          "3004476631",
          "3006157842",
          "3002806988",
          "3013187142",
          "3006606984",
          "1220253",
          "3005519819",
          "3027977504",
          "3002808696",
          "3017520736",
          "3002808128",
          "3005515211",
          "3044994756",
          "3009747810",
          "3011707784",
          "3003332988",
          "3003862188",
          "3015168843",
          "3021244152",
          "3013095415",
          "3007337939",
          "3011987967",
          "3009271244",
          "1000112968",
          "3018116884",
          "3002673652",
          "3004122598",
          "3010042047",
          "3003949221",
          "3010838917",
          "3020614",
          "3023294726",
          "3011503227",
          "3002963732",
          "3009746061",
          "3004365956",
          "3045816902",
          "3000268902",
          "3002558692",
          "3003399703",
          "3008693658",
          "3008311916",
          "3002807267"
        ],
        "device_name": "Filter, Bacterial, Breathing-Circuit",
        "medical_specialty_description": "Anesthesiology",
        "regulation_number": "868.5260",
        "device_class": "2"
      }
    }
  ]
}