      "cfres_id": "192323",
      "product_res_number": "Z-0900-2022",
      "event_date_initiated": "2022-02-17",
      "event_date_created": "2022-04-09",
      "recall_status": "Terminated",
      "event_date_terminated": "2024-05-06",
      "res_event_number": "89735",
      "product_code": "MMY",
      "pma_numbers": [
        "P910018"
      ],
      "product_description": "LIPOSORBER System, Model: LA-15, consisting of: \n\nApheresis machine, product# 101445 model: MA-03, and \n\ndisposables (product#, model) \n[#101447, KP-05 Sulflux Plasma Separator], \n[#101473, HDE KP-05 Sulflux Plasma Separator], \n[#101785, LA-15 (AU) LDL Adsorption Column (Luer Lock Type)], \n[#101786, NK-M3R(UL) Tubing (Luer Lock Type)], \n[#101788, HDE LA-15(AU) LDL Adsorption Column (Luer Lock Type)], \n[#101789, HDE NK-M3R(UL) Tubing (Luer Lock Type)]. \n\nAssociated labeling:      \n1) Instructions for Use (6): IFU LIPOSORBER LA-15 LDL Adsorption Columns FH, IFU SULFLUX KP-05 PLASMA SE-PARATOR FH, IFU TUBING SYSTEM FOR PLASMAPHERESIS NK-M3R(UL) FH, IFU LIPOSORBER LA-15 LDL Adsorption Columns FSGS, IFU SULFLUX KP-05 PLASMA SE-PARATOR FSGS, IFU TUBING SYSTEM FOR PLASMAPHERESIS NK-M3R(UL) FSGS,\n2) Operators Manuals for the following (2): LIPOSORBER LA-15 SYSTEM OPERATOR's MANUAL FH, LIPOSORBER LA-15 SYSTEM OPERATOR's MANUAL FSGS     3) Patients Guide (1): A PATIENT GUIDE TO THE LIPOSORBER LA-15 SYSTEM FSGS",
      "code_info": "All product lot and UDI numbers are affected.",
      "firm_fei_number": "2435151",
      "recalling_firm": "Kaneka America Corp",
      "address_1": "546 5th Ave Fl 21 546 5th Ave",
      "address_2": "Fl 21",
      "city": "New York",
      "state": "NY",
      "postal_code": "10036-5000",
      "additional_info_contact": "Ahmad Al-Sattari\n212-7054355",
      "reason_for_recall": "Labeling change to due to the risk of severe anaphylactoid reactions including shock in patients treated with the Liposorber LA-15 LDL Adsorption Column under concomitant ACE inhibitor medication treatment.",
      "root_cause_description": "Error in labeling",
      "action": "On 02/17/2022 a recall notification letter with revised labeling was emailed regarding contraindication in patients who are being treated with angiotensin-converting enzyme (ACE) inhibitors. Customers are to locate, destroy and replace all operator's manuals, patient guides, and IFUs with those attached to the recall notice, ensuring they are distributed to all end users.  All device model, lot, and UDI numbers for systems and disposables are affected. Submit the Acknowledgement and Receipt Form to Kaneka Medical America, LLC via email. If you have any question or concerns about this recall please contact Kaneka Medical America LLC Monday through Friday, 8:00 AM to 4:30 PM, Eastern Time, at: Phone: (212) 705-4355, Email: Ahmad.Al-Sattari@kaneka.com",
      "product_quantity": "138 Systems, 54703 disposables",
      "distribution_pattern": "US only: CA, NH, CT, MA, FL, DE, NY, NJ, PA, OH, MD, VA, NC, SC, GA, MS, TN, TX, IN, AZ, MI, WI, MN, IL, MO, NE, CO, WA, OR, UT, KS, OK, AL, ME",
      "openfda": {
        "device_name": "Lipoprotein, Low Density, Removal",
        "medical_specialty_description": "Unknown",
        "regulation_number": "",
        "device_class": "3",
        "pma_number": [
          "P910018",
          "P940016"
        ],
        "registration_number": [
          "9614654",
          "3014680791",
          "8031561",
          "3008229990",
          "8010002",
          "3031408734"
        ],
        "fei_number": [
          "3014680791",
          "3002808904",
          "3002807721",
          "3002806407",
          "3008229990",
          "3031408734"
        ]
      }
    }
  ]
}